Congratulations. You have chosen a quality instrument manufactured
by Hermann Sewerin GmbH.
Our equipment will provide you with the highest standards of perfor-
mance, safety and efciency. They correspond with the national and
international guide-lines.
Please read and understand the following operating instructions before
using the equipment; they will help you to use the instrument quickly and
competently. If you have any queries we are available to offer advice
and assistance at any time.
Yours
Hermann Sewerin GmbH
Robert-Bosch-Straße 3
33334 Gütersloh, Germany
Tel.: +49 5241 934-0
Fax: +49 5241 934-444
www.sewerin.com
info@sewerin.com
SEWERIN SARL
17, rue Ampère – BP 211
67727 HOERDT CEDEX, France
Tél. :+33 3 88 68 15 15
Fax :+33 3 88 68 11 77
www.sewerin.fr
sewerin@sewerin.fr
SEWERIN IBERIA S.L.
Centro de Negocios “Eisenhower”
Avenida Sur del Aeropuerto
de Barajas 28, Of. 2.1 y 2.2
28042 Madrid, España
Tel.: +34 91 74807-57
Fax: +34 91 74807-58
This product may only be operated by appropriately-trained persons who
are familiar with the relevant operating manual.
It may only be used for its designated purpose, i.e. for industrial and commercial use.
Repair work may only be carried out by specialists or by persons who have
undergone appropriate training.
Any alterations or modications to the product require the prior approval of
Hermann Sewerin GmbH. In the event of unauthorised alterations to the
product the manufacturer accepts no liability for damage.
Only Hermann Sewerin GmbH accessories may be used with the product.
Only spare parts approved by us may be used for repairs.
Hermann Sewerin GmbH accepts no liability for damage resulting from
non-compliance with the foregoing. The guarantee and liability provisions
in the Hermann Sewerin GmbH terms of sale and supply are not extended
by the foregoing.
We reserve the right to make changes in the context of continued technical development.
In addition to these instructions, please comply with generally applicable
safety and accident-prevention regulations.
Symbols used:
CAUTION:
Note:
This symbol warns of dangers that may threaten
the safety of the user or maty damage or destroy
the product.
This symbol ags information and hints extending
beyond the actual operation of the product .
Page 7
Contents Page
1 EX-TEC OD 4/EX-TEC GM 4 system ......................................1
1.1 The EX-TEC OD 4 system ........................................................1
1.1.1 Before starting work ...............................................................1
1.1.2 The need for gas odorisation .................................................1
13.3 EX-TEC OD 4/GM 4 operating instructions in brief ................. 92
III
Page 10
1 EX-TEC OD 4/EX-TEC GM 4 system
1 EX-TEC OD 4/EX-TEC GM 4 system
The EX-TEC OD 4 is a combined warning and measuring instru-
ment for determining odorant concentration. It is available with
either a THT or a TBM sensor.
The EX-TEC GM 4 is a warning and measuring instrument that
detects oxygen and various toxic gases.
1.1 The EX-TEC OD 4 system
1.1.1 Before starting work
Adequate gas odorisation is an important criterion for safe network
operation. Please take careful note of the following information.
1.1.2 The need for gas odorisation
Natural gas is odourless. So that even the smallest leaks in indoor
installations or elsewhere in a gas-distribution network can be
immediately detected, the gas is odorised: a strong-smelling substance is added to it. The substances used are organic sulphur
compounds and acrylates. The most widespread currently being
tetrahydrothiophene (THT) and tertiary butyl mercaptan (TBM).
DVGW worksheet G 280 sets the minimum values for odorant
concentration at 10 mg of THT / m3 and 3.8 mg of TBM / m3. This
worksheet also requires the operator of a distribution network to
check and document the odorant concentration at least once a
year.
1.1.3 Measurement principle
The EX-TEC OD 4 contains an electrochemical sensor that reacts
to the smallest traces of sulphur compounds. This produces elec-
trical signals, which are then processed by the integral electronic
system. The intensity of the signal varies directly with the strength
of the concentration.
1
Page 11
1 EX-TEC OD 4/EX-TEC GM 4 system
The sensor is lled with a liquid (electrolyte), which makes it very
sensitive to high temperatures. The maximum storage temperature of 40 °C must therefore not be exceeded. Electrochemical
processes constantly take place inside the measurement cell,
requiring the EX-TEC OD 4 to be permanently supplied with battery power at an adequate voltage. For this reason the integral
batteries must not be permitted to reach a state of deep discharge.
When the instrument is not in use it should therefore be stored
in charge-maintenance operation (see section 4.1: Charging and
charge maintenance when using NiMH batteries).
1.1.4 Boundary conditions for measurement and sampling
As well as all sulphur compounds, the electrochemical measurement cell also reacts very strongly to moisture. Even though this
is rarely important when carrying out measurements in a gasdistribution network, operators should not blow through sampling
hoses to clear them.
When sampling it is essential to ensure that there are no connecting pieces made of rubber or synthetic substances between the
EX-TEC OD 4 and the connection to the gas main unless these
consist of inert materials (e.g. Viton or nylon).
A further factor that affects the measurement of sulphur com-
pounds with the EX-TEC OD 4 is the ow rate through the mea-
surement chamber, which must always exceed 20 l/h. Only the
connection hoses supplied with the instrument should therefore
be used for sampling.
2
Page 12
1.1.5 EX-TEC OD 4
The following sensors are available:
1 EX-TEC OD 4/EX-TEC GM 4 system
The EX-TEC OD 4 is a hand-held
instrument for the measurement
of tetrahydrothiophene (THT) or
tertiary butyl mercaptan (TBM). It
is available in pump and diffusion
versions.
Sensor
THT
TBM
rangeprinciple
0 –100 mg/m
0 –100 mg/m
3
3
electrochemical
electrochemical
3
Page 13
1 EX-TEC OD 4/EX-TEC GM 4 system
1.1.6Testcerticates
Passive explosion protection
The EX-TEC OD 4 has been tested for explosion protection in
accordance with European norms (CENELEC):
EC prototype test certicate: TÜV 01 ATEX 1657
Identication: II2G Ex e ib IIB T4 Gb
Testing institution: TÜV NORD CERT GmbH,
Hannover
4
Page 14
1.2 EX-TEC GM 4 system
The EX-TEC GM 4 is a gas-measurement and warning instru-
ment. It is available in pump and diffusion versions. A different
sensor is used for each gas.
