ScienTECH Vector Series, Vector S310D, Vector S310 Operating Manual

Page 1
Vector
Series
Laser Power and
Operating Manual
Energy Meters
Thank you for choosing a Scientech Vector calorimeter detector. Scientech’s employees are pleased to provide you with a product designed for years of reliable and accurate service. Please read this operating manual before using your detector and power meter. This reference information will allow you to fully understand the capabilities of the product. The detector is intended to be used only in the manner outlined in this manual. Operation not within specifications for the product may cause product damage.
PN10190T
1
Page 2
Table of Contents
Detector Operating Parameters .................................................................................................................................................... 3
Environmental Requirements ...................................................................................................................................................... 4
CE Mark Certification ......................................................................................................................................................... 4
Vector™ S310 Power Meter Specifications ................................................................................................................................ 4
Unpacking .................................................................................................................................................................................... 6
Quick Setup ................................................................................................................................................................................. 6
Group Settings ............................................................................................................................................................................. 9
To Select a Group................................................................................................................................................................ 9
Group Settings for Vector Calorimeters ........................................................................................................................... 9
Group Settings for Vector Pyroelectric Detectors .......................................................................................................... 12
Operating Procedures ................................................................................................................................................................. 14
Using the S310 with Vector Pyroelectric Detectors ....................................................................................................... 14
9. External T rigger .......................................................................................................................................... 20
To Perform a Transfer Calibration ......................................................................................................................... 20
HR Battery Installation .......................................................................................................................................... 20
Set Electrical Time Constants for Model PHF02, PHF05 and PHF09 .................................................................. 21
Correcting Pyroelectric Detector Operating Parameters for Use at Different Wavelengths ............................................. 21
Using S310 with Vector Series Calorimeters ................................................................................................................. 22
Correcting FX Calorimeter Operating Parameters for Use at Different Wavelengths ...................................................... 27
Analog Output ................................................................................................................................................................... 28
Calibration of VectorCalorimeters Using Electric Substitution Heaters ............................................................................... 30
Calibration of Large Aperture Calorimeters Using Electric Substitution Heating ..................................................................... 31
Calibration with an Interface Module and S310 Power Meter .......................................................................................... 32
Calibration without an Interface Module and S310 Power Meter ..................................................................................... 32
Detector Operation Without a Power Meter .............................................................................................................................. 34
Pyroelectric Detector Operation without Power Meter ..................................................................................................... 34
Vector ™ Calorimeter Operation without Power Meter ..................................................................................................... 34
Operation of VectorCalorimeters with a Digital Volt Meter ......................................................................................... 35
Operation of VectorCalorimeters with an Analog Recorder ......................................................................................... 35
Calorimeter Re sponse: ........................................................................................................................................... 35
Numerical Integration: ........................................................................................................................................... 36
Initial Voltage Interpolation: ................................................................................................................................. 36
Peak Voltage Estimate: .......................................................................................................................................... 37
Remote Interface Language ....................................................................................................................................................... 37
Scientech Calibration Service .................................................................................................................................................... 45
Limited Warranty ....................................................................................................................................................................... 45
Return Material Procedure ......................................................................................................................................................... 45
Disposal of Electronic Equipment ............................................................................................................................................. 46
Schematics ................................................................................................................................................................................. 47
PN10190T
2
Page 3

DETECTOR OPERATING PARAMETERS

All detectors are calibrated at a specifi c wavelen gth and the d etector’s op erating param eters are d erived for that wavelength. This information is recorded below and on the detector’s serial tag. When a detector is used at a wavelength other than the calibration wavelength some of the operating parameters may need to be adjusted. For specific instructions please refer to the Operating Procedures section for the type of detector you are using.
Power Meter Serial Number___________________
Calorimeter 1: Calorimeter 2:
Model No: ___________ Model No: __________ Serial No: ___________ Serial No: __________ Group No: ___________ Group No: __________ Calibration Wavelength ___________nm or µm Calibration Wavelength: __________nm or µm Output Sensitivity (S): ___________V/W Output Sensitivity (S): __________V/W Time Constant (1/e): ___________sec. Time Constant (1/e): __________sec. Calibration Temp: ___________°C Calibration Temp: __________°C Calibration Humidity: ___________%R.H. Calibration Humidity: __________% R.H. Sub. Heater Resistance (Rc): ___________ohms Sub. Heater Resistance (Rc): __________ohms
Sub. Heater Voltage (Vh): ___________volts Sub. Heater Voltage (Vh): __________volts Sub. Heater Wattage (Wh) ___________watts Sub. Heater Wattage (Wh): __________watts
Pyroelectric Detector 1:
Model No: __________ Serial No: __________ Group No: __________ Calibration Wavelength: __________nm or µm Output Sensitivity: Output Sensitivity:1 V/J (Scope) Calibration Temp: __________°C Calibration Humidity: ___________%R.H.
Pyroelectric Detector 2:
Model No: __________ Serial No: __________ Group No: __________ Calibration Wavelength: __________nm or µm Output Sensitivity: Output Sensitivity: Calibration Temp: __________°C Calibration Humidity: ___________%R.H.
The V/J is used t o calibrate the power meter to a known NIST traceable standard. This is the value the meter must use. The V/J (Scope) is the “true” Volts per Joule as seen at an oscilloscope.
1
__________V/J or __________V/mJ S_________ I__________ L_________
1
__________V/J or __________V/mJ S_________ I_________ L__________
V/J (Scope)
PN10190T
3
Page 4

ENVIRONMENTAL REQUIREMENTS

This product is intended for indoor use at altitudes up to 2000 meters, Pollution Degree 2 in accordance with IEC 664 and transient over voltages according to Installation Categories (Overvoltage Categories) II. Note that each of the above detectors will not pass the IEC 801 Publication, Part 3, Radiated Electromagnetic Field Requirements. The system, meter and detector, is designed to measure radiation within the test's radiation band. The detectors were held outside the radiate d elec tromagnetic f ield during this test. It is up to the user to be aware of RF fields present during measurements and their effects if any on those measurements.

CE Mark Ce rtification

All of the power meters listed in this manual have been certified for the European CE mark.

VECTOR S310 POWER METER SPECIFICATIONS

The Models S310 and S31 0D are identical to each other i n every respect except the an alog meter. The S310 has both analog and digital displays. The S310D has only a digital display. All references to the S310 are intended to include the S310D except where noted. Also, the words "indicator" and "meter" are synonymous. Specifications are summarized in Table 1.
PN10190T
4
Page 5
Model
S310
S310D
Display
4 Digit backlight LCD with analog
4 Digit backlight, LCD
Operating Modes, Power or
Pyroelectric: energy, average energy,
Pyroelectric: energy, avg energy, power,
Pyroelectric Statistics Mode
Pulses collected, avg energy, min energy,
Pulses collected, average energy, min Calorimeter max repetition rate,
Unlimited
Unlimited
Calorimeter max repetition rate,
1 pulse every 60 to 90 sec, calorimeter
1 pulse every 60 to 90 sec, calorimeter
For 25mm Calorimeters:
For 50mm Calorimeters:
For Pyroelectric Detectors:
For HR Pyroelectric Detectors
Not supported
3 µJ, 30 µJ, 300 µJ, 3 mJ, 30 mJ, Auto
For 100mm Calorimeter:
150J
500mW, 5W, 50W, Auto or 1.5mJ, 15J,
For 200mm Calorimeter:
Use Interface Module, PN 10769
1W, 10W, 100W, Auto or 3J, 30J, 300J
Power Requirements
120 Volts, 60 Hz ± 10 % or
120 Volts, 60 Hz ± 10 % or
Dimensions D x L, inches
8.25 x 4.0 x 1.5
4.7 x 8.8 x 7.8
Other Specs:
LabVIEW™ drivers,
correction, IEEE
LabVIEW™ drivers,
Table 1. S310 Power Meter Specifications
indicator
Energy
power mode
energy mode
Power Full Scale Ranges Energy Full Scale Ranges
Power Full Scale Ranges Energy Full Scale Ranges
Power Full Sca le Ranges Energy Full Scale Ranges
power, statistics Calorimeter: energy, power, calibration
max energy, standard deviation, coefficient of variation
dependent
10mW, 100mW, 1W, 10W, Auto 10mJ, 100mJ, 1J, 10J
300mW, 3W, 30W, Auto 300mJ, 3J, 30J
3mW, 30mW, 300mW, 3W, 30W, Auto 3mJ, 30mJ, 300mJ, 3J, 30J, Auto
statistics Calorimeter: energy, power, calibration
energy, max energy, standard deviation, coefficient of variation
dependent
10mW, 100mW, 1W, 10W, Auto 10mJ, 100mJ, 1J, 10J
300mW, 3W, 30W, Auto 300mJ, 3J, 30J
3mW, 30mW, 300mW, 3W, 30W, Auto 3mJ, 30mJ, 300mJ, 3J, 30J, Auto
Use Interface Module, PN 10735
Use Interface Module, PN 10748
Use Interface Module, PN 10747
cm
PN10190T
300mW, 3W, 30W, Auto or 300mJ, 3J, 30J 500mW, 5W, 50W, Auto or 1.5mJ, 15J,
300mW, 3W, 30W, Auto or 300mJ, 3J, 30J
220 Volts, 50 Hz ± 10 %
20.96 x 10.16 x 3.81
Selectable attenuation 488 interface option
300mW, 3W, 30W, Auto or 300mJ, 3J, 30J
150J
300mW, 3W, 30W, Auto or 300mJ, 3J, 30J 1W, 10W, 100W, Auto or 3J, 30J, 300J
220 Volts, 50 Hz ± 10 %
11.9 x 22.5 x 19.9
Selectable attenuation correction, IEEE 488 interface option
5
Page 6

