Seneca MSC User Manual

Page 1
USER MANUAL
1MI004962-EMSC
Page 2
Page 3
Page 4
Page 5
TABLE OF CONTENTS
PRODUCT DESCRIPTION P.06
SAFETY INFORMATION P.10
USER GUIDE P.12
CALIBRATOR FUNCTIONS P.15
ERRORS TABLE APP P.35
CONNECTION DIAGRAMS P.36
USING THE APP P.40
SUMMARY OF FEATURES P.50
TECHNICAL SPECIFICATIONS P.52
USER INTERFACE P.54
TROUBLESHOOTING P.56
FIRMWARE UPDATE P.59
Page 6
PRODUCT DESCRIPTION
The MSC calibrator is a portable multi-function instrument for the generation and measurement of signals commonly found in auto­mation environments. It was designed to help professionals who have to verify and de­velop signal processing equipment connected to automatic control devices. Unlike what the market offers, the MSC calibrator combines simulations and measurements in one instrument using the user’s smartphone, tablet or Windows PC (via the USB cable) for an intuitive and modern interface.
The MSC multifunction process calibrator is a hand-held instru­ment, powered by rechargeable batteries, used to measure and generate electrical quantities. In addition to the functions described in the "FUNCTIONS" para­graph (page 15), the calibrator has the following characteristics:
Input/output terminal for thermocouples (TC) and isothermal internal block with automatic temperature compensation of the reference junction.
4 standard 4mm diameter bushings for 2-3-4 wire sizes.
Protection against overvoltages up to 250 Vac.
Lithium battery power supply, with up to 14h autonomy.
Bluetooth connection with tablet and smartphone.
Management of universal signals (analogue, digital, pulse, electrical, weight, temperature)
High accuracy class: 0.05% basic class for each type of input/ output.
Data always available thanks to the dedicated app.
6 MI004962-E MSC
Page 7
Micro USB connector for power supply and PC connection.
The analogue signals managed both in measurement and genera­tion are:
• Milliamperes (dc)
• Volts, millivolts (dc)
• Thermocouples, two/three or four-wire heating elements
• 350 Ω bridge extensometer (LOAD CELL)
The following are also managed:
• Pulse generation with duration, voltage amplitude and variable duty cycle
• Pulse measurement with adjustable threshold
• Frequency measurement and generation
The device is protected against damage due to incorrect connec­tions such as short circuits and overvoltages applied to the inputs (e.g. 230 Vac). The simplicity of use of the connections and the protections against
overload make the device suitable even for the less experienced
and in a hobby and educational environment. It is possible to connect multiple calibrators to a smartphone and control them individually.
Electrical and electronic waste disposal (applicable in the European Union and other countries with
recycling). The symbol on the product or its packaging shows the product must be surrendered to a collection centre authorized to recycle electrical and electronic waste.
Damaged parts must be replaced by the manufacturer, who is also responsible for repairing the device. The product is sensitive to electrostatic discharges. Take appropriate measures during any operation.
7MI004962-EMSC
Page 8
PRODUCT DESCRIPTION
Socket for thermocouple measurement/generation
1
Measurement/generation bushing -EX
2
Measurement/generation bushing -SN
3
4
Measurement/generation bushing +SN
Measurement/generation bushing +EX
5
ON/OFF button
6
Bluetooth RESET button
7
PWR ON Led
8
Established connection Bind Led
9
Battery status indicator Led
10
Bluetooth/USB communication Led
11
Data recording Led
12
Error message Led
13
Power/communication micro USB connector
14
RESET button
15
Battery charge warning Led
16
8 MI004962-E MSC
Page 9
1
-EX -SN +SN +EX
10
12
2
3
6
8
4
5
7
9
11
13
14
16
15
9MSC
Page 10
SAFETY INFORMATION
This manual contains safety rules that must be complied with to safeguard personal safety and to prevent damage to property. The indications to be followed to guarantee personal safety are high­lighted by a triangle with the following meanings:
WARNING
the symbol associated with the word ‘warning’ indicates conditions or actions that put the user's safety at risk.
CAUTION
The symbol associated with the word ‘caution’ indicates conditions or actions that might damage the calibrator or equipment tested.
Qualied personnel
The product covered by this documentation may only be used by
qualied personnel for the respective assigned task, in compliance
with the documentation relating to the task, and especially with the safety warnings and the precautions contained therein. By virtue of
their training and experience, qualied personnel can recognize the
risks related to the use of this product and avoid possible dangers.
Disclaimer
We have checked that the contents of this documentation corre­spond to the hardware and software described. However, since we cannot rule out any differences, we cannot guarantee a perfect match. However, the content of this documentation is periodically checked and any corrections or changes are inserted in subse­quent editions.
10 MI004962-E MSC
Page 11
WARNING
• Do not apply different voltages or higher voltages than indicated
between terminals, or between any terminal and earth (voltages greater than 50Vdc)
• Do not use the calibrator if damaged, or if signs of possible dam-
age are visible on it.
• Do not remove the internal battery.
• Do not use the calibrator if it runs abnormally.
• Use the reset button if there is an anomaly in the operation.
• Do not touch the calibrator terminals during use.
• To carry out the measurements, use the cables supplied or any-
way adequate for the measurements to be taken.
Select the appropriate range for the desired measurement or generation.
CAUTION
• Make sure the internal battery is charged if you intend to use the
calibrator without power from the USB port.
• Use the calibrator as described in this manual.
• Before each measurement or generation, refer to the electrical
connections shown in the manual.
• Do not use the calibrator in environments containing combustible/
explosive dust, gas or vapour.
