LOW-FLOW MAGNETIC FLOW METER INSTRUCTIONS
EM 100/102 - 025 EM 100/102 - 038 EM 100/102 - 050
The EM100 is a 24 VDC-powered electromagnetic flow meter with chemically-resistant plastic wetted parts, in nominal maximum flow rates of 4, 11, and 30 liters per minute. It uses bipolar DC technology at a frequency high enough to allow for measurement of pulsating flows from diaphragm­type chemical metering pumps. All electronics are contained in a single compact, chemical-resistant housing. This configuration, combined with the standard pulse and analog outputs, makes the EM100 well suited for OEM applications. Other applications include remote monitoring of chemical volume or rate using a computer or PLC.
EM100/102 LOW FLOW
MAGNETIC FLOW METER
INSTRUCTIONS
SPECIFICATIONS
MATERIALS
Wetted Materials -
Housing -● ABS
POWER
ACCURACY
MAXIMUM FLUID TEMPERATURE
MAXIMUM PRESSURE
MINIMUM CONDUCTIVITY
OUTPUTS
Body: Kynar™ PVDF
Electrodes: Platinum over titanium
Internal Seals: Viton™
12-24 VDC, 180 mA
+/-1% of reading, above 10% of full scale
+/-3% of reading, below 10% of full scale
185º F
150 PSI (@ 75º F)
20 microsiemens
Frequency opto isolated 1000 pulse/liter (3785 p/g)
Optional 4-20 mA opto isolated passive
FLOW RANGE
SIZE NOMINAL FULL SCALE MIN. FOR 1% ACCURACY LOW FLOW CUTOFF
L/MIN. GAL./MIN. GAL./HR. L/MIN. GAL./MIN. GAL./HR. L/MIN. GAL./MIN. GAL./HR
-025 3.8 1 60 .38 0.1 6 .04 .009 .54
-038 11 3 180 1.1 .3 18 0.1 .025 1.5
-050 30 8 480 3.0 .8 48 .25 .065 3.9
1- 800-975-8153
INSTALLATION
The EM100 can be supported by its piping connections if the piping is rigid and there is no vibration. Alternatively, the grounding studs can be used for mounting by inserting them through pre-drilled holes. Note the centerline distance in the dimensional drawing below.
MOUNTING
CAUTION: Although this meter has
an empty pipe detection function, under certain conditions of empty or partially-full pipe the meter may read a flow when there is none.
If this is a hazardous condition, mount the meter in such a way as to ensure the meter will always be full of liquid.
Table 1: Fittings and Length
Size
-025
-038
-050
Fittings
3/8 NPT
3/8 NPT
1/2 NPT
L (Length)
3.75
3.75
4.90
MAGNETIC FLOW METER INSTRUCTIONS
L
MALE PIPE FITTING, PVDF PLASTIC, 2 PLACES; SEE TABLE 1
EM 100/102 SERIES
ABS HOUSING
3.75
2.00
6 FT. LONG CONTROL CABLE, 6 CONDUCTOR
.83
www.seametrics.com
#10-24 STUDS (2), 316
.60
12 FT. LONG GROUNDING WIRE WITH LUGS (SUPPLIED)
5.70
2.48
.62
2.000
2 EACH #10 LOCKWASHERS & #10-24 HEX NUTS (SUPPLIED)