Secomam Basic User manual

BASIC
70VB0358 70VB0370
User’s manual
Ref. 0NVBAU
Manual version: 1.19
91, Avenue des Pins d’Alep 30319 ALES FRANCE Tel: +33 4 66 54 35 60 Fax: +33 4 66 54 35 69 E-mail: info@secomam.fr URL: www.secomam.fr
SYSTEME QUALITE CERTIFIE
BASIC
SUMMARY
1 REMARKS......................................................................................................................................................... 5
2 WARRANTY.....................................................................................................................................................7
3 INFORMATION................................................................................................................................................ 9
4 PRECAUTIONS OF USE............................................................................................................................... 11
5 PRELUDE........................................................................................................................................................ 13
6 DESCRIPTION................................................................................................................................................ 15
6.1 GENERAL OVERVIEW 15
6.2 SCREEN AND KEYBOARD 15
6.3 DISPLAY DESIGN 17
7 SET-UP............................................................................................................................................................. 19
7.1 GENERAL SET-UP 19
7.2 LANGUAGE SET-UP 20
7.3 DATE AND CLOCK SET-UP 21
7.4 PUMP SET-UP 21
7.5 PRINTER SET-UP 23
7.6 TRACEABILITY SET-UP 24
7.6.1 Downloading to a PC : F1 « BASIC –PC » 25
7.6.2 Downloading to a printer : F2 « BASIC-PRINTER » 26
7.6.3 KINETICAL REPORT 28
7.7 QUALITY CONTROL SET-UP 28
8 PROGRAMMATION ..................................................................................................................................... 33
8.1 PROGRAM KEYS 33
8.2 RECOMMANDATION FOR DAILY ANALYSIS 41
8.3 ERROR MESSAGES 43
8.4 SAMPLE NUMBERING 45
9 MAINTENANCE............................................................................................................................................. 47
9.1 TECHNICAL CHARACTERISTICS 47
10 MEMO............................................................................................................................................................ 49
11 GLOSSARY................................................................................................................................................... 53
- 3 -
BASIC
1 REMARKS
The Automatic aspiration system kit is manufactured by SECOMAM. The information contained in this document may be the object of patents or patent applications by
SECOMAM. The possession of this document does not confer any license to such patents. The following names are SECOMAM’s registered trademarks throughout the world.
SECOMAM BASIC
This manual is updated periodically. The updates are included in the new editions. All information supplied in this edition of the manual may be amended before the products described
herein are available.
All reproduction rights reserved.
Reference Date of last
modification
Version Date of document. Writer Controller Approving
0NVBAU 07/12/04
1.19
JM
- 5 -
BASIC
2 WARRANTY
The new equipment and material sold by SECOMAM is guaranteed against any manufacturing defects for one year (unless otherwise stated by SECOMAM) with effect:
From the technical acceptance of the equipment in the factory by the buyer or his designee,
or failing this :
* For Metropolitan France: from the date on the delivery note. * For other destinations: from the date of factory shipment certified by air waybill,
consignment note or bill of lading.
The SECOMAM company guarantee applies exclusively to defectiveness arising from a design fault or from a concealed defect. It is strictly limited to the free dispatching of replacement parts (except for consumable items) or to the repairing of the equipment in our workshops within a deadline of 10 working days (shipping delay not included).
By express agreement, the following are strictly excluded from our guarantee:
All damages, notably for staff costs, loss of earnings, business trouble, etc
Any breakdown due to an incorrect use of the equipment (non adapted mains, fall, attempt at
transformation, etc) or to a lack of maintenance by the user or to poor storage conditions.
Any breakdown due to the use of parts not supplied by SECOMAM, on SECOMAM equipment
Any breakdown due to the transporting of the equipment in packaging which is not its original
packaging
The lamps, the cells and generally any item which appears in the "accessories" section on the price list.
Our customers are kindly asked to apply for our consent before returning any instrument for repair. No return of materials may be accepted without the prior written consent of our Sales Management which will precise the terms of such return. If the above consent is given, articles shall be returned in their original packaging on a prepaid basis to the following address:
SECOMAM - 91 avenue des Pins d’Alep – 30100 ALES FRANCE
We reserve the right to reship all instruments received collect failing such consent. Whatever method and conditions of transport are chosen for the shipment of the equipment to be
repaired under guarantee, in the original packaging, the corresponding costs and the insurance costs will be payable by the customer.
Any damage connected to the return transport of the equipment falls within the framework of the guarantee on the express condition that the customer has sent his complaint within forty-eight hours by registered letter with acknowledgement of receipt to the carrier. A copy of the letter should be sent to SECOMAM.
For equipment with a guaranty card, this is only applicable if the card delivered with the equipment is returned to SECOMAM duly completed.
- 7 -
BASIC
SOFTWARE GUARANTEE
The software is guaranteed by the designer or the distributor of the software under the conditions specified in the literature accompanying the aforementioned software packages. Under no circumstances whatsoever will SECOMAM supply any type of guarantee for software packages. By express agreement, all damages, notably for staff costs, lost of earnings; business trouble, etc are strictly excluded from our guarantee. The customer is informed that the software cannot be guaranteed exempt from defects or bugs.
