APPLICATIONS
The 3-inch Platform accepts single or multiple test tubes, beakers and various flasks. The removable cover, made of thermoplastic rubber, is highly impervious to chemicals. It is, however recommended that spills are blotted quickly. Do not attempt to use the 3-inch Platform without the rub­ber cover.
The Pop-off Cup can accommodate tubes of virtually any size or shape from micro tubes to those up to 40mm in diameter and 200mm in length. In most instances, tubes without caps may be mixed with proper speed control. Tubes must be capped when potential aerosols may be hazardous.
Catalog numbers:
0K-0500-902 3-inch Platform with Rubber Cover 0K-0500-901 3-inch Platform without Rubber Cover 580-2013-00 Rubber Cover for 3-inch Platform
OPERATING INSTRUCTIONS
We recommend you retain the original packaging for 90 days in case you need to return the product for any reason to your distributor or Scientific Industries.
Your VORTEX-GENIE 2 Mixer is supplied with a Pop-off Cup located in the white foam box insert and a 3-inch Platform. After you unpack your VORTEX-GENIE 2 Mixer from its protective plastic bag, install either the Pop-off Cup or the 3-inch Platform by first aligning the "flat" part of the mixer shaft with the "flat" part of the attachment. YOU MUST PRESS DOWN HARD WITH YOUR HAND UNTIL YOU HEAR THE ATTACHMENT SNAP IN PLACE. This will ensure that the attachment will not come off the mixer during operation.
1.0 - Plug the cord into a properly grounded, three pronged outlet (120 VAC for
the G560 model, 230 VAC for the G560E model and 100V for the G560­100V model).
2.0 - The VORTEX-GENIE
®
2 Mixer has the following modes of operation:
2.1 - TOUCH MODE: Place the toggle switch in the TOUCH position and set the speed control to any position between 3 and 10. A setting of 3 is least vigorous mixing and 10 is the most vigorous mixing. When you touch a vessel to the 3-inch Platform or Pop­off Cup, the mixer will start automatically. Releasing the pressure on the Platform or Cup will stop the mixer.
2.2 - HANDS FREE MODE: Place the toggle switch in the ON position and set the speed control to any position between 3 and 10. Hands Free mode allows for unattended operation that will continue indefinitely until you stop the mixer by placing the toggle