Sewerin Ex-tec Combi User Manual

Operating
Instructions
Measurable success by Sewerin equipment
Congratulations. You have chosen a quality instrument manufactured by Hermann Sewerin GmbH. Our equipment will provide you with the highest standards of perfor­mance, safety and efficiency. They correspond with the national and international guide-lines.
Please read and understand the following operating instructions before using the equipment; they will help you to use the instrument quickly and competently. If you have any queries we are available to offer advice and assistance at any time.
Yours
Hermann Sewerin GmbH
Robert-Bosch-Straße 3 33334 Gütersloh, Germany Tel.: +49 5241 934-0 Fax: +49 5241 934-444 www.sewerin.com info@sewerin.com
Sewerin Ltd.
8, Walsworth Road Hitchin Hertfordshire SG4 9SP, UK phone: +44 1462-634363 www.sewerin.co.uk info@sewerin.co.uk
SEWERIN Sarl
17, rue Ampère - BP 211 67727 HOERDT CEDEX, France Tél. : +33 3 88 68 15 15 Fax : +33 3 88 68 11 77 www.sewerin.com sewerin@sewerin.fr
Sewerin USA, LLC
13551 W. 43rd Drive, Unit R Golden, CO 80403-7272 phone: +1 303-424-3611 fax: +1 303-420-0033 www.sewerin.net jerry.palmer@sewerin.net
SEWERIN IBERIA S.L.
c/ Cañada Real de Merinas, 17 Centro de Negocios „Eisenhower“ Edificio 5; Planta 2 - C 28042 Madrid, España Tel.: +34 91 74807-57 Fax: +34 91 74807-58 www.sewerin.com info@sewerin.es

Illustration EX-TEC Combi

EX-TEC® Combi
EX-TEC® Combi
100%
Device variation 1
Device variation 2

Brief operating instructions

EX-TEC Combi: brief operating instructions
Switch on/off
Toggle between GAS WARNING/GAS MEASUR E M ENT op e rat i n g mo d es (Device variation 2)
Device variation 1: Toggle between
GA S WAR NING/GAS MEASURE ­MENT operating modes Switch pump on/off
(in GAS MEASUREMENT mode)
Toggle between automatic/manual gas switching
Change the gas displayed (in manual operation)
Zero-point correction for the gas dis­played
Display gas alarm thresholds AL1 and AL2
Acoustic clearance of the AL1 alarm
Switch LCD illumination on/off (switches off automatically after about 4
Operating Instructions
EX-TEC® Combi
15.06.2005 – V2.X – 103401 – en

For your safety

This product may only be operated by appropriately-trained persons who are familiar with the relevant operating manual.
It may only be used for its designated purpose, i.e. for industrial and com­mercial use.
Repair work may only be carried out by specialists or by persons who have undergone appropriate training.
Any alterations or modifications to the product require the prior approval of Hermann Sewerin GmbH. In the event of unauthorised alterations to the product the manufacturer accepts no liability for damage.
Only Hermann Sewerin GmbH accessories may be used with the product.
Only spare parts approved by us may be used for repairs.
Hermann Sewerin GmbH accepts no liability for damage resulting from non-compliance with the foregoing. The guarantee and liability provisions in the Hermann Sewerin GmbH terms of sale and supply are not extended by the foregoing.
We reserve the right to make changes in the context of continued techni­cal development.
In addition to these instructions, please comply with generally applicable safety and accident-prevention regulations!
Symbols used:
()
CAUTION!
This symbol warns of dangers that may threaten the safety of the user or maty damage or destroy the product.
Note:
This symbol flags information and hints extending beyond the actual operation of the product .
This symbol will hereinafter identify all operating modes and measuring techniques not tested by DMT!

Contents Page

1 EX-TEC Combi system ...........................................................1
1.1 Excess-gas warning and measuring instrument .......................1
1.2 Fields of application ..................................................................2
1.3 Test certificates .........................................................................4
1.4 Carrying equipment ...................................................................5
2 Safety .......................................................................................6
2.1 Safety notes .............................................................................. 6
3 Measuring operation ...............................................................8
3.1 Instrument description ...............................................................8
3.2 Switching on ..............................................................................9
3.3 Modes .....................................................................................13
3.4 Switching gases ...................................................................... 15
3.5 Alarms .....................................................................................16
3.6 Instantaneous value alarms (AL1, AL2, AL3) ..........................17
3.7 Short-term value alarm (KZW), long-term value alarm (LZW),
minimum and maximum values (MIN/MAX) ............................23
3.8 Zero point adjustment .............................................................26
3.9 Illumination and contrast ......................................................... 26
3.10 Operating hours display and battery alarm ............................. 27
3.11 Switching off ............................................................................28
4 Charging ................................................................................29
4.1 Charging and charge maintenance .........................................29
4.2 Self-discharge ......................................................................... 30
5 Testing/maintenance .............................................................31
5.1 Function testing, testing display accuracy, maintenance ........32
5.2 Test set ....................................................................................34
5.3 Test gases ...............................................................................35
5.4 Testing the pump power, zero point and sensitivity .................37
I
Contents Page
6 Info menu ...............................................................................39
6.1 Menu structure ........................................................................ 39
6.2 Overview ................................................................................. 40
7 Adjustment menu ..................................................................43
7.1 Menu structure ........................................................................ 43
7.2 Setting the H2S sensor ............................................................44
7.3 Setting the CO sensor .............................................................45
7.4 Setting the O2 sensor ..............................................................46
7.5 Setting the CO2 sensor ............................................................47
7.6 Setting the CH4 sensor ............................................................48
7.7 Setting the measurement quantity ..........................................50
7.8 Setting the test gas concentration ...........................................51
7.9 Inspection confirmation ........................................................... 54
7.10 Leaving the adjustment menu .................................................55
8 Memory menu ........................................................................56
8.1 Menu structure ........................................................................ 56
8.2 Clearing memory .....................................................................57
8.3 Setting the memory interval ....................................................58
8.4 Setting the memory mode .......................................................60
8.5 Setting the short-term value interval .......................................61
8.6 Leaving the memory menu ......................................................62
9 System menu .........................................................................63
9.1 Menu structure ........................................................................ 63
9.2 Setting the date/time ............................................................... 64
9.3 Setting the inspection interval ................................................. 66
9.4 Setting the inspection block ....................................................69
9.5 Setting the 100 vol.% range ....................................................70
9.6 Setting the PIN code ............................................................... 71
9.7 Setting the alarm thresholds ..................................................73
9.8 Checking the LCD ...................................................................77
9.9 Leaving the system menu ....................................................... 77
II
Contents Page
10 Application hints ...................................................................78
10.1 Application hints from report of suitability examination
PFG-No. 41300401P ...............................................................78
11 Technical aspects ................................................................. 80
11.1 Technical notices ..................................................................... 80
11.2 Technical data .........................................................................82
11.3 Error messages .......................................................................93
11.4 Error and Alarm properties ......................................................97
11.5 Wearing Parts .........................................................................98
11.6 Spare parts ..............................................................................98
11.7 EC-Sensor Disposal ................................................................98
12 Hints on Disposal ..................................................................99
13 Delivery variants and accessories ....................................100
13.1 Delivery variants ....................................................................100
13.2 Accessories ...........................................................................101
Appendix ............................................................................................108
EC-type-examination certificates .........................................................108
Declaration of Conformity .................................................................... 118
Inspection protocols ............................................................................ 119
III
1

1 EX-TEC Combi system

1 EX-TEC Combi system

1.1 Excess-gas warning and measuring instrument

The EX-TEC Combi is a com­bined warning and measuring instrument for a number of diffe­rent gases. It consists of:
the basic instrument, incorpo­ra ting a pum p and a dat a memory for documentation purposes
4 sensor sockets for the mea­surement of up to 5 different gases
The following sensors are availa­ble:
methane CH
4
or
propane C3H8
or nonane C9H20 (*)
carbon dioxide CO
2
oxygen O
2
hydrogen sulphide H2S
carbon monoxide CO
combined H2S/CO sensor
(*) This operating manual only describes methane mea-
surement!
2
1 EX-TEC Combi system

