Secomam UviLine Series, UviLine 9400C, UviLine 8100C, UviLine 9600, UviLine 9100C Operating Manual

...
Page 1
OPERATING MANUAL
MANUEL D’UTILISATION
Ref : N0005B
UviLine Series
Gamme Uviline
SPECTROPHOTOMETERS / SPECTROPHOTOMETRE
Page 2
Page 3
UviLine Series
- 3 -
The most current version of the present operating manual is available on our web site:
http://www.aqualabo-group.com/
La version la plus récente du présent mode d'emploi est disponible sur notre site web :
http://www.aqualabo-group.com/
REMARKS:
The information contained in this document may be the object of patents or patent applications by Aqualabo Analyse. The possession of this document does not confer any license to such patents.
The following names are Aqualabo Analyse’s registered trademarks throughout the world: UviLine
This manual is updated periodically. The updates are included in the new editions. All information supplied in this edition of the manual may be amended before the products described herein are available.
All reproduction rights reserved.
REMARQUES:
Les informations contenues dans ce document peuvent faire l'objet de brevets ou demandes de brevets par Aqualabo Analyse. La possession de ce document ne confère aucune licence sur ces brevets.
Les noms suivants sont des marques déposées Aqualabo Analyse à travers le monde:
UviLine
Ce manuel est mis à jour périodiquement. Les mises à jour sont incluses dans les nouvelles éditions. Toutes les informations fournies dans la présente édition du manuel peuvent être modifiées avant que les produits décrits soient
disponibles. Tous droits de reproduction réservés.
English version: page 4 to page 116 Version francaise : page 117 to page 207
Copyright © Ales 2016, Aqualabo Analyse
Reproduction in whole - or even in part - is prohibited without the express written permission of Aqualabo.
Page 4
UviLine Series
- 4 -
CONTENTS
1 WARRANTY...............................................................................................................9
2 INFORMATION.........................................................................................................11
3 SAFETY....................................................................................................................13
3.1 SAFETY INFORMATION ................................................................................................................13
3.1.1 S
AFETY INFORMATION IN THE OPERATING MANUAL
........................................................................13
3.1.2 S
AFETY SIGNS ON THE PRODUCT
.................................................................................................13
3.2 SAFE OPERATION.........................................................................................................................13
3.2.1 A
UTHORIZED USE
........................................................................................................................13
3.2.2 R
EQUIREMENTS FOR SAFE OPERATION
.........................................................................................13
3.2.3 U
NAUTHORIZED USE
...................................................................................................................13
3.3 HANDLING OF HAZARDOUS SUBSTANCES...............................................................................14
3.4 GENERAL NOTES ON HANDLING................................................................................................14
3.5 PACKING.........................................................................................................................................14
4 OVERVIEW...............................................................................................................15
4.1 FRONT OF THE INSTRUMENT .....................................................................................................15
4.2 SOCKET FIELD ON THE REAR PANEL........................................................................................15
4.2.1 USB-A P
ORT
.............................................................................................................................16
4.2.2 USB-B P
ORT
.............................................................................................................................16
4.3 KEYBOARD.....................................................................................................................................17
4.3.1 O
VERVIEW
.................................................................................................................................17
4.3.2 K
EY FUNCTIONS
..........................................................................................................................17
4.3.3 F
UNCTION KEYS
..........................................................................................................................18
4.3.4 A
LPHANUMERIC KEYBOARD
.........................................................................................................18
4.3.4.1 Characters sets ..................................................................................................................................18
4.3.4.2 Command principle............................................................................................................................. 18
4.3.4.3 Special characters..............................................................................................................................18
4.3.4.4 How to correct a bad character...........................................................................................................18
4.4 DISPLAY..........................................................................................................................................19
4.4.1 D
ISPLAY ELEMENTS
....................................................................................................................19
4.4.2 S
YMBOLS IN THE STATUS LINE
.....................................................................................................19
4.5 CUVETTE COMPARTMENT...........................................................................................................20
5 TECHNICAL DATA ..................................................................................................21
5.1 MEASUREMENT CHARACTERISTICS..........................................................................................21
5.2 MEASURMENT MODE ...................................................................................................................24
5.2.1 C
ONCENTRATION
........................................................................................................................24
5.2.2 A
BSORBANCE / % TRANSMISSION
................................................................................................24
5.2.3 M
ULTI WAVELENGTHS
.................................................................................................................24
5.2.4 S
PECTRUM
.................................................................................................................................24
5.2.5 K
INETICS
....................................................................................................................................24
5.3 MEMORY FOR MEASURED VALUES ...........................................................................................25
5.3.1 M
EMORY CAPACITY
.....................................................................................................................25
5.3.2 O
UTPUT OPTIONS
.......................................................................................................................25
5.3.3 F
ILE FORMATS
............................................................................................................................25
5.4 GENERAL INSTRUMENT DATA ....................................................................................................25
5.5 AVAILABLES LANGUAGES: ..........................................................................................................26
6 COMMISSIONING....................................................................................................27
6.1 BUFFER BATTERIES INSTALLATION ..........................................................................................27
6.2 CONNEXION OF POWER SUPPLY...............................................................................................27
6.3 CHOICE OF THE LANGUAGE .......................................................................................................28
6.4 DATE AND TIME SETTING UP ......................................................................................................28
Page 5
UviLine Series
- 5 -
6.4.1 S
ELFT-TEST
................................................................................................................................29
6.4.2 W
ARM-UP TIME
...........................................................................................................................29
7 OPERATION.............................................................................................................31
7.1 SWITCHING ON OR OFF THE PHOTOMETER ............................................................................31
7.2 GENERAL OPERATING PRINCIPLES...........................................................................................32
7.2.1 N
AVIGATING
...............................................................................................................................32
7.2.2 C
USTOMER METHODS AND PROFILE IMPORTATION/EXPORTATION
..................................................33
7.2.3 P
DF FILE PRINTING
......................................................................................................................33
8 CONCENTRATION MODE.......................................................................................35
8.1 GENERAL POINT............................................................................................................................35
8.2 METHOD PROGRAMMING............................................................................................................35
8.3 METHOD SAVING...........................................................................................................................38
8.4 METHOD DELETING......................................................................................................................39
8.5 METHOD EXECUTION...................................................................................................................39
8.5.1 WITH STANDARD SOLUTIONS..............................................................................................39
8.5.1.1 Measure.............................................................................................................................................. 39
8.5.1.2 Calibration curve displaying................................................................................................................ 41
8.5.1.3 Point deleting......................................................................................................................................42
8.5.1.4 Calibration curve printing....................................................................................................................43
8.5.1.5 Sample measurement ........................................................................................................................43
8.5.1.6 Results printing...................................................................................................................................44
8.5.1.7 Sample dilution...................................................................................................................................44
8.5.1.8 Sample blank value ............................................................................................................................45
8.5.2 W
ITH VALUE PAIRS
......................................................................................................................46
8.5.3 W
ITH A FORMULA
........................................................................................................................47
8.6 METHOD EDITING..........................................................................................................................47
9 ABSORPTION / TRANSMISSION MODE.................................................................49
9.1 GENERAL POINTS.........................................................................................................................49
9.2 MEASUREMENT.............................................................................................................................49
9.3 RESULTS PRINTING......................................................................................................................49
10 SPECIAL/ MULTI-WAVELENGTHS.........................................................................51
10.1 BASIC INFORMATION................................................................................................................51
10.2 METHOD PROGRAMMING ........................................................................................................51
10.3 METHOD DELETING ..................................................................................................................53
10.4 METHOD SELECTION................................................................................................................53
10.5 METHOD EXECUTION ...............................................................................................................54
11 SPECTRUM MODE..................................................................................................56
11.1 GENERAL POINTS .....................................................................................................................56
11.2 METHOD PROGRAMMING ........................................................................................................56
11.3 METHOD SAVING.......................................................................................................................57
11.4 METHOD DELETING ..................................................................................................................57
11.5 METHOD SELECTION................................................................................................................57
11.6 MEASUREMENT EXECUTION...................................................................................................57
11.7 SPECTRA TREATMENT.............................................................................................................58
11.7.1 D
OWNLOADING OF STORED SPECTRUM
........................................................................................58
11.7.2 C
URSOR
.....................................................................................................................................58
11.7.3 Z
OOM
.........................................................................................................................................58
11.7.4 MANUAL CHANGE OF ABSORBANCE SCALE......................................................................60
11.7.5 CALCULATIONS......................................................................................................................60
11.7.5.1 Picks and valleys detection.............................................................................................................60
11.7.5.2 Mark points..................................................................................................................................... 60
11.7.5.3 Deleting of mark points...................................................................................................................61
11.7.5.4 Area calculation.............................................................................................................................. 62
11.7.5.5 Derivative calculation...................................................................................................................... 62
11.7.5.6 Spectra comparison........................................................................................................................ 63
Page 6
UviLine Series
- 6 -
11.7.5.7 Addition of two spectra ...................................................................................................................63
11.7.5.8 Subtraction of one spectrum from another one...............................................................................64
11.7.5.9 Division of a spectrum ....................................................................................................................64
11.7.5.10 Addition of a constant..................................................................................................................... 64
11.7.5.11 Multiplication by a constant.............................................................................................................65
12 KINETICS MODE......................................................................................................67
12.1 GENERAL POINTS .....................................................................................................................67
12.2 METHOD PROGRAMMING ........................................................................................................67
12.3 METHOD SELECTION................................................................................................................68
12.4 METHOD SAVING.......................................................................................................................68
12.5 METHOD DELETING ..................................................................................................................69
12.6 LIMITATION OF METHODS LIST ...............................................................................................69
12.7 METHOD EXECUTION ...............................................................................................................70
12.8 METHOD EDITING......................................................................................................................71
12.9 KINETICS TREATMENT .............................................................................................................71
12.9.1 K
INETIC CURVE PRINTING
............................................................................................................71
12.9.2 C
URSOR
.....................................................................................................................................71
12.9.3 K
INETIC CURVE SAVING
...............................................................................................................71
12.9.4 D
OWNLOADING OF STORED KINETIC CURVE
..................................................................................71
12.9.5 « E
DIT » MENU
...........................................................................................................................72
12.9.5.1 Slope ..............................................................................................................................................72
12.9.5.2 Comparison of kinetic curves..........................................................................................................72
12.9.5.3 Subtraction of a kinetic curve from another one ............................................................................. 73
12.9.6 M
ANUAL CHANGE OF ABSORBANCE SCALE
....................................................................................73
12.10 CURVE PRINTING ......................................................................................................................73
13 MEMORY..................................................................................................................75
13.1 OVERVIEW..................................................................................................................................75
13.2 INSTRUCTIONS ON USING USB MEMORY DEVICES ............................................................76
13.3 MEASURMENT DATASETS .......................................................................................................76
13.3.1 E
LEMENT OF A MEASUREMENT DATASET
......................................................................................76
13.3.2 O
PERATIONS WITH MEASUREMENT DATASETS
..............................................................................76
13.3.3 R
ESULT RECORDING
...................................................................................................................77
13.3.3.1 Manual recording............................................................................................................................77
13.3.3.2 Automatic recording with « AutoStore » function............................................................................ 77
13.3.4 R
ESULT DISPLAYING
...................................................................................................................78
13.3.5 R
ESULT FILTERING
......................................................................................................................80
13.3.5.1 Filtering of a results group.............................................................................................................. 80
13.3.5.2 Filters inversion ..............................................................................................................................80
13.3.5.3 Erase stored measurement dataset................................................................................................80
13.3.5.4 Result deleting................................................................................................................................81
13.3.5.5 Saving Kinetic recordings, spectrUM.............................................................................................. 81
13.3.5.6 Saving data as a pdf file .................................................................................................................81
13.3.6 S
AVING / EXPORTING FILES
..........................................................................................................81
13.3.6.1 Copying all measurement data files to a USB memory device....................................................... 81
13.3.6.2 Copying user-defined methods / profiles to a USB memory device................................................82
13.3.6.3 Copy of individual file on USB memory .......................................................................................... 82
13.3.7 P
RINTING THE DATA
(USB)..........................................................................................................83
13.3.8
SETTING FOR DATA TRANSMISSION
...............................................................................................83
13.3.9 I
MPORTING FILES
........................................................................................................................84
13.3.9.1 Importing spectra or kinetic recordings from a USB memory device .............................................. 84
13.3.9.2 Importing methods / profiles from a USB memory device...............................................................84
14 USERS MANAGEMENT...........................................................................................85
14.1 USER LEVEL AND USER RIGHTS ............................................................................................85
14.2 ACTIVATION/DESACTIVATION OF USER MANAGEMENT .....................................................85
14.2.1 A
CTIVATION OF USER MANAGEMENT FUNCTION
.............................................................................86
14.2.2 D
ESACTIVATION OF USER MANAGEMENT FUNCTION
.......................................................................86
14.3 CREATION, MODIFICATION OR DELETING A USER ACCOUNT ...........................................86
14.3.1 U
SER ACCOUNT CREATION
..........................................................................................................86
Page 7
UviLine Series
- 7 -
14.3.2 U
SER ACCOUNT MODIFICATION
....................................................................................................87
14.3.3 P
ASSWORD MODIFICATION
..........................................................................................................87
14.3.4 U
SER ACCOUNT SUPPRESSION
....................................................................................................87
15 VARIOUS ADJUSTMENT ......................................................................................105
15.1 LANGUAGE CHOICE................................................................................................................105
15.2 DATE AND TIME ADJUSTMENT..............................................................................................105
15.3 CONTRAST ADJUSTMENT......................................................................................................105
15.4 RESET.......................................................................................................................................106
15.5 LAMP SERVICING COUNT DOWN..........................................................................................106
15.6 SPECTROPHOTOMETER INFORMATION ([INFO])................................................................106
15.7 TIMER........................................................................................................................................107
15.8 SOFTWARE UPDATING...........................................................................................................107
16 MAINTENANCE......................................................................................................109
16.1 LAMP CHANGING.....................................................................................................................109
16.2 BUFFER BATTERIES CHANGING ...........................................................................................110
16.3 CLEANING.................................................................................................................................111
16.3.1 C
LEANING YOUR SEPCTROPHOTOMETER
....................................................................................111
16.3.2 C
LEANING THE LENSES
.............................................................................................................111
16.4 WHAT CAN YOU DO, IF…........................................................................................................111
16.4.1 W
ARNING IF CUVETTE IS BROKEN
...............................................................................................111
16.5 TROUBLESHOOTING...............................................................................................................112
17 ACCESSORIES......................................................................................................113
17.1 10MM CELL HOLDER (REF. 70VI0604)...................................................................................113
17.2 PORTE CUVE JUSQU’A 100MM (REF. 70VI0605)..................................................................113
17.3 PORTE CUVE 10MM THERMOSTATABLE (REF. 70VI0602) .................................................113
17.4 PORTE TUBE 16MM (REF. 70VI0607).....................................................................................113
17.5 AUTOMATIC CELL CHANGER TURRET (REF. 70VI0600).....................................................113
17.6 SIPPER (REF. 70VI0601)..........................................................................................................114
17.7 EFFET PELTIER (REF. 70VI0603) ...........................................................................................114
17.8 SIPPER WITH PELTIER EFFECT (REF. 70VI0610) ................................................................114
17.9 DIAPHRAGMS FOR MICRO-CELLS (REF. 70VI0608) ............................................................115
Page 8
UviLine Series
- 8 -
Page 9
UviLine Series
- 9 -
1 WARRANTY
UviLine Series instruments sold by Aqualabo Analyse is guaranteed against any manufacturing defects for three years (unless otherwise stated by AQUALABO ANALYSE) with effect:
From the technical acceptance of the equipment in the factory by the buyer or his designee,
or failing this :
* For Metropolitan France: from the date on the delivery note. * For other destinations: from the date of factory shipment certified by air waybill, consignment note or bill of lading.
