SHARP R1850A, R-1851A, R-1852A Service Manual

---
---
SUPPLEMENTAL
SERVICE MANUAL
R-l
R-1851A R-l852A
S98M121 R1850A
850A
This is a This model is
Use this supplemental
supplemental
quite
Service Manual for
similar to the
manual
togetherwith the Base Model Service Manualsforcomplete operation and service
R-1851A
Model
Models R-l 850A/R-1851A/R-1852A.
Model R-l 850 (S/M# S6804R185OXDJ
Base

TABLE OF CONTENTS

OVER
THE
RANGE
MICROWAVE OVENS
R-l 850A/R-1851 A/R-l 85249
to be
user-safety the oven
interest
In the
original condition
its
to
of
and only the parts identical to
those specified should be used.
is
information.
restored
SECTION
PRECAUTIONS TO BE OBSERVED BEFORE AND DURING
SERVICE AVOID POSSIBLE EXPOSURE TO EXCESSIVE MICROWAVE ENERGY. Before Servicing
Forewordand
Warning to Service Personnel Microwave Measurement Procedure Wiring/SchematicDiagram
.............................................................
Warning
............................................................
......................................................
...........................................................
...................................................................
Parts ......................................................................................
PackingandAccessories
....................................................................
TO
SHARP ELECTRONICS CORPORATION
Service Headquarters: Sharp Plaza, Mahwah, New Jersey, 07430-2135
...........
PAGE
Inside Front Cover Inside Front Cover
1 2 3 4 6
13
R-1850A R-185lA R-l 852A
Do not operate or allow the oven to be operated with the door open.
(a)
Make the following safety checks on all ovens to be serviced before activating the magnetron
04
or other microwave source, and make repairs as necessary: (1) Interlock operation, (2) proper door closing, (3) seal and sealing surfaces (arcing, wear, and other damage), (4) damage to
or loosening of hinges and latches, (5) evidence of dropping or abuse.
Before turning on microwave power for any service test or inspection within the microwave gen-
(d
erating compartments, check the magnetron, waveguide or transmission line, and cavity for proper alignment, integrity, and connections.
Any defective or misadjusted components in the interlock, monitor, door seal, and microwave
(4
generation and transmission systems shall be repaired, replaced, or adjusted by procedures described in this manual before the oven is released to the owner.
A microwave leakage check to verify compliance with the Federal performance standard
03
should be performed on each oven prior to release to the owner.
Before servicing an operative unit, perform a microwave emission check as per the Microwave
Measurement Procedure outlined in the base model service manual. If microwave emissions level is in excess of the specified limit, contact SHARP ELECTRONICS CORPORATION immediately Q l-800-237-4277.
(RD21101U
If the unit operates with the door open, service person should 1) tell the user not to operate the oven
and 2) contact SHARP ELECTRONICS CORPORATION and DHHS immediately.
Service personnel should inform SHARP ELECTRONICS CORPORATION of any certified unit found with emissions in excess of 4mW/cm2. The owner of the unit should be instructed not to use the unit until the oven has been brought into compliance.
I
(RD81001 U)
SERVICE MANUAL
SHARC,
Over The Range
Microwave Ovens
R-l 850A/R-1851 A/R-l 852A
Foreword
This manual has been prepared to provide Sharp Electronics Corporation personnel with complete operation and service information for Sharp microwave oven models R-l 850A/R-1851 AIR-1 852A
It is recommended that service personnel carefully study the entire text of this manual so they will be qualified to render satisfactory customer service.
Check interlock switches and the door seal carefully. Special attention must be given to avoid electrical shock and microwave radiation hazards.
This supplemental service manual contains up­date information only. Please refer to base model service manual (R-1850) for complete service in­formation.
WARNING
Never operate the oven until the following points are ensured:
(A)The door is tightly closed. (B)The door brackets and hinges are not defective. (C)The door packing is not damaged. (D)The door is not deformed or warped. (E)There are no other visible signs of damage to the
oven.
DANGER
Certain initial parts are intentionally not grounded and present a risk of electrical shock only during servicing. Service personnel - Do not contact the following parts while the appliance is energized; High Voltage Capacitor, Power Transformer, Mag­netron, High Voltage Rectifier Assembly and High voltage Harness. If provided, Vent Hood, Fan Assembly and Cooling Fan motor.
All of the parts marked ages more than 250V.
Removal of the outer case cabinet gives access to voltage above 25OV.
All of the parts marked “A” on parts list may cause un­due microwave exposure, by themselves, or when they are damaged, loosened or removed.
“*I’ on parts list are used at volt-
Sharp Electronics Corporation
Sharp Plaza
Mahwah, New Jersey 07430-2135
I

WARNING TO SERVICE PERSONNEL

Microwave ovens contain circuitry capable of producing very high voltage and current. Con-
tact with following parts may result in a
possibly fatal, electrical
(EXAMPLE)
High Voltage Capacitor, Transformer, Magnetron, Assembly, High Voltage Harness etc..
Read
the Service
instructions.
Manual carefully and
shock.
High Voltage Power
High Voltage Rectifier
severe,
follow all

