Thank you for choosing a Scientech Vector™ calorimeter detector. Scientech’s employees are pleased to
provide you with a product designed for years of reliable and accurate service. Please read this operating
manual before using your detector and power meter. This reference information will allow you to fully
understand the capabilities of the product. The detector is intended to be used only in the manner outlined
in this manual. Operation not within specifications for the product may cause product damage.
CE Mark Certification ......................................................................................................................................................... 4
Vector™ S310 Power Meter Specifications ................................................................................................................................ 4
Group Settings ............................................................................................................................................................................. 9
To Select a Group................................................................................................................................................................ 9
Group Settings for Vector Calorimeters ........................................................................................................................... 9
Group Settings for Vector Pyroelectric Detectors .......................................................................................................... 12
Using the S310 with Vector Pyroelectric Detectors ....................................................................................................... 14
9. External T rigger .......................................................................................................................................... 20
To Perform a Transfer Calibration ......................................................................................................................... 20
Set Electrical Time Constants for Model PHF02, PHF05 and PHF09 .................................................................. 21
Correcting Pyroelectric Detector Operating Parameters for Use at Different Wavelengths ............................................. 21
Using S310 with Vector Series Calorimeters ................................................................................................................. 22
Correcting FX Calorimeter Operating Parameters for Use at Different Wavelengths ...................................................... 27
Analog Output ................................................................................................................................................................... 28
Calibration of Vector Calorimeters Using Electric Substitution Heaters ............................................................................... 30
Calibration of Large Aperture Calorimeters Using Electric Substitution Heating ..................................................................... 31
Calibration with an Interface Module and S310 Power Meter .......................................................................................... 32
Calibration without an Interface Module and S310 Power Meter ..................................................................................... 32
Detector Operation Without a Power Meter .............................................................................................................................. 34
Pyroelectric Detector Operation without Power Meter ..................................................................................................... 34
Vector ™ Calorimeter Operation without Power Meter ..................................................................................................... 34
Operation of Vector Calorimeters with a Digital Volt Meter ......................................................................................... 35
Operation of Vector Calorimeters with an Analog Recorder ......................................................................................... 35
Calorimeter Re sponse: ........................................................................................................................................... 35
Initial Voltage Interpolation: ................................................................................................................................. 36
Peak Voltage Estimate: .......................................................................................................................................... 37
Remote Interface Language ....................................................................................................................................................... 37
Scientech Calibration Service .................................................................................................................................................... 45
Return Material Procedure ......................................................................................................................................................... 45
Disposal of Electronic Equipment ............................................................................................................................................. 46
All detectors are calibrated at a specifi c wavelen gth and the d etector’s op erating param eters are d erived
for that wavelength. This information is recorded below and on the detector’s serial tag. When a detector
is used at a wavelength other than the calibration wavelength some of the operating parameters may need
to be adjusted. For specific instructions please refer to the Operating Procedures section for the type of
detector you are using.
Power Meter Serial Number___________________
Calorimeter 1: Calorimeter 2:
Model No: ___________ Model No: __________
Serial No: ___________ Serial No: __________
Group No: ___________ Group No: __________
Calibration Wavelength ___________nm or µm Calibration Wavelength: __________nm
or µm
Output Sensitivity (S): ___________V/W Output Sensitivity (S): __________V/W
Time Constant (1/e): ___________sec. Time Constant (1/e): __________sec.
Calibration Temp: ___________°C Calibration Temp: __________°C
Calibration Humidity: ___________%R.H. Calibration Humidity: __________%
R.H.
Sub. Heater Resistance (Rc): ___________ohms Sub. Heater Resistance (Rc): __________ohms
Sub. Heater Voltage (Vh): ___________volts Sub. Heater Voltage (Vh): __________volts
Sub. Heater Wattage (Wh) ___________watts Sub. Heater Wattage (Wh): __________watts
Pyroelectric Detector 1:
Model No: __________
Serial No: __________
Group No: __________
Calibration Wavelength: __________nm or µm
Output Sensitivity:
Output Sensitivity:1 V/J (Scope)
Calibration Temp: __________°C Calibration Humidity: ___________%R.H.
Pyroelectric Detector 2:
Model No: __________
Serial No: __________
Group No: __________
Calibration Wavelength: __________nm or µm
Output Sensitivity:
Output Sensitivity:
Calibration Temp: __________°C Calibration Humidity: ___________%R.H.
The V/J is used t o calibrate the power meter to a known NIST traceable standard. This is the value the
meter must use.
The V/J (Scope) is the “true” Volts per Joule as seen at an oscilloscope.
1
__________V/J or __________V/mJ S_________ I__________ L_________
1
__________V/J or __________V/mJ S_________ I_________ L__________
V/J (Scope)
PN10190T
3
Page 4
ENVIRONMENTAL REQUIREMENTS
This product is intended for indoor use at altitudes up to 2000 meters, Pollution Degree 2 in accordance
with IEC 664 and transient over voltages according to Installation Categories (Overvoltage Categories)
II. Note that each of the above detectors will not pass the IEC 801 Publication, Part 3, Radiated
Electromagnetic Field Requirements. The system, meter and detector, is designed to measure radiation
within the test's radiation band. The detectors were held outside the radiate d elec tromagnetic f ield during
this test. It is up to the user to be aware of RF fields present during measurements and their effects if any
on those measurements.
CE Mark Ce rtification
All of the power meters listed in this manual have been certified for the European CE mark.
VECTOR™S310POWER METER SPECIFICATIONS
The Models S310 and S31 0D are identical to each other i n every respect except the an alog meter. The
S310 has both analog and digital displays. The S310D has only a digital display. All references to the
S310 are intended to include the S310D except where noted. Also, the words "indicator" and "meter" are
synonymous. Specifications are summarized in Table 1.
PN10190T
4
Page 5
Model
S310
S310D
Display
4 Digit backlight LCD with analog
4 Digit backlight, LCD
Operating Modes, Power or
Pyroelectric: energy, average energy,
Pyroelectric: energy, avg energy, power,
Pyroelectric Statistics Mode
Pulses collected, avg energy, min energy,
Pulses collected, average energy, min
Calorimeter max repetition rate,
The meter, detect ors, and accessories are sh ipped in custom packing ma terials. All packing materials
should be saved for future damage free shipments.
Before making any connections, verify that the power (VAC) requirement shown on the power entry
module is compatible with the actual AC power outlet to which the power meter will be connected. To
change the power meter’s voltage, proceed as follows:
1. Refer to Figure 1. Locate the power entry module and the fuse holder in the center of the module.
2. Remove the fuse holder by inserting a slotted screwdriver in the slot on the right side and prying
it out.
3. Slide the voltage selector out, flip it over and re-insert it into the fuse holder.
4. Re-insert the fuse holder into the power entry module.
Figure 1. S310 Rear Panel
QUICK SETUP
For detailed instructions for each type of detector, refer to the Operating Procedures section.
1. Power On the Meter
For the most accurate measurements possible, the S310 should be turned on and warmed up for 30
minutes.
Press the ON switch located in the rear pan el of the meter. T he S310 will imm ediately turn on with its
operational state based on the last used detector Group Setting. For information on Group Settings, refer
PN10190T
6
Page 7
Analog Needle Zero Adjustment
to the Group Settings section. If you purchased the S310 with one detector, this detector’s Group settings
will be active and you are ready to take measurements.
If you purchased more than one detector with the S310, you must choose the Group setting for the detector
you are connecting to the S310 in step 4. The Group Setting for each detector purchased with your S310
is shown on page 2. To select a Group Setting, press the SETUP button repeatedly until the desire Group
number appears in the display. Wait 3 seconds and that Group will be automatically selected and made
active.