TRADE SECRET AND PROPERTY RIGHTS
This document is protected by a SECOMAM copyright (c) 2003 and the copying rights are explicitly reserved. The software supplied with the equipment or referenced contains trade secrets and confidential information which are SECOMAM's property. It is legally protected by the international copyright (c) laws.
SECOMAM grants a license to use its software to the user. This may not be disclosed, used or duplicated with the intention to save it, without SECOMAM's written permission. The beneficiary must attach a copy of this document to all authorized partial or total reproductions.
- 8 -
BASIC
3 INFORMATION
The SECOMAM equipment has been designed, manufactured, tested and inspected according to the ISO 9001 standards.
If the unit is not immediately installed, it should be stored in a dry and clean area. The storage temperature should be between 10 and 35°C.
SECOMAM equipment is carefully inspected before it is packed. As soon as you receive your equipment, check the condition of the packaging and if you notice any problems, notify your carrier within 48 hours. Then consult the packing list and check that everything is in order. Finally, if you discover that something is missing, or if the goods are damaged immediately notify SECOMAM.
IMPORTANT: In order to benefit from SECOMAM’s service (application notes, SECOMAM information, technical assistance, etc.) immediately complete the attached guarantee card and return it to the following address:
SECOMAM
Service PRODUIT
91 Avenue des Pins d’Alep
30319 ALES Cedex
FRANCE
Tel: +33 4 66 54 35 60
Fax: +33 4 66 54 35 69
E-mail: info@secomam.com
Service Department: +33 4 66 54 35 63
Technical support: +33 1 39 35 42 12
- 9 -
BASIC
4 PRECAUTIONS OF USE
Always make sure that the instrument is connected on the good voltage.
(Between 100 – 240V 50-60Hz)
Always disconnect the mains plug before starting any work inside the
instrument.
When dangerous substances for health and environment are used, the
laboratory or site rules, where the instrument is installed must be followed.
Take all the necessary precautions, during the use the instrument, to protect the operator from eventual liquids leaks or spills or possible radiations (protective gloves, anti radiation glasses , protected clothes, etc)
All operations made inside the instrument, must be done by SECOMAM or by
SECOMAM’s authorized technicians.
- 11 -
BASIC
5 PRELUDE
The BASIC was designed within the framework of In Vitro Diagnostic to cover the need of semi-automatization in clinical biochemistry analysis.
It includes the principal functions required by this kind of analysis such as:
- Wavelength range 340-700 nm,
- Peltier thermostated cell holder
- Calculation method adapted to enzymatic reading of results and parameters
- Quality control management
- Partial management of laboratory practice
BASIC complies with the requirements in CE marking of medical devices for In Vitro Diagnostics (98/79/EEC)
- 13 -
BASIC
6 DESCRIPTION
6.1 GENERAL OVERVIEW
Front view :{
{ XE "Screen" }{ XE "Keyboard" }
6.2 SCREEN AND KEYBOARD
XE "Front view" }
2
1
3
5
4
Program keys +
number keys.
Control keys. Screen. Function keys. ° Cursors.
Function and screen keys
Thanks to these keys, working with BASIC is simplified. Screen instructions as well as the function keys guide you throughout the analysis.
On the display, Function keys correspond to an icon or a value. Simply validate by a light stroke.
:
- 15 -
BASIC
Cursor keys other functions of the keyboard such as :
{ XE "Cursor" }{ XE "Program keys" }
Program keys
secondary position (secondary position are M1 to M47). These program keys are preassigned to give you an idea of the program capacity and calculation method the instrument is capable of. You can modify any given test method at any time, experiment your own parameters, place it at a different key than the original and call it just by one stroke. Before modifying, it is recommended to print the existing test method.
Number keys
(date, hour, sample values, sample numbers, ...). { XE "Number keys" }
: each program key can contain two test methods : on the primary position and the
: otherwise indicated on the display, numbers are to be entered by way of program keys
: Use these keys to have access to
- elevator mouvement : BASIC test methods on the primary and secondary positions, next screen display,
- right arrow for double rinse
- 16 -
BASIC
)
BASIC’s software control keys :
6.3 DISPLAY DESIGN
Programmed
temperature
Validate an operation
Quit active screen
Reach software configuration screen
Erase & correct data
Test name
Calculating method
Concentration value,
measuring unit
Name of sample measured
(standard, sample X ...)
When using cuvette by cuvette method, you will see :
While for pumping aspiration :
Absorbance value
(blank value
deducted
- 17 -
BASIC
7 SET-UP
7.1 GENERAL SET-UP
{ XE "General set-up" } Step-by-step set-up of BASIC is done through the « CONF »key
CLOCK
Use (F1) to (F5) keys to modify
the values. Abort modification
with ESC key, register with VAL
key.
CONF
PRINTER
Access to set-up menu for the printer
STAND-BY
Access to a sub-menu
allowing to choose the
duration before starting the
stand-by (screen
DATA FILES
Access to : -Patient identification –Patient report –Data archiving to PC or Printer – Memory reset.
saver)
- 19 -
BASIC
Access to program
NOTE :
modification
PUMP
.
