Sensa Core Gluco Spark User Manual

ISO
15197:2013
ISO
9001:2015
ISO
13485:2016
SC-GLSPL-0219A
IVD
Table of Contents
Important Information About Your
Gluco Spark Blood Glucose Monitoring System ....................................... 1
Alternate Test Site.......................................................................................... 2
Gluco Spark Meter Display........................................................................... 4
Gluco Spark Test Strip ................................................................................... 5
Testing Your Blood Glucose
Testing Your Blood Glucose.... .................................................................... 6
1. Performing a Blood Glucose Test ........................................................... 6
2. Filling the Gluco Spark test strip with blood ...........................................7
3. Waiting for result ....................................................................................... 8
Strip Ejection
Strip Ejection .................................................................................................. 8
Lancing device handling procedure
Lancing device handling procedure ........................................................
Obtaining a blood sample .......................................................................10
Safety Lancets handling Procedure ........................................................11
Low and Hi Readings
Low & High Readings .................................................................................12
Meter Maintenance
Replacing the battery ...............................................................................13
Time and Date set-up ................................................................................14
Important Notes
Caring & Cleaning for your meter ............................................................15
Handling & Storage ....................................................................................15
Error Message ..............................................................................................16
Troubleshooting ...........................................................................................17
Technical Specifications
Gluco Spark System Specifications ..........................................................18
Explanation of Symbols ..............................................................................19
Log book tables ......................................................................................... 20
System trouble shooting
9
.
Important information about your Gluco Spark blood glucose monitoring system
The Gluco Spark system is intended for use outside the body (in other words, in vitro diagnostic use only).
The Gluco Spark system is used for measuring blood glucose (also known as blood sugar) from samples taken from the forearm, upper arm, hand, thigh, calf or fingers.
Important:
The Gluco Spark meter should only be used with Gluco Spark test strips. Using other test strips with this meter can produce inaccurate results.
Do use the Gluco Spark system for:
Testing blood glucose.
Testing only with fresh capillary whole blood samples.
Warning:
The Gluco Spark system contains small parts that may be dangerous if swallowed.
Caution:
Please read all the instructions provided in this owner 's booklet and practice the testing procedures before using the Gluco Spark system. Blood glucose monitoring should be done with the guidance of a healthcare professional.
Important health-related information
Severe dehydration and excessive water loss may cause false low results. If you believe you are suffering from severe dehydration, consult your physician or doctor immediately.
Test results below 70 mg/dL mean low blood glucose (hypoglycemia).
Test results greater than 180 mg/dL mean high blood glucose (hyperglycemia).
If you get results below 70 mg/dL or above 180 mg/dL and do not have symptoms of hypoglycemia or hyperglycemia, repeat the test. If you have symptoms or continue to get results that fall below 70 mg/dL or above 180 mg/dL, follow the treatment/advice of your healthcare professional.
If you are experiencing symptoms that are not consistent with your blood glucose test and you have followed all instructions described in the Gluco Spark owner's booklet, consult your healthcare professional.
01
Gluco Spark ystem uses 0. sample. s 5 µL
So you can test on a variety of locations
Alternate Test Sites
Suggested Test Areas for the Hand
The human figure above shows the areas where you can test with the Gluco Spark System. Avoid moles, veins, bones, and tendons.
Warning:
We recommend that you test on your fingers for better results.
02
Upper arm
Forearm
Hand
Fingertips
Thigh
Calf
Gluco Spark meter and how it works
This screen always appears when the meter is turned on so that you can make sure the
14:33 05-01-19
display is working properly. Do not use the meter if the system check screen does not match the example here.
(NO) Button
Exit
Strip Ejector
Press this button to release the used strip after testing.
Sample Area
Apply blood to sample area at end of the strip only.
Carbon Electrode
(for accurate results)
Reference Electrode
Plasma Etching
(for excellent precision)
Fast Draw Technology
(fast and easy)
Sample Application Area
03
mg/dL
14:33 05-01-19
(Yes) Button
Previous Readings.
Display Screen
Di spl ays yo ur t est re su lt s a nd ot he r important information.
Test Strip Port
Insert the top end of a new Gluco Spark test strip here. The meter powers on when you insert the test strip.
