The ORBITAL-GENIE™ is a large capacity, multi vessel orbital shaker.Speed and time are digitally controlled for reproducibility and
g
g
accuracy, with true speed display.
OPERATING INSTRUCTIONS
We recommend you retain the original packaging for 90 days in case you need to return the product for any reason to your distributor or
Scientific Industries.
Plug the line cord into a properly grounded electrical outlet. (120 VAC for the SI-1500, SI-1600 and SI-1700 models; 220-240 VAC
for the SI-1501, SI-1502, SI-1503, SI-1504, SI-1506, SI-1601, SI-1602, SI-1603, SI-1604, SI-1606, SI-1701, SI-1702, SI-1703, SI-1704
and SI-1706 models; 100 VAC for the SI-1506, SI-1606 and SI-1706 models).
Illuminated numbers will appear in the TIME and SPEED windows indicating that the unit is ON and ready for use.
The work surface on which the shaker is placed must be level, stable and non-slip.
. 2Fig. 3
Fi
. 1
Fi
ORBITAL-GENIE™ with Universal Ratcheting Clamps (SI-1500)
The shaker is supplied with two ratcheting and one non-ratcheting cushioned clamps (Fig.1, shown
Fig. 4
with additional ratcheting clamp). The non-ratcheting clamp acts as a stationary stop for securing
the vessels. It should be located at the far end or middle of the platform rods and locked in place
with the thumbscrews. Position the two ratcheting clamps on the platform rods with sufficient space
to fit the vessels. Place the vessels so that they do not touch each other and are seated firmly
against the non-ratcheting clamp. Position vessels evenly distributed on the platform. Uneven
loading may result in excessive vibrating of the shaker and support table. Should this occur, reduce
the speed or turn the shaker off and redistribute the vessels. The clamping platform can
accommodate 250mL – 2 liter vessels with a 26 lbs load capacity. It is very important to regularly
monitor shaking cycles to ensure that the vessels remain firmly secure. Press the ratcheting clamps
until the foam is compressed against the vessels (Fig.4). To release the ratcheting clamps, pull the
release arm and slide away (Fig.5).
Fig. 5
ORBITAL-GENIE™ for Standard Flask Clamps (SI-1600)
The shaker is supplied with a platform for accepting various size flasks or beakers (Fig.2). Flask
clamps attach to the platform with captive mounting screws. Flask clamps are sold separately
Position clamps evenly distributed on the platform. Uneven loading may result in excessive vibrating
of the shaker and support table. Should this occur, reduce the mixing speed or turn the shaker off and
redistribute the load of the vessels.
ORBITAL-GENIE™ with Adhering Mat (SI-1700)
The shaker is supplied with an adhering rubber mat for use with bottles and dishes that cannot be
held with conventional clamps (Fig.3). Patented material is naturally adhering. Rinse clean to renew
adhering properties. Flasks clamps can be used on this unit by removing the adhering mat.
OPERATION
Power on the ORBITAL-GENIE™ (rocker switch on lower right side of housing). The software version number will briefly be shown on
the display. After 5 seconds the display will show time and speed settings.
TIME
Time is displayed in the left window. Press the UP/DOWN arrow button beneath the
TIME window to set the desired cycle time from 1 minute to 99 hours 59 minutes.
Press the RUN/STOP button to begin the shaking action. The shaker will ramp up to
speed and run until the set time has elapsed. Upon time elapse, an alarm will beep
every 15 seconds and the TIME set value will blink. Beeping will end after 5 minutes
but the TIME display will continue to blink. Press any button to exit the cycle end
state.
Fig. 6
NON-TIMED (CONTINUOUS)
Press and hold the DOWN arrow button beneath the TIME window until the illuminated display reaches “00.00”. Press the RUN/STOP
button to begin the shaking action. The timer will count up from 00.00 to 99.59 and then wrap back to 00.00.
SPEED
Set the speed control by pressing the UP/DOWN arrow beneath the SPEED window to any value between 10 RPM and 300 RPM.
REVERSING and ACCELERATION/DECELERATION
To modify the ACCELERATION/DECELERATION ramp speed and REVERSING
TIME, press and hold the NORMAL/REVERSE button for 2 seconds. The TIME and
SPEED displays will flash (Fig.7). The left display is reversing time. The right display
is acceleration/deceleration rate. Use the UP/DOWN arrows on the left side to change
the time between reversing (minutes and seconds – from one second to 99 minutes)
and the arrows on the right side to change the acceleration/deceleration speed value
from 10 to 100 RPM/SEC. The new value will be applied immediately and memorized upon key release. This mode
(flashing) will be exited automatically after 10 seconds or by pressing the NORMAL/REVERSE button once. By modifying these values,
you can change the nature of mixing action from smooth vortexing to sloshing aeration, or anything in between.