QUALITY CONTROL
Confirm (F1) or skipping (F4) the
quality control function. Display (F3)
values and dates of previous control.
Access to pump set-
up menu
CONFIGURATION screen is active when you see an elevator in the middle of the screen. Use upward and downward cursor keys to activate and have access to the next display screen.{ XE "Elevator" }
7.2 LANGUAGE SET-UP
This would be the first configuration to check upon starting the instrument. Successively press « CONF », « downward cursor key », (F1) « Language set-up » then (F3) « English ». { XE "Language set-up" }
- 20 -
BASIC
Y
7.3 DATE AND CLOCK SET-UP
D+1
M+1
+1
Use the corresponding function key to increment date and time. Then « VAL » to { XE "Date set-up" }{ XE "clock set-up" }store, « ESC » to return to previous display.
7.4 PUMP SET-UP
{ XE "Pump set-up" }
Access to pump functions & set-up : debit , dead volume, aspiration volume. For your convenience, these volumes are factory preadjusted. (Note : upon changing pump and tubings , you have to readjust setting).
Press (F3) « PUMP SET-UP » , you will obtain :
T+1
M+1
Press (F5) « NO » for manual cuvette by cuvette operation (recommended to use on highly­contaminant tests). By answering (F2) « YES », you obtain three choices :
- 21 -
BASIC
Pump rate (calibration) :
Prepare ahead :
four test tubes with 2000 µl of distilled water each,
one test tube with 1000 µl of distilled water to use as standard.
One graduated pipette (vol. up to 1000 µl)
To adjust pump volume, you need to: { XE "Pump calibration" }
Return to « CONFIGURATION ».
Go to menu « PUMP ».
Use pump « YES ».
Exit from « CONFIGURATION » (ESC twice).
Rince twice consecutively (using right cursor button) to fill the tube with water.
Then :
Again return to « CONFIGURATION »,« PUMP »,« YES ».,« PUMP RATE »
Introduce the first tube containing 2000 µl of distilled water to the aspiration noozle.
Press (F3) « SUCK » and « VALidate » the message of 1000 µl. The pump should have aspirated
1000 µl. Compare the remaining volume with the volume of standard tube. Or better yet, aspirate the remaining distilled water with the pipette and verify the volume. If the remaining volume is more (or less) than 1000 µl, « DELete» the message of 1000 µl on display and enter the real volume sucked by the pump using the number keys (calculation example : initial volume of 2000 µl – remaining volume = new volume to program). Introduce the second tube containing 2000 µl of distilled water.
Press (F3) « SUCK » and « VALidate ». The remaining volume should be 1000 µl, same as the volume of standard tube (or verify again with the pipette). If another adjustment is to be done, repeat this process untill the instrument aspirated the right 1000 µl volume.
Always validate at the end of process by (F6) «OK». (To keep previous volume, press « ESC »).
Delay :
Is the frequency (in seconds) to program for sample introduction following one re ading. You can work faster by shortening pump aspiration delay when you are more familiar with the system.
- 22 -
BASIC
beginning of the dead volume
Dead volume
Sample
Joint
Flow through cell
.{ XE "Volume" } 200 µl is the factory preprogrammed volume:
Dead volume :
Dead volume is needed to avoid contamination of samples. It is represented by a certain volume of air. Dead volume is influenced by the length of aspiration tube. It is recommended to program dead volume in such a way that the end of the aspirated sample remains approximately 2 cm on top of the cell tip (see drawing). { XE "Dead volume" }
Upon selecting F3 Dead Volume, you will get a choice of 6 volumes.
7.5 PRINTER SET-UP
{ XE "Printer set-up" }
Select « PRINTER OPTION » by pressing F3. Printer should be an 80 column or has an 80 column set-up.
When using a printer, select (F2)« YES », (F1) « TYPE » , enter the type of printer. A selection of 4 printers are proposed : a KYOLINE, CITIZEN IDP562S, HEWLETT PACKARD and
EPSON. Choose one. Next, check the type of port « connection » either serial RS232C or Parallel. Connect IDP or KYOLINE printer using its serial
printer cable to BASIC’s serial RS232C port.
- 23 -
BASIC
Warning : never connect a serial printer with a 25-pin cable to the parallel port. This will do serious damage to BASIC and the instrument will emit a continuous beeping noise. This type of error is not covered by the warranty.
Exit from set-up by « ESC ».
7.6 TRACEABILITY SET-UP
{ XE "Trace-ability set-up" }
Data identification, filing, report etc. are given in this (F6) TRACE-ABILITY function : F1 BASIC - PC : this function allows you to download data (patient reports) to the PC (see chapter 5.6.1) F2 BASIC – PRT : use this function to download data to a printer. Two types of printed documents can be obtained :
global daily report of all patients (enter Nr 0) or
patient by patient report (selectable : example Nr 114, 121).
