Sewha SI 4010 Instruction Manual

Digital Weighing Controller
SI 4010
Instruction Manual
Version 3.20 (May 2011)
DIGITAL WEIGHING INDICATOR
SI 4010
2
CONTENTS
1. Before Installation
---------------
3 page
2. Introduction
---------------
4 page
3. Specification
---------------
5 page
4. Installation
4-1. External Dimension & Cutting Size 4-2. Installation Components 4-3. Load Cell Installation
---------------
---------------
---------------
---------------
11 page
11 page 12 page 12 page
5. Set-up
5-1. Calibration 5-2. TEST Weight Calibration Mode 5-3. Simulation Calibration Mode(Without Test Weight) 5-4. Set-up 5-5. F-Function
---------------
---------------
---------------
---------------
---------------
---------------
13page
13 page 13 page 16 page 19 page 20 page
6. Interface
6-1. Serial Interface (RS-232C) 6-2. Current Loop Interface 6-3. Analog Output Interface (0~10V) 6-4. Analog Output Interface (4~20mA) 6-5. Serial Interface (RS-232C / 422 / 485) 6-6. BCD Input ( BCD INPUT : P/N Change) 6-7. BCD Output 6-8. Print Interface 6-9. Command Mode
---------------
---------------
---------------
---------------
---------------
---------------
---------------
---------------
---------------
---------------
27 page
27 page 31 page 32 page 33 page 34 page 35 page 36 page 37 page 38 page
7. Error & Treatment
7-1. Load Cell Installation 7-2. Calibration Process 7-3. Indicator Error & Treatment 7-4. Indicator Test Mode
---------------
---------------
---------------
---------------
---------------
40 page
40 page 40 page 41 page 42 page
Warrantee Certification
---------------
43 page
DIGITAL WEIGHING INDICATOR
SI 4010
3
1. BEFORE INSTALLATION
1-1. Caution / Warning Marks
This mark warns the possibility to arrive death or serious injury in case of wrongly used.
This mark cautions the possibility to arrive serious human body injury or product lose in case of wrongly used.
1-2. Other Marks
Warning for Electric Shock or Damage. Please do not touch by hand
Protective Ground(Earth) terminal
Prohibition of Operation process
1-3. Copy Rights
1). All Right and Authority for this Manual is belonged to Sewhacnm Co.,Ltd.
2). Any kinds of copy or distribution without Sewhacnm Co.,Ltd’s permission will be prohibited.
1-4. Inquiries
If you have any kinds of inquiries for this model, please contact with your local agent or Head Office. Head Office : Sewhacnm Co.,Ltd. Website : http://www.sewhacnm.co.kr Email : sales@sewhacnm.co.kr, info@sewhacnm.co.kr
DIGITAL WEIGHING INDICATOR
SI 4010
4
2. INTRODUCTION
2-1. Introduction
Thank you for your choice, this “SI 4010” Industrial Digital Weighing Indictor.. This “SI 4010” model is simple weight display purpose with Large display(1inch),
and powerful communication performance. With 2ports serial port communication and High Speed A/D conversion performance will lead you to precise weighing process. Please review this instruction Manual and learn more about information about “SI 4000”. Enjoy your process efficiency with “SI 4010” Weighing Indicator..
2-2. Cautions
1) Don’t drop on the ground or avoid serious external damage on item.
2) Don’t install under sunshine or heavy vibrated condition.
3) Don’t install place where high voltage or heavy electric noise condition.
4) When you connect with other devices, please turn off the power of item.
5) Avoid from water damage.
6) For the improvement of function or performance, we can change item specification without prior notice or permission.
7) Item’s performance will be up-dated continuously base on previous version’s performance.
2-3. Features
1) All Modules and Option Cards are isolated to maximize accuracy and performance.
2) External input terminal inside.(4pcs:Can be set by F11 mode)
3) By using “Photo-Coupler” on each module (Option, Analog board, In/Out), we improved “Impedance
problem”, “Isolation ability among inputs”, “Leading power problem”, and “Noise covering function”.
4) Data back-up function, when the sudden power off.
5) Polycarbonate film panel, strong against dust and water.
6) 2port Serial Interface - RS-232C (Com. Port1) is standard installed.
7) HIGH / LOW Set points and Free Fall function for HIGH Set point.
8) Weight Unit selection Function added. (“g”, “kg”, “t” selectable – F40)
9) Variable options(Order in advance, Refer Chapter 5. Interface)
Cautions
!