The following electrochemical sensors are available:
Sensor range
Carbon monoxide CO 0 – 500 ppm
Hydrogen H2 0 – 1.0 vol.%
Hydrogen sulphide H2S 0 – 100 ppm
Hydrogen sulphide H2S 0 – 2000 ppm
Ammonia NH3 0 – 100 ppm
Oxygen O2 0 – 25 vol.%
Hydrogen chloride HCI
1.2.1Testcerticates
Passive explosion protection
The EX-TEC GM 4 has been tested for explosion protection in
accordance with European norms (CENELEC):
1 EX-TEC OD 4/EX-TEC GM 4 system
0 – 30 ppm
EC prototype test certicate: TÜV 01 ATEX 1657
Identication 1: II2G Ex e ib IIB T4 Gb
basic instrument without leather case for all gases
except hydrogen H
2
Identication 2: II2G Ex e ib IIC T4 Gb
basic instrument with leather
case for all gases includinghydrogen H
2
Testing institution: TÜV NORD CERT GmbH,
Hannover
5
Page 15
2 Safety
2 Safety
2.1 Safety instructions
CAUTION!
Always use original SEWERIN accessories with the
EX-TEC OD 4/GM 4!
CAUTION!
Observe the permissible operating temperature
range of minus 10 °C to plus 40 °C!
(with some sensors the temperature range may be
wider - see section 11.2: Technical data)!
CAUTION!
The EX-TEC OD 4/GM 4 must not be recharged or
its battery compartment opened in areas exposed
to the danger of explosion!
CAUTION!
Test gases may only be used in well-ventilated
spaces!
CAUTION!
The EX-TEC OD 4/GM 4 satises the limits of the
EMV regulation. When using it near (mobile) radio
equipment please also observe the instructions in
the manuals for that equipment!
6
Page 16
CAUTION!
Please note that when the EX-TEC OD 4/GM 4 is
in measuring operation the gas sample is released
into the ambient air through its outlet.
Special care must be taken to ensure that this does
not produce an explosive or toxic gas mixture, espe-
cially during use in conned spaces. It may therefore
be necessary to use a gas-warning device to monitor
the ambient air.
CAUTION!
The EX-TEC GM 4 with the hydrogen sulphide H2S
sensor (0 – 2000 ppm) is a pure measuring instru-
ment: the „warning“ operating mode is not available.
Please note that the human sense of smell fails at
concentrations of 100 ppm of H2S and above. Concentrations of over 500 ppm of H2S lead to very seri-
ous damage to health, uncoordinated movements,
unconsciousness, respiratory failure and death. It is
thus essential to ensure that the lower toxic limit of
10 ppm in the ambient air is not exceeded!
2 Safety
CAUTION!
When using the H2 sensor the HG4 must be in its
leather case (to provide explosion protection)!
7
Page 17
3 Measuring operation
V1.001
EX-TEC GM4
H2
3 Measuring operation
3.1 Switching on
Note:
This operating manual describes the functions
of software version 1.XXX. Other versions may
involve changes.In this event please consult any
page insertions.
z always switch the instrument
on in „fresh air“
z hold the on/off key down for
about 3 seconds
z the optical and acoustic control
signals (items 1 and 7) operate
for about 3 seconds
z the LC display illumination
automatically switches on
z pump instruments: the inte-
gral pump runs at constant
power
Opening screen
z display of:
- version number (V1.001)
- instrument type (GM 4)
- sensor type (H2)
8
27.11.2000
12:37
Stadtwerke GT
Heinz Muster
Abt. TDLL 24.4
Date/time
z display of:
- date (27.11.2000)
- time (12:37)
- user name
z properly-set values are impor-
tant for the documentation of
your readings
Page 18
3 Measuring operation
Sensor install.
12.06.2002
durability
24 Monate
H2
0...1,00 Vol%
z you can correct any variances
(cf. section 8.2: Setting the
date/time)
Instrument settings
The EX-TEC OD 4/GM 4 defaults
to a set operating mode at switchon, which you can change (cf.
section 9.4: Autostart). The
operating mode is factory-set to
„Measurement“:
z display of:
- battery-charge level
- gas (here: H2)
- operating mode
(here: measurement)
- range
(here: 0 – 1.00 vol.%)
Other possible settings can be
found in section 9.4: Autostart.
z display of the day the sensor
was tted
z display of expected sensor
life
Symbols
z available operating hours are
displayed (5 hours = 5 bars)
symbol: battery
z measurement is under way
(symbol eggtimer)
z all alarm thresholds are acti-
vated
symbol: alarm threshold
9
Page 19
3 Measuring operation
Next scheduled inspection
(display optional)
z if you have set an inspec-
tion interval, the next scheduled inspection date (e.g.
11.03.2004) is displayed for
about 3 seconds (cf. section
8.3: Setting the inspection
interval))
z depending on the current and
inspection dates, the intermittent or continuous alarm may
also be triggered (items 1
and 7)
z clearing the alarm with the
on/off key or waiting for 15
seconds switches to measuring operation
Note:
If the EX-TEC OD 4/GM 4 now automatically
switches off, the inspection date has passed with
the inspection block switched on (cf. section 8.0:
System menu).
The instrument does not revert to measuring opera-
tion until an inspection has been carried out and
conrmed.
10
Page 20
3.2 Operating modes
WARNING
z Warning of high gas concentration in buildings and the work-
ing environment
z Testing enclosed spaces
H2
1500
PPM
3 Measuring operation
z all alarm thresholds are acti-
vated
NB:
z the current gas concentration
is displayed:
- as a gure
(e.g.: 1500 PPM H2)
- as a trend bar
z every 5 seconds an operating
signal - both acoustic (item 7)
and optical (item 1) - is emitted
to conrm that the instrument
in functioning in the warning
operating mode. If this operating signal is not emitted,
you cannot be sure that the
gas concentration is being
properly monitored and must
leave the endangered area
without delay.