1.2 Fields of application

The EX-TEC Combi is suitable for use in the following fields:
WARNING mode
Workplace (atmospheric) monitoring in shafts and chambers in these fields:
drinking-water supply (metering
and transfer shafts)
district-heating systems
telecommunications shafts
effluent-treatment systems
(sewage works, pump sumps, digestion-tank areas, rain­overflow basins)
traffic areas
accessible culverts
Warning of explosive*) mixtures
- due to leaking gas pipes near the shaft
- due to the proximity of oil, coal, natural-gas or LPG storage facilities
- due to the proximity of landfill sites, marshland, chemical works, filling stations or refineries
- due to cleaning or coating work with substances containing solvents
- due to the prohibited introduction of combustible substances into the canal network (e.g. petrol leaks)
*) the EC prototype test certificate DMT01ATEX G 002 includes measurement functions
of the LEL methane/propane range
Warning of an oxygen defect /oxygen excess
- due to an increase in other gas components
- due to the decomposition of organic waste in shafts (e.g. wet leaves)
1 EX-TEC Combi system
- due to welding and heating processes with naked flames
- due to air depletion
Warning of toxic gases
- due to the formation of carbon dioxide by bacterial conversion processes
- due to the formation of carbon dioxide in areas where there is mineral water
- due to the formation of carbon dioxide in exhaled air
- due to the formation of hydrogen sulphide in effluent
- due to the formation of carbon monoxide from incomplete combustion (e.g. poorly-adjusted gas-heating systems)
- due to the formation of carbon monoxide near highways or car-parks
GAS-MEASURING mode *) Determining gas concentrations
Localisation
- measuring gas concentra­tion in probe holes
- determining which probe hole has the maximum con­centration
Gas injection
- monitoring filling with natural gas and oxygen extraction
- measuring the methane and oxygen concentration
Inertisation
- monito r ing flus h ing with nitrogen and natural-gas extraction
- measuring the methane and oxygen concentration
*) No measuring function for explosion protection in accordance with guideline 94/4/EG.
3
4
1 EX-TEC Combi system
1.3 Test certificates
Passive explosion protection
The EX-TEC Combi has been tested for explosion protection in accordance with the European norm (CENELEC):
EC prototype test certificate:
Identification: II 2 G EEx ib d IIB T3
Testing institution: Physikalisch-Technische Bundesan-
Active explosion protection (measuring function)
The EX-TEC Combi has also been undergone a measurement­function test in gas-warning mode:
EG prototype test certificate:
Test report: PFG n° 41300401
Testing institution: Deutsche Montan-Technologie GmbH
PTB 96 ATEX 2166, supplements 1 and 3
stalt, Braunschweig
DMT 01 ATEX G002, supplement 1 (measurement range 0 – 100 %LEL methane/propane)
measurement range 0 – 25 vol.%
oxygen, measurement of oxygen defect and excess,
measurement range 0 – 5 vol.% of
carbon dioxide,
measurement range 0 – 500 ppm of
carbon monoxide
measurement range 0 – 100 ppm of
hydrogen sulphide
(DMT), Essen
The test certificates can be found in the appendix.
front view side view back view
front view side view back view

1.4 Carrying equipment

TRIANGEL carrying system
a quick and easy way of carrying the instrument, consisting of a carrying strap and neck-pad
1 EX-TEC Combi system
CROSS-STRAP carrying system
a comfortable way to carry the instrument for longer periods, consisting of 2 carrying straps crossed at the back
5
6

2 Safety

2 Safety

2.1 Safety notes

CAUTION!
Always use original SEWERIN accessories with the
EX-TEC Combi.
CAUTION!
Always use a probe hose with a hydrophobic filter.
CAUTION!
Do not use other than the original SEWERIN fine dust filters with the EX-TEC Combi. Never use ac- tivated carbon filters, as these will collect H2S gas particles, thus causing the warning feature to stop working properly.
CAUTION!
Observe the permissible operating temperature of
-10 °C to +40 °C.
CAUTION!
The EX-TEC Combi may only be recharged in an area not exposed to the danger of explosion.
2 Safety
CAUTION!
Use the test gases only in well-ventilated areas, as some concentrations exceed the pertinent MAK values.
CAUTION!
The EX-TEC Combi satisfies the limits of the EMV regulation. When using it near mobile radio equip­ment please also observe the instructions in their manuals.
7
8

3 Measuring operation

3 Measuring operation

3.1 Instrument description

Note:
Fold out the illustration inside the front cover!
Item description function
1 alarm lamp optical warning on:
2 LCD display of:
3 probe connection connection for:
4 keypad instrument operation
5 attachment for carrying systems:
6 outlet for the gas sample
7 buzzer acoustic warning on:
8 interface serial RS-232-interface for con-
breaching alarm thresholds
display of error messages
gas concentrations
menu items
operating conditions
error messages
probe hose
test set
Triangel
cross-strap
breaching alarm thresholds
display of error messages
nection to a PC
()

3.2 Switching on

3 Measuring operation
always switch the instrument
on in “fresh air”
press the on/off key for about
3 seconds
the optical and acoustic control
signals (items 1 and 7) operate for about 3 seconds
the LCD illumination automati-
cally switches on for about 4 minutes
available operating hours are
displayed in the form of the battery symbol and bars (e.g.: 5 hours = 5 bars)
the built-in pump runs at con-
stant power
the software version number
(e.g. 2.1) and instrument type (Combi) are displayed
Note:
All the following instrument displays assume that the EX-TEC Combi is fully equipped for the mea­surement of 5 gases (CH4 - CO2 - O2 - H2S - CO).
9
10
3 Measuring operation
CH4 - methane
the measurement range for
methane is displayed:
0.00 – 4.40 %VOL
depending on your last setting
a display in the %UEG, %LEL, %LIE, %VOL, %GAZ quanti-
ties may also be possible (cf. section 7.6: Adjusting the CH4 sensor)
CO2 - carbon dioxide
the measurement range for
carbon dioxide is displayed:
0.00 – 5.00 %VOL
depending on your last setting
a display in the %VOL, %GAZ quantities may be possible (cf. section 7.7: Setting the measurement quantity)
O2 - oxygen
the measurement range for
oxygen is displayed:
0.0 – 25.0 %VOL
depending on your last setting
a display in the %VOL, %GAZ quantities may be possible (cf. section 7.7: Setting the measurement quantity)
display of the sensor lifetime
in the form of bars (cf. section
11.2: Technical data):
8 bars = 100 %
3 Measuring operation
H2S - hydrogen sulphide
the measurement range for
hydrogen sulphide is displayed:
0 – 100 PPM
display of the sensor lifetime
in the form of bars (cf. section
11.2: Technical data):
8 bars = 100 %
CO - carbon monoxide
the measurement range for
carbon monoxide is displayed:
0 – 500 PPM
display of the sensor lifetime
in the form of bars (cf. section
11.2: Technical data):
8 bars = 100 %
Time/date
the current time (e.g.17:49)
and date (e.g. 24.02.2000) are displayed
properly-set values are impor-
tant for the documentation of your readings
you can correct any variances
(cf. section 9.2: Setting the date/time)
11
12
3 Measuring operation
Next scheduled inspection (display optional)
if you have set an inspection
interval, the next scheduled inspection (e.g. 17.04.2000) is displayed for about 3 seconds (cf. section 9.3: Setting the inspection interval)
depending on the date and
set schedule, the interval or continuous alarm may also be triggered (items 1 and 7)
clearing the alarm with the
buzzer key or waiting for 15 seconds switches to measuring operation
Note:
If the EX-TEC Combi now automatically switches off, the inspection date has passed with the inspec­tion block switched on (cf. section 9.4: Setting the inspection block).
The instrument does not revert to measuring opera­tion until an inspection has been carried out and confirmed.

3.3 Modes

GAS-WARNING mode
Monitoring the atmosphere in shafts and chambers (cf. sec­tion 1.2: Fields of application)
CH4 - CO2 - O2 - H2S - CO
3 Measuring operation
im m edi at ely af t er bei ng
switched on the EX-TEC Combi is always in warning mode
Indication:
all a la r m t hr es h ol ds ar e
activated
Indication:
each avail able gas is dis-
played for about 3 seconds before switching to the next
Sequence:
Every 5 seconds an operating
signal is activated as a marker. It is both, acoustic (item 7) and optical (item 1). The optical signal acts as a control that the instrument is functioning in gas-warning mode.
If the op e r a ting si g n al is switched off, the monitoring of the ga s conc e n t ra t i on is no long er insur ed. The endangered area must be left immediately.
13
14
3 Measuring operation
GAS-MEASURING mode (⊗)
Determining gas concentrations (cf. section 1.2: Fields of application)
pressing the operating-mode
key (instruments variant 2) for about 3 seconds switches between the two modes
on instruments variant 1, the
switching is carried out by pressing the pump-key for about 3 seconds
in this mode all alarm thresholds
and the operating signal are switched off
the pump now runs at maxi-
mum power
NB: (no symbols)
each avail able gas is dis-
played for about 3 seconds before switching to the next
Sequence:
CH4 - CO2 - O2 - H2S - CO
in this mode, depending on
the task at hand, you can also switch the instrument‘s pump on and off by briefly pressing the pump-key
Note:
If you wish to use the EX-TEC Combi purely as a warn- ing device, you can deactivate GAS-MEASURING mode (cf. section 9.5: Setting the 100 vol.% range)!