The MSC calibrator has been designed according to EN 6101-1 standards
11MI004962-EMSC
Page 12
USER GUIDE
The calibrator is a device that can measure or generate electrical quantities to test devices. To proceed with the correct use of the instrument, make sure to follow the following steps:
• Turn on the device and connect with the APP via Bluetooth.
• Wait 5 minutes for the appliance temperature to stabilize.
• Connect the cables correctly, according to the type of meas­urement/generation to be carried out: refer to the connection diagrams on p. 36.
It is however necessary, before use, to make sure that the battery is fully charged so as to guarantee its life for the generations or measurements that will be carried out. Before each use it is advis­able to recharge the instrument using the USB cable supplied, until
the LED battery charge indicator on the bottom of the calibrator
goes off, as shown in the diagram on page 9. The calibrator can maintain the signals to be generated (set by the user) within a determined error range. During generation, the calibrator will constantly check the generat­ed signal, warning about any generation error (where provided for)
by notifying it on the APP.
CAUTION
The calibrator is protected against the risk of applying temporary overvoltages. However, it is good to remember that in this case the device needs 15-20 minutes to restore the normal conditions. If this period of time is not observed, the accuracy of the measure­ments and/or generations cannot be guaranteed.
12 MI004962-E MSC
Page 13
Be careful when using the calibrator with devices that have earth
connections. Connections 1 and 2, shown in FIG. 1, can exist, but
not simultaneously. This working condition can interfere with the correct operation of the calibrator and the Bluetooth connection. If this connection is made during the generation of an electrical quantity, internal protections may intervene, in which case, once the correct electrical connection between the devices has been restored, it is necessary to leave the calibrator inactive for about 15-20 minutes to go back to normal conditions.
In order to achieve proper PC//measurement systems (EUT)
isolation and to protect the USB port of your PC, you must use the USB-ISO accessory as shown in Fig. 2.
1 2
USB-ISO
FIG. 1
FIG. 2
13MI004962-EMSC
Page 14
USING THE CALIBRATOR
-EX -SN +SN +EX
CABLES SUPPLIED
BLACK CABLE WITH 4MM BLACK PLUG
14 MI004962-E MSC
WHITE CABLE WITH 4MM BLACK PLUG
WHITE CABLE WITH 4MM RED PLUG
RED CABLE WITH 4MM RED PLUG
Page 15
CALIBRATOR FUNCTIONS
MEASUREMENT FUNCTIONS:
The MSC calibrator allows measuring the following signals:
passive current
active current
HiRange voltage
LowRange voltage
thermocouple
resistance thermometer
load cell
frequency/pulses
For correct measurement, connect the cables as shown in the diagrams on page 36 and page 37.
GENERATION FUNCTIONS:
The MSC calibrator allows measuring the following signals:
passive current
active current
HiRange voltage
LowRange voltage
thermocouple
resistance thermometer
load cell
frequency/pulses
For correct measurement, connect the cables as shown in the diagrams on page 38 and page 39.
15MI004962-EMSC
Page 16
MEASUREMENTS
Checks common to all measurements
The following are the checks that are common to all measure­ments:
Sample screen from the APP:
The images and diagrams shown must be considered as indicative and not binding
since they are subject to possible updates.
16 MI004962-E MSC
Page 17
In the sample screen it is possible to distinguish the checks that are common to all the measurements (red) and those common to analogue measurements (orange) only. In red you can identify the checks that have the following mean­ings:
• Zeroing of MIN / AVG / MAX measurements, for analogue
measurements
• Count zeroing for pulse measurement
• Pausing of analogue or digital measurements
Sharing the measurements taken. This button allows for exam­ple sharing a memo of the measurements just made.
• Datalogger: the activation of this function allows starting to record the measurements that are being taken.
The MIN / AVG / MAX measurements are identied in orange and
the check allows changing the display from primary to second­ary measurement and vice versa.
17MI004962-EMSC
Page 18
CURRENT / VOLTAGE MEASUREMENTS
The calibrator can measure voltages and currents that are applied to the two innermost bushings (SN+ SN-) according to the following ranges.
For the voltages, two measurement scales are available:
• Voltage -10÷90 mV ( LowRange)
• Voltage 0÷27 V ( HiRange)
To carry out the voltage measurement, connect the calibrator according to gures M5 or M7 while, to make a current measure­ment, connect the calibrator according to FIG. M1 or M3 (page 36).
Once the calibrator is connected and the Bluetooth connection is
working, press M (Measurement) on the Home screen of the APP and select "CURRENT / VOLTAGE".
Then select the desired size. Since the measurements are homogeneous in the way they are
presented by the APP, a measurement is then taken into considera­tion; then choose "VOLTAGE 0.27 V".
The calibrator starts displaying the values related to the voltage applied to the input bushings in Volts. The default value displayed is in Volts. The measurement can be displayed as a percentage of the full scale by pressing the P9 button (double arrow symbol)
If the measurement is unstable, ltering can be applied. To activate
it, follow the following steps by pressing:
1. Settings (refer to image on page 47)
2. AVG lter settings
Then select the desired ltering time between: 0.25, 0.5, 1, 2, 4, 8
seconds.
18 MI004962-E MSC
Page 19
FREQUENCY MEASUREMENTS
The calibrator allows the measurement of the voltage pulses applied in the two innermost bushings (SN+ SN-). The electrical connections are shown in FIG. M8
Once the calibrator is connected and the Bluetooth connection
works, press M (Measurement) on the APP Home screen and then choose "FREQUENCY"
To correctly detect a frequency of a voltage signal applied to the bushings, it is important to choose an appropriate trigger threshold by going to:
1 Settings 2 Set trigger level
The calibrator shows the frequency in Hz.
19MI004962-EMSC
Page 20
THERMOCOUPLE MEASUREMENTS
The calibrator allows the temperature measurement with thermo­couples which are connected via connector 1 (diagram on page 9). The electrical connections are shown in FIG. M2 (page 36).