Our company guarantee applies exclusively to defectiveness arising from a design fault or from a concealed defect. It is strictly limited to the free dispatching of replacement parts (except for consumable items) or to the repairing of the equipment in our workshops within a deadline of 10 working days (shipping delay not included).
By express agreement, the following are strictly excluded from our guarantee:
All damages, notably for staff costs, loss of earnings, business trouble, etc
Any breakdown due to an incorrect use of the equipment (non adapted mains, fall, attempt at transformation, etc) or to a lack of
maintenance by the user or to poor storage conditions.
Any breakdown due to the use of parts not supplied by AQUALABO ANALYSE, on AQUALABO ANALYSE equipment
Any breakdown due to the transporting of the equipment in packaging which is not its original packaging
The lamps, the cells and generally any item which appears in the "accessories" section on the price list.
Our customers are kindly asked to apply for our consent before returning any instrument for repair. No return of materials may be accepted without the prior written consent of our Servicing after Sales Management which will precise the terms of such return. If the above consent is given, articles shall be returned in their original packaging on a prepaid basis to the following address:
AQUALABO ANALYSE – USINE SECOMAM - 91 avenue des Pins d’Alep 30100 ALES - France
We reserve the right to reship all instruments received collect failing such consent. Whatever method and conditions of transport are chosen for the shipment of the equipment to be repaired under guarantee, in the
original packaging, the corresponding costs and the insurance costs will be payable by the customer. Any damage connected to the return transport of the equipment falls within the framework of the guarantee on the express condition
that the customer has sent his complaint within forty-eight hours by registered letter with acknowledgement of receipt to the carrier. A copy of the letter should be sent to AQUALABO ANALYSE.
For equipment with a guaranty card, this is only applicable if the card delivered with the equipment is returned to AQUALABO ANALYSE duly completed.
SOFTWARE GUARANTEE
The software is guaranteed by the designer or the distributor of the software under the conditions specified in the literature accompanying the aforementioned software packages. Under no circumstances whatsoever will AQUALABO ANALYSE supply any type of guarantee for software packages. By express agreement, all damages, notably for staff costs, lost of earnings; business trouble, etc are strictly excluded from our guarantee. The customer is informed that the software cannot be guaranteed exempt from defects or bugs.
TRADE SECRET AND PROPERTY RIGHTS
This document is protected by AQUALABO ANALYSE copyright (c) 2003 and the copying rights are explicitly reserved. The software supplied with the equipment or referenced contains trade secrets and confidential information which are AQUALABO ANALYSE's property. It is legally protected by the international copyright (c) laws.
AQUALABO ANALYSE grants a license to use its software to the user. This may not be disclosed, used or duplicated with the intention to save it, without AQUALABO ANALYSE's written permission. The beneficiary must attach a copy of this document to all authorized partial or total reproductions.
Page 10
UviLine Series
- 10 -
Page 11
UviLine Series
- 11 -
2 INFORMATION
The AQUALABO ANALYSE equipment has been designed, manufactured, tested and inspected according to the ISO 9001 standards (Version 2000).
If the unit is not immediately installed, it should be stored in a dry and clean area. The storage temperature should be between 10 and 35°C.
AQUALABO ANALYSE equipment is carefully inspected before it is packed. As soon as you receive your equipment, check the condition of the packaging and if you notice any problems, notify your carrier within 48 hours. Then consult the packing list and check that everything is in order. Finally, if you discover that something is missing, or if the goods are damaged immediately notify AQUALABO ANALYSE.
The main differences between UviLine 9300 and UviLine 9600 spectrophotometers are the wavelengths ranges and the light sources.
UviLine 8100C / 9100C / 9300
Wavelengths range: 320 – 1100 nm Light source: Halogen lamp 5VDC 10W
UviLine 9400C / 9600
Wavelengths range: 190 – 1100 nm Light source: Xenon lamp.
UviLine Series instruments are fully designed and manufactured by AQUALABO ANALYSE in Ales France.
AQUALABO ANALYSE / SECOMAM 91 Avenue Des Pins D’Alep • 30100 Alès • France
Tel : +33 (0)4 66 54.26.95 Fax : +33 (0)9 72 35 16 80 E-mail : info@aqualabo.fr Web: www.aqualabo-group.com
Page 12
UviLine Series
- 12 -
Page 13
UviLine Series
- 13 -
3 SAFETY
1.1 SAFETY INFORMATION
1.1.1 S
AFETY INFORMATION IN THE OPERATING MANUAL
This operating manual provides important information on the safe operation of the product. Read this operating manual thoroughly and make yourself familiar with the product before putting it into operation or working with it. The operating manual must be kept in the vicinity of the product so you can always find the information you need.
Important safety instructions are highlighted in this operating manual. They are indicated by the warning symbol (triangle)
in the left column. The signal word (e.g. "CAUTION") indicates the level of danger:
WARNING Indicates a possibly dangerous situation that can lead to serious (irreversible) injury or
death if the safety instruction is not followed.
1.1.2 S
AFETY SIGNS ON THE PRODUCT
Note all labels, information signs and safety symbols on the product. A warning symbol (triangle) without text refers to safety information in this operating manual.
1.2 SAFE OPERATION
1.2.1 A
UTHORIZED USE
The authorized use of the photometer consists exclusively of the carrying out of photometric measurements according to this operating manual. Follow the technical specifications of the cells. Any other use is considered unauthorized.
1.2.2 R
EQUIREMENTS FOR SAFE OPERATION
Note the following points for safe operation:
- The product may only be operated according to the authorized use specified above.
- The product may only be supplied with power by the energy sources mentioned in this operating manual.
- The product may only be operated under the environmental conditions mentioned in this operating manual.
- The product may not be opened.
1.2.3 U
NAUTHORIZED USE
The product must not be put into operation if:
- it is visibly damaged (e.g. after being transported)
- it was stored under adverse conditions for a lengthy period of time
.
User qualification
Carrying out photometric determinations with the aid of test sets frequently requires the handling of hazardous substances. We assume that the operating personnel know how to handle hazardous substances due to their professional training and experience. The operating personnel must particularly be able to understand and correctly implement the safety labels and safety instructions on the packages and inserts of the test sets.
Page 14
UviLine Series
- 14 -
1.3 HANDLING OF HAZARDOUS SUBSTANCES
When developing test sets, Orchidis carefully sees that the tests can be carried out as safely as possible. Some hazards by dangerous substances, however, cannot always be avoided. If self-produced tests or solutions are used, the responsibility concerning any risks caused by those tests or soulutions lies with the user (personal responsibility).
WARNING Improper handling of certain reagents can cause damage to your health. In any case follow the safety labels on the packing and the safety instructions of the package insert. Protective measures specified there have to be followed exactly.
Safety datasheets
The safety datasheets of the chemicals comprise all instructions on safe handling occurring hazards, preventive actions and actions to take in hazardous situations. Follow these instructions in order to work safely.
1.4 GENERAL NOTES ON HANDLING
The UviLine instruments are an optical precision meters. Therefore, it should always be handled with care, especially in mobile use.
Always protect the meter from conditions that could damage the mechanical, optical and electronic components.
Heed the following points especially:
The temperature and humidity during operation and storage must be within the limits specified in chapter
TECHNICAL DATA.
The following influences always have to be avoided with the meter:
– Extreme dust, moisture and wetness – Exposure to intensive light and heat – Fumes that are corrosive or contain high concentrations of solvents.
For measuring, the meter must be placed on a flat surface. Spilled liquid or other material should be removed immediately If a cell has broken in the cell shaft, the cell shaft should be cleaned immediately The cell shaft should always be closed when the photometer is not used. For the instrument to be transported the cell shaft has to be empty. For mobile used we recommend using the suitable transport case Always make sure that the instrument is connected on the good voltage. (Between 100 – 240V 50-60Hz) When dangerous substances for health and environment are used, the laboratory or site rules, where the
instrument is installed must be followed.
Take all the necessary precautions, during the use the instrument, to protect the operator from eventual liquids
leaks or spills or possible radiations (protective gloves, anti-UV radiation glasses , protected clothes, etc)
Keep the sample compartment clean. The Xenon lamp used in the UviLine 9400C & 9600 emits UV radiation. Install the instrument in a ventilated area because it is likely to generate ozone, which, beyond the limits below,
can harm health.
Exposure average value = 100 ppb Exposure limits value = 200 ppb
All operations made inside the instrument, must be done by AQUALABO ANALYSE or by AQUALABO
ANALYSE’s authorized technicians.
1.5 PACKING
The spectrophotometer is sent out in a protective transport packing. Keep the original packing including the inner packing to protect the instrument against hard shocks if it has to be transported.The original packing is also required for the proper return of the instrument if it has to be repaired. Note that damage caused by improper transport voids all warranty claims.
Page 15
UviLine Series
- 15 -
1. LCD
graphic screen.
2. Membrane keyboard.
3. Cuvette compartment.
4. Cover of the cuvette well.
5. Connection for power supply.
6. Ethernet RJ45.
7. USB-A connection.
8. USB-B connection.
4 OVERVIEW
1.6 FRONT OF THE INSTRUMENT
1.7 SOCKET FIELD ON THE REAR PANEL
You can connect the following accessories to the photometer:
- PC (USB-B)
- Printer (USB-A The data is printed out with <PRINT>)
- USB memory device (USB-A)
- USB-PC keyboard (USB-A)
Suitable are all printers that can interpret the PCL-3 printer control language.
If you want to connect several USB devices such as a USB-PC keyboard and a USB memory device to the meter, you can increase the number of USB-A sockets by a commercially available USB-2 hub with separate power supply.
3
4
1
2
5
1
6
7
8
Page 16
UviLine Series
- 16 -
1.7.1 USB-A P
ORT
The USB-A port allows:
To print data when a printer is connected. With <PRINT> key data are sent on the printer.
To records data on USB memory (USB key) when it is connected.
To update spectrophotometer software and methods data.
To connect an external USB keyboard.
1.7.2 USB-B P
ORT
Allows the spectrophotometer and PC to be connected directly by emulating a RS232 port. A USB  External RS232 adapter is not required to connect the spectrophotometer to a free PC RS232C port. Please contact Orchidid Aqualabo Analyse for the driver package.
Page 17
UviLine Series
- 17 -
1. Function keys F1 to F4 (functions menu depending)
2. Fixed key functions.
3.
Alphanumeric keys block
.
1.8 KEYBOARD
1.8.1 O
VERVIEW
1.8.2 K
EY FUNCTIONS
Key Designation Function
<ON/OFF>
Switch the spectrophotometer ON and OFF
<HOME>
Switches to the main menu from any operating situation. Actions that are not completed are cancelled.
<PRINT>
Downloads the displayed value to an interface.
<STORE>
Saves a displayed value or spectrum or kinetic curve.