3.4 Switching gases

3 Measuring operation
repeatedly pressing the hand
key toggles between auto­matic and manual gas-type switching
Automatic switching
each avail able gas is dis-
played for about 3 seconds before switching to the next:
Sequence:
CH4 - CO2 - O2 - H2S - CO
Manual switching
each gas is displayed until a
arrow key is pressed
pressing a arrow key then
switches to the next gas:
Sequence (cursor up):
CH4 - CO2 - O2 - H2S - CO
Note:
In WARNING mode automatic switching always has priority!
If the instrument is in manual switching mode and no key has been pressed for about 10 seconds, it shifts back to automatic switching!
15
16
3 Measuring operation

3.5 Alarms

The „Technical Rules for Hazardous Substances (TRGS 402 and
900)“ require certain limit values to be monitored and adhered to in the air in the workplace.
In WARNING mode only the EX-TEC Combi thus has three different types of alarm: one optical (item 1), one acoustic (item
7) and one that takes the form of the LCD illumination (item 2) switching on:
AL1, AL2, AL3 The instantaneous value alarm appears with
KZW The short-term value alarm appears with toxic
LZW The long-term value alarm appears with
all gases:
CH4 - CO2 - O2 - H2S - CO
when the measured gas concentration exceeds or falls below a fixed threshold value.
gases:
CO2 - H2S - CO
when the gas concentration averaged over an adjustable time period (e.g. 15 minutes) has exceeded an assessed limit value.
toxic gases:
CO2 - H2S - CO
when the gas concentration averaged over an 8-hour working shift has exceeded a fixed
limit value.
3 Measuring operation

3.6 Instantaneous value alarms (AL1, AL2, AL3)

when in gas-warning mode
the EX-TEC Combi is con­stantly in alarm readiness
this is displayed in the LCD
by two symbols: Ex-area and alarm threshold
the alarm is triggered as soon
as a threshold is breached
Display of alarm thresholds
holding down the threshold
value key displays the set alarm thresholds for the per­tinent gas
Methane CH
alarm thresholds (factory set-
4
tings):
AL1 = 0.90 %VOL = 20 %LEL
AL2 = 2.20 %VOL = 50 %LEL
Carbon dioxide CO
alarm thresholds (factory set-
4
tings):
AL1 = 0.50 %VOL (MAK value)
AL2 = 1.00 %VOL
17
18
3 Measuring operation
Oxygen O
alarm thresholds (factory set-
2
tings):
AL1 = 18.0 %VOL
AL2 = 23.0 %VOL
the AL1 alarm threshold is
triggered when the value falls below the threshold
Hydrogen sulphide H2S
alarm thresholds (factory set-
tings):
AL1 = 10 PPM (MAK value)
AL2 = 20 PPM
Carbon monoxide CO
alarm thresholds (factory set-
tings):
AL1 = 30 PPM (MAK value)
AL2 = 60 PPM
Note:
You can alter the factory-set thresholds to your own individual limits (cf. section 9.7: Setting the alarm thresholds)!
Triggering the alarm thresholds
AL1 alarm (Example: CH4)
AL2 alarm (Example: CH4)
3 Measuring operation
exceeding this alarm threshold triggers:
- the optical alarm (item 1)
- the acoustic alarm (item 7)
- the appearance of AL1 in the LCD
the interval tone is clearly dis- tinguishable from the operating signal
the AL1 alarm can be cleared, th e optical ala rm (it em 1) persists
if the concentration falls below this alarm threshold, the opti­cal and acoustic alarms (items 1 and 7) switch off
exceeding this alarm threshold triggers:
- the optical alarm (item 1)
- the acoustic alarm (item 7)
- the appearance of AL2 in the LCD
the continuous tone is clearly distinguishable from the ope­rating signal
the AL 2 alarm cannot be cleared
if the concentration falls be­low this alarm threshold, the alarm that sounds (AL1) can be cleared
19
20
3 Measuring operation
AL3 alarm (Example: CH4)
if the end of the measurement
range is exceeded, the optical and acoustic alarms (items 1 and 7) are triggered and AL3 appears in the LCD
the continuous tone is clearly
distinguishable from the ope­rating signal
the AL 3 alarm cannot be
cleared either
if the gas concetration falls be-
low the end of measurement range, the AL3 alarm is reset and the AL2 alarm activated
Note:
The AL3 alarm can be suppressed by switching to GAS-MEASURING mode or switching off the instrument!
Note:
In the GAS-WARNING mode, 0 – 100 %LEL methane concetrations exceeding the upper measuring range value can influence the accuracy of measurement.
After triggering the AL3 alarm in this measuring range, the instrument must be inspected before any further use (cf. section 5.1: Function testing, testing display accuracy, maintenance). It is necessary to repeat this operation the following days.
3 Measuring operation
Note:
In the GAS-WARNING mode the following remarks should be observed during operation at degrees be­low zero, high temperatures and high humidity:
Degrees below zero lead to a diminished display
(up to 4 % LEL at -10 °C) at the zero-point of the methane and propane sensor. This can be com­pensated by a correspondingly lower adjustment of the alarm thresholds.
When adjusting the methane sensor with dry gas,
high temperatures and high relative humidity (> 90 %) could lead to a diminished display (up to 30 % of measuring value). This can be compensated by a correspondingly lower adjustment of the alarm thresholds.
Note:
After triggering the AL3 alarm in the CO- or H2S­measuring range, the instrument must be inspected before any further use (cf. section 5.1: Function testing, testing display accuracy, maintenance).
21
22
3 Measuring operation
CH
C H C H
4
3 8
9 20
Confirmation
CO
2
O *
2
H S
2
CO
End
of range
AL3
Main alarm AL2
Pre-alarm Al1
Alarm display
0.90 %VOL 20 %LEL
0.34 %VOL
2.20 %VOL 50 %LEL
0.86 %VOL
4,40 %VOL
100 %LEL
1.7 %VOL
0.50 %VOL 1.00 %VOL 5.00 %VOL
18.0 %VOL 23.0 %VOL 25.0 %VOL
10 PPM 20 PPM 100 PPM
30 PPM 60 PPM 500 PPM
X
X
NH
3
50 PPM 75 PPM 100 PPM
0.15 %VOL 0.35 %VOL 0.7 %VOL
Summary
All alarm thresholds (factory settings) and clearance facilities in brief:
* Exception:
With oxygen O2 both thresholds take the form of a main alarm. The AL1 alarm is triggered if the value falls below the threshold of 18.0 vol.%.
3 Measuring operation

3.7 Short-term value alarm (KZW), long-term value alarm (LZW), minimum and maximum values (MIN/MAX)

briefly press the on/off key in
gas-warning mode
MIN/MAX - minimum and maximum values
Often it is only the display of the minimum or maximum con­centrations encountered so far that is of interest. These can be displayed for all gases since the last time the instrument was switched on.
MIN value (Example: CO2)
the minimum concentration
value (e.g. 0.25 %VOL) since the last time the instrument was switched on is displayed
MIN appears in the LCD
pressing the arrow-up key
brings you to the next display
MAX value (Example: CO2)
the maximum concentration
value (e.g. 0.75 %VOL) since the last time the instrument was switched on is displayed
MAX appears in the LCD
pressing the arrow-up key
brings you to the next display
23
24
3 Measuring operation
KZW - short-term value alarm
This monitoring mode is based on an averaging period, usu­ally 15 minutes (cf. section 8.5: Setting the short-term value
interval).
The total concentrations of the pertinent toxic gas over the averaging period must not exceed the product of the MAK value and the overshoot factor.
Gas MAK value Factor
carbon dioxide CO
hydrogen sulphide H2S 10 ppm 1
carbon monoxide CO 30 ppm 2
Example: a constant reading > 2.00 vol.% (= 4 x 0.50 vol.%) over
2
0.50 vol.% 4
15 minutes triggers the KZW alarm (main alarm), which cannot be cleared.
KZW (Example: CO2)
the current short-term value
(e.g. 0.30 %VOL) since the last time the instrument was switched on is displayed
KZW appears in the LCD
pressing the arrow-up key
brings you to the next display
3 Measuring operation
LZW - long-term value alarm
This monitoring mode is based on a shift length of 8 hours.
The total concentrations of the pertinent toxic gas over the length of the shift must not exceed the MAK value.
Example: a constant reading > 0.50 vol.% over 8 hours trig-
gers the LZW alarm (main alarm), which cannot be cleared
LZW (Example: CO2)
the current long-term value
(e.g. 0.10 %VOL) since the last time the instrument was switched on is displayed
LZW ap p ears in the LCD
(item 2)
the display of these values
disappears:
- if the on/off key is briefly pressed
- if no key is pressed for about 10 seconds
Note:
Both alarms (KZW and LZW) can be suppressed by switching to GAS-MEASURING mode or turning the instrument off.
25
26
3 Measuring operation

3.8 Zero point adjustment

if the EX-TEC Combi deviates
from its zero point after flush­ing with „fresh air“ (observe tolerances), this can be cor­rected manually
pressing the zero point key
sets the zero point for the gas displayed (max. 5% from the end of the measurement range)
if the zero point cannot be set,
the sensor must be adjusted (cf. section 7: Adjustment menu)
CO2 zero point
the zero point for carbon di-
oxide is 0.04 %VOL, as this is the concentration found in normal fresh air

3.9 Illumination and contrast

Illumination
repeatedly pressing the light
key switches the LCD illumi­nation on and off
after about 4 minutes the illu-
mination automatically switch­es off again
3 Measuring operation
Contrast
simultaneously pressing the
light key and a arrow key increases or reduces the con­trast of the LCD
your last setting is preserved
even when the instrument is switched off

3.10 Operating hours display and battery alarm

Operating hours display
simultaneously pressing both
arrow keys during measuring operation displays the number of operating hours remaining (e.g. 5 hours)
this display (battery symbol and
bars) disappears automatically after about 10 seconds
Battery alarm
when the charge falls below
a threshold value the battery symbol appears, the alarm lamp flashes and an acoustic signal sounds.
the battery alarm ca n be
cleared
when the battery symbol is
triggered, there is at least another 15 minutes‘ operating time remaining.
27
28
3 Measuring operation

3.11 Switching off

this period of time is empha-
sized acoustically by a double beep tone
after that the instrument must
be recharged (cf. section 4: Charging)
press the on/off key for about
3 seconds
the optical and acoustic control
signals (items 1 and 7) operate for about 3 seconds
remaining operating hours are
displayed in the form of the battery symbol and bars (e.g. 5 hours = 5 bars)
4 Charging