For the types of supported thermocouples, refer to TAB 1 (page
50).
Once the calibrator is connected and the Bluetooth connection
works, press M (Measurement) on the APP Home screen and then choose "THERMOCOUPLE". Then select the type of thermocouple
you wish to take the measurement with.
In measurement mode the temperature value in °C (and mV) is shown. The displayed temperature value is the real temperature measured by the thermocouple, including the cold junction com-
pensation (depending on the set conguration). This last value is
shown on the display with the wording “C.J. Read”. On the other hand, the value in mV is the real voltage value read by the terminals without any compensation.
If the measurement is unstable, ltering can be applied.
To activate it, follow the following steps by pressing:
1. Settings
2. AVG lter settings
Then select the desired ltering time between: 0.25, 0.5, 1, 2, 4, 8
seconds.
20 MI004962-E MSC
Page 21
CAUTION
Wait until the device reaches a stable thermal condition before taking measurements with the thermocouples (typically at least 30 minutes).
CAUTION
If the thermocouple does not have its own connector, use the inner­most bushings SN+ SN- as an alternative, bearing in mind that the compensation of the cold junction in this case may be less precise.
21MI004962-EMSC
Page 22
RESISTANCE THERMOMETER MEASUREMENTS (RTD)
The calibrator allows the measurement of the temperature with resistance thermometers connected with 2, 3 or 4 wires that are connected to the bushings, according to the types indicated in
TAB 1 (page 50). The electrical connections are shown in FIG. M4
(page 36).
Once the calibrator is connected and the Bluetooth connection
works, press M (Measurement) on the APP Home screen and then
choose "RTD". Then select the type of resistance thermometer you want to measure.
In RTD measurement mode, the temperature value in °C is shown according to the type of resistance thermometer selected, along
with the value measured in Ω.
If the measurement is unstable, ltering can be applied. To activate
it, follow the following steps by pressing:
1. Settings
2. AVG lter settings
Then select the desired ltering time between: 0.25, 0.5, 1, 2, 4, 8
seconds.
22 MI004962-E MSC
Page 23
PULSE MEASUREMENTS
The calibrator allows the measurement of the voltage pulses applied in the two innermost bushings (SN+ SN-). The electrical connections are shown in FIG. M8 (page 37). Once the calibrator is connected and the Bluetooth connection
works, press M (Measurement) on the APP Home screen and then choose "PULSE" followed by the mode in which the pulse is to be detected: "POSITIVE FRONT SIGNAL WITH PULSES" or "NEG­ATIVE FRONT SIGNAL WITH PULSES" (the pulse is detected on
the rising or falling edge when crossing the set trigger threshold)
To correctly detect the pulses it is important to choose an appropri­ate trigger threshold by going to:
1. Settings
2. Set trigger level
It is possible to lter the pulses applied to the input bushings by ap­plying the pulse minimum duration lter. The pulse is then detected
only if its duration is greater than or equal to the set value.
To change this parameter follow the following steps:
1. Settings
2. Set minimum pulse duration
then set the value expressing it in micro seconds.
23MI004962-EMSC
Page 24
MEASUREMENTS WITH LOAD CELLS
The calibrator is compatible with load cells having a rated resist-
ance of 350 Ω.
Be sure to follow the guidelines for use to take measurements correctly.
By activating this measurement function, the device supplies the load cell through the outermost terminals (+EX -EX) while the gen­erated signal is read via the innermost terminals (+SN -SN)
To take the measurement, connect the calibrator according to g.
M6 (page 37); on the Home screen, press M (Measurement) and then "Load Cell". The measurement of the load cell is displayed in two ways:
• mV/V ratio
• weight in Kg
To set the sensitivity of the load cell sensor or set the tare and full
scale in Kg press:
1. Settings
2. Options
In the "load cell settings" screen you can:
• Set the unit of measurement
• Set cell sensitivity [mV/V]
• Set the load cell capacity [Kg]
• Zero the tare: before proceeding to zero the tare, make sure
24 MI004962-E MSC
Page 25
that the measurement on the display is stable then proceed with resetting using the reset button:
TARE
If the measurement is unstable, ltering can be applied. To activate
it, follow the following steps by pressing:
3. Settings
4. AVG lter settings
Then select the desired ltering time between: 0.25, 0.5, 1, 2, 4, 8
seconds.
CAUTION
The 4-wire measurement of the load cell is inuenced by the resist­ance of the conductors supplying it. If it is not possible to shorten the connections between the load cell and the MSC calibrator, by measuring the resistance only of the conductors that supply the
350+R
K=
+EX +R-EX
350
load cell (conductors +EX and -EX), the correction coefcient can
be calculated: The obtained coefcient must be multiplied with the value meas­ured by the calibrator to obtain the real value.
25MI004962-EMSC
Page 26
GENERATION / SIMULATION
The following are the checks that are common to all measure­ments:
APP sample screen:
The images and diagrams shown must be considered as indicative and not binding
since they are subject to possible updates.
26 MI004962-E MSC
Page 27
In the sample screen, a distinction is made between the checks common to all analogue generations (in orange) and those of the pulse generation only (in red).
In Orange it is possible to identify the area where the generation value can be set by using the up/down swipe and the keyboard to enter the desired value. The other checks are listed below:
• Button indicating the status of the generation (active or not). Pressing it, the output is deactivated and the button changes status.
• Command to reverse the primary and secondary measure­ments and vice versa
• Ramp function start button.