2. Turn the Analog Needle On or Off (does not apply to the S310D)
To turn the analog meter on or off, press and keep holding down the RANGE button. Then release the
button after the meter appears or disappears.
3. Zero the Analog Needle (does not apply to the S310D)
Refer to Figure 2. The black slotted button located just below the display allows screwdriver adjustment
to set the analog needle to zero. This adjustment should be made before connecting the detector.
PN10190T
Figure 2. S310 Front Panel
(under serial tag)
7
Page 8
4. Connect a Detector
Do not connect more than one detector to the power meter at any time.
Refer to the drawing of the S310 rear panel in Figure 1. A 3-meter mini-DIN type cable with "D" shaped
connectors comes with Vector calorimete rs, incl uding the large aperture calorimeters. A 3 -meter BNC
type cable comes with Vector pyroelectric detectors and Vector HR pyroelectric detectors. To connect the
Vector detectors, the input connectors on the rear panel of the S310 are labeled as follows:
"Ultra" for Ultra Detectors (no longer manufactured)
"Vector" for all Vector Pyroelectric detectors
"Astral” for all Vector calorimeters, including the large aperture calorimeters
Note that the flat side of the DIN type cables should be oriented up when plugging into the S310.
Detectors have a 1/2" diameter mounting post hole for installing a detector post. Mounting bases and posts
are available for purchase at Scientech to attach the detector to a working surface.
5. Select a Range
On the first press of the RANGE button, the analog scale (if active) will disappear. Then briefly press the
RANGE button each time another range is desired. Press the SELECT button when the range you desire
appears in the display (failure to press the SELECT button within 3 seconds will activate the range shown
in the display and return the S310 to its operation state).
6. Select a Mode
The MODE button allows selection of the type of measurement to be made (watts, joules, etc). It also
gives access to the scientific notation unit of measure. The different measurement modes available are:
A. When configured for a pyroelectric detector: Energy (J), Avg. Energy (J AVG), Power ( W AVG),
and Volts (V).
B. When configured for a calorimeter: Power (W AVG), Energy (J).
Briefly press the MODE button each time a different mode setting is wanted. Press the SELECT button
when the mode you desire appears in the display (failure to press the SELECT button within 3 seconds
will activate the mode shown in the display and return the S310 to its operation state).
To select the scien tific notation annunciator, pr ess and hold down the MODE button and only release
when the annunciator appears. To exit the scientific notation mode, press and hold down the MODE
button again.
7. Zero the Display (only for calorimeters with Watts mode selected)
Press the CANCEL button to zero the display when using a calorimeter in watts mode.
PN10190T
8
Page 9
8. Take your measurement
Direct the laser beam onto the absorbing surface of the detector.
GROUP SETTINGS
The operating parameters for the detectors used with the S310 power meter must be stored in the power
meter’s memory. This information is stored in Group Settings. The typical configuration of the Group
Settings is as follows:
Group #1 – Vector Calorimeters or Large Aperture Calorimeters
Group #2 – Vector Pyroelectric Detectors
Group #3 – Vector HR Pyroelectric Detectors
However, the Group Settings may be configured differently depending on what detectors are used with
the power meter. The operating parameters and group number for the detectors purchased with the S310
Power meter are recorded on page 2 of this manual.
To Select a Group
A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again.
C. To select a Group, press the SELECT button when the Group you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the Group shown in the d isplay
and return the S310 to its operational state).
Group Settings for Vector Calorimeters
The factory default Group for Vector 8mm, 25mm, 16mm, 50mm or large aperture calorimeters is Group
#1.
Note: Pressing the CANCEL button at any time during this process will terminate this process with no
changes made.
Note: Holding down the SELECT button for 3 seconds at any time durin g this process will save any
changes made up to that point and return the S310 to its operational state for the Group selected.
PN10190T
9
Page 10
To review, change, or setup a Group Setting for Vector calorimeters, proceed as follows:
A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again.
C. To select a Group, press the SELECT button when the Group you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the Group shown in the d isplay
and return the S310 to its operational state).
D. Press the SETUP button until the calorimeter annunciator “CAL” appears in the display . Press the
SELECT button to activate the calorimeter configuration program.
E. The "tc" (time constant) annunciator now appears in the display. The time constant is a measure
of the length of time the calorimeter takes to respond to a laser beam.
Press the count up (RANGE) and count down (MODE) buttons to change the number in the display
to match the time constant number listed on the serial tag of your Vector calorim eter or int erface
module for large aperture calorimeters. Press the SELECT button to enter the time constant value
to memory.
F. The "SP" (speed) annunciator now appears in the display. The speed setting allows you to control
the power met er’s display rate. The best value will cause a slight overshoot then a quick settling
on the final value. Too high of a setting will cause the display to overshoot then slowly drift back
down to the final value. A slow setting will cause the display to slowly count up to the final value.
Press the count up (RANGE) button to speed up or count down (MODE) button to slow down the
response time of the meter. Press the SELECT button when the desired number appears in the
display. The following settings can be fine tuned to your preference of speed versus overshoot. We
recommend these initial settings:
G. The "Cd" (calorimeter delay) annunciator now appears in the display. The calorimeter delay
feature prohibits the display of energy if a pulse is fired before the entered time (1 to 255 seconds)
elapses. The calorimeter must reach environmental thermal equilibrium before a subsequent pulse
is fired or low energy measurements will occur.
PN10190T
10
Page 11
To set the time delay between pulses use the count up (RANGE) and count down buttons (MODE)
buttons to enter the time in seconds. Press the SELECT button after the time has been entered to
save the setting. The following time delays are recommended:
H. The "At" (attenuation) annunciator now appears in the display. If no attenuator is b eing used in
conjunction with the calorimeter, the attenuation factor in the display must be set at 1.000 since
this value is a display multiplier. Attenuation multipliers from .0001 to 9999 can be entered into
the displayed value.
The attenuation factors of optics can be entered into the S310 so the displayed value will
automatically compensate for the amount of attenuation. For example, assume a beam splitter is
being used that transmits 75% and reflects 25% of the beam. If the S310 is set up to measure the
reflected beam the attenuation could be set up as follows:
• An attenuation factor of 1 would display the value of the reflected beam.
• An attenuation factor of 3 would display the value of the transmitted beam.
• An attenuation factor of 4 would display the value of the source.
Use the count up (RANGE), count down (MODE) buttons to change the attenuation factor. Press
the SELECT button to enter the attenuation factor to memory.
I. If the S310 has the optional IEEE488 interface, the REMOTE annunciator will appear along with
the digital interface previously selected; r232 (RS232) or IEEE (IEEE488). Press the SETUP
button to toggle the interface between r232 (RS232) and IEEE (IEEE488). Press the SELECT
button when the desired interface appears in the display.
J. If the optional IEEE488 interface was not installed, the "br" (baud rate) annunciator for setting up
the RS232 interface appears in the display along with the baud rate previously selected. To change
the baud rate press SETUP button until the desired baud rate appears in the display. One of the
following baud rates can be selected: 300, 1200, 2400, 9600, and 19,200. When the preferred
baud rate appears in the display press the SELECT button.
K. The "PA" (parity) annunciator now appears in the display. Press the SETUP button to select none,
even, or odd parity. Press the SELECT button to enter the parity of choice to memory
PN10190T
11
Page 12
L. The "HS" (handshake) annunciator now appears in the display. Press the SETUP button until your
choice of none (nOnE), on/off (onoF), or clear to send (CtS) appears in the display. Press the
SELECT button when your choice is displayed to enter the handshake to memory.