4.1 Charging and charge maintenance

Charging
When fully charged the instrument has a maximum of 9.5 hours‘ operating time with the pump running.
To charge the instrument you will need the docking station HS
1.2 A (see illustration), which can be used in the workshop or the emergency vehicle.
The following connection sockets can be found on the side of the docking station HS 1.2 A:
AC/DC adapter M4 100 – 240 V~,
Car cable M4 12 V= mounting,
Car cable M4 12 V= mobile,
Car cable M4 24 V= mounting.
Switch the EX-TEC Combi off and place it in the charger. A dis­play on the following lines appears:
the instrument still has 5 ope-
rating hours left (= 5 bars) and will take another 3 hours to be fully recharged
if it is fully charged, all the bars
appear and the numerical dis­play disappears

4 Charging

Charge maintenance
As soon as the instrument is fully charged it automatically switches to charge maintenance. It an be left in the charging adapter until the next time it is needed.
29
30
4 Charging

4.2 Self-discharge

If the instrument is not placed in the charging adapter when it is switched off, this will cause the nickel-cadmium battery to self­discharge, reducing the remaining operating time.
After a maximum of 30 days the instrument indicates no remaining operating hours and it must be recharged.
Note:
Brief periods of use and protracted disuse may in the long term lead to the co-called „memory effect“, which means that the actual battery capacity availa­ble is less than what is shown in the display!
You can counteract this by fully discharging the EX-TEC Combi regularly (e.g. once a month): leave it switched on until it switches itself off, then recharge it!
5 Testing/maintenance
Note:
Gas warning instruments must always be tested by the user before use or before each shift. But at a minimum all 4 month. The test must cover:
the battery charge
the display with environmental gas conditions
and test gases
(Technical bulletin BGI 518)
Note:
DIN EN 50073 and EN 45544-4 provides that porta­ble gas warning instruments (of which the EX-TEC Combi is one) must be tested as stipulated in their operating instructions and immediately before use. Items that the test is required to cover include the zero point and display sensitivity with a field calibra­tion instrument and test gas.

5 Testing / maintenance

Note:
In case of calibration data to multiple gas types are stored, tests of display accuracy must be carried out for all existing calibrations using corresponding test gases (see section 5.3).
31
32
5 Testing / maintenance

5.1 Function testing, testing display accuracy, maintenance

Please consider the necessary and prescribed instruments con­trol according to:
EN 50073
EN 45544-4
BGI 518 (T023)
BGI 836 (T021)/Applicable only in Germany
DVGW worksheet G 465-4/Applicable only in Germany
Testing must also cover accessories used. Tests carried out and other activities must be documented and the documentation re­tained for at least one year. The instrument tests required and prescribed by DVGW G 465-4 (Technical Communications, re­ference) are divided into the following sections:
What ? Who ? When ?
function testing user prior to start working
testing display accu­racy (adjustment)
servicing (maintenance, any repairs required)
specialist or special­ist company
SEWERIN, specialist or author­ised company
daily to half-yearly
annual; on defective instru­ments
Function testing
This is the simplest form of instrument test, which must be car­ried out by the user before starting work. It covers the following items:
external condition including probe systems
function-testing the controls
battery charge
inspecting the pump and suction inlet
pump function
zero point
display of test gas feeding
5 Testing / maintenance
Testing display accuracy (adjustment)
The frequency of the tests must be fixed depending on the sen­sors built-in and intensity of use.
The test must be carried out by a gas-supply company specialist, a specialist company or by SEWERIN itself.
The test gas described (see section 5.3: Test gases) must be used. The guidelines on the selection, installation, application and maintenance of appliances for the detection and measuring of flammable gases, oxygen or toxic gases (EN 50073 and EN 45544-4) need to be observed
Moreover, national regulations BGI 518 and BGI 836 must be complied with!
A full function test must also be carried out.
Servicing - maintenance and repair
Instruments must be serviced (and any necessary repairs car­ried out) at least once a year, by SEWERIN Service, a specialist company authorised by SEWERIN or an authorised specialist of the gas-supply company.
These activities must be documented in writing.
The test plaque on the instru­ment gives the date of the last service and the scheduled date of the next (e.g. 5/00 = May
2002).
Annual maintenance and repair must include at least the spe­cialist care of the instrument, its settings and the replacement of components with a limited lifetime.
Note:
Where instruments feature explosion protection
the applicable regulations must be observed!
Servicing operatives must be SEWERIN-
trained!
33
34
5 Testing / maintenance

5.2 Test set

The test set SPE VOL is available to test the pump power, zero point and sensitivity:
E
C
D
A
G
F
B
(Fig. 1)
Item description function
A device connection connection with:
probe connection
test heads
B test gas connection connection for:
test gas cans
pressure hose adapter
(in conjunction with pressure cyl­inder and pressure reducer)
C manometer Display of remaining pressure
inside the test gas container
D release button release of test gas
E fresh air supply opening for:
aspirating fresh air
fresh air hose
F needle valve with
flowmeter
reading the pump power in litres per hour (l/h)
G connection hose connected to:
device

5.3 Test gases

The following test gases are used in conjunction with the test set SPE VOL tester to test zero point and sensitivity:
Gas Test gas
Methane CH
zero point: - fresh air
sensitivity 1:- 2.2 vol.% (50 %LEL) CH4 in synth. air
sensitivity 2: - 100 vol.% CH
Propane C3H
zero point:- fresh air
sensitivity- 1.0 vol.% (59 %LEL) propane (C3H8)
Nonane C9H
zero point:- fresh air
sensitivity- repla cement test gas 0.30 vol .%
Carbon dioxide CO
zero point:- fresh air
sensitivity- 2.0 vol.% (CO2) in synth. air
Oxygen O
zero point:- 100 vol.% methane (CH4)
sensitivity- fresh air
Hydrogen sulphide H2S
zero point:- fresh air
sensitivity- 40 ppm hydrogen sulphide (H2S) in
Carbon monoxide CO
zero point:- fresh air
sensitivity- 40 ppm carbon monoxide (CO) in
Ammonia NH
zero point:- fresh air
sensitivity- 50 ppm Ammonia (NH3) in Nitrogen
5 Testing / maintenance
4
4
8
in synth. air
(⊗)
20
propane (C3H8) in synth. air (reading to be a returned when replacement test gas: 0.35 vol.%)
2
2
synth. air
synth. air
()
3
35
36
5 Testing / maintenance
These gases are supplied in 6 test gas cans:
2.2 vol.% CH4, 2.0 vol.% CO2, 17.5 vol.% O2, 40 ppm CO
1.0 vol.% C3H
0.3 vol.% C3H
40 ppm H2S
50 ppm NH
100 vol.% CH
8
8
3
4
5 Testing / maintenance

5.4 Testing the pump power, zero point and sensitivity

Proceed as follows:
screw the selected test gas can onto the test set SPE VOL as
far as it will go (fig. 1, item B)
connect the instrument‘s probe connection (item 3) with the
hose of the test set SPE VOL (fig. 1, item A)
switch the instrument on; the pump aspirates fresh air through
the test set SPE VOL (fig. 1, item E)
enter your test results in the Inspection protocol
Pump power
use the needle valve (fig. 1, item F) to set the flow to maximum;
it must be greater than 30 l/h
Zero point
wait for the instrument to settle at a stable zero point; permis-
sible tolerances with fresh air are:
Gas
Methane CH
4
Tolerance
-0.15 – +0.15 vol.% (-3 – +3 %LEL)
Propane C3H
8
-0.06 – +0.06 vol.% (-3 – +3 %LEL)
Nonane C9H
20
()
-0.05 – +0.05 vol.% or (-3 – +3 %LEL)
Carbon dioxide CO
2
+0.04 – +0.10 vol.%
Hydrogen sulphide H2S -3 – +3 ppm
Carbon monoxide CO -3 – +3 ppm
Ammonia NH
3
-3 – +3 ppm
()
In a fresh air environment, the display of the oxygen sensor will indicate a nominal value of 20.9 vol.%.
For zero calibration, the sensor is exposed to a pure CH4 envi­ronment (100 % by volume).
Tolerances:
Oxygen O2 -0.5 – +0.5 vol.%
37
38
5 Testing / maintenance
If display values are outside these tolerances the pertinent sensor must be readjusted (cf. section 7: Adjustment menu).
Sensitivity
press the release key (fig. 1, item D) on the test set SPE VOL
and adjust the flow to the fresh air value (fig. 1, item F)
hold down the release key until the displayed concentration
has settled at a stable value; permissible tolerances with test gases:
Gas
Methane CH
Propane C3H
Nonane C9H
4
20
8
()
Carbon dioxide CO
2
Tolerance
2.0 – 2.4 vol.% (45 – 55 %LEL)
0.90 – 1.10 vol.% (54 – 64 %LEL)
0.35 – 0.40 vol.% (50 – 57 %LEL)
1.8 – 2.2 vol.%
Hydrogen sulphide H2S 37 – 43 ppm
Carbon monoxide CO 37 – 43 ppm
Ammonia NH
3
47 – 53 ppm
()
The oxygen sensor is tested for its sensitivity in a fresh air en­vironment.
Tolerances:
Oxygen O2 20.4 – 21.4 vol.%
If display values are outside these tolerances the pertinent sensor must be readjusted (cf. section 7: Adjustment menu).
INFO V2.1
JUSTAGE
H2S 0 PPM H2S 40 PPM CO 0 PPM CO 30 PPM O2 0 VOL.% O2 20,9 VOL.% CO2 0 VOL.% CO2 2,00 VOL.% CH4 0 %UEG CH4 50 %UEG CH4 100 %VOL CH4 %UEG %LEL %LIE %VOL %GAZ %VOL %GAZ PRÜFGAS INS-OK EXIT >
MEM-CLR MEM-INT MEM-STOP KZW-INT EXIT >
DAT/UHR INS-INT INS-STOP 100
%VOL PIN-CODE ALARM LCD EXIT
>
MEMORY
SYSTEM
EXIT
6 Info menu