UNPACKING

The meter, detect ors, and accessories are sh ipped in custom packing ma terials. All packing materials should be saved for future damage free shipments.
Before making any connections, verify that the power (VAC) requirement shown on the power entry module is compatible with the actual AC power outlet to which the power meter will be connected. To change the power meter’s voltage, proceed as follows:
1. Refer to Figure 1. Locate the power entry module and the fuse holder in the center of the module.
2. Remove the fuse holder by inserting a slotted screwdriver in the slot on the right side and prying
it out.
3. Slide the voltage selector out, flip it over and re-insert it into the fuse holder.
4. Re-insert the fuse holder into the power entry module.
Figure 1. S310 Rear Panel

QUICK SETUP

For detailed instructions for each type of detector, refer to the Operating Procedures section.
1. Power On the Meter
For the most accurate measurements possible, the S310 should be turned on and warmed up for 30 minutes.
Press the ON switch located in the rear pan el of the meter. T he S310 will imm ediately turn on with its operational state based on the last used detector Group Setting. For information on Group Settings, refer
PN10190T
6
Page 7
Analog Needle Zero Adjustment
to the Group Settings section. If you purchased the S310 with one detector, this detector’s Group settings will be active and you are ready to take measurements.
If you purchased more than one detector with the S310, you must choose the Group setting for the detector you are connecting to the S310 in step 4. The Group Setting for each detector purchased with your S310 is shown on page 2. To select a Group Setting, press the SETUP button repeatedly until the desire Group number appears in the display. Wait 3 seconds and that Group will be automatically selected and made active.
2. Turn the Analog Needle On or Off (does not apply to the S310D)
To turn the analog meter on or off, press and keep holding down the RANGE button. Then release the button after the meter appears or disappears.
3. Zero the Analog Needle (does not apply to the S310D)
Refer to Figure 2. The black slotted button located just below the display allows screwdriver adjustment to set the analog needle to zero. This adjustment should be made before connecting the detector.
PN10190T
Figure 2. S310 Front Panel
(under serial tag)
7
Page 8
4. Connect a Detector
Do not connect more than one detector to the power meter at any time.
Refer to the drawing of the S310 rear panel in Figure 1. A 3-meter mini-DIN type cable with "D" shaped connectors comes with Vector calorimete rs, incl uding the large aperture calorimeters. A 3 -meter BNC type cable comes with Vector pyroelectric detectors and Vector HR pyroelectric detectors. To connect the Vector detectors, the input connectors on the rear panel of the S310 are labeled as follows:
"Ultra" for Ultra Detectors (no longer manufactured) "Vector" for all Vector Pyroelectric detectors "Astralfor all Vector calorimeters, including the large aperture calorimeters
Note that the flat side of the DIN type cables should be oriented up when plugging into the S310. Detectors have a 1/2" diameter mounting post hole for installing a detector post. Mounting bases and posts are available for purchase at Scientech to attach the detector to a working surface.
5. Select a Range
On the first press of the RANGE button, the analog scale (if active) will disappear. Then briefly press the RANGE button each time another range is desired. Press the SELECT button when the range you desire appears in the display (failure to press the SELECT button within 3 seconds will activate the range shown in the display and return the S310 to its operation state).
6. Select a Mode
The MODE button allows selection of the type of measurement to be made (watts, joules, etc). It also gives access to the scientific notation unit of measure. The different measurement modes available are:
A. When configured for a pyroelectric detector: Energy (J), Avg. Energy (J AVG), Power ( W AVG),
and Volts (V).
B. When configured for a calorimeter: Power (W AVG), Energy (J).
Briefly press the MODE button each time a different mode setting is wanted. Press the SELECT button when the mode you desire appears in the display (failure to press the SELECT button within 3 seconds will activate the mode shown in the display and return the S310 to its operation state).
To select the scien tific notation annunciator, pr ess and hold down the MODE button and only release when the annunciator appears. To exit the scientific notation mode, press and hold down the MODE button again.
7. Zero the Display (only for calorimeters with Watts mode selected)
Press the CANCEL button to zero the display when using a calorimeter in watts mode.
PN10190T
8
Page 9
8. Take your measurement
Direct the laser beam onto the absorbing surface of the detector.

GROUP SETTINGS

The operating parameters for the detectors used with the S310 power meter must be stored in the power meter’s memory. This information is stored in Group Settings. The typical configuration of the Group Settings is as follows:
Group #1 – Vector Calorimeters or Large Aperture Calorimeters
Group #2 – Vector Pyroelectric Detectors
Group #3 – Vector HR Pyroelectric Detectors
However, the Group Settings may be configured differently depending on what detectors are used with the power meter. The operating parameters and group number for the detectors purchased with the S310 Power meter are recorded on page 2 of this manual.