<ZERO BLANK>
Starts one of the following measurements, depending on the operating situation:
- Zero adjustment
- Blank value measurement
- Baseline measurement
<TIMER>
Open “Timer” menu.
<ESC>
Cancels the running action.
- Entries that have not yet been accepted are discarded.
- Switches to the next higher menu level.
<START ENTER - Starts an action (e.g. measurement)
- Opens a selected menu
- Confirms a selection or entry
« » or « ». Moves the selection in menus and lists one position up or down.
« »
- Deletes the character left of the cursor during character entries
- Moves the cursor to the left in a spectrum or kinetic diagram.
3
1
2
Page 18
UviLine Series
- 18 -
« »
– Moves the cursor to the right in a spectrum or kinetic diagram.
1.8.3 F
UNCTION KEYS
The function keys F1 to F4 have different functions depending on the operating situation. The current functions are displayed in the function key menu at the bottom edge of the display.
1.8.4 A
LPHANUMERIC KEYBOARD
Numerals, letters, punctuation marks and special characters are entered with the alphanumeric keypad of the meter or using an external keyboard.
1.8.4.1 Characters sets
The following characters are available:
Numerals (0 ... 9)
Letters (A ... Z) and (a ... z).
Punctuation marks (. and -)
Special characters ° / + , Σ, µ, #, %, : et ()
1.8.4.2 Command principle
Entering characters is always possible if there is an input field on the display. The numerals and characters (expect for the small letters) assigned to the keys of the alphanumeric keypad are printed on
the keys. Example: With the <2/ABC> key you can enter the following characters: 2, A, B, C, a, b, c.
Select the required character by pressing the key several times (similar to a mobile phone). When pressing a key that is assigned to several characters once, the respective numeral appears first. To enter a numeral, one key pressing is always sufficient. When pressing the key for the first time a line pops up that displays all characters possible with this key. The currently selected character is highlighted.
A character is taken over in the input field if
the character is highlighted for more than one second,
the character is confirmed with <START·ENTER>,
Another alphanumeric key is pressed.
Remark
During mere number entries (such as entering a wavelength), the keys of the alphanumeric keypad are assigned to the respective numeral only. Each pressing key directly enters the numeral (like a pocket calculator).
1.8.4.3 Special characters
To enter special characters, use <1/*> key.
1.8.4.4 How to correct a bad character
Use <> key, to delete all characters until THE INCORRECT one and start again the entries from there.
Page 19
UviLine Series
- 19 -
1.9 DISPLAY
1.9.1 D
ISPLAY ELEMENTS
1. Status line (current state, date and time)
2. Display range for menus and measurement results
3. Function keys menu
1.9.2 S
YMBOLS IN THE STATUS LINE
Symbol Designation Function
Save
The <STORE> key is active. You can store the displayed data with
<STORE>
Printer
The <PRINT> key is active. You can output to an interface the displayed data with <PRINT>
1
2
3
Page 20
UviLine Series
- 20 -
1.10 CUVETTE COMPARTMENT
The cuvette holder is quickly install thank to a special system named “Quick lock”
- Set in position the cuvette holder (Quick lock system above the hole inside the cuvette well).
- Push, with finger, the “quick lock” system.
- Release the “quick lock” system.
- The cuvette holder is installed. And ready to receive a cuvette
- The beam is going from the left to the right.
« Quick lock »
Cuvette holder
Cuvette compartment
Page 21
UviLine Series
- 21 -
5 TECHNICAL DATA
1.11 MEASUREMENT CHARACTERISTICS
Uviline 8100C
Page 22
UviLine Series
- 22 -
Page 23
UviLine Series
- 23 -
The specifications above are valid with the cuvette cover closed. To ensure accuracy and performances make measurements in low and stable light envirement (as dark as possible).
Page 24
UviLine Series
- 24 -
1.12 MEASURMENT MODE
1.12.1 C
ONCENTRATION
– Measurement with permanently programmed methods, adapted to the aqualabo test set program – Program support for the creation of additional user-defined methods (max. 100) – Citation forms and units method dependent – Display of the absorbance value can be added
1.12.2 A
BSORBANCE / % TRANSMISSION
– Measurement against own reference absorbance value possible
1.12.3 M
ULTI WAVELENGTHS
– Freely definable calculations from absorbance values at up to 10 different wavelengths – Calculations can be stored as methods (max. 499)
1.12.4 S
PECTRUM
– Absorbance or % transmission mode – Limits freely selectable within the wavelength range – Increment: 1 nm – Settings can be stored as profiles (max. 20) – Evaluation functions: Cursor scanning, zoom, min./max.recognition, peak area determination, derivation, smoothing, multiplication by constants, addition of constants, addition and subtraction of spectra, formation of the quotient of two spectra
1.12.5 K
INETICS
– Absorbance or % transmission mode – Minimal adjustable scan interval: 1 s (if the absorbance of the test sample is high, the scan interval is extended due to the longer duration of the individual measurements) – Settings can be stored as profiles (max. 20) – Evaluation functions: Cursor scanning, zoom, min/max determination, slope calculation (for an interval or total), enzymatic activity
Page 25
UviLine Series
- 25 -
1.13 MEMORY FOR MEASURED VALUES
1.13.1 M
EMORY CAPACITY
5000 single measured values from the measuring modes, concentration, absorbance/ % transmission and multi wavelengths
40 MByte internal memory, sufficient for approx. 500 spectra and 400 kinetic curves (sample values based on the following assumptions: All spectra over a wavelength range of 600 nm and all kinetic curves with 150 single values each)
1.13.2 O
UTPUT OPTIONS
USB memory device, printer, PC
1.13.3 F
ILE FORMATS
ASCII, *.csv
1.14 GENERAL INSTRUMENT DATA
Dimensions 404 x 197 x 314 mm (width x height x depth)
Weight approx. 4.5 kg (without plug-in power supply)
Housing type of protection IP 30
Electrical protective class III
Test certificates CE
Temperature allowed
Operation: +10 °C to + 35 °C
(41 °F to 95 °F)
Storage: -25 °C to + 65 °C
(-13 °F to 268 °F)
Humidity allowed 80 %
Climatic class 2 / VDI/VDE 3540
Power pack
Type: EDACPOWER EA1036R
Input: 100 - 240 V ~ / 50 - 60 Hz /
1 A
Output: 12 V = / 3 A
EMC
EC directive 2004/108/EC
EN 61326-1
– Interference emission: Class B
– Interference immunity:
IEC 61000-4-3
– Tolerance extension: 0.008 E
FCC Class A
Meter safety
EC directive 2006/95/EC
EN 61010-1
Page 26
UviLine Series
- 26 -
1.15 AVAILABLES LANGUAGES:
German
English
Français
Español
Italiano
Bulgarian/Български
Česko
Simplified Chinese/ 中文
Traditional Chinese/ 繁體中文
Greek/Ελληνικά
Indonesian/Indonesia
Japanese/ 日本語
Magyar
Malay/Melayu
Norsk
Polski
Portuguése
Russian/Русский
Slovenščina
Thai/ ภาษาไทย
Turkish/Turkce
Dansk
Română
Nederlands
Page 27
UviLine Series
- 27 -
6 COMMISSIONING
For the fisrt starting up, oerform the following activities:
Insert the buffer batteries Connect the power supply Switch on the photometer Set the language Set the date and time
1.16 BUFFER BATTERIES INSTALLATION
The buffer batteries supply the integrated clock while the photometer is switched off. Four alkaline manganese batteries (type AA or Mignon) separately included in the scope of delivery are used as the buffer batteries. Insert the batteries as follows:
1. Turn the photometer upside down and place it on a soft surface.
2. Open the lid of the battery compartment (1).
3. Insert the four batteries in the battery compartment. Make sure that the poles of the batteries are in the correct position.
4. The ± signs on the batteries must correspond to the ± signs in the battery compartment.
5. Close the lid of the battery compartment.
Batteries time life: The power consumption of the clock is very low. The lifetime of high quality batteries is at least 5 years.
1.17 CONNEXION OF POWER SUPPLY
The power is supplied via the enclosed plug-in power pack. The power pack supplies the photometer with low voltage (12 VDC).
CAUTION The line voltage of the usage location must fulfil the specifications stated on the power pack. Always use the supplied 12 V original power pack only. Before plugging in the power cable check whether it is undamaged.If the power cable is damaged, the instrument must not be operated.
1) Connect the miniplug of the power pack to the socket (1) of the spectrophotometer.
2) Connect the power pack to an easily accessible power socket. The display illumination switches on and then off again.
1
Page 28
UviLine Series
- 28 -
1.18 CHOICE OF THE LANGUAGE
During the initial commissioning, the spectrophotometer automatically guides the user through the setting of the meter language, date and time after switching on (see following sections).
- Press key. <ON/OFF> the spectrophotometer is switched on.
1. With « » and « » keys, select the language.
2. Confirm the choice with <START·ENTER> key. The language has been set.The currently selected language is marked by a check.
NOTE: After the initial commissioning, the operator can change the language at any time. From the main menu <HOME> enter in the menu “General setup”, then “Language” menu.
1.19 DATE AND TIME SETTING UP
During the initial commissioning, the instrument automatically guides the operator to the setting of the time and date after the setting of the language.
The menu « Date / Time » is opened.
1. With « » and « » keys, select one option from the menu. Valid this option with <START·ENTER> key.
2. The input field for the current date pops up.
3. Enter the current date with <0...9> keys and confirm.
4. The input field closes. The date is accepted.
5. Select « Time » and confirm.
6. The input field for the current hour pops up.
Page 29
UviLine Series
- 29 -
7. Enter the current hour with <0...9> keys and confirm. The input field closes. The time is accepted.
NOTE: After the initial commissioning, the operator can change the date and time at any time. From the main menu <HOME> enter in the menu “General setup”, then “Date/time” menu.
1.19.1 S
ELFT-TEST
IMPORTANT
During self-test, the cuvette compartment must be empty (without anymore cuvette inside the cuvette well) and the cover of the cuvette compartment must be closed.
- Press <START ENTER> key to run the self-test.
During self-test, different parts of the spectrophotometer are checked. If the test is OK, the symbol is displayed on the corresponding line.
This self-test includes: –memory, processor, interfaces, filter and lamp tests. – checking of the wavelength calibration.
It is possible to see and print the self-test results by pressing F4 key [Info]
1.19.2 W
ARM-UP TIME
After switching on the photometer requires a warm-up time of 15 minutes. Reproducibility of measurement data is restricted during the warm-up time. Therefore, do not measure during the warm-up time. During the warm-up time, a progress bar appears on the display next to the date. The progress bar disappears as soon as the warm-up time is over.
Page 30
UviLine Series
- 30 -
Progress bar during warm-up time
Page 31
UviLine Series
- 31 -
7 OPERATION
1.20 SWITCHING ON OR OFF THE PHOTOMETER
Switch the photometer on with <ON/OFF>.
- The Self test dialog (if the user management is not active) or the Login dialog (if the user management is active
- Enter user name and password or register as a guest then press <START·ENTER>
Switching off: To switch off the instrument, press the <ON/OFF> key.
Page 32
UviLine Series
- 32 -
1.21 GENERAL OPERATING PRINCIPLES
1.21.1 N
AVIGATING
After the end of the self-test, the screen displays the main menu. 5 modes are available:
- Concentration
- Absorbance / % Transmission
- Special / Multi wavelengths
- Spectrum
- Kinetics
Press <F2> key from the main menu to access in general setup options
Press « » and « » keys, to slect the wanted
Page 33
UviLine Series
- 33 -
1.21.2 C
USTOMER METHODS AND PROFILE IMPORTATION/EXPORTATION
This function gives you possibility to export and import methods from an instrument to another one. Therefore, you can program all your methods in one instrument and duplicate them on all other same instrument easily with an USB stick.
Configuration menu:
You have the choice to either export all data of your instrument or to select on a list which emthods or profile you want to
1.21.3 P
DF FILE PRINTING
This new function gives you possibility to print on a virtual printer by generating a printed report on a PDF file. This is useful when a physical printer is not available or you want to store numerically printed result directly.
General setup menu to configure printer parameters
With parameter Function of <PRINT> key, you can select if you want to use a real USB printer or the virtual printer generating PDF file printed report.
Page 34
UviLine Series
- 34 -
Page 35
UviLine Series
- 35 -
8 CONCENTRATION MODE
1.22 GENERAL POINT
The concentration displays results according to existing standard methods, or specific method defined by the user.
1.23 METHOD PROGRAMMING
In the “concentration” mode use F1 [New methods] for a new set up.
* Number
Automatic classification of methods (From 1001 to 1100).
* Wavelength
320 – 1100 nm for UviLine 8100C / 9100C / 9300 190 – 1100 nm for UviLine 9400C / 9600
* Resolution
From 0 to 3
Page 36
UviLine Series
- 36 -
It is the number of figures after comma for the result.
* Unit
i.e.: mg/L (10 characters maximum).
* Citation form
(18 characters maximum). Enter the chemical formula of the compound to be measured. i.e.: NO3
* Designation
(18 characters maximum). Enter the compound name. i.e.: Nitrate
* Version
(18 characters maximum). Enter the parameter concentration range. i.e.: 0.9 – 75.3 mg/L
* Calibration curve
There are several possibilities:
- Measure the standard solutions.