6.1 Menu structure

6 Info menu

39
40
6 Info menu

6.2 Overview

The info menu is accessible only when the EX-TEC Combi is switched off.
now simultaneously press the
following 3 keys
you are now in the INFO menu
item (cf. menu structure)
the software version number
(e.g. V2.1) is displayed and the LCD illumination automati­cally switches on for about 4 minutes
you must now enter your PIN
code (cf. section 9.6: Setting the PIN code)
0001 = factory settings
only now do you have access
to all menu items
the pump runs at constant
power; it can be switched on or off with the pump key at any time
the arrow-up key returns you
to the menu structure
6 Info menu
you are now in the ADJUST-
MENT menu item (cf. section 7: Adjustment menu)
the arrow-up key brings you
to the next display
you are now in the MEMORY
me n u item (cf. secti o n 8: Memory menu)
the arrow-up key brings you
to the next display
you are now in the menu item
SYSTEM (cf. section 9: Sys­tem menu)
the arrow-up key brings you
to the next display
41
42
6 Info menu
... or
you are now in the EXIT menu
item
there are now 2 ways of con-
tinuing to navigate through the menu structure
pressing the arrow-up key
returns you to the ADJUST­MENT menu item (cf. menu
structure)
briefly pressing the on/off key
leaves the menu structure and the instrument switches to measuring operation
JUSTAGE
H2S 0 PPM H2S 40 PPM CO 0 PPM CO 30 PPM O2 0 VOL.% O2 20,9 VOL.% CO2 0 VOL.% CO2 2,00 VOL.% CH4 0 %UEG CH4 50 %UEG CH4 100 %VOL CH4 %UEG %LEL %LIE %VOL %GAZ %VOL %GAZ PRÜFGAS INS-OK EXIT >
7 Adjustment menu

7.1 Menu structure

7 Adjustment menu

you are in the ADJUSTMENT
menu item
briefly pressing the on/off key
brings you to the adjustment menu
43
44
7 Adjustment menu

7.2 Setting the H2S sensor

H2S - zero point 0 PPM
release fresh air from the
SPE VOL
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
pressing the arrow-up key
brings you to the next display
H2S - sensitivity 40 PPM
now release test gas 40 PPM
H2S from the SPE VOL (cf. section 7.8: Setting the test gas concentration)
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
once this has happened, turn
off the test gas feed
pressing the arrow-up key
brings you to the CO-sensor setting facility

7.3 Setting the CO sensor

7 Adjustment menu
CO - zero point 0 PPM
now release fresh air from the
SPE VOL
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
pressing the arrow-up key
brings you to the next display
CO - sensitivity 40 PPM
now release test gas 40 PPM
CO from the SPE VOL (cf. sec­tion 7.8: Setting the test gas concentration)
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
once this has happened, turn
off the test gas feed
pressing the arrow-up key
brings you to the O2-sensor setting facility
45
46
7 Adjustment menu

7.4 Setting the O2 sensor

O2 - zero point 0 %VOL
now release test gas 100 vol.%
CH4 from the SPE VOL
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
once this has happened, turn
off the test gas feed
pressing the arrow-up key
brings you to the next display
O2 - sensitivity 20.9 %VOL
now release fresh air from
the SPE VOL (cf. section 7.8: Setting the test gas concen­tration)
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
pressing the arrow-up key
brings you to the CO2-sensor setting facility

7.5 Setting the CO2 sensor

7 Adjustment menu
CO2 - zero point 0 %VOL
now release fresh air from the
SPE VOL through a carbon dioxide filter (Sewerin acces­sory, see section 13.2)
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
pressing the arrow-up key
brings you to the next display
CO2 - sensitivity 2.00 %VOL
now release test gas 2.00
%VOL CO2 from the SPE VOL (cf. section 7.8: Setting the test gas concentration)
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
once this has happened, turn
off the test gas feed
pressing the arrow-up key
brings you to the CH4-sensor setting facility
47
48
7 Adjustment menu

7.6 Setting the CH4 sensor

CH4 - zero point 0 %VOL
now release fresh air from the
SPE VOL
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
pressing the arrow-up key
brings you to the next display
CH4 - sensitivity 2.20 %VOL
now release test gas 2.20 %VOL
(50 %UEG) CH4 from the SPE VOL (cf. section 7.8: Setting the test gas concentration)
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
once this has happened, turn
off the test gas feed
pressing the arrow-up key
brings you to the next display
7 Adjustment menu
CH4 - sensitivity 100 %VOL
now release test gas 100 %VOL
CH4 from the SPE VOL
alternatively you can dispense
with the SPE VOL and use
natural gas from the local supply instead
wait for the display to settle at
a stable value and confirm the adjustment with the on/off key (OK appears in the LCD)
once this has happened, turn
off the test gas feed
Note:
Before switching to the next step, please wait until the displayed concentration has reached the value of 0 vol.%!
pressing the arrow-up key
brings you to the next display
CH4 - %UEG-range language
repeatedly pressing the on/off
key switches between the fol­lowing displays in the %UEG range:
%UEG - display in Untere Explosionsgrenze (D)
%LEL - display in Lower Explosive Limit (GB)
%LIE - display in Limite Inférieure d´Explosion (F)
%VOL - display in %VOL (D/GB)
%GAZ - display in %GAZ (F)
49
50
7 Adjustment menu
conf i r m th e d i s pl a y, e. g .
%VOL, with the on/off key (OK appears in the LCD)
this se t t in g is pre s er v e d
even when the instrument is switched off
pressing the arrow-up key
brings you to the measure­ment-quantity setting facility

7.7 Setting the measurement quantity

CH4, CO2, O2 ­%VOL-range language
repeatedly pressing the on/off
key switches between the fol­lowing displays in the vol.% range:
%VOL - display in %VOL (D/GB)
%GAZ - display in %GAZ (F)
conf i r m th e d i s pl a y, e. g .
%VOL, with the on/off key (OK appears in the LCD)
this se t t in g is pre s er v e d
even when the instrument is switched off
pressing the arrow-up key
brings you to the test gas con­centration setting facility
CH4 GAS CO2 GAS O2 GAS H2S GAS CO GAS EXIT >>
PRÜFGAS

7.8 Setting the test gas concentration

if you use use test gases
other than those supplied by SEWERIN, the concentration must be set accordingly
you are in the TEST GAS
menu item
7 Adjustment menu
briefly pressing the on/off
key brings you to the test gas menu
Example: CH
you are now in the CH4 GAS
menu item
4
51
52
7 Adjustment menu
briefly pressing the on/off key
brings you to the CH4 test gas concentration setting facility
the current concentration is
displayed (e.g.: 50 %UEG =
2.20 %VOL)
you can set the desired con-
cen t r at i on by re p ea t ed ly pressing or holding down a arrow key
the selected concentration is
displayed (e.g.: 45 %UEG =
2.00 %VOL)
confirm the concentration with
the on/off key
this se t t in g is pre s er v e d
even when the instrument is switched off
the adjustment ranges of the
individual test gas concentra­tions are:
7 Adjustment menu
test gas 2.20 vol.% (50%LEL) methane CH
adjustment range:- 1.75 – 3.50 vol.% (40 – 80 %LEL)
step size- 0.05 vol.% (1 %LEL)
test gas 1.0 vol.% (59 %LEL) propane C3H
adjustment range- 0.68 – 1.36 vol.% (40 – 88 %LEL)
step size- 0.02 vol.% (1 %LEL)
test gas
for nonane C9H
20
()
replacement test gas 0.3 vol.% propane (C3H8)
adjustment range- 0.30 – 0.50 vol.% (40 – 80 %LEL)
step size- 0.05 vol.% (1 %LEL)
test gas 2.00 vol.% carbon dioxide CO
2
adjustment range- 0.90 – 4.10 vol.%
tep size- 0.02 vol.%
test gas 20.9 vol.% oxygen O
2
adjustment range- 17.0 – 22.0 vol.%
tep size- 0.1 vol.%
test gas 40 ppm hydrogen sulphide H2S
adjustment range- 10 – 100 ppm
step size- 1 ppm
test gas 40 ppm carbon monoxide CO
adjustment range- 10 – 150 ppm
step size- 1 ppm
test gas
50 ppm Ammoniak NH
()
3
adjustment range- 10 – 100 ppm
step size- 1 ppm
4
8
repeatedly pressing a arrow
key brings you to the exit from the test gas menu
53
54
7 Adjustment menu
7.9 Inspection confirmation
EXIT >>
signpost (>>) to menu level 2
pressing the on/off key leaves
the test gas menu
pressing the arrow-up key
brings you to the inspection
INSPECTION OK
the EX-TEC Combi can re-
mind you of scheduled inspec­tion and adjustment dates
this requires the inspection
interval and the inspection block to be set in the system
menu (cf. sections 9.3 „Set­ting the inspection interval“ and 9.4 „Setting the inspection block“)
confirm the inspection or ad-
justment you have carried out with the on/off key (OK appears in the LCD):
this date is stored as a function
of the set date (cf. section 9.2 „Setting the date/time“)

