Model 204
High Accuracy Gauge & Absolute
Pressure Transducer
Setra’s Model 204 is the “standard” for measuring gauge and
absolute pressure in the Test and Measurement industry.
Decades worth of installations have helped the 204 build a
reputation of reliability and remains the trusted choice for
critical installations. The 204 delivers a high performance
±0.073% FS accuracy over a wide temperature range which
outperforms competitive transducers in the mid to high
pressure market. The 204 oers multiple options to meet
both simple and demanding application requirements that
are not provided on competitive transducers.
Long-Term Reliability
The Model 204 pressure transducer uses a simple and reliable variable capacitance sensor design. The 204 provides
repeatable and dependable readings in rugged applications
through its ecient sensor design.
Accuracy & Performance
The Model 204 is a Test and Measurement grade transducer
for mid to high pressure ranges. The 204 covers a large selection of pressure ranges with ±0.073% FS accuracy over a wide
temperature range. The Model 204 provides response time
of <1 ms, exceeding the performance of many competitors.
Customization is Standard
Unlike many competitors, the 204 oers many mechanical
and electrical options that can be integrated into existing
system designs. These options reduce engineering design
time, allowing for earlier project completion and quicker
time to market.
n
Ideal for High Accuracy Applications
n
Excellent Thermal Effects
n
Highly Congurable Design
Model 204 Features:
• ±0.073% FS Accuracy
• 0-5 VDC
• Fast Response, Less than 1 ms
• Low Output Noise
• Solid One-Piece Stainless Steel Sensor
• Meets CE Conformance Standards
Applications:
• High Accuracy General Purpose
• R&D Test and Measurement
• Vacuum Systems
• Dynamometers
• Engine Test Cells
Model 204
High Accuracy Gauge & Absolute Pressure Transducer
ORDERING INFORMATION
2 0 4 1
Model Pressure Ranges Pressure Fitting Output Electrical Termination Accuracy1Options
2041=Model 204 Gauge Pressure Absolute Pressure 2F 1/4” NPT Female 2B20-5 VDC 02 2’ Cable
025PG 0-25 PSIG 025PA 0-25 PSIA 2Y 0-2.5 VDC
050PG 0-50 PSIG 050PA 0-50 PSIA 27 1-5 VDC
100PG 0-100 PSIG 100PA 0-100 PSIA 28 1-6 VDC
250PG 0-250 PSIG 250PA 0-250 PSIA 2C 0-10 VDC
500PG 0-500 PSIG 500PA 0-500 PSIA 2U 1-10 VDC
10CPG 0-1000 PSIG 10CPA 0-1000 PSIA
30CPG 0-3000 PSIG 30CPA 0-3000 PSIA
50CPG 0-5000 PSIG 50CPA 0-5000 PSIA
10KPG 0-10000 PSIG R4Remote Calibration (Adjustable)
Z01PV 0-14.7 PSI
(VACUUM)
Ordering Example: 2041025PG2F2B02WNN, Model 204, Ranges 0-25 PSIG, 1/4” NPT Female, 0-5 VDC, 2’ Cable, ±0.1% FS Accuracy.
2 F
10 10’ Cable 9 ± 0.073% FS 7 Clean for Oxygen
25 25’ Cable D Mate with Datum
Y1 2’ Red Cable 9-Conductor 30 AWG E Special Excitation Voltage ± -24 VDC
1
Units with pressure range > 5,000 PSI have accuracy of ±0.14% FS only.
2
Both boxes must be lled in alphanumeric order:
-If No options: N + N
-If 1 option: Option Code + N
-If 2 options: Option Code + Option Code
3
2x Thermal Eects Specication
4
Options M, R and S will have Y1 Cable as STD.
Note: Setra adheres to strict quality standards including
ISO 9001 and ANSI-Z540-1. The calibration of this product is NIST traceable.
W ± 0.11% FS
2
33Compensated Temperature Range (-65 to 250°F )
F NEMA4 Enclosure
G Special Excitation Voltage ± -15 VDC
L Etched SS Tags
4
Remote Full Scale Sensitivity
M
N None
4
S
Remote Calibration Adjustable (Fixed)
DIMENSIONS GENERAL SPECIFICATIONS
Performance Data Physical Description
PROOF PRESSURE
Pressure Ranges
0 PSIA or
0 PSIG to:
100 150 500 3.5
250 375 1000 5.0
500 750 1500 8.0
1000 1250 3000 11.0
3000 3750 4500 15.0
5000 6000 7500 25.0
10,000 PSIG only 11,000 12,500 30.0
0-14.7 PSIV 50 150 2.0
Proof Pressure
25 50 150 2.0
50 75 200 2.5
(PSI)
Burst Pressure
Rating (PSI)
Approx. Natural
Frequency (KHz)
Accuracy RSS1
(at constant temperature)
Non-Linearity (BFSL) ±0.07% FS Excitation 22 to 30 VDC, 24 VDC (Normal)
Hysteresis 0.08% FS
Non-Repeatability 0.02% FS Power Consumption 10 mA (0.25 Watts)
Thermal Eects
2
Zero Shift3 %FS/100°F (%FS/50°C) <±0.4 (<±0.36) Output Noise <100 Microvolts RMS (0 Hz to 10 KHz)
Span Shift %FS/100°F (%FS/50°C) <±0.3(<±0.27)
Static Acceleration Eect <0.05 PSI/G (Typ.)
Volume Increase Due to FS Pressure 5 x 10-5 cu. in.
Warm-Up Shift ±0.5% Total (±0.1% Residual
Environmental Data
Operating Temperature °F(°C) 0 to +175 (-18 to +80)
Storage Temperature °F(°C) -65 to +250 (-55 to +120)
Vibration 2g from 5 Hz to 500 Hz
Shock 50g
Acceleration 10g Maximum
±0.11% FS
±0.14% for 10,000 PSIG
0.1% for 10,000 PSIG
(Pressure Port Axis)
Shift after 5 Minutes)
Pressure Fitting 1/4” - 18 NPT Internal
Reverse Excitation Protected
Output
4
0 to 5 VDC
5
Output Impedance <10 Ohms
1
RSS of Non-Linearity, Hysteresis and Non-Repeatability
2
Units calibrated at nominal 70°F
3
Approximately 50% higher for 0-14.7 psiv range
4
Calibrated into 50K ohm load. Operable into 5000 ohms or greater.
5
Zero output factory set to within ±10mV. Span (Full Span) output factory
set to within ±10mV. Note: Both output leads are normally 1.6 VDC above
the negative excitation lead at zero pressure. Either negative excitation
or negative output should be connected to case (ground). But both leads
cannot be connected to case (ground). Unit is calibrated at the factory with
the negative excitation connected to case (ground.)
6
Calibrated at factory with a 24 VDC loop supply voltage and a 250 ohm load.
Specications subject to change without notice.
SSP204 REV. I 07/17