The MSC calibrator is a portable multi-function instrument for the
generation and measurement of signals commonly found in automation environments.
It was designed to help professionals who have to verify and develop signal processing equipment connected to automatic control
devices.
Unlike what the market offers, the MSC calibrator combines
simulations and measurements in one instrument using the user’s
smartphone, tablet or Windows PC (via the USB cable) for an
intuitive and modern interface.
The MSC multifunction process calibrator is a hand-held instrument, powered by rechargeable batteries, used to measure and
generate electrical quantities.
In addition to the functions described in the "FUNCTIONS" paragraph (page 15), the calibrator has the following characteristics:
• Input/output terminal for thermocouples (TC) and isothermal
internal block with automatic temperature compensation of the
reference junction.
• 4 standard 4mm diameter bushings for 2-3-4 wire sizes.
• Protection against overvoltages up to 250 Vac.
• Lithium battery power supply, with up to 14h autonomy.
• Bluetooth connection with tablet and smartphone.
• High accuracy class: 0.05% basic class for each type of input/
output.
• Data always available thanks to the dedicated app.
6MI004962-EMSC
Page 7
• Micro USB connector for power supply and PC connection.
The analogue signals managed both in measurement and generation are:
• Milliamperes (dc)
• Volts, millivolts (dc)
• Thermocouples, two/three or four-wire heating elements
• 350 Ω bridge extensometer (LOAD CELL)
The following are also managed:
• Pulse generation with duration, voltage amplitude and variable
duty cycle
• Pulse measurement with adjustable threshold
• Frequency measurement and generation
The device is protected against damage due to incorrect connections such as short circuits and overvoltages applied to the inputs
(e.g. 230 Vac).
The simplicity of use of the connections and the protections against
overload make the device suitable even for the less experienced
and in a hobby and educational environment.
It is possible to connect multiple calibrators to a smartphone and
control them individually.
Electrical and electronic waste disposal (applicable in the European Union and other countries with
recycling). The symbol on the product or its packaging shows the product must be surrendered to a
collection centre authorized to recycle electrical and electronic waste.
Damaged parts must be replaced by the manufacturer, who is also responsible for repairing the
device. The product is sensitive to electrostatic discharges. Take appropriate measures during any
operation.
7MI004962-EMSC
Page 8
PRODUCT DESCRIPTION
Socket for thermocouple measurement/generation
1
Measurement/generation bushing -EX
2
Measurement/generation bushing -SN
3
4
Measurement/generation bushing +SN
Measurement/generation bushing +EX
5
ON/OFF button
6
Bluetooth RESET button
7
PWR ON Led
8
Established connection Bind Led
9
Battery status indicator Led
10
Bluetooth/USB communication Led
11
Data recording Led
12
Error message Led
13
Power/communication micro USB connector
14
RESET button
15
Battery charge warning Led
16
8MI004962-EMSC
Page 9
1
-EX-SN+SN+EX
10
12
2
3
6
8
4
5
7
9
11
13
14
16
15
9MSC
Page 10
SAFETY INFORMATION
This manual contains safety rules that must be complied with to
safeguard personal safety and to prevent damage to property. The
indications to be followed to guarantee personal safety are highlighted by a triangle with the following meanings:
WARNING
the symbol associated with the word ‘warning’ indicates conditions
or actions that put the user's safety at risk.
CAUTION
The symbol associated with the word ‘caution’ indicates conditions
or actions that might damage the calibrator or equipment tested.
Qualied personnel
The product covered by this documentation may only be used by
qualied personnel for the respective assigned task, in compliance
with the documentation relating to the task, and especially with the
safety warnings and the precautions contained therein. By virtue of
their training and experience, qualied personnel can recognize the
risks related to the use of this product and avoid possible dangers.
Disclaimer
We have checked that the contents of this documentation correspond to the hardware and software described. However, since
we cannot rule out any differences, we cannot guarantee a perfect
match. However, the content of this documentation is periodically
checked and any corrections or changes are inserted in subsequent editions.
10MI004962-EMSC
Page 11
WARNING
• Do not apply different voltages or higher voltages than indicated
between terminals, or between any terminal and earth (voltages
greater than 50Vdc)
• Do not use the calibrator if damaged, or if signs of possible dam-
age are visible on it.
• Do not remove the internal battery.
• Do not use the calibrator if it runs abnormally.
• Use the reset button if there is an anomaly in the operation.
• Do not touch the calibrator terminals during use.
• To carry out the measurements, use the cables supplied or any-
way adequate for the measurements to be taken.
•
Select the appropriate range for the desired measurement or generation.
CAUTION
• Make sure the internal battery is charged if you intend to use the
calibrator without power from the USB port.
• Use the calibrator as described in this manual.
• Before each measurement or generation, refer to the electrical
connections shown in the manual.
• Do not use the calibrator in environments containing combustible/
explosive dust, gas or vapour.
The MSC calibrator has been designed according to EN 6101-1 standards
11MI004962-EMSC
Page 12
USER GUIDE
The calibrator is a device that can measure or generate electrical
quantities to test devices.
To proceed with the correct use of the instrument, make sure to
follow the following steps:
• Turn on the device and connect with the APP via Bluetooth.
