This high-resolution color display with touch screen use is
specifically designed to meet the rigorous performance standards needed
for diagnostic, interventional radiology, and other medical applications.
To guarantee image integrity, features include accurate signal
conversion and a wide range of interfacing options.
Compact design -Low weight and small size with improved performance make
the color flat panel display HL1916S SERIAL preferable to conventional
CRT monitors.
Embedded LUT(Look Up Table )-This monitor is factory calibrated to
achieve Gamma2.2 compliance and Linear gray level reproduction at the
factory set point. Six different settings are stored within the display.
Screen resolution- HL1916S SERIAL is equipped with a panel with Twisted
Nematic technology. The optimal picture resolution is 1280 x 1024 pixels.
Video signals with other resolutions typical to medical engineering are
optimally zoomed in or out to the screen size.
Fast backlight stability- The luminance stabilization circuit employs
a built in photo sensor to keep the back-light lamps at a constant
luminance for consistent calibration over the life of the display and
can control the back light system automatically to extend the life of
the monitor
Multi-interface for video inputs- Support DVI-D, VGA, DisplayPort.
Embedded full functional stand base - The stand base is embedded inside
the monitor and can be easily removed. The stand base is up/down and tilt
adjustable.
2. Declarations
Safety precautions
Medical Equipment
With respect to electric shock, Fire and mechanical
hazards only in accordance with ANSI/AAMI
ES60601-1:2005&CSA C22.2 No.60601-1:2008
WARNING:
To avoid risk of electric shock, this equipment must only be connected
to a supply mains with protective earth.
Page 1 of 26
Appliance coupler
or separable plug of is used as isolation means to
isolate the equipment from mains supply.
Accessory equipment connected to the analog and digital interfaces
must be certified according to the respective IEC/EN standards
(e.g.IEC/EN950 for data processing equipment and IEC/EN 60601-1 for
medical equipment).
Furthermore all configurations shall comply with the valid version of
the System standard IEC/EN 60601-1-1.Everybody who connects
additional equipment to the signal input connector or signal output
connector
Configures a medical system, and therefore responsible that the system
Complies with the requirements of the valid version of the system
standard IEC/EN 60601-1-1. If in doubt, consult our technical service
department or your local distributor.
Regular maintenance and calibration are recommended
Please note that liquid crystal displays such as the HL1916S SERIAL
do not have a failure rate of zero and image parameters may change over
time (e.g. luminance or discoloration). Please ensure that all
measures are taken to prevent injuries or incorrect diagnoses. Regular
maintenance and calibration are recommended.
Correct and safe operation of the flat panel displays is dependent on
proper transport, storage, installation and assembly, as well as
careful operation and maintenance. The units must only be used for
applications for which monitors are normally used. The information in
the Section "Technical data" must be observed exactly.
For the sake of safety, the following precautions must be observed:
Only use a perfect power supply cable
A damaged power supply cable may result in a fire or electric shock.
When disconnecting the power supply cable, always do so by holding the
plug.
Only use the same type of fuse
T2A/250V
Do not insert any objects into the housing
Objects inserted into the housing may result in damage to the unit or
personal injury.
Page 2 of 26
Do not place any objects on top of the units
Penetrating liquids may result in a fire or electric shock.
Connection
No contact to a patient must occur when handling the cables.
Do not hurt yourself, when moving the display
The display can be tilted backwards and forwards. Please, pay attention
not to hurt yourself, when moving the display. Fingers or small objects
may get stuck at the bottom of the display.
When moving the display up and down (height adjustment), make sure you
do not squeeze your hand or any other object. The minimum distance
between the display edge and the bottom is only 59 mm.
Caution
Incorrect installation may result in extensive damage to property.
Installation should be carried out by trained personnel
When installing your medical electrical system with our products in
an environment with patients, please observe the safety requirements
of EN 60601-1 (IEC 60601-1) for "Specifications for the safety of
medical electrical systems" in order to prevent injury to patients and
users of your systems.
Take appropriate measures to particularly ensure that discharge
currents remain below the required limits: Appropriate measures:
- Disconnecting devices for signal input or output unit
- Use of a safety transformer
- Use of additional PE conductor
Only use the signal cables and interface cables specified by the
manufacturer for the installation.
Use power cables with a PE contact. Only insert into sockets with a
PE contact.
For certain applications, the video earth can be separately connected
to the PE via the additional PE connection in the plug panel (observe
IEC 60601-1).
Close the plug panel using the provided cover , and secure using the
Page 3 of 26
screws.
Turn switch off and then remove power cord.