To Select a Group

A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again.
C. To select a Group, press the SELECT button when the Group you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the Group shown in the d isplay and return the S310 to its operational state).
Group Settings for Vector Calorimeters
The factory default Group for Vector 8mm, 25mm, 16mm, 50mm or large aperture calorimeters is Group #1.
Note: Pressing the CANCEL button at any time during this process will terminate this process with no changes made.
Note: Holding down the SELECT button for 3 seconds at any time durin g this process will save any changes made up to that point and return the S310 to its operational state for the Group selected.
PN10190T
9
Page 10
To review, change, or setup a Group Setting for Vector calorimeters, proceed as follows:
A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again.
C. To select a Group, press the SELECT button when the Group you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the Group shown in the d isplay and return the S310 to its operational state).
D. Press the SETUP button until the calorimeter annunciator “CAL” appears in the display . Press the
SELECT button to activate the calorimeter configuration program.
E. The "tc" (time constant) annunciator now appears in the display. The time constant is a measure
of the length of time the calorimeter takes to respond to a laser beam.
Press the count up (RANGE) and count down (MODE) buttons to change the number in the display to match the time constant number listed on the serial tag of your Vector calorim eter or int erface module for large aperture calorimeters. Press the SELECT button to enter the time constant value to memory.
F. The "SP" (speed) annunciator now appears in the display. The speed setting allows you to control
the power met er’s display rate. The best value will cause a slight overshoot then a quick settling on the final value. Too high of a setting will cause the display to overshoot then slowly drift back down to the final value. A slow setting will cause the display to slowly count up to the final value.
Press the count up (RANGE) button to speed up or count down (MODE) button to slow down the response time of the meter. Press the SELECT button when the desired number appears in the display. The following settings can be fine tuned to your preference of speed versus overshoot. We recommend these initial settings:
AC2500, AC25HD, ACX25HD 103.0 AC2501, ACX2501, AC25UV, AC2504, AC25FX, ACX25FX 136.0 AC5000, AC50HD ACX50HD, AC50FX, ACX50FX 100.0 AC5001, ACX5001, AC50UV, AC5004, 120.0 360401 with interface module 150.0 380401, 380402, 384UV5 with interface module 245.0 360801 with interface module 170.0 380801, 380802, 388UV5 with interface module 280.0
G. The "Cd" (calorimeter delay) annunciator now appears in the display. The calorimeter delay
feature prohibits the display of energy if a pulse is fired before the entered time (1 to 255 seconds) elapses. The calorimeter must reach environmental thermal equilibrium before a subsequent pulse is fired or low energy measurements will occur.
PN10190T
10
Page 11
To set the time delay between pulses use the count up (RANGE) and count down buttons (MODE) buttons to enter the time in seconds. Press the SELECT button after the time has been entered to save the setting. The following time delays are recommended:
AC2500, AC25FX, ACX25FX, AC25HD, ACX25HD 60 seconds AC2501, AC25UV, AC2504 60 seconds AC5000, AC50FX, ACX50FX, AC50HD, ACX50HD 90 seconds AC5001, AC50UV, AC5004 90 seconds 360401 with interface module 105 seconds 380401, 380402, 384UV5 with interface module 180 seconds 360801 with interface module 125 seconds 380801, 380802, 388UV5 with interface module 200 seconds
H. The "At" (attenuation) annunciator now appears in the display. If no attenuator is b eing used in
conjunction with the calorimeter, the attenuation factor in the display must be set at 1.000 since this value is a display multiplier. Attenuation multipliers from .0001 to 9999 can be entered into the displayed value.
The attenuation factors of optics can be entered into the S310 so the displayed value will automatically compensate for the amount of attenuation. For example, assume a beam splitter is being used that transmits 75% and reflects 25% of the beam. If the S310 is set up to measure the reflected beam the attenuation could be set up as follows:
An attenuation factor of 1 would display the value of the reflected beam.
An attenuation factor of 3 would display the value of the transmitted beam.
An attenuation factor of 4 would display the value of the source.
Use the count up (RANGE), count down (MODE) buttons to change the attenuation factor. Press the SELECT button to enter the attenuation factor to memory.
I. If the S310 has the optional IEEE488 interface, the REMOTE annunciator will appear along with
the digital interface previously selected; r232 (RS232) or IEEE (IEEE488). Press the SETUP button to toggle the interface between r232 (RS232) and IEEE (IEEE488). Press the SELECT button when the desired interface appears in the display.
J. If the optional IEEE488 interface was not installed, the "br" (baud rate) annunciator for setting up
the RS232 interface appears in the display along with the baud rate previously selected. To change the baud rate press SETUP button until the desired baud rate appears in the display. One of the following baud rates can be selected: 300, 1200, 2400, 9600, and 19,200. When the preferred baud rate appears in the display press the SELECT button.
K. The "PA" (parity) annunciator now appears in the display. Press the SETUP button to select none,
even, or odd parity. Press the SELECT button to enter the parity of choice to memory
PN10190T
11
Page 12
L. The "HS" (handshake) annunciator now appears in the display. Press the SETUP button until your
choice of none (nOnE), on/off (onoF), or clear to send (CtS) appears in the display. Press the SELECT button when your choice is displayed to enter the handshake to memory.
M. If IEEE488 has been selected, the "bA" (bus address) annunciator appears with the default bus
address. Bus addresses from 0 to 30 may be selected by pressing the count up (RANGE) count down (MODE) buttons followed by the SELECT button. Once the remote interface has been setup, the group annunciator will appear. Press the SELECT button or simply wait a few seconds and the group will automatically be selected for operation.
Group Settings for Vector Pyroelectric Detectors
Note: The factory default Group for Vector Pyroelectric Detectors is Group #2.
Note: The factory default Group for Vector HR Pyroelectric Detectors is Group #3.
Note: Pressing the CANCEL button at any time during this process will terminate this process with no
changes made.
Note: Holding down the SELECT button for 3 seconds at any time during this process will save any changes made up to that point and return the S310 to its operational state for the Group selected.
Note: If you want to use the transfer calibration function, y ou must select the watts mode prior to entering the Group configuration program.
To review, change, or setup a Group Setting for Vector Pyroelectric or Vector HR Pyroelectric detectors, proceed as follows:
A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again. C. To select a Group, press the SELECT button when the Group you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the Group shown in the display and return the S310 to its operational state).
D. Press the SETUP button until the "PYro" annunciator appears. Immediately press the SELECT
button to select the pyroelectric detector configuration program.
E. Either the "V/J" annunciator or the "V/mJ" annunciator and a detector sensitivity number will
appear in the display. V/J is used for standard p yroelectric detectors. V/mJ is used only for HR pyroelectric detectors. Press the SETUP button to toggle between the V/J and V/ mJ entries . The output sensitivity of the pyroelectric detector in V/J or V/mJ is listed on the serial tag of your detector. Use the count up (RANGE) and count down (MODE) buttons to enter the V/J or V/mJ value. Press the SELECT button to enter the value to memory. Do not push the SELECT button
PN10190T
12
Page 13
before entering the sensitivity number as this takes you to the next setup step without the proper sensitivity number.
F. The "SP" (speed) annunciator now appears in the display. Press the SETUP button to toggle
between bL (black or HD coated detectors) or hF (high frequency) detector. If the detector Model No. includes the letters “HF”, press the SELECT button when the hF annunciator appears in the display. If the detector’s surface is black (does not have the let ters “ HF” in the Model No.) press the SELECT button when the "bL" annunciator appears in the display.
G. The "AUTO, SET CAL" annunciators now appear in the display (only if watts mode was selected
prior to group configuration). This is the Transfer Calibration function. The Transfer Calibration function is designed to transfer the calibration from a NIST certified calorimeter to a Vector pyroelectric detector. This function allows you to adjust the output sensitivit y o f your Vector pyroelectric detector in combination with your S310 meter (in the average power mode) in order to match the avera ge po wer r eadin g of a N IST c erti fied s ystem. T ypicall y, a 50/50 beam splitter is used with the Vector pyroelectric detector to be calibrated in one beam path and the NIST certified calorimeter in the other beam path. You are to adjust the output sensitivity of the pyroelectric detector while it is operating by using the count up (RANGE), count down (MODE) buttons, which change the V/J or V/mJ settings, to make the displayed value of the S310 power reading the same as the NIST standard. Press the SELECT button when the readings match.
H. The "At" (attenuation) annunciator now appears in the display. This allows you to enter a value
from 0.0001 to 9999 which will become a multiplier of the actual displayed value. I f no attenuator or correction factor is to be used in conjunction with the pyroelectric detector then the attenuation factor must be set to 1.000.
The attenuation factors of optics can be entered into the S310 so the displayed value will automatically com pensat e for the amou nt of atten uatio n. For exam ple, assu me a beam spl itter is being used that transmits 75% and reflects 25% of the beam. If the S310 is set up to measure the reflected beam the attenuation could be set up as follows:
An attenuation factor of 1 would display the value of the reflected beam.
An attenuation factor of 3 would display the value of the transmitted beam.
An attenuation factor of 4 would display the value of the source.
Use the count up (RANGE), count down (MODE) buttons to enter the attenuation value. Press the SELECT button to enter the value to memory.
I. The REMOTE annunciator now appears along with the digital interface previously selected; r232
(RS232) or IEEE (IEEE488). Press the SETUP button to change the interface between r232 (RS232) and IEEE (IEEE488). Press the SELECT button when the desired interface appears in the display.
J. If RS232 has been selected, the "br" (baud rate) annunciator appears in the display along with the
baud rate previously selected. To change the baud rate press the SETUP button repeatedly until the desired baud rate appears in the display. One of the following baud rates can be selected: 300,
PN10190T
13
Page 14
1200, 2400, 9600, and 19,200. When the preferred baud rate appears in the display press the SELECT button.
K. The "PA" (parity) annunciator now appears in the display. Press the SETUP button to select none,
even, or odd parity. Press the SELECT button to enter the parity choice to memory
L. The "HS" (handshake) annunciator now appears in the display. Press the SETUP button until your
choice of none (nOnE), on/off (onoF), or clear to send (CtS) appears in the display. Press the SELECT button when your choice is displayed to enter the handshake to memory.
M. If IEEE488 has been selected, the "bA" (bus address) annunciator appears with the default bus
address. Bus addresses from 0 to 30 may be selected by pressing the count up (RANGE) count down (MODE) buttons followed by the SELECT button. Once the remote interface has been setup, the group annunciator will appear. Press the SELECT button or simply wait a few seconds and the group will automatically be selected for operation.