• Wait 5 minutes for the appliance temperature to stabilize.
• Connect the cables correctly, according to the type of measurement/generation to be carried out: refer to the connection
diagrams on p. 36.
It is however necessary, before use, to make sure that the battery
is fully charged so as to guarantee its life for the generations or
measurements that will be carried out. Before each use it is advisable to recharge the instrument using the USB cable supplied, until
the LED battery charge indicator on the bottom of the calibrator
goes off, as shown in the diagram on page 9.
The calibrator can maintain the signals to be generated (set by the
user) within a determined error range.
During generation, the calibrator will constantly check the generated signal, warning about any generation error (where provided for)
by notifying it on the APP.
CAUTION
The calibrator is protected against the risk of applying temporary
overvoltages. However, it is good to remember that in this case the
device needs 15-20 minutes to restore the normal conditions.
If this period of time is not observed, the accuracy of the measurements and/or generations cannot be guaranteed.
12MI004962-EMSC
Page 13
Be careful when using the calibrator with devices that have earth
connections. Connections 1 and 2, shown in FIG. 1, can exist, but
not simultaneously.
This working condition can interfere with the correct operation of
the calibrator and the Bluetooth connection.
If this connection is made during the generation of an electrical
quantity, internal protections may intervene, in which case, once
the correct electrical connection between the devices has been
restored, it is necessary to leave the calibrator inactive for about
15-20 minutes to go back to normal conditions.
In order to achieve proper PC//measurement systems (EUT)
isolation and to protect the USB port of your PC, you must use the
USB-ISO accessory as shown in Fig. 2.
12
USB-ISO
FIG. 1
FIG. 2
13MI004962-EMSC
Page 14
USING THE CALIBRATOR
-EX-SN+SN+EX
CABLES SUPPLIED
BLACK CABLE WITH 4MM BLACK PLUG
14MI004962-EMSC
WHITE CABLE WITH 4MM BLACK PLUG
WHITE CABLE WITH 4MM RED PLUG
RED CABLE WITH 4MM RED PLUG
Page 15
CALIBRATOR FUNCTIONS
MEASUREMENT FUNCTIONS:
The MSC calibrator allows measuring the following signals:
• passive current
• active current
• HiRange voltage
• LowRange voltage
• thermocouple
• resistance thermometer
• load cell
• frequency/pulses
For correct measurement, connect the cables as shown in the
diagrams on page 36 and page 37.
GENERATION FUNCTIONS:
The MSC calibrator allows measuring the following signals:
• passive current
• active current
• HiRange voltage
• LowRange voltage
• thermocouple
• resistance thermometer
• load cell
• frequency/pulses
For correct measurement, connect the cables as shown in the
diagrams on page 38 and page 39.
15MI004962-EMSC
Page 16
MEASUREMENTS
Checks common to all measurements
The following are the checks that are common to all measurements:
Sample screen from the APP:
The images and diagrams shown must be considered as indicative and not binding
since they are subject to possible updates.
16MI004962-EMSC
Page 17
In the sample screen it is possible to distinguish the checks that
are common to all the measurements (red) and those common to
analogue measurements (orange) only.
In red you can identify the checks that have the following meanings:
• Zeroing of MIN / AVG / MAX measurements, for analogue
measurements
• Count zeroing for pulse measurement
• Pausing of analogue or digital measurements
• Sharing the measurements taken. This button allows for example sharing a memo of the measurements just made.
• Datalogger: the activation of this function allows starting to
record the measurements that are being taken.
The MIN / AVG / MAX measurements are identied in orange and
the check allows changing the display from primary to secondary measurement and vice versa.
17MI004962-EMSC
Page 18
CURRENT / VOLTAGE MEASUREMENTS
The calibrator can measure voltages and currents that are applied
to the two innermost bushings (SN+ SN-) according to the following
ranges.
For the voltages, two measurement scales are available:
• Voltage -10÷90 mV ( LowRange)
• Voltage 0÷27 V ( HiRange)
To carry out the voltage measurement, connect the calibrator
according to gures M5 or M7 while, to make a current measurement, connect the calibrator according to FIG. M1 or M3 (page 36).
Once the calibrator is connected and the Bluetooth connection is
working, press M (Measurement) on the Home screen of the APP
and select "CURRENT / VOLTAGE".
Then select the desired size.
Since the measurements are homogeneous in the way they are
presented by the APP, a measurement is then taken into consideration; then choose "VOLTAGE 0.27 V".
The calibrator starts displaying the values related to the voltage
applied to the input bushings in Volts. The default value displayed
is in Volts. The measurement can be displayed as a percentage of
the full scale by pressing the P9 button (double arrow symbol)
If the measurement is unstable, ltering can be applied. To activate
it, follow the following steps by pressing:
1. Settings (refer to image on page 47)
2. AVG lter settings
Then select the desired ltering time between: 0.25, 0.5, 1, 2, 4, 8
seconds.
18MI004962-EMSC
Page 19
FREQUENCY MEASUREMENTS
The calibrator allows the measurement of the voltage pulses
applied in the two innermost bushings (SN+ SN-). The electrical
connections are shown in FIG. M8
Once the calibrator is connected and the Bluetooth connection
works, press M (Measurement) on the APP Home screen and then
choose "FREQUENCY"
To correctly detect a frequency of a voltage signal applied to the
bushings, it is important to choose an appropriate trigger threshold
by going to:
1 Settings
2 Set trigger level
The calibrator shows the frequency in Hz.