- Enter the equation of the calibration curve or the factor value.
- Simulate measurements while entering of the couples of Absorbance/Concentration values * Cell changer mode
Usefull if the instrument is equiped with a cell changer
* Standard ID
Enter the identification number of the standards if this one exists.
* Standard manufacturer
Enter the standard manufacturer name if this one exists
* Blank value
YES or NO.: Measure done or not of the reagent blank
* Standard conc 1
Enter the concentration of standard 1.
Page 37
UviLine Series
- 37 -
MEASURE OF STANDARD SOLUTIONS
Remark
It is possible to add other standards (Maximum 10) by pressing F2 key [Add] and to enter their concentration. To delete a standard, select it with « » and « » keys and press F3 key [Delete]. When all standards are entered, pres F4 key [Next] to continue.
ENTER VALUE PAIRS
ENTER FORMULA
Page 38
UviLine Series
- 38 -
Enter a function as: C = a5.A5 + a4.A4 + a3.A3 + a2.A2 + a1.A + a0 With C = Concentration
A = Absorbance a0 to a5 free factors (from 0.000 to 9999.000)
Example: Use of a K factor K (C = a1A)
- a0 = 0
- a1 = K (factor)
- a2, a3, a4, a5 = Zero
Example: straight line regression (C = a1A + a0)
- a0 = bias
- a1 = slope
- a2, a3, a4, a5 = Zero
Example: third degree curve (Nitrate test in sea water) : C = 0.1493A3 – 2.5154A2 + 36.524A + 1.865
- a0 = 1.865
- a1 = 36.524
- a2 = - 2.5154
- a3 = 0.1493
- a4 & a5 = 0
TIMER If your methods needs timers, you can edit vaues by pressing <F4>, [NEXT]
1.24 METHOD SAVING
The methods are automatically saved under the classification name from 1001 to 1100.
Remark: However, methods using standards will be saved only after standards measurement.
* Lower and upper limit of measuring range
Choice between zero and the maximum value of measured parameter.
Page 39
UviLine Series
- 39 -
1.25 METHOD DELETING
- Select method to be deleted in the method List with « » and « » keys and press <START/ENTER> key. Then press F1 key [Setup].
- Select « Delete method » with « » and « » keys and press <START/ENTER> key.
1.26 METHOD EXECUTION
1.26.1 WITH STANDARD SOLUTIONS
1.26.1.1 Measure
When standard (s) are entered, press F4 key [Next]. Then - press F3 key [Measurement] to start the measure.
- Press on <ZERO/BLANC> key to perform the zero of the meter.
- Install the blank (if necessary) inside the cuvette holder.
- Start the blank measurement (if this one was programmed) by pressing <START/ENTER> key.
- It is possible, in the event of error, to refuse the value and to measure again the solution. Press then on F2 key [Discard].
- Then remake measurement while pressing on <START/ENTER> key.
Page 40
UviLine Series
- 40 -
- It is also possible, if the solution is unstable, to measure it again several times. The software will make the average of the various measured values. Remark: The number of times to measure again the solution is not limited.
- Press, then F4 key [Apply].
- The operator can choose to preserve the last measured value by validating the option "Last measured absorbance" or to use the average value calculated by choosing and validating the option "Median".
- Install the first standard to be measure. Then press F3 key [Measurement].
- Start measurement of the first standard by pressing <START ENTER> key.
- As for the blank, the operator can reject measurement (F2 key [Discard]) or make several times the measurement of the same standard (F1 key [Next meas.]) and preserve the average value.
- Make, in the same way, the measurement of the other standards.
-Press F4 key [Next.] to proceed
NOTE:
If the programming of standards concentration is not increasing or decreasing in a monotonous way or if the measurement of the blank and standards absorbances is not, also increasing or decreasing in a monotonous way (see example below), the spectrophotometer will display an error prompt (see below)
Page 41
UviLine Series
- 41 -
1.26.1.2 Calibration curve displaying
F1 key [Back]
- It allows to the user, in the event of error, to return back to the analysis programming and to modify the acquiring way of the values to make the curve (standards measurement, use of the curve equation or entry of the Concentration/absorbance couples). That also enables him to change the standards concentration.
F2 key [Curve type] It allows changing the curve calculation mode.
The equation of the calculated curve is displayed on the top with the coefficient of R² determination
F3 key [Meas.range]
It allows to the operator to reduce the parameter measuring range.
Lower limit entered by the operator Upper limit entered by the operator
These limits are materialized by an arrow « » on the concentration axis.
Straight line
The straight line calibration curve is consisted of line segments connecting the calibration points between them.
Linear regression
Curve whose equation is form ax + B.
Quadratic regression
Curve which equation is: ax² + bx + c.
Page 42
UviLine Series
- 42 -
Any value under or above limits fixed by the operator will be declare as > or < to the limits, but absorbances will be displayed (see example above).
1.26.1.3 Point deleting
To delete a point press F1 key [Back], then Press F4 key [Next].
-- Select the point to delete with « » and « » keys and delete the point by pressing F3 key [Delete]. Then press on F4 key [Next].
During this stage, it is always possible to measure a standard once again.
- Select the standard to measure again.
- Press F3 key [Measurement] to start the standard measurement.
- Press F4 key [Next] to display the curve.
- To set back the removed standard, press F4 key [Back].
- Add the standard.
- Enter its concentration.
- Measure the concentration of this standard once again, or enter its absorbance following the curve building mode selected.
Page 43
UviLine Series
- 43 -
1.26.1.4 Calibration curve printing
When the curve is displayed, press the <PRINT> key.
UviLine 9600 7449001 1.14 Mathieu
1000 Nitrate 0.9 / 75.3 mg/L 500 nm
f(x) = 0.03207x + 0.0427 R² = 0.9993 Meas. range: 0.00 _ 30.00 mg/L
1.26.1.5 Sample measurement
- Press <ZERO BLANK> key to start the blank measurement. It performs the zero
- Press < START ENTER> key to start sample measurement.
If you want that the samples absorbance is displayed at the same time as the concentration, he has to select it in the [Setup] menu and unselect it. The symbol means that the absorbance will be displayed at the same time as their
concentration.
Page 44
UviLine Series
- 44 -
1.26.1.6 Results printing
- Press <PRINT> key to send data on printer.
UviLine 9600 7449001 1.16 Mathieu
Time of measurement: 3/20/2015 16:28:23 Method: Nitrate Measured value: >30.00 mg/l Sample absorbance: 1.119
Time of measurement: 3/20/2015 16:28:23 Method: Nitrate Measured value: 17.50 mg/l Sample absorbance: 0.604
1.26.1.7 Sample dilution
It is possible to program a dilution factor (From 0 to 999) that will be taken in account during sample measurement. Before sample measurement. Press F1 key [Setup]. Then press on <START ENTER> key.
When the dilution facto is entered, confirm with <START ENTER> key. The dilution factor (i.e.: [1 + 1]) is displayed on the top left part of the screen.
Enter the dilution factor.
1 + 0 = No dilution. 1 + 1 = Dilution by 2.
Page 45
UviLine Series
- 45 -
NOTE: This dilution factor will be maintained with the value entered by the operator for all measurements to come. At the
time of the exit of the mode of analysis or analysis in progress it will be given to 0 (no dilution).
1.26.1.8 Sample blank value
It is possible to subtract a sample blank for each measured sample. Press F1 key [Setup] after zero measurement. It displays:
Remark: The symbol at the end of « Sample blank value » means that the option is selected. The symbol [SB] appears at top left corner of the screen and stays here and will remain all the time that this option is
selected or that the operator did not leave the analysis in progress or the mode of concentration measurement. Insert the first sample blank and run the measurement by pressing <START ENTER> key.
As for the blank, the operator can reject measurement (F2 key [Discard]) or make several times the measurement of the same sample blank (F1 Key [Next meas.]) and preserve the average value.
- Press F4 key [Apply] to run the sample measurement:
- Insert the sample corresponding to the measured sample blank and start the measurement by pressing <START ENTER> key.
Page 46
UviLine Series
- 46 -
It displays the sample absorbance read against the zero (i.e.: 0,562) and the sample concentration calculated as following:
Sample concentration = sample absorbance read against the zero minus sample blank absorbance read against the zero multiply by the factor or reported to the calibration curve.
1.26.2 W
ITH VALUE PAIRS
Enter, with keyboard, the absorbance values of the blank (if necessary) and the one of each standard and press F4 key [Next]. It displays then the calibration curve. From this point, the continuation proceeds is the same as previous chapter “standards.Solutions”
Page 47
UviLine Series
- 47 -
1.26.3 W
ITH A FORMULA
When the formula is entered, press F4 key [Next], then proceed as previous chapter “standards.Solutions”
1.27 METHOD EDITING
- Enter in « Concentration » mode.
- Select a method with « » and « » keys, then confirm with <START ENTER> key.
- Press F1 key [Setup].
- Select « Edit method » with « » and « » keys, then confirm with <START ENTER> key.
Page 48
UviLine Series
- 48 -
Page 49
UviLine Series
- 49 -
9 ABSORPTION / TRANSMISSION MODE
1.28 GENERAL POINTS
The absorbance or transmittance respectively is measured without the use of any methods. All settings are configured during measurement. The absorbance or transmittance can alternatively be measured against the air or against a reference solution determined by the operator. The settings of the last measurement are active.
1.29 MEASUREMENT
- Press F2 key [Wavelength] to enter a new wavelength.
- Press <ZERO BLANK> key to start the zero measurement.
- Press <START ENTER> key to start the sample measurement (Absorbance and transmittance are displayed).
- Press <START ENTER> key to start a new measurement.
When there were several measurements, the four last are displayed at the same time as measurement in progress
- Press F4 key [Continuous] to measure the same sample several time again.
- Press F4 key [Stop] to stop the continuous measurement.
1.30 RESULTS PRINTING
- Press <PRINT> key to print the results. UviLine 9300 7449001 1.14
Administrator Time of measurement: 07/01/2015 14:05:46
Wavelength: 525 nm Measured value: 0.278 52.8 %
Page 50
UviLine Series
- 50 -
Page 51
UviLine Series
- 51 -
10 SPECIAL/ MULTI-WAVELENGTHS
1.31 BASIC INFORMATION
In the Special / Multi wavelengths mode of the UviLine Series, you can carry out measurements with special methods and functions. You can use the following functions for these methods:
Measurements at different wavelengths
Use of procedure variables.Procedure variables provide a value that has to be entered prior to each
measurement on the photometer (e.g. volume, pH value or temperature)
Check whether a value meets a condition. With a condition you can check a value for validity (e.g. absorbance value, procedure variable or the result of a formula).
Formula editor for the convenient programming of any user-defined methods
1.32 METHOD PROGRAMMING
Remark
For multi wavelength methods, you can use the method numbers 2001 to 2499. All special methods can also be selected in the method list of the concentration mode.
The creation of a user-defined method is done in the following steps:
Enter the general method data Method number, method name, version, unit etc.
Enter the wavelengths for absorbance measurements (Ax nm) Minimum 1, maximum 10
Define the procedure variables (KX ) (optional) Procedure variables are used to take into account any influence quantities that cannot be measured by the
photometer. The values for these procedure variables have to be entered for all measurements with the method, e.g. the temperature or pH value.
Enter the formula to calculate the measurement result Enter the formula with which you want to calculate the measurement result in the formula editor.
Enter an additional condition (optional) Conditions are used to check the measurement result for validity.
The condition is entered with the formula editor.
To creat a new method:
- Press F1 key [New method].
- Enter here general data of the method. The next number of method available is already registered as next number of analysis to be used. To fill the fields with entry, it is possible to proceed as follows:
- Fill the entire empty fields with entry ones after the others.
- Press on the key F2 [Method list] to select a method already recorded
- Press on the F3 key [Delete] to remove the method completely.
- Press on the F4 key [Next] to validate all the entries and to pass in the following page.
Page 52
UviLine Series
- 52 -
- Press on F2 key [Add] to add another wavelength.
- Press F3 key [Effacer] to delete the last wavelength.
- Press on the F4 key [Next] to validate all the entries and to pass in the following page.
- Add varaibale if necessary (for exemple ‘sample temperature’)
- Press on the F4 key [Next] to validate all the entries and to pass in the following page.
- Start writing the formula by using operators (F2) and variables (F3)
- Press on the F4 key [Next] to validate all the entries and to pass in the following page.
-You can define a condition formula wish validate or not the measure (for exemple: limit absorbance value)
- Press on the F4 key [Next] to save and start the method.
Page 53
UviLine Series
- 53 -
1.33 METHOD DELETING
- Select the method to be deleted in the mutli-wavelenght” mode with « » and « » keys and confirm with <START ENTER> key.
- Press F1 key [Setup].
- Select « Edit method » with « » and « » keys and confirm with <START ENTER> key.
- Press F3 key [Delete].
1.34 METHOD SELECTION
The method list is displayed.
Methods are ordered by method number.
Method selection:
- Select a method with « » and « » keys. The active selection is displayed in reverse video.
LIMITATION OF METHODS LIST
If the list is very long, the operator can narrow down the method list and thus make the search easier as follows:
- Press F4 key [Last used], it is possible to restrict the method list to the ten methods last used.
- With the search function, the operator can search certain character strings in the list. The search takes place as a full­text search of the entire list contents. The operator can search for a method number or certain citation form.