Sensa Core
®
Gluco Spark Meter Display
Blood Drop Symbol
This symbol works with st ri p s ym bol in an animation to tell you when the meter is ready to apply blood.
Test Result Area
Displays test results.
14:33 05-01-19
Time and Date
Shows Time and Date of the Test
Battery Symbol
Strip Symbol
This symbol works wi th Blood Drop symbol in an animation to indicate that the meter is ready to apply blood.
mg/dL (U.S. Standard)
Indicates unit of measurement.
04
Gluco Spark test strip
Apply blood to strip sample area per test.
Important:
Test strips may be used only once. Discard used test strips. You can continue to fill the test strip for up to 60 seconds.
Warning:
The cap or vial contains drying agents that may be harmful if inhaled or swallowed and may cause skin or eye irritation.
Important:
The Gluco Spark Meter should only be used with Gluco Spark Test Strips. Using other test strips with this meter can produce inaccurate results.
Important Test Strip Information
The Gluco Spark system has an operating range of 2° to 50°C.
Store the test strip package in a cool, dry place between 0° to 50°C.
Use test strips only within the system operating temperature range.
Keep away from direct sunlight and heat.
Store your test strips in their original vial only; never transfer them to another vial or any other container.
Never store individual test strips outside the vial.
After removing a Gluco Spark test strip from the vial, immediately close the vial cap tightly.
With clean, dry hands, you may touch the test strip anywhere when removing it from the vial or inserting it into the meter.
Do not use test strips beyond the expiration date. This may cause inaccurate results.
Do not bend, cut, or alter Gluco Spark test strips.
Note:
For further information, refer to the Gluco Spark Test Strip package insert.
05
Test Strip
Sample Area
Testing your blood glucose
14:33 05-01-19
mg/dL
Important:
To ensure accurate results, wash your hands and test site (for example, your finger) with soap and water. Make sure
is no cream
there
your hands and test site.
With the Gluco Spark meter, you can test your blood glucose on your forearm, fingers, upper arm, hand, thigh, or calf.
Testing w ith the Glu co Spark me ter requi res a very sm all sampl e size
0.5 µL of blo od; about t he size of pi nhead
Note
We recommend that you test on your fingers for better results.
or lotion on the test site. Thoroughly dry
Times when you may be checking for low blood glucose include:
When you are feeling symptoms such as weakness, sweating, nervousness, headache, or confusion.
When you have delayed a meal after taking insulin. When your healthcare professional advises you to do so.
1 Performing a blood glucose test
Prepare your meter
Insert a Gluco Spark Test Strip into the meter till it allows. The meter will power on.
Note:
If you do not start the test within one minute, the meter will turn off. To restart the meter, take out the unused Gluco Spark test strip and reinsert it into the meter.
System check screen
14:33 05-01-19
System Check
Blood Drop and Test Strip Symbols
When the meter powers on, this screen will appear so that you can make sure the display is working properly.
The blood drop and test strip symbols will appear on the display screen. Your meter is now read y to ap ply blo od sa mple.
Important:
The Gluco Spark meter should only be used with Gluco Spark test strips. using other test strips with this meter can produce inaccurate results
06
Obtain a blood sample
2
Select a test site. There are differences in testing on fingers versus other alternative sites. Use your lancing device to obtain a blood sample. Refer to the detailed instructions on how to use the Gluco Spark lancing device. ( Page 9 to 11 )
Note
We recommend that you test on your fingers for better results.
3 Filling the Gluco Spark test strip with blood
a.
Make sure that the Gluco Spark Test Strip is i n the meter and the meter is powered on. (If the meter is off, take out the test strip and reinsert it into the meter.) You are now ready to apply the blood sample.
b.
Bring the Gluco Spark Test Strip to the blood sample at a slight angle.
c.
The Gluco Spark Test Strip acts like a sponge and pulls the blood into the strip through the sample area.
Do not press the Gluco Spark test strip against the test site.
Do not scrape the blood onto the Gluco Spark test strip.
Do not apply blood to the top side of the Gluco Spark test strip.
Do not apply blood to the Gluco Spark test strip when the Gluco Spark test strip is out of the meter.