11
Page 21
3 Measuring operation
H2
PPM
1500
MEASUREMENT
EX-TEC GM 4:
z Testing the gas concentration in buildingsz Measuring gas concentrations in pipes and elsewhere
z Searching for leaks in pipes in buildings
z Checking joints, e.g. weld seams
z Carrying out building tests
z Checking enclosed spaces
EX-TEC OD 4:
z Determining the odorant concentration in natural gasz Measuring the odorant concentration in the air in odorising
plants
z Before sampling observe the sufcient ush of intake
z the current gas concentration
is displayed:
- as a gure
(e.g.: 1500 PPM H2)
- as a trend bar
12
Sampling hoses
To ensure that the sample always passes through the
EX-TEC OD 4 at an adequate rate, there are different sampling
hoses for each of the possible pressure ranges. These are colourcoded as follows:
this alarm threshold, the optical and acoustic alarms (items
1 and 7) switch off
16
Page 26
3 Measuring operation
H2
AL2
VOL%
0,8
AL2-alarm (example: H2)
z if this alarm threshold is
exceeded, the result is:
- optical alarm (item 1)
- acoustic alarm (item 7) at
5 Hz
- LCD (item 2) lights up
- AL2 appears in the LCD
z the rapid intermittent tone at
5 Hz is quite different from the
operating signal
z the AL2 alarm cannot be
cleared
z if the concentration falls below
this alarm threshold, the alarm
becomes an AL1 alarm, which
can be cleared
17
Page 27
3 Measuring operation
0
3.4 Illumination and operating-hours display
z pressing any key switches the
LCD illumination on
z ou can also adjust the illumina-
tion time individually (cf. sec-
tion 9: Hardware menu)
z when the illumination comes
on the remaining operating
hours (e.g. 5 hours) are displayed
z this display (battery symbol
and bars) and the illumination
switch off automatically after
about 10 seconds
z the battery symbol appears in
the LC display and the operating signal sounds at twice
its normal rate
z you now have at least another
15 minutes' operating time
available
z after that the instrument must
be charged (cf. section 4:
Charging and battery opera-
tion)
1500
3.5 Battery alarm
H2
PPM
H2
PPM
18
Page 28
3.6 Switching off
3 Measuring operation
z hold down the on/off key for
about 3 seconds
z the optical and acoustical
control signals (items 1 and
7) are activated for about 3
seconds
z remaining operating hours are
displayed in the form of the
battery symbol and bars (e.g.
5 hours = 5 bars)
19
Page 29
4 Charging and battery operation
4 Charging and battery operation
CAUTION!
The EX-TEC OD 4/GM 4 must not be recharged or
its battery compartment opened in an area exposed
to the danger of explosion
4.1 Charging and charge maintenance when using NiMH batteries
Batteries should be approximately at room temperature for charging.
Charging
After 12 hours‘ charging the instrument has about 20 hours of oper-
ation with the pump running or about 50 hours of operation in diffusion operation.
The docking station HG4 is required for charging. It can be used
either in the workshop or in the emergency vehicle.
The docking station has the following connections on its side:
z AC/DC adapter M4 for
100 – 240 V~
z Vehicle cable M4 12 V= mount-
ing
z Vehicle cable M4 12 V=
mobile
z Vehicle cable M4 24 V= mount-
ing
z You can operate up to 5 dock-
ing stations with one 100 – 240
V~ AC/DC adapter M4
z With more than 5 stations the
charg. volt.
too low
charg. volt.
too high
charging voltage is insufcient.
An error message is displayed
to that effect
z if you connect the docking sta-
tion directly to a 24-V network in
the vehicle, the voltage will be
too high for proper charging
20
Page 30
4 Charging and battery operation
04
Place the EX-TEC OD 4/GM 4 in the docking station and a display
on the following lines appears:
z the instrument now has 5
operating hours left (= 5 bars)
and needs another 4 hours to
reach full charge
z if it is fully charged, all the bars
appear and the digital display
disappears
Charge maintenance
Once the instrument is fully charged, it switches automatically to
charge maintenance. It can be left in the docking station until the
next time it is needed.
21
Page 31
4 Charging and battery operation
4.2 Self-dischargeand battery care
If the instrument is not placed in the charging adapter when it is
switched off, this will cause the nickel/metal-hydride battery to
self-discharge, reducing the remaining operating time.
After a maximum of 30 days the instrument indicates no remain-
ing operating hours and it must be recharged.
Note:Brief periods of use and protracted disuse may in
the long term lead to the so-called „memory effect“,
which means that the actual battery capacity available is less than what is shown in the display.
You can counteract this by fully discharging the
EX-TEC OD 4/GM 4 regularly (e.g. once a month)
and then recharging it.
Switch the instrument on and place it in the docking
station. When it is fully discharged, charging begins
automatically. The entire process takes:
22
20 h (typical discharge, pump instrument)
+ 12 h (full charge)
= 32 h
Page 32
4.3 Battery operation
charge aborted
Alcalinebattery
With new non-rechargeable alkaline batteries the EX-TEC OD 4/
GM 4 will operate for 20 hours with the pump running or 50 hours
in diffusion operation.
Changing the batteries
When the batteries need to be changed, the battery compartment
can be opened with the screwdriver supplied (the two lower screws
on the back of the instrument). Insert the new cells the right way
round (as marked) and close the battery compartment again.
Once the old batteries are removed the date and time are preserved in the instrument for about 120 seconds. If the battery
change takes longer than this, you will have to re-enter both values
next time it is switched on. No other data will be lost.
The value for the operating-hours display for nickel/metal-hydride
batteries is lost if these are replaced by alkaline batteries, so if
nickel/metal-hydride batteries are inserted again they must rst
be fully discharged and then recharged. This ensures that the
available operating hours will be correctly displayed.
4 Charging and battery operation
z An EX-TEC OD 4/GM 4 tted
with non-rechargeable alkaline
batteries cannot be charged.
If you try to do so a warning
message is displayed and the
instrument must be removed
from the docking station.
23
Page 33
4 Charging and battery operation
4.4 Instruments showing serial number 062 00
WARNING!
To ensure that the device remains explosion-proof
as per ATEX 100a, only the following disposable/
rechargeable batteries may be used:
– Those supplied by SEWERIN,– Others offered by SEWERIN, provided that
compliance with standard EN 60079-7:2003 (especially chapter 5.7.2.1.17; explanation stated
below) is guaranteed.
The types used in a battery compartment must
always be identical in terms of sort (disposable/rechargeable), capacity and manufacturer.
Disposable battery requirements
z Battery size: Mignon AAz The creepage distance and air gap between the poles must
not be less than 0.5 mm in accordance with (EN 60079-7:2003;
Kap. 5.7.2.1.17).
z Alkaline batteries must comply with EN 60086-1 type LR6.