19MI004962-EMSC
Page 20
THERMOCOUPLE MEASUREMENTS
The calibrator allows the temperature measurement with thermocouples which are connected via connector 1 (diagram on page 9).
The electrical connections are shown in FIG. M2 (page 36).
For the types of supported thermocouples, refer to TAB 1 (page
50).
Once the calibrator is connected and the Bluetooth connection
works, press M (Measurement) on the APP Home screen and then
choose "THERMOCOUPLE". Then select the type of thermocouple
you wish to take the measurement with.
In measurement mode the temperature value in °C (and mV) is
shown. The displayed temperature value is the real temperature
measured by the thermocouple, including the cold junction com-
pensation (depending on the set conguration). This last value is
shown on the display with the wording
“C.J. Read”.
On the other hand, the value in mV is the real voltage value read
by the terminals without any compensation.
If the measurement is unstable, ltering can be applied.
To activate it, follow the following steps by pressing:
1. Settings
2. AVG lter settings
Then select the desired ltering time between: 0.25, 0.5, 1, 2, 4, 8
seconds.
20MI004962-EMSC
Page 21
CAUTION
Wait until the device reaches a stable thermal condition before
taking measurements with the thermocouples (typically at least 30
minutes).
CAUTION
If the thermocouple does not have its own connector, use the innermost bushings SN+ SN- as an alternative, bearing in mind that the
compensation of the cold junction in this case may be less precise.
21MI004962-EMSC
Page 22
RESISTANCE THERMOMETER MEASUREMENTS (RTD)
The calibrator allows the measurement of the temperature with
resistance thermometers connected with 2, 3 or 4 wires that are
connected to the bushings, according to the types indicated in
TAB 1 (page 50). The electrical connections are shown in FIG. M4
(page 36).
Once the calibrator is connected and the Bluetooth connection
works, press M (Measurement) on the APP Home screen and then
choose "RTD". Then select the type of resistance thermometer you
want to measure.
In RTD measurement mode, the temperature value in °C is shown
according to the type of resistance thermometer selected, along
with the value measured in Ω.
If the measurement is unstable, ltering can be applied. To activate
it, follow the following steps by pressing:
1. Settings
2. AVG lter settings
Then select the desired ltering time between: 0.25, 0.5, 1, 2, 4, 8
seconds.
22MI004962-EMSC
Page 23
PULSE MEASUREMENTS
The calibrator allows the measurement of the voltage pulses
applied in the two innermost bushings (SN+ SN-). The electrical
connections are shown in FIG. M8 (page 37).
Once the calibrator is connected and the Bluetooth connection
works, press M (Measurement) on the APP Home screen and then
choose "PULSE" followed by the mode in which the pulse is to be
detected: "POSITIVE FRONT SIGNAL WITH PULSES" or "NEGATIVE FRONT SIGNAL WITH PULSES" (the pulse is detected on
the rising or falling edge when crossing the set trigger threshold)
To correctly detect the pulses it is important to choose an appropriate trigger threshold by going to:
1. Settings
2. Set trigger level
It is possible to lter the pulses applied to the input bushings by applying the pulse minimum duration lter. The pulse is then detected
only if its duration is greater than or equal to the set value.
To change this parameter follow the following steps:
1. Settings
2. Set minimum pulse duration
then set the value expressing it in micro seconds.
23MI004962-EMSC
Page 24
MEASUREMENTS WITH LOAD CELLS
The calibrator is compatible with load cells having a rated resist-
ance of 350 Ω.
Be sure to follow the guidelines for use to take measurements
correctly.
By activating this measurement function, the device supplies the
load cell through the outermost terminals (+EX -EX) while the generated signal is read via the innermost terminals (+SN -SN)
To take the measurement, connect the calibrator according to g.
M6 (page 37); on the Home screen, press M (Measurement) and
then "Load Cell".
The measurement of the load cell is displayed in two ways:
• mV/V ratio
• weight in Kg
To set the sensitivity of the load cell sensor or set the tare and full
scale in Kg press:
1. Settings
2. Options
In the "load cell settings" screen you can:
• Set the unit of measurement
• Set cell sensitivity [mV/V]
• Set the load cell capacity [Kg]
• Zero the tare: before proceeding to zero the tare, make sure
24MI004962-EMSC
Page 25
that the measurement on the display is stable then proceed with
resetting using the reset button:
TARE
If the measurement is unstable, ltering can be applied. To activate
it, follow the following steps by pressing:
3. Settings
4. AVG lter settings
Then select the desired ltering time between: 0.25, 0.5, 1, 2, 4, 8
seconds.
CAUTION
The 4-wire measurement of the load cell is inuenced by the resistance of the conductors supplying it. If it is not possible to shorten
the connections between the load cell and the MSC calibrator, by
measuring the resistance only of the conductors that supply the
350+R
K=
+EX +R-EX
350
load cell (conductors +EX and -EX), the correction coefcient can
be calculated:
The obtained coefcient must be multiplied with the value measured by the calibrator to obtain the real value.