You can print previous day’s result (if it is not crushed/erased yet) by reprogramming the date (see CONFIG, DATE). Note :For two samples bearing the same number, only the latest will be edited. F3 INIT is used to clear result memory. Be sure to save/print results before deleting. F4 YES : by choosing this function, BASIC will report kinetical reading points per segment. F5 NO : by this function, BASIC will give brief kinetical report. F6 ORDER is to select the printing order of a given patient, for example Nr 114 first then Nr 121.
- 24 -
BASIC
7.6.1 Downloading to a PC : F1 « BASIC –PC »
Connect BASIC to the PC using ref : 0X6807 cable and if needed, an adaptator.
{ XE "Download to a PC (Win 95)" }
FOR WINDOWS 95
Click on « START », PROGRAM », « ACCESSORIES », « HYPERTERM », windows 95 opens
a window where you will find «
Open « HYPERTRM.EXE » which will give access to other window « NEW CONNECTION »
and «
DESCRIPTION OF A CONNECTION ».
Choose the name and icone of this connection (bearing in mind the type of instrument
which is your source of data) then «
You’ll see « DESCRIPTION » disappear to leave space to a new windows called « TEL
NUMBER », also « NEW CONNECTION HYPERTERMINAL » will appear as « NAME OF INSTRUMENT
plugged. Validate «
-HYPERTERMINAL ». Select the COM number where the connecting cable is
OK ».
When « PROPERTY COMx » shows up, and you should feed the following informations
:
Bits / second : 9600 Data bit : 8 Parity : none Stop bit : 1 Flux control : none
Then validate « OK ». Check that option « ADD LINE MODIFICATION AT THE END OF ENTERING LINES », in menu
«
FILES », « PROPERTY » index « PARAMETERS » button « CONFIG ASCII » at framed
level « ASCII
RECEPTION » is activated.
To save the parameters, select « FILES » then « SAVE ». Windows register selected
parameters under predefined name.
To receive result file from the instrument, click on « TRANSFER » then « CAPTURE
» menu.
TEXT
HYPERTRM.EXE ».
OK ».
- 25 -
BASIC
Windows opens a dialog box « CAPTURE TEXT ». Choose in which repertory you will
save the data result file also under which name, either by entering the pathway or the name with « .
Type « start », and your terminal will be on reception mode. On BASIC itself, press « BASIC-PC » button, you will see its data on the PC screen.
Open « TERMINAL » program which is usually under program group « ACCESSORIES ». To create acquisition program, do the following :
Select « option. Then select «
Continue by selecting « follows :
{
XE "Download to a PC (Windows 3.x)" }
Then choose COM number where the connecting cable is plugged. To register the parameters, select «
box « name with « .
To receive result file from the instrument click menu «
TEXT FILE
A dialog box « save the data result file, also under which name with « . your terminal will be on reception mode.
On BASIC itself, press « BASIC-PC », data will appear on PC’s screen.
SAVE ». Choose in which repertory you want to save the program, also give a
TXT » extension, or by selecting « FOLLOW » icone.
FOR WINDOWS 3.x
PARAMETERS », « TERMINAL EMULATION » and select « DEC VT 100 (AINSI) »
PARAMETERS », « TERMINAL PARAMETERS », then the following options :
Character terminal : terminal column : 80
CR > CR/LF Entry Exit
PARAMETERS » « COMMUNICATION » and define the page as
transmission rate : 9600 bauds Data bits : 8 bits Stop bits : 1 bit Parity : none Flux control : none Parity control : (non selected) Beared detection : (non selected)
FILE » then « SAVE ». Windows will open a dialog
TRM » then enter « OK ». That file name is shown on PC’s title bar.
TRANSFER » then « RECEIPT A
».
RECEIPT A TEXT FILE » is opened. Choose in which repertory you will
TXT » extension. Press « OK »,
7.6.2 Downloading to a printer : F2 « BASIC-PRINTER »
{ XE "Download to a printer" }
Connect BASIC to a printer. Press « B You will obtain :
ASIC-PRT ».
- 26 -
BASIC
1 2 3 4 5
M 47;E; 0 0.0;%; 29/08/1997;08H17; -;? M 47;E; 1 210.4;%; 29/08/1997;08H19; L;? M 47;E; 2 -0.0;%; 29/08/1997;08H19; ;? M 47;E; 3 -0.0;%; 29/08/1997;08H20; ;? M 47;E; 4 9.9;%; 29/08/1997;08H20; ;? M 47;E; 5 9.9;%; 29/08/1997;08H20; ;? M 47;E; 6 9.8;%; 29/08/1997;08H21; ;? M 47;E; 7 9.8;%; 29/08/1997;08H22; ;?
Column 1 : Sample number + characteristics :
B = Reagent blank, S = Standard, E = Sample, C = Control, b = Blank of reagent blank, s = Standard blank, e = Sample blank, c = control blank.
Column 2 : Sample number. Colonne 3 : Concentration value and unit. Column 4 : Date and time. Column 5 : Status :
- : No problem,
> : Result superior to normal value, < : Result inferior to normal value,
L : result superior to linearity limit,
: negative concentration value B : Blank value superior to linearity limit, Q : Control non performed,
! : Control badly performed,
- : Control normal. Memory initialisation{ XE "Memory initialisation" } To erase the results stored in memory, press « init ».
