High Accuracy Gauge & Absolute
Pressure Transducer
Setra’s Model 204 is the “standard” for the measuring gauge and
absolute pressure in the test and measurement industry. Decades worth
of installations have helped the 204 build a reputation of reliability and
remains the trusted choice for critical installations. The 204 delivers a high
performance ±0.073% FS accuracy over a wide temperature range which
outperforms competitive transducers in the mid to high pressure market.
The 204 offers multiple options to meet both simple and demanding
application requirements that are not provided on competitive transducers.
Long-term reliability
The Model 204 pressure transducer uses a simple and reliable variable
capacitance sensor design. The 204 provides repeatable and dependable
readings in rugged applications through its efcient sensor design.
Accuracy & performance
The Model 204 is a test and measurement grade transducer for mid to
high pressure ranges. The 204 covers a large selection of pressure ranges
with ±0.073% FS accuracy over a wide temperature range. The Model 204
provides response time of <1 ms, exceeding the performance of many
competitors.
Features
• Ideal for high accuracy applications
• Excellent thermal effects
• Highly congurable design
• ±0.073% FS accuracy
• 0-5 VDC
• Fast response, less than 1 ms
• Low output noise
• Solid one-piece stainless steel sensor
• Meets CE conformance standards
Applications
• Hight accuracy general purpose
• R&D test and measurement
• Vacuum systems
• Dynamometers
• Engine test cells
Customization is standard
Unlike many competitors, the 204 offers many mechanical and electrical
options that can be integrated into existing system designs. These options
reduce engineering design time, allowing for earlier project completion
and quicker time to market.
The Setra Systems name and logo are registered trademarks of Setra Systems Inc.
Model 204 I High Accuracy Gauge & Absolute Pressure Transducer
Specications
Performance data
Accuracy RSS
(at constant temp)
Non-linearity (BFSL)±0.07% FS
Hysteresis
Non-repeatability0.02% FS
1
±0.14% for 10,000 PSIG
0.1% for 10,000 PSIG
±0.11% FS
0.08% FS
Physical description
Pressure ttings1/4” -18 NPT internal
Excitation
Output0 to 5 VDC
Power consumption10 mA (0.25 Watts)
Output impedance<10 Ohms
Output noise<100 Microvolts RMS (0 Hz to KHz)
RSS of Non-Linearity, Hysteresis and Non-Repeatability
2
Units calibrated at nominal 70°F
3
Approximately 50% higher for 0-14.7 psiv range
4
Calibrated into 50K ohm load. Operable into 5000 ohms or greater.
5
Zero output factory set to within ±10mV. Span (Full Span) output factory set to within ±10mV.
Note: Both output leads are normally 1.6 VDC above the negative excitation lead at zero pressure.
Either negative excitation or negative output should be connected to case (ground). But both leads
cannot be connected to case (ground). Unit is calibrated at the factory with the negative excitation
connected to case (ground.)
6
Calibrated at factory with a 24 VDC loop supply voltage and a 250 ohm load.
Specications subject to change without notice.
22 to 30 VDC, 24 VDC (normal)
Reverse excitation protected
Environmental data
Operating temperature3 °F(°C)0 to +175 (-18 to +80)
Storage temperature °F(°C)-65 to +250 (-55 to +120)
Model 204 I High Accuracy Gauge & Absolute Pressure Transducer
Ordering information
Example part number: 2391005PB1F2S02WNN
Model 239, ±5 PSID pressure range, 1/8” NPT Int. tting, ±2.5 VDC, 2’ Cable Length, ±0.14% FS Accuracy, No Options.
242F01
[1][2][3][5][7][6]
[1][6][7][3][4]
Model
2041 Model 204
1
Units with pressure range > 5,000 PSI have accuracy of ±0.14% FS only.
2
Both boxes must be lled in alphanumeric order:
-If No options: N + N
-If 1 option: Option Code + N
-If 2 options: Option Code + Option Code
3
2x Thermal Effects Specication
4
Options M, R and S will have Y1 Cable as STD.
Note: Setra adheres to strict quality standards including
ISO 9001 and ANSI-Z540-1. The calibration of this product is NIST traceable.