Search for a character string:
- Enter the character string to be searched for in the search window with <A...9>. The list appearing below shows all hits containing the character string. The hit list is updated with each character that is entered.
Page 54
UviLine Series
- 54 -
1.35 METHOD EXECUTION
- Select a method with « » and « » keys and confirm with <START·ENTER> key in the mutli-wavelenght” mode.
- Enter de variables values by using F4
- Press <ZERO BLANK> key
- Zero is performed at the programmed wavelengths.
- Press <START ENTER> key to start the sample measurement.
It displays at the same time the absorbance values (or transmittance values) of the sample read at the different wavelengths.
- Press on F3 key [Transmission] to have the transmission values (%).
- Press on <START ENTER> key to run a new measurement.
- If the condition is not met, folowing message is displayed
Page 55
Page 56
UviLine Series
- 56 -
11 SPECTRUM MODE
1.36 GENERAL POINTS
The Spectrum mode makes it possible to measure and record the Absorption and Transmission values according to the wavelength. The wavelength range can be freely selected inside the spectrophotometer measuring range.
Baseline
A baseline has to be recorded before a spectrum is recorded. The baseline has to cover at least the wavelength range of the spectrum to be recorded. Once the baseline is measured, it remains stored in the photometer until a new baseline is recorded, the Spectrum mode is exited or the photometer is switched off.
1.37 METHOD PROGRAMMING
In the “spectrum” mode, press on F1 key [Setup].
* Start wavelength
From 320 nm for the UviLine 8100C / 9400C / 9300 From 190 nm for the UviLine 9400C / 9600
* Stop wavelength
Up to 1100 nm for all UviLine * Scan speed Select « Low », « Medium » or « High »
* Scanning interval 1, 2, 5, or 10nm interval point resolution.
* Mode
Absorbance or Transmission
* Smoothing
YES or NO
Cell changer mode
YES or NO
* Colour of the graph
Brown; Blue; Orange; Green; Red; Purple
* Scaling
Page 57
UviLine Series
- 57 -
Auto or Manual
1.38 METHOD SAVING
- Enter in the « Spectrum » mode.
- Press on F1 key [Setup] to enter inside the programming parameters sub-menu.
- Program parameters.
- Press on F2 key [Save].
- Press on F1 key [Location].
- Select the place where the method will store (Internal folder DataB or USB memory)
- Give a name to the method and confirm with <START/ENTER> key.
1.39 METHOD DELETING
- Enter in the « Spectrum » mode.
- Press on F1 key [Setup].
- Press on F1 key [Open]. It displays:
- Select the method to be deleted with « » and « » keys.
- Press on F2 key [Delete].
1.40 METHOD SELECTION
- Enter in the « Spectrum » mode.
- Press on F1 key [Setup] to enter inside the programming parameters sub-menu.
- Press on F1 key [Open].
- Press on F1 key [Location] if the method is stored inside a USB memory.
- Select the method and confirm with <START/ENTER> key.
1.41 MEASUREMENT EXECUTION
- Enter in the « Spectrum » mode.
- Start the baseline measurement by pressing <ZERO BLANK> key. The spectrophotometer memorizes the baseline. Wait until the end of measurement.
- Insert the sample inside the cuvette holder, close the cover the ress <START ENTER> key to start the measurement.
The cursor appears at the absolute maximum of the spectrum. The operator has the following possibilities:
- Immediately realized some calculations on the spectrum (see following paragraph).
Information curseur
Page 58
UviLine Series
- 58 -
- To print, with <PRINT> key, the spectrum on a connected printer as a graphic.
- To record, with <STORE> key, the spectrum as a “*.csv” file. As the memory location, it is possible to select the spectrophotometer (Internal DataB folder) or a connected USB memory at the USB-A connection (USB memory).
1.42 SPECTRA TREATMENT
1.42.1 D
OWNLOADING OF STORED SPECTRUM
Enter in the « Spectrum » mode then, press on F4 key [Open].
- If necessary, it is possible to select a different memory location for the spectrum with F1 key [Location] (USB medium at the USB-A connection).
- Select the required spectrum. The original view of the curve is displayed.
1.42.2 C
URSOR
The cursor consists of a horizontal and vertical line that crosses each other at a point of the curve. A box names the x and y values of the point of the curve.
Move the cursor along the x axis (wavelength) with « » and « » keys. It is possible to trace and evaluate the curve point by point.
1.42.3 Z
OOM
- Press on F3 key [Zoom].
Page 59
UviLine Series
- 59 -
The zoom window appears. – Press F1 key [Original], at any time, to come back to the original view of the spectrum.
The bottom left corner of the zoom window is marked by a small black square.
Adjusting the zoom window:
– Define the bottom left corner of the zoom window with « » and « » keys and with « » and « » keys. – Use F4 [xy max] key to pass from the bottom left corner of the zoom window to the top right corner of the zoom window.
– Define the top right corner of the zoom window (small black square) with « » and « » keys and with « » and « » keys.
– Press on <START·ENTER> key. The zoom window is scaled up on the entire diagram area.
- Press on <ESC> key to come back to the spectrum original view.
Page 60
UviLine Series
- 60 -
1.42.4 MANUAL CHANGE OF ABSORBANCE SCALE
Press on F1 key [Setup], select « Scaling » and then « Manual » to determine manually the absorbance scale (Y axis). Enter manually the minimum and the maximum of absorbance.
1.42.5 CALCULATIONS
1.42.5.1 Picks and valleys detection
Press on F2 key [Edit].
Select « Etreme values » with « » and « » keys. Confirm with <START ENTER> key. There is to way for the peak to valley detection:
1 - Automatic
2- Manual input
- Enter the minimum distance between minimum and maximum (More the difference between minimum and maximum will be small, more peaks and valleys will be listed)
- Press F1 key [OK] to come back to the curve.
- Press F4 key [Mark] to mark, on the curve, minimum and maximum.
1.42.5.2 Mark points
This function makes it possible to the operator to mark on the spectrum a particular point with its coordinates (Absorbance Wavelength).
- Press on F2 key [Edit].
- Select « Mark points » with « » and « » keys. Confirm with <START ENTER> key.
Page 61
UviLine Series
- 61 -
- Move the cursor along the curve with « » and « » keys.
- Press on F4 key [Mark] to display the points coordinates.
1.42.5.3 Deleting of mark points
Deleting of individual mark point
- Move the cursor along the curve with « » and « » keys. When cursor arrives on marked point, the [Delete] key appears beside [Mark] key.
- Press F3 key [Delete] to remove the selected point.
Deleting of all mark points
Page 62
UviLine Series
- 62 -
When there are marked points, the command « Delete all marks » appears inside the « Edit » menu.
- Select « Delete all marks » with « » and « » keys. Confirm with <START ENTER> key. All marked points will be deleted.
1.42.5.4 Area calculation
Calculate the surface under the curve between two wavelengths [X1, X2] freely chosen by the operators.
- Press on F2 key [Edit].
- Select « Surface calculation » with « » and « » keys. Confirm with <START ENTER> key.
- Press on F3 key [X1] to select left cursor.
- Press on F4 key [X2] to select right cursor. The surface value and the selected wavelength range are displayed at the top right corner of the screen.
1.42.5.5 Derivative calculation
Calculate the derivative of the complete spectrum. For the calculation of the second and the third derivative, it is possible to carry out the function on several times.
- Press on F2 key [Edit].
- Select « Derivative » with « » and « » keys. Confirm with <START ENTER> key.
Back to the original curve
- Press on F2 key [Edit].
- Select « Original values » with « » and « » keys. Confirm with <START ENTER> key.
Page 63
UviLine Series
- 63 -
NOTE: The functions as « Zoom », « Mark points », « Surface calculation » and « Derivative » are also available for a
derivative curve. A derivative curve can be also recorded in memory with <STORE> key.
1.42.5.6 Spectra comparison
Load a second spectrum into the same diagram for direct comparison. NOTE: The comparisons of spectra can be done only on spectra which do not result from calculations (derivative,
multiplication...)
- Press F4 key [Open] and select the first spectrum
- Press F2 key [Edit].
- Select « Compare spectrum » with « » and « » keys. Confirm with <START ENTER> key.
- Select the second spectrum and confirm with <START ENTER> key.
To come back to a normal screen (first selected spectrum):
- Press on F2 key [Edit].
- Select « Original values » with « » and « » keys. Confirm with <START ENTER> key.
1.42.5.7 Addition of two spectra
- Press F4 key [Open] and select the first spectrum or use the acquired spectrum
- Press F2 key [Edit].
- Select « Add spectrum » with « » and « » keys. Confirm with <START ENTER> key.
- Select the second spectrum and confirm with <START ENTER> key. It displays the summary of the two spectra.
Page 64
UviLine Series
- 64 -
To come back to a normal screen (first selected spectrum):
- Press on F2 key [Edit].
- Select « Original values » with « » and « » keys. Confirm with <START ENTER> key.
1.42.5.8 Subtraction of one spectrum from another one
- Press F4 key [Open] and select the first spectrum or use the acquired spectrum
- Press F2 key [Edit].
- Select « Subtract spectrum » with « » and « » keys. Confirm with <START ENTER> key.
- Select the second spectrum and confirm with <START ENTER> key.
To come back to a normal screen (first selected spectrum):
- Press on F2 key [Edit].
- Select « Original values » with « » and « » keys. Confirm with <START ENTER> key.
1.42.5.9 Division of a spectrum
Divide the absorbance or % transmission values of the current spectrum by the values of a stored spectrum
- Press F4 key [Open] and select the first spectrum or use the acquired spectrum
- Press F2 key [Edit].
- Select « Divide spectrum (ratio) » with « » and « » keys. Confirm with <START ENTER> key.
- Select the second spectrum and confirm with <START ENTER> key.
To come back to a normal screen (first selected spectrum):
- Press on F2 key [Edit].
- Select « Original values » with « » and « » keys. Confirm with <START ENTER> key.
Note: This function always applies to the common wavelength range of both spectra only.
1.42.5.10 Addition of a constant
- Press F2 key [Edit].
- Select « Add fixed value » with « » and « » keys. Confirm with <START ENTER> key.
- Enter the positive or negative fixed value and confirm with <START ENTER> key.
Page 65
UviLine Series
- 65 -
To come back to a normal screen:
- Press on F2 key [Edit].
- Select « Original values » with « » and « » keys. Confirm with <START ENTER> key.
1.42.5.11 Multiplication by a constant
Multiplies the absorbance or % transmission values of the current spectrum by a constant value.
- Press F4 key [Open] and select the first spectrum or use the acquired spectrum
- Press F2 key [Edit].
- Select « Multiply fixed value » with « » and « » keys. Confirm with <START ENTER> key.
- Enter the constant value and confirm with <START ENTER> key.
To come back to a normal screen:
- Press on F2 key [Edit].
- Select « Original values » with « » and « » keys. Confirm with <START ENTER> key.
Page 66
UviLine Series
- 66 -
Page 67
UviLine Series
- 67 -
12 KINETICS MODE
1.43 GENERAL POINTS
The Kinetics function enables the temporal tracing of the absorbance or transmission of a sample at a certain wavelength. For recording, the spectrophotometer carries out single measurements at regular intervals (measuring interval) and stores the measured values as a time function. All measuring parameters are entered by the operator in order to create methods which can be stored, edited and deleted.
1.44 METHOD PROGRAMMING
- Press on F1 key [New method].
* Number
4001…4020
* Name
Name of the method (18 characters).
* Mode
Absorbance or Transmission
* Wavelength
320 – 1100 nm for l’UviLine 8100C / 9100C / 9300 190 – 1100 nm for l’UviLine 9400C / 9600
* Duration (Delay + Measure)
Total duration of the kinetics including the waiting time (Delay). Format hh:mm:ss (hours:minutes:seconds)
* Interval
It is the time interval between two successive single measurements. Format hh:mm:ss (hours:minutes:seconds)
* Delay
Time between the start of the recording and the start of the first single measurement. This time allows avoiding the no linear part of a reaction. Format hh:mm:ss (hours:minutes:seconds)
Page 68
UviLine Series
- 68 -
* Scaling
Auto or Manual Scaling: Auto **
The spectrophotometer adjusts the scaling of the axes (minimum and maximum value of the axis) to the measured values
while measuring. The entire curve is always visible.
Scaling: Manual Abs min. Abs max.
The scaling of the axes (minimum and maximum value of the axis) is permanently set manually.
* Measurements/interval
1/interval or Max./interval Here, it is possible to define how many measurements will be carried out per interval. The setting affects the calculation of the slope of the individual intervals.
The displayed slope for an interval is determined as follows, depending on the profile:
Measurements/interval Slopes
1/interval Difference of two measured values (2 points calculation) Max./interval Slope of the straight line determined in an interval by linear regression
* Katalytic activity
YES or NO
Katalytic activity Result
NO Total slope report to the minute YES Total slope report to the minute multiplied by a factor
* Factor
Factor multiplying the average slope report to the minute.
* Units
i.e.: µkat (10 characters maximum).
* Resolution
From 0 to 3 Number of digits, after comma, for the result.
1.45 METHOD SELECTION
- Select a method in the “Kinetcis” mode then press on <START·ENTER> key to confirm the method choice.
1.46 METHOD SAVING
Methods are automatically saved in memory.