7.10 Leaving the adjustment menu

7 Adjustment menu
the next inspection or adjust-
ment date is calculated in ac­cordance with the set inspec­tion interval
pressing the arrow-up key
brings you to the exit from the adjustment menu
EXIT >
signpost (>) to menu level 1
pressing the on/off key leaves
the adjustment menu
you are now back at the top
ma i n -menu level an d can switch between the following menu items:
- JUSTAGE
- MEMORY
- SYSTEM
- EXIT
55
56

8 Memory menu

MEM-CLR MEM-INT MEM-STOP KZW-INT EXIT >
MEMORY
8 Memory menu

8.1 Menu structure

the EX-TEC Combi continu-
ously stores readings from the sensors that are present
these can later be read out
with the appropriate evalua­tion software (separate user manual) via the RS-232-inter­face (item 8)
you are in the MEMORY menu
item
briefly pressing the on/off
key brings you to the memory menu
8 Memory menu

8.2 Clearing memory

(⊗)
MEMORY CLEAR
if you have set the memory
mode = stack memory (cf. section 8.4: Setting memory mode), this function enables you to clear the entire readings memory
confirm the clearance with
the on/off key (OK appears in the LCD)
pressing the arrow-up key
brings you to the memory-inter­val setting facility
57
58
8 Memory menu

8.3 Setting the memory interval

(⊗)
MEMORY INTERVAL
briefly pressing the on/off key
brings you to the memory-in­terval setting facility
by repe atedly pres sing or
holding down a arrow key you can select the following memory intervals:
Seconds range:
- 1 second
- 10 seconds
- 20 seconds
- 30 seconds
Minutes range:
- 1 minute
- 2 minutes
- 3 minutes
- 5 minutes
- 10 minutes
- 20 minutes
- 30 minutes
8 Memory menu
Data-memory capacity
the EX-TEC Combi stores the following values in its data
memory:
- readings for each gas (instantaneous values at the moment of scanning)
- event values for each gas (MIN, MAX, KZW, LZW values and alarm overshoots)
- special error displays, alarms
depending on the set memory interval and the number of gases,
data memory can continuously record the following periods (times in hh:mm):
M e m o r y
3 gases 4 gases 5 gases
interval
1 sec 01:02 00:41 00:29
10 sec 10:28 06:58 04:51
20 sec 20:56 13:56 09:43
30 sec 31:25 20:54 14:35
1 min 62:50 41:48 29:11
2 min 125:40 83:37 58:23
3 min 188:30 125:25 87:34
5 min 314:10 209:02 145:58
10 min 628:20 418:05 291:56
20 min 1256:40 836:10 583:52
30 min 1885:00 1254:15 875:48
Example: when the instrument is set to warn for 4 gases and with
the memory interval set to 1 minute you can record event values over a period of 41 hours 48 minutes.
59
60
8 Memory menu
confirm the interval with the
on/off key
this se t t in g is pre s er v e d
even when the instrument is switched off
pressing the arrow-up key
brings you to the memory­mode setting facility

8.4 Setting the memory mode

(⊗)
MEMORY STOP
you have the possibility to
choose the memory mode between the ring memory or the stack memory
briefly pressing the on/off key
brings you to the memory­mode setting facility
by repeatedly pr e s s i ng a
arrow key you can select the following memory modes:
OFF (ring memory)
readings are continuously written to memory; when memory is full the oldest values are over­written
ON (stack memory)
readings ar e only written to memory until it is full, thus write­protecting the oldest values
confirm the memory mode with
the on/off key
this se t t in g is pre s er v e d
even when the instrument is switched off
pressing the arrow-up key
brings you to the short-term value-interval setting facility

8.5 Setting the short-term value interval

SHORT-TERM VALUE INTERVAL
for the meaning of the KZW
alarm (cf. section 3.7 Short­term value alarm (KZW), long­term value alarm (LZW), mini­mum and maximum values MIN / MAX)
8 Memory menu
briefly pressing the on/off key
brings you to the short-term value-interval setting facility
by repeatedly pressing or hold-
ing down a arrow key you can select the following short-term value-interval settings:
Adjustment range:
- 1 – 30 minutes
- 1 minute step size
61
62
8 Memory menu

8.6 Leaving the memory menu

the usual setting is for a 15
minutes averaging interval
confirm the interval with the
on/off key
this se t t in g is pre s er v e d
even when the instrument is switched off
pressing the arrow-up key
brings you to the exit from the memory menu
EXIT >
signpost (>) to menu level 1
pressing the on/off key leaves
the memory menu
you are now back at the top
ma i n -menu level an d can switch between the following menu items:
- JUSTAGE
- MEMORY
- SYSTEM
- EXIT
DAT/UHR INS-INT INS-STOP 100 %VOL PIN-CODE ALARM LCD EXIT
>
SYSTEM
9 System menu

9.1 Menu structure

9 System menu

you are in the SYSTEM menu
item
briefly pressing the on/off
key brings you to the system menu
63
64
9 System menu

9.2 Setting the date/time

DATE/TIME
briefly pressing the on/off key
brings you to the date/time setting
Date
the last day (24 - flashing)
and month (02) to be set are displayed
by repe atedly pres sing or
holding down a arrow key ...
... and confirming with the
on/off key you can first set the day and then the month to the current date
Year
the last year (2000 - flashing)
to be set is displayed
9 System menu
by repe atedly pres sing or
holding down a arrow key ...
... and confirming with the
on/off key you can set the current year
Time
the last hours (17 - flashing)
and minutes (49) to be set are displayed
by repe atedly pres sing or
holding down a arrow key ...
... and confirming with the
on/off key you can first set the hours and then the minutes to the current time
these settings are preserved
even when the instrument is switched off
pressing the arrow-up key
brings you to the inspection­interval setting facility
65
66
9 System menu

9.3 Setting the inspection interval

INSPECTION INTERVAL
the EX-TEC Combi can re-
mind you of regular scheduled tests (e.g. inspections, adjust­ments)
this reminder is based on the
inspection interval
briefly pressing the on/off key
brings you to the inspection­interval setting facility
Inspection interval = 0 – 52 CW
the last interval to be set is
displayed in CW (calendar weeks), e.g.:
- 0 CW = function inactive
- 4 CW = monthly
- 52 CW = annual
by repe atedly pres sing or
holding down a arrow key ...
... and confirming with the
on/off key you can set the desired interval
this se t t in g is pre s er v e d
even when the instrument is switched off
pressing the arrow-up key
brings you to the inspection­block setting facility
9 System menu
Example: inspection interval
selected inspection interval:
04 weeks
inspection routine started (i.e. inspection confirmed, see section 7.9):
10.02.2004
from these settings follows: the next inspection has to be effected between 09. – 15.03.2004
During the next 3 weeks the EX-TEC Combi will display the fol­lowing references to the inspection date:
24.02.2004 – 01.03.2004 (> 1 week before)
LCD: the coming inspection date is
displayed for about 3 seconds on switch-on
lamp/buzzer: inactive
instrument: the instrument then automati-
ca lly swit ches to meas uring operation
02.03.2004 08.03.2004 (1 week before)
LCD: the coming inspection date is
displayed for about 3 seconds on switch-on
lamp/buzzer interval lamps / interval tone
instrument: the instrument then automati-
ca lly swit ches to meas uring operation
67
68
9 System menu
09.03.2004 – 15.03.2004 (scheduled)
LCD: the due inspection date is dis-
lamp/buzzer: interval light/sound
instrument: when the display is cleared with
16.03.2004 ... (from 1 week later)
LCD: the overdue inspection date is
lamp/buzzer: interval light/sound
instrument: depending on the setting of the
played on switch-on
the buzzer key (item 4) or after waiting for about 15 seconds the instrument automatically switches to measuring operation.
displayed on switch-on
INS-STOP function (cf. section
9.4: Setting the inspection block) the following conditions are pos­sible:
INS-STOP = OFF
when the display is cleared with the buzzer key (item 4) or after waiting for about 15 seconds the instrument automatically switches to measuring operation
INS-STOP = ON
when any key is pressed (item
4) or after waiting for about 15 seconds the instrument automati­cally switches off.