24
Rechargeable battery requirements
z Battery size: Mignon AAz The creepage distance and air gap between the poles must
not be less than 0.5 mm in accordance with (EN 60079-7:2003;
Kap. 5.7.2.1.17).
z The accumulators must comply with DIN EN 61951-2 type HR6
and adhere to the temperature range.
CAUTION!
A device operated with disposable alkaline batteries
cannot be charged. A notice to this effect is shown
on the display.
The device comes with nickel metal hydride rechargeable batteries. The corresponding settings are saved.
Page 34
5 Testing the instrument
5.1 Testing/upkeep
The required and prescribed instrument tests must be carried out
in accordance with the following norms:
z EN 50073 (Guidelines for the selection, installation, operation
and maintenance of instruments for the detection and mea-
surement of combustible gases and oxygen)
z EN 45544-4 (Electrical workplace-atmosphere instruments for
the direct detection and direct concentration measurement of
toxic gases and vapours, part 4: Guidelines for selection, installation, operation and upkeep)
z BGI 518 (T023)z DVGW worksheet G 465-4
The instrument tests prescribed by DVGW worksheet G 465-4
(Technical information, note) are divided into the following sections. Testing also covers accessories used with the instru-
ments.
What ?Who ?When ?
function testuserbefore use
display-accuracy
test (adjustment)
upkeep
(maintenance, re-
pair if necessary)
expert or special-
ist rm
SEWERIN, expert,
authorised special-
ist rm
5 Testing the instrument
daily to half-yearly
annually or whenever a defect occurs
Function test
This is the simplest form of instrument test. Carried out by the
user before use, it comprises the following points:
z external condition including probe systemsz function of the operating controlsz function of the operating controlsz inspection of the pump and the intake channelz pump functionz zero-point check
25
Page 35
5 Testing the instrument
Checking display sensitivity (adjustment)
Testing frequency must be specied as a function of the sensors
tted and the use of the instrument. It can be anywhere between
daily and half-yearly.
Testing must be carried out by an expert on the operator‘s own
staff, by a specialist rm or by Sewerin itself.
The function test should be carried out at the same time.
Upkeep - maintenance and repair
The instrument must be maintained at least once a year by
SEWERIN Service, a specialist rm authorised by SEWERIN or
a SEWERIN-authorised expert.
Certicates must be issued accordingly.
The test disc on the instrument
conrms when maintenance was
last carried out and indicates the
next scheduled date.
26
Annual maintenance and repair must cover at least the specialist
care and adjustment of the instrument and the replacement of
components with a limited useful life.
Note:
Where instruments have explosion protection the
applicable regulations must be be observed!
Note:
Technicians responsible for upkeep must have been
trained and instructed by Sewerin!
Page 36
5.2 Test set
The EX-TEC GM 4 and EX-TEC OD 4 each have their own test
sets.
5.2.1 EX-TEC GM 4
The test set SPE HG and SPE-Y check display accuracy:
C
B
E
5 Testing the instrument
D
A
Fig.: Test set SPE HG
Item.DescriptionFunction
Ainstrument hose
connection with:
z probe connectionz test heads
Btest gas connectionconnection for:
z test gas cansz pressure-hose adapter (in
conjunction with pressure
reducer
Cpressure indicationDisplay of actual pressure inside
the test gas can
Drelease keyrelease of test gas
Econnection hoseconnection with:
z instrument hose
27
Page 37
5 Testing the instrument
5.2.2 EX-TEC OD 4
The test set SPE OD checks display accuracy:
ED
C
A
B
Fig.: Test set SPE OD
Item.DescriptionFunction
Ainstrument hose
connection with:
z probe connectionz test heads
Btest gas connectionconnection for:
z test gas cansz pressure-hose adapter (in
conjunction with pressure
cylinder and pressure reducer
reducer
Cpressure indicationDisplay of actual pressure inside
the test gas can
Drelease keyrelease of test gas
Econnection hoseconnection with:
z instrument hose
28
Page 38
5.3 Test gases
The following test gases are used in conjunction with the test set
SPE HG/SPE OD to check the zero point and display accuracy
in the factory. Test gas concentrations other than those listed
here can also be used - see section 7.3: Setting the test gas
concentration.
Gastest gases
THT
-zero point
-sensitivity
TBM
-zero point
-sensitivity
Carbon monoxide CO
-zero point
-sensitivity
Hydrogen H2
-zero point
-sensitivity
Hydrogen sulphide H2S
-zero point
-sensitivity
Ammonia NH3
-zero point
-sensitivity
Oxygen O2
-zero point
-sensitivity
Hydrogen chloride HCI
-zero point
-sensitivity
5 Testing the instrument
fresh air
e. g. 15 mg/m3 in synthetic air
fresh air
z. B. 10 mg/m3 in synthetic air
fresh airt
z. B. 30 ppm in synthetic air
fresh air
1 % vol. in synthetic air
(0 – 100 ppm and 0 – 2000 ppm)
fresh airt
40 ppm in N
2
fresh air
50 ppm in N
100 % vol. CH4 or CO
2
2
fresh air
fresh air
5 ppm in synthetic air
29
Page 39
5 Testing the instrument
5.4 Testing the pump, zero point and display accuracy
Pump (pump instruments only)
The power of the instrument‘s integral pump is 5 – 15 l/h. You
can function-test it by carrying out a simple impermeability test
as follows:
z swith the instrument on in fresh air
z place the test head on the EX-TEC OD 4/GM 4 and cover its
nipple with a nger
if the pump is working properly, error message F100 (cf. section
11.3: Error messages) appears after about 10 seconds
This error message is triggered if the volume ow falls below
half its nominal level.
The error message can be cleared by holding down any key.
z we recommend, replacing the pump lter of pump instruments
on a regular basis (cf. section 11.4: Consumables).
Undo and remove the sensor cap and extract the sensor from
its holder. Now remove the pump lter (the white disc 4 mm in
diameter) and replace it with a new one. The sensor (with rubber
seal) and sensor cap can now be replaced. Do not overtighten
the sensor cap screws.
30
For further testing proceed as follows:
z screw the selected test gas can by hand onto the test set you
are using as far as it will go
z EX-TEC GM 4: connect the test head HG4 to the hose of the
SPE HG or SPE-Y
z EX-TEC OD 4: connect the test and probe head OD4 (plug
connection) to the hose of the SPE OD
z enter your results in the test record (cf. appendix)
Page 40
5 Testing the instrument
Zero point
wait for the instrument to warm up and reach a stable zero
point:
You can activate an inspection block to make sure your EX-TEC
OD 4/GM 4 is regularly checked. This block does not become
active until the next inspection date has passed (cf. section 8.3:
Setting the inspection interval)..