Page 69
UviLine Series
- 69 -
1.47 METHOD DELETING
- Select the method to be deleted with « » and « » keys. Confirm with <START ENTER> key. Then press F1 key [Setup].
- Select “Edit method” with « » and « » keys. Confirm with <START ENTER> key.
- Press F3 key [Delete].
1.48 LIMITATION OF METHODS LIST
If the list is very long, it is possible to narrow down the methods list and thus make the search easier as follows:
- Press on F4 key [Last used]; it is possible to restrict the method list to the ten methods last used.
- With the search function, it is possible to search certain character strings in the list. The search takes place as a full-text search of the entire list contents. Thus the operator can search for a method number or name.
Search for a character string: Enter the character string to be searched for in the search window with <A...9>. The list appearing below shows all methods containing the character string. This methods list is updated with each character that is entered.
Page 70
UviLine Series
- 70 -
1.49 METHOD EXECUTION
In the “Kinetics mode” select a method with « » and « » keys and- Press on <START·ENTER> key to confirm the method choice. Then -press <ZERO BLANK> key
- Insert sample.
- Press <START ENTER> key to start the measurement. The spectrophotometer starts recording automatically.
- Wait the end of measurement.
Stopping the recording:
- Press on F4 key [Stop] to terminate the recording prematurely. The curve recorded up to this point can be stored and edited (see paragraph 12.8.3).
- Press on <ESC> key to completely cancel measurement. The curve recorded up to this point is discarded. After the specified Duration has expired, the cursor appears.
- Press on <START ENTER> key to run a new measurement.
Page 71
UviLine Series
- 71 -
1.50 METHOD EDITING
- Enter in « Kinetics » mode.
- Select a method with « » and « » keys, confirm with <START ENTER> key.
- Press F1 key [Setup].
- Select « Edit method » with « » and « » keys, confirm with <START ENTER> key.
1.51 KINETICS TREATMENT
1.51.1 K
INETIC CURVE PRINTING
- Press <PRINT> key to output the kinetic curve to a connected printer as a graphic.
1.51.2 C
URSOR
The cursor consists of a horizontal and vertical line that crosses each other on a point of the curve. A box names the x and y values of the point of the curve.
Move the cursor along the x axis (time axis) with « » and « » keys. By this way, it is possible to trace and evaluate the curve point by point. These points are corresponding to the measurement interval.
1.51.3 K
INETIC CURVE SAVING
Press <STORE> key to save the kinetic curve. If the operator forgot to save the curve, this one is nevertheless kept in memory under the name of "date kineticsBackup.csv".
1.51.4 D
OWNLOADING OF STORED KINETIC CURVE
- Press F4 key [Open].
The list with the stored kinetic records (Internal DataB folder) is displayed.
- Press F1 key [Location] to select another memory location of the kinetic record (USB memory)
- Select the kinetic curve and validate.
Page 72
UviLine Series
- 72 -
1.51.5 « E
DIT » MENU
1.51.5.1 Slope
Select « Slope & catalytic activity».
« Interval » column It is the number of measurement interval.
« Slope [/min] » column It is the slope report to the minute per interval.
« Time [s] » column It is the time at the end of each interval.
RESULT:
It is the slope, report to the minute, of the straight regression line passing through various points of measurement, multiplied by the factor and expressed with the resolution and unit selected by the operator.
1.51.5.2 Comparison of kinetic curves
Load a second kinetics curve into the same diagram for direct comparison.
- Press on F3 key [Edit].
- Select « Compare kinetics ».
- Select another kinetics curve. Confirm with <START ENTER> key
Page 73
UviLine Series
- 73 -
To come back to a normal screen (first selected kinetics curve):
- Press on F1 key [Back].
1.51.5.3 Subtraction of a kinetic curve from another one
Subtract a stored kinetics curve from the current kinetics curve.
- Press on F3 key [Edit]
- Select « Subtract kinetics ».
- Select a kinetics curve. Confirm with <START ENTER> key. The curve resulting of this operation can’t be saved in memory. To come back to a normal screen (first selected kinetics curve):
- Press on F1 key [Back].
NOTE: To subtract one kinetics curve from another one, the intervals have to be identical.
1.51.6 M
ANUAL CHANGE OF ABSORBANCE SCALE
Press on F1 key [Setup], select « Scaling » and then « Manual » to determine manually the absorbances scales (Y axis). Enter the absorbance minimum and maximum.
1.52 CURVE PRINTING
- Press <PRINT> key to print the curve on a connected printer.
UviLine 9300 7449001 1.14 Administrator 04/10/2015 14:05:46 4003
Page 74
UviLine Series
- 74 -
Page 75
UviLine Series
- 75 -
13 MEMORY
1.53 OVERVIEW
Page 76
UviLine Series
- 76 -
Measurement data Saving and exporting
Concentration Absorption / % transmission Multi-wavelengths
Measurement datasets of these modes are saved with <STORE> or AutoStore in the measured value memory of the spectrophotometer first (5000 locations).
The measured value memory is available via the Measurement data memory menu.
Here it is possible to view or filter stored measurement datasets or export them into a PC readable file (*.csv) (<STORE>).
Csv files of these modes cannot be imported back into the photometer.
Spectrum Kinetics
Measurement datasets of these modes are saved and exported as a PC readable file (*.csv) with <STORE>.
Csv-files of these modes can be re-imported and viewed on the photometer.
As the location for the PC readable files (*.csv) to be stored, it is possible to select for each export either the spectrophotometer (Internal DataB folder) or an external memory (USB memory).
Later it is possible to transmit the files stored in the photometer (Internal DataB folder) to a connected PC or an external memory (USB memory).
It is possible to transmit to a USB memory either individual files with measuring data or all files stored in the Internal DataB folder.
1.54 INSTRUCTIONS ON USING USB MEMORY DEVICES
The safety of data stored on USB memory devices depends on the quality of the memory device and the data transmission. Data is stored partly or not at all if for example:
The power supply of the external memory device is interrupted during the write process, or
The external memory device is prematurely disconnected from the photometer during the data backup.
To prevent a data loss we recommend the following:
Save all data internally in the photometer first.
After performing a backup leave the USB memory device connected to the photometer for some time.
Check whether the stored data is complete, e.g. on a PC.
Use the USB memory device for data transport but not for permanent data storage.
1.55 MEASURMENT DATASETS
1.55.1 E
LEMENT OF A MEASUREMENT DATASET
A complete measurement dataset consists of:
- Consecutive number (is automatically assigned by the spectrophotometer).
- Date/time.
- Identification (e.g. ID or "AutoStore").
- User name.
- Measured parameter, e.g. method number, dilution, wavelength (depending on the measuring mode).
- Measured value with unit and if necessary citation form.
1.55.2 O
PÉRATIONS WITH MEASUREMENT DATASETS
Measurement datasets can be
Stored
Displayed and printed
Filtered, i.e. selected or hidden based on certain criteria
Deleted
Copied
Page 77
UviLine Series
- 77 -
Remark:
If the memory is full, you can erase measurement datasets, or overwrite the oldest dataset with the next storing procedure. A security prompt appears before a dataset is overwritten. To backup the measurement data, you can transmit the measurement datasets from the measurement data memory to the internal DataB folder or a USB memory device connected to the USB-A connection and archive them further from there
1.55.3 R
ESULT RECORDING
1.55.3.1 Manual recording
After each measurement, it is possible to store the measurement data manually with the <STORE> key. It is stored in the measurement data memory. The memory symbol in the header indicates that the measurement data displayed on
the screen is ready to be stored. With the measuring modes, Concentration, Absorbance / % Transmission, and Multi wavelengths there is the additional option to automatically store all new measured values at the time of the measurement
When storing manually, an input field for the identification (ID) appears after pressing the <STORE> key. Here it is possible to enter an individual combination of alphanumeric characters for later easier identification of the measurement datasets. 30 digits are available for this.
WARNING:
In « Absorption / % transmission » mode, the « AutoStore » function is not operational for data acquired during this measurement. If the operator wants to store data, he has to use <STORE> key. So, all results acquired during last measurement will be stored under the names given by the operator.
1.55.3.2 Automatic recording with « AutoStore » function
- Press on F1 key [Setup].
- Select « AutoStore » with « » and « » keys and confirm with <START ENTER> key.
The symbol at the end of « AutoStore » indicates that the function is active. If this function is not active, it means that data will be saved manually only with <STORE> key.
Remark
For the measuring modes, Concentration, Absorbance / % Transmission, and Multi wavelengths, it is possible to save every measured value automatically with (AutoStore) function. All automatically saved measurement datasets are given the label "AutoStore". If the same measurement dataset is manually saved using (<STORE>) key, the "AutoStore" label is overwritten by ID code given by the operator.
The AutoStore setting is valid for all three measuring modes, Concentration, Absorbance / % Transmission, and Multi wavelengths.
Page 78
UviLine Series
- 78 -
1.55.4 R
ESULT DISPLAYING
From main menu
From concentration menu
From Absorption/Transmission menu
From Multi-wavelengths menu
Each one of these possibilities displays the contents of the measurement data memory in the form of list in the following way:
- Select « Measurement data memory » with « » and « » keys and confirm with <START ENTER> key.
<HOME>
– [General setup]
– Measured value memory
<HOME>
- Concentration – Choice of a memorized method
- [Setup]
– Measured value memory
<HOME>
- Absorption/Transmission
- [Setup]
– Measured value memory
<HOME>
- Multi-wavelengths – Choice of a memorized method
- [Setup]
– Measured value me
mory
Page 79
UviLine Series
- 79 -
If there are more datasets available than can be displayed, the arrows « » and « » displayed additionally. Filter indicates that filter settings are active. In this case, only those datasets are displayed that correspond to the
selected filter criteria.
Options:
Measurement datasets can be:
displayed in short form as a list or in details as individual values ([List] <–> [Single value])
filtered (see section 4.12.7 and section 4.12.8)
deleted (see section 4.12.9).
with <STORE>, you can store the entire displayed list as a *.csv file in the internal DataB folder or on a USB
memory device connected to the USBA connection. The filter settings apply to the storing process. You can freely select the file name. Thus you can, e. g. store in a separate file and systematically archive measurement data of a certain period.
with <PRINT>, the entire displayed list can be printed. The filter settings apply to the print process.
Page 80
UviLine Series
- 80 -
1.55.5 R
ESULT FILTERING
1.55.5.1 Filtering of a results group
It is possible to select a result or a group of results with the “Filter” function. The functions display, delete and download stored measurement datasets refer to all stored measurement datasets that
correspond to the specified filter criteria.
Filter criteria: The following filter criteria can be set:
Mode (Absorption / % transmission and Multi-wavelengths)
User
ID (identification)
Date (date from ... to ...)
Method (for the measured parameters, Concentration and Multi-wavelength)
Remark: Alternatively, it is possible to hide measurement datasets that meet the specified filter criteria with the “Selected values: invert selection” function).
1.55.5.2 Filters inversion
With the “Selected values: invert selection” function, it is possible to hide all measurement datasets that correspond to the specified criteria of the filter.
Remark: It is possible to use this function to select and delete measurement datasets no longer used.
1.55.5.3 Erase stored measurement dataset
It will delete only results corresponding to the normal or invert selection criteria.
- Select « Delete memory (selected values only or all values) » with « » and « » keys and confirm with <START ENTER> key.
- Select « OK » or « Cancel »
Page 81
UviLine Series
- 81 -
1.55.5.4 Result deleting
It is possible to delete only one result.
- Select the result to be deleted with « » and « » keys in the data memory.
- Press F3 key [Delete].
- Select « OK » or « Cancel »
1.55.5.5 Saving Kinetic recordings, spectrUM
After the following measurements, the Save dialog opens and prompts you to save the data in a *.csv file:
Kinetics
Spectrum
If the data are not saved in *.csv format, they are lost when the measuring mode is terminated.
Remark:
During a kinetic recording, the current measurement is always saved in the file, "KineticsBackup.csv" for safety reasons.
1.55.5.6 Saving data as a pdf file
All data that can be printed (printer symbol on the display) can also be saved as a pdf file. The pdf file contains the data that are also output to a USB printer. Kinetic recordings and spectra are stored in the pdf file as a graphic.
Saving as a pdf file and printing is done with the <PRINT> key. Prerequisite is that pdf printing is set as the printer in the menu <HOME>/General setup/ Data transfer/Printer/Function of PRINT key. Subsequently, enter a file name and select the storage location (internal folder DataB or USB memory device).
1.55.6 S
AVING / EXPORTING FILES
If you want to back up or process measurement data files outside the photometer, you can copy them to external media.
1.55.6.1 Copying all measurement data files to a USB memory device
Page 82
UviLine Series
- 82 -
You can very simply transfer all measurement data files from the photometer (Internal DataB folder) to a connected USB memory device.
All measurement data files from the photometer (Internal DataB folder) have been transferred to the USB memory device. The complete folder structure from the photometer is created on the USB memory device. The individual measurement data files are stored in subfolders sorted by measurement data types.
1.55.6.2 Copying user-defined methods / profiles to a USB memory device
A list is displayed that includes all user-defined methods and profiles available on the photometer. All methods and profiles are checked off with a checkmark. All methods and profiles checked off are saved.