9.4 Setting the inspection block

9 System menu
INSPECTION STOP
to make sure your EX-TEC
Combi is regularly checked you can activate an inspec­tion block
this block does not become
active until the next inspection date has passed (cf. section
9.3: Setting the inspection interval)
after that you cannot use the
instrument until the inspection has been carried out and confirmed (cf. section 7.9: Inspection confirmation)
briefly pressing the on/off key
brings you to the inspection­block setting facility
Inspection-block condition
the last setting is displayed,
e.g.:
- OFF = block inactive
- ON = block active
by repeatedly pr e s s i ng a
arrow key ...
... and confirming with the on/
off key you can set the desired condition
69
70
9 System menu

9.5 Setting the 100 vol.% range

this se t t in g is pre s er v e d
even when the instrument is switched off
pressing the arrow-up key
brings you to the operating­signal setting facility
100 %VOL (GAS-MEASURING)
th e EX- T EC Com b i has
two modes (cf. section 3.3: Modes):
WARNING mode
monitoring the atmosphere (w o r kplace moni t o ring) in shafts and chambers
GAS-MEASURING mode determining gas concentra­tions
if you wish to use the instrument
purely as a warning device, you can switch off the 100 %VOL (GAS-MEASURING) range
briefly pressing the on/off key
brings you to the 100 %VOL range setting
100 %VOL range condition
the last setting is displayed:
- OFF = 100 %VOL inactive
- ON = 100 %VOL active

9.6 Setting the PIN code

9 System menu
by repeatedly pr e s s i ng a
arrow key ...
... and confirming with the on/
off key you can set the desired condition
this se t t in g is pre s er v e d
even when the instrument is switched off
pressing the arrow-up key
brings you to the setting of the PIN-code
PIN CODE
you can set your EX-TEC
Combi so that only authorised persons, e.g.:
- instrument technicians
- experts
have access to the info menu with all its subfunctions
this involves setting a PIN
code that must be entered every time the info menu is called
when an incorrect PIN code is
entered the instrument reverts to its switch-on routine
briefly pressing the on/off key
brings you to the PIN-code setting facility
71
72
9 System menu
Setting the PIN CODE
the last PIN code to be set
(0001 = factory setting) ap­pears in the LCD
we recommend you to use a
different PIN code
by repe atedly pres sing or
holding down a arrow key ...
... and confirming with the
on/off key you can set each of the 4 digits from left to right to the desired PIN code
PIN code = 0000
the function is inactive, every
user has access to the info menu
PIN code = 0001 – 9999
the function is active, only
persons who know the set PIN code have access to the info menu
this se t t in g is pre s er v e d
even when the instrument is switched off
pressing the arrow-up key
brings you to the alarm-thresh­old setting facility
Note:
Make a note of your PIN code and only give it to authorised persons.
If you forget your PIN, please contact SEWERIN Service.
CH4 AL1 CH4 AL2 CO2 AL1 CO2 AL2 O2 AL1 O2 AL2 H2S AL1 H2S AL2 CO AL1 CO AL2 EXIT >>
ALARM

9.7 Setting the alarm thresholds

9 System menu
you can set your own alarm
thresholds if you do not wish to use those pre-set by SEWERIN
you are in the ALARM menu
item
briefly pressing the on/off
key brings you to the alarm­thresholds menu
73
74
9 System menu
Example: CH4 alarm threshold 2
you are now in the CH4 AL2
menu item
briefly pressing the on/off key
brings you to the facility to set alarm threshold 2 for methane CH
4
the current alarm threshold is
displayed (e.g.: 50 %UEG = 2.20 %VOL)
by repe atedly pres sing or
holding down a arrow key you can set the desired alarm threshold
the selected alarm threshold
is displayed (e.g. 40 %UEG =
1.75 %VOL)
the setting ranges for the in-
dividual alarm thresholds are: (factory settings in bold):
9 System menu
Gas AL1 threshold AL2 threshold Methane CH4 (vol.%) 0.90 vol.% 2.20 vol.%
adjustment range:- 0.45 – 4.15 vol.% 0.50 – 4.20 vol.%
step size:- 0.05 Vol.% 0.05 Vol.%
Methane CH4 (%LEL) 20 %LEL 50 %LEL
adjustment range:- 10 – 94 %LEL 11 – 95 %LEL
step size:- 1 %LEL 1 %LEL
Propane C3H8 (vol.%) 0.34 vol.% 0.86 vol.%
adjustment range:- 0.18 – 1.60 vol.% 0.20 – 1.62 vol.%
step size:- 0.02 vol.% 0.02 vol.%
Propane C3H8 (%LEL) 20 %LEL 50 %LEL
adjustment range:- 10 – 94 %LEL 11 – 95 %LEL
step size:- 1 %LEL 1 %LEL
Nonane C9H20 (%LEL) (⊗) 0.15 vol.% 0.35 vol.%
adjustment range- 0.05 – 0.60 vol.% 0.10 – 0.65 vol.%
step size:- 0.05 vol.% 0.05 vol.%
Carbon dioxide CO
0.50 vol.% 1.00 vol.%
2
adjustment range:- 0.10 – 4.90 vol.% 0.20 – 5.00 vol.%
step size:- 0.10 vol.% 0.10 vol.%
Oxygen O
2
18.0 vol.% 23.0 vol.%
adjustment range:- 15.0 – 24.9 vol.% 15.1 – 25.0 vol.%
step size:- 0.1 vol.% 0.1 vol.%
Hydrogen sulphide H2S 10 ppm 20 ppm
adjustment range:- 5 – 99 ppm 6 – 100 ppm
step size:- 1 ppm 1 ppm
Carbon monoxide CO 30 ppm 60 ppm
adjustment range:- 5 – 199 ppm 6 – 200 ppm
step size:- 1 ppm 1 ppm
Ammonia NH3 (⊗) 50 ppm 75 ppm
adjustment range- 5 – 99 ppm 6 – 100 ppm
step size:- 1 ppm 1 ppm
75
76
9 System menu
confirm the alarm threshold
with the on/off key
this setting is preserved even
when the instrument is switched off
repeatedly pressing a arrow
key brings you to the exit from the alarm-threshold menu
EXIT >>
signpost (>>) to menu level 2
pressing the on/off key leaves
the alarm thresholds menu
pressing the arrow-up key
brings you to the LCD-check facility

9.8 Checking the LCD

9.9 Leaving the system menu

9 System menu
with this function you can carry
out a visual check that all segments of the LCD are in working order
confirm the LCD check with
the on/off key
all the possible LCD characters
and symbols are activated
pressing the arrow-up key
brings you to the exit from the menu
EXIT >
signpost (>) to menu level 1
pressing the on/off key leaves
the memory menu
you are now back at the top
ma i n -menu level an d can switch between the following menu items:
- JUSTAGE
- MEMORY
- SYSTEM
- EXIT
77
78

10 Application hints

10 Application hints

10.1 Application hints from report of suitability examination PFG-No. 41300401P

On the basis of the measurement results and remarks contained in the test report PFG-No 41300401P, the EX-TEC portable gas warning device from Hermann Sewerin GmbH is suitable for measuring carbon monoxide up to 500 ppm CO, hydrogen sulphide up to 100 ppm H2S, oxygen up to 25 % O2 (with respect to the use of the device to monitor the ambient air for oxygen deficit or excess) and carbon dioxide at concentrations up to 5 % CO2, provided that its properties and version correspond to the documents specified in the test report PFG-No 41300401P, and that it is operated accordingly and that the following requirements are fulfilled:
The operating manual submitted and tested by the DMT must
be followed in every respect.
When using the gas warning device it is essential to check
whether the response times are short enough to enable its warning function to be triggered quickly enough to avoid potentially hazardous situations. If necessary the alarm threshold should be set substantially below the safety threshold.
Please keep in mind that if the device is used in a location with
high temperatures and high humidity, the measured value in the oxygen-mesurement range will be considerably increased. If necessary, recalibrate the device before use under these conditions.
The device must be checked regularly, particularly for the
measurement ranges 0 – 100 ppm H2S and 0 – 500 ppm CO, to insure that its display for dummy and test gases is accurate.
The devices are to be provided with a permanent name plate
where the manufacturer, type and serial number and the following inscription is to be included:
„PFG-No. 41300401“
10 Application hints
Other indentification regulations, especially those of ElexV, re­mains unaffected. By attaching this name plate, the manufacturer confirms that the devices meet the properties and technical re­quirements described in this report. Any device not bearing this name plate does not conform to the present report.
Upon request, the operator must be supplied with a copy of this
report and test report PFG-No. PFG-Nr. 41300401P.
79
80

11 Technical aspects

11 Technical aspects

11.1 Technical notices

Heavy gas
When working in probe holes (GAS-MEASURING mode), you may find that carbon dioxide CO2 (heavy gas) overlays the methane CH4 (light gas). In this event the instrument will indicate no concentration (0 vol.%).
If it is carbon dioxide (monitoring over the CO2 measurement range), we recommend the use of a CO2 filter (accessory kit) to suppress the heavy-gas component.
High CO2 concentrations (from approx. 5 vol.% on) will have a damaging effect on the CO2 sensor.
Gas injection/blanketing
If you wish to carry out a gas injection (an increase in concentra­tion to 100 vol.%) or blanketing (a reduction in concentration to 0 vol.%), select GAS-MEASURING mode (device variation 1) with the pump key (device variation 2) (cf. section 9.5: Setting the 100 vol.% range).
Only in this mode can measurement be reliably monitored.
Probe hoses
The rule of thumb is: the concentration display is delayed by about 1 second per metre of hose.
Always use SEWERIN probe hoses 1 m, 2 m or 6 m long. The maximum length of 12 m should not be exceeded.
Sensitivity of the thermal-shading sensor
The sensor sensitivity of the thermal-shading sensor can be falsi­fied under the following conditions:
an atmosphere low in oxygen will lead to the reduction of the
value displayed (the sensor will choke).
the device may not be operated in an oxygen-enriched atmos-
phere due to the explosion protection.
11 Technical aspects
The operation of the sensors is impaired by gaseous components, e.g. of silicons, oils and phosphate esters. These irreversibly reduce sensitivity.
Sensor sensitivity is also reduced by pollution of the measuring environment, e.g. by halogens, burnt neoprene, PVC or trichloroethylene - but not irreversibly.
Cleaning
The instrument should be cleaned with a damp cloth. Use no solvents, benzene or similar substances.
Static charge
Electrostatic charges should generally be avoided. Electrostatically floating objects (like metallic housings with no earth connection, for example) are unprotected against charges transferred from dust, aerosols and the like.
Fine dust filters
There are fine dust filters in the removable probe connection (item 3) and in most probes.
Note:
Heavily-soiled filters should be replaced (cf. section
11.5: Wearing parts)!
CAUTION!
Do not use other than the original SEWERIN fine dust filters with the EX-TEC Combi. Never use ac- tivated carbon filters, as these will collect H2S gas particles, thus causing the warning feature to stop working properly.
81
82
11 Technical aspects