M. If IEEE488 has been selected, the "bA" (bus address) annunciator appears with the default bus
address. Bus addresses from 0 to 30 may be selected by pressing the count up (RANGE) count
down (MODE) buttons followed by the SELECT button. Once the remote interface has been setup,
the group annunciator will appear. Press the SELECT button or simply wait a few seconds and
the group will automatically be selected for operation.
Group Settings for Vector Pyroelectric Detectors
Note: The factory default Group for Vector Pyroelectric Detectors is Group #2.
Note: The factory default Group for Vector HR Pyroelectric Detectors is Group #3.
Note: Pressing the CANCEL button at any time during this process will terminate this process with no
changes made.
Note: Holding down the SELECT button for 3 seconds at any time during this process will save any
changes made up to that point and return the S310 to its operational state for the Group selected.
Note: If you want to use the transfer calibration function, y ou must select the watts mode prior to entering
the Group configuration program.
To review, change, or setup a Group Setting for Vector Pyroelectric or Vector HR Pyroelectric detectors,
proceed as follows:
A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again.
C. To select a Group, press the SELECT button when the Group you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the Group shown in the display
and return the S310 to its operational state).
D. Press the SETUP button until the "PYro" annunciator appears. Immediately press the SELECT
button to select the pyroelectric detector configuration program.
E. Either the "V/J" annunciator or the "V/mJ" annunciator and a detector sensitivity number will
appear in the display. V/J is used for standard p yroelectric detectors. V/mJ is used only for HR
pyroelectric detectors. Press the SETUP button to toggle between the V/J and V/ mJ entries . The
output sensitivity of the pyroelectric detector in V/J or V/mJ is listed on the serial tag of your
detector. Use the count up (RANGE) and count down (MODE) buttons to enter the V/J or V/mJ
value. Press the SELECT button to enter the value to memory. Do not push the SELECT button
PN10190T
12
Page 13
before entering the sensitivity number as this takes you to the next setup step without the proper
sensitivity number.
F. The "SP" (speed) annunciator now appears in the display. Press the SETUP button to toggle
between bL (black or HD coated detectors) or hF (high frequency) detector. If the detector Model
No. includes the letters “HF”, press the SELECT button when the hF annunciator appears in the
display. If the detector’s surface is black (does not have the let ters “ HF” in the Model No.) press
the SELECT button when the "bL" annunciator appears in the display.
G. The "AUTO, SET CAL" annunciators now appear in the display (only if watts mode was selected
prior to group configuration). This is the Transfer Calibration function. The Transfer Calibration
function is designed to transfer the calibration from a NIST certified calorimeter to a Vector
pyroelectric detector. This function allows you to adjust the output sensitivit y o f your Vector
pyroelectric detector in combination with your S310 meter (in the average power mode) in order
to match the avera ge po wer r eadin g of a N IST c erti fied s ystem. T ypicall y, a 50/50 beam splitter
is used with the Vector pyroelectric detector to be calibrated in one beam path and the NIST
certified calorimeter in the other beam path. You are to adjust the output sensitivity of the
pyroelectric detector while it is operating by using the count up (RANGE), count down (MODE)
buttons, which change the V/J or V/mJ settings, to make the displayed value of the S310 power
reading the same as the NIST standard. Press the SELECT button when the readings match.
H. The "At" (attenuation) annunciator now appears in the display. This allows you to enter a value
from 0.0001 to 9999 which will become a multiplier of the actual displayed value. I f no attenuator
or correction factor is to be used in conjunction with the pyroelectric detector then the attenuation
factor must be set to 1.000.
The attenuation factors of optics can be entered into the S310 so the displayed value will
automatically com pensat e for the amou nt of atten uatio n. For exam ple, assu me a beam spl itter is
being used that transmits 75% and reflects 25% of the beam. If the S310 is set up to measure the
reflected beam the attenuation could be set up as follows:
• An attenuation factor of 1 would display the value of the reflected beam.
• An attenuation factor of 3 would display the value of the transmitted beam.
• An attenuation factor of 4 would display the value of the source.
Use the count up (RANGE), count down (MODE) buttons to enter the attenuation value. Press
the SELECT button to enter the value to memory.
I. The REMOTE annunciator now appears along with the digital interface previously selected; r232
(RS232) or IEEE (IEEE488). Press the SETUP button to change the interface between r232
(RS232) and IEEE (IEEE488). Press the SELECT button when the desired interface appears in
the display.
J. If RS232 has been selected, the "br" (baud rate) annunciator appears in the display along with the
baud rate previously selected. To change the baud rate press the SETUP button repeatedly until
the desired baud rate appears in the display. One of the following baud rates can be selected: 300,
PN10190T
13
Page 14
1200, 2400, 9600, and 19,200. When the preferred baud rate appears in the display press the
SELECT button.
K. The "PA" (parity) annunciator now appears in the display. Press the SETUP button to select none,
even, or odd parity. Press the SELECT button to enter the parity choice to memory
L. The "HS" (handshake) annunciator now appears in the display. Press the SETUP button until your
choice of none (nOnE), on/off (onoF), or clear to send (CtS) appears in the display. Press the
SELECT button when your choice is displayed to enter the handshake to memory.
M. If IEEE488 has been selected, the "bA" (bus address) annunciator appears with the default bus
address. Bus addresses from 0 to 30 may be selected by pressing the count up (RANGE) count
down (MODE) buttons followed by the SELECT button. Once the remote interface has been setup,
the group annunciator will appear. Press the SELECT button or simply wait a few seconds and
the group will automatically be selected for operation.
OPERATING PROCEDURES
Using the S310 with V ec tor Pyroelectric Detectors
Pyroelectric detector models P25, P50, SP25, SP50, P05, and P09 are coated with a special black
absorbing material which provides a very flat spectral response over a broad wavelength band.
Pyroelectric detector models PHD25, PHDX25, PHDX25UV, PHD50, PHDX50, PHDX50UV, SPHD25,
and SPHD50 are coated with a special high damage absorbing material which provides absorption over a
broad wavelength band. Models PHF25, PHF50, SPHF25, SPHF50, PHF02, PHF05, and PHF09 have a
partially absorbing, partially reflecting chromium coating. The relative spectral responses of these
detectors are shown in the following graph. Please be aware of the absorption differences between the
detector’s calibration wavelength and your operational wavelength. Detailed absorption information is
contained in the charts at the end of this manual.
PN10190T
14
Page 15
RELAT I VE SPEC TRAL R ESPON SE
20
30
40
50
60
70
80
90
100
0.1110100
WAVELENGTH [m icrons]
PERCENT ABSORPTION
P and SP
HR-P
PHD, PH
and SPH
HR-P
PHF a nd SPHF
*HDX Limi ted to 0.4 to 2.0
PHDXUV
Figure 3. Pyroelectric Spectral Response
Before using your Vector jouleme ter sy stem, plea se review the energy density for mulas given in the chart
at the front of this man ual. Familia rize yo urself with all of th e spec ifications of the detecto r models which
you are using. A damage test slide is provided with each P and PH F type det ectors, but not with PHD type detectors. Fire the beam at the test slide before using the detector to be sure you are operating under
safe conditions.
The default Group Setting for Vector Pyroelectric Detectors is Group #2. The default Group Setting for
Vector HR Pyroelectric Detectors is Group #3. Either select Group #2, #3, or configure another Group
for pyroelectric detectors and select that Group. For Group configuration, refer to the Group Settings
section.