25MI004962-EMSC
Page 26
GENERATION / SIMULATION
The following are the checks that are common to all measurements:
APP sample screen:
The images and diagrams shown must be considered as indicative and not binding
since they are subject to possible updates.
26MI004962-EMSC
Page 27
In the sample screen, a distinction is made between the checks
common to all analogue generations (in orange) and those of the
pulse generation only (in red).
In Orange it is possible to identify the area where the generation
value can be set by using the up/down swipe and the keyboard to
enter the desired value. The other checks are listed below:
• Button indicating the status of the generation (active or not).
Pressing it, the output is deactivated and the button changes
status.
• Command to reverse the primary and secondary measurements and vice versa
• Ramp function start button.
In red we can identify the elements that concern the generation of
pulses; in the example we can identify the 8 remaining pulses to be
generated and the three buttons:
• START: starts the generation of voltage pulses
• PAUSE: temporarily interrupts the generation of pulses
• STOP: cancels the current pulse generation session
27MI004962-EMSC
Page 28
CURRENT / VOLTAGE GENERATION
The calibrator can operate as a voltage or current generator. These
quantities are supplied by the innermost bushings (SN+ SN-)
according to the following ranges.
For the voltages two generation scales are available:
• Voltage -10÷90 mV
• Voltage 0÷27 V
To carry out voltage generation, connect the calibrator according
to gure G7 (page 39), while to generate a current, connect the
calibrator according to gures G1 or G3 (page 38) depending on
whether you want to generate a passive or active current.
Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and
choose "CURRENT / VOLTAGE".
Then select the desired voltage or current generation.
Since the generations are homogeneous in the way they are
presented by the APP, the following shows a sample generation.
Select "VOLTAGE 0.27 V".
The generation screen is displayed and the calibrator starts supplying the output bushings with the preset voltage of 0 V. To change
this value, simply enter the desired value using the numeric keypad
and conrm with OK.
28MI004962-EMSC
Page 29
THERMOCOUPLE SIMULATION
The calibrator can simulate a thermocouple. To see which types
of thermocouples are supported, refer to
simulation of thermocouples is made possible by the CONN. TC,
of the diagram on page 9 or through the SN+ and SN- terminals.
The electrical connections are shown in FIG G2 (page 38). Once
the calibrator is connected and the Bluetooth connection works,
press G (signal Generator) on the APP Home screen and choose
"THERMOCOUPLE".
Then select the type of thermocouple you want to simulate; the
following is an example of the simulation of a type K thermocouple.
A screen relative to the simulation of type K thermocouples is displayed on the APP. It is possible to enable or disable the compen-
sation in the cold junction, by default automatically set on internal
temperature. It is possible to change this setting by going to:
1 Settings
2 Cold junction setting
3 Enable the "Manual setting” control
4 Set the temperature value for the cold junction and then
conrm
CAUTION
Wait until the device reaches a stable thermal condition before
simulating a thermocouple.
If the thermocouple does not have its own connector, use the innermost bushings SN+ SN- as an alternative, bearing in mind that the
compensation of the cold junction in this case may be less precise.
TAB 1 (page 50).
The
29MI004962-EMSC
Page 30
RESISTANCE THERMOMETER SIMULATION (RTD)
The calibrator is able to simulate a resistance thermometer (RTD)
connected to the 2, 3 or 4 wire bushings. The types of supported
RTDs are listed in TAB 1 (page 50).
To carry out RTD simulation, connect the calibrator according to
FIG.G4 (page 38).
Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and
choose "RTD".
Then select the type of RTD that you want to simulate; the simulation of a PT100-type resistance thermometer is shown as an
example.
Then enter the desired value in °C and conrm; it is possible to
switch the display of the unit of measurement from °C to Ω.
Thermal resistance generation can be divided into two scales; the
PT100 scale and the PT1000 scale (which includes the PT500).
For these two scales the excitation current from the external instrument must remain within a certain range (see table on page 50) . If
this condition is not met, an Generation error is signalled, indicating
the possibility that the simulated resistive value is not correct.
CAUTION
30MI004962-EMSC
Page 31
LOAD CELL SIMULATION
The calibrator can simulate a load cell (LOAD CELL) with a nominal resistance of 350 Ω. The connection is made by 4 wires through
the bushings EX+ EX-, SN+ SN-. The supply voltage applicable
to the external bushings has a range from 5V to 10V. To simulate
a load cell, connect the calibrator according to FIG. G6 (page 39).
Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and
then select "LOAD CELL".
From the load cell screen, proceed to set the parameters by following the steps below:
1 Settings
2 Options
3 Set the parameters "Unit of measurement", "Load cell gain" in
mV/V, and then the full scale of the load cell.
For instance, once the sensitivity is set to 2 mV/V, the scale to
Kg and the full scale to 1000 Kg, simply conrm the parameters,
then return to the main page of the load cell simulation and enter
the value in Kg that you wish to simulate. The value can also be
entered as mV/V
CAUTION
Wait until the device reaches a stable thermal condition before
simulating a load cell.
31MI004962-EMSC
Page 32
FREQUENCY GENERATION
The calibrator can generate a square wave voltage signal with 50%
duty cycle and settable frequency and peak voltage values. The
generated signal is taken out to the bushings (SN+ SN-).