- 27 -
BASIC
7.6.3 KINETICAL REPORT
On methods using KINETICS, you have a choice of having a short one line final reading report by choosing F5 « NO » or segment per segment reading through F4 « YES ».
7.7 QUALITY CONTROL SET-UP
Each Quality Control runned is memorized automatically. F4« NO » gives you the choice not to run QC. F3« DISPLAY » allows to display on screen as graph the values and dates of previous control (only after at least two controls of the same analysis method have been made). Printout of QC can be obtained after programming TRACE-ABILITY function, YES. { XE "Quality control set-up" }
« NO ».
Upon validating (F1) « YES » , you will be reminded to run Quality Control later on upon running a test method (CHECK ANALYSIS) and flags will be given according to the « MAXIMUM & MINIMUM TOLERANCE VALUES » entered. Q preceding result indicates that QC was not performed.
Note : Upon using a test method for the first time in the day, you will be suggested to run a control. However, once QC is done for the day, no need to run another one, hence F6
- 28 -
BASIC
This control can be done at a regular interval, defined in days :
To do so, return to « PROGRAMMING », F1« MODIFY », select the corresponding analysis method, and validate till you obtain the following display :
{ XE "Control frequency" } Enter the frequency (in days) and validate . This procedure determines QC periodicity. Control can be done daily, for 31 days.
- 29 -
BASIC
This way, daily/periodically, the instrument memorizes analysis results & quality controls.
Status of control
Use upward/downward
Chronological filing of
control
Printing control results
Exemple of print-out (analysis comprising one quality control per day :
SECOMAM - Polyphotometer UV-Vis BASIC V1.11 Date : 04/06/1997 Hour : 09H25
User :
Analysis : ALBUMIN 37°C
Sample Abs. Concentration Status
arrows to visualize
controls .
Reagent Bl. 0.000 Standard 0.100 Control 0.150 OK Control - Le 05/06/97
Normal value: 38.00 à 54.00 g/l
S. # 1 S. # 2
Std. # 3 Std. # 4 Std. # 5
0.150
0.300
0.700
1.300
1.250
0.00 g/l
1.00 g/l
1.50 g/l OK
1.50 g/l
3.00 g/l
7.00 g/l
13.00 g/l
12.50 g/l
< <
< < <
- 30 -
BASIC
SECOMAM - Polyphotometer UV-Vis BASIC V1.11 Date : 05/06/1997 Hour : 09H26
User :
Analysis : ALBUMIN 37°C
Sample Abs. Concentration Status
nd
2
day
Reagent Bl. 0.000 Contrôle 2.100 Control Pb Le 06/06/97
Normal value : 38.00 à 54.00 g/l
S. # 1 S. # 2
0.100
1.600
0.00 g/l
21.00 g/l PB
1.00 g/l
16.00 g/l
< <
SECOMAM - Polyphotometer UV-Vis BASIC V1.11 Date : 06/06/1997 Hour : 09H27
User :
Analysis : ALBUMIN 37°C
Sample Abs. Concentration Status
Reagent Bl. 0.000 Control 0.990
Control 0.990 Control Pb Le 07/06/97
0.00 g/l
9.90 g/l PB
9.90 g/l PB
rd
3
day
Normal value: 38.00 à 54.00 g/l
Status :
> : Sample concentration superior to upper limit. < : Sample concentration inferior to lower limit..
PB : Problem on control. OK : satisfactory control. Q : Appears next to analysis result when control is programmed but not runned. ! : Appears next to analysis result when the control is out of tolerance range. L : The result in concentration is no longer within the linearity limit of the analysis.
− : The result in concentration is negative.
Upon modifying analysis method, memorized results are erased.
- 31 -
BASIC
8 PROGRAMMATION
Through configuration, select « PROGRAMMING » to modify a preset analysis method or to create a new one.
Programmable parameters are :
suction volume,
temperature,
unit
accuracy,
calculation mode,
analysis mode
time (lag time, interval time and number of intervals),
wavelength,
linearity limit,
normal range (high/low)
check frequency,
Note : At any case, arbitrary values are labeled by small arrows on the right or the left side of the value.
{ XE "Program keys" }
8.1 PROGRAM KEYS
To program an analysis test, simply return to set-up procedure : « CONFIGURATION », then « PROGRAM », PROGRAMMING : F1« MODIFY » :
Next you will obtain the « ANALYSIS » selector panel : simply press or or cursor key for access to program keyboard.
previous results & standard control will be crushed upon saving a new method.
Primary position Secondary position
- 33 -
BASIC
Select one of the program key to place the new test method.
ADJUSTMENTS RELATED TO ANALYSIS
Name the analysis method should be done first. It is recommended not to use the same
name as the one found in the primary position. DELete test method name and replace. Choose letters and numbers with F1 to F6 keys and or cursor. Validate at the end.
Aspiration volume :
800 µl is the recommended pump aspiration volume. However a selection of six volumes from 500
to 1500 µl are available.