OPERATING PROCEDURES

Using the S310 with V ec tor Pyroelectric Detectors
Pyroelectric detector models P25, P50, SP25, SP50, P05, and P09 are coated with a special black absorbing material which provides a very flat spectral response over a broad wavelength band. Pyroelectric detector models PHD25, PHDX25, PHDX25UV, PHD50, PHDX50, PHDX50UV, SPHD25, and SPHD50 are coated with a special high damage absorbing material which provides absorption over a broad wavelength band. Models PHF25, PHF50, SPHF25, SPHF50, PHF02, PHF05, and PHF09 have a partially absorbing, partially reflecting chromium coating. The relative spectral responses of these detectors are shown in the following graph. Please be aware of the absorption differences between the detector’s calibration wavelength and your operational wavelength. Detailed absorption information is contained in the charts at the end of this manual.
PN10190T
14
Page 15
RELAT I VE SPEC TRAL R ESPON SE
20
30
40
50
60
70
80
90
100
0.1 1 10 100
WAVELENGTH [m icrons]
PERCENT ABSORPTION
P and SP
HR-P
PHD, PH and SPH
HR-P
PHF a nd SPHF
*HDX Limi ted to 0.4 to 2.0
PHDXUV
Figure 3. Pyroelectric Spectral Response
Before using your Vector jouleme ter sy stem, plea se review the energy density for mulas given in the chart at the front of this man ual. Familia rize yo urself with all of th e spec ifications of the detecto r models which you are using. A damage test slide is provided with each P and PH F type det ectors, but not with PHD type detectors. Fire the beam at the test slide before using the detector to be sure you are operating under safe conditions.
The default Group Setting for Vector Pyroelectric Detectors is Group #2. The default Group Setting for Vector HR Pyroelectric Detectors is Group #3. Either select Group #2, #3, or configure another Group for pyroelectric detectors and select that Group. For Group configuration, refer to the Group Settings section.
Note: The automatic trigger threshold of the S310 is 7 % of full scale.
1. To Turn On the Meter
Note: For the most accurate measurements possible, the S310 should be turned on and warmed up for 30
minutes.
Press the ON switch located in the rear panel of the meter. The S310 will immediately turn on with its operational state based on the last used detector Group Setting. If you purchased the S310 with one detector, this detector’s Group settings will be active and you are ready to take measurements.
If you purchased more than one detector with the S310, you must choose the Group setting for the detector you are connecting to the S310 as described in step 4. The Group Setting for each detector purc hased with your S310 is shown on page 2.
PN10190T
15
Page 16
2. To Turn the Analog Needle On or Off (does not apply to the S310D)
To turn the analog meter on or off, press and keep holding the RANGE button down. Then release after the meter appears or disappears.
3. To Zero the Analog Needle (does not apply to the S310D)
Refer to Figure 3. The black slotted button located just below the display allows screwdriver adjustment to set the analog needle to zero. This adjustment should be made before connecting a pyroelectric detector.
4. To Connect a Pyroelectric Detector
Note: Only one detector should be plugged in at any time.
Refer to the drawing of the S310 rear panel in Figure 2. A 3 mete r BNC type cabl e comes wit h Vector pyroelectric detectors and Vector HR pyroelectric detectors. One of the input connectors on the rear panel of the S310 is labeled "Vector" for hook up of the pyroelectric detectors.
5. To Select a Group
A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again.
B. To select the appropriate Group, press the SELECT button when the Group you desire appears in
the display (failure to press the SE LECT button within 3 seconds will activate the Group shown in the display and return the S310 to its operational state).
6. To Select a Range
Note: AUTO range may be selected if the energy levels of repetitive pulses are to be measured. However,
do not select AUTO range if you want to measure single pulse energy or pulses running at repetition rates lower than 10 Hz.
A. Press the RANGE button. On the first press of the RANGE button, the analog scale (if active) will
disappear. Then briefly press the RANGE button each time another range is desired.
B. To select a range, press the SELECT button when the range you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the range shown in the display and return the S310 to its operational state). The ranges available for pyroelectric detectors are in the following table.
PN10190T
16
Page 17
Model
Standard or SP
PHF02 - L
PHF02 – S or I
PHF05 – L
PHF05 – S or I
Mode
Power
Energy
Power
Energy
Power
Energy
Power
Energy
Power
Energy
3mW
3mJ
30.00μW
30.00μJ
3.000μW
AUTO
3.000μJ
AUTO
300.0μW
300.0μJ
3.000μW
AUTO
3.000μJ
AUTO
Model
PHF09 – L
PHF09 – S or I
P05
P09
Mode
Power
Energy
Power
Energy
Power
Energy
Power
Energy
3.000m
3.000m
30.00μW
30.00μJ
3.000μW
AUTO
3.000μJ
AUTO
3.000μW
AUTO
3.000μJ
AUTO
Table 2. S310 Ranges for Pyroelectric
Range s
Range s
30mW
300mW
3W
AUTO
W
30.00m W
AUTO
30mJ
300mJ
3J
AUTO
J
30.00m J
AUTO
3.000μW
3.000m W
30.00m W
AUTO
300.0μW
3.000m W
30.00m W
AUTO
3.000μJ
3.000m J
30.00m J
AUTO
300.0μJ
3.000m J
30.00m J
AUTO
30.00μW
300.0μW
3.000m W
30.00m W
30.00μW
300.0μW
3.000m W
30.00m W
30.00μJ
300.0μJ
3.000m J
30.00m J
30.00μJ
300.0μJ
3.000m J
30.00m J
3.000m W
30.00m W
AUTO
30.00μW
300.0μW
3.000m W
30.00m W
3.000m J
30.00m J
AUTO
30.00μJ
300.0μJ
3.000m J
30.00m J
30.00μW
300.0μW
3.000m W
30.00m W
30.00μJ
300.0μJ
3.000m J
30.00m J
7. To Select a Mode
Available pyroel ectric modes are: Energ y (J), Avg. Energy (J AVG), A verage Power (W AVG), and Volts (V).
A. To Measure Energy (J):
Note: The update rate of the display is 20Hz.
i. Press the MODE button (repeatedly, if necessary) until the “J” annunciator appears in the
display.
ii. Press the SELECT button (or wait 3 seconds) and the joules mode will automatically be
activated.
iii. The energy level of each laser pulse will be displayed on the LCD.
B. To Measure Average Energy (J AVG)
PN10190T
17
Page 18
Note: The maximum repetition rate for average energy is 300Hz.
i. Press the MODE button (repeatedl y, if necessary) until the “J AVG” annunciator app ears in
the display.
ii. Press the SELECT button (or wait 3 seconds) and the average joules mode will be
automatically activated. The number of pulses to be averaged will now appear in the display. You may select the number of pulses to average from 2 to 9999.
iii. To change the number of pulses to be averaged, press the RANGE (count up) and/or M ODE
(count down) buttons. When the desired number of pulses to be averaged appears in the display, press the SELECT button.
iv. The average energy of the number of pulses you selected will be displayed after the number of
pulses entered in step ii is received by the pyroelectric detector. This is not a running average, but is the average for the pulse population selected in step ii. Nothing is displayed until the full pulse population is delivered. This average is displayed until another full population of pulses is delivered at which time the display is updated with the average for that population of pulses.
C. To Measure Average Power (W AVG)
Note: The maximum re petition rate for average power is 300Hz. The minimum repe tition rate for average
power is 10Hz.
Note: The average power mode displays the average power (watts) of repetitively pulsed lasers. Pyroelectric detectors will not work with continuous wave lasers.
i. Press the MODE button repeatedly until the “W AVG” annunciator appears.
ii. Press the SELECT button (or wait 3 se conds) and the S310 will automatically enter into its
operational state in the average power mode.
iii. The average power will be displayed.
D. To Measure Volts (V)
Note: The maximum repetition rate for volts is 300Hz.
i. Press the MODE button repeatedly until the “V” annunciator appears.
ii. Press the SELECT button (or simpl y wait 3 seconds) and the S310 will automatically enter
into its operational state in the volts mode.
iii. The volts per pulse will be displayed.
8. To Measure a Statistical Run of Energy Pulses
PN10190T
18
Page 19
Note: Do not use AUTO range when making a statistical run.
Note: Do not go from the Average Power Mode to Stats since the range will be too high. Select the range
manually.
Note: When using the PHF09 pyroelectric detector in the long pulse setting, a manual range must be selected when running statistics.
Note: Each time a new stats run begins, data from the previous run is lost. If statistical data is to be saved, it must be done though the digital interface.
Note: To exit the statistical mode at any time, press the CANCEL button.
Note: Statistics mode can collect data at repetition rates of up to 750 Hz depe nding on the detector model.
The statistics mode will collect data on a pulse population of up to 1000 pulses. At your prompting, the power meter will display the number of pulses deli vered, average energy, minimum energy, maximum energy, standard deviation, and coefficient of variation. When the statistics mode is selected, the energy mode is automati cally activated regardless of t he m od e prev io us l y sele cted. Select the appropri at e r ange for the pulse energy level to be measur ed. It is very importan t to select the most appropriate range. I f you have selected a manual range and the laser pulse(s) has overflowed the maximum energy of the range, OF will be displayed when the data is recalled. You should then select a higher range.
To enter into the statistics mode:
A. Press the STATS button. The number of pulses in the last statistics run will appear in the display.
B. Use the RANGE (count up) and MODE (count down) buttons to change the display to the desired
number of pulses to include in the statistics run (up to 1000).
C. Press the SELECT button to enter the pulse population to memory.
D. Press the SELECT button to begin the run. The power meter will automatically stop once the data
has been collected.
B. Press the STATS button to recall the data to the display. The STATS button must be pressed each
time to recall each of the following statistical calculations:
Number of pulses collected Average energy (AVG) Minimum energy (MIN) Maximum energy (MAX) Standard deviation (SIGMA) Coefficient of variation (CV%)
PN10190T
19
Page 20
F. After recalling the statistic al inf or mation , the fla shing STA TS a nnunc ia tor indica tes that th e S310
is ready for a new run.
A. Press the SELECT button to start a new run (each time a new run begins, data from the previous
run is lost) or press the CANCEL button to return the power meter to the mode of operation in effect prior to statistics mode.

9. External Trigger

The automatic trigger threshold of the S310 is 7 % of full scale. The external trigger function is designed for single pulse energ y measurement with a calorimeter. It does not improve the trigger circuit of the S310 with pyroelectric detectors.

To Perform a Transfer Calibration

You can transfer a calibration from a calori meter to a Vector pyroelect ric detector using the Tran sfer Calibration function of the S310. To do this, please refer to the Group Settings for Vector Pyroelectric Detectors, parag raph #G. You must follow the Group Settings steps to get into the Transfer Calibration function. This function allows you to adjust the output sensitivity of your Vector pyroelectric detector in combination with your S310 meter in order to match the average power reading from the S310 to that of a NIST certified system. Typically, a 50/50 beam splitter is used with the Vector pyroelectric detector to be calibrated in one beam path and the NIST certified calorimeter in the other beam path.

HR Battery Installation

A 9 Volt alkaline battery comes installed with all HR Series detect ors. Alwa ys select the OFF positi on of the power switch when the detector is not in use to spare the lifetime of the battery. Repla ce the battery when the low battery LED indicator lights up.
Figure 4. HR Detector Battery Access
PN10190T
20
Page 21
Refer to Figure 4. Remove the two slotted 4-40 binder head screws located on the underside of the detector. Pull off the outer housing to expose the battery. Remove the used battery from the battery holder and snap in the new battery. Slide the outer housing back in to place and secure with the screws.
Do not touch the delicate pyroelectric crystals in the HR Series detectors. The y should only be cleaned with a stream of clean air, nitrogen or CO2.

Set Electrical Time Constants for Model PHF02, PHF05 and PHF09

The pulse width switch, located on the rear of the PHF02, PHF05, and PHF09 detectors, selects one of three electrical time constants and should be set as follows:
S (Short Pulses): Select for pulse durations of 5 µsec. or less. Repetition rates up to 4
kHz can be measured.
I (Intermediate Pulses): Select for pulse durations of 50 µsec. or less. Repetition rates of up to
400 Hz can be measured.
L (Long Pulses): Select for pulse durations of 250 µsec. or less. Repetition rates of up to
80 Hz can be measured.
There are no pulse width settings on the P05 and P09 detectors which utilize a highly absorbing material on the crystal. The pulse duration considerations merely function in the determination of damage thresholds. Repetition rates up to 100 Hz using the P05 and 50 Hz using the P09 can be measured.