In red we can identify the elements that concern the generation of
pulses; in the example we can identify the 8 remaining pulses to be
generated and the three buttons:
• START: starts the generation of voltage pulses
• PAUSE: temporarily interrupts the generation of pulses
• STOP: cancels the current pulse generation session
27MI004962-EMSC
Page 28
CURRENT / VOLTAGE GENERATION
The calibrator can operate as a voltage or current generator. These quantities are supplied by the innermost bushings (SN+ SN-) according to the following ranges. For the voltages two generation scales are available:
• Voltage -10÷90 mV
• Voltage 0÷27 V
To carry out voltage generation, connect the calibrator according
to gure G7 (page 39), while to generate a current, connect the calibrator according to gures G1 or G3 (page 38) depending on
whether you want to generate a passive or active current.
Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and choose "CURRENT / VOLTAGE".
Then select the desired voltage or current generation.
Since the generations are homogeneous in the way they are
presented by the APP, the following shows a sample generation. Select "VOLTAGE 0.27 V".
The generation screen is displayed and the calibrator starts supply­ing the output bushings with the preset voltage of 0 V. To change this value, simply enter the desired value using the numeric keypad
and conrm with OK.
28 MI004962-E MSC
Page 29
THERMOCOUPLE SIMULATION
The calibrator can simulate a thermocouple. To see which types of thermocouples are supported, refer to simulation of thermocouples is made possible by the CONN. TC, of the diagram on page 9 or through the SN+ and SN- terminals. The electrical connections are shown in FIG G2 (page 38). Once the calibrator is connected and the Bluetooth connection works,
press G (signal Generator) on the APP Home screen and choose "THERMOCOUPLE".
Then select the type of thermocouple you want to simulate; the
following is an example of the simulation of a type K thermocouple. A screen relative to the simulation of type K thermocouples is dis­played on the APP. It is possible to enable or disable the compen-
sation in the cold junction, by default automatically set on internal temperature. It is possible to change this setting by going to:
1 Settings 2 Cold junction setting
3 Enable the "Manual setting” control
4 Set the temperature value for the cold junction and then
conrm
CAUTION
Wait until the device reaches a stable thermal condition before simulating a thermocouple.
If the thermocouple does not have its own connector, use the inner­most bushings SN+ SN- as an alternative, bearing in mind that the compensation of the cold junction in this case may be less precise.
TAB 1 (page 50).
The
29MI004962-EMSC
Page 30
RESISTANCE THERMOMETER SIMULATION (RTD)
The calibrator is able to simulate a resistance thermometer (RTD) connected to the 2, 3 or 4 wire bushings. The types of supported
RTDs are listed in TAB 1 (page 50).
To carry out RTD simulation, connect the calibrator according to FIG.G4 (page 38). Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and
choose "RTD". Then select the type of RTD that you want to simulate; the sim­ulation of a PT100-type resistance thermometer is shown as an
example.
Then enter the desired value in °C and conrm; it is possible to switch the display of the unit of measurement from °C to Ω.
Thermal resistance generation can be divided into two scales; the PT100 scale and the PT1000 scale (which includes the PT500). For these two scales the excitation current from the external instru­ment must remain within a certain range (see table on page 50) . If this condition is not met, an Generation error is signalled, indicating the possibility that the simulated resistive value is not correct.
CAUTION
30 MI004962-E MSC
Page 31
LOAD CELL SIMULATION
The calibrator can simulate a load cell (LOAD CELL) with a nomi­nal resistance of 350 Ω. The connection is made by 4 wires through the bushings EX+ EX-, SN+ SN-. The supply voltage applicable to the external bushings has a range from 5V to 10V. To simulate
a load cell, connect the calibrator according to FIG. G6 (page 39). Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and then select "LOAD CELL".
From the load cell screen, proceed to set the parameters by follow­ing the steps below:
1 Settings 2 Options 3 Set the parameters "Unit of measurement", "Load cell gain" in mV/V, and then the full scale of the load cell.
For instance, once the sensitivity is set to 2 mV/V, the scale to
Kg and the full scale to 1000 Kg, simply conrm the parameters,
then return to the main page of the load cell simulation and enter
the value in Kg that you wish to simulate. The value can also be
entered as mV/V
CAUTION
Wait until the device reaches a stable thermal condition before simulating a load cell.
31MI004962-EMSC
Page 32
FREQUENCY GENERATION
The calibrator can generate a square wave voltage signal with 50% duty cycle and settable frequency and peak voltage values. The generated signal is taken out to the bushings (SN+ SN-).
The denable properties of the pulses are:
Value ON: value in Volts assigned to the square wave when the signal is at high logic level
Value OFF: value in Volts assigned to the square wave when the signal is at low logic level.
To generate a voltage square wave signal, connect the calibrator according to FIG. G8 (page 39). Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and then select "FREQUENCY".
To proceed with the generation, set the band parameters of the square wave by following the following steps:
1 Settings 2 Set Pulse Parameter
3 Enter the desired Value ON / Value OFF band values and conrm with OK
Then set the desired frequency using the numeric keypad and make sure that the output enable button is turned to ON.
The primary measurement unit of the signal is in Hz, switchable to seconds to show the length of the period.
32 MI004962-E MSC
Page 33
CAUTION
It is important to note that the time resolution according to which
the square wave signal is generated is 50 μs. This implies an error
that is dependent on the numeric value set. For instance, you want to generate a signal with a frequency of 150 Hz whose corresponding period is 6.666 ms; the number of counts necessary to obtain this frequency is obtained through:
6.666 mHz /50 µs=133.33. The system then rounds off to the nearest whole number, 133, which it will use to generate the output signal. The output frequency will have the value:
FOUT= 1 = 150.37 Hz
133X50 µs
The APP will automatically show the value actually generated.
33MI004962-EMSC
Page 34
PULSE GENERATION
The calibrator can generate a train of voltage pulses available at the two innermost bushings (SN+ SN-).
The denable properties of the pulses are:
• Value ON / Value OFF: values in V relative to the pulse value and to the value of the steady state of OFF respectively.