DIGITAL WEIGHING INDICATOR
SI 4010
5
3. SPECIFICATION
3-1. Analog Input & A/D Conversion
Input Sensitivity
0.2 / Digit
Load Cell Excitation
DC 10V ( - 5V ~ + 5V )
Max. Input Signal
Max.3.2mV/V
Temperature Coefficient
[Zero] ±20PPM/ [Span] ±20PPM/
Input Noise
±0.3 P.P
Input Impedance
Over 10
A/D Conversion Method
Sigma-Delta
A/D Resolution(Internal)
520,000 Count(19bit)
A/D Sampling Rate
Max. 500times / Sec
Non-Linearity
0.005% FS
Display Resolution(External)
1/20,000
3-2. Digital Part
Display
Parts
Specification
Display
Main Display
7Segments, RED FND Display Size :24.5mm(H) Large Display (1inch)
Min. Division
×1, ×2, ×5, ×10, ×20, ×50
Max. display value
+999,950
Under Zero value
"-" (Minus display)
Status lamp
Steady, Zero, Tare
Green LED Display(2Ø )
kg, g, ton
K e y
Number Key : 10 ea , Function : 3 ea , CAL. Lock key : 1 ea
3-3. General Specification
Power Supply
AC110/220V±10%), 50/60Hz, about 30VA
Operating Temperature Range
-10 ~ 40
Operating Humidity Range
Under 85% Rh (non-condensing)
External Dimension
200mm(W)105mm(H)165mm(L)
Net Weight(kg)
About 2.3kg
Gross Weight(kg)
About 3.0kg
AC 110V, Power supply is an optional before ex-factory.
DIGITAL WEIGHING INDICATOR
SI 4010
6
3-4. Option Card
Option No.1
Printer Interface : Centronics Parallel
Option No.2
Analog Output (0~10V or 0~5V)
Option No.3
Analog Output (4~20mA)
Option No.4
Serial Interface : RS-232C / 422 / 485
Option No.5
BCD Input : Part No. Change Purpose
Option No.6
BCD Output
Option No.7
Ethernet
Serial Interface (RS-232C) or Current Loop is Standard installed.
In the Optional Serial port, there is no Current Loop function.
3-5. Front Panel (Display / Key Pad)
3-5-1. Status Lamp (ANNUNCIATORS) : “LED” Lamp is “ON”
Steady
When the weight is Steady, “LED” Lamp is “ON”
Zero
When the current weight is Zero, “LED” Lamp is “ON”
Tare
Tare function is set, “LED” Lamp is “ON”
(Tare Reset “LED” Lamp is “OFF”.)
Weight Display
Status Lamp
Weight Unit LED
Calibration Lock Key
Function key
DIGITAL WEIGHING INDICATOR
SI 4010
7
3-5-2. Key Operation
Make Weight value as Zero. Under F08, you can set the Zero key operation range, as 2%, or 5%, or 10%, or 20% of Max. Capacity. Under “Tare” key input, Zero key will not be activate within operation range.
Make Weight value as Zero, including Tare Weight. Under F09, you can set the Tare key operation range, as 10%, 20%, 50%, or 100% of Max. Capacity. Whenever pressing “Tare” key, you can set the Tare continuously.
TARE RESET
1. Remove the Set TARE function.
- If you press this key, TARE set value will be removed and display gross weight.
Hold the Weight display when indicator detects “Peak Hold”, or “Sample Hold”. You can select “Hold” function on F10. First input : HOLD is set Second input : HOLD Reset
Under “TARE” setting, you can select weight display mode. First input, Gross Weight will be displayed, second input, Net weight will be displayed. This key will be activated only under “TARE” set.
You can set each weighing process as a certain P/N. And you can call certain P/N with pressing this key. P/N save : Select P/N and Enter key input. P/N call : P/N + Number key + Enter
Print out All Part Nos’ weighing data will be printed. (Grand Total Data Print Out)
Current Part No.’s weighing data will be printed. (Sub Total Data Print Out)
DIGITAL WEIGHING INDICATOR
SI 4010
8
No. specific function
1. Manual Print
Whenever press this key, you can print out.
2. Calibration mode
- Digit setting : Whenever pressing “0”key, digit will be change 1, 2, 5, 10, and 50.
- Decimal point position : Whenever pressing “0”key, decimal point will be change. Decimal Point set will be done in the calibration mode.
1. Enter F-Function / Test mode
2. set the Function No, in F-Function mode.
1. Modify the set value during setting process.
2. Calibration mode - Move back to previous step.
3. F-function setting mode - Change F-function No. F-function no.(number key) + Clear directly move to that F-function
4. Function key : Sub-total, Grand-total manual delete.
1. Save set value during setting process.
2. Calibration mode - Save current setting and move to next step.
3. F-Function mode- Save current F-function setting, and move to next F-function
Enter/Exit to “Calibration” mode.
First time press “CAL. Button”, enter the calibration mode. And automatically exit, when calibration is finished.
Function Keys (Combined Key functions)
Delete “Gross-Total data”
Delete “Sub-Total data”
DIGITAL WEIGHING INDICATOR
SI 4010
9
Warning
!
