EN Using additional functions
Using additional functions
The following additional functions are not required for normal use of the wireless headphone system. You can, however, extend the potential use of the system and adjust some of the functions to your individual needs.
Pairing one or several stethoset receivers with the transmitter
The stethoset receiver and the transmitter of your wireless headphone system are factory pre-paired. The following steps are only necessary if you want to pair additional ste­thoset receivers with the transmitter.
You can simultaneously use up to 4 paired stethoset receivers with your transmitter. All other paired and switched-on receivers cannot connect to the transmitter.
If more than one stethoset receiver is connected to the transmitter, the remaining operating time/charge status and the hearing profile cannot be displayed. In addition, the balance and a hearing profile cannot be adjusted. These functions can only be used/adjusted when only one stethoset receiver is connected to the transmitter.
Place the stethoset receiver into the transmitter’s charging
compartment. The stethoset receiver is paired with the transmitter.
+
Advanced settings
The advanced settings allow you to adjust the wireless head­phone system to your individual needs and to protect the set­tings against accidental changes.
Activating the protected settings mode
You can change the advanced settings by activating the pro­tected settings mode:
24 | RS 5000
Page 2
RS_5000_570727_0117_PRT.book Seite 25 Montag, 10. Oktober 2016 2:52 14
Using additional functions EN
1 Disconnect the transmitter from the power supply system
by unplugging the connector of the power supply unit.
2 Keep the Hearing profile button pressed and connect the
transmitter to the power supply system by connecting the connector of the power supply unit to the yellow socket of the transmitter.
21
DEFRES EN
The protected settings mode is activated. The
LED flashes blue. The LED segments indicate the cur-
status
Transmitter
rent settings as specified in the following chapters:
• battery status display (see page 25)
• charging reminder function (see page 26)
• hearing profile and balance protection (see page 27)
Deactivating the protected settings mode and saving the advanced settings
To save the advanced settings and to deactivate the pro­tected settings mode:
3 Disconnect the transmitter from the power supply system
by unplugging the connector of the power supply unit.
3
Battery status display
This setting is saved on the transmitter, independent of the stethoset receiver you are using.
To change the battery status display:
Make sure that the protected settings mode is activated
(see page 24).
RS 5000 |25
Page 3
RS_5000_570727_0117_PRT.book Seite 26 Montag, 10. Oktober 2016 2:52 14
EN Using additional functions
Press the Balance button L.
Status and display of the
Battery
LED
permanently activated (operation and charging)*
Function/behavior
The current charge status is permanently displayed during charging.
The remaining charge status is
st
1
segment lit
activated for 30 sec­onds (operation and charging)
permanently displayed during operation.
The current charge status is displayed for 30 seconds during charging. Then all LEDs go off. The charging process is continued.
st
1
segment off
The remaining charge status is displayed for 30 seconds during operation. The battery status display then goes off.
* factory setting
Save the setting by disconnecting the transmitter from the
power supply system (see page 25).
Charging reminder function
This setting is saved on the transmitter, independent of the stethoset receiver you are using.
To activate or deactivate the function:
Make sure that the protected settings mode is activated
(see page 24).
Press the Balance button R.
26 | RS 5000
Page 4
RS_5000_570727_0117_PRT.book Seite 27 Montag, 10. Oktober 2016 2:52 14
Using additional functions EN
Status and display of the charging
Function/behavior
reminder function
activated*
When you switch the stethoset receiver off and the transmitter switches to standby mode, the
rd
3
segment lit
battery status display will flash for 10 minutes to remind you to charge the stethoset receiver.
deactivated
After switching the wireless head­phone system off, all LEDs remain off.
rd
3
segment off
* factory setting
Save the setting by disconnecting the transmitter from the
power supply system (see page 25).
Hearing profile and balance protection
This setting is saved on the stethoset receiver, independent of the transmitter you are using.
Make sure that your stethoset receiver’s balance setting
and hearing profile are correctly adjusted (see page 17 and
18).
Make sure that the protected settings mode is activated
(see page 24).
Place the stethoset receiver into the transmitter’s charging
compartment.
Press the volume button + on the stethoset receiver to acti­vate or deactivate the hearing profile and balance protection.
DEFRES EN
+
Status and display of the protection function
activated
nd
segment lit
2
+
Function/behavior
The activated hearing profile and the balance setting cannot be changed/ deactivated.
RS 5000 |27
Page 5
RS_5000_570727_0117_PRT.book Seite 28 Montag, 10. Oktober 2016 2:52 14
EN Using additional functions
Status and display of the protection
Function/behavior
function
deactivated*
The hearing profiles can be changed at any time by pressing the Hearing profile button for 2 seconds.
nd
segment off
2
Save the setting by disconnecting the transmitter from the
The balance setting can be changed.
* factory setting
power supply system (see page 25).