Correcting Pyroelectric Detector Operating Parameter s for Use at Different Wavelengths

Due to variability in the manufacturing process the absorption characteristics of the HD and P model detectors can vary in the UV region (190 to 400nm). Scientech r ecommends optica l calibra tion at 266nm if the detector is to be used in the UV region instead of re lying on an absorption ra te in the chart at the end of this manual.
All pyroelectric d etectors are cal ibrated at a spec ific wavelength and the detector’s output sensitivity is derived for that wavelength. The output sensitivity and calibration wavelength is recorded in the Operating Parameters section at the front of the manual and on the detector’s serial tag. When a pyroelectric det ector is used at a wavelength other than the calibration wavelength, its output sensitivity can be adjusted to compensate for the absorption rate at the new wavelength. The new output sensitivity is calculated as follows:
1. Find the absorption rate from the chart at the end of this manual for the calibration wavelength of
your pyroelectric detector.
2. Find the absorption rate for the wavelength where you will be working. PN10190T
21
Page 22
3. Determine the new output sensitivity using the following formula:
absorption rate of new wavelength x output sensitivity (V/J) = output sensitivity for new
absorpt ion rate of calibration wavelength wavelength from serial tag
This new output sensitivity can be entered into the S310 as described in Group Settings or when using the pyroelectric detector without a S310 power meter as discussed later in this manual.
Using S310 with Vector Series Calorimeters
The calorimeter sele cted needs to be t he appropriate mod el for the planned l aser measurements. Please familiarize yourself with the operating specifications which are given in the front of this manual.
The default Group Setting for calorimeters is Group #1. Either select Group #1 or configure another Group for calorimeters and select that Group.
Vector™ calorimeters are sensitive to all types of thermal input. Due to the handling of the calorimet er during setup and possible environmental temperature differences, thermal gradients may exist in the calorimeter. Allow the calorimeter to sit undisturbed for several minutes to, reach thermal equilibrium, before using.
Note: When using a 25mm calorimeter for measuring average power levels below 30mW and single pulse energy levels below 30mJ, a Scientech Model 360203A isoperibol™ Enclosure is highly recommended. The isoper ibol enclosure should not be used at average power levels above 30mW, and single pulse energy levels above 100mJ because heat build-up will occur.
Note: Large Ap erture calorimet ers and the Interface M odul es that th ey are c alibrat ed with a re matched sets and must be used together.
1. To Turn-On the Meter
Note: For the most accurate measurements possible, the S310 should be turned on and warmed up for 30
minutes.
Press the ON switch located in the rear pan el of the meter. T he S310 will imm ediately turn on with its operational state based on the last used detector Group Setting. If you purchased the S310 with one detector, this detector’s Group settings will be active and you are ready to take measurements. If you purchased more than one detector with the S310, you must choose the Group setting for the detector you are connecting to the S310 as described in step 4. The Group Setting for each detector purchased with your S310 is shown in the Group Setting section.
2. To Turn the Analog Needle On or Off (does not apply to the S310D)
To turn the analog meter on or off, press and keep holding the RANGE button down. Then release after the meter appears or disappears.
PN10190T
22
Page 23
Model
Vector 25mm
Vector 50mm
Large Aperture 100mm
Module*
Large Aperture 200mm
Module**
Mode
Power
Energy
Power
Energy
Power
Energy
Power
Energy
10mW
AUTO
10mJ
300mW
300mJ
300mW
300mJ
300mW
300mJ
3. To Zero the Analog Needle (does not apply to the S310D)
Refer to Figure 2. The black slotted button located just below the display allows screwdriver adjustment to set the analog needle to zero. This adjustment should be made before connecting a calorimeter.
4. To Connect a Calorimeter
Note: Only one detector should be plugged in at any time.
Refer to the drawing of the S310 rear panel in Figure 1. A 3 meter mini-DIN type cable with "D" shaped connectors comes with Vector c alorimeters. For large aper ture calo rimeters, th e interconnect cables for both the calorimeter and power meter are hardwired to the interface module. One of input connectors on the rear panel of the S310 is labeled "Astral" for the connection of Vector (previously called Astral) calorimeters and including the large aperture (100 & 200mm apertures) calorimeters. Note that the flat side of the DIN type cables should be oriented up when plugging in to the S310.
5. To Select a Group
A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again.
B. To select the appropriate Group, press the SELECT button when the Group you desire appears in
the display (failure to press the SELECT button within 3 seconds will activate the Group shown in the display and return the S310 to its operational state).
6. To Select a Range
Note: AUTO range is not available in the energy mode for calorimeters
A. Press the RANGE button. On the first press of the RANGE button, the analog scale (if active) will
disappear. Then briefly press the RANGE button each time another range is desired.
B. To select a range, press the SELECT button when the range you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the range shown in the display and return the S310 to its operational state). The ranges available for calorimeters are specified in Table 3.
Table 3. S310 Ranges for Vector™ Calorimeters
Range
PN10190T
100mW
1W
10W
100mJ
1J
10J
3W
30W
AUTO
3J
30J
with PN10735 Interfac e
3W
30W
AUTO
3J
30J
with PN10747 Interface
3W
30W
AUTO
3J
30J
23
Page 24
* With a PN10748 – 10X attenuator, the actual power or energy is 10 times the displayed value up to 50W
or 150J.
** With a PN10769 – 10X attenuator, the actual power or energy is 10 times the displayed value up to
100W or 300J.
Note: When using PN10748 or PN10769, 10X attenuators, you may want to use an attenuation factor of 10 so the displayed reading on the S310 is correct. However, the range designation may be incorrect if the power or energy reading exceeds the upper limit of the ran ge. See the Group Settings section for details.
7. To Select a Mode
The modes available for the calorimeters are: Average Power (W AVG) or Energy (J).
A. To Measure Average Power (W AVG):
Note: The speed-up circuit (differentiator) in watts mode is controlled by software and was adjusted at the factory to accom modat e the t ype of calorim eter bein g used. You can ch ange the sp eed-up setting to your preference. For changing a Group setting, refer to the Group Setting discussion in the previous section on Group Settings for Vector or Lar ge Aper ture Cal orim eters . The speed-up circuit is not active in AUTO range.
Note: The average power mode displays the average power (watts) of repetitively pulsed lasers or continuous wave lasers.
i. Press the MODE button repeatedly until the “W AVG” annunciator appears.
ii. Press the SELECT button (or simply wait 3 seconds) and the S310 will automatically enter
into its operational state in the average power mode.
iii. Press the CANCEL button to zero the display.
iv. The average power will be displayed.
A. To Measure Energy (J)
Note: Calorimeters can only measure single shot energy pulses (time between pulses is dependent on the calorimeter delay “CD” setting in the Group Setting for calorimeters [see previous section on Group Settings for Vector or Large Aperture Calorimeters]). With the calorimeter delay entered, the S310 will display the "trig" annunciator and the single pulse energy after the first pulse is delivered. The "trig" annunciator will then disappe ar after the calorimeter de lay time has elapsed prompting you to fire another pulse. Do not fire another pulse until the annunciator disappears. If you do, the S310 resets the time delay and ignores the sequential pulse altogether.
i. Press the MODE button (repeatedly, if necessary) until the “J” annunciator appears in the
display.
PN10190T
24
Page 25
ii. Press the SELECT button (or simply wait 3 seconds) and the joules mode will automatically
be activated.
iii. The energy level of each laser pulse will be displayed on the LCD.
8. To Measure a Statistical Run of Single Shot Energy Pulses
Note: Calorimeters can only measure single shot energy pulses (time between pulses is dependent on the calorimeter delay “CD” setting in the Group Setting for calorimeters [see previous section on Group Settings for Vector or Large Aperture Calorimeters]). With the calorimeter delay entered, the S310 will display the "trig" annunciator and the single pulse energy after the first pulse is delivered. The "trig" annunciator will then disappear after the calorimeter delay time has elapsed prompting you to fire another pulse. Do not fire another pulse until the annunciator disappears. If you do, the S310 resets the time delay and ignores the sequential pulse altogether.
Note: Do not use AUTO range when making a statistical run.
Note: Do not go from the Average Power Mode to Stats since the range will be too high. Select the range
manually.
Note: Each time a new stats run begins, data from the previous run is lost. If statistical data is to be saved, it must be done though the digital interface.
Note: To exit the statistical mode at any time, press the CANCEL button.
The statistics mode will collect data on a pulse population of up to 1000 pulses. At your prompting, the power meter will displ ay the number of pulses delivered, average energy, minimum energy, maximum energy, standard deviation, and coefficient of variation. When the statistics mode is selected, the energy mode is automatically activated regardless of the mode previously selected . S el ect t he ap pro pri ate range for the pulse energy level to be measur ed. It is very importan t to select the most appropriate range. If you have selected a manual range and the laser pulse(s) has overflowed the maximum energy of the range, OF will be displayed when the data is recalled. You should then select a higher range.
To enter into the statistics mode:
A. Press the STATS button. The number of pulses in the last statistics run will appear in the display.
B. Use the RANGE (count up) and MODE (count down) buttons to change the display to the desired
number of pulses to include in the statistics run (up to 1000).
C. Press the SELECT button to enter the pulse population to memory.
D. Press the SELECT button to begin the run. The power meter will automatically stop once the data
has been collected.
E. P ress the STATS button to recall the data to the display. The STATS button must be pressed to
recall each of the following statistical calculations:
PN10190T
25
Page 26
Number of pulses collected Average energy (AVG) Minimum energy (MIN) Maximum energy (MAX) Standard deviation (SIGMA) Coefficient of variation (CV%)
F. After recalling the statistic al inf or mation , the fla shing STA TS a nnunc ia tor indica tes that th e S310
is ready for a new run.
G. Press the SELECT button to start a new run (each time a new run begins, data from the previous
run is lost) or press the CANCEL button to return the power meter to the mode of operation in effect prior to statistics mode.
9. External Trigger
The external trigger input is located on the rear panel of the S310 as shown in Figure 1. The external trigger function is designed to ensure the entire pulse energy of a single pulse is captured by the S310 power meter. An extern al tri gger “awakens” the S310’s circuitry so that it is ready for the arrival of the laser pulse. The external trigger pulse must have the following characteristics:
1. 2.5 volts min. to 5 volts max.
2. A minimum pulse width of 10 µsec. to a maximum of the laser pulse repetition rate.
3. A maximum rise/fall time of 500 nsec.
4. Delivered 1 to 3 milliseconds prior to the laser pulse.
The trigger is generated on the rising edge of the incoming pulse.
The following are viable external triggers.
The following pulses will not trigger the S310.
PN10190T
26
Page 27