Do not put blood or foreign objects into the Gluco Spark test strip port.
07
Waiting for result
14:33 05-01-19
mg/dL
4
Strip Ejection
Do not pull the Gluco Spark test strip away until you hear one beep or see the timer starting on your meter screen. this means you have enough blood and the meter is reading your glucose.
If after five seconds your meter does not show a timer starting on the display screen, the sample may be too small or strip might be faulty. Replace the strip.
The test result is complete when your blood glucose test result is shown on the display screen and you will hear two beeps.
On a ver ag e, yo ur r ead in g wil l disp lay i n 5 se con ds .
After getting the result, Press the strip ejector button near a dustbin so that the used strip gets released from the meter and falls in the dustbin.
08
Instructions for use of Safety Lancets
How to use
1
2
3
11
LO and HI Readings
Important:
Low or high blood glucose readings can indicate a potentially serious medical condition. The Gluco Spark Meter displays results from 20 to 800 mg/dL
LO Reading with Symptoms
If you have a reading below 70 mg/dL a nd have symptoms such as w eakness, sweati ng, nervousness , he adache o r confusi on, then fo llow your d octor's r ecommen dation to t reat hypogly cemia.
HI Reading with Symptoms
If your test result is above 800 mg/dL, this indicates severe hyperglycemia (high blood glucose).
If you feel symptoms such as fatigue, thirst, excess urination, or blurry vision, then follow your doctor's recommendation to treat hyperglycemia.
HI Reading without Symptoms
If you g et a HI read ing, b ut have no s ympt oms of hig h bloo d gluc ose, th en re -test wi th a new t est str ip. I f you st ill g et a HI read ing, follo w you r doc tor's reco mmendat ion t o tre at hype rglyce mia.
12
Meter Maintenance
Battery
Your Gluco Spark meter comes with one preinstalled, 3 volt, lithium battery. The battery provides enough power for the meter to perform about 1000 Tests. If your battery runs low, the battery needs to be replaced.
Replacing the Battery
Open the battery cover as shown in the image
Tilt the meter to sideways to release the battery
13
Insert a new battery with “+” symbol on the top
Close the battery cover
Date and Time Set-Up (On Replacement of Battery)
Your Gluco Spark meter allows you to set the Date and Time as you like. You can view 24 hour time.
To get into set-up mode, remove the battery and replace it. A complete display will briefly appear, then numbers will flash in the bottom left corner of the display.
Steps to set the Year, Date :
Remove battery and replace it to set date and time.
5-01-19
0
14:33
5-01-19
0
14:33
5-01-19
0
14:33
Setting the Time
05-01-19
00: 00
: 02
05-01-19
00
Note:
To quickly adjust a certain number, keep pressing the > button until the one you
1.
want appears.
2.
At any steps if you want to shut the meter off, Press< button until you complete the cycle and all setting still cab be automatically saved.
With the number flashing in the position of Year on the display. Press “>” (forward) button until you reach the current year. Press “<” (backward) button to set.
The number in the month flashes on the display. Press “>” (forward) button until you reach the current month. Press “<” (backward) button to set.
The number in the day position flashes on the display. Press “>” (forward) button until you reach the current day. Press “<” (backward) button to set.
The minute will now flash. Press “>” (forward) button to select the current minute. Press “<” (backward) button to set.
The number in the hour position will now flash. Press “>” (forward) button until you reach the current hour. Press “<” (backward) button to set.
14
Important Notes
Caring for your meter
Taking care of your meter is very simple. Just follow the guidelines shown below
1.
Do not disassemble your meter. If you have any questions, please contact your local authorized distributor of Gluco Spark Blood Glucose Monitoring System.
2.
Please use your meter carefully. Avoid throwing, shaking, or dropping it, which may damage its internal parts.
3.
Clean the meter’s surface with a soft tissue slightly dampened with water. Do not use alcohol or abrasive solutions.
4.
Do not wash or pour liquid on the meter.
Cleaning of your meter
Avoid getting dirt, dust, blood, water, or any other liquid in the meter's test strip and data ports. Clean the outside of the meter using a cloth dampened with one of the following:
Mild detergent/mild soap and water. 70% isopropyl alcohol.