3-6. Rear Panel
POWER AC IN
- Power switch : Power on/off switch.
- Fuse : AC250V / 0.5A , φ5.25 , 20mm.
- AC IN : Available Input AC 110V / 220V.
The standard power supply is AC 220V(Fixed when ex-warehouse), if you want to have AC
110V, please inform in advance.
Option Card 1 Option Card 2
Option Card Connector installed for Optional Interface or Output.
(Printer I/F, Analog out, RS-422/485, or RS-232C(two port)
LOAD CELL Connector (N16-05) SERIAL I/F
“RS-232C” or “CURRENT LOOP”(9Pin, D-Type Female) are built-in as standard
External Input : External control input for wired remote control.
Refer to F-Function F11 to select desired function mode.
○3
1
○6
○5
○4
DIGITAL WEIGHING INDICATOR
SI 4010
10
4. INSTALLATION
4-1. External Dimension & Cutting Size
(External Dimension) (unit : mm)
(Cutting Size) (unit : mm)
4-2. Installation Components
Power Cable
Communication Connector (D-SUB 9P)
Load-cell Cable
186
165
105
200
92
3
94
162
188
DIGITAL WEIGHING INDICATOR
SI 4010
11
4-3. Load Cell Installation
4-3-1. Load Cell Connector Specification
4-3-2. Load Cell Installation
1) You can connect Max. 8pcs of same capacity Load cells at once. (350Ω)
2) You have to make horizontal balance on the ground.
3) If you install more than 2pcs of Load cells, use Summing box and adjust output signal difference as minimum. It can make wrong weighing process caused by each load cell’s variation.
4) If there is some temperature difference around Load cell, it can cause wrong weight measurement.
5) Don’t do Welding job or Arc discharge around installation place. But, there is no choice, please disconnect power cable and Load cell cable.
6) If you measure static electricity material, please make earth between down part and up part of Load cell.
Red(EX+) White(EX-) Green(SIG+) Blue(SIG-) Yellow(SHLD)
DIGITAL WEIGHING INDICATOR
SI 4010
12
4-3-3. Formula to plan the precise weighing system
This “SI 4010” weighing controller’s Max. input sensitivity is 0.2 / Digit. And for precise weighing system, the following formula must be satisfied. Caution : “Input sensitivity” means Min. output voltage variation of weighing part to change 1digit. So,
please do not make large input voltage to make reliable weighing system.
Single Load cell use
0.2 EBD
A : Load cell capacity(kg) B : Load cell Voltage(mV) D : Digit E : affirmation Voltage of Load cell N : Number of Load cell
A
Plural Load cells use
0.2 EBD
AN
Example1)
Number of Load cell : 1pcs Load cell capacity : 500kg Load cell Voltage : 2mV/V Digit : 0.05kg Affirmation Voltage of Load cell : 5V Max. Capacity of Weighing System : 300kg
Then, estimation result for this weighing system with formula,
5000ⅹ2ⅹ0.05
= 1 ≥ 0.2
The calculated value is larger than 0.2, so this system has no problem.
500
Example2)
Number of Load cell : 4pcs Load cell capacity : 500kg Load cell Voltage : 2mV/V Digit : 0.10kg Affirmation Voltage of Load cell : 5V Max. Capacity of Weighing System : 1,000kg
Then, estimation result for this weighing system with formula,
5000ⅹ2ⅹ0.10
= 0.5 ≥ 0.2
The calculated value is larger than 0.2, so this system has no problem.
500ⅹ4
Cautions
!
DIGITAL WEIGHING INDICATOR
SI 4010
13
5. SET-UP
5-1. Calibration
Adjust weight balance between “Real weight” on the load cell(Weight Part) and “Displayed weight of Indicator”. When you replace LOAD CELL or Indicator, you have to do Calibration process once again
5-2. Test Weight Calibration Mode (Using Test weight)
remarks : In case that P-W is displayed, you have input the pass word to start
calibration mode.
Prepare At least 10% of Max. capacity of your weighing scale, and remove “CAL-BOLT” on the Front panel
and press “CAL - LOCK S/W” inside.
1. At normal mode, remove “CAL-BOLT” on the Front panel
123
2. And press “CAL - LOCK S/W” inside.
Check the “SET-CAL. Massage on display.
5et-Cal
For the save the each step, press key, for the cancel or move back, press key.
3. If you press key, Calibration Mode starts.
After displaying “CAPA”,
C999999
4. Please input Max capacity of your weighing scale,
And press key. Ex) Load cell CAPA : 20kg, division : 0.001 Input 20000
C 20000
5. Define the optimal position or Decimal point
Whenever pressing key, Decimal point will be changed. Ex) Load Cell CAPA : 20kg, division : 0.001kg input 20.000
p 20.000