If you are using several stethoset receivers and want to protect the activated hearing profiles from accidental adjustment, you can place the stethoset receivers, one after the other, into the transmitter’s charging compart­ment to activate the hearing profile and balance protec­tion.
Restoring the factory default settings
To restore the transmitter to the factory settings:
1 Disconnect the transmitter from the power supply system
by unplugging the connector of the power supply unit.
2 Simultaneously press and hold the buttons L, and R. 3 Connect the transmitter to the power supply system by
connecting the connector of the power supply unit to the yellow socket of the transmitter while keeping the 3 but­tons pressed for at least 3 seconds. All LEDs on the transmitter flash. The transmitter settings are restored to the factory settings.
1
2
3
28 | RS 5000
Page 6
RS_5000_570727_0117_PRT.book Seite 29 Montag, 10. Oktober 2016 2:52 14
Using additional functions EN
To restore a stethoset receiver to the factory settings:
Place the stethoset receiver into the transmitter’s charging
compartment. The
Receiver status
LED flashes green. The settings of the
stethoset receiver are also restored to the factory settings.
+
If you place the stethoset receiver into the transmitter’s charging compartment at the very beginning of the restore procedure, the factory settings of the transmit­ter and the receiver are restored simultaneously.
Disconnect the transmitter from the power supply system
by unplugging the connector of the power supply unit. The factory settings are restored and you can use the wire­less headphone system again.
DEFRES EN
RS 5000 |29
Page 7
RS_5000_570727_0117_PRT.book Seite 30 Montag, 10. Oktober 2016 2:52 14
EN If a problem occurs ...
If a problem occurs ...
Problem
does not
Transmitter
Stethoset receiver
cannot be switched on
No sound
No sound when a digital
audio source is connected
Occasional sound dropouts,
it might be that you hear 5 beeps
Possible cause Possible solution
No connection to the power supply system.
switch on
The rechargeable battery is empty.
The rechargeable battery is deep discharged.
No connection to the power supply system.
The audio plug is not properly connected.
The audio source is switched off.
The volume of the analog audio source is either set to the minimum or is on mute.
The audio cable is defective. Replace the audio cable. – The stethoset receiver is not
correctly paired with the transmitter (e.g. additional stethoset receiver).
Cables are connected to both the digital and the analog audio input.
The connected digital audio source uses an incompatible audio format. The
Transmitter status
flashes white and blue.
The stethoset receiver is out of the transmitter’s range.
The signal is shielded.
There are interfering devices (e.g. WiFi router, Bluetooth devices or microwaves) in
in the stethoset receiver
the vicinity.
LED
Check the connections of the power supply unit.
Recharge the rechargeable battery.
Recharge the rechargeable battery for several hours. If the rechargeable battery is deep discharged, the charging process might take a while to start (up to 1 hour).
Check the connections of the power supply unit.
Check the audio plug connection.
Switch the audio source on.
Increase the volume on the audio source to at least a medium level/deactivate the muting function on the audio source.
Pair the stethoset receiver with the transmitter again.
Remove the cable from the analog audio input.
Set the audio format of your audio source to “PCM”, with a sampling rate of max. 96 kHz (see the instruction manual of the audio source).
Reduce the distance between the stethoset receiver and the transmitter.
Remove obstacles between the transmitter and the stethoset receiver.
Change the position of the transmitter or of the receiver.
Separate the transmitter from other wireless devices by at least 50 cm/20”.
Page
10
11
10
6
24
6
6
30 | RS 5000
Page 8
RS_5000_570727_0117_PRT.book Seite 31 Montag, 10. Oktober 2016 2:52 14
If a problem occurs ... EN
Problem
Possible cause Possible solution
The volume on the stethoset receiver is adjusted to low.
The volume of the analog audio source is adjusted to low.
Sound too low
The balance is misadjusted. Adjust the balance. 17 The audio cable is not
properly connected.
The TV has only a mono
on one ear
Sound only
headphone socket
The audio cable is defective. Replace the audio cable. – The signal of the audio
source is distorted.
Sound is
The volume on the stethoset
distorted
receiver is adjusted to high.
Some TVs can adjust a delay or latency of the audio sig­nal for the digital audio out­put.
receiver with a TV
using the stethoset
You hear echoes when
The battery status display is set so that it goes off after 30 seconds.
More than one stethoset receiver is currently
goes off
connected to the
Battery status/
transmitter.
charge status display
Increase the volume on the stethoset receiver.
Increase the volume of the analog audio source to at least a medium level (approx.
).
1 V
rms
Check the audio plug connection.
Use an adapter (mono to stereo jack socket) to connect the transmitter to the TV.
Reduce the volume of the audio source.
Reduce the volume on the stethoset receiver.
Check the audio output set­tings of your TV and set the latency to “0”.