The denable properties of the pulses are:
• Value ON: value in Volts assigned to the square wave when
the signal is at high logic level
• Value OFF: value in Volts assigned to the square wave when
the signal is at low logic level.
To generate a voltage square wave signal, connect the calibrator
according to FIG. G8 (page 39).
Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and
then select "FREQUENCY".
To proceed with the generation, set the band parameters of the
square wave by following the following steps:
1 Settings
2 Set Pulse Parameter
3 Enter the desired Value ON / Value OFF band values and
conrm with OK
Then set the desired frequency using the numeric keypad and
make sure that the output enable button is turned to ON.
The primary measurement unit of the signal is in Hz, switchable to
seconds to show the length of the period.
32MI004962-EMSC
Page 33
CAUTION
It is important to note that the time resolution according to which
the square wave signal is generated is 50 μs. This implies an error
that is dependent on the numeric value set.
For instance, you want to generate a signal with a frequency of 150
Hz whose corresponding period is 6.666 ms; the number of counts
necessary to obtain this frequency is obtained through:
6.666 mHz /50 µs=133.33. The system then rounds off to the
nearest whole number, 133, which it will use to generate the output
signal. The output frequency will have the value:
FOUT= 1 = 150.37 Hz
133X50 µs
The APP will automatically show the value actually generated.
33MI004962-EMSC
Page 34
PULSE GENERATION
The calibrator can generate a train of voltage pulses available at
the two innermost bushings (SN+ SN-).
The denable properties of the pulses are:
• Value ON / Value OFF: values in V relative to the pulse value
and to the value of the steady state of OFF respectively.
• Time ON / Time OFF: values in ms that dene the period T of
the pulse. ( T=TimeON + TimeOFF).
Time ON is the time in which the pulse remains at the Value ON
value in ms.
To generate voltage pulses, connect the calibrator according to
FIG. G8 (page 39).
Once the calibrator is connected and the Bluetooth connection
works, press G (signal Generator) on the APP Home screen and
then select "PULSE".
To proceed with the generation, set the time and band parameters
of the pulse by following the following steps:
1. Settings
2. Set Pulse Parameter
3. Enter the desired band Value ON / Value OFF values and the
times corresponding to the Time ON / Time OFF pulse period
and conrm with OK
Then set the number of pulses and press start to start the generation; the remaining pulses are displayed at the same time. During
this phase it is possible to pause and then resume the generation
by pressing PAUSE or cancel it by pressing STOP.
34MI004962-EMSC
Page 35
ERRORS APP
ERROR SIGNALS ON APP / LEDSTATUS
ERROR MESSAGEMEANINGLED STATUS
Generation Error
Measure Error
Loop Error
The set generation
is not guaranteed
Signal out of scale
Only for current
supplied by an ex-
ternal generator.
(if V
> 29 V)
Loop
Low Battery
Error Led ON
(Red)
Error Led ON
(Red)
Error Led ON
(Red)
Low Battery Led ON
blinking (Amber)
35MI004962-EMSC
Page 36
-EX-SN+SN+EX
FIGURA M5: MISURATORE DI TENSIONE HiRange
V
+
-EX-SN+SN+EX
Kg
FIGURA M6: MISURATORE CELLA DI CARICO
-EX-SN+SN+EX
FIGURA M7: MISURATORE DI TENSIONE LowRange
mV
+
-EX
-SN+SN
+EX
FIGURA M8: MISURATORE DI FREQUENZA/IMPULSI
Hz
+
EXCITATION -
SENSE -
SIGNAL -
SIGNAL +
SENSE +
EXCITATION +
FIGURA M1:
MISURATORE DI CORRENTE PASSIVO
(trasmettitore a 4 fili)
-EX-SN+SN+EX
mA
FIGURA M3:
MISURATORE DI CORRENTE ATTIVO
(trasmettitore a 2 fili)
FIGURA M2:
MISURATORE TEMP. CON TERMOCOPPIA
-EX -SN+SN +EX
°C
FIGURA M4:
MISURATORE DI TERMORESISTENZA
3 W
2 W
2 W
4 W
-EX-SN+SN+EX
mA
36
-SN+SN
-EX
°C
+EX
Page 37
FIGURA M5:
MISURATORE DI TENSIONE HiRange
+
FIGURA M6:
MISURATORE CELLA DI CARICO
-EX-SN+SN+EX
V
FIGURA M7:
MISURATORE DI TENSIONE LowRange
+
-EX-SN+SN+EX
EXCITATION -
SENSE -
SIGNAL -
-EX -SN+SN +EX
SIGNAL +
SENSE +
EXCITATION +
Kg
FIGURA M8:
MISURATORE DI FREQUENZA/IMPULSI
+
-SN+SN
-EX
+EX
mV
Hz
37
Page 38
-EX-SN+SN+EX
FIGURA G5: GENERATORE DI TENSIONE HiRange
-EX-SN+SN+EX
FIGURA G6: SIMULATORE CELLA DI CARICO
-EX-SN+SN+EX
FIGURA G7: GENERATORE DI TENSIONE LowRange
-EX
-SN+SN
+EX
FIGURA G8: GENERATORE FREQUENZA/IMPULSI
V
+
+
mV
Hz
Kg
FIGURA G1:
GENERATORE DI CORRENTE PASSIVO
FIGURA G2:
SIMULATORE TERMOCOPPIA
mA
-EX-SN+SN+EX
FIGURA G3:
GENERATORE DI CORRENTE ATTIVO
mA
-EX-SN+SN+EX
°C
-EX -SN+SN +EX
FIGURA G4:
SIMULATORE DI TERMORESISTENZA
°F
°C
3 W4 W
2 W2 W
-SN+SN
-EX
+EX
38
Page 39
FIGURA G5:
GENERATORE DI TENSIONE HiRange
FIGURA G6:
SIMULATORE CELLA DI CARICO
V
-EX-SN+SN+EX
+
FIGURA G7:
GENERATORE DI TENSIONE LowRange
mV
-EX-SN+SN+EX
Kg
-EX -SN+SN +EX
FIGURA G8:
GENERATORE FREQUENZA/IMPULSI
Hz
-SN+SN
-EX
+EX
+
39
Page 40
USING THE APP FROM A MOBILE
To use the MSC calibrator, you need to install the free "MSC by
Seneca" APP available for iOs and Android devices on your smartphone or tablet and register your details.