To modify the volume, press the F key corresponding to the volume to program.
Temperature :
{XE "Temp"} BASIC is equiped with a Peltier-effect temperature control cell-holder. Recommended temperature to use is given by the test method technical sheet. Use F key to program the temperature. Note :during warm-up, the programmed temperature digits will blink, rather fast at first, slowing down afterwards. It takes 15 minutes to stabilize the cell-holder and cuvette to 37°C Once temperature is reached, the digits will be displayed permanently. Peltier-effect is deactivated when programmed at XX (Room temperature/F1) Note on work organization : it is adviced to run a ll test methods using the same temperature in one session to avoid warm-up delay. For your convenience, the cell-holder will remain on the last
Room temp
temperature used.
Unit :
Choose measuring unit according to the recommendation of the reagent technical sheet.
- 34 -
BASIC
{ XE "Unit" }
Result resolution (decimals) : Concentration result is given by the number of decimals
programmed.
{ XE "Result resolution" }
Calculation mode :
{ XE "Calcul mode" } A choice of three calculation modes are proposed :
F1 Factor
Enter factor value indicated on the reagent technical sheet (DELete, then enter numbers). Results will be calculated using this factor.
: in Kinetic mode, enzymatic factor (coefficient) value is used instead of calibration factor.
- 35 -
BASIC
F2 Linear : In End-point, End-point with blank and Two-points-kinetics analysis modes,
calibration is needed (which means, the calculation is done using standard(s)). Enter standard(s) value(s) (DELete, then enter numbers).
F3 Non-linear
: E.I.A tube methods use this calculation mode (polynomial, log/logit, spline, multi-
points). Enter the number of standards used, then the values of each. Up to 6 standards can be used.
Wavelength :
{ XE "Wavelength" }
Wavelength to use is indicated by the reagent technical sheet. You may not find the exact same wavelength among those proposed by BASIC. In that case, select one closest to the ideal wavelength .
{ XE "Analysis mode" }
- 36 -
BASIC
∑∑∑
Analysis mode :
End point without
blank serum
End point with
blank serum
Enzymes kinetics
.F1. End-point analysis mode : When using this mode, reading is done at the final absorbance (optical density) point (at the end of the reaction process). Zero is adjusted using reagent alone (reage nt blank) or distilled water.
F2. End point with A number of environments, in particular biological ones, may be the object of spectral interferences at the wavelength selected for the dosage. This is the case for colored molecules (haemoglobin, bilirubin, ...) These interferences which are independent of the parameter to be measured, will be eliminated by measuring the sample alone or with a reagent which does not trigger a coloring effect. : this is called a sample/serum blank. When there is only one reagent and such a reagent may add a parasitical color to the wavelength of the reading, it is deducted from such reading; this is referred to as reagent blank.
End point with serum blank:
Conc. Std = (abs. Std – Abs. reagent Blank) x coef. Calibr.
End point without serum blank:
Conc. Std. (Abs. Std.- Abs. Serum Blank )- (Abs. Blank - Abs Blank blank)= Coef .
F3 Enzyme kinetics{ method calculates the evolution of a sample, in Absorbance, against a time period defined in a certain number of intervals. BASIC computes the straight regression line which will average the means of the points. BASIC allows you to select the duration of analysis reading as well as the number of intervals to be measured. Also lag time between each intervals should be indicated.
[]
XE "Enzyme kinetics"} : Called also zero order kinetics. This
: {XE "End point with serum blank "}
.{ XE "End point without blank serum" }
slope
N : Total number of point of the kinetic curve (nb of segments x segment time).
ψX : 1+2+3+...+N x (N+1) /2 ψY : total of absorbances at points 1, 2, 3, ... N=A+A2+A3+...+AN ψXY : (A1 x 1)+(A2 x 2)+(A3 x 3)+...+(AN x N)
2
ψX
: 1+4+9+...+N2 = N(N+1)(2N+1) /6
The slope thus calculated is expressed in delta of Abs./min.
F4 Kinetics with standards{ first order kinetics. This analysis method allows to perform measurements of Concentration in Kinetics mode. That is, to calculate the absorbance variation of a solution within the time
and to multiply the differences of Absorbance by a coefficient or to compare to a standard measured with the same measuring conditions. Thus, global waiting time of measurement should be indicated. Then BASIC requires confirmation on the time consumed between the two point reading.
blank:
()( )
Nx XY Xx Y
()()
Nx X X
XE "Kinetics with standards"} : Called also two-point kinetics or
22
Kinetics with
standards
Bichromatism
x=
60
- 37 -
BASIC
ABS ABS x f x
R
=
ABS t1 : Abs of the solution at the end of the lag time. ABS t2 : Abs of the solution at the end of the measuring time. factor : Coefficient of molecular absorption. tm : Measuring time. 60/tm : Ratio reducing the result to one minute.
The slope thus calculated is expressed in delta of Abs./min.
F5 Bichromatism Reading is done using 2 wavelengths multiplied by a factor.
Abs Smpl Abs. Smpl.
.