Set the battery status display so that it is permanently displayed.
This function is only possible when only one stethoset receiver/receiver is connected to the transmitter.
Page
16
6
DEFRES EN
16
25
The rechargeable battery is almost empty.
during
2 beeps
operation
The rechargeable battery is exhausted.
time
decreases
Operating
The hearing profile and balance protection is activated.
More than one stethoset receiver is currently connected to the transmitter.
A hearing profile or the balance
cannot be activated or changed
Recharge the rechargeable battery.
Have the rechargeable battery replaced by an authorized Sennheiser service partner.
Deactivate the hearing profile and balance protection.
This function is only possible when only one stethoset receiver/receiver is connected to the transmitter.
RS 5000 |31
11
23
27
Page 9
+
RS_5000_570727_0117_PRT.book Seite 32 Montag, 10. Oktober 2016 2:52 14
EN If a problem occurs ...
Problem
Possible cause Possible solution
The stethoset receiver is not paired with the transmitter.
The distance to the transmitter is too large.
4 stethoset receivers are currently connected to the transmitter.
connect to the transmitter
Stethoset receiver does not
Pair the stethoset receiver with the transmitter.
Reduce the distance between the stethoset receiver and the transmitter.
Up to 4 stethoset receivers/ receivers can be simultaneously used with the transmitter.
Page
24
If a problem occurs that is not listed in the above table or if the problem cannot be solved with the proposed solutions, visit the RS 5000 product page at www.sennheiser.com for the most up-to-date list of frequently asked questions.
You may also contact your local Sennheiser partner for assis­tance. To find a Sennheiser partner in your country, search at www.sennheiser.com.
Leaving the range of the transmitter
The range of the transmitter largely depends on environmen­tal conditions such as wall thickness, wall composition etc. If the headphones leave the range of the transmitter, the audio first cuts out occasionally until it finally cuts out completely. You hear alerts in the stethoset receiver and the
LED flashes green (or red if the rechargeable battery is
status
Receiver
almost empty). If you re-enter the transmission range within 5 minutes, the
connection is automatically re-established. If you spend more than 5 minutes outside the transmission
range, the stethoset receiver switches off automatically.
32 | RS 5000
Page 10
RS_5000_570727_0117_PRT.book Seite 33 Montag, 10. Oktober 2016 2:52 14
Specifications
RS 5000 system
Carrier frequency 2.40 to 2.48 GHz Modulation 8-FSK Digital
analog input:
Signal-to-noise ratio
RF output power max. 10 dBm Class 1 Audio latency < 60 ms
Range
Number of stethoset receivers that can be simultaneously connected to the transmitter
Temperature range
Relative air humidity (non­condensing)
typ. 85 dBA at 1 V
rms
digital input: > 90 dBA
up to 70 m (line of sight)
4
operation: 5 °C to +40 °C storage: -25 °C to +70 °C
operation: 10 to 80 % storage: 10 to 90 %
Specifications EN
DEFRES EN
RR 5000 stethoset receiver
Type intra-aural, closed Transducer principle dynamic, neodymium magnets Ear pad material silicone Frequency response 15 Hz to 15 kHz Sound pressure level 125 dB at 1 kHz, 1 V
rms
THD < 0.5 % at 1 kHz, 100 dB SPL Operating time up to 12 hours Charging time of recharge-
able battery
up to 3 hours
built-in lithium polymer
Power supply
rechargeable battery BAP 800,
3.7 V, 350 mAh
Weight (incl. rechargeable battery)
Dimensions (switched off, W x H x D)
approx. 61 g
approx. 102 x 265 x 23 mm
RS 5000 |33
Page 11
10
5
0
-5
-10
-15
-20
0 Hz 0.25 0.5 1 21.5 3 4
Frequency/kHz
6
Gain/dB
RS_5000_570727_0117_PRT.book Seite 34 Montag, 10. Oktober 2016 2:52 14
EN Specifications
TR 5000 transmitter
digital input: optical supported data streams: PCM, 16 - 96 kHz/16 - 24 bit
Connections
analog input: 3.5 mm jack socket input voltage range:
0.15 to 4.0 V
pk
Power supply 5 V , 1 A
operation: typ. 4 W
Power consumption
standby:
0.5 W (without
charging)
Dimensions (W x H x D) approx. 50 x 42 x 270 mm
NT 5-10AW power supply unit
Nominal input voltage 100 – 240 V~ Power frequency 50 – 60 Hz Nominal output voltage 5 V Nominal output current 1 A
Temperature range
Relative air humidity (non-condensing)
operation: 0 °C to +40 °C storage: -25 °C to +70 °C
operation: 10 to 80 % storage: 10 to 90 %
Hearing profiles 1-3
Lows Midrange Highs
Profile 1 Profile 2 Profile 3
34 | RS 5000