Scilogex MX-RD-PRO, MX-RL-PRO User Manual

MX-RD-PRO/MX-RL-PRO
MX-RD-Pro LCD digital Rotator MX-RL-Pro LCD digital Rotator
User Manual
12300183
Please read follow all operating and safety instructions!
www.scilogex.com
the User Manual carefully before use, and
Contents
Preface ................................................................................. 1
Service ................................................................................. 1
Warranty .............................................................................. 1
1. Safety Instructions ....................................................... 2
2. Proper Use ................................................................... 3
3. Inspection .................................................................... 3
3.1 Receiving Inspection ............................................. 3
3.2 Listing of Items ..................................................... 3
4. Function ...................................................................... 4
4.1 Control .................................................................. 4
4.2 Display .................................................................. 5
5. Operation ..................................................................... 6
6. Faults ........................................................................... 6
7. Maintenance and Cleaning .......................................... 7
8. Associated standards and regulations .......................... 7
9. Specifications .............................................................. 8
10. Ordering Informations ................................................. 8
Rotator
Preface
Welcome to the Rotator User Manual. Users should read this manual carefully, follow the instructions and procedures, and beware of all the cautions when using this instrument.
Service
Please provide the customer care representative with the following information:
Serial number ( on the rear panel )
Certification
Description of problem ( i.e., hardware or software )
Methods and procedures adopted to resolve the
problems
Your contact information
When help needed, you can always contact the service department of manufacturer for technical support in the following ways:
SCILOGEX, LLC
500 Four Rod Road Suite 122 Berlin, CT 06037 USA Tel: 1-860-828-5289 Fax: 1-860-828-5389 E-mail: info@scilogex.com Http://www.scilogex.com
Warranty
This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a period of 24 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts which have been damaged on account of improper installation, improper connections, misuse, accident or abnormal conditions of operation. For claims under the warranty please contact your local supplier. You may also send the instrument directly to manufacturer, enclosing the invoice copy and by giving reasons for the claim.
1
Rotator
1. Safety Instructions
Warning!
Read the operating instructions carefully
before use.
Ensure that only trained staff works with the
instrument.
Protective ground contact
Ensure the socket must be grounded
(protective ground contact) before use.
When working wear personal safety guards to avoid
the risk of:
Splashing liquids Broken glass containers
Follow the safety instructions, guidelines and accident
prevention regulations.
Do not touch the running parts, moving instrument
care not rolling your fingers.
Set up the instrument in a spacious area on a stable,
clean, non-slip, dry and fireproof surface. Do not operate the instrument in explosive atmospheres, with hazardous substances or under water.
If the instrument does not run smoothly, please
decrease the motor speed.
Firmly secure the accessories and vessels in place to
avoid damage or risk.
Preparation of samples may lead to dangerous
flammable. Only process samples that will not react dangerous.
Use the standard accessories listed in the “accessories”
section, and follow the instructions to use accessories to ensure safety. Pleaseswitch off the power before assembly of accessories, confirm the instrument and accessories are intact before switch on each time.
The instrument only be opened by expert, please
switch off before use.
The voltage stated on the nameplate must correspond
to the mains voltage.
Do not cover the instrument during running. Prevent
the collision and extrusion to instrument and accessories.
Keep away from high magnetic field.
2
Rotator
2. Proper Use
The instrument is designed for mixing sticky substance in schools, laboratories or factories. It can be installed on a variety of impeller, for different viscosity of the medium. This instrument is not suitable for using in residential areas or other constraints mentioned in Chapter 1. Do not use the accessories recommended by the manufacturer, or failure to use the instructions, may be caused unsafe situation.
3. Inspection
3.1 Receiving Inspection
Unpack the instrument carefully and check for any damages which may have arisen during transport. If it happens, please contact manufacturer/supplier for technical support.
Note:
If there is any apparent damage to the system, please do not connect to the power line.
3.2 Listing of Items
Items
Main unit 1 Power cable 1 User manual 1
T able 1
Qty
3
4. Function
4.1 Control
Rotator
Power switch
Angle control knob
Figure 1
Figure 2
Figure 3
4
Rotator
Items Default settings
Mode switch button
Mode
Speed /timer control knob
Speed/timer light
Speed/Timer
Power/Overload protection light
Power/Overload
Shift speed and time display.
Set speed and time by rotating speed/timer control knob. Push on/off to start/stop rotating function.
Green/Red LED display light. Different colors of LED light show the value for speed or time that LCD screen currently displays. Green LED light shows that LCD screen currently displays speed. Red LED light shows that LCD screen currently displays time.
Green/Red LED display light. LED light shows green when switch
on, LED light shows red when starts overload protection. When the torque reaches limited value, overload protection function will be started. At the same time overload protection light flashes, while the
system stops running.
LCD display
Power switch I/O
LCD displays the real working state and all setting values.
Switch ON or OFF the instrument.
Table 2
4.2 Display
Figure 4
Display Descriptions
Set Display when set target value. Err Display in case of error happening.
Display area When Set displays, this area shows
setting value; When Set disappears, this area shows running value.
Table 3
5
Rotator
5. Operation
Place the Rotator in safe and stable surface, ensure the
required operating voltage and power supply voltage matched, and the socket must be earthed reliably.
Connect the power cable.
Switch ON instrument.
The instrument begins self-checking.