After that the instrument cannot be operated until the inspection
has been carriedoutandconrmed (cf. section 7.4: Inspection
conrmation):
z briey pressing the on/off key
brings you to the inspectionblock routine
z the last setting is displayed,
e.g.:
No = block inactive
Yes = block active
z you can set the desired con-
dition by pressing one of the
cursor keys and conrming
with the on/off key
46
Page 56
PIN001
0
PIN code
8.5 Setting the PIN
You can set up your EX-TEC OD 4/GM 4 so that only authorised
persons, e.g. instrument technicians and experts, have access
to the information menu with all its subfunctions. This involves
setting a PIN that must be entered every time the information
menu is called.
When an incorrect PIN is entered the instrument reverts to its
switch-on routine:
PIN = 0000
the function is inactive, every user has access to the Info menu
PIN = 0001 – 9999
the function is active, only persons who know the set PIN have
access to the Info menu
8 System menu
z briey pressing the on/off key
brings you to the PIN-setting
routine
z the last PIN to be set (0001 =
factory setting) appears in the
LC display
z we recommend you to use a
different PIN
z by pressing or holding down
one of the cursor keys and
and conrming with the on/off
key you can set each of the 4
digits from left to right to the
desired PIN
Note:
Make a note of your PIN and only give it to authorised
persons!
If you forget your PIN, please contact SEWERIN
service!
47
Page 57
8 System menu
AL1 alarm
PPM4000
0
AL2 alarm
PPM8000
0
8.6 Setting the alarm thresholds
Alarm thresholds other than those pre-set by SEWERIN can be
set as follows.
AL1-alarm
AL2-alarm
z briey pressing the on/off key
brings you to the AL1 alarm
threshold routine
z the current alarm threshold is
displayed, e.g. the factory setting for H2: 4000 PPM
z press one of the cursor keys
to set the desired threshold
and the on/off key to conrm
z briey pressing the on/off key
brings you to the AL2 alarm
threshold routine
z the current alarm threshold is
displayed, e.g. the factory setting for H2: 8000 PPM
48
z press one of the cursor keys
to set the desired threshold
and the on/off key to conrm
Page 58
8 System menu
The AL1 threshold is always lower than the AL2 threshold. The
adjustment ranges of the individual alarm thresholds (factory
settings in bold) are as follows:
GasAL1 thresholdAL2 threshold
THT/TBM
-adjustment range
-increment
CO
-adjustment range
-increment
H
2
-adjustment range
-increment
H2S (0 – 100 ppm)
-adjustment range
-increment
NH
3
-adjustment range
-increment
O
2
-adjustment range
-increment
HCI
-adjustment range
-increment
0.1 – 20.0 mg/m
0.1 mg/m
3
30 ppm
5 – 199 ppm
1 ppm
0.4 % vol.
40 – 9.980 ppm
1 ppm
10 ppm
5 – 99 ppm
1 ppm
20 ppm
1 – 99 ppm
1 ppm
18.0 % vol.
15.0 – 24.9 % vol.
0.1% vol.
5 ppm
9.5 ppm
0.5 ppm
3
20.1 – 100.0 mg/m
0,1 mg/m
60 ppm
6 – 200 ppm
1 ppm
0.8 % vol.
60 – 10.000 ppm
1 ppm
20 ppm
6 – 100 ppm
1 ppm
150 ppm
2 – 100 ppm
1 ppm
23.0 % vol.
15.1 – 25.0 % vol.
0,1 % vol.
10 ppm
1.0 – 10 ppm
0.5 ppm
3
3
49
Page 59
8 System menu
unit VOL%
%GAZ
%Vol
VOL%
% vol. unit
z briey pressing the on/off key
brings you to the volume-unit
setting routine
z he current setting is displayed,
e.g.:
VOL% - display in VOL% (D/GB)
%VOL - display in VOL% (I)
%GAZ - display in %GAZ (F)
z press one of the cursor keys
to set the required unit and the
on/off key to conrm
50
Page 60
8.7 Setting the user name
You can also enter your name, your department or other personal
information in the EX-TEC OD 4/GM 4. These will then be dis-
played every time the instrument is switched on to avoid doubt
about who it is assigned to.
Three lines of 16 characters each are available for entries, e.g.:
Gassupply- line 1
Anytown- line 2
Otto Mustermann - line 3
User name
The user name is entered with buttons and . All existing
signs must be conrmed.
There are two possibilities of entering:
1. possibility: new entry, no existing sign to be overwritten.
Entry starts with a blanc „ „.
Letters A - Z can be chosen in alphabetical order with button
.
After reaching letter Z the menu starts again with letter A.
Letters Z - A can be chosen downwards with button .
After letter A the additional characters
@>=<;:9876543210/.-,+*)(‚&%$#“ are displayed!
8 System menu
Note:
The additional characters can only be chosen with
button .
The chosen letter is conrmed with button .
The instrument automatically goes to the next letter.
After conrming the last letter of the user name, the instrument
returns to the system menu.
51
Page 61
8 System menu
2. possibility: an existing letter has to be overwritten.
When switching to the letter, the already existing letter is displayed.
The blanc appears when pressing button , thereafter letters
A - Z can be chosen as described on the previous page.
After reaching letter Z the menu starts again with letter A.
When pressing button the previous letter of the alphabet ap-
pears, all others appear in descending order.
After letter A the additional characters @>=<;:9876543210/.-,+*)
(‚&%$#“ are displayed!
The chosen letter is conrmed with the button.
The instrument automatically goes to the next letter.
After conrming the last letter of the user name, the instrument
returns to the system menu.
Note:
It could occur that during entry blancs a displayed
by a black block. This only happens in entry mode.
The actual display shows the correct blancs.