• Time ON / Time OFF: values in ms that dene the period T of
the pulse. ( T=TimeON + TimeOFF). Time ON is the time in which the pulse remains at the Value ON value in ms.
To generate voltage pulses, connect the calibrator according to FIG. G8 (page 39). Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and then select "PULSE".
To proceed with the generation, set the time and band parameters of the pulse by following the following steps:
1. Settings
2. Set Pulse Parameter
3. Enter the desired band Value ON / Value OFF values and the
times corresponding to the Time ON / Time OFF pulse period
and conrm with OK
Then set the number of pulses and press start to start the genera­tion; the remaining pulses are displayed at the same time. During this phase it is possible to pause and then resume the generation
by pressing PAUSE or cancel it by pressing STOP.
34 MI004962-E MSC
Page 35
ERRORS APP
ERROR SIGNALS ON APP / LEDSTATUS
ERROR MESSAGE MEANING LED STATUS
Generation Error
Measure Error
Loop Error
The set generation
is not guaranteed
Signal out of scale
Only for current
supplied by an ex-
ternal generator.
(if V
> 29 V)
Loop
Low Battery
Error Led ON
(Red)
Error Led ON
(Red)
Error Led ON
(Red)
Low Battery Led ON
blinking (Amber)
35MI004962-EMSC
Page 36
-EX -SN +SN +EX
FIGURA M5: MISURATORE DI TENSIONE HiRange
V
+
-EX -SN +SN +EX
Kg
FIGURA M6: MISURATORE CELLA DI CARICO
-EX -SN +SN +EX
FIGURA M7: MISURATORE DI TENSIONE LowRange
mV
+
-EX
-SN +SN
+EX
FIGURA M8: MISURATORE DI FREQUENZA/IMPULSI
Hz
+
EXCITATION -
SENSE -
SIGNAL -
SIGNAL +
SENSE +
EXCITATION +
FIGURA M1: MISURATORE DI CORRENTE PASSIVO
(trasmettitore a 4 fili)
-EX -SN +SN +EX
mA
FIGURA M3: MISURATORE DI CORRENTE ATTIVO
(trasmettitore a 2 fili)
FIGURA M2: MISURATORE TEMP. CON TERMOCOPPIA
-EX -SN +SN +EX
°C
FIGURA M4: MISURATORE DI TERMORESISTENZA
3 W
2 W
2 W
4 W
-EX -SN +SN +EX
mA
36
-SN +SN
-EX
°C
+EX
Page 37
FIGURA M5: MISURATORE DI TENSIONE HiRange
+
FIGURA M6: MISURATORE CELLA DI CARICO
-EX -SN +SN +EX
V
FIGURA M7: MISURATORE DI TENSIONE LowRange
+
-EX -SN +SN +EX
EXCITATION -
SENSE -
SIGNAL -
-EX -SN +SN +EX
SIGNAL +
SENSE +
EXCITATION +
Kg
FIGURA M8: MISURATORE DI FREQUENZA/IMPULSI
+
-SN +SN
-EX
+EX
mV
Hz
37
Page 38
-EX -SN +SN +EX
FIGURA G5: GENERATORE DI TENSIONE HiRange
-EX -SN +SN +EX
FIGURA G6: SIMULATORE CELLA DI CARICO
-EX -SN +SN +EX
FIGURA G7: GENERATORE DI TENSIONE LowRange
-EX
-SN +SN
+EX
FIGURA G8: GENERATORE FREQUENZA/IMPULSI
V
+
+
mV
Hz
Kg
FIGURA G1: GENERATORE DI CORRENTE PASSIVO
FIGURA G2: SIMULATORE TERMOCOPPIA
mA
-EX -SN +SN +EX
FIGURA G3: GENERATORE DI CORRENTE ATTIVO
mA
-EX -SN +SN +EX
°C
-EX -SN +SN +EX
FIGURA G4: SIMULATORE DI TERMORESISTENZA
°F
°C
3 W 4 W
2 W 2 W
-SN +SN
-EX
+EX
38
Page 39
FIGURA G5: GENERATORE DI TENSIONE HiRange
FIGURA G6: SIMULATORE CELLA DI CARICO
V
-EX -SN +SN +EX
+
FIGURA G7: GENERATORE DI TENSIONE LowRange
mV
-EX -SN +SN +EX
Kg
-EX -SN +SN +EX
FIGURA G8: GENERATORE FREQUENZA/IMPULSI
Hz
-SN +SN
-EX
+EX
+
39
Page 40
USING THE APP FROM A MOBILE
To use the MSC calibrator, you need to install the free "MSC by Seneca" APP available for iOs and Android devices on your smart­phone or tablet and register your details. Ensure that your tablet and / or smartphone is equipped with Blue­tooth 4.1 or higher.
To install the App on a smartphone or tablet, scan the QR code
shown and enter the required details in the registration page. When the application starts, select "Search" and, in the list of avail­able devices, select the desired device to start the synchronization with your smartphone. To ensure the calibrator is working properly, make sure you have selected the correct device.
40 MI004962-E MSC
Page 41
MAIN SCREEN:
Navigation menu
1
“Search” command
2
When the application starts, select the “Search” command to view the devices to be associated.
1
2
41MI004962-EMSC
Page 42
DEVICE SEARCH
Devices detected
1
When the app is opened, the device will automatically recognize the terminal to join. To complete the operation simply select the device you want to interface with.
Stops the search
2
1
2
42 MI004962-E MSC
Page 43
Activates the sound signal in
1
the event of an alarm
Activates the sound signal
2
on overow in generation
Activates the sound signal
3
on turning on
Activates the sound signal
4
during data recording
If selected, once the
5
available memory is full, the instrument will write the data on the recordings already made
Auto power OFF function:
6
this function allows setting a measurement time or signal generation, at the end of the operation the instrument will automatical-
ly turn o󰀨.