Correcting FX Calorimeter Operating Parameters for U se at Different Wavelengths

Note: FX calo ri meters are no long er avai labl e f or sal e. Th is i nformati on i s pro vided fo r users w ho previously purchased FX calorimeters to understand how they operate with Vector power meters.
Note: Due to variability in the manufacturing process the absorption characteristics of th e FX model
detectors can vary in the UV region (190 to 400nm). Scientech r ecommends optica l calibra tion at 266nm if the detector is to be used in the UV region instead of re lying on an absorption ra te in the chart at the end of this manual.
Scientech calorim eters in gene ral have a flat response to al l wavelengths within their sp ecified spectr al response. FX and FXX calorimeters are an exception to that rule and are calibrated at a specific wavelength by adjusting the ca lor imeter’s gain circuitry for that wavelength. The calibration wavelength is recorded in the Operating Parameters section at the front of the manual and on the detector’s serial tag. When a FX or FXX calor im eter i s used at a w avel ength o th er t han t he c al i brat io n wavelength, the power meter’s displayed value can be adj usted to compens ate for t he absorp tion rate at t he new w avelength by using an attenuation factor. The attenuation factor is calculated as follows:
1. Find the absorption rate from the chart at the end of this manual for the calibration wavelength of your
calorimeter.
2. Find the absorption rate for the wavelength where you will be working.
3. Determine the attenuation factor using the following formula:
absorption rate of calibration wavelength = attenuation factor
absorption rate of the new wavelength
The attenuation factor can be entered into the S310 as described in the Group Setting section.
PN10190T
27
Page 28

Analog Output

The analog output is an uncalibrated output accessible via the 50 ohm terminated BNC connector located on the power meter's r ear panel (see Figure 1). Although the analog output voltage level is uncalibrated, it is representative of the power and energy readings displayed on the S310. The analog output voltage level is approximately 3 volts full scale. For high accuracy, you must determine the relationship between the analog output voltage level and the power and energy readings on the S310 display.
With a calorimeter connected to the S310 in watts mode, the analog output voltage readings will rise to a steady level when power is applied to the calorimeter. After the voltage readings become stable, note the analog output voltage reading and the watt reading on the S310 display. All subsequent voltage readings will have the same V/W value s. In the joules mode, the voltage w ill rise to a peak value relatively quickly (seconds) followed by a rapid return to baseline. The peak voltage is representative of the joules value. Again, this can be determined by noting the peak voltag e value on the analog output, and the joules reading on the S310 display.
With a pyroelectric det ecto r connected t o the S31 0, determi ne the relati onsh ip between t he peak vo ltage value and the energy (joules) or power (watts) reading on the S310 display in the same manner. The analog output with the various detectors and operating modes should appear as follows:
A. Calorimeter Watts Mode
Compare voltage to the watt reading displayed on the S310 to determine V/W constant to use when looking at the analog output.
In power (watts) mode, the analog output is nominally 3 V full scale for each range. Therefore, to calculate watts from the analog output voltage:
W = V/S
where:
V = analog output voltage (V) S = analog output sensitivity (V/W) 3 V/range (W) (for example, on the 10 mW range, S 3 V/.010 W = 300 V/W)
B. Calorimeter Joules Mode
PN10190T
28
Page 29
Compare the peak voltage to the energy (joules) reading displayed on the S310 to determine V/J constant to use when looking at the analog output.
In the energy (joules) mode the analog output is an amplified voltage signal proportional to the voltage generated by the calorimeter thermopile. To calculate energy (joules) from the analog output use the following equation.
J = 0.3(Vpk)(TC)
(S)(R)
where:
Vpk = peak voltage from the analog output
TC = calorimeter time constant from the calorimeter serial tag
S = calorimeter output sensitivity as follows:
0.5 V/W for 25 mm models
0.1667 V/W for 50 mm models and large aperture calorimeters
R = power meter range multiplier as follows:
10.0 for the 10 or 30 range
100.0 for the 1 or 3 range 1,000.0 for the .1 or .3 range 10,000.0 for the .01 or .03 range
C. Pyroelectric Watts and Joules Modes
Compare peak voltage to energy (joules) or power (watts) reading displayed on the S310 to determine V/J or V/W constant to use when looking at the analog output.
PN10190T
29
Page 30
In power (watts) and ener gy (joules) mo de, th e peak volt age of a laser p ulse is directl y proportional to t he power energy level:
W = Vpk/S
where:
Vpk = peak voltage from the analog output
S = analog output sensitivity (V/W) 3 V/range (W) (for example, on the 3mW range, S 3 V/.003 W = 1000 V/W)
J = Vpk/S
where:
Vpk = peak voltage from the analog output
S = analog output sensitivity (V/J) 3 V/range (J) (for example, on the 3 mJ range, S 3 V/.003 J = 1000 V/J)
CALIBRATION OF VECTOR CALORIMETERS USING ELECTRIC SUBSTITUTION HEATERS
For Vector™ calorimeters the electric substitution heating option must be ordered and installed at the factory when th e calorimeter is purchased. It cannot be retrofitted to a calorim eter at a later time. To calibrate using electric substitution heating proceed as follows.
PN10190T
30
Page 31
Figure 5. Calorimeter Ci rcu it Board
A. Remove the screws h olding the calorimeter 's ID tag and remove the plat e to expose the circuit
board as shown in Figure 5.
B. Connect the calorimeter to the power meter, turn on the power and let the system equilibrate. C. Connect a DVM to the test points labeled SUB and HTR on the calorimeter circuit board. D. Measure the resistance of the substitution heater making sure to subtract the resistance of the patch
cables from the total resistance measurement. Compare this resistance to Rc in the calibration data in the front of the manual. The two should agree within 2%. If not contact Scientech.
E. Remove the DVM. Connect a power supply to the SUB and HTR test points and connect the
DVM to monitor the power supply.
F. Set up the power meter in the Watts Mode and the 10W range for 25 mm calorimeters or the 3W
range for 50 mm calorimeters.
G. Apply Vh volts, stated in the calib ration da ta you receiv ed with the ca lorimeter, to the substitutio n
heater.
H. If needed, adjust the calibration trim pot, R4 on the calorimeter circuit board, until Wh Watts, from
the calibration data, is displayed by the power meter.

CALIBRATION OF LARGE APERTURE CALORIMETERS USING ELECTRIC SUBSTITUTION HEATING

Electrical substitution heating is a standard feature of large aperture calorimeters.
PN10190T
31
Page 32
Figure 6. Large Calorimeter Connector Panel

Calibration with an Int erf ac e Modu le an d S310 Power Me ter

A. Connect a DVM to the white jacks of the calorimeter. Refer to Figure 6. B. Measure the resistance of the substitution heater making sure to subtract the resistance of the patch
cables from the total resistance measurement.
Note: When measuring the substitution heater resistance or a 200 mm calorimeter, R1 and R2 must be connected together in series.
Compare this resistance to Rc in the calibration data in the front of the manual. The two should agree within 2%. If not contact Scientech.
C. Remove the DVM. Connect a power suppl y to the white jacks and connect the DVM to monitor
the power supply.
D. Set up the power meter in the Watts Mode and the 30W range. E. Remove the screws holding the interface module’s ID tag and remove the plate to expose the
circuit board. Refer to Figure 5.
F. Apply Vh volts, stated in the calibration data you received with the calorimeter, to the substitu tion
heater.
G. If needed, adjust the calibration trim pot, R4 on the calorimeter circuit board, until Wh Watts, from
the calibration data, is displayed by the power meter.

Calibration with out an Interface Module and S310 Power Meter

For this procedure you will need to make an adapter cable to go from the calorimeter’s DIN connector to the DVM. The voltage output is on pin 1 of the DIN connector and should be connected to the posit ive side of the DVM. Ground is on pin 3 and should be connected to the negative side. Pin 2 is not used. Refer to Figure 6.
A. Connect a DVM to the white jacks of the calorimeter. Refer to Figure 5.
PN10190T
32
Page 33
B. Measure the resistance of the substitution heater making sure to subtract the resistance of the patch
cables from the total resistance measurement.
Note: When measuring the substitution heater resistance or a 200 mm calorimeter, R1 and R2
must be connected together in series.
Compare this resis tance to Rc in the calibration data in the front of the manual. The two should agree within 2%. If not contact Scientech.
C. Calculate the voltage equivalent to laser power using the following formula:
V = (Rc x C x W)
1/2
where:
V = voltage applied to the heater coil
Rc = substitution heater resistance from step B
C = Cal coefficient 360401 = 1.018 360801 = 1.000 380401 = 0.974 380801 = 1.008 380402 = 1.024 380802 = 1.008 384UV5 = 1.021 388UV5 = 1.002
W = desired laser power in watts
D. Connect the DVM to the calorimeter’s DIN connector. E. Apply the calculated voltage (V) to the electrical substitution heater. F. Record the voltage reading of the DVM (Vc). G. Calculate the calorimeter’s output sensitivity (S) as follows:
S = Vc/W
where:
S = calorimeter’s output sensitivity
PN10190T
33
Page 34
Vc = voltage output from the calorimeter in mV
W = desired laser power output.
The measured sensitivity should be ± 3 % of the calorimeter’s original sensitivity value.