Mounting information: The stability of the display must be guaranteed
following mounting of the foot/holder. The immersion depth of the
mounting screws has to be 10 to 12 mm including a 3 mm VESA mounting
plate. (See also table "Mounting screws" on the following). All these
requirements are satisfied when using the original foot. All
requirements must be observed when using customer-specific mounting
solutions.
Notice for users: The plug panel closed by the cover, must not be opened
by users.
Servicing information: If housing components have to be removed for
servicing, this must not be carried out in the presence of patients,
the user, or other persons not involved with servicing.
The following applies to installations in the USA and Canada: Molded
power supply plugs must comply with the requirements for "Hospital
Grade Attachments" UL 498.
Caution
Failure to observe the warnings may result in substantial damage to
property.
Provide sufficient heat dissipation
Holes are provided at the rear of the housing. The display must be
placed or secured on a hard, level surface at least 10cm from the wall
and 15cm away from other devices. Several displays can be butt-mounted
horizontally and vertically.
The following must be observed when mounting (VESA connection):
Mounting screws
Number 4
Thread M4
Strength 12
Immersion depth Min. 10 mm; Max. 12 mm
Torque Max. 3 Nm
Please see 4.1.4 for details.
Page 4 of 26
The permissible ambient temperature range (5 °C ... 35 °C) must n ot
be violated. Do not subject device to unnecessary shocks. Take care
when transporting! Use the original packaging! The panel in particular
should be protected against shocks.
When touching the panel surface, the mechanical contact or an
electrical discharge may cause a brief disturbance in the picture
quality.
Care of unit / cleaning agents
− The front panel is extremely sensitive to mechanical damage. Avoid
all scratches, knocks etc.!
− Remove water drops immediately; extended contact with water
discolors the surface.
Clean the front panel when dirty, using a micro fiber cloth and, if
necessary, a glass cleaning agent. Only clean housing parts using a
cleaning agent for plastics.
Note:
Do not use cleaning agents containing solvent, e.g. petroleum spirit!
Explanation of the symbols
Alternation current (AC)
Protective earth
China Rohs symbol
EC WEEE symbol
European comformity
China Compulsory Certification
T TUV approval mark
Attention: Consult the accompanying documents
Page 5 of 26
3. Installation
Provide adequate ventilation
Ventilation slots are located on the rear of the housing.
Ambient temperature
The permissible ambient temperature range must not be violated.
Minimize reflections
The display should be positioned so that reflections of lights, windows,
furniture with shiny surfaces or light-colored walls do not appear on
the screen.
Minimize mirroring
In order to reduce mirroring on the unit, ceiling lighting or reflected
light (no dazzling) should be used. Mirroring can only be eliminated
if the screen is clean and free of grease. Clean the display using a
suitable micro fiber cloth.
Change of environment
If the unit is brought into a warm environment from a cold one, water
may condense upon it. The unit should not be switched on until all the
condensed water has evaporated, including that inside the unit. This
may take several hours, depending on the conditions.
4. Start-up
Caution
In order to ensure safe operation of the equipment, close attention
must be paid to the information contained in this Instruction Manual
as well as the warnings in Section 2 "Safety precautions".
Caution Information for end customer
None of the settings must be changed on site by the user, otherwise
the guarantee is canceled. This also applies to settings made using
the HL1916S SERIAL keys. These are therefore locked for certain
applications. If settings have to be changed, please contact the
responsible servicing department.
The display is designed for individual connection to a graphics card
with a power supply of 100 or 240 Volt (TN-S system with PE conductor).
Page 6 of 26
If the display is to be used in a sequence of several displays, or if
it is not exactly known whether the graphics card standard can be output
by the display, refer to Section 5.1 "Connection of the flat panel
display".
In order to start the unit properly, the following steps should be
carried out in the given sequence.
4.1 Connecting the power and signal cables
Warning The display can be tilted backwards and forwards. Please, pay
attention not to hurt yourself, when moving the display. Fingers or
small objects may get stuck at the bottom of the display.
Caution
Use a power cable with PE conductor corresponding to the safety
requirements of the respective country of use. Note for North America:
Molded power supply plugs must comply with the requirements for
hospitals with respect to CSA Std. C22.2 No. 21 and UL 498.The power
supply and signal connections are located on the rear of the color flat
panel display.
Note
Note that the cables are already positioned when you receive the
display (power cable and DVI-D cable). The following steps are only
necessary if you need to connect/disconnect the cables of the scope
of supply.
4.1.1 Little cover (removing)
Remove the one screws with a M4 Slot screwdriver (one turn
suffices).Pull down the little cover and then open the little cover
and remove it.
Page 7 of 26