Note: The automatic trigger threshold of the S310 is 7 % of full scale.
1. To Turn On the Meter
Note: For the most accurate measurements possible, the S310 should be turned on and warmed up for 30
minutes.
Press the ON switch located in the rear panel of the meter. The S310 will immediately turn on with its
operational state based on the last used detector Group Setting. If you purchased the S310 with one
detector, this detector’s Group settings will be active and you are ready to take measurements.
If you purchased more than one detector with the S310, you must choose the Group setting for the detector
you are connecting to the S310 as described in step 4. The Group Setting for each detector purc hased
with your S310 is shown on page 2.
PN10190T
15
Page 16
2. To Turn the Analog Needle On or Off (does not apply to the S310D)
To turn the analog meter on or off, press and keep holding the RANGE button down. Then release after
the meter appears or disappears.
3. To Zero the Analog Needle (does not apply to the S310D)
Refer to Figure 3. The black slotted button located just below the display allows screwdriver adjustment
to set the analog needle to zero. This adjustment should be made before connecting a pyroelectric detector.
4. To Connect a Pyroelectric Detector
Note: Only one detector should be plugged in at any time.
Refer to the drawing of the S310 rear panel in Figure 2. A 3 mete r BNC type cabl e comes wit h Vector
pyroelectric detectors and Vector HR pyroelectric detectors. One of the input connectors on the rear panel
of the S310 is labeled "Vector" for hook up of the pyroelectric detectors.
5. To Select a Group
A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again.
B. To select the appropriate Group, press the SELECT button when the Group you desire appears in
the display (failure to press the SE LECT button within 3 seconds will activate the Group shown
in the display and return the S310 to its operational state).
6. To Select a Range
Note: AUTO range may be selected if the energy levels of repetitive pulses are to be measured. However,
do not select AUTO range if you want to measure single pulse energy or pulses running at repetition rates
lower than 10 Hz.
A. Press the RANGE button. On the first press of the RANGE button, the analog scale (if active) will
disappear. Then briefly press the RANGE button each time another range is desired.
B. To select a range, press the SELECT button when the range you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the range shown in the display
and return the S310 to its operational state). The ranges available for pyroelectric detectors are in
the following table.
PN10190T
16
Page 17
Model
Standard or SP
PHF02 - L
PHF02 – S or I
PHF05 – L
PHF05 – S or I
Mode
Power
Energy
Power
Energy
Power
Energy
Power
Energy
Power
Energy
3mW
3mJ
30.00μW
30.00μJ
3.000μW
AUTO
3.000μJ
AUTO
300.0μW
300.0μJ
3.000μW
AUTO
3.000μJ
AUTO
Model
PHF09 – L
PHF09 – S or I
P05
P09
Mode
Power
Energy
Power
Energy
Power
Energy
Power
Energy
3.000m
3.000m
30.00μW
30.00μJ
3.000μW
AUTO
3.000μJ
AUTO
3.000μW
AUTO
3.000μJ
AUTO
Table 2. S310 Ranges for Pyroelectric
Range
s
Range
s
30mW
300mW
3W
AUTO
W
30.00m
W
AUTO
30mJ
300mJ
3J
AUTO
J
30.00m
J
AUTO
3.000μW
3.000m
W
30.00m
W
AUTO
300.0μW
3.000m
W
30.00m
W
AUTO
3.000μJ
3.000m
J
30.00m
J
AUTO
300.0μJ
3.000m
J
30.00m
J
AUTO
30.00μW
300.0μW
3.000m
W
30.00m
W
30.00μW
300.0μW
3.000m
W
30.00m
W
30.00μJ
300.0μJ
3.000m
J
30.00m
J
30.00μJ
300.0μJ
3.000m
J
30.00m
J
3.000m
W
30.00m
W
AUTO
30.00μW
300.0μW
3.000m
W
30.00m
W
3.000m
J
30.00m
J
AUTO
30.00μJ
300.0μJ
3.000m
J
30.00m
J
30.00μW
300.0μW
3.000m
W
30.00m
W
30.00μJ
300.0μJ
3.000m
J
30.00m
J
7. To Select a Mode
Available pyroel ectric modes are: Energ y (J), Avg. Energy (J AVG), A verage Power (W AVG), and
Volts (V).
A. To Measure Energy (J):
Note: The update rate of the display is 20Hz.
i. Press the MODE button (repeatedly, if necessary) until the “J” annunciator appears in the
display.
ii. Press the SELECT button (or wait 3 seconds) and the joules mode will automatically be
activated.
iii. The energy level of each laser pulse will be displayed on the LCD.
B. To Measure Average Energy (J AVG)
PN10190T
17
Page 18
Note: The maximum repetition rate for average energy is 300Hz.
i. Press the MODE button (repeatedl y, if necessary) until the “J AVG” annunciator app ears in
the display.
ii. Press the SELECT button (or wait 3 seconds) and the average joules mode will be
automatically activated. The number of pulses to be averaged will now appear in the display.
You may select the number of pulses to average from 2 to 9999.
iii. To change the number of pulses to be averaged, press the RANGE (count up) and/or M ODE
(count down) buttons. When the desired number of pulses to be averaged appears in the
display, press the SELECT button.
iv. The average energy of the number of pulses you selected will be displayed after the number of
pulses entered in step ii is received by the pyroelectric detector. This is not a running average,
but is the average for the pulse population selected in step ii. Nothing is displayed until the
full pulse population is delivered. This average is displayed until another full population of
pulses is delivered at which time the display is updated with the average for that population of
pulses.
C. To Measure Average Power (W AVG)
Note: The maximum re petition rate for average power is 300Hz. The minimum repe tition rate for average
power is 10Hz.
Note: The average power mode displays the average power (watts) of repetitively pulsed lasers.
Pyroelectric detectors will not work with continuous wave lasers.
i. Press the MODE button repeatedly until the “W AVG” annunciator appears.
ii. Press the SELECT button (or wait 3 se conds) and the S310 will automatically enter into its
operational state in the average power mode.
iii. The average power will be displayed.
D. To Measure Volts (V)
Note: The maximum repetition rate for volts is 300Hz.
i. Press the MODE button repeatedly until the “V” annunciator appears.
ii. Press the SELECT button (or simpl y wait 3 seconds) and the S310 will automatically enter
into its operational state in the volts mode.
iii. The volts per pulse will be displayed.
8. To Measure a Statistical Run of Energy Pulses
PN10190T
18
Page 19
Note: Do not use AUTO range when making a statistical run.
Note: Do not go from the Average Power Mode to Stats since the range will be too high. Select the range
manually.
Note: When using the PHF09 pyroelectric detector in the long pulse setting, a manual range must be
selected when running statistics.
Note: Each time a new stats run begins, data from the previous run is lost. If statistical data is to be saved,
it must be done though the digital interface.
Note: To exit the statistical mode at any time, press the CANCEL button.
Note: Statistics mode can collect data at repetition rates of up to 750 Hz depe nding on the detector model.
The statistics mode will collect data on a pulse population of up to 1000 pulses. At your prompting, the
power meter will display the number of pulses deli vered, average energy, minimum energy, maximum
energy, standard deviation, and coefficient of variation. When the statistics mode is selected, the energy
mode is automati cally activated regardless of t he m od e prev io us l y sele cted. Select the appropri at e r ange
for the pulse energy level to be measur ed. It is very importan t to select the most appropriate range. I f you
have selected a manual range and the laser pulse(s) has overflowed the maximum energy of the range, OF
will be displayed when the data is recalled. You should then select a higher range.