Before Servicing

1, Disconnect the power supply cord and then remove
outer
case cabinet.
2. Open the door and block it open.
3. Discharge the high voltage capacitor.
WAl?NING:RISK OF ELECTRICAL SHOCK. DISCHARGE THE HIGH VOLTAGE CAPACITOR BEFORE SERVICING.
The high-voltage capacitor remains charged about 60 se­conds after the oven has been switched off. Wait for 60 se­conds and then short-circuit the connection of the high volt­age capacitor (that is the connecting lead of the high voltage rectifier) against the chassis with the use of an insulated screwdriver.
Whenever troubleshooting is performed, the powersup­ply must be disconnected. In some cases it may be nec­essary to connect the power supply after the outer case has been removed, in this event:
Disconnect the power supply cord and then remove
1. outer case cabinet. Open the door and block it open.
2. Discharge the high voltage capacitor.
3. Disconnect the leads to the primary of the power
4.
transformer. Ensure that the leads remain isolated from other
5.
components and oven chassis by using insulation tape. After the above procedure, reconnect the power sup-
6.
ply cord.
Don’t
Touch
!
Danger High Voltage

When the testing is completed

Disconnect the power supply cord and then remove
1. outer case cabinet. Open the door and block it open.
2. Discharge the high voltage capacitor.
3.
4.
Reconnect the leads to the primary of the power transformer. Re-install the outer case
5. Reconnect the power supply cord after the outer
6. case cabinet is installed. Start the oven and check all functions.
7.
cabinet.

After repairing

Reconnect all Lads removed from components dur-
1. ing testing. Re-install the outer case (cabinet).
2. Reconnect the power supply cord after the outer
3. case is Reconnect the power supply cord after the outer
4. case is installed. Run the oven and check all functions.
5.
Microwave ovens should not be operated empty. To test for the presence of microwave energy within a cavity, place a cup of cold water on the oven turntable, close the door and set the power to HIGH and set the microwave timer for two (2) minutes. When the two minutes has elapsed (timer at zero), carefully check to see if the wa­ter is hot.
Servicing
the component being tested. When all service work is completed and the oven is fully
assembled, the microwave power output should be checked and the microwave leakage test should be car­ried out.
installed.
If the water remains cold, carry out
procedure and re-examine the connection to
Before
2
R-l
85OA
;:;gs
MICROWAVE
After
adjustment switches are completed individually or collectively, a IN­TERLOCK SWITCH AGE TEST must be performed with
DHHS
wave Ovens.
of the door, interlock and monitor
TEST and
(CDRH) Performance Standards for Micro-
MICROWAVE LEAK-
to
assure
MEASUREMENT
compliance

Interlock Switch Test

Make
sure that the door terlock switch by checking Refer to the “Test Procedure” of the secondary
and
with
interlock
an
sensing
monitor
switch are operating properly
ohmmeter.
switch and monitor switch.
switch, secondary in-
door sensing switch,

Microwave Leakage Test

Requirements
1. Microwave leakage limit (Power density limit).
The power density of microwave radiation emitted by a microwave point 5cm or more from the external surface of the oven, measured prior to acquisition-by the purchaser, and thereafter (through the useful life of the oven) 5mW/cm* at any point 5cm or more from surface of the oven.
2. Safety interlock switches.
Primary interlock relay and door sensing switch will prevent microwave radiation emissions in excess of the requirement as above mentioned, secondary interlock switch shall prevent microwave radiation emission in excess of 5 mW/cm* at any point 5cm or more from the external surface of the oven.
Preparation Before beginning the actual measurement of leakage,
proceed
1. Make sure that the actual instrument is operating nor-
Note:
as follows:
mally as specified in its instruction booklet.
Survey instruments that comply with the require­ment for instrumentatoin 21 CFR1030.1 O(c)(3)(i), as prescribed by the performance standard for microwave ovens must be used for testing.
oven shall not exceed 1 mW/cm* at any
the external
For Testing
PROCEDURE
2.
Place the
3. Place tially at 20 z!z
ity. The water container oz.) beaker with an inside
8.5 cm (3 % ductive material such as glass or plastic. The placing of this standard load in the oven is impor­tant any leakage is measured accurately.
4.
Set
5.
Close several survey is completed, ter.
Leakage Test
Closed-door leakage test (microwave measurement).
1. Grasp the probe of the survey instrument perpendicular to the gap between the door and the body of the oven.
2. Move the probe slowly, not faster than lin./sec (2.5cm/sec.) along the gap, watching for the maxi­mum indication of the meter.
3. Check for leakage at the door screen, sheet metal seams and other accessible positions where the con­tinuity of the metal has been switches, indicator and vents). While testing for leakage around the door pull the door away from the front of the oven as far as is per­mitted
4. Measure carefully at the point of the highest leakage and make sure that the highest leakage is than 4 mW/cm* to allow for measurement uncertainty, and that the secondary interlock switch does turn the oven OFF before any
Note:
oven tray
the load of 275 &
5OC
in.)
not only to protect the oven, but
the cooking control on
the door and
minutes.
by the closed latch assembly.
After servicing, record data on service invoice and microwave leakage report.
in
the oven cavity.
15ml(9.8
(68OF) in the center of the oven cav-
shall be a low form of
and made of an electrically noncon-
put the oven
If the water
replace it with 275ml of cool
door movement.
oz.) of tap water ini-
diameter
Full power
begins to
of approximately
also to insure
into
a cook
breached
600
ml
(20
that
cooking mode.
cycle
for
boil before the
wa-
and hold it
(eg., around
no greater
Loading...
+ 11 hidden pages