Sealey MM18 Instruction Manual

1.1. PERSONAL PRECAUTIONS
When using this multimeter, please observe all normal safety rules concerning: Protection against the dangers of electrical current. Protection of the meter against misuse. Full compliance with safety standards can only be guaranteed if used with the test leads supplied. If necessary, they must be replaced with genuine Sealey leads with the same electrical ratings. Failure to do so will invalidate the warranty.
DO NOT use leads if damaged or if the wires are bared in any way.
Familiarise yourself with the application and limitations of the multimeter as well as the potential hazards.
IF IN ANY DOUBT CONSULT A QUALIFIED ELECTRICIAN.
USE EXTREME CAUTION when working with high voltages.  DO NOT use the meter if it is damaged. Before you use it inspect the case. Pay particular attention to the insulation
surrounding the connectors. Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. Replace damaged test leads before you use the meter.  DO NOT use the meter if it operates abnormally. Protection may be impaired. When in doubt have the meter inspected by your Sealey Authorised Dealer.
 DO NOT operate the meter where explosive gas, vapour or dust is present.  DO NOT apply more than the rated voltage, as marked on the meter, between the probes or between any probe and earth
ground.
 Before use verify the meter's operation by measuring a known voltage.  When measuring current, turn off circuit power before connecting the meter in the circuit. Remember to place the meter in
series with the circuit.  WARNING Use caution when working with voltage above 30V ac rms, 42V peak, or 60V dc. Such voltages pose a shock hazard.
 When using the probes keep your fingers behind the finger guards.  When making connections, connect the black test lead before you connect the red test lead. When you disconnect the test
leads, disconnect the red one first.
 Remove the test leads from the circuit under test before you open the back cover.  DO NOT operate the meter with the back cover removed or loosened.  To avoid false readings, which could lead to possible electric shock or personal injury, replace the button cells as soon as
the low battery indicator ( ) appears.  DO NOT touch any naked conductor with your hand or skin, to avoid electric shock, DO NOT ground yourself whilst using the meter.
 When a probe is connected to a dangerous live potential, it must be noted that this potential can occur at the other probe. NOTE CAT II measurement category is for measurements performed on circuits directly connected to low voltage
installations. (Examples are measurements on household appliances, portable tools and similar equipment.) DO NOT use the meter for measurements within Measurement Categories III and IV.
 Disconnect circuit power and discharge all capacitors before testing resistance, diode or continuity.  Use the proper function for your measurements.  Before measuring current, check the meter's fuse and turn off power to the circuit before connecting the meter to the
circuit.
 Before moving the function switch or rotating the range switch, disconnect the test leads from the circuit under test. The user shall ensure that test probes are correctly selected in order to prevent danger. Probes shall be selected
to ensure that adequate barriers guard against inadvertent hand contact with live conductors under test and that probes have minimal exposed probe tips. Where there is a risk of the probe tip short circuiting with other live conductors under test, it is recommended that the exposed tip length shall not exceed 4mm. NOTE:The warnings, cautions and instructions referred to in this manual cannot cover all possible conditions and situations that may occur. It must be understood that common sense and caution are factors which cannot be built into this product, but must be applied by the operator.
INSTRUCTIONS FOR:
POCKET MULTIMETER
Model No: MM18
Thank you for purchasing a Sealey product. Manufactured to a high standard this product will, if used according to these instructions and properly maintained, give you years of trouble free performance.
1. SAFETY INSTRUCTIONS
IMPORTANT:
PLEASE READ THESE INSTRUCTIONS CAREFULLY. NOTE THE SAFE OPERATIONAL REQUIREMENTS, WARNINGS & CAUTIONS. USE THE PRODUCT CORRECTLY AND WITH CARE FOR THE PURPOSE FOR WHICH IT IS INTENDED. FAILURE TO DO SO MAY CAUSE DAMAGE AND/OR PERSONAL INJURY AND WILL INVALIDATE THE WARRANTY. PLEASE KEEP THESE INSTRUCTIONS SAFE FOR FUTURE USE.
Original Language Version
MM18 Issue No:2(I)-09/09/14
© Jack Sealey Limited
1.3. ELECTRICAL SYMBOLS
Alternating Current.
Direct Current.
Both direct and alternating current.
Caution risk of danger refer to operating manual before use.
Caution risk of electric shock.
Earth (ground) terminal.
Fuse.
The equipment is protected throughout by double insulation or reinforced insulation.
Handy compact unit, small enough to fit in your pocket, making it ideal for basic circuit testing on the move. Features flip lid for added screen protection and storage of leads. Conforms to EN 61010-1 CAT II safety requirements. Supplied with test leads.
Measures:
• AC and DC Voltage
• DC Current
• Resistance
• Audible Continuity
2. FEATURES
© Jack Sealey Limited
3. SPECIFICATION
Model No: ......................... MM18
Measures:
Voltage DC: ......................2V, 20V, 200V, 250V (±1.0%)
Voltage AC: ....................... 250V (±1.2%)
Current DC: ...................... 200mA (±1.2%)
Resistance: ....................... 2k Ω, 20k Ω, 200k Ω, 2000k Ω (±1.2%)
Continuity Audible: ............ Ye s
Digits x Height: ................ 5 x 11mm
Batteries (Supplied): ......... 2 x 1.5V
Size (L x W x H): ............... 114 x 56 x 23mm
Weight: ............................. 101g
Conformity: ....................... EN 61010-1
Operation Environment:.... 0°C to 40°C
Relative Humidity:............. <75%
Storage Temperature: ....... -10°C to 50°C
Relative Humidity:............. <85%
1. Display 3½ digit LCD, with a max. reading of 1999.
2. Function Used to select desired function.
3. Range Switch Used to select desired range.
4. Test Probes
DC VOLTAGE
Range Resolution Accuracy
2V 0.001V
20V 0.01V
200V 0.1V ±(1.0% +5 digits)
250V 1V
Input Impedance: 1MΩ Max. Allowable Input Voltage: 250V dc
AC VOLTAGE
Range Resolution Accuracy
200V 0.1V
250V 1V ±(1.2% +10 digits)
Input Impedance: About 500kΩ Frequency Range: 40Hz to 400Hz Max. Allowable Input Voltage: 250V ac rms
Response: Average, calibrated in rms of sine wave
DC CURRENT
Range Resolution Accuracy
200mA 0.1mA ±(1.2% + 5 digits)
Overload Protection: 250mA/250V Fast fuse
RESISTANCE
Range Resolution Accuracy
2kΩ 0.001kΩ
20kΩ 0.01kΩ
200kΩ 0.1kΩ ±(1.2% +5 digits)
2000kΩ 1kΩ
Max Open Circuit Voltage: About 0.4V Overload Protection: 250V DC/AC rms
Original Language Version
MM18 Issue No:2(I)-09/09/14