11.2 Technical data

Instrument data
Serial no.: 045 01 0001 (type - model - number)
Dimensions: 129 x 192 x 65 mm (W x H x D)
Weight: 1500 g (depending on sensor equipment)
Type of protection: IP 54
Fields of application
Operating temperature: -10 °C – +40 °C
Storage temperature: -20 °C – +40 °C
Humidity range: 15% r.h. – 90% r.h. (not condensing)
Pressure range: 800 hPa – 1200 hPa
Explosion protection (CENELEC)
Testing institution: Physikalisch-Technische Bundesanstalt,
Test number: PTB 96 ATEX 2166, 1. + 3. supplement
Identification mark: II 2 G EEx ib d IIB T3
(5 % r.h. – 90% r.h. short-time)
Braunschweig
Measuring function
Testing institution: DMT GmbH, Essen
Test number DMT 01 ATEX G 002, 1. supplement
Measuring range 0 – 100 %LEL methane/ propane
PFG-Number 41300401
Power supply
Operating time: min. 9,5 h
Battery type: NiCd, rechargeable
Charging voltage: 12 V=
Charging current: 380 mA
Charging time: 13 h
11 Technical aspects
Pump performance
Gas-warning: > 30 l/h and >150 mbar
Gas-measuring: > 50 l/h and >150 mbar
Sensors warm-up time
for all sensors: < 60s
83
84
11 Technical aspects
Sensor - Methane CH
4
Sensor data
measurement principle:- combined sensor, thermal shad-
ing (TS) and thermal conductivity (TC)
measurement range:- 0 – 4.40 vol.% (AL3) in 0.05 vol.%
steps or 0 – 100 %LEL (AL3) in 1 %LEL steps
t90 time:- < 30 seconds
storage temperature:- -20 °C – +60 °C (for repl. sensors)
Alarm thresholds (factory settings)
CH4:- AL1 = 0.90 vol.% AL2=2.20 vol.%
or
AL1 = 20 %LEL AL2 = 50 %LEL
Lifetime
garanteed:- 1 year
expected:- 5 years
Test gases
zero point:- fresh air
sensitivity 1:- 2.20 vol.% (50 %LEL) CH4 in synth. air
sensitivity 2:- 100 vol.% CH
4
11 Technical aspects
Sensor - Carbon dioxide CO
2
Sensor data
measurement principle:- infra-red sensor (IR)
measurement range:- 0.06 – 5.00 vol.% (AL3) in 0.02 vol.%
steps
zero point drift:- 0.12 vol.%
t90 time:- < 20 seconds
storage temperature:- -20 °C – +60 °C (for repl. sensors)
Alarm thresholds (factory settings)
CO2:- AL1 = 0.50 vol.% AL2=1.00 vol.%
Temperature influence
zero point:- up to 40 ppm CO2/°C
sensitivity- up to 1.3 % of the measuring value/°C
Time drift: < 70 ppm/month
Cross-sensivity
none known-
Lifetime
garanteed:- 2 years
expected:- 5 years
Test gases
zero point:- fresh air
sensitivity:- 2.00 vol.% CO2 in synth. air
85
86
11 Technical aspects
Sensor - Oxygen O2
Sensor data
measurement principle:- electrochemical sensor (EC)
measurement range:- 0 – 25.0 vol.% (AL3) in 0.1 vol.%
t90 time:- < 30 seconds
storage temperature:- -0 °C – +20 °C (for repl. sensors)
Alarm thresholds (factory settings)
O2:- AL1 = 18.0 vol.% AL2=23.0 vol.%
Temperature influence
sensitivity- < 0.3 % Signal/°C
Cross-sensivity
CO2:- at 5 vol.% CO2 in 23 vol.% O2 <= 1 % O
Lifetime
garanteed:- 20 months
expected:- 24 months
steps
2
Test gases
zero point:- 100 vol.% CH
sensitivity:- fresh air
4
11 Technical aspects
Sensor - Hydrogen sulphide H2S
Sensor data
measurement principle:- electrochemical sensor (EC)
measurement range:- 2 – 100 ppm (AL3) in 1 ppm steps
zero point drift:- 3 ppm
t90 time:- < 30 seconds
storage temperature:- -0 °C – +20 °C (for repl. sensors)
Alarm thresholds (factory settings)
H2S:- AL1 = 10 ppm AL2 = 20 ppm
Temperature influence
zero point- no temperature influence
sensitivity- < 0.5 % Signal/°C
Time drift < 0.3 ppm/month
Cross-sensivity at 20 °C
100 ppm CO:- ~ +3 ppm H2S
20 ppm CL2:- ~ -1 ppm H2S
500 ppm C2H4:- ~ +2 ppm H2S
2 Vol.-% H2:- ~ +100 ppm H2S
10 ppm SO2:- ~ +3 ppm H2S
none known-
Lifetime
garanteed:- 2 years
expected:- 3 years
Test gases
zero point:- fresh air
sensitivity:- 40 ppm H2S in synthetic air
87
88
11 Technical aspects
Sensor - Carbon monoxide CO
Sensor data
measurement principle:- electrochemical sensor (EC)
measurement range:- 2 – 500 ppm (AL3) in 1 ppm steps
zero point drift:- 4 ppm
t90 time:- < 30 seconds
storage temperature:- -0 °C – +20 °C (for repl. sensors)
Alarm thresholds (factory settings)
CO:- AL1 = 30 ppm AL2 = 60 ppm
Temperature influence
zero point- < 5 ppm
sensitivity- 1.4 % Signal/°C
Time drift < 0.3 ppm/month
Cross-sensivity at 20 °C
1000 ppm H2:- ~ +450 ppm CO
100 ppm NO:- ~ +25 ppm CO
none known-
Lifetime
garanteed:- 2 years
expected:- 3 years
Test gases
zero point:- fresh air
sensitivity:- 40 ppm CO in synthetic air
11 Technical aspects
Sensor - Hydrogen sulphide H2S/Carbon monoxide CO
Sensor data
measurement principle:- electrochemical combined sensor
(EC) measurement range H2S:- 2 – 100 ppm (AL3) in 1 ppm steps measurement range CO:- 2 – 500 ppm (AL3) in 1 ppm steps zero point drift H2S:- 3 ppm zero point drift CO:- 4 ppm t90 time:- < 60 seconds for H2S
< 50 seconds for CO storage temperature:- -0 °C – +20 °C (for repl. sensors)
Alarm thresholds (factory settings)
H2S:- AL1 = 10 ppm AL2 = 20 ppm CO:- AL1 = 30 ppm AL2 = 60 ppm
Temperature influence
H2S zero point:- no influence for H2S H2S sensivity:- < 0.5 % Signal/°C CO zero point:- < 5 ppm CO sensivity:- 1.4 % Signal/°C
Time drift
H2S:- < 0.3 ppm/month CO:- < 0.3 ppm/month
Cross-sensivity at 20 °C
300 ppm CO:- < +3 ppm on H2S
~ +300 ppm on CO 15 ppm H2S:- ~ +15 ppm on H2S
< +11 ppm on CO 100 ppm H2:- ~ +30 ppm on CO 5 ppm NO2:- ~ -1 ppm on H2S
~ -1 ppm on CO none known-
Lifetime
garanteed:- 3 years expected:- 3 years
Test gases
zero point:- fresh air H2S sensivity:- 40 ppm H2S in synthetic air CO sensivity:- 40 ppm CO in synthetic air
89
90
11 Technical aspects
Sensor - Propane C3H
8
Sensor data
measurement principle:- combined sensor catalytic com-
bustion and thermal conductivity
measurement range:- 0 – 1.70 vol.% (AL3) in 0.05 vol.%
steps or 0 – 100 %LEL (AL3) in 1 %LEL steps
t90 time:- < 30 seconds
storage temperature:- -20 °C – +60 °C (for repl. sensors)
Alarm thresholds (factory settings)
C3H8:- AL1 = 0.35 vol.% AL2=0.85 vol.%
or AL1 = 20 %LEL AL2 = 50 %LEL
Lifetime
garanteed:- 1 year
expected:- 5 years
Test gases
zero point:- fresh air
sensitivity 1:- 1.00 vol.% (50 %LEL) C3H8 in
synth. air
sensitivity 2:- 100 vol.% C3H
8
11 Technical aspects
Sensor - Nonane C9H20 (⊗)
Sensor data
measurement principle:- combined sensor catalytic combus-
tion and thermal conductivity
measurement range:- 0 – 0.70 vol.% (AL3) in 0.05 vol.%
steps or
0 – 100 %LEL (AL3) in 1 %LEL
steps
t90 time:- < 30 seconds
storage temperature:- -20 °C – +60 °C (for repl. sensors)
Alarm thresholds (factory settings)
C9H20:- AL1 = 0.15 vol.% AL2=0.35 vol.%
or
AL1 = 20 %LEL AL2 = 50 %LEL
Lifetime
garanteed:- 1 year
expected:- 5 years
Test gases
zero point:- fresh air
sensitivity 1:- 0.35 vol.% (50 %LEL)
at 0.30 vol.% propane (replace-
ment gas)
Note:
100 vol.% measurement range not available!
91
Loading...