52
Page 62
language
Englisch
Fran ais
Italiano
ç
Deutsch
system
exit
AL1 alarm
AL2 alarm
unitVOL%
user name
language
8.8 Setting the language
You can set the language of the individual menu items:
8.9 Leaving the system menu
8 System menu
z briey pressing the on/off key
brings you to the languagesetting routine
z the available settings are
displayed
z press one of the cursor keys
to set the required language
and the on/off key to conrm
z this function switches back to
the Info menu
z press the on/off key to con-
rm
53
Page 63
9 Hardware menu
hardwarebatteryaccu NiMH
hardware
9 Hardware menu
9.1 Menu structure
9.1.1 EX-TEC GM 4
Alkaline
accu capacitymAh 1100...2300
back lightsec. 010
contrast55%
autostartwarning
rangePPM
type of gasCO 300 PPM
sensor install.12.06.2002
pumpyes
LCD test
resetno
exit
measuring
mg/m3
10000 PPM
H
2
S 100 PPM
H
2
S 2000 PPM
H
2
100 PPM
NH
3
25.0 Vol%
O
2
no
yes
54
Page 64
9.1.2 EX-TEC OD 4
hardwarebatteryaccu NiMH
hardware
9 Hardware menu
Alkaline
accu capacitymAh 1100...2300
back lightsec. 010
contrast55%
autostartwarning
rangePPM
type of gasTHT 0...100 mg/m3
sensor install.12.06.2002
pumpyes
LCD test
resetno
exit
measuring
mg/m3
TBM 0...100 mg/m3
no
yes
55
Page 65
9 Hardware menu
0..10000 PPM
battery
Accu NiMH
Alcaline
NullpunktNullpunkt
Accu capacity
mAh1600
1
9.2 Setting the battery type
The EX-TEC OD 4/GM 4 can be operated with both rechargeable
and non-rechargeable batteries (primary cells).
For the correct display of operating hours it is important for the
current battery type to be properly set:
If you have selected NiMH as the battery type, you can enter the
capacity (see the information on the battery itself):
z the last battery type set is dis-
played
z press one of the cursor keys
to select between settings:
NiMH batteries
(rechargeable)
alkaline batteries
(non-rechargeable)
z press the on/off key to save
the battery type set
z the last value set is displayed
56
z press the cursor keys to
change the value
(range: 1100 – 2300 mAh)
z press the on/off key to save
the set value
Page 66
contrast
22 %
back light
sec.10
0
9.3 Setting the illumination time and contrast
During measuring operation the LCD illumination can be switched
on by pressing any key (cf. section 3.4: Illumination and operating-hours display).
The period for which the illumination stays on can be set individually:
z the last period set is dis-
played
z press the cursor keys to
change the value in seconds
(range: 10 – 240 seconds)
z press the on/off key to save
the set value
The contrast of the LCD can be set individually:
z the last contrast set is dis-
played
9 Hardware menu
z press one of the cursor keys
to change the value in percent
(range: 0 – 100 %); 55 % is a
good working value
z press the on/off key to save
the set value
57
Page 67
9 Hardware menu
autostart
warning
measuring
9.4 Autostart
The EX-TEC OD 4/GM 4 enables you to set the default operating
mode after switch-on. This function is preserved even when the
instrument is switched off.
z briey pressing the on/off key
brings you to the Autostart
routine
z the last set operating mode
is highlighted, e.g. the factory
setting „Measurement“
z press one of the cursor keys
to select one of the Autostart
functions
z briey press the on/off key to
conrm the selected Autostart
function
58
Page 68
range
PPM
mg/m3
9.5 Range
The EX-TEC OD 4/GM 4 enables you to dene the measurement
units for some gases.
With THT/TBM sensors, for example, you can switch between a
display in ppm and mg/m3.
9.6 Changing the sensor
The sensor in the EX-TEC GM 4 can be changed. The following
sensors are available: CO, H2, H2S (100 ppm), H2S (2000 ppm),
NH3, O2 and HCl.
Sensors not in use have to be stored in a special storage box.
9 Hardware menu
z briey pressing the on/off key
brings you to the Autostart
routine
z the last set unit is highlighted
z press one of the cursor keys
to select one of the measurement units
Note:
The sensor in the EX-TEC OD 4 cannot be changed,
and if an EX-TEC GM 4 is tted with an HCl sensor
that cannot be changed either.
When you have changed the sensor you must enter the gas type
and then the date on which the sensor was tted.
CAUTION!
Whenever you change a sensor, you must readjust
the instrument following a suitable test routine (see
section 7).
59
Page 69
9 Hardware menu
sensor install.
12.06.2002
type of gas
CO300 PPM
H2 10000 PPM
9.6.1 Gas
9.6.2 Sensor installation
The sensor installation date can be entered.
z briey pressing the on/off key
brings you to the gas-setting
routine
z the possible sensors are dis-
played
z press one of the cursor keys to
select one of the sensors
z briefly pressing the on/off
key brings you to the sensor
installation-date routine
z the date when the sensor was
installed is displayed
60
z press one of the cursor keys
to enter the date and the on/
off key to conrm
Page 70
pump
No
Yes
9.7 Setting pump operation
This function only applies to pump instruments. Pump instruments
have the letter P on the identication sticker on the back.
If the pump of an EX-TEC OD 4/GM 4 (pump) fails, you can switch
it off and continue to use it as a diffusion instrument (though the
probes cannot be used):
9.8 Carrying out an LCD test
The LCD of your EX-TEC OD 4/GM 4 has a matrix display. This
test checks whether it is fully functional.
The test pattern that it generates enables the failure of individual
pixels to be rapidly detected.
9 Hardware menu
z the last set condition is dis-
played
z press one of the cursor keys
to select one of the condi-
tions:
Yes (pump active)
No (pump inactive)
z press the on/off key to save
the set condition
z briey pressing the on/off key
starts the LCD test
61
Page 71
9 Hardware menu
9.9 Restoring the factory settings
If all instrument settings (e.g. gas, sensor type, pump condition,
contrast, illumination time etc.) have become totally disordered,
the EX-TEC OD 4/GM 4 can be restored to a dened initial state
(factory setting):
CAUTION!
„After the instrument has been restored to its original
settings, the adjustment must be checked!“
z the date of the last restora-
tion to factory settings is displayed
z press one of the cursor keys
to select one of the options:
Noleave settings as they
are
Yesrestore factory settings
z press the on/off key to conrm
the selected option
62
Page 72
9.10 Leaving the hardware menu
This function switches back to the Info menu:
9 Hardware menu
z press the on/off key to con-
rm
63
Page 73
10 Memory menu
memoryclear
memory
10 Memory menu
10.1 Menu structure
10.2 Clearing memory
interval
exit
z this function clears all saved
readings
z press one of the cursor keys
to select one of the options:
Noleave the memory con-
tent as it is
Yesclear all saved read-
ings
64
z press the on/off key to conrm
the selected option.