When initialization is over, LCD displays the last
running values of speed.
Rotate the speed/timer control knob to set targe t spee d.
a. Press the speed/timer control knob to start rotating function. Press the knob again to stop rotating function. b. Press the mode switch button, LCD displays time, and then rotating the speed/timer control knob to set target time. Press the speed/timer control knob to start rotating function at the set time. During operation, motion can be stopped at any time by pressing the speed/timer control knob. If the knob is pressed again, motion will start again and the timer will restart countdown. When the timer reaches zero, the unit will
be automatically halted and an alert will sound. If these operations above are normal, the instrument is ready to operate. If not, the instrument may be damaged during transportation, please contact technical support of manufacturer/supplier.
Note:
The set values can be altered at any time.
6. Faults
Instrument cannot be powered ON when start rotating
function.
Check whether the power cable is connected.
Rotating function suddenly stop
Power/overload protection light changed to red,
LCD displays “Er 03”, indicate the current failure is “overload protection”.
Pressing the speed/timer control knob first to
close rotating function, and then restart rotating function after lowered load. If rotating function cannot start, then repeat the process and gradually reduce load.
6
Rotator
7. Maintenance and Cleaning
Proper maintenance can keep instrument working
properly and lengthen its lifetime.
Do not spray cleanser into the instrument when
cleaning.
Do not remove the power line when cleaning. Only
use recommended cleansers: Dyes Isopropyl alcohol Construction
materials Cosmetics Water containing tenside /
Foodstuffs Water containing tenside Fuels Water containing tenside
Before using other method for cleaning or
decontamination, the user must ascertain with the manufacturer that this method will not damage the instrument. Wear the proper protective gloves during cleaning of the instrument.
Water containing tenside / Isopropyl alcohol
Isopropyl alcohol
Note:
Electronic device cannot clean with cleanser.
If you require maintenance service, must be
cleaned the instrument in advance to avoid pollution of hazardous substances, and to send back into original packing.
If the instrument will not use for a long time,
please switch off and place in a dry, clean, room temperature and stable location.
8. Associated standards and regulations
Construction in accordance with the following safety standards: EN 61010-1
UL 3101-1 CAN/CSA C22.2(1010-1) EN 61010-2-10
Construction in accordance with the following EMC standards: EN 61326-1
7
Rotator
Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
9. Specifications
Items Specifications
Voltage [VAC] 100~240
Items Specifications
Angle range 0°~ 90° Motor type DC motor Speed range [rpm] 10~70 Speed display LCD Timer range [min] 1~1199 Timer display LCD
Run type
Overall Dimensions [mm] 512×190×250 Weight [kg] 6.5
Permissible ambient temperature [°C]
Permissible relative humidity 80% Protection class acc. to DIN
EN60529
Timer/Continuous operation
540
IP21
Table 4 Frequency [Hz] 50/60 Power [W] 40
8
Rotator
10. Ordering Informations
Cat. No. Descriptions
MX-RD-Pro, LCD Digital Rotator, used
824232019999
824232119999
824232219999
824232319999
with disk accessories, variable speed, adjustable mixing angle, US plug, 100V-240V, 50Hz/60Hz
MX-RD-Pro, LCD Digital Rotator, used with disk accessories, variable speed, adjustable mixing angle, Cn plug, 100V-240V, 50Hz/60Hz MX-RD-Pro, LCD Digital Rotator, used with disk accessories, variable speed, adjustable mixing angle, Euro plug, 100V-240V, 50Hz/60Hz
MX-RD-Pro, LCD Digital Rotator, used with disk accessories, variable speed, adjustable mixing angle, UK plug, 100V-240V, 50Hz/60Hz
824222019999
824222119999
824222219999
824222319999
Accessories
MX-RL-Pro, LCD Digital Rotator, used with rotisserie accessories, variable speed,adjustable mixing angle, US plug, 100V-240V, 50Hz/60Hz
MX-RL-Pro, LCD Digital Rotator, used with rotisserie accessories, variable speed,adjustable mixing angle, Cn plug, 100V-240V, 50Hz/60Hz MX-RL-Pro, LCD Digital Rotator, used with rotisserie accessories, variable speedadjustable mixing angle, Euro plug, 100V-240V, 50Hz/60Hz
MX-RL-Pro, LCD Digital Rotator, used with rotisserie accessories, variable speed,adjustable mixing angle, UK plug, 100V-240V, 50Hz/60Hz
Cat. No Descriptions
18900137
Disk, 1.5ml centrifuge tube, capacity: 60ea
9
Rotator
Cat. No Descriptions
18900138
18900139
18900140
Disk, 15ml centrifuge tube, capacity: 16ea
Disk, 50ml centrifuge tube, capacity: 8ea
Post for installing disk together
1.5ml Centrifuge tube,
18900142
capacity: 48ea, horizontal mount
15ml Centrifuge tube,
18900143
capacity: 24ea, horizontal mount
50ml Centrifuge tube,
18900144
capacity: 24ea, horizontal mount
Cat. No Descriptions
18900145
1.5ml Centrifuge tube, capacity: 32ea, vertical mount
15ml Centrifuge tube,
18900146
capacity: 16ea, vertical mount
50ml Centrifuge tube,
18
900147
capacity: 16ea, vertical mount
Table 5
10
Rotator
Inst
alling rotisserie attachment
Changing attachment
11