To enter into the statistics mode:
A. Press the STATS button. The number of pulses in the last statistics run will appear in the display.
B. Use the RANGE (count up) and MODE (count down) buttons to change the display to the desired
number of pulses to include in the statistics run (up to 1000).
C. Press the SELECT button to enter the pulse population to memory.
D. Press the SELECT button to begin the run. The power meter will automatically stop once the data
has been collected.
B. Press the STATS button to recall the data to the display. The STATS button must be pressed each
time to recall each of the following statistical calculations:
Number of pulses collected
Average energy (AVG)
Minimum energy (MIN)
Maximum energy (MAX)
Standard deviation (SIGMA)
Coefficient of variation (CV%)
PN10190T
19
Page 20
F. After recalling the statistic al inf or mation , the fla shing STA TS a nnunc ia tor indica tes that th e S310
is ready for a new run.
A. Press the SELECT button to start a new run (each time a new run begins, data from the previous
run is lost) or press the CANCEL button to return the power meter to the mode of operation in
effect prior to statistics mode.
9. External Trigger
The automatic trigger threshold of the S310 is 7 % of full scale. The external trigger function is designed
for single pulse energ y measurement with a calorimeter. It does not improve the trigger circuit of the
S310 with pyroelectric detectors.
To Perform a Transfer Calibration
You can transfer a calibration from a calori meter to a Vector pyroelect ric detector using the Tran sfer
Calibration function of the S310. To do this, please refer to the Group Settings for Vector Pyroelectric
Detectors, parag raph #G. You must follow the Group Settings steps to get into the Transfer Calibration
function. This function allows you to adjust the output sensitivity of your Vector pyroelectric detector in
combination with your S310 meter in order to match the average power reading from the S310 to that of
a NIST certified system. Typically, a 50/50 beam splitter is used with the Vector pyroelectric detector to
be calibrated in one beam path and the NIST certified calorimeter in the other beam path.
HR Battery Installation
A 9 Volt alkaline battery comes installed with all HR Series detect ors. Alwa ys select the OFF positi on
of the power switch when the detector is not in use to spare the lifetime of the battery. Repla ce the battery
when the low battery LED indicator lights up.
Figure 4. HR Detector Battery Access
PN10190T
20
Page 21
Refer to Figure 4. Remove the two slotted 4-40 binder head screws located on the underside of the
detector. Pull off the outer housing to expose the battery. Remove the used battery from the battery holder
and snap in the new battery. Slide the outer housing back in to place and secure with the screws.
Do not touch the delicate pyroelectric crystals in the HR Series detectors. The y should only be cleaned
with a stream of clean air, nitrogen or CO2.
Set Electrical Time Constants for Model PHF02, PHF05 and PHF09
The pulse width switch, located on the rear of the PHF02, PHF05, and PHF09 detectors, selects one of
three electrical time constants and should be set as follows:
S (Short Pulses): Select for pulse durations of 5 µsec. or less. Repetition rates up to 4
kHz can be measured.
I (Intermediate Pulses): Select for pulse durations of 50 µsec. or less. Repetition rates of up to
400 Hz can be measured.
L (Long Pulses): Select for pulse durations of 250 µsec. or less. Repetition rates of up to
80 Hz can be measured.
There are no pulse width settings on the P05 and P09 detectors which utilize a highly absorbing material
on the crystal. The pulse duration considerations merely function in the determination of damage
thresholds. Repetition rates up to 100 Hz using the P05 and 50 Hz using the P09 can be measured.
Correcting Pyroelectric Detector Operating Parameter s for Use at Different
Wavelengths
Due to variability in the manufacturing process the absorption characteristics of the HD and P model
detectors can vary in the UV region (190 to 400nm). Scientech r ecommends optica l calibra tion at 266nm
if the detector is to be used in the UV region instead of re lying on an absorption ra te in the chart at the end
of this manual.
All pyroelectric d etectors are cal ibrated at a spec ific wavelength and the detector’s output sensitivity is
derived for that wavelength. The output sensitivity and calibration wavelength is recorded in the
Operating Parameters section at the front of the manual and on the detector’s serial tag. When a
pyroelectric det ector is used at a wavelength other than the calibration wavelength, its output sensitivity
can be adjusted to compensate for the absorption rate at the new wavelength. The new output sensitivity
is calculated as follows:
1. Find the absorption rate from the chart at the end of this manual for the calibration wavelength of
your pyroelectric detector.
2. Find the absorption rate for the wavelength where you will be working.
PN10190T
21
Page 22
3. Determine the new output sensitivity using the following formula:
absorption rate of new wavelength x output sensitivity (V/J) = output sensitivity for new
absorpt ion rate of calibration wavelength wavelength from serial tag
This new output sensitivity can be entered into the S310 as described in Group Settings or when using the
pyroelectric detector without a S310 power meter as discussed later in this manual.
Using S310 with Vector Series Calorimeters
The calorimeter sele cted needs to be t he appropriate mod el for the planned l aser measurements. Please
familiarize yourself with the operating specifications which are given in the front of this manual.
The default Group Setting for calorimeters is Group #1. Either select Group #1 or configure another
Group for calorimeters and select that Group.
Vector™ calorimeters are sensitive to all types of thermal input. Due to the handling of the calorimet er
during setup and possible environmental temperature differences, thermal gradients may exist in the
calorimeter. Allow the calorimeter to sit undisturbed for several minutes to, reach thermal equilibrium,
before using.
Note: When using a 25mm calorimeter for measuring average power levels below 30mW and single pulse
energy levels below 30mJ, a Scientech Model 360203A isoperibol™ Enclosure is highly recommended.
The isoper ibol enclosure should not be used at average power levels above 30mW, and single pulse energy
levels above 100mJ because heat build-up will occur.
Note: Large Ap erture calorimet ers and the Interface M odul es that th ey are c alibrat ed with a re matched
sets and must be used together.
1. To Turn-On the Meter
Note: For the most accurate measurements possible, the S310 should be turned on and warmed up for 30
minutes.
Press the ON switch located in the rear pan el of the meter. T he S310 will imm ediately turn on with its
operational state based on the last used detector Group Setting. If you purchased the S310 with one
detector, this detector’s Group settings will be active and you are ready to take measurements. If you
purchased more than one detector with the S310, you must choose the Group setting for the detector you
are connecting to the S310 as described in step 4. The Group Setting for each detector purchased with
your S310 is shown in the Group Setting section.
2. To Turn the Analog Needle On or Off (does not apply to the S310D)
To turn the analog meter on or off, press and keep holding the RANGE button down. Then release after
the meter appears or disappears.
PN10190T
22
Page 23
Model
Vector 25mm
Vector 50mm
Large Aperture 100mm
Module*
Large Aperture 200mm
Module**
Mode
Power
Energy
Power
Energy
Power
Energy
Power
Energy
10mW
AUTO
10mJ
300mW
300mJ
300mW
300mJ
300mW
300mJ
3. To Zero the Analog Needle (does not apply to the S310D)
Refer to Figure 2. The black slotted button located just below the display allows screwdriver adjustment
to set the analog needle to zero. This adjustment should be made before connecting a calorimeter.
4. To Connect a Calorimeter
Note: Only one detector should be plugged in at any time.