Ensure that your tablet and / or smartphone is equipped with Bluetooth 4.1 or higher.
To install the App on a smartphone or tablet, scan the QR code
shown and enter the required details in the registration page.
When the application starts, select "Search" and, in the list of available devices, select the desired device to start the synchronization
with your smartphone.
To ensure the calibrator is working properly, make sure you have
selected the correct device.
40MI004962-EMSC
Page 41
MAIN SCREEN:
Navigation menu
1
“Search” command
2
When the application starts,
select the “Search”
command to view the devices
to be associated.
1
2
41MI004962-EMSC
Page 42
DEVICE SEARCH
Devices detected
1
When the app is opened,
the device will automatically
recognize the terminal to join.
To complete the operation
simply select the device you
want to interface with.
Stops the search
2
1
2
42MI004962-EMSC
Page 43
Activates the sound signal in
1
the event of an alarm
Activates the sound signal
2
on overow in generation
Activates the sound signal
3
on turning on
Activates the sound signal
4
during data recording
If selected, once the
5
available memory is full,
the instrument will write
the data on the recordings
already made
Auto power OFF function:
6
this function allows setting
a measurement time or
signal generation, at the
end of the operation the
instrument will automatical-
ly turn o.
Allows setting the desired
7
sampling time
OPTIONS
1
2
3
4
5
6
7
8
9
1110
Space dedicated to setting
8
the desired sampling time
Unit of measurement setting
9
Setting conrmation
10
Cancels the settings
11
43MI004962-EMSC
Page 44
Navigation menu
1
Meter mode
2
HOME SCREEN
1
Generator/simulator mode
3
Instrument name
4
Calibrator characteristics
5
Device disconnect
6
button
23
4
5
6
44MI004962-EMSC
Page 45
MEASUREMENT CONTROLS
Calibrator battery level
1
Access to the last measure-
2
ment taken
Current / voltage meter
3
Thermocouple meter
4
Resistance temperature meter
5
Load cell meter
6
Pulse meter
7
Frequency meter
8
1
2
3
4
5
6
7
8
45MI004962-EMSC
Page 46
GENERATION/SIMULATION CONTROLS
Access to the last generation
1
carried out
Current / voltage generator
2
Thermocouple simulator
3
1
Resistance temperature
4
simulator
Load cell simulator
5
Pulse generator
6
2
3
4
5
7
Frequency generator
6
7
46MI004962-EMSC
Page 47
METER READING SCREEN
Meter setting
1
Calibrator battery level
2
Reading screen: primary
3
(a)
measurement
Reading screen: second-
3
(b)
ary measurement
Command to reverse the pri-
4
mary and secondary measurements and vice versa
TARE command
5
Minimum measurement
6
Medium measurement
7
Maximum measurement
8
Datalogger function:
9
by selecting this function, measurement
recording starts.
