Copyright @ SSME (SIEMENS
Medical Equipment Ltd.) All
rights reserved
HL1916
Page 3 of 32
Federal Communications Commission (FCC) Statement
You are cautioned that changes or modifications not expressly approved by the
part responsible for compliance could void the user’s authority to operate the
equipment
FCC- Class B
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference in a communications.
However, there is no guarantee that interference will not occur in particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and on,
the user is encouraged to try to correct the interference by one or more of the
following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
1. SCOPE
This document defines the performance requirements for a 19.0-inch TFT LCD color
monitor for medical use. This product is controlled by model name; any change will be
recorded in the listand confirmed by SIEMENS MED.
This high-resolution color display is specifically designed to meet the rigorous
performance standards needed for diagnostic, interventional radiology, and other
medical applications. To guarantee image integrity, features include accurate signal
conversion and a wide range of interfacing options.
This monitor is factory calibrated to achieve DICOM part 3.14 compliance and
Gamma CIE at the factory set point. The luminance stabilization circuit employs a built
in photo sensor to keep the back-light lamps at a constant luminance for consistent
calibration over the life of the display and can control the back light system
automatically to extend the life of the monitor and achieve very short warming up time.
The surface of the monitor has an anti-glare coating to minimize reflection and a hard
coating to reduce scratch.
With the deliberate designed bracket the monitor can stick on the desktop firmly.
Last change: 27
th
, Oct. 2005
Copyright @ SSME (SIEMENS
Medical Equipment Ltd.) All
rights reserved
HL1916
Page 4 of 32
2. ELECTRICAL PERFORMANCE
2.1 Power Supply
-Input Voltage
: AC100-240 ± 10%
- Current (max) : 0.25A(AC220-240V); 0.55A(AC100-120V)
- Frequency
: 50/60Hz ± 3Hz
- Power Consumption : <54W
2.2 Power Management
Power Management condition and status for ANALOG Input mode
State
Horizontal Vertical Video Consumption
ON ON ON Active <54W Green
Active off
Power Management condition and status for DIGITAL Input mode
State
OFF ON Blanked Orange
ON OFF Blanked Orange
OFF OFF Blanked
DE HorizontalVertical Video Consumption
SIGNALS Power
SIGNALS Power
LED
Status
<2W
Orange
LED
Status
Active off NO
2.3 Signal Interface
Last change: 27
th
ON Pulses ON ON Active <54W Green
N/A N/A Blanked <2W Orange
Pulses
IN LET 3P IEC TYPE
Copyright @ SSME (SIEMENS
, Oct. 2005
Medical Equipment Ltd.) All
GND
15-PIN MIN D-TYPE VGA
FEMALE CONNECTOR
DVI-D
5V SWITCH
9-PIN D-SUB RS-232
FEMALE CONNECTOR
HL1916
Page 5 of 32
ONOFF
rights reserved
2.3.1 Signal Specifications
Item SPEC
Analog:
Frequency
Signal Input
(Analog)
D-SUB
CVS Signal
DVI-I Input Via DVI-I to VGA connector
DVI-D Input
2.3.2 D-SUB Connector and Pin Assignment (Figure1)
Pixel clock 25--165MHz
Video Input
Signal Input Separate Sync, TTL (N or P)
H 30 ~ 82kHz
V 50 ~ 85Hz
Analog 0.7Vpp
Input Impedance 75 Ohm
Video Level: 0.5---1.0V
Sync level: 0.2---0.3V
DVI-Digital
DDC via DVI
Last change: 27
Figure 1
15-pin Min D-type Female Connector
Table 1
Pin - Assignment of 15-pin D-sub:
1 Red Video 6 Red Ground 11 Monitor Ground
2 Green Video 7 Green Ground 12 DDC-Serial Data
3 Blue Video 8 Blue Ground 13 H-Sync.
4 N/C 9 NC 14 V-Sync.
5 GND 10 Logic Ground 15 DDC-Serial Clock
th
, Oct. 2005
Copyright @ SSME (SIEMENS
Medical Equipment Ltd.) All
HL1916
rights reserved
Page 6 of 32
2.3.3 Digital Visual Interface and Pin Assignment (DVI)
Figure 2
29-pin DVI-D Female Connector
Table 2
Pin - Assignment of 29-pin DVI-D/DVI-I Female Connector
Pin 1 - TMDS Data 2- Pin 12 - TMDS Data 3- Pin 22 - TMDS Clock Shield
Pin 2 - TMDS Data 2+ Pin 13 - TMDS Data 3+ Pin 23 - TMDS Clock+
Pin 3 - TMDS Data 2/4 Shield Pin 14 - +5 V Power Pin 24 - TMDS ClockPin 4 - TMDS Data 4- Pin 15 – Ground
Pin 5 - TMDS Data 4+ Pin 16 - Hot Plug Detect
Pin 6 - DDC Clock Pin 17 - TMDS Data 0-
Pin 7 - DDC Data Pin 18 - TMDS Data 0+
Pin 8 - Analog Vertical Sync Pin 19 - TMDS Data 0/5 S h ield
Pin 9 - TMDS Data 1- Pin 20 - TMDS Data 5-
Pin 10 - TMDS Data 1+ Pin 21 - TMDS Data 5+
Pin 11 - TMDS Data 1/3
Shield
2.3.4 Control Interface
Beside the OSD, we also support RS-232 interface to update and control the monitor by
software. When the monitor is connected with a CT or other iatrical diagnosis instruments, we
provide additional interface for the service technician to get the status of the monitor and update
software. When use RS-232 for software update, first put 5V switch (near the RS-232 input) to on
position. And then load the soft ware from computer. After complete software update, slide the
switch to off position. And then power off the monitor. Power on again the software update is
completed.
Last change: 27
th
, Oct. 2005
Copyright @ SSME (SIEMENS
Medical Equipment Ltd.) All
HL1916
Page 7 of 32
rights reserved
Figure 3 Female 9-PIN D-SUB connector
Table3
Pin Signal Pin Signal
1 Data Carrier Detect 6 Data Set Ready
2 Received Data 7 Request to Send
3 Transmitted Data 8 Clear to Send
4 Data Terminal Ready9 Ring Indicator
5 Signal Ground
Last change: 27
th
, Oct. 2005
Copyright @ SSME (SIEMENS
Medical Equipment Ltd.) All
rights reserved
HL1916
Page 8 of 32
2.4 Product Features
Item
LCD
Specification
Analog Input Digital Input
Panel Module LG.PHILIPS LM190E05-SL02
Size 19.0” (48 cm diagonal)
Active Display Area 376.32 (H) x 301.056 (V) mm
Resolution 1280 x 1024 dots (SXGA)
Pixel Pitch 0.294(H) x 0.294(V) mm
Color Depth 16.7M true 8 bit
Luminance 280 cd/m2 (typical); 230cd/m2 (min.)
Viewing Angle
(Type.)
H
L 89 degree
R 89 degree
CR>10
H 89 degree
V
L 89 degree
Contrast Ratio 600:1 (typical); 400:1 (min.)
Back Light CCFL x 6pcs.
Horizontal frequency 31kHz -82kHz 31kHz -82kHz
Vertical frequency
50.0Hz - 85.0 Hz
(Non-Interlaced)
50.0Hz - 85.0 Hz
(Non-Interlaced)
Video Signal Analog RGB Digital RGB
Input
Signals
Sync. Signal
Separate Sync. (TTL)
Composite Sync.
TMDS
Sync on green
Pixel Clock 25.0MHz -165.0MHz
Input connector Mini D-sub 15Pin
25.0MHz
-165.0MHz
DVI-I (D) &
DVI-D
Preset Timings Factory preset: 39 / User preset: 10 *
Control key Select, up, down, exit
Functions
Regulations
OSD
Safety IEC60601-1, CCC
Backlight Brightness, Contrast, Color control,
Position, Size, Phase, Gamma, etc.