Seametrics FT415 Instructions Manual

FT415 Flow Computer
ARCHIVED
Instructions
SpecificationsGeneral Information
The FT415 is a microcontroller-based rate/total indica­tor powered by a lithium battery. It displays rate and total, and provides a programmable pulse output for use in metering pump pacing or datalogging.
The rugged cast-aluminum housing is gasketed for maxi­mum environmental protection, and the electronics are potted into a solid block of urethane. A membrane key­pad allows settings to be changed without removing the cover. The wall mount version comes with brackets for wall mounting, and a kit (MK10) is available for field in­stallation on TX80-series flow sensors. The FT415 can also be factory or field installed on a TX100/200 series sensor or WT meter. (Micropower sensors are required to prolong battery life.)
The standard output is a programmable 0.1 second open­collector transistor (TTL) pulse which can be used to trigger most externally-paced metering pumps.
Features
Sturdy gasketed cover
with electronics
Display
Rate 6-digit autorange, 1/2"
character height
Total 8-digit, 5/16" character
height
Pulse Output 0.1 second open collector
pulse
Input Micro-power GMR sensor
(square wave)
Input Range Minimum: 20 mV peak-peak
Maximum: 6V peak-peak
K factor Range 0.050 - 999999.9 Pulse Output Range 0.1 - 200,000 units/pulse Battery Type Lithium "C", 3V, replaceable Normal Battery Life 3-5 years Temperature 0 C - 70 C Environmental NEMA 4X
Tested for Immunity
and Emissions standards
Removeable foot brackets
for wall mounting
2341678.9
SET K RESET
SET
Easy set-up
prompts
Large rate numerals
10.5
Triple bosses for
watertight input/output
connections
LT-10621-B Page 1 of 4
Installation
!
ARCHIVED
To mount the FT415W on a wall, hold the unit in the desired position, mark the holes in the mounting feet, drill, and mount with screws.
This unit can also be mounted directly on a flow sensor for local reading. Note that the FT415 requires the optional micropower sensor, in order to conserve bat­tery life. For loop or other external power, the FT420 is recommended, and can be installed on any standard sensor.
If the unit was ordered as an FT415M, factory mounted to the top of a 100 or 200 Series flow sensor, no further installation is required.
To mount the unit on a flow sensor, an MK10 mounting kit is required. Then remove the housing cover. Place the lower half of the housing in the groove on the top of the flow sensor. An adapter from the kit is included in the lower housing, which threads into the top of any 80­Series flow sensor. Pull the sensor wires through the hole in the center of this adapter, then thread the adapter into place. Tighten the adapter using a screwdriver. (See diagram). Connect the wires, then replace the housing cover, orienting it any direction as desired for viewing convenience.
Connection
To connect to the FT415 to a flow sensor, follow the diagram marked “Connections”. This diagram also shows the connections for an external device, such as a chemical metering pump.
Programming
Caution: If pulse output is being
used to control an external device, such as a metering pump, do not connect the pulsed device until pro-
gramming is completed. If malfunc­tion or incorrect programming of the pulse output could cause personal injury or property damage, separate safeguards must be installed to prevent such injury or damage.
At a minimum, every FT415 must be programmed with the “K-factor”. This is simply the number of pulses per gallon which the meter or flow sensor produces. (If you wish the FT415 to read in another unit, enter the num­ber of pulses per that unit. For example, if the desired units are liters, the K-factor is the number of pulses per liter that the flow sensor produces.) To find the K-factor on any SeaMetrics flow sensor fit­ting or in-line meter, look on the model-serial label. The line reading K = xxxx is the desired number. For adjust­able sensors (101,201,115,215) look in the instruction manual under your pipe size.
Set K. Begin by pressing the key SET once. The prompt SET K should appear on the display. The digit to the far right will be blinking. Use the up arrow key to reach your desired digit. Then press the left arrow key to move to the next digit. Repeat the process until the entire num­ber is entered. (Note that the decimal is fixed at three places. If you only have two decimal places for your K­factor, enter a zero for the third digit.)
Chemical
FT415
on TX81
Flow
2 of 4
Control Pulses
Proportional
Chemical
Set P. If the pulse output is to be used, set the desired pulse rate. Note: If the pulse output feature is not being used, this step can be skipped. The P (pulse output) setting does not affect anything if it is not being used. This is the number of gallons (or other
units, if you are using them) between pulses. The sec­ond time the SET key is pressed, SET P will appear. Follow the same process as above to enter the desired number.
Pressing all three buttons at once will put the unit into low alarm. Pressing all three again will put the unit into high alarm.
Press the ARROW LEFT keys to set the flow rate at which you want to have the output pulse switch on.
and ARROW UP
Sensor Mount: Step 1
ARCHIVED
Sensor Wires
Sensor Mount: Step 2
Sensor Wires
Standard Connections
Red White Black
Micropower Sensor Only
Battery Type:
Lithium "C, 3V, replaceable
+
Sensor
Input
s
-
- +
Outputs
- +
Sensor Pulse
To Other Control
Pulse Responsive
Metering Pump
3 of 4
Set Decimal Point. The third time the SET key is
ARCHIVED
pressed, a D prompt appears. Pressing the up arrow key switches back and forth between no decimal place and one decimal place.
Set Time Unit. The fourth time the SET key is pressed, a blinking arrow appears. Press the up arrow key to select S (seconds), M (minutes), H (hours) or D (days).
Operation
To return to normal operation after entering settings, press SET for a fourth time. The rate and total indicator numbers should appear in the display. If the unit is con­nected to an operating flow sensor, the larger-sized digits indicate flow rate and the smaller-sized digits in­dicate total.
Unless the unit has been ordered with the non-reset option, a RESET prompt is constantly visible in the lower right corner, above the up arrow key. Press this key at any time to reset the totalizer to zero.
Battery Change
The expected average life of the battery is approximately four years, with a range of 3-5 depending on the fre­quency of the input. Pull the battery from its holder and replace it with the new one. Note that when the battery is removed, all of the settings previously entered will be retained. However, the totalizer will reset to a previ­ous total, which represents the last auto-backup (ap­proximately one hour intervals.) If it is necessary to save the exact current total at the time of the battery change, save before removing the battery, by simultaneously pressing the SET and up arrow keys. Then press SET again, then again simultaneously press the SET and up arrow keys.
Dimensions
2341678.9
SET
3.93"
4.52"
10.5
with Wall Mount Brackets
FT415
3.93"
2.06"
FT415
Sensor-Mount
2.57"
2.78"
4 of 4
20419 80th Ave. So., Kent WA. 98032 USA
Phone: 253-872-0284 Fax: 253-872-0285
www.seametrics.com 1-800-975-8153
Loading...