SENSIT Technologies CO, HCN Quick Start Instructions

Page 1
Quick-Start Instructions
IMPORTANT: After testing ue samples
it may be necessary to remove condensate collected in the water/particle lter that is included with the probe.
WARNINGS:
To prevent ignition of ammable or combustible atmospheres, disconnect power before servicing.
This is not a substitute for the instruction manual. It is for reference only.
Read and understand the instruction manual before use.
MADE IN THE USA
WITH GLOBALLY SOURCED COMPONENTS
Distributed by:
851 Transport Drive Valparaiso, IN 46383-8432
Phone: 219 465 2700 Web: www.gasleaksensors.com
Innovative Detection Solutions
Page 2
1. Install the batteries by pushing down the locking tab and sliding the handle away from the top of the instrument. When replacing handle be sure the tab is securely in place.
2. IMPORTANT: CAREFULLY OBSERVE POLARITY WHEN CHANGING BATTERIES. Incorrect installation can damage an internal “factory service only” fuse.
3. Locate button “A” – push & hold until the unit powers up, then release the power button.
5. If FAIL is displayed for any of the sensor readings, make sure the instrument is in clean air; push and hold the “C” button until AUTO ZERO is displayed. If this process does not clear the fail on the display, this could indicate a problem with the instrument or sensor.
6. Look at the display –CO (and optionally O2) readings are displayed.
7. Attach the hot air probe assembly to the ¼ turn twist connection on the back.
8. Place your nger over the inlet of the probe and wait ~5 seconds for “FLOW BLOCKED” to appear on the display. Check for tight connections if it does not show “FLOW BLOCKED”.
9. You are now ready to use the instrument. You can now enter the area and detect gases.
10. When performing CO checks in high temp areas do not exceed prolonged exposures (over 5 min) at 500°F.
11. Press and hold the “C” button to zero the instrument. (only in a gas free area).
12. When your investigation is complete push button “A” and hold for 5 seconds until the instrument displays “POWER OFF” then release to shut off.
SENSIT® CO
QUICK-START INSTRUCTIONS