DETECTOR OPERATION WITHOUT A POWER METER

Pyroelectric D et ector Operation without Power Meter

A. Standard and SP Models
Standard and SP model pyroelectric detectors can be operated with a 1MΩ input oscilloscope. The peak
voltage shown on the oscilloscope can be divided by the V/J output sensitivity of the detector to calculate energy.
B. HR Models
HR pyroelectric detectors can be operated with a 50Ω input oscilloscope. The peak voltage shown on the
oscilloscope can be divided by the V/mJ output sensitivity of the detector to calculate energy.
Vector
Vector (previous called Astral) calorimeters are powered up by the power meter. To use a Vector calorimeter with ou t a S cientech power met er, but with a volt meter or chart recorder, you must app l y + / ­8VDC to the mini DIN connector as shown in Figure 7. The voltage output of the calorimeter, from pin 8, should be connected to the positive side of the DVM or chart recorder. All 3 of the grounds should be tied together at the negative side. Pins 2 and 3 are not used.
When large aperture calorimeters are used without an power meter their interface module is not used. The output of the calorimeter is connected directly to the DVM or chart recorder. Large aperture calorimeters do not require any power. The voltage output is on pin 1 of the DIN connector and should be connected to the positive side of the DVM or chart recorder. Ground is on pin 3 and should be connected to the negative side. Pin 2 is not used. Refer to Figure 6.
Calorimeter Operation with ou t Power Met er
A. Cable Requirements
Figure 7. Calorimeter mini DIN Connector
PN10190T
34
Page 35
Operation of Vector Calorimeters with a D igital Volt Meter
Note: Whenever a l arge ap erture cal orimete r is used without a power meter the interface module is not
required.
The calorimeters may be used with any digital volt meter (DVM) capable of reading 5 volts full scale. Connect the output of the calorimeter to the DVM.
A. Select the DC volts mode. B. Direct the laser beam on to the absorbing surface of the calorimeter. C. When the displa y of the DVM has stabilized (about 2 minutes), calculate the laser power using the
formula:
W = V/S
where:
W = Laser power in watts
V = Voltage reading of the DVM in volts
S = Sensitivity of the calorimeter from page 2.
Operation of Vector Calorimeters with an Analog Recorder
Note: Whenever a large apertur e calorim eter is used with out a power meter the interface module is not
used.

Calorimeter Response:

The response of a calorimeter to a single pulse input as displayed by a chart recorder appears below.
The output voltage from a chart recorder can be converted to wattage at any time by:
PN10190T
35
Page 36
W = V/S, Wi = Vi/S V = Chart recorder voltage level in mV S = Calorimeter sensitivity in mV/W
The total energy (E) in the pulse can be found by integrating the instantaneous wattage over time:
E = W(t) dt
0
The following methods may be used to compute the total integrated energy:

Numerical Integration:

Finding the area under the curve in figure 7 is the equivalent procedure for determining pulse energy.
Choose an appropriate time interval, dt, and perform the summation:
N N E = Wixdt = (dt/S) Vi I=1 i=1
The error caused by this procedure is:
N dE = (dt/S) Σ dVi i=1
The error, in theory, is only dependent upon the value of dVi, that is the cumulative random error of Vi. This number should approach zero if data is carefully taken. The accurac y is also increased if the time interval, dt, is minimized. Numerical integration can yield accurate results, but is a tedious task.

Initial Voltage Interpolation:

A method used to eliminate the tedious nume rical i ntegration ta sk is to project the thermal decay envelope on to the voltage axis, determine the 1/e decay time constant T, and estimate the total energy value (E):
E = (Vo/S) x T
The change from thermal absorption to thermal transport phenomena near the peak causes difficulty in accurately projecting the envelope on to the voltage axis introducing an error, dVo. Further, the determination of the time constant T, introduces another error, dT. The total error is the sum of the two errors.
dE = (Vo/S)dT + (T/S)dVo
The difficulty in eliminating the potential error makes this method typically less accurate than numerical integration, but much faster in application.
PN10190T
36
Page 37

Peak Voltage Estimate:

The peak voltage method requires using an independent determination of total energy and referencing it back to the peak voltage value, Vp.
For a given pulse, use the numerical integration method to obtain E. Note the peak voltage, Vp. Compute the value, F
F = E/Vp
For the next pulse compute the total energy: E = F x Vp
The error in using this method yields: dE = FdVp + VpdF
The accuracy of this measurement depends upon the error in the original calibration, dF, and the error in the peak voltage dVp. A careful numerical integration yields a value for dF near zero. The value of dVp can be minimized by maintaining the geometry of the system (i.e. beam intensity, beam profile, wavelength and environment) during operation to be the same as during calibration. Under controlled circumstances, the peak method accuracy usually falls between the numerical integration and initial voltage interpolation methods.