()
: {XE "Bichromatism"}is recommended when sample is cloudy (turbid).
F
12
λλ
() t t2 actor 160
tm
- 38 -
BASIC
For both KINETIC modes, enter Lag time (the waiting time before reaction starts), number of intervals and time segments of each interval. Indicate also whether it is an increasing/decreasing type of reaction.
5
10
- 39 -
BASIC
Linearity limit :
{ XE "Linearity limit" }
The maximum measuring limit is given in the reagent technical sheet. If result is beyond this limit, it is recommended to dilute the sample or rerun the test. CAUTION : in hyper-active samples having decreasing kinetic-type of reaction, total substrate can be consumed before first measure starts. (Shown by reading result close to zero, below the depletion limit). It is user’s responsibility to interprete reading result taking into account this element.
Normal high : Normal High & Low values border the acceptable pathological range. It is also
called the usual expected values. Flags will be shown on abnormal result.
{ XE "Normal high" }
Normal low:
{ XE "Normal low" }
CHECK FREQUENCY:
At the end of programming, you are asked the frequency of analysis quality control checking : whether daily or periodically. A signal will remind you to do the quality control every day or once every x days. Result of the QC can be found at « CONFIGURATION », « QUALITY CONTROL », and you can ask for « DISPLAY ».
- 40 -
BASIC
8.2 RECOMMANDATION FOR DAILY ANALYSIS
Zero measurement will be asked at the beginning of each day.
SAMPLE ASPIRATION :
Introduce tube to the noozle upon instructed by display. Audio beeper indicates aspiration. Remove tube as soon as aspiration ceased. Resume with next sample simply by introducing the tube. In case of aspiration detector failure,
MODIFICATION OF STANDARD VALUE : Standard concentration value can be modified during analysis. SINGLE STANDARD : use F3 « MODIFY ».{XE "Single standard"} Enter new standard value.
use F1 button to start aspiration.
MULTI-STANDARDS : { XE "Multi-standard" }
Confirm whether standard values to use are as programmed (« OK ») or remeasurement are to be done (« Repeat »).
- 41 -
BASIC
If you confirm the latter, you will see :
Modification of each standard value can be done through the corresponding F button.
If the values on display are correct, validate. Otherwise modify using numerical keys, then validate.
Start with reagent blank reading. This will be deducted automatically from other readings :
Measuring results will be displayed as :
The standards will be called one after another to establish standard curve to calculate sample values.
- 42 -
BASIC
As soon as the standards are read , the curve will be displayed. Absorbance values of standards are also displayed.
By validating F6 « OK », the values are stored in memory.
8.3 ERROR MESSAGES
{ XE "Error messages" }
Sample absorbance value is superior to 2,5 (above the reading limit of BASIC) : in this case the software will show « -,--- ». Check any possible manipulation errors : the sample not prepared correctly or the cleanlinest of the cuvette.
The result in Concentration is superior or inferior to the normal (high or low) value. In this case, a sign « < » or « > » will be displayed following the result. Negative Concentration value will be indicated by the sign .
The result in Concentration is no longer in the linearity limit of the analysis : another sign « < » shows up in front of the result (in this case the result is also beyond the normal high or low range).
Non executed quality control procedure will be signaled by « Q».
During decreasing kinetics method, if the difference of Optical Density between the initial O.D. at T0 and at T1 (end of the latent time) is superior to the depletion limit, the software does not display any result.
- 43 -
BASIC
If the initial O.D. value of the blank is superior or inferior to the limit , a warning sign appears following the concentration result. If the Quality Control value is beyond the specified limit, the sign « ! » appears in front of the Concentration result. In case of Peltier failure Tsign appears instead of the temperature programmed.
PATIENT REPORT & PRINT-OUT
PATIENT REPORT : Daily analysis are memorized automatically by BASIC and indicated at the bottom of the result display as a blinking line. User is recommended to empty the memory at the end his working session. This can be done by : transferring all data to print ( see chapter 5.6.2.) or downloading all data to file on a PC ( see chapter 5.6.1.) or selecting particular data to print (see chapters 5.6.2 & 5.6.3.). In this case, remember to use patient number as selection criteria. Note : the latest of sample bearing identical number will be reported.
{XE "Print out"} When quality control is programmed, error messages will be given on « status » column.
SECOMAM - Polyphotometer UV-Vis BASIC V1.11 Date : 04/06/1997 Hour : 09H24
User :
Analysis : ALBUMIN 37°C
Regeant Bl. 0.000 0.00 g/l Standard 1.000 1.00 g/l Control 0.900 0.90 g/l
Sample
Abs.
Concentration
Status
PB
- 44 -
BASIC
SECOMAM - Polyphotometer UV-Vis BASIC V1.11 Date : 04/06/1997 Hour : 09H24
User :
Analysis : ALBUMIN 37°C
Reagent Bl. 0.000 0.00 g/l Std 1.000 1.00 g/l Control 0.600 0.60 g/l Control 0.990 0.99 g/l Control 0.990 0.99 g/l S. # 0.500 0.50 g/l S. # 0.500 0.50 g/l S. # 1.200 1.20 g/l Std. # 2.000 2.00 g/l
When everything goes well OK will be indicated:
SECOMAM - Polyphotometer UV-Vis BASIC V1.11 Date : 04/06/1997 Hour : 09H24
User :
Analysis : ALBUMIN 37°C
Bl. Reagent 0.000 0.00 g/l Sample 1.000 1.00 g/l Control 1.500 1.50 g/l
Sample
Standard
Abs.