Refer to the drawing of the S310 rear panel in Figure 1. A 3 meter mini-DIN type cable with "D" shaped
connectors comes with Vector c alorimeters. For large aper ture calo rimeters, th e interconnect cables for
both the calorimeter and power meter are hardwired to the interface module. One of input connectors on
the rear panel of the S310 is labeled "Astral" for the connection of Vector (previously called Astral)
calorimeters and including the large aperture (100 & 200mm apertures) calorimeters. Note that the flat
side of the DIN type cables should be oriented up when plugging in to the S310.
5. To Select a Group
A. Press the SETUP button. The Group Setting last used will appear in the display.
B. To move to the next Group, press SETUP button again.
B. To select the appropriate Group, press the SELECT button when the Group you desire appears in
the display (failure to press the SELECT button within 3 seconds will activate the Group shown
in the display and return the S310 to its operational state).
6. To Select a Range
Note: AUTO range is not available in the energy mode for calorimeters
A. Press the RANGE button. On the first press of the RANGE button, the analog scale (if active) will
disappear. Then briefly press the RANGE button each time another range is desired.
B. To select a range, press the SELECT button when the range you desire appears in the display
(failure to press the SELECT button within 3 seconds will activate the range shown in the display
and return the S310 to its operational state). The ranges available for calorimeters are specified in
Table 3.
Table 3. S310 Ranges for Vector™ Calorimeters
Range
PN10190T
100mW
1W
10W
100mJ
1J
10J
3W
30W
AUTO
3J
30J
with PN10735 Interfac e
3W
30W
AUTO
3J
30J
with PN10747 Interface
3W
30W
AUTO
3J
30J
23
Page 24
* With a PN10748 – 10X attenuator, the actual power or energy is 10 times the displayed value up to 50W
or 150J.
** With a PN10769 – 10X attenuator, the actual power or energy is 10 times the displayed value up to
100W or 300J.
Note: When using PN10748 or PN10769, 10X attenuators, you may want to use an attenuation factor of
10 so the displayed reading on the S310 is correct. However, the range designation may be incorrect if
the power or energy reading exceeds the upper limit of the ran ge. See the Group Settings section for
details.
7. To Select a Mode
The modes available for the calorimeters are: Average Power (W AVG) or Energy (J).
A. To Measure Average Power (W AVG):
Note: The speed-up circuit (differentiator) in watts mode is controlled by software and was adjusted at
the factory to accom modat e the t ype of calorim eter bein g used. You can ch ange the sp eed-up setting to
your preference. For changing a Group setting, refer to the Group Setting discussion in the previous
section on Group Settings for Vector or Lar ge Aper ture Cal orim eters . The speed-up circuit is not active in AUTO range.
Note: The average power mode displays the average power (watts) of repetitively pulsed lasers or
continuous wave lasers.
i. Press the MODE button repeatedly until the “W AVG” annunciator appears.
ii. Press the SELECT button (or simply wait 3 seconds) and the S310 will automatically enter
into its operational state in the average power mode.
iii. Press the CANCEL button to zero the display.
iv. The average power will be displayed.
A. To Measure Energy (J)
Note: Calorimeters can only measure single shot energy pulses (time between pulses is dependent on the
calorimeter delay “CD” setting in the Group Setting for calorimeters [see previous section on Group
Settings for Vector or Large Aperture Calorimeters]). With the calorimeter delay entered, the S310 will
display the "trig" annunciator and the single pulse energy after the first pulse is delivered. The "trig"
annunciator will then disappe ar after the calorimeter de lay time has elapsed prompting you to fire another
pulse. Do not fire another pulse until the annunciator disappears. If you do, the S310 resets the time delay
and ignores the sequential pulse altogether.
i. Press the MODE button (repeatedly, if necessary) until the “J” annunciator appears in the
display.
PN10190T
24
Page 25
ii. Press the SELECT button (or simply wait 3 seconds) and the joules mode will automatically
be activated.
iii. The energy level of each laser pulse will be displayed on the LCD.
8. To Measure a Statistical Run of Single Shot Energy Pulses
Note: Calorimeters can only measure single shot energy pulses (time between pulses is dependent on the
calorimeter delay “CD” setting in the Group Setting for calorimeters [see previous section on Group
Settings for Vector or Large Aperture Calorimeters]). With the calorimeter delay entered, the S310 will
display the "trig" annunciator and the single pulse energy after the first pulse is delivered. The "trig"
annunciator will then disappear after the calorimeter delay time has elapsed prompting you to fire another
pulse. Do not fire another pulse until the annunciator disappears. If you do, the S310 resets the time delay
and ignores the sequential pulse altogether.
Note: Do not use AUTO range when making a statistical run.
Note: Do not go from the Average Power Mode to Stats since the range will be too high. Select the range
manually.
Note: Each time a new stats run begins, data from the previous run is lost. If statistical data is to be saved,
it must be done though the digital interface.
Note: To exit the statistical mode at any time, press the CANCEL button.
The statistics mode will collect data on a pulse population of up to 1000 pulses. At your prompting, the
power meter will displ ay the number of pulses delivered, average energy, minimum energy, maximum
energy, standard deviation, and coefficient of variation. When the statistics mode is selected, the energy
mode is automatically activated regardless of the mode previously selected . S el ect t he ap pro pri ate range
for the pulse energy level to be measur ed. It is very importan t to select the most appropriate range. If you
have selected a manual range and the laser pulse(s) has overflowed the maximum energy of the range, OF
will be displayed when the data is recalled. You should then select a higher range.
To enter into the statistics mode:
A. Press the STATS button. The number of pulses in the last statistics run will appear in the display.
B. Use the RANGE (count up) and MODE (count down) buttons to change the display to the desired
number of pulses to include in the statistics run (up to 1000).
C. Press the SELECT button to enter the pulse population to memory.
D. Press the SELECT button to begin the run. The power meter will automatically stop once the data
has been collected.
E. P ress the STATS button to recall the data to the display. The STATS button must be pressed to
recall each of the following statistical calculations:
PN10190T
25
Page 26
Number of pulses collected
Average energy (AVG)
Minimum energy (MIN)
Maximum energy (MAX)
Standard deviation (SIGMA)
Coefficient of variation (CV%)
F. After recalling the statistic al inf or mation , the fla shing STA TS a nnunc ia tor indica tes that th e S310
is ready for a new run.
G. Press the SELECT button to start a new run (each time a new run begins, data from the previous
run is lost) or press the CANCEL button to return the power meter to the mode of operation in
effect prior to statistics mode.
9. External Trigger
The external trigger input is located on the rear panel of the S310 as shown in Figure 1. The external
trigger function is designed to ensure the entire pulse energy of a single pulse is captured by the S310
power meter. An extern al tri gger “awakens” the S310’s circuitry so that it is ready for the arrival of the
laser pulse. The external trigger pulse must have the following characteristics:
1. 2.5 volts min. to 5 volts max.
2. A minimum pulse width of 10 µsec. to a maximum of the laser pulse repetition rate.
3. A maximum rise/fall time of 500 nsec.
4. Delivered 1 to 3 milliseconds prior to the laser pulse.
The trigger is generated on the rising edge of the incoming pulse.
The following are viable external triggers.
The following pulses will not trigger the S310.
PN10190T
26
Page 27
Correcting FX Calorimeter Operating Parameters for U se at Different Wavelengths
Note: FX calo ri meters are no long er avai labl e f or sal e. Th is i nformati on i s pro vided fo r users w ho
previously purchased FX calorimeters to understand how they operate with Vector power meters.
Note: Due to variability in the manufacturing process the absorption characteristics of th e FX model
detectors can vary in the UV region (190 to 400nm). Scientech r ecommends optica l calibra tion at 266nm
if the detector is to be used in the UV region instead of re lying on an absorption ra te in the chart at the end
of this manual.