Page 74
interval
5 sec.
10.3 Setting the memory interval
Data-memory capacity
Intervalcapacity
1 styp.7 h
2 styp. 15 h
5 styp. 39 h
10 styp. 78 h
20 s typ. 156 h
30 styp. 234 h
60 styp. 470 h
10 Memory menu
z you can set the frequency at
which readings are saved,
e.g.:
5 sec (factory setting)
readings are saved every 5
seconds
z press one of the cursor keys
to change the value
z press the on/off key to save
the new value
10.4 Leaving the memory menu
This function switches back to the Info menu:
z press the on/off key to con-
rm
65
Page 75
11 Technical aspects
11 Technical aspects
11.1 Technical notes
Identicationsticker
The identication sticker on the back of the instrument includes
a pictogram of a crossed-out screwdriver. This means that the
battery compartment may not be opened in an area exposed to
the danger of explosion.
Notes
If the EX-TEX OD 4 is not left in the docking station after
charging, it will have to be fully recharged again after 4 weeks
at the most.
If the instrument will not switch on because of lack of power, charge it up and recalibrate it after a 3-day recovery period.
66
Accessories
The instrument accessories fall into three main groups: charging
equipment, test equipment and sampling hoses (see section 12.2
„Accessories“).
Cleaning
The instrument should only be cleaned with a damp cloth: use
no solvents, benzene, silcon-containing Cockpit-Spray or similar
substances.
Page 76
11 Technical aspects
Static charge
Electrostatic charges should generally be avoided. Electrostatically oating objects (like metallic housings with no earth connection,
for example) are unprotected against charges transferred from
dust, aerosols and the like.
Battery change
After a change of batteries the EX-TEC OD 4 and the EX-TEC GM 4
with an HCl sensor require a minimum warm-up period of 15 mi-
nutes for a stable zero point to be reached.
67
Page 77
11 Technical aspects
11.2 Technical data
Instrument data
- Dimensions (W x H x D)60 x 144 x 35 mm
- Weight:about 300 g
EX-TEC GM 4 (Serial number)
- diffusion instrument:062 00 (type - version - number)
- pump instrument:062 01 (type - version - number)
EX-TEC OD 4 (Serial number THT)
- diffusion instrument:062 02 (type - version - number)
- pump instrument:062 03 (type - version - number)
EX-TEC OD 4 (Serial number TBM)
- diffusion instrument:062 04 (type - version - number)
- pump instrument:062 05 ((type - version - number)
Explosion protection (CENELEC)
- Testing institution:TÜV NORD CERT GmbH, Hannover
- Test number:TÜV 01 ATEX 1657
- Identication 1:
- Identication 2:
II2G Ex e ib IIB T4 Gb
basic instrument without leather case
for all gases except hydrogen H
2
II2G Ex e ib IIC T4 Gb
basic instrument with leather case for
all gases including hydrogen H
2
68
Pump power
- negative pressure:> 150 mbar
- volume ow:typically 5 – 15 l/h
Page 78
11 Technical aspects
Power supply
- Batteries:3 rechargeable NiMH batteries or
3 alkaline primary cells
(cf. section 4.3: Battery operation)
- Charging voltage:12 V=
- Charging current:360 mA (pulsed)
- Charging period:about 12 hours (depending on battery
capacity)
Measurement
principle:
electrochemical measurement cell,
temperature compensation,
zero-point correction
Sensor life:see sensor data!
Operating time:20 h (pump operation)
50 h (diffusion operation)
alarms and illumination reduce the operating time
Connection pressure
(EX-TEC OD 4)
20 mbar - 2 bar
(through different hoses: see page 12)
Operating and
storage temperature:see sensor data!
An error message is displayed in case of any failure during operation. The error No. and error description will be indicated.
11 Technical aspects
Error codeLCD (error
description)
F09no adjustment: no adjustment data
F10 adjustment
error:
F13adjustment
error:
F51 – F54system error component errorwitch instrument off
F 59voltage supply voltage outside the
F100pump powerpump power too lowcheck lters in the in-
CauseRemedy
present
zero point in the ppm
range
sensitivity in the ppm
range
permitted range
carry out an adjustment
check test gas or repeat adjustment
check test gas or repeat adjustment
and on again or con-
tact SEWERIN service
contact Sewerin service only
strument and probes.
Please observe the
instructions in section
9.9: Adjusting pump
operation
Note:
If any other error codes appear, please contact
SEWERIN service!
Cause:
Remedy:
F09 – F15 temporary errors
F40, F100clearable errors
F51 – 54, F59 – F62 system errors; switch instrument off
Programm defect
SEWERIN service
79
Page 89
11 Technical aspects
11.4 Consumables
Pumplterbuilt into the instrument to protect the pump
H2Sltreinside
the senor head
(EX-TEC OD 4)
Test gas cansvarious concentrations for monitoring and
Note:
Test gas cans are under pressure: do not store
above 50 °C!
(to replace it, the sensor cap must be removed with the screwdriver supplied with
the instrument and the sensor removed)
Change in SEWERIN service
adjustment purposes
80
Page 90
11.5 Hints on Disposal
The disposal of instruments and accessories is governed by the
European Waste Catalogue (EWC).
Type of WasteCorresponding EWC Code
Instrument16 02 13
Test gas can16 05 05
Battery, accu16 06 05
Old Instruments
Old instruments can be returned to Hermann Sewerin GmbH.
We will arrange the qualied disposal free of charge through
certied specialists.
11 Technical aspects
81
Page 91
12 Delivery variants and accessories
12 Delivery variants and accessories
12.1 Delivery variants
EX-TEC OD 4
(diffusion device)
THT:
Part no.: OD04-10001
Serial number:
062 02 ....
(Type - Model - Number)
TBM:
Part no.: OD04-11001
Serial number:
062 04 ....
(Type - Model - Number)
EX-TEC OD 4
(pump-device)
THT:
Part no.:OD04-10101
Serial number:
062 03 ....
(Type - Model - Number)
82
TBM:
Part no.: OD04-11101
Serial number:
062 05 ....
(Type - Model - Number)
z All devices are equipped with an integral memory to store
measuring data
z Suitable for connecting accessories
Page 92
12 Delivery variants and accessories
EX-TEC GM 4
(diffusion device)
Part no.: GM04-10001
Serial number:
062 00 ....