REVERSE mode toggles the shaker between operating in only one direction, and
switching directions at regular intervals. To initiate the REVERSE mode, press the
NORMAL/REVERSE button. The REVERSE LED on the display will turn on (Fig.8).
Press the RUN/STOP button to begin the mixing action.
Please note that if you set acceleration to lower values, in combination with short
reversing or cycle times, the unit will limit the maximum speed setting to values that
are achievable (ie: reversing time of 5 seconds with acceleration/deceleration of 10 RPM/second – unit will only allow maximum speed
of 25 RPM – 2.5 seconds of acceleration at 10 RPM/second and 2.5 seconds of deceleration at 10 RPM/second).
Also note that, for safety reasons, when pressing the RUN/STOP button to end a cycle (instead of letting the timer stop the unit),
maximum deceleration (100RPM/second) is used .
In between shaking cycles, the platform can move freely. To lock the movement of the
platform (for ease of loading vessels) press and hold the RUN/STOP button and
simultaneously press the NORMAL/REVERSE button. The shaking platform will lock.
See Fig. 9 for display readout. Initiating a shaking cycle will exit this mode.
SPECIFICATIONS
The ORBITAL-GENIE™ Shaker is classified as “Installation Category 2”
Environmental: 0°C – 38°C (32°F-100°F), 95% Humidity max.
MODEL POWER REQUIRED AMPS MODEL POWER REQUIRED AMPS
SI-1500 120V 2.0 SI-1703 230V British Plug 1.0
SI-1600 120V 2.0 SI-1504 230V Swiss Plug 1.0
SI-1700 120V 2.0 SI-1604 230V Swiss Plug 1.0
SI-1501 230V No Plug 1.0 SI-1704 230V Swiss Plug 1.0
SI-1601 230V No Plug 1.0 SI-1506 230V Australian Plug 1.0
SI-1701 230V No Plug 1.0 SI-1606 230V Australian Plug 1.0
SI-1502 230V Euro Plug 1.0 SI-1706 230V Australian Plug 1.0
SI-1602 230V Euro Plug 1.0 SI-1505 100V 2.5
SI-1702 230V Euro Plug 1.0 SI-1605 100V 2.5
SI-1503 230V British Plug 1.0 SI-1705 100V 2.5
SI-1603 230V British Plug 1.0
Weight: 21 Kg (46 lbs) Base Dimensions: (DxW) 590 x 500mm, (23 x 20in) Orbit: 35mm diameter
Care and Handling
Your ORBITAL-GENIE™
fumes. The finish can be washed with water (after unplugging) and soap or detergents, using a cloth or sponge. Keep the unit clean by immediately
blotting any spills. Replacement parts are available through your laboratory equipment distributor or directly from Scientific Industries. Refer to “Parts
Assembly List”.
Shaker should be given the care normally required for any electrical appliance. Avoid wetting or unnecessary exposure to
CAUTION! Do not lift your ORBITAL-GENIE™ Shaker by the clamping platform.
CAUTION! Failure to follow operating instructions can compromise the user’s safety.
CAUTION! Unplug from power before cleaning. Do not immerse
Fig. 7
Fig. 8
Fig. 9
Parts Assembly List
o
To order parts for the ORBITAL-GENIE™ Shaker:
Contact your local distributor or visit www.scientificindustries.com. Please specify Part No., quantity and electric voltage.
Item N
4A 318-0510-02 120V Line Cord 17 SI-1617 Adhering Mat
4B 0K-0246-901 230V Line Cord, No Plug 18 SI-1616 Double Sided Adhesive
4C 0K-0256-901 230V Line Cord, Euro Plug 19 SI-1510 Universal Ratcheting Clamp
4D 0K-0266-901 230V Line Cord, British Plug 20 SCP0074 ¼-20 x 2 inch Screw
4E 0K-0276-901 230V Line Cord, Swiss Plug 21 EMP0010 Driver, Motor
4F 0K-0286-901 100V Line Cord 22 EMP0009 Stepper Motor
4G ECP0021 Australian Plug (plug only) 23 EL-1500-575 Keypad
10 0M-1500-204 Top Platform 29 BRP001 Bearing
11 0M-1500-215 Threaded Spacer 30 SCP0076 6-32 x 3/8” long screw
12 0M-1500-205 Rod, Ratchet 31 WAP0031 Washer
13 SI-1575 Non-ratcheting Clamp 32 WAP0038 Special Flat Washer
14 0M-1500-214 Non-slip Rubber Pad 33 HWP0075 Ribbed Bushing