Scientech calorim eters in gene ral have a flat response to al l wavelengths within their sp ecified spectr al
response. FX and FXX calorimeters are an exception to that rule and are calibrated at a specific
wavelength by adjusting the ca lor imeter’s gain circuitry for that wavelength. The calibration wavelength
is recorded in the Operating Parameters section at the front of the manual and on the detector’s serial tag.
When a FX or FXX calor im eter i s used at a w avel ength o th er t han t he c al i brat io n wavelength, the power
meter’s displayed value can be adj usted to compens ate for t he absorp tion rate at t he new w avelength by
using an attenuation factor. The attenuation factor is calculated as follows:
1. Find the absorption rate from the chart at the end of this manual for the calibration wavelength of your
calorimeter.
2. Find the absorption rate for the wavelength where you will be working.
3. Determine the attenuation factor using the following formula:
absorption rate of calibration wavelength = attenuation factor
absorption rate of the new wavelength
The attenuation factor can be entered into the S310 as described in the Group Setting section.
PN10190T
27
Page 28
Analog Output
The analog output is an uncalibrated output accessible via the 50 ohm terminated BNC connector located
on the power meter's r ear panel (see Figure 1). Although the analog output voltage level is uncalibrated,
it is representative of the power and energy readings displayed on the S310. The analog output voltage
level is approximately 3 volts full scale. For high accuracy, you must determine the relationship between
the analog output voltage level and the power and energy readings on the S310 display.
With a calorimeter connected to the S310 in watts mode, the analog output voltage readings will rise to a
steady level when power is applied to the calorimeter. After the voltage readings become stable, note the
analog output voltage reading and the watt reading on the S310 display. All subsequent voltage readings
will have the same V/W value s. In the joules mode, the voltage w ill rise to a peak value relatively quickly
(seconds) followed by a rapid return to baseline. The peak voltage is representative of the joules value.
Again, this can be determined by noting the peak voltag e value on the analog output, and the joules reading
on the S310 display.
With a pyroelectric det ecto r connected t o the S31 0, determi ne the relati onsh ip between t he peak vo ltage
value and the energy (joules) or power (watts) reading on the S310 display in the same manner.
The analog output with the various detectors and operating modes should appear as follows:
A. Calorimeter Watts Mode
Compare voltage to the watt reading displayed on
the S310 to determine V/W constant to use when
looking at the analog output.
In power (watts) mode, the analog output is nominally 3 V full scale for each range. Therefore, to calculate
watts from the analog output voltage:
W = V/S
where:
V = analog output voltage (V)
S = analog output sensitivity (V/W) ≈ 3 V/range (W)
(for example, on the 10 mW range, S ≈ 3 V/.010 W = 300 V/W)
B. Calorimeter Joules Mode
PN10190T
28
Page 29
Compare the peak voltage to the energy (joules)
reading displayed on the S310 to determine
V/J constant to use when looking at the analog
output.
In the energy (joules) mode the analog output is an amplified voltage signal proportional to the voltage
generated by the calorimeter thermopile. To calculate energy (joules) from the analog output use the
following equation.
J = 0.3(Vpk)(TC)
(S)(R)
where:
Vpk = peak voltage from the analog output
TC = calorimeter time constant from the calorimeter serial tag
S = calorimeter output sensitivity as follows:
0.5 V/W for 25 mm models
0.1667 V/W for 50 mm models and large aperture calorimeters
R = power meter range multiplier as follows:
10.0 for the 10 or 30 range
100.0 for the 1 or 3 range
1,000.0 for the .1 or .3 range
10,000.0 for the .01 or .03 range
C. Pyroelectric Watts and Joules Modes
Compare peak voltage to energy (joules) or power (watts) reading displayed on the S310 to determine V/J
or V/W constant to use when looking at the analog output.
PN10190T
29
Page 30
In power (watts) and ener gy (joules) mo de, th e peak volt age of a laser p ulse is directl y proportional to t he
power energy level:
W = Vpk/S
where:
Vpk = peak voltage from the analog output
S = analog output sensitivity (V/W) ≈ 3 V/range (W)
(for example, on the 3mW range, S ≈ 3 V/.003 W = 1000 V/W)
J = Vpk/S
where:
Vpk = peak voltage from the analog output
S = analog output sensitivity (V/J) ≈ 3 V/range (J)
(for example, on the 3 mJ range, S ≈ 3 V/.003 J = 1000 V/J)
CALIBRATION OF VECTORCALORIMETERS USING ELECTRIC
SUBSTITUTION HEATERS
For Vector™ calorimeters the electric substitution heating option must be ordered and installed at the
factory when th e calorimeter is purchased. It cannot be retrofitted to a calorim eter at a later time. To
calibrate using electric substitution heating proceed as follows.
PN10190T
30
Page 31
Figure 5. Calorimeter Ci rcu it Board
A. Remove the screws h olding the calorimeter 's ID tag and remove the plat e to expose the circuit
board as shown in Figure 5.
B. Connect the calorimeter to the power meter, turn on the power and let the system equilibrate.
C. Connect a DVM to the test points labeled SUB and HTR on the calorimeter circuit board.
D. Measure the resistance of the substitution heater making sure to subtract the resistance of the patch
cables from the total resistance measurement. Compare this resistance to Rc in the calibration data
in the front of the manual. The two should agree within 2%. If not contact Scientech.
E. Remove the DVM. Connect a power supply to the SUB and HTR test points and connect the
DVM to monitor the power supply.
F. Set up the power meter in the Watts Mode and the 10W range for 25 mm calorimeters or the 3W
range for 50 mm calorimeters.
G. Apply Vh volts, stated in the calib ration da ta you receiv ed with the ca lorimeter, to the substitutio n
heater.
H. If needed, adjust the calibration trim pot, R4 on the calorimeter circuit board, until Wh Watts, from
the calibration data, is displayed by the power meter.
CALIBRATION OF LARGE APERTURE CALORIMETERS USING ELECTRIC
SUBSTITUTION HEATING
Electrical substitution heating is a standard feature of large aperture calorimeters.
PN10190T
31
Page 32
Figure 6. Large Calorimeter Connector Panel
Calibration with an Int erf ac e Modu le an d S310 Power Me ter
A. Connect a DVM to the white jacks of the calorimeter. Refer to Figure 6.
B. Measure the resistance of the substitution heater making sure to subtract the resistance of the patch
cables from the total resistance measurement.
Note: When measuring the substitution heater resistance or a 200 mm calorimeter, R1 and R2 must be
connected together in series.
Compare this resistance to Rc in the calibration data in the front of the manual. The two should agree
within 2%. If not contact Scientech.
C. Remove the DVM. Connect a power suppl y to the white jacks and connect the DVM to monitor
the power supply.
D. Set up the power meter in the Watts Mode and the 30W range.
E. Remove the screws holding the interface module’s ID tag and remove the plate to expose the
circuit board. Refer to Figure 5.
F. Apply Vh volts, stated in the calibration data you received with the calorimeter, to the substitu tion
heater.
G. If needed, adjust the calibration trim pot, R4 on the calorimeter circuit board, until Wh Watts, from
the calibration data, is displayed by the power meter.
Calibration with out an Interface Module and S310 Power Meter
For this procedure you will need to make an adapter cable to go from the calorimeter’s DIN connector to
the DVM. The voltage output is on pin 1 of the DIN connector and should be connected to the posit ive
side of the DVM. Ground is on pin 3 and should be connected to the negative side. Pin 2 is not used.