If necessary, select individual methods/profiles with and remove the checkmark with <STAR/ENTER>. These methods/profiles will not be saved. Start the save process with <STORE>. A message appears when the data have been saved. Confirm the message with <START/ENTER>. The save process is completed. The data are stored in the Exchange_Method_Profile folder on the USB memory device. The individual files with the methods/ profiles are in subfolders. Already existing files with identical names are overwritten without confirmation prompt
1.55.6.3 Copy of individual file on USB memory
1.1.1.1.1 Spectrum and kinetic curve after acquisition
After measurements of spectrum or kinetics, the user can save his curve spectral or kinetic in a “*.csv” file:
- Press F4 key [Open] to enter in the safeguard sub-menu.
- Press F1 key [Location].
- Select « USB memory ».
- Give a name.
- Validate with <START/E NTER> key.
If the safeguard, in “*.csv” format were not carried out, the measurement data are lost when the user leaves the measurement mode.
Page 83
UviLine Series
- 83 -
Remark
With kinetic recordings, the current measurement is always saved in the "KineticsBackup.csv" file for safety reasons.
1.1.1.1.2 Spectrum and kinetic curve already memorized inside the « Internal file DataB »
- Press F4 key [Open].
- Select a spectrum or kinetics curve.
- Confirm with <START/E NTER> key.
- Press <STORE> key to open save dialog box.
- Press F1 key [Location].
- Select « USB memory ».
- Change the name if necessary.
- Validate with <START ENTER> key.
1.1.1.1.3 Absorption/Transmission, Concentration, Multi-wavelengths mode
The measurement data of Absorbance/ Transmission, Concentration and multi-wavelengths are stored in the measurement data memory first, either automatically or manually (with the <STORE> key). Data stored in the measurement data memory can be filtered with filter criteria and then exported in the PC readable in “*.csv” format.
- Press F1 key [Setup] inside one of the measuring mode or [General setup] from the main menu
- If necessary, set any filter criteria with [Setup].
- Press <STORE> key to open a save dialog box. The spectrophotometer automatically suggests the location, Internal DataB folder, and a file name.
- If necessary, use F1 key [Location] to change the location (USB memory).
- If necessary, change the suggested file name.
- Press <START ENTER> to save the selected results.
1.55.7 P
RINTING THE DATA
(USB)
Usable printers Data can be printed with standard printers (ink-jet or laser) connected to the USB-A interface. Suitable are all printers that
can interpret the PCL-3 printer control language.
The printer symbol ndicates that the display contents can be printed. To print, press <PRINT>. Pdf file As an alternative, you can also output the print data to a pdf file.
Remark
In den following paragraphs, "Print" means:
- output to a USB printer
- output to a pdf file.
1.55.8
SETTING FOR DATA TRANSMISSION
Decimal separator for csv-Files For the output of CSV files, it is possible to select a comma or point as the decimal separator. Data format When printing measurement datasets, it is possible to select a short or long version with different information contents. Function of PRINT Key Her you can choose between a USB printer & PDF file
Page 84
UviLine Series
- 84 -
1.55.9 I
MPORTING FILES
You can import to an UviLine spectrophotometer the data that were created with the same or another UviLine spectrophotometer, and were then saved to a USB memory device. You can import the following data:
Spectra
Kinetic recordings
User-defined methods
Profiles
1.55.9.1 Importing spectra or kinetic recordings from a USB memory device
You can import to the photometer any spectrum or kinetic recording by opening an externally stored spectrum or kinetic recording with the Open function of the photometer.
1.55.9.2 Importing methods / profiles from a USB memory device
A list is displayed including all user-defined methods and profiles stored in the corresponding subfolders of the Exchange directory on the USB memory device. All methods and profiles are checked off with a checkmark. All methods and profiles checked off are imported.
Start the import with [Import]. A confirmation prompt appears before any data on the photometer are overwritten. A message appears when the data have been imported. Confirm the message with <START·ENTER>. The import is completed. The imported methods / profiles are available on the photometer.
Remark
When importing methods make sure that your photometer supports the wavelengths of the imported methods.
Page 85
UviLine Series
- 85 -
14 USERS MANAGEMENT
There are a user name and a password by fault:
- User name: Administrator
- Password: admin Respect the spaces and the letters size (Capital or small letter). The functions of the user management are only available for users of the user group, Administrator.
An administrator can:
Activate or deactivate the user management for the spectrophotometer.
Create, change or delete individual user accounts.
The spectrophotometers UviLine Series allow the management of up to 100 users. Each user is a member of a user group with defined user rights.
1.56 USER LEVEL AND USER RIGHTS
There are three hierarchical user groups:
Administrator (top level)
User (user account registered by the administrator)
Guest (user without user account)
Administrators and users log in to the photometer with their user name and password. Guests can optionally enter a name for their login. Thus, documented measured values can later be assigned to the user.
Action Administrator User Guest
Select methods
Execution of methods
Record of results
Deactivate automatic records of the results
Activate or deactivate « users management » function
Edit and modify a method stored in memory
Delete a method stored in memory
Date and time setting
Delete of results stored in memory
Users management
Reset of spectrophotometer setting
Update of the software
Remark
You can also switch off the user management and reactivate it asnecessary. To do so, you need administrator rights. If the user management is switched off, the user name and password do not have to be entered. Each user has full rights.
1.57 ACTIVATION/DESACTIVATION OF USER MANAGEMENT
Each user can activate the user management function. If the function is deactivated, each user has administrator rights. Only members of the user group, administrator can deactivate the user management function.
Page 86
UviLine Series
- 86 -
If the function is active, each user has to log in the spectrophotometer. After the login, the user has certain rights depending on the user group.
1.57.1 A
CTIVATION OF USER MANAGEMENT FUNCTION
User management activation is accessible in the “general setup” menu
1.57.2 D
ESACTIVATION OF USER MANAGEMENT FUNCTION
Select « Deactivate user management » and confirm. Then, select « OK » and confirm. The user management function is desactivated. Each user has administrator rights.
REMARK
If the user management is deactivated by a user of the Administrator user group, all user accounts that were set up are lost. The password for the administrator is reset to "admin".
1.58 CREATION, MODIFICATION OR DELETING A USER ACCOUNT
When the user management function is active, a user with administrator rights can manage user accounts.
1.58.1 U
SER ACCOUNT CREATION
During the creation of a user account, the Name, User group and Password are defined.
Page 87
UviLine Series
- 87 -
1.58.2 U
SER ACCOUNT MODIFICATION
At the time of the modification of a user account, it is possible to modify the membership of User group and the Password. Only the user having an Administrator account is authorized to make these changes.
1.58.3 P
ASSWORD MODIFICATION
The administrator creates user accounts and assigns a password on each user account. As soon as a user opened a session successfully with his user account, it can modify itself the password of his user account.
1.58.4 U
SER ACCOUNT SUPPRESSION
1. Select a user account.
2. Press F3 key [Delete] to delete the user account: A security prompt appears.
3. Confirm the security prompt: The user account is deleted.
Page 88
Gamme UviLine
- 104 -
Page 89
Gamme UviLine
- 105 -
15 VARIOUS ADJUSTMENT
1.59 LANGUAGE CHOICE
In “general setup” menu, select “Language”, then validate the selected language with <START·ENTER> key.
The language selected is noted with the symbol .
1.60 DATE AND TIME ADJUSTMENT
In “general setup” menu, select “Date/Time”. Enter the current date with <0...9> keys and confirm.The input field closes. The date is accepted. Enter the current hour with <0...9> keys and confirm. The input field closes. The time is accepted.
1.61 CONTRAST ADJUSTMENT
It is possible to adapt the contrast to the lighting conditions. Select and confirm « Contrast ». An adjustment system using a cursor is displayed to adjust the screen contrast.
Adjust the screen contrast by pressing « » and « » keys and confirm with <START/ENTER> key. Come back to the main menu with <ESC> key.
Page 90
Gamme UviLine
- 106 -
1.62 RESET
It is possible to reset (initialize) the measurement settings or all settings.
Remark
The reset function is only available for users belonging to the user group « Administrator ». There are different possibilities to reset the instruments settings.
Parameters reset All settings are deleted except the memory which contains
measurements results, methods defined by the user and the measured blank values.
Delivery conditions All settings (including measurement data memory and user-
defined methods) are deleted and the spectrophotometer is reset to the delivery condition.
Service lamp counter Allow to reset to zero the service lamp counter after the
lamp changing
1.63 LAMP SERVICING COUNT DOWN
The photometer counts the operating hours of the lamp. The information on the operating hours of the lamp is given in the Info menu. The number quoted there corresponds to the number of measurements. After the lamp was exchanged you should reset the lamp counter (previous section).
1.64 SPECTROPHOTOMETER INFORMATION ([INFO])
Following information relating to the spectrophotometers are displayed by pressing F4 “info” on the main menu:
Instrument designation.
Instrument serial number.
Number of the instrument software version.
Updating date of the software.
Hardware version
Hardware status (for servicing).
Number of lamp working hour.
Results of the tests of the most important parts of the spectrophotometer.
Memory status.
Page 91
Gamme UviLine
- 107 -
Registered user.
1.65 TIMER
This « Timer » menu allows storing in memory a waiting time before to start a measurement (Ex: Incubation time of a react ional mixture). This « Timer » acts like a stop watch.
- Press
<TIMER> key
- Press F3 key [Edit] to enter a delay.
- Press F1 key [Start] to start the time count down.
- Press F2 key [Stop] to stop the time count down. When the programmed delay is ended,
The timer displays 00:00:00,
Its status commutes from « Running » to « Expired » and
One can hear a “bip”. Press F2 key [Stop] to deactivate the timer.
NOTE:
The [Add] function key is only displayed if a method is selected for which analysis timers were programmed in concentration mode. Press < timer > in the concentration mode.
1.66 SOFTWARE UPDATING
The software and method update is used to continuously update the spectrophotometer.
Remark
Only users of the user group, Administrator are allowed to carry out a software and method update. The update comprises
The newest firmware (meter software)
New or changed method data
Data transfer to the spectrophotometer can be made with USB medium as a temporary storage To realize the update, new software version or new methods have to be recording on USB medium. This one has to be connected to
the spectrophotometer.
Execution
1. Connect the USB medium to the PC.
2. Unpack the contents and complete folder structure of the downloaded exe or zip file in the main directory (top level) of the USB medium (If an unpacking program such as WinZip is used, the option, "Nutze Ordnernamen" or "Use Folder Names" must be set).
Page 92
Gamme UviLine
- 108 -
3. Connect the USB medium to the spectrophotometer.
4. Switch on the spectrophotometer if necessary.
Press « » and « » keys to select « USB memory » as a source and press <START·ENTER> key. The update process takes about five minutes.The spectrophotometer switches itself off and then on again.
Remark
If the update cannot be carried out, an error message appears on the display. Check whether the "Update" folder with its subfolders is stored on the USB memory device (top level). If on the photometer there is not enough free memory capacity for the update, you can create memory capacity by erasing
measurement data. Save your data to a USB memory device before erasing them on the photometer.
Page 93
Gamme UviLine
- 109 -
16 MAINTENANCE
1.67 LAMP CHANGING
In this chapter only the change of the UviLine 9300 halogen lamp is explained. Indeed, UV-Vis UviLine spectrophotometer works with xenon lamp which time life approaches the one of the apparatus. Any time, if it prematurely stops working, then contact AQUALABO ANALYSE servicing.
The tungsten halogen lamp is a wear part with a certain average service life. It has to be replaced if defective. The spectrophotometer has a service hour counter for the lamp module.
NOTE
The replacement lamp is readily assembled as a lamp module and optically adjusted in the factory. Therefore, treat it with utmost care. Fingerprints on the glass will shorten the service life of the lamp. Do not touch the bulb of the new lamp module with bare fingers. If you have touched the bulb inadvertently, carefully clean it with a clean cloth soaked in alcohol.
The halogen lamp is behind a protection cover (Aluminum sheet) at the rear part of the spectrophotometer. To change it, proceed as follows:
WARNING The lamp becomes very hot during operation. Do not touch the hot lamp because it can cause burns! The lamp should cool down for approx. 10 minutes before it is exchanged.
1. Switch off the spectrophotometer and disconnect it from the main power.
2. Remove the two screws (1) and remove the lamp protection cover (2).
3. Disconnect the electrical plug connection (3). Disconnecting it might take a lot of effort.
4. Unscrew the two knurled-head screws (4) and remove the defective lamp module (5).
5. Fix the new lamp module with the knurled-head screws. When doing so, the metal-plated side of the PCB must point outward, toward the knurled-head screws.
6. Reconnect the power supply cable.
7. Reinstall the protection lamp cover.
8. Reset the service hour counter for the lamp module.
1
2
3
4
5
Page 94
Gamme UviLine
- 110 -
1.68 BUFFER BATTERIES CHANGING
Information
Only use leak proof alkaline manganese batteries. If you leave the spectrophotometer switched on during the exchange or insert the new batteries within a minute after taking out the old ones, the date and time are retained in the spectrophotometer.
1. Turn on the spectrophotometer upside down and place it on a soft surface.
2. Remove the cover (1) of batteries compartment.
3. Remove out of the compartment the old batteries.
4. Insert the four new batteries in the battery compartment. Make sure that the poles of the batteries are in the correct position.