REMOTE INTERFACE LANGUAGE

The remote interface language for the S310 is compatible with standard IEEE488.2 and it also works with the RS-232 remote interface.
A. RS-232 Connector
The RS-232 connector is a 9 pin subminiature connector on the instrument rear panel. The pin out and pin descriptions are shown below:
Pin 1 Unused Pin 2 Data In (RXD) Pin 3 Data Out (TXD) Pin 4 DTR (is generated) Pin 5 Ground Pin 6 DSR (is ignored) Pin 7 RTS (is marking) Pin 8 CTS (is evaluated if requested) Pin 9 Unused
PN10190T
37
Page 38
B. RS-232 Specifications
Type: EIA-RS232C Method: Half-duplex, Asynchronous Transmission: Bi-directional Format: 300, 1200, 2400, 9600, 19200 baud rate selectable Data bits: 7 Parity bit: Even, Odd, or None Stop bit: 1 Code: ASCII Total no. of bits: 10
* The start bit counts for the first bit. Therefore, if you choose no parity you must have 2 stop bits.
C. Remote Interface Language Syntax
Remote interface m essages consis t of zero or mor e commands or queries, separated by semicolons and terminated by a linefeed (IEEE488) or a carriage return (RS232). A command or query consists of a command or query header followed by zero or more arguments separated by commas. Messages must be less than 75 characters.
Example:
cmd1 arg1;cmd2arg1,arg2;...cmdN arg1
The queries RPT?, SND?, COL?, and *OPC?, and the commands COL and *OPC are intended to be placed as the last command in the message. Placing them elsewhere will not result in harm to the instrument, but it may produce results which seem unusual.
Queries which have not finished will be aborted by the receipt of additional commands or queries. This will result in Query Errors in the IEEE488 interface. If a RP T?, COL?, or SND? query is immediately followed by another command, it is likely tha t no data will be transmitte d. If COL is followed by another command, statistics gathering will be halted unless the command *WAI appears between the two commands.
D. Remote Interface Language Common Commands and Queries
Most commands and queries may be used with either the RS232 interface or the IEEE488 interface. Some commands, however, work only with the IEEE488 interface. Some commands require that a password has been entered. A few commands are archaic, but were left in because no harm was done, and they may be needed again if multi-channel units are constructed again.
The commands which work in both interfaces are presented first , followed by the password protected commands peculiar to the IEEE488 interface, followed finally by the archaic commands. In the descriptions which follow, the command (or query) will be presented first, followed optionally by one or more arguments, separ ated by commas. When the vertical bar "|" is used i n an argument, it denotes the word "or", and signifies that one and only one of the items separated by bars ma y be inserted into the
PN10190T
38
Page 39
space of the argument. For example, the command: XXXA|B,C has two arguments, the first of which may be `A' or `B', and the second of which is t he letter `C'. W here a space ch aracter is required by the syntax, it is represented with an underline character.
Commands in this group are available from both interfaces: *IDN?
This query takes no arguments, and returns a comma separated collection of four strings, describing respectively the manufacturer of the instrument (Scientech, Inc.), the model number (S310), the serial number, and the firmware version number.
*SAV_<dec num>
This is the IEEE488.2 common command for saving the instrument setup. The number may be an integer from 1 through 4. When used from either remote interface, it saves the current configuration to the numbered save area. The next time the instrument is powered up or the *RST command is issued with the same save area as the default, or the RCL command is executed for the same save area, the configuration will be restored to the values saved.
SENS_PYRO|CAL|PHR|ULTRA
Sets the detector type.
SENS?
Returns the detector type (PYRO|CAL|PHR|ULTRA).
SND?
The SND? query causes the instrument to send the next reading to the remote interface. When using a calorimeter in the joules mode, the SND? query must be sent to the S310 before the calorimeter is pulsed by the laser.
RPT?
This query causes the remote interface to begin sending a sequence of comma-separated readings. A new reading is sent each time one is taken by the instrument. This activity will continue until the remote interface is interrupted by a new command or query, or the instrument is turned off.
COL _<dec num>
This command causes the instrument in energy mode to begin collecting readings for statistical analysis indicated by the response COLLECTING... This command continues until <dec num> data points are collected, or until it is interrupted by another command. If it is interrupted, no statistics are computed. If it terminates normally, a carriage return and line feed occurs, and it computes values for the mean, max, standard deviation, and coefficient of variation of the data points collected. Those values may then be
PN10190T
39
Page 40
PHR 2, 5, 9 mm
Long Mode
1
1.000 m
300.0 m
150.0
3.000 m
300.0 μ
2
10.00 m
300.0 m
150.0
30.00 m
300.0 μ
3
100.0 m
300.0 m
150.0
300.0 m
300.0 μ
4
1.000
3.000
150.0
3.000
3.000 m
5
10.00
30.00
150.0
30.00
30.00 m
PHR 2 mm
Short/Int Mode
PHR 5 mm
Short/Int Mode
PHR 9 mm
Short/Int Mode
PHR 5 mm
Painted
PHR 9 mm
Painted
1
3.000 μ
3.000 μ
3.000 μ
3.000 μ
3.000 μ
2
30.00 μ
30.00 μ
30.00 μ
30.00 μ
30.00 μ
3
300.0 μ
300.0 μ
300.0 μ
300.0 μ
300.0 μ
4
300.0 μ
3.000 m
3.000 m
3.000 m
3.000 m
5
300.0 μ
30.00 m
30.00 m
3.000 m
30.00 m
accessed with the queries mean?, max?, sigma?, and cv?. The individual data points may be dumpe d with the dump? query.
RANGE_<dec num>|auto
Sets the range to the decimal number supplied, or to the closest legal range allowed for the current detector and configuration. If the word auto is supplied, this command activates the auto-range feature i f it is supported for the current detector and configuration.
RANGE?
The RANGE query returns the decimal number between 1 and the maximum range of the instrument. The meaning of the number varies, depending upon the type of detector, and the configuration of the instrument. (Generally speaking, there are 5 ranges, though not all of them are used for every detector).
Table 4. S310 Range Selection Using Remote Interface Language
Range Cal 25 mm Cal 50 mm Ultra Pyro 25/50 mm
Range
MEAN?
If stat istics have been collected from the front panel, or by the COL or COL? commands, this command returns the mean value of the collection. If no statistics have been collected, it returns to zero.
MIN?
If statistics have been co llected from the fro nt panel, or b y the COL or COL? commands, this command returns the minimum value of the collection. If no statistics have been collected, it returns to zero.
MAX?
If statistics have been collected from the front panel, or by the COL or COL? commands, this command returns the maximum value of the collection. If no statistics have been collected, it returns to zero.
PN10190T
40
Page 41
SIGMA?
If statistics have been collected from the front panel, or by the COL or COL? commands, this command returns the standard deviation of the collection. If no statistics have been collected it returns zero.
CV?
If statistics have been collected from the front panel, or by the COL or COL? commands, this command returns the coefficient of variation of the collection. If no statistics have been collected it returns zero.
PULSES?
Returns the number of pulses collected in the last run.
DUMP?
If statistics have been collected from the front panel, or by the COL or COL? commands, this command returns all of the data values in the collection. If no statistics have been collected, it does nothing.
BAUD_<dec num>
If the <dec num> takes any of the values 300, 1200, 2400, 9600, or 19200; or any of their abbreviations 3, 12, 24, 96, 192, this command sets the RS232 baud rate to the corresponding value.
PARITY EVEN|ODD|NONE
Sets the parity of the RS232 interface to even parity, odd parity, or no parity, as specified.
HANDS XON|CTS|NONE
Sets the handshake method of the RS232 interface to XON/XOFF,CTS, or NONE as specified.
IO RS232|IEEE
On units with both IEEE and RS232 interfaces, this command makes the named interface the S AVED active interface. The next time the instrument is powered up, or its configuration is restored from the configuration save set active when this command was issued, the specified interface becomes the acti ve remote interface.
*RST
This is the IEEE488.2 common command by the same name. When executed from the RS232 interface, it has the effect of restoring the saved instrument configuration from the current save area. It has the additional function in the IEEE488 interface of forcing the interface into the OCIS state and the OQIS state.
METER ON|OFF
PN10190T
41
Page 42
Activates or deactivates the analog meter.
MODE VOLTS|ENERGY|AVGE|AVGP
If the specified mode is valid for the detector in use, the software changes mode to the one specified.
MODE?
Returns the instrument's operating mode. The possible responses are VOLTS, ENERGY, AVGE, and AVGP.
*TST?
This is the IEEE488.2 common query. From either interface, it causes ROM checksum, and non­destructive RAM test to be run. If the tests succeed, this query returns zero. If the ROM test fails, a 1 is returned. If the RAM test fails, a 2 is returned.
*OPC?
This is the IEEE488.2 common query. From either interface, it waits until no overlapping command is in progress, and then it returns a 1.
*WAI
This is the IEEE488.2 common command. From either interface, it causes the command processor to wait until any overlapping command is finished before continuing to process commands.
CLR
This command halts RPT?, SND?, or COL? commands in progres s. I t a lso te rmin ate s sta tistics gathering started by the COL command. This command is issued internally whenever a carriage return that is not preceded by a command is typed into the RS232 interface. When the IEEE488 interface receives a linefeed terminated message with no commands, it too generates the CLR command.
*RCL_<dec num>
This is the IEEE488.2 common command. It takes the single numeric argument [1-4], and restores the instrument configuration to that configuration stored in the corresponding save area. It also places the IEEE interface into the OCIS and OQIS states.
COUNT_<dec num>
Sets the number of pulses making up each average in average energy mode.
COUNT?
Returns the number of pulses making up each average in average energy mode.
PN10190T
42
Page 43
CALIB_<dec num>
Sets the calibration constant (Volts/Joule) for a pyro detector, (Volt/Watt) for an Ultra Series detector or the time constant for a calorimeter. Note: This command is context sensitive; the detecto r type mus t be set by the SENS command before this command is executed.
CALIB?
Returns the constant for a pyro detector, or the time constant for a calorimeter. Note: This command is context sensitive; the returned value depends on the detector type set by the SENS command.
ATTEN_<dec num>
Sets an attenuation factor for the current detector.
ATTEN?
Returns the attenuation factor for the current detector.
ZERO
This command zeros the power baseline for a calorimeter.
PSWD_<dec num>
Enables password protection commands if the password is entered correctly.
PSPEED HF|BL
Sets the electronic response speed according to the repetition rate capabilit y of the pyroelectric detector type - a "black" coated absorbing crystal, or uncoated "high frequency" absorbing crystal.
PSPEED?
Returns the pyro speed, possible responses are BL and HF.
CSPEED_<dec num> Sets the watts mode display response speed for calorimeters and Ultra Series detectors.
CSPEED? Returns the watts mode display response speed for calorimeters and Ultra Series detectors.
CDELAY_<dec num>
Sets the "Calorimeter Delay" function of the meter.
PN10190T
43
Page 44
CDELAY?
Returns the "Calorimeter Delay" function of the meter.
E. IEEE488 Specific Commands
These commands may be used only from the IEEE488.2 interface. They are all members of the collection of the so-called "common Commands" described in the IEEE488.2 standard.
*CLS Clears the Standard Event Status Register and forces the device into Operation Complete Command Idle state.
*ESR?
Returns a decimal number which is the value of the Standard Event Status Register. Reading the register clears it.
*ESE_<dec num>
Sets the bits of the Standard Event Status Enable Register to the binary representation of the decimal integer supplied.
*ESE?
Returns a decimal number representing the contents of the Standard Event Status Enable Register.
*SRE_<dec num>
Sets the bits of the Standard Request Enable Register to the binary representation of the decimal integer supplied.
*SRE?
Returns a decimal number which represents the contents of the Service Request Enable Register.
*OPC
Sets the "Operation Complete" event bit in the Standard Event Status Enable Register when pending device operations have been completed.
*STB?
Returns a decimal number which is the value of the IEEE488.1 status byte and the Master Summary Status message.
PN10190T
44
Page 45

SCIENTECH CALIBRATION SERVICE

Scientech recommends that a complete calibration be performed annually to verify system accuracy. Please visit Scientech’s website at www.scientechinc.com to obtain an RMA (returned material authorization) number and complete the form noting that you are requestin g calibration of your detector. You may also contact our Product Service Department at (800)525-0522 or (303)444-1361 or Fax (303)444-9229 or email inst@scientechinc.com. Be prepared to provide model number, serial number along with contact information. to arrange for a NIST traceable, factory calibration. Scientech calibrates the detector with its power meter for the same price.

LIMITED WARRANTY

Scientech warran ts and r epres ents t hat th e las er p ower m easur ement p rodu ct w ill b e free f rom d efe cts in design, materials and workmanship and conform with applicable Scientech product specifications for a period of three (3) years. The product warranty period begins on the date of shipment from Scientech. Scientech warrants that its products shall conform to applicable Scientech specifications and drawings and will meet all the functional and performance requirements when properly installed, operated, and maintained in accordance with Scientech’s operating manual. Warranty does not extend to any Scientech products that have been subjected to misuse, abuse, or accidents, or improper installation, maintenance or applications, repaired by unauthorized personnel, or Products in which the tamper proof sticker has been removed or broken.
During the warranty period, Scientech will repair, or at its option replace at no charge, components that prove to be defective. The product must be returned, shipping prepaid, to Scientech's authorized repair facility. Products repaired by Scientech’s authorized repair personnel/facilities will be warranted against defects in the repaired component and workmanship for a period of 365 days from the date of shipment of the repaired Product.

RETURN MATERIAL PROCEDURE

Should it become necessary to return any product to Scientech for any reason including calibration, please visit Scientech’s website at www.scientechinc.com to obtain an RMA (returned material authoriz ation) number and complete the form. You may also contact our Product Service Department at (800)525-0522 or (303)444-1361 or Fax (303)444-9229 or email inst@scientechinc.com. Be prepared to provide model number, serial number, and a description of the problem along with contact information. Frequently we can provide self-help information which will eliminate the need for returning the product.
If product return is required, please pack the items in the original box and packing material. As an alternate, place the equipment in a snug-fitting box, and then pack that box in a larger box with at least four inches of packing material. Scientech does not assume responsibility for products damaged during shipping and shipping damage will not be treated as a warranty repair. Please include a point of contact, email address, and phone number of the person we should contact regarding repair questions.
Normally, products are repaired and shipped within five (5) business days following receipt of the product at the authorized service facility. The repair turn-around time could vary depending on the workload.
Shipping Address: Scientech, Inc.
PN10190T
45
Page 46
Product Service Department 5649 Arapahoe Ave. Boulder, Colorado 80303 U.S.A.

DISPOSAL OF ELECTRONIC EQUIPMENT

Scientech recommends the following for disposal of electrical and electronic equipment:
1. The best option is to reuse the equipment in its entirety.
2. Where the equipment cannot be reused in its entirety, priority should be given to reuse of
its subassemblies and components.
3. Where reuse is not appropriate, electrical and electronic equipment, including batteries,
should be recycled according to local ordinances.
4. Waste electrical and electronic equipment should never be mixed with municipal waste.
PN10190T
46
Page 47

SCHEMATICS

PN10190T
47
Page 48
PN10190T
48
Page 49
PN10190T
49
Page 50
PN10190T
50
Page 51
PN10190T
51
Loading...