SHARP CX164 Technical Report

Date: Oct. 21, 1994
COLOR IMAGE SCANNER: JX-325
No. : CX-164
[Service information]
SERVICE MANUAL CORRECTIONS
1.Model name: JX-325
2.General: There were false descriptions in the service manual. We have corrected them.
3.Description: (1)There were missed descriptions in the page 3 “• Maintenance mode setting”
Therefore we have added Panel display test and Heater test. These descriptions have been in the item 5. Diagnostics.
Maintenance mode setting
Switch
SW1
(*) The maintenance mode has the following options.
.
No
1 00000000Standard
2 11110000
3 01010101
4 10100001
5 00000011
6 00001011
7 00101011
8 01001 011
9
10 00011111
(*1) When you carry out this test, the DIP SW2 has to be set as following diagram.
Switch No.
8 0 1 Maintenance mode (*)
Mode
Transit, packing
Panel display test (*1)
Heater test
Aging test
White balance test
Correction circuit test
Line buffer test
MTF test
Transparent unit test
DIP SW2 12345678 10101010
DIP SW1
12345678
00011011
Content
Normal mode
Function
Normal operation Shifts the scanner
head to the fixed position for transit.
Checks the operation of DIP SW1, SW2, and display LED.
Checks the power supply to the heater.
Lights the three lamps in sequence, and performs all area scanning.
Repeats white balance lighting after white balance.
Checks shading and the dark-component correction circuit function.
Checks the MTF peripheral line memory operation.
Checks MTF edge emphasis.
Checks the transparent unit operation.
1/2
Orange
C
(2)In the “5-2. Diagnostic execution and cancellation” it was described that the white
balance test, the correction circuit test and the line buffer test were executed by turning off the DIP SW1 No. 8. However this was false. There is no need to turn the switch off. We have made a new flow chart this time.
5-2. Diagnostic execution and cancellation5-2. Diagnostic execution and cancellation
Power is turned OFF
Set DIP SW1 and SW2 to the desired mode (Refer to 5-1)
Power is turned ON
Execution of tests
Power is turned OFF
All of DIP SW1 are set to OFF
Execution of diagnostic
Refer to 5-3-1 turn 5-3-6
Cancel lation of diagnostic
Caution: About the MTF test and the transparent unit test, testing methods are shown in
the item 5-3, the test diagnostic contents.
2/2