3
(a)
5
TARE
3
(b)
678
1
2
4
Share minimum, medium and
10
maximum value measurements
Suspend minimum, medium
11
and maximum value measurements
Update minimum, medium
12
and maximum value measurements
12
11
9
10
47MI004962-EMSC
Page 48
GENERATOR/SIMULATOR READING SCREEN
Generator settings
1
Set value
2
Ramp start
3
Unit of measurement
4
management
Status indicator
5
Quick output levels
6
2
3
1
4
Setting keypad
7
output
5
6
7
48MI004962-EMSC
Page 49
RAMP READING SCREEN
Generated signal
1
Ramp generation start/
2
pause
Ramp management
3
Table to dene the points
4
for ramp creation
1
3
2
4
49MI004962-EMSC
Page 50
FEATURE SUMMARY
PULSES / FREQUECY
CMRR-NMRR
Note
RANGEPRECISION * RESOLUTIONRULE
Note
GENERATIONMEASURE
0÷240,04 + 0,011 uA18>60 dB
4,21,22
*:% of the measure + % of the scale
MI004961-IMSC
FUNCTIONU.MRANGEPRECISION * RESOLUTION
ACTIVE CURRENT [dc mA]0,1÷240,04 + 0,011 uA30÷240,04 + 0,011 uA17>60 dB
VOLTAGE [hi range] [dc V]0÷26 min 0,02 + 0,011 mV10÷26 min0,02 + 0,011 mV9>100 dB
VOLTAGE [lo range] [dc mV]-10÷900,03 + 0,015 uV2-10÷900,03 + 0,015 uV9>100 dB
PASSIVE CURRENT [dc mA]0,1÷240,04 + 0,011 uA
Pt100[°C]-200÷8590,02 + 0,010,15,19-200÷8500,03 + 0,020,1IEC 607516>140 dB
Pt500[°C]-200÷8590,03 + 0,010,15,20-200÷8500,03 + 0,020,1IEC 607516>140 dB
Pt1000[°C]-200÷8590,03 + 0,010,15,20-200÷8500,03 + 0,020,1IEC 607517>140 dB
CU50, CU100[°C]-180÷2000,02 + 0,010,115,19-180÷2000,03 + 0,020,1GOST 6651-2009 6,15>140 dB
NI100, NI120[°C]-80÷2600,02 + 0,010,116,19-60÷2500,03 + 0,020,1DIN 437606,16>140 dB
THERMOCOUPLE J [°C]-210÷12000,03 + 0,010,112,2-210÷12000,03 + 0,010,1EN 60584-1:1997 9,12>100 dB
THERMOCOUPLE T [°C]-270÷4000,03 + 0,010,112,2-200÷4000,03 + 0,010,1EN 60584-1:1997 9,12>100 dB
THERMOCOUPLE K [°C]-270÷13720,03 + 0,010,112,2-200÷13720,03 + 0,010,1EN 60584-1:1997 9,12>100 dB
THERMOCOUPLE E [°C]-270÷10000,03 + 0,010,112,2-200÷10000,03 + 0,010,1EN 60584-1:1997 9,12>100 dB
THERMOCOUPLE S [°C]-50÷17680,03 + 0,010,512,2-50÷17680,03 + 0,010,5EN 60584-1:1997 9,12>100 dB
THERMOCOUPLE N [°C]-270÷13000,03 + 0,010,112,2-200÷13000,03 + 0,010,1EN 60584-1:1997 9,12>100 dB
THERMOCOUPLE R [°C]-50÷17680,03 + 0,010,512,2-50÷17680,03 + 0,010,5EN 60584-1:1997 9,12>100 dB
THERMOCOUPLE L [°C]-200÷8000,03 + 0,010,112,2-200÷8000,03 + 0,010,1Gost 8.585-2001 9,12>140 dB
THERMOCOUPLE B [°C]0÷18200,03 + 0,010,512,2250÷18200,03 + 0,010,5EN 60584-1:1997 9,12>100 dB
Current that can be supplied in voltage generation:
Overload buzzer and impossibility to simulate
Mini plug (7.9mm) for thermocouple measure-
for fw update or modbus communication (virtual
com); charge / supply / PC connection
Bluetooth Low Energy 4.1 to Windows PC,
Smartphone and Android or Ios Tablet
On / O
Bluetooth reset
Start-up Led
Communication Led
Error Led
PAIRING BT Led
Battery status Led
the value requested
ment and simulation
25
20
15
10
I Load (mA)
55
0
051015202530
Massima
Max
Guaranteed
Garantita
53MI004962-EMSC
Page 54
USER INTERFACE
the instructions on page 59
The MSC calibrator has LED indicators and a buzzer to indicate
different operating states.
Explanatory table of the buzzer signals:
NUMBER OF BEEPSMEANING
2
3Correct start-up
INTERMITTENT CONTINUOUS
TAB. 4
Led operation table:
FUNCTIONLED COLOURSTATUS
PWRGREENON: instrument ON
BATTERYAMBER
DATA LOGGERRED
BINDBLUE
COMMAMBER
ERRORRED
RECHARGE (NEAR USB
SOCKET)
Corrupted parameters, correct operation is not
Firmware update mode activated.
To activate the rmware update mode follow
AMBERON: battery charging
guaranteed
ON: battery charged
Flashing: low battery
ON: data logger enabled
OFF: data logger disabled
Flashing: connected
OFF: disconnected
Flashing: communicating
OFF: no communication
ON: error underway
OFF: no error
TAB. 5
54MI004962-EMSC
Page 55
TROUBLESHOOTING
The following section lists the solutions to some common problems:
Switching on/Recharging:
PROBLEMCAUSESOLUTION
The yellow LED (named "16"
in the "product description"
diagram on page 11) does not
light up while the battery is
charging
The MSC calibrator does not
turn on
The MSC calibrator does not
turn o
TAB. 6
- The battery may already be
charged
- The supplied USB power unit
may be faulty
- the battery may be at- charge the battery
- the MSC calibrator may be
blocked
- check that the charger works
with another device
- press the reset button and
check that the MSC calibrator
turns on and o correctly using
the dedicated button.