A mix of one part household bleach, nine parts water.
Important:
Never immerse the meter in water or any other liquid.
Handling & Storage
Handle your meter with care and protect it from direct sunlight or
1.
extremely high or low temperature. Do not expose your meter and test strips to an environment of high
2.
humidity, such as in bathrooms, kitchens, etc. Keep your meter free of dust.
3.
It is recommended that you use the carrying case that is designed to
4.
store and protect your Blood Glucose Monitoring System. Put your system in an appropriate operating environment at least
5.
30 minutes before testing.
15
System troubleshooting
Error messages
When an error occurs, an error message appears on the display with continuous beep sound. Turn the meter off by ejecting the strip.
If your meter has been dropped, this can also lead to error messages. If error messages occur frequently, contact your customer support and service centre.
Error Message Cause and Solution
• You slid a used test strip into the meter.
• You applied blood or control solution to the test strip too soon, i.e. before the drop symbol flashed on the display. Start the blood glucose test from the
beginning with a new test strip.
When strip is ejected from the meter after applying the blood sample
Start the blood glucose test from the beginning with a new test strip and wait until the result appears on the screen.
16
Troubleshooting
Meter does not enter the test mode after inserting a test strip.
Probable cause
The battery is dead. Replace the battery.
The battery is installed incorrectly or there is no battery in the meter.
Test strip inserted upside down, wrong end in, or incompletely inserted into the meter.
Test does not start after applying the blood sample.
Probable cause
Blood sample too small.
Defective test strip.
Sample applied after meter times out (one minute) and turns off.
Sample applied before the appearance of blood sign
What to do
Check that the battery is installed correctly with the positive (+) sign facing upward, towards you.
Insert the Gluco Spark Test Strip with the white sheet with sample inlet side up and the top of the
strip in the meter.
What to do
Replace the Strip and apply enough Sample. Do not apply blood to the strip with small Sample.
Repeat the test with a new Gluco Spark test strip
Repeat the test using a new test strip. Wait until you see the Blood and Test Strip symbols on the display screen before you apply the blood sample.
17
Technical Specifications
Gluco Spark system specifications
Test Principle
Assay Method
Automatic Shutoff
Hematocrit
Measurement Units
Strips Storage Temp.
Operating Temperature
Operating Relative Humidity
Memory
Power Source
Battery Life
Result Range
Sample
Sample Size
Meter Size (Centimeters)
Display Size (Millimeters)
Test Time
Weight
Strip Ejection
Construction
Protection Class
Determination of glucose in fresh capillary whole blood
Amperometric Bio sensor
After 5 seconds, depending on operating status
15 % to 65 %
mg/dL
0 0
0 to 50 C (Do not refrigerate)
0 0
2 to 50 C
10 % to 95 % (non-condensing)
500 Previous Readings
One 3 volt lithium battery (CR2032), replaceable
Approximatly 1000 tests
20 to 800 mg/dL
Capillary whole blood
0.5 µL (500 nanoliters)
Length 10.0 x Width 3.1 x Depth 1.4
48 (diagonally) LCD
5 seconds
34 grams
Yes
Handheld
B
18
Blood
OTHER NOTES
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Target Range for Blood Sugar set with your Healthcare Professional
Wakeup Time _____________ Before Meals _____________ After Meals _____________ At Bed Time _____________
BREAKFAST LUNCH DINNER BEDTIME
Blood
Date
Medication
Glucose
M
T
W
20
F
T
S
S
Avg.
Blood
OTHER NOTES
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Target Range for Blood Sugar set with your Healthcare Professional
Wakeup Time _____________ Before Meals _____________ After Meals _____________ At Bed Time _____________
BREAKFAST LUNCH DINNER BEDTIME
Blood
Date
Medication
Glucose
M
T
W
21
F
T
S
S
Avg.
SENSA CORE
Medical Instrumentation Pvt. Ltd.
Plot No: 3, Export Promation Industrial Pack, Pashamylaram, Sangareddy (Dist.), Hyderabad - 502307, Telangana, INDIA.
Tel: +08455 - 223400, www.sensacore.com
Customer Care Number
1800 200 2880