Abs.
Concentration
Concentration
Status
Status
PB PB PB
< < < <
OK
8.4 SAMPLE NUMBERING
Sample numbers will be incremented automatically. However you can assign another number through F2 key. Example : Sample # 3 is to be modified. Press F2, and you will see Sample Nr.1 on the screen. Press Delete and replace by the number you want using number keys.
- 45 -
BASIC
9 MAINTENANCE
9.1 TECHNICAL CHARACTERISTICS
{ XE "Technical characteristics" }
PARAMETERS IN MEMORY 94 READING MODE SIMPLE ABSORBANCE END POINT WITH SERUM BLANK END POINT WITHOUT SERUM BLANK REAL KINETICS KINETIC TWO POINTS BICHROMATISM CALIBRATION LINEAR / NON LINEAR 6 STANDARDS LIGHT SOURCE HALOGEN LAMP MONOCHROMATOR HOLOGRAPHIC GRATING DETECTOR SILICIUM DIODE ARRAY SAMPLE COMPARTMENT DRY TEMPERATURE PELTIER SYSTEM OPERATING TEMPERATURE AMBIENT - 25 - 30 - 37°C TYPE OF CUVETTE FLOW-THROUGH µCUVETTE 30 µl
10 mm OPTICAL PATH ASPIRATION VOLUME PROGRAMMABLE, 500 to 1500µl QUALITY CONTROL MANAGEMENT OF QC RESULTS/MONTHS POWER REQUIREMENTS 110 / 220V INTERFACE SERIAL RS232C AND PARALLEL CERTIFICATION AND
APPROVALS LANGUAGES MULTILINGUAL
ISO 9001 AFAQ N°1995/4102 - CE - EMC
- 47 -
BASIC
10 MEMO
BASIC
- 49 -
Note down the analysis names you have modified.
SUBSTRATES ENZYMES
M1
..................................
M4
..................................
M7
..................................
M10
..................................
M13
..................................
M16
..................................
M19
..................................
M22
..................................
M2
.....................................
M5
.....................................
M8
.....................................
M11
.....................................
M14
.....................................
M17
.....................................
M20
.....................................
M3
.....................................
M6
.....................................
M9
.....................................
M12
.....................................
M15
.....................................
M18
.....................................
M21
.....................................
M23
..................................
M25
..................................
M27
..................................
M29
..................................
M31
..................................
M33
..................................
M24
.....................................
M26
.....................................
M28
.....................................
M30
.....................................
M32
.....................................
M34
.....................................
Note down the analysis names you have modified.
SUBSTRATES ENZYMES
Albumin
..................................
Anti Thrombin
..................................
Calcium
..................................
HDL Choles.
..................................
Glucose
..................................
RBC Magnesium
..................................
T. Protein
..................................
Uric Acid
..................................
Alcohol
.....................................
Bicarbonate
.....................................
Chloride
.....................................
LDL Choles.
.....................................
Iron
.....................................
Phos. Lipids
.....................................
Triglycerides
.....................................
Ammonia
.....................................
Bilirubin
.....................................
Cholesterol
.....................................
Creatinine
.....................................
Magnesium
.....................................
Phosphorus
.....................................
Urea
.....................................
Acid Phosphat.
..................................
Amylase
..................................
Apo-B
..................................
Alat / GPT
..................................
Gamma GT
..................................
Lipase
..................................
Alk. Phosphat.
.....................................
Apo-A1
.....................................
Asat / GOT
.....................................
Ck. NAC
.....................................
LDH
.....................................
Nucleotidase
.....................................
BASIC
11 GLOSSARY
A
Analysis mode 30
M
Memory initialisation 21 Multi-standard 35
B
Bichromatism 32
C
Calcul mode 29 clock set-up 15 Control frequency 23 Cursor 10
D
Date set-up 15 Dead volume 17 Download to a PC (Win 95) 19 Download to a PC (Windows 3.x) 20 Download to a printer 21
N
Normal high 34 Normal low 34 Number keys 10
P
Print out 38 Printer set-up 17 Printer set-up 17 Program keys 10, 27 Pump calibration 16 Pump set-up 15
Q
Quality control set-up 22
E
Elevator 14 End point with serum blank 31 End point without blank serum 31 Enzyme kinetics 31 Error messages 37
F
Front view 9
R
Result resolution 29
S
Screen 9 Single standard 35
T
G
General set-up 13
Technical characteristics 41 Temp 28 Trace-ability set-up 18
K
Keyboard 9 Kinetics with standards 32
L
Language set-up 14 Linearity limit 34
U
Unit 29
V
Volume 17
W
Wavelength 30
- 53 -
Loading...