55MI004962-EMSC
Page 56
to which it is connected
TROUBLESHOOTING
Bluetooth/APP connection:
PROBLEMCAUSESOLUTION
The MSC calibrator does not
connect to the smartphone
device via Bluetooth
- The device may not be turned
on
- the distance between the
smartphone and the MSC calibrator is excessive
- the smartphone returns "assignation rejected" when direct
connection is attempted
- turn the MSC calibrator on
- reduce the distance between
the MSC calibrator and the
smartphone
- make sure you have downloaded and installed the APP
on your smartphone or tablet
The smartphone is connected to the MSC calibrator but,
when connecting the cables to
the terminals, the connection is
interrupted
The MSC calibrator does not
respond to the commands set
on the APP
The MSC calibrator only responds to some commands
The MSC calibrator is disconnected from the APP and
a new connection cannot be
made
The calibrator runs abnormally- in case of malfunction, press
- possible electrostatic charge
accumulated in your body or in
the device to be tested that is
transferred to the MSC calibrator
- the area in which you intend
to operate is subject to strong
electromagnetic interference or
interference from other Bluetooth or WiFi devices
- connection was made to another MSC calibrator
- discharge the excessive electrostatic charge from yourself
or the device in order not to
transfer it via the wiring to the
MSC calibrator
- reduce the distance between
the smart phone and the MSC
calibrator or, if possible, move
to a less disturbed area
- connect to the correct MSC
calibrator
-1 Restart the MSC calibrator
-2 Restart the smartphone or
tablet or PC
-3 make a new connection
the BT Reset button to release
the calibrator from the device
TAB. 7
56MI004962-EMSC
Page 57
TROUBLESHOOTING
Measurement and generation/simulation problems
PROBLEMCAUSESOLUTION
The measurement displayed
by the APP is not consistent
with the actual input /output
signal to/from the MSC calibrator
The message "out of scale" is
displayed on the APP screen
The message "generation
error" is displayed on the APP
screen
- the area in which you intend
to operate is subject to strong
electromagnetic interference or
interference from other Bluetooth or WiFi devices
- the battery may be at
- the electrical signal or quantity to be measured is outside
the measuring range of the
MSC calibrator
- the load connected to the
output terminals of the MSC
calibrator is excessive
- reduce the distance between
the smartphone and the MSC
calibrator or, if possible, move
to a less disturbed area
- charge the battery
- bring the electrical signal to
be measured within the ranges
listed in the technical specications
- connect a load or device
compatible with what is stated
in the technical specications
The message "Supply voltage"
is displayed on the APP screen
The measured value is not
consistent with the measurement taken or the generated /
simulated signal
TAB. 8
- the voltage supplied externally to the MSC calibrator termi-
nals is too high and aects the
internal supply voltages
- possible connection error
- possible presence of a
ground connection between
USB power supply and device
to be tested
- possible external power
supply not intended for power
generation
- if intended, reduce the value of the external voltage
according to what is stated in
the technical specications
and wait 15-20 minutes for the
internal protections to reach
normal temperature conditions
- make sure that the electrical
connections are consistent
with the type of simulation /
generation to be carried out
(see diagrams on page 36)
- use the USB power supply or
make sure you have connected the calibrator in compliance
with the instructions given on
page 13
57MI004962-EMSC
Page 58
General problems
PROBLEMCAUSESOLUTION
The APP screen does not update the measurements taken
or to be generated
The measures, or generations,
do not match what was expected
TAB. 9
TROUBLESHOOTING
- the area in which you intend
to operate is subject to strong
electromagnetic interference or
interference from other Bluetooth or WiFi devices
- the battery may be at
- electrical connections may be
incorrect
- electrical connections may
not be allowed
- connection made to another
MSC calibrator
- reduce the distance between
the smart phone and MSC calibrator or, if possible, move to a
less disturbed area
- charge the battery
- make sure that the electrical
connections are consistent
with the type of simulation /
generation to be carried out
(see diagrams on page 38)
-Use the USB power supply or
make sure you have connected the calibrator as shown in
Figure 1
- connect to the correct MSC
calibrator
For further information, please go to www.seneca.it/msc.
For technical support, please call: +39.049.8705359
or send an email to: supporto@seneca.it
58MI004962-EMSC
Page 59
FIRMWARE UPDATE
To perform the rmware update of the MSC device, follow the
procedure below:
• Download and install the MSC desktop APP from www.seneca.it/
msc;
• Connect the calibrator to the PC using the USB cable;
• Turn the instrument on by pressing the ON / OFF button and keep
it pressed for about 5/6 seconds until the buzzer starts to emit an
intermittent and continuous sound;
• At this point release the button;
• During the update phase, the calibrator will have the PWR, ERROR and COMM LEDs lit;
• Start the MSC desktop program;
• Select the “ Upgrade” button
• Select the serial communication port;
• Start the "Upgrade" procedure using the “Upgrade” button;
• Wait for the end of the rmware upload by looking at the Log
buffer;
• Once the rmware "Upload" has been completed, restart the
calibrator using the PWR key.
If the procedure has been carried out correctly when the instrument
is started up, three beeps will sound as indicated in table 4 on page
The information in this document can be modied or supplemented without notice for technical or commercial reasons. The gures and diagrams are
indicative and not binding. Neither can inconsistencies or imperfections be excluded, in spite of the continuous search for perfection. The content of this
document undergoes a periodical update anyway. Any unauthorised reproduction is strictly forbidden.
SENECA s.r.l.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.