5. The ± signs on the batteries must correspond to the ± signs in the battery compartment.
6. Close the batteries compartment cover.
Batteries time life
The power consumption of the clock is very low. The lifetime of high quality batteries is at least 5 years.
Page 95
Gamme UviLine
- 111 -
1.69 CLEANING
WARNING
The housing components are made out of synthetic materials (ABS, PMMA and PC). Thus, avoid contact with acetone, ethyl alcohol and similar detergents that contain solvents. Any splashes must be wiped off immediately.
1.69.1 C
LEANING YOUR SEPCTROPHOTOMETER
Clean your photometer as follows:
If the housing surface is dirty, wipe it with a soft cloth and mild soapy water.
Remove any chemicals splashes as soon as possible.
For disinfection, you can use isopropanol for cleaning for a short time.
1.69.2 C
LEANING THE LENSES
AQUALABO ANALYSE advises to regularly check that the 2 lenses inside the cell holder are clean. In order to access the lenses, use the Quick –Lock system to take out the accessory inserted in the cell holder (see picture below). If necessary, then proceed to the cautious cleaning of the 2 lenses using a cotton bud lightly soaked with isopropanol. Check there are no deposit/traces before use.
1.70 WHAT CAN YOU DO, IF…
1.70.1 W
ARNING IF CUVETTE IS BROKEN
WARNING Cells can contain dangerous substances. If the contents are released, follow the safety instructions in the package insert. If necessary, take corresponding protective measures (protective goggles, protective gloves etc.).
CAUTION Do not turn the photometer upside down to remove the liquid! When doing so, the liquid could come into contact with electronic components and damage the photometer.
The photometer has a drain device through which the contents of a broken cell can drain off without causing any damage.
Proceeding after a cuvette is broken:
1. Switch of the spectrophotometer and disconnect it from the main power supply.
2. Let the liquid drain off into a suitable container and dispose of it properly according to the instructions of the reagent package.
3. Carefully clean the cuvette well using a wet, lint-free cloth.
4. Let dry the cuvette well.
Lenses
Quick Lock
Page 96
Gamme UviLine
- 112 -
Remark
After recommissioning, check out the spectrophotometer for all measurements.
1.71 TROUBLESHOOTING
The spectrophotometer doesn’t switch on
Cause Remedy
– The power cable is not connected – Connect correctly the power cable
Measurement range undercut or exceeded
Cause Remedy
– Method not suitable. - Select another method which suitable measuring range.
- Dilute the sample.
Not stable results
Cause Remedy
- The cover of cuvette compartment is opened - Close the cover of the cuvette compartment
- The halogen lamp arrives at its end life - Change the halogen lamp
Incorrect measured values
Cause Remedy
- Cuvette dirty - Clean the cuvette
- Wrong dilution - Make a dilution again
- The selected method is not appropriated - Select another method
- Erroneous zero measurement - Perform the zero again – Erroneous blank value – Measure the blank again
Connected printer does not print
Cause Remedy
- Printer not suitable
- Connect a printer that can interpret the printer control language PCL-3
Data transmission to USB memory device does not work
Cause Remedy
-Connected USB memory device was not recognized – The USB memory device has been formatted to a file system which is not supported, e. g. (NTFS)
- Use other USB memory device – Format the USB memory device to the FAT 32 file system
Incorrect measured values
Cause Remedy
- Cuvette dirty - Clean the cuvette
- Wrong dilution - Make a dilution again
- The selected method is not appropriated - Select another method
- Erroneous zero measurement - Perform the zero again – Erroneous blank value – Measure the blank again
Page 97
Gamme UviLine
- 113 -
17 ACCESSORIES
List of manual and automatic accessories available on UviLine instruments
1.72 10MM CELL HOLDER (REF. 70VI0604)
Single cell holder up to 10mm optical path length. The 10mm cell holder guarantees a perfect positioning for all types of cells. It can be used with micro-cells up to 50µl. Quick and easy installation thanks to the Quick-Lock system.
1.73 PORTE CUVE JUSQU’A 100MM (REF. 70VI0605)
Tank holder up to 100mm of optical thickness for tanks (comes as standard with PASTEL UviLine devices). Suitable for 5/10/20/50/100mm tanks and 16mm tubes, it rigorously covers all manual uses of the UviLines that require high reading sensitivity. Quick and easy installation due to the Quick-Lock system.
1.74 PORTE CUVE 10MM THERMOSTATABLE (REF. 70VI0602)
Thermostatable 10mm cell holder with integrated thermostatable function. Allows all standard cell with a 10mm optical path length. Quick and easy installation thanks to the Quick-Lock system. Thermostatable by external bath.
1.75 PORTE TUBE 16MM (REF. 70VI0607)
16mm tube holder. Can be used in combination with micro-methods and especially designed for the use of 16mm diameters tubes. Quick and easy installation thanks to the Quick-Lock system and the sketch indicating how to position the tube.
1.76 AUTOMATIC CELL CHANGER TURRET (REF. 70VI0600)
Designed to analyze 5 samples and 1 blank, the 5+1 turret find its place in automation applications and guarantees measurements stability with unique advantages:
Removable and easy to re-insert turret for convenient cells set up  Accurate positioning for low volume cells
Page 98
Gamme UviLine
- 114 -
1.77 SIPPER (REF. 70VI0601)
Compact, the sipper is delivered standard with a peristaltic pump integrated to the cell holder allowing:
Programmable sucking from 500µl to 3000µl  Use of flowthrough cells from 30µl to 450µl  It makes manipulation safer and higher in productivilty
1.78 EFFET PELTIER (REF. 70VI0603)
T°C set up between 10°C and 60°C  Accuracy: + 0.5°C  Sipper option  Fast, ultra compact and fully instrument driven
1.79 SIPPER WITH PELTIER EFFECT (REF. 70VI0610)
Compact, this accessory is equipped with a peristaltic pump cassette integrated to the cell holder and a Peltier module temperature regulating system. Temperature programmation between 10°C and 60°C
• Accuracy: + 0.5°C
• Aspiration programmation: from 250µl to 2000µl
• Cells: from 30µl to 450µl
• It secures and increases productivity handling
• Very compact, fast and completly controlled by the device
Page 99
Gamme UviLine
- 115 -
1.80 DIAPHRAGMS FOR MICRO-CELLS (REF. 70VI0608)
Reference: 70VI0608 (optional accessory) Designation: Batch of 5 diaphragms for micro-cell
Reference 70VI0608 includes the following 5 diaphragms: 0M8582 : size 1,8 x 2,3 mm ; 0M8583 : size 3,8 x 2,3 mm ; 0M8584 : size 1,8 x 4,8 mm ; 0M8580 : size 1,8 x 11 mm ; 0M8581 : size 0,9 x 11 mm
Use: these diaphragms are to be used with micro-cells without black walls. They help concentrate the beam on the sample through the whole in the metal strip, without the light being refracted in all directions.
You will find here below the list of SECOMAM cells for which we advise to use the diaphragms. This list indicates the reference of the diaphragm to be used with the micro-cells.
Cell reference
Diaphragm reference
Cell reference
Diaphragm reference
Cell reference
Diaphragm reference
Cell reference
Diaphragm reference
0GQ53Z0 0M8580 0GQ244Z0 0M8580 0GQ424Z0 0M8581
0GQ644Z8.5 0M8583
0GG53Z0 0M8580 0GG244Z0 0M8580 0GG424Z0 0M8581
0GQ644Z15 0M8583
0GQ54Z0 0M8580 0GQ253Z0 0M8580 0GQ434Z0 0M8580
0GG644Z8.5 0M8583
0GG54Z0 0M8580 0GG253Z0 0M8580 0GG434Z0 0M8580
0GG644Z15 0M8583
0GQ123Z0 0M8581 0GQ254Z0 0M8580 0GQ444Z0 0M8580
0GQ653Z8.5 0M8582
0GG123Z0 0M8581 0GG254Z0 0M8580 0GG444Z0 0M8580
0GQ653Z15 0M8582
0GQ124Z0 0M8581 0GQ283Z0 0M8580 0GQ464Z0 0M8581
0GG653Z8.5 0M8582
0GG124Z0 0M8581 0GG283Z0 0M8580 0GG464Z0 0M8581
0GG653Z15 0M8582
0GQ133Z0 0M8580 0GQ284Z0 0M8580 0GQ474Z0 0M8580
0GQ654Z8.5 0M8583
0GG133Z0 0M8580 0GG284Z0 0M8580 0GG474Z0 0M8580
0GQ654Z15 0M8583
0GQ134Z0 0M8580 0GQ373Z0 0M8581 0GQ484Z0 0M8580
0GG654Z8.5 0M8583
0GG134Z0 0M8580 0GG373Z0 0M8581 0GG484Z0 0M8580
0GG654Z15 0M8583
0GQ233Z0 0M8581 0GQ374Z0 0M8581 0GQ634Z8.5 0M8582
0GQ674Z8.5 0M8582
0GG233Z0 0M8581 0GG374Z0 0M8581 0GQ634Z15 0M8582
0GQ674Z15 0M8582
0GQ234Z0 0M8581 0GQ383Z0 0M8580 0GG634Z8.5 0M8582
0GG674Z8.5 0M8582
0GG234Z0 0M8581 0GG383Z0 0M8580 0GG634Z15 0M8582
0GG674Z15 0M8582 0GQ243Z0 0M8580 0GQ384Z0 0M8580 0GG243Z0 0M8580 0GG384Z0 0M8580
i
Page 100
Gamme UviLine
- 116 -
SECOMAM also sells a large range of cells for which you will find here below the list of our best-sellers. For our full list of available cells, contact us on analytics.info-fr@xyleminc.com
Our Best-sellers :
References
Description (packaging: one cell)
0GQ4Z0 Standard quartz cell with lid 10mm 3,5ml 45*12,5*12,5 0GG4Z0 Standard glass cell with lid 10mm 3,5ml 45*12,5*12,5 0GQ5Z0 Standard quartz cell with lid 20mm 7ml 45*12,5*22,5 0GG5Z0 Standard glass cell with lid 20mm 7ml 45*12,5*22,5 0GQ8Z0 Standard quartz cell with lid 50 mm 17,5ml 45*12,5*52,5 0GG8Z0 Standard glass cell with lid 50 mm 17,5ml 45*12,5*52,5 0GQ14Z0 Standard quartz cell with teflon lid 10mm 3,5ml 45*12,5*12,5 0GG14Z0 Standard glass cell with teflon lid 10mm 3,5ml 45*12,5*12,5 0GQ15Z0 Standard quartz cell with teflon lid 20mm 7ml 45*12,5*22,5 0GG15Z0 Standard glass cell with teflon lid 20mm 7ml 45*12,5*22,5 0GQ18Z0 Standard quartz cell with teflon lid 50 mm 17,5ml 45*12,5*52,5 0GG18Z0 Standard glass cell with teflon lid 50 mm 17,5ml 45*12,5*52,5 0GQ74Z8.5 Glass flowthrough cell 10 mm 0,48ml 45*12,5*12,5 0GQ74Z15 Glass flowthrough cell 10 mm 0,48ml 45*12,5*12,5 0GG74Z8.5 Glass flowthrough cell 10 mm 0,48ml 45*12,5*12,5 0GG74Z15 Glass flowthrough cell 10 mm 0,48ml 45*12,5*12,5 0GQ244Z0 Micro quartz cell with frosted walls and with lid 10 mm 0,7ml 45*12,5*12,5 0GG244Z0 Micro glass cell with frosted walls and with lid 10 mm 0,7ml 45*12,5*12,5 0GQ245Z0 Micro quartz cell with frosted walls and with lid 20 mm 1,4ml 45*12,5*22,5 0GG245Z0 Micro glass cell with frosted walls and with lid 20 mm 1,4ml 45*12,5*22,5 0GQ254Z0 Micro quartz cell with frosted walls and with lid 10 mm 1ml 45*12,5*12,5 0GG254Z0 Micro glass cell with frosted walls and with lid 10 mm 1ml 45*12,5*12,5 0GQ255Z0 Micro quartz cell with frosted walls and with lid 20 mm 2,1ml 45*12,5*22,5 0GG255Z0 Micro glass cell with frosted walls and with lid 20 mm 2,1ml 45*12,5*22,5 0GQ264Z0 Semi micro quartz cell with frosted walls and with lid 10 mm 1,4ml 45*12,5*12,5 0GG264Z0 Semi micro glass cell with frosted walls and with lid 10 mm 1,4ml 45*12,5*12,5 0GQ265Z0 Semi micro quartz cell with frosted walls and with lid 20 mm 2,8ml 45*12,5*22,5 0GG265Z0 Semi micro glass cell with frosted walls and with lid 20 mm 2,8ml 45*12,5*22,5 0GQ268Z0 Semi micro quartz cell with frosted walls and with lid 50 mm 7ml 45*12,5*52,5 0GG268Z0 Semi micro glass cell with frosted walls and with lid 50 mm 7ml 45*12,5*52,5 0GQ653Z8.5 Quartz micro cell with black walls 10 mm 0,05 ml 13*12,5*12,5 0GQ653Z15 Quartz micro cell with black walls 10 mm 0,05 ml 13*12,5*12,5 0GG653Z8.5 Glass micro cell with black walls 10 mm 0,05 ml 13*12,5*12,5 0GG653Z15 Glass micro cell with black walls 10 mm 0,05 ml 13*12,5*12,5
Loading...