(Type - Model - Number)
EX-TEC GM 4
(pump device)
Part no.: GM04-10101
Serial number:
062 01 ....
(Type - Model - Number)
z All devices are equipped with an integral memory to store
measuring data
z Suitable for connecting accessoriesz For available sensors please note chapter 11.2
83
Page 93
12 Delivery variants and accessories
12.2 Accessories
Docking station HG4
Part-No.: LP10-10001
Docking station HG4 with interface
Part-No.: LP10-10101
z to hold the instrument
z with a connection socket for
the AC/DC adapter or a vehicle cable
z with apertures for mounting in
the vehicle or elsewhere
AC/DC adapter M4
Part-No.: LD10-10001
z to connect the docking stati-
on HG4 to a 100 – 240-volt~
mains supply
84
Vehicle cable M4 12 V= mounting
Part-No.: ZL07-10000
z to connect the docking stati-
on HG4 to a 12-volt= vehicle
supply
z with built-in fuse and blade
receptacles
Vehicle cable M4 12 V= mobile
Part-No.: ZL07-10100
z to connect the docking stati-
on HG4 to a 12-volt= vehicle
supply
z with built-in fuse and connec-
tor for cigarette lighter
Page 94
12 Delivery variants and accessories
Vehicle cable M4 24 V= mounting
Part-No.: ZL09-10000
z to connect the docking stati-
on HG4 to a 24-volt= vehicle
supply
z with voltage transformer and
blade receptacles for a fix
connection
Carrying bag HG4
Part-No.: 3204-0034
z leather case with viewing win-
dow and eyelets for the strap
z tab with press-stud to secure
the instrument
z clip to wear device at the belt
z can be used in areas exposed
to the danger of explosion
(conducting front foil)
z must be used in areas where
there is hydrogen and with the
H2-sensor
Carrying strap
Part-No.: 3209-0003
z Leather strap to carry the
instrument
z adjustable between 0.5 – 1.0 m
Storage box for electrochemical sensors
Part-No.: GM04-Z0100
z designed for the safe storage
of electrochemical sensors
85
Page 95
12 Delivery variants and accessories
EX-TEC GM 4 probe equipment
Diffusion device
Pump device
Flexible hand probe HG4 with
ball
Part-No.: ZS42-10000
z for two-handed operation in re-
latively inaccessible locations:
probe tip with suction globe,
ne dust lter, probe hose and
probe head with attachment
screws
Flex Probe HG4
Part-No.: ZS40-10200
z for one-handed detection and
measurement in relatively in-
accessible locations: exible
swan-neck, lter insert in the
head and adapter head with
attachment screws
Flexible hand probe HG4
Part-No.: ZS40-10100
z for two-handed detection and
measurement in relatively in-
accessible locations: hand grip
with a exible probe tip, lter
insert in the head, probe hose
and probe head with attachment screws
86
Note:
The probes for the pump instrument cannot be used
in diffusion operation.
Probes for other gases are available on request.
Page 96
Adapter EX-TEC OD 4
12 Delivery variants and accessories
Test and probe head OD4
Part-No.: PP01-B1300
z the EX-TEC OD 4 is operated
in conjunction with the test and
probe head, which has a plug
connection for the sampling
hoses
Adapter hoses
z Please note that different ho-
ses must be used for different
pressure ranges
A detailed listing can be found
on page 12
Mini measuring adapter
Part-No.: ZZ20-10000
z for connecting adapter hose
and gas pipe
87
Page 97
12 Delivery variants and accessories
Prüftechnik
EX-TEC OD 4
EX-TEC GM 4
Test set SPE OD
Part-No.: PP01-50100
z for mobile and in-vehicle usez with a connection for SEWERIN
test gas cans, ow-rate control,
release key and connecting
hose, in conjunction with the
Test set and probe head OD4
Test set SPE HG
Part-No.: PP01-10201
z for mobile and in-vehicle usez with a connection for SEWERIN
test gas cans, ow-rate control,
release key and connecting
hose, in conjunction with the
probe head HG4
88
Test set SPE-Y
Part-No.: PP01-20001
z for mobile and in-vehicle use
z with a connection for
SEWERIN test gas cans,
ow-rate control, release key
and connecting hose, for the
probe head HG4
Page 98
12 Delivery variants and accessories
Test gas cans
z for testing display accuracy
and adjustment
z various test gas concentra-
tions in 1-litre cans under
pressure of approx. 12 bar
Pressure cylinders
z for testing and adjusting dis-
play sensitivity
z various test gas concentra-
tions in 0.4/2.0/10.0-litre steel
cylinders under pressure of
100 – 150 bar
Case HG4
Part-No.: ZD18-10000
z to hold:
- instrument and charging
equipment
- probe equipment
- Test gas can and test set
SPE-Y
z delivery does not include con-
tents shown in the picture
89
Page 99
Appendix
13 Appendix
13.1 EU-declaration of conformity
Hermann Sewerin GmbH hereby declares that the EX-TEC
OD 4 / GM 4 fulls the requirements of the following guidelines:
z 2014/30/EU
z 2014/34/EU
Gütersloh, 2016-04-20
Dr. S. Sewerin
(General Manager)
The complete declaration of conformity can be found online.
®
90
Page 100
13.2 Inspection protocol
Inspection protocolEX-TEC® GM 4 / OD 4
Sensor (e.g. CO):
test gas conc. (e.g. 40 ppm CO):
display tolerance zero point (e.g. -3 – +3 ppm)(Display tolerances according to
display tolerance test gas (e.g. 27 – 33 ppm) chapter 5.4 of operating instruction)
Serial no. (e.g. 062 01 0001):
1.0 Device status
1.1 - status correct (e.g.: Y / N)
1.2 - remaining operating hours (e.g.: 5 h)
2.0 Pump test
2.1 - fault F 100 when sealing
2.2 - pump filter changed (e.g.: Y / N)
3.0
Display accuracy
3.1 - display (zero point)
- within tolerance (e.g.: Y / N)
3.2 - display (test gas)
- within tolerance (e.g.: Y / N)
Appendix
22.07.2008
4.0 AL 1 alarm triggering
4.1 optical alarm (e.g.: Y / N)
4.2 audible alarm (e.g.: Y / N)
5.0 Observations
- housing broken
- adjustment, repair
- factory inspection
- or the like
6.0 Test
- day
- month
- year
- signature
91
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.