Refer to Figure 6.
A. Connect a DVM to the white jacks of the calorimeter. Refer to Figure 5.
PN10190T
32
Page 33
B. Measure the resistance of the substitution heater making sure to subtract the resistance of the patch
cables from the total resistance measurement.
Note: When measuring the substitution heater resistance or a 200 mm calorimeter, R1 and R2
must be connected together in series.
Compare this resis tance to Rc in the calibration data in the front of the manual. The two should
agree within 2%. If not contact Scientech.
C. Calculate the voltage equivalent to laser power using the following formula:
D. Connect the DVM to the calorimeter’s DIN connector.
E. Apply the calculated voltage (V) to the electrical substitution heater.
F. Record the voltage reading of the DVM (Vc).
G. Calculate the calorimeter’s output sensitivity (S) as follows:
S = Vc/W
where:
S = calorimeter’s output sensitivity
PN10190T
33
Page 34
Vc = voltage output from the calorimeter in mV
W = desired laser power output.
The measured sensitivity should be ± 3 % of the calorimeter’s original sensitivity value.
DETECTOR OPERATION WITHOUT A POWER METER
Pyroelectric D et ector Operation without Power Meter
A. Standard and SP Models
Standard and SP model pyroelectric detectors can be operated with a 1MΩ input oscilloscope. The peak
voltage shown on the oscilloscope can be divided by the V/J output sensitivity of the detector to calculate
energy.
B. HR Models
HR pyroelectric detectors can be operated with a 50Ω input oscilloscope. The peak voltage shown on the
oscilloscope can be divided by the V/mJ output sensitivity of the detector to calculate energy.
Vector
Vector (previous called Astral) calorimeters are powered up by the power meter. To use a Vector
calorimeter with ou t a S cientech power met er, but with a volt meter or chart recorder, you must app l y + / 8VDC to the mini DIN connector as shown in Figure 7. The voltage output of the calorimeter, from pin
8, should be connected to the positive side of the DVM or chart recorder. All 3 of the grounds should be
tied together at the negative side. Pins 2 and 3 are not used.
When large aperture calorimeters are used without an power meter their interface module is not used. The
output of the calorimeter is connected directly to the DVM or chart recorder. Large aperture calorimeters
do not require any power. The voltage output is on pin 1 of the DIN connector and should be connected
to the positive side of the DVM or chart recorder. Ground is on pin 3 and should be connected to the
negative side. Pin 2 is not used. Refer to Figure 6.
™
Calorimeter Operation with ou t Power Met er
A. Cable Requirements
Figure 7. Calorimeter mini DIN Connector
PN10190T
34
Page 35
Operation of Vector Calorimeters with a D igital Volt Meter
Note: Whenever a l arge ap erture cal orimete r is used without a power meter the interface module is not
required.
The calorimeters may be used with any digital volt meter (DVM) capable of reading 5 volts full scale.
Connect the output of the calorimeter to the DVM.
A. Select the DC volts mode.
B. Direct the laser beam on to the absorbing surface of the calorimeter.
C. When the displa y of the DVM has stabilized (about 2 minutes), calculate the laser power using the
formula:
W = V/S
where:
W = Laser power in watts
V = Voltage reading of the DVM in volts
S = Sensitivity of the calorimeter from page 2.
Operation of Vector Calorimeters with an Analog Recorder
Note: Whenever a large apertur e calorim eter is used with out a power meter the interface module is not
used.
Calorimeter Response:
The response of a calorimeter to a single pulse input as displayed by a chart recorder appears below.
The output voltage from a chart recorder can be converted to wattage at any time by:
PN10190T
35
Page 36
W = V/S, Wi = Vi/S
V = Chart recorder voltage level in mV
S = Calorimeter sensitivity in mV/W
The total energy (E) in the pulse can be found by integrating the instantaneous wattage over time:
∞
E = ∫ W(t) dt
0
The following methods may be used to compute the total integrated energy:
Numerical Integration:
Finding the area under the curve in figure 7 is the equivalent procedure for determining pulse energy.
Choose an appropriate time interval, dt, and perform the summation:
N N
E = ∑ Wixdt = (dt/S)∑ Vi
I=1 i=1
The error caused by this procedure is:
N
dE = (dt/S) Σ dVi
i=1
The error, in theory, is only dependent upon the value of ∑dVi, that is the cumulative random error of Vi.
This number should approach zero if data is carefully taken. The accurac y is also increased if the time
interval, dt, is minimized. Numerical integration can yield accurate results, but is a tedious task.
Initial Voltage Interpolation:
A method used to eliminate the tedious nume rical i ntegration ta sk is to project the thermal decay envelope
on to the voltage axis, determine the 1/e decay time constant T, and estimate the total energy value (E):
E = (Vo/S) x T
The change from thermal absorption to thermal transport phenomena near the peak causes difficulty in
accurately projecting the envelope on to the voltage axis introducing an error, dVo. Further, the
determination of the time constant T, introduces another error, dT. The total error is the sum of the two
errors.
dE = (Vo/S)dT + (T/S)dVo
The difficulty in eliminating the potential error makes this method typically less accurate than numerical
integration, but much faster in application.
PN10190T
36
Page 37
Peak Voltage Estimate:
The peak voltage method requires using an independent determination of total energy and referencing it
back to the peak voltage value, Vp.
For a given pulse, use the numerical integration method to obtain E. Note the peak voltage, Vp. Compute
the value, F
F = E/Vp
For the next pulse compute the total energy: E = F x Vp
The error in using this method yields: dE = FdVp + VpdF
The accuracy of this measurement depends upon the error in the original calibration, dF, and the error in
the peak voltage dVp. A careful numerical integration yields a value for dF near zero. The value of dVp
can be minimized by maintaining the geometry of the system (i.e. beam intensity, beam profile,
wavelength and environment) during operation to be the same as during calibration. Under controlled
circumstances, the peak method accuracy usually falls between the numerical integration and initial
voltage interpolation methods.
REMOTE INTERFACE LANGUAGE
The remote interface language for the S310 is compatible with standard IEEE488.2 and it also works with
the RS-232 remote interface.
A. RS-232 Connector
The RS-232 connector is a 9 pin subminiature connector on the instrument rear panel. The pin out and
pin descriptions are shown below:
Pin 1 Unused
Pin 2 Data In (RXD)
Pin 3 Data Out (TXD)
Pin 4 DTR (is generated)
Pin 5 Ground
Pin 6 DSR (is ignored)
Pin 7 RTS (is marking)
Pin 8 CTS (is evaluated if requested)
Pin 9 Unused
PN10190T
37
Page 38
B. RS-232 Specifications
Type: EIA-RS232C
Method: Half-duplex, Asynchronous
Transmission: Bi-directional
Format: 300, 1200, 2400, 9600, 19200 baud rate selectable
Data bits: 7
Parity bit: Even, Odd, or None
Stop bit: 1
Code: ASCII
Total no. of bits: 10
* The start bit counts for the first bit. Therefore, if you choose no parity you must have 2 stop bits.
C. Remote Interface Language Syntax
Remote interface m essages consis t of zero or mor e commands or queries, separated by semicolons and
terminated by a linefeed (IEEE488) or a carriage return (RS232). A command or query consists of a
command or query header followed by zero or more arguments separated by commas. Messages must be
less than 75 characters.
Example:
cmd1 arg1;cmd2arg1,arg2;...cmdN arg1
The queries RPT?, SND?, COL?, and *OPC?, and the commands COL and *OPC are intended to be
placed as the