Allows setting the desired
7
sampling time
OPTIONS
1 2 3 4 5
6
7
8
9
11 10
Space dedicated to setting
8
the desired sampling time
Unit of measurement setting
9
Setting conrmation
10
Cancels the settings
11
43MI004962-EMSC
Page 44
Navigation menu
1
Meter mode
2
HOME SCREEN
1
Generator/simulator mode
3
Instrument name
4
Calibrator characteristics
5
Device disconnect
6
button
2 3
4
5
6
44 MI004962-E MSC
Page 45
MEASUREMENT CONTROLS
Calibrator battery level
1
Access to the last measure-
2
ment taken
Current / voltage meter
3
Thermocouple meter
4
Resistance temperature meter
5
Load cell meter
6
Pulse meter
7
Frequency meter
8
1
2
3
4
5
6
7
8
45MI004962-EMSC
Page 46
GENERATION/SIMULATION CONTROLS
Access to the last generation
1
carried out
Current / voltage generator
2
Thermocouple simulator
3
1
Resistance temperature
4
simulator
Load cell simulator
5
Pulse generator
6
2
3
4
5
7
Frequency generator
6
7
46 MI004962-E MSC
Page 47
METER READING SCREEN
Meter setting
1
Calibrator battery level
2
Reading screen: primary
3
(a)
measurement
Reading screen: second-
3
(b)
ary measurement
Command to reverse the pri-
4
mary and secondary measure­ments and vice versa
TARE command
5
Minimum measurement
6
Medium measurement
7
Maximum measurement
8
Datalogger function:
9
by selecting this func­tion, measurement recording starts.
3
(a)
5
TARE
3
(b)
6 7 8
1
2
4
Share minimum, medium and
10
maximum value measure­ments
Suspend minimum, medium
11
and maximum value measure­ments Update minimum, medium
12
and maximum value meas­urements
12
11
9
10
47MI004962-EMSC
Page 48
GENERATOR/SIMULATOR READING SCREEN
Generator settings
1
Set value
2
Ramp start
3
Unit of measurement
4
management
Status indicator
5
Quick output levels
6
2
3
1
4
Setting keypad
7
output
5
6
7
48 MI004962-E MSC
Page 49
RAMP READING SCREEN
Generated signal
1
Ramp generation start/
2
pause
Ramp management
3
Table to dene the points
4
for ramp creation
1
3
2
4
49MI004962-EMSC
Page 50
FEATURE SUMMARY
PULSES / FREQUECY
CMRR-NMRR
Note
RANGE PRECISION * RESOLUTION RULE
Note
GENERATION MEASURE
0÷24 0,04 + 0,01 1 uA 18 >60 dB
4,21,22
*: % of the measure + % of the scale
MI004961-I MSC
FUNCTION U.M RANGE PRECISION * RESOLUTION
ACTIVE CURRENT [dc mA] 0,1÷24 0,04 + 0,01 1 uA 3 0÷24 0,04 + 0,01 1 uA 17 >60 dB
VOLTAGE [hi range] [dc V] 0÷26 min 0,02 + 0,01 1 mV 1 0÷26 min 0,02 + 0,01 1 mV 9 >100 dB
VOLTAGE [lo range] [dc mV] -10÷90 0,03 + 0,01 5 uV 2 -10÷90 0,03 + 0,01 5 uV 9 >100 dB
PASSIVE CURRENT [dc mA] 0,1÷24 0,04 + 0,01 1 uA
Pt100 [°C] -200÷859 0,02 + 0,01 0,1 5,19 -200÷850 0,03 + 0,02 0,1 IEC 60751 6 >140 dB
Pt500 [°C] -200÷859 0,03 + 0,01 0,1 5,20 -200÷850 0,03 + 0,02 0,1 IEC 60751 6 >140 dB
Pt1000 [°C] -200÷859 0,03 + 0,01 0,1 5,20 -200÷850 0,03 + 0,02 0,1 IEC 60751 7 >140 dB
CU50, CU100 [°C] -180÷200 0,02 + 0,01 0,1 15,19 -180÷200 0,03 + 0,02 0,1 GOST 6651-2009 6,15 >140 dB
NI100, NI120 [°C] -80÷260 0,02 + 0,01 0,1 16,19 -60÷250 0,03 + 0,02 0,1 DIN 43760 6,16 >140 dB
THERMOCOUPLE J [°C] -210÷1200 0,03 + 0,01 0,1 12,2 -210÷1200 0,03 + 0,01 0,1 EN 60584-1:1997 9,12 >100 dB
THERMOCOUPLE T [°C] -270÷400 0,03 + 0,01 0,1 12,2 -200÷400 0,03 + 0,01 0,1 EN 60584-1:1997 9,12 >100 dB
THERMOCOUPLE K [°C] -270÷1372 0,03 + 0,01 0,1 12,2 -200÷1372 0,03 + 0,01 0,1 EN 60584-1:1997 9,12 >100 dB
THERMOCOUPLE E [°C] -270÷1000 0,03 + 0,01 0,1 12,2 -200÷1000 0,03 + 0,01 0,1 EN 60584-1:1997 9,12 >100 dB
THERMOCOUPLE S [°C] -50÷1768 0,03 + 0,01 0,5 12,2 -50÷1768 0,03 + 0,01 0,5 EN 60584-1:1997 9,12 >100 dB
THERMOCOUPLE N [°C] -270÷1300 0,03 + 0,01 0,1 12,2 -200÷1300 0,03 + 0,01 0,1 EN 60584-1:1997 9,12 >100 dB
THERMOCOUPLE R [°C] -50÷1768 0,03 + 0,01 0,5 12,2 -50÷1768 0,03 + 0,01 0,5 EN 60584-1:1997 9,12 >100 dB
THERMOCOUPLE L [°C] -200÷800 0,03 + 0,01 0,1 12,2 -200÷800 0,03 + 0,01 0,1 Gost 8.585-2001 9,12 >140 dB
THERMOCOUPLE B [°C] 0÷1820 0,03 + 0,01 0,5 12,2 250÷1820 0,03 + 0,01 0,5 EN 60584-1:1997 9,12 >100 dB
[Hz] 0,1÷1000 0,02 see manual 1 0,1÷10000 0,02 see manual 10,11
LOAD CELL 350 Ohm [mV/V] -0,2÷2,4 0,05 0,001 13,8 -0,2÷2,4 0,05 0,001 9,14 >100 dB
TAB. 1
OTHER DATA
Typical temperature drift: 50 ppm/fs
Output response times: < 100 ms between displayed command and 90% output
Operating temperature: -20..50°C (not under charge), 0-45°C during charging
Storage temperature: 0..35°C
Weight: 330 g ca.
50
Page 51
NOTES ON TABLE 1
1 - IoutMAX = 20 mA per V < 20; IoutMAX =10 mA per 20 < V <25; IoutMAX = 5mA per V > 25; output impedance < 50 mΩ
2 - IoutMAX = 10 mA; output impedance < 100 mΩ
3 - RLoad < 1 KΩ
4 - 3V minimum loop voltage, 28 V maximum
5 - α=0.00385
6 - Measurement current: 1 mA
7 - Measurement current: 250 uA
8 - Maximum excitation voltage: from 5 to 10 V
9 - Input impedance: ~ 10 MΩ
10 - Input impedance: 1 MΩ; Voltage from 1 to 24 V
11 - Max frequency measurement 1 kHz;
12 - Cold junction error: 1°C between 10 and 35°C ambient, 2°C between -20 and 10°C and between 35 and 50°C ambient
13 - Output impedance: 300 Ohm
14 - Excitation voltage < 8V
15 - α=0.00428
16 - α=0.006178
17 - Voltage drop < 3.5 V
18 - Minimum voltage supplied: 24 V
19 - Accuracy dened with 1 mA measurement current and polarity respected
20 - Accuracy dened with 250 uA measurement current
21 - Output impedance: 3.75 MΩ
22 - Precision dened with 26 V loop power supply voltage
MI004961-I MSC
51
Page 52
TECHNICAL SPECIFICATIONS
GENERAL DATA
Power supply from the network
Battery power supply
Protection rating IP20
Operating temperature -20..50°C (not under charge), 0-40°C during charging
Storage temperature 0..35°C
Humidity 30..90 % non condensing
Insulation
Overvoltage protection 240 Vac max without permanent damage
Rejection 50/60 Hz
Freq. Sampling 20 Hz
Dimensions 147 x 88 x 25 mm
Weight 330 g
Supplied Connecting cables (4), battery charger, USB cable
Calibration report On request
From 100-240 Vac network via standard USB battery
3400 mAh LiPo battery; 8 hours’ autonomy (minimum @
Intrinsically insulated battery powered instrument
charger
max load), 20 hours (max)
No isolation of the USB port
Type approval CE
Standards EN61326-1; EN61010-1
TAB. 2
52 MI004962-E MSC
Page 53
TECHNICAL SPECIFICATIONS
Vout (V)
INTERFACES AND WARNINGS
Buttons
LED
Buzzer
Standard bushings No. 4 x 4mm bushings
Thermocouple attachment
Micro USB
Wireless communication
TAB. 3
Current that can be supplied in voltage generation:
Overload buzzer and impossibility to simulate
Mini plug (7.9mm) for thermocouple measure-
for fw update or modbus communication (virtual
com); charge / supply / PC connection
Bluetooth Low Energy 4.1 to Windows PC,
Smartphone and Android or Ios Tablet
On / O󰀨
Bluetooth reset
Start-up Led
Communication Led
Error Led
PAIRING BT Led
Battery status Led
the value requested
ment and simulation
25
20
15
10
I Load (mA)
55
0
0 5 10 15 20 25 30
Massima
Max
Guaranteed
Garantita
53MI004962-EMSC
Page 54
USER INTERFACE
the instructions on page 59
The MSC calibrator has LED indicators and a buzzer to indicate
different operating states.
Explanatory table of the buzzer signals:
NUMBER OF BEEPS MEANING
2
3 Correct start-up
INTERMITTENT CONTINUOUS
TAB. 4
Led operation table:
FUNCTION LED COLOUR STATUS
PWR GREEN ON: instrument ON
BATTERY AMBER
DATA LOGGER RED
BIND BLUE
COMM AMBER
ERROR RED
RECHARGE (NEAR USB
SOCKET)
Corrupted parameters, correct operation is not
Firmware update mode activated.
To activate the rmware update mode follow
AMBER ON: battery charging
guaranteed
ON: battery charged Flashing: low battery
ON: data logger enabled OFF: data logger disabled
Flashing: connected OFF: disconnected
Flashing: communicating OFF: no communication
ON: error underway OFF: no error
TAB. 5
54 MI004962-E MSC
Page 55
TROUBLESHOOTING
The following section lists the solutions to some common problems:
Switching on/Recharging:
PROBLEM CAUSE SOLUTION
The yellow LED (named "16" in the "product description" diagram on page 11) does not light up while the battery is charging
The MSC calibrator does not turn on
The MSC calibrator does not
turn o󰀨
TAB. 6
- The battery may already be charged
- The supplied USB power unit may be faulty
- the battery may be at - charge the battery
- the MSC calibrator may be blocked
- check that the charger works with another device
- press the reset button and check that the MSC calibrator
turns on and o󰀨 correctly using
the dedicated button.
55MI004962-EMSC
Page 56
to which it is connected
TROUBLESHOOTING
Bluetooth/APP connection:
PROBLEM CAUSE SOLUTION
The MSC calibrator does not connect to the smartphone device via Bluetooth
- The device may not be turned on
- the distance between the smartphone and the MSC cali­brator is excessive
- the smartphone returns "as­signation rejected" when direct connection is attempted
- turn the MSC calibrator on
- reduce the distance between the MSC calibrator and the smartphone
- make sure you have down­loaded and installed the APP on your smartphone or tablet
The smartphone is connect­ed to the MSC calibrator but, when connecting the cables to the terminals, the connection is interrupted
The MSC calibrator does not respond to the commands set on the APP
The MSC calibrator only re­sponds to some commands
The MSC calibrator is dis­connected from the APP and a new connection cannot be made
The calibrator runs abnormally - in case of malfunction, press
- possible electrostatic charge accumulated in your body or in the device to be tested that is transferred to the MSC calibra­tor
- the area in which you intend to operate is subject to strong electromagnetic interference or interference from other Blue­tooth or WiFi devices
- connection was made to an­other MSC calibrator
- discharge the excessive elec­trostatic charge from yourself or the device in order not to transfer it via the wiring to the MSC calibrator
- reduce the distance between the smart phone and the MSC calibrator or, if possible, move to a less disturbed area
- connect to the correct MSC calibrator
-1 Restart the MSC calibrator
-2 Restart the smartphone or tablet or PC
-3 make a new connection
the BT Reset button to release the calibrator from the device
TAB. 7
56 MI004962-E MSC
Page 57
TROUBLESHOOTING
Measurement and generation/simulation problems
PROBLEM CAUSE SOLUTION
The measurement displayed by the APP is not consistent with the actual input /output signal to/from the MSC calibra­tor
The message "out of scale" is displayed on the APP screen
The message "generation error" is displayed on the APP screen
- the area in which you intend to operate is subject to strong electromagnetic interference or interference from other Blue­tooth or WiFi devices
- the battery may be at
- the electrical signal or quan­tity to be measured is outside the measuring range of the MSC calibrator
- the load connected to the output terminals of the MSC calibrator is excessive
- reduce the distance between the smartphone and the MSC calibrator or, if possible, move to a less disturbed area
- charge the battery
- bring the electrical signal to be measured within the ranges listed in the technical specica­tions
- connect a load or device compatible with what is stated
in the technical specications
The message "Supply voltage" is displayed on the APP screen
The measured value is not consistent with the measure­ment taken or the generated / simulated signal
TAB. 8
- the voltage supplied external­ly to the MSC calibrator termi-
nals is too high and a󰀨ects the
internal supply voltages
- possible connection error
- possible presence of a ground connection between USB power supply and device to be tested
- possible external power supply not intended for power generation
- if intended, reduce the val­ue of the external voltage according to what is stated in
the technical specications
and wait 15-20 minutes for the internal protections to reach normal temperature conditions
- make sure that the electrical connections are consistent with the type of simulation / generation to be carried out (see diagrams on page 36)
- use the USB power supply or make sure you have connect­ed the calibrator in compliance with the instructions given on page 13
57MI004962-EMSC
Page 58
General problems
PROBLEM CAUSE SOLUTION
The APP screen does not up­date the measurements taken or to be generated
The measures, or generations, do not match what was expect­ed
TAB. 9
TROUBLESHOOTING
- the area in which you intend to operate is subject to strong electromagnetic interference or interference from other Blue­tooth or WiFi devices
- the battery may be at
- electrical connections may be incorrect
- electrical connections may not be allowed
- connection made to another MSC calibrator
- reduce the distance between the smart phone and MSC cali­brator or, if possible, move to a less disturbed area
- charge the battery
- make sure that the electrical connections are consistent with the type of simulation / generation to be carried out (see diagrams on page 38)
-Use the USB power supply or make sure you have connect­ed the calibrator as shown in Figure 1
- connect to the correct MSC calibrator
For further information, please go to www.seneca.it/msc.
For technical support, please call: +39.049.8705359
or send an email to: supporto@seneca.it
58 MI004962-E MSC
Page 59
FIRMWARE UPDATE
To perform the rmware update of the MSC device, follow the
procedure below:
• Download and install the MSC desktop APP from www.seneca.it/
msc;
• Connect the calibrator to the PC using the USB cable;
• Turn the instrument on by pressing the ON / OFF button and keep it pressed for about 5/6 seconds until the buzzer starts to emit an intermittent and continuous sound;
• At this point release the button;
• During the update phase, the calibrator will have the PWR, ER­ROR and COMM LEDs lit;
• Start the MSC desktop program;
• Select the “ Upgrade” button
• Select the serial communication port;
• Start the "Upgrade" procedure using the “Upgrade” button;
• Wait for the end of the rmware upload by looking at the Log
buffer;
• Once the rmware "Upload" has been completed, restart the
calibrator using the PWR key.
If the procedure has been carried out correctly when the instrument is started up, three beeps will sound as indicated in table 4 on page
54.
59MI004962-EMSC
Page 60
Page 61
Page 62
Via Austria, 26 – 35127 – PADOVA – ITALY
Tel. +39.049.8705359 - 8705355 - Fax +39.049.8706287
The information in this document can be modied or supplemented without notice for technical or commercial reasons. The gures and diagrams are
indicative and not binding. Neither can inconsistencies or imperfections be excluded, in spite of the continuous search for perfection. The content of this
document undergoes a periodical update anyway. Any unauthorised reproduction is strictly forbidden.
SENECA s.r.l.
Loading...