EN Using additional functions
Using additional functions
The following additional functions are not required for normal
use of the wireless headphone system. You can, however,
extend the potential use of the system and adjust some of
the functions to your individual needs.
Pairing one or several stethoset receivers with
the transmitter
The stethoset receiver and the transmitter of your wireless
headphone system are factory pre-paired. The following
steps are only necessary if you want to pair additional stethoset receivers with the transmitter.
You can simultaneously use up to 4 paired stethoset receivers
with your transmitter. All other paired and switched-on
receivers cannot connect to the transmitter.
If more than one stethoset receiver is connected to the
transmitter, the remaining operating time/charge
status and the hearing profile cannot be displayed. In
addition, the balance and a hearing profile cannot be
adjusted. These functions can only be used/adjusted
when only one stethoset receiver is connected to the
transmitter.
왘 Place the stethoset receiver into the transmitter’s charging
compartment.
The stethoset receiver is paired with the transmitter.
+
Advanced settings
The advanced settings allow you to adjust the wireless headphone system to your individual needs and to protect the settings against accidental changes.
Activating the protected settings mode
You can change the advanced settings by activating the protected settings mode:
24 | RS 5000
Page 2
RS_5000_570727_0117_PRT.book Seite 25 Montag, 10. Oktober 2016 2:52 14
Using additional functions EN
1 Disconnect the transmitter from the power supply system
by unplugging the connector of the power supply unit.
2 Keep the Hearing profile button pressed and connect the
transmitter to the power supply system by connecting the
connector of the power supply unit to the yellow socket of
the transmitter.
21
DEFRESEN
The protected settings mode is activated. The
LED flashes blue. The LED segments indicate the cur-
status
Transmitter
rent settings as specified in the following chapters:
• battery status display (see page 25)
• charging reminder function (see page 26)
• hearing profile and balance protection (see page 27)
Deactivating the protected settings mode and saving the
advanced settings
To save the advanced settings and to deactivate the protected settings mode:
3 Disconnect the transmitter from the power supply system
by unplugging the connector of the power supply unit.
3
Battery status display
This setting is saved on the transmitter, independent of the
stethoset receiver you are using.
To change the battery status display:
왘 Make sure that the protected settings mode is activated
(see page 24).
RS 5000 |25
Page 3
RS_5000_570727_0117_PRT.book Seite 26 Montag, 10. Oktober 2016 2:52 14
EN Using additional functions
왘 Press the Balance button L.
Status and display
of the
Battery
LED
permanently
activated (operation
and charging)*
Function/behavior
The current charge status is
permanently displayed during
charging.
The remaining charge status is
st
1
segment lit
activated for 30 seconds (operation and
charging)
permanently displayed during
operation.
The current charge status is
displayed for 30 seconds during
charging. Then all LEDs go off. The
charging process is continued.
st
1
segment off
The remaining charge status is
displayed for 30 seconds during
operation. The battery status display
then goes off.
* factory setting
왘 Save the setting by disconnecting the transmitter from the
power supply system (see page 25).
Charging reminder function
This setting is saved on the transmitter, independent of the
stethoset receiver you are using.
To activate or deactivate the function:
왘 Make sure that the protected settings mode is activated
(see page 24).
왘 Press the Balance button R.
26 | RS 5000
Page 4
RS_5000_570727_0117_PRT.book Seite 27 Montag, 10. Oktober 2016 2:52 14
Using additional functions EN
Status and display of
the charging
Function/behavior
reminder function
activated*
When you switch the stethoset
receiver off and the transmitter
switches to standby mode, the
rd
3
segment lit
battery status display will flash for
10 minutes to remind you to charge
the stethoset receiver.
deactivated
After switching the wireless headphone system off, all LEDs remain
off.
rd
3
segment off
* factory setting
왘 Save the setting by disconnecting the transmitter from the
power supply system (see page 25).
Hearing profile and balance protection
This setting is saved on the stethoset receiver, independent
of the transmitter you are using.
왘 Make sure that your stethoset receiver’s balance setting
and hearing profile are correctly adjusted (see page 17 and
18).
왘 Make sure that the protected settings mode is activated
(see page 24).
왘 Place the stethoset receiver into the transmitter’s charging
compartment.
왘
Press the volume button + on the stethoset receiver to activate or deactivate the hearing profile and balance protection.
DEFRESEN
+
Status and display
of the protection
function
activated
nd
segment lit
2
+
Function/behavior
The activated hearing profile and the
balance setting cannot be changed/
deactivated.
RS 5000 |27
Page 5
RS_5000_570727_0117_PRT.book Seite 28 Montag, 10. Oktober 2016 2:52 14
EN Using additional functions
Status and display
of the protection
Function/behavior
function
deactivated*
The hearing profiles can be changed
at any time by pressing the Hearing
profile button for 2 seconds.
nd
segment off
2
왘 Save the setting by disconnecting the transmitter from the
The balance setting can be changed.
* factory setting
power supply system (see page 25).
If you are using several stethoset receivers and want to
protect the activated hearing profiles from accidental
adjustment, you can place the stethoset receivers, one
after the other, into the transmitter’s charging compartment to activate the hearing profile and balance protection.
Restoring the factory default settings
To restore the transmitter to the factory settings:
1 Disconnect the transmitter from the power supply system
by unplugging the connector of the power supply unit.
2 Simultaneously press and hold the buttons L, and R.
3 Connect the transmitter to the power supply system by
connecting the connector of the power supply unit to the
yellow socket of the transmitter while keeping the 3 buttons pressed for at least 3 seconds.
All LEDs on the transmitter flash. The transmitter settings
are restored to the factory settings.
1
2
3
28 | RS 5000
Page 6
RS_5000_570727_0117_PRT.book Seite 29 Montag, 10. Oktober 2016 2:52 14
Using additional functions EN
To restore a stethoset receiver to the factory settings:
왘 Place the stethoset receiver into the transmitter’s charging
compartment.
The
Receiver status
LED flashes green. The settings of the
stethoset receiver are also restored to the factory settings.
+
If you place the stethoset receiver into the transmitter’s
charging compartment at the very beginning of the
restore procedure, the factory settings of the transmitter and the receiver are restored simultaneously.
왘 Disconnect the transmitter from the power supply system
by unplugging the connector of the power supply unit.
The factory settings are restored and you can use the wireless headphone system again.
DEFRESEN
RS 5000 |29
Page 7
RS_5000_570727_0117_PRT.book Seite 30 Montag, 10. Oktober 2016 2:52 14
EN If a problem occurs ...
If a problem occurs ...
Problem
does not
Transmitter
Stethoset receiver
cannot be switched on
No sound
No sound when a digital
audio source is connected
Occasional sound dropouts,
it might be that you hear 5 beeps
Possible causePossible solution
No connection to the power
supply system.
switch on
The rechargeable battery is
empty.
The rechargeable battery is
deep discharged.
No connection to the power
supply system.
The audio plug is not
properly connected.
The audio source is switched
off.
The volume of the analog
audio source is either set to
the minimum or is on mute.
The audio cable is defective. Replace the audio cable.–
The stethoset receiver is not
correctly paired with the
transmitter (e.g. additional
stethoset receiver).
Cables are connected to
both the digital and the
analog audio input.
The connected digital audio
source uses an incompatible
audio format. The
Transmitter status
flashes white and blue.
The stethoset receiver is out
of the transmitter’s range.
The signal is shielded.
There are interfering devices
(e.g. WiFi router, Bluetooth
devices or microwaves) in
in the stethoset receiver
the vicinity.
LED
Check the connections of the
power supply unit.
Recharge the rechargeable
battery.
Recharge the rechargeable
battery for several hours. If the
rechargeable battery is deep
discharged, the charging
process might take a while to
start (up to 1 hour).
Check the connections of the
power supply unit.
Check the audio plug
connection.
Switch the audio source on.–
Increase the volume on the
audio source to at least a
medium level/deactivate the
muting function on the audio
source.
Pair the stethoset receiver with
the transmitter again.
Remove the cable from the
analog audio input.
Set the audio format of your
audio source to “PCM”, with a
sampling rate of max. 96 kHz
(see the instruction manual of
the audio source).
Reduce the distance between
the stethoset receiver and the
transmitter.
Remove obstacles between the
transmitter and the stethoset
receiver.
Change the position of the
transmitter or of the receiver.
Separate the transmitter from
other wireless devices by at
least 50 cm/20”.
Page
10
11
10
6
–
24
6
–
–
–
6
30 | RS 5000
Page 8
RS_5000_570727_0117_PRT.book Seite 31 Montag, 10. Oktober 2016 2:52 14
If a problem occurs ... EN
Problem
Possible causePossible solution
The volume on the stethoset
receiver is adjusted to low.
The volume of the analog
audio source is adjusted to
low.
Sound too low
The balance is misadjusted. Adjust the balance. 17
The audio cable is not
properly connected.
The TV has only a mono
on one ear
Sound only
headphone socket
The audio cable is defective. Replace the audio cable. –
The signal of the audio
source is distorted.
Sound is
The volume on the stethoset
distorted
receiver is adjusted to high.
Some TVs can adjust a delay
or latency of the audio signal for the digital audio output.
receiver with a TV
using the stethoset
You hear echoes when
The battery status display is
set so that it goes off after
30 seconds.
More than one stethoset
receiver is currently
goes off
connected to the
Battery status/
transmitter.
charge status display
Increase the volume on the
stethoset receiver.
Increase the volume of the
analog audio source to at
least a medium level (approx.
).
1 V
rms
Check the audio plug
connection.
Use an adapter (mono to
stereo jack socket) to connect
the transmitter to the TV.
Reduce the volume of the
audio source.
Reduce the volume on the
stethoset receiver.
Check the audio output settings of your TV and set the
latency to “0”.
Set the battery status display
so that it is permanently
displayed.
This function is only possible
when only one stethoset
receiver/receiver is connected
to the transmitter.
Page
16
–
6
DEFRESEN
–
–
16
–
25
–
The rechargeable battery is
almost empty.
during
2 beeps
operation
The rechargeable battery is
exhausted.
time
decreases
Operating
The hearing profile and
balance protection is
activated.
More than one stethoset
receiver is currently
connected to the
transmitter.
A hearing profile or the balance
cannot be activated or changed
Recharge the rechargeable
battery.
Have the rechargeable battery
replaced by an authorized
Sennheiser service partner.
Deactivate the hearing profile
and balance protection.
This function is only possible
when only one stethoset
receiver/receiver is connected
to the transmitter.
RS 5000 |31
11
23
27
–
Page 9
+
RS_5000_570727_0117_PRT.book Seite 32 Montag, 10. Oktober 2016 2:52 14
EN If a problem occurs ...
Problem
Possible causePossible solution
The stethoset receiver is not
paired with the transmitter.
The distance to the
transmitter is too large.
4 stethoset receivers are
currently connected to the
transmitter.
connect to the transmitter
Stethoset receiver does not
Pair the stethoset receiver with
the transmitter.
Reduce the distance between
the stethoset receiver and the
transmitter.
Up to 4 stethoset receivers/
receivers can be
simultaneously used with the
transmitter.
Page
24
–
–
If a problem occurs that is not listed in the above table or if
the problem cannot be solved with the proposed solutions,
visit the RS 5000 product page at www.sennheiser.com for
the most up-to-date list of frequently asked questions.
You may also contact your local Sennheiser partner for assistance. To find a Sennheiser partner in your country, search at
www.sennheiser.com.
Leaving the range of the transmitter
The range of the transmitter largely depends on environmental conditions such as wall thickness, wall composition etc. If
the headphones leave the range of the transmitter, the audio
first cuts out occasionally until it finally cuts out completely.
You hear alerts in the stethoset receiver and the
LED flashes green (or red if the rechargeable battery is
status
Receiver
almost empty).
If you re-enter the transmission range within 5 minutes, the
connection is automatically re-established.
If you spend more than 5 minutes outside the transmission
range, the stethoset receiver switches off automatically.
32 | RS 5000
Page 10
RS_5000_570727_0117_PRT.book Seite 33 Montag, 10. Oktober 2016 2:52 14
Specifications
RS 5000 system
Carrier frequency2.40 to 2.48 GHz
Modulation8-FSK Digital
analog input:
Signal-to-noise ratio
RF output powermax. 10 dBm Class 1
Audio latency< 60 ms
Range
Number of stethoset
receivers that can be
simultaneously connected
to the transmitter
Temperature range
Relative air humidity (noncondensing)
typ. 85 dBA at 1 V
rms
digital input: > 90 dBA
up to 70 m
(line of sight)
4
operation: 5 °C to +40 °C
storage: -25 °C to +70 °C
operation: 10 to 80 %
storage: 10 to 90 %
Specifications EN
DEFRESEN
RR 5000 stethoset receiver
Typeintra-aural, closed
Transducer principledynamic, neodymium magnets
Ear pad materialsilicone
Frequency response15 Hz to 15 kHz
Sound pressure level125 dB at 1 kHz, 1 V
rms
THD< 0.5 % at 1 kHz, 100 dB SPL
Operating timeup to 12 hours
Charging time of recharge-
able battery
up to 3 hours
built-in lithium polymer
Power supply
rechargeable battery BAP 800,
3.7 V, 350 mAh
Weight (incl. rechargeable
battery)
Dimensions (switched off,
W x H x D)
approx. 61 g
approx. 102 x 265 x 23 mm
RS 5000 |33
Page 11
10
5
0
-5
-10
-15
-20
0 Hz0.250.5121.534
Frequency/kHz
6
Gain/dB
RS_5000_570727_0117_PRT.book Seite 34 Montag, 10. Oktober 2016 2:52 14
EN Specifications
TR 5000 transmitter
digital input: optical
supported data streams:
PCM, 16 - 96 kHz/16 - 24 bit
Connections
analog input: 3.5 mm jack socket
input voltage range:
0.15 to 4.0 V
pk
Power supply5 V , 1 A
operation: typ. 4 W
Power consumption
standby:
≤ 0.5 W (without
charging)
Dimensions (W x H x D)approx. 50 x 42 x 270 mm
NT 5-10AW power supply unit
Nominal input voltage100 – 240 V~
Power frequency50 – 60 Hz
Nominal output voltage5 V
Nominal output current1 A
Temperature range
Relative air humidity
(non-condensing)
operation: 0 °C to +40 °C
storage: -25 °C to +70 °C
operation: 10 to 80 %
storage: 10 to 90 %
Hearing profiles 1-3
LowsMidrangeHighs
Profile 1
Profile 2
Profile 3
34 | RS 5000