Setra Systems Multi-Sense 231RS User Manual

®
Multi-Sense
Model 231RS
Multi-Congurable, Wet-to-Wet
Differential Pressure Transducer
Setra’s Model 231RS is the industry’s rst multi-congurable, wet-to-wet dierential pressure transducer utilizing remote sensors. This design reduces labor and material costs versus traditional copper piping installations. The 231RS has a robust, NEMA 4 enclosure with an LCD display and a hinged, captive cover allowing for easy access to switches in order to adjust range and output. An optional display is available that allows users to view high, low, and dierential pressure readings on a simple rotating cycle.
Advantages of Remote Sensors
Remote sensors provide multiple advantages. By connecting the high and low side transducers at the point of measurement instead of running copper piping back to the transducer, the labor and material costs are cut by one-third.
All Inclusive Field Selectable Design
The 231RS has a multi-congurable design, providing the user with eld selectable ranges and outputs as well as push button or remote zero. This design gives the user total exibility to make changes on the job site.
Multiple Connector Options for Added Flexibility
The 231RS oers remote sensors that connect to the unit via armored jacket, cable or conduit tting available in 10, 20, 30, 40 and 50 foot lengths. With the remote sensors, there is no need for a 3 or 5 valve manifold and no risk to compromising the electronics.
Display Options Available
The 231RS has an optional LCD display which gives the user the ability to view the high, low and dierential pressure out­puts locally at the device. The visual indicator gives instant feedback providing performance validation to the user.
n Wet-to-Wet w/ Remote Sensors n Armored Jacket, Conduit, Cable Versions n Optional LCD Display w/ Hinged Cover
Model 231RS Features:
• Remote Sensor Design
• Labor and Material Costs Cut By One Third
• Field Selectable Ranges
• Field Selectable Outputs
• Field Accessible Push Button & Remote Zero
• Jumper Selectable Port Swap
• All Cast Aluminum, NEMA 4 Rated Housing
• CE & RoHS Compliant
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Applications:
• Energy Management Systems
• Process Control Systems
• Flow Measurement of Various Gases or Liquids
• Liquid Level Measurement of Pressurized Vessels
• Pressure Drop Across Filters
®
Multi-Sense
Model 231RS
Multi-Congurable, Wet-to-Wet Differential Pressure Transducer
REMOTE SENSOR DIMENSIONS
Transducer w/Conduit (3M)
1/2 inch Conduit Opening
91.78
3.61
19.4 .76
1/2" CONDUIT
41.07
1.617
1/4" NPT MALE
GENERAL SPECIFICATIONS
Electrical Data (Voltage) Performance Data
Circuit 3-Wire Accuracy RSS4 (at constant temp.)
Excitation
1
Output
Output Impedance 30 Ohms Pressure Ranges (Selection Example, Pg 4.)
Circuit Consumption
Electrical Data (Current)
Circuit 2-wire (Reverse Excitation Protected) RS3 100 50 20 10 100
2
Output
External Load 0 to 250 Ohms RS5 250 125 50 25 250
COVER
Min. Supply Voltage
Max. Supply Voltage
Physical Description Thermal Eects
Case Die Cast Aluminum, Powder Coated Compensated Range °F (°C) +32 to +130 (0 to +54)
Pressure Fittings 1/4-18 NPT Male
Electrical Connection 1/2 in. Conduit Warm-up Shift <0.12% FS
Size 4.0 x 6 x 2 in. (102 x 152 x 51 mm) Response Time 1 to 5 sec. (selectable)
Weight 1.3 lb Proof Pressure 2 x Full Scale
Environmental Data
Operating3 Temperature °F (°C) -4 to +185 (-20 to -85)
Storage Temperature °F (°C) -4 to +185 (-20 to +85)
Vibration 10g from 50Hz to 2000 Hz
Shock 200g
15 to 30 VDC/18 to 30 VAC6 (Reverse Excitation Protected)
Pressure Ranges A, B, C ±1.0% FS
0 to 5 VDC, 0 to 10 VDC, 1 to 5 VDC Pressure Ranges D ±2.0% FS
8 mA (typ.) at 5 VDC, 8 mA (typ) at 10 VDC, 40 mA (typ.) at 18-30 VAC
Range Code A B C D
RS1 50 25 10 5 50
RS2 75 37.5 15 7.5 75
4 to 20 mA RS4 150 75 30 15 150
15 VDC + 0.02 x Resistance of receiver plus line)
30 VDC + 0.004 x Resistance of receiver plus line)
Pressure Media
Liquids or Gases Compatible with 17-4 PH Stainless Steel
Note: Hydrogen not recommended for use with 17-4 PH stainless steel
5
Zero/Span Shift %FS/100°F
(50°C)
Burst Pressure
1
Calibrated into a 50K ohm load, operable into a 5000 ohm load or greater.
2
Calibrated at factory with a 24 VDC loop supply voltage and a 250 ohm load.
3
Operating temperature limits of the electronics only. Pressure media temperatures may be
considerably higher or lower.
4
RSS of Non-Linearity, Hysteresis, and Non-Repeatability.
5
Units calibrated at nominal 70˚F. Maximum thermal error computed from this datum.
Specications subject to change without notice.
6
NOT RECOMMENDED TO CONNECT VAC EXCITATION TO EARTH (SAFETY) GROUND
2.0 (1.8)
15 x Full Scale (50 psi), 10 x Full Scale (75 x 150 psi), 8 x Full Scale (250 PSI)
Max. Line
Pressure
Transducer w/Hirschmann Connector (AJ)
7.3 .29
31.8
1.25
19.4 .76
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
41.07
1.617
1/4" NPT MALE
HIRSCHMAN CONNECTOR
Transducer w/Packard Connector (4M)
3 PIN PACKKARD CONNECTOR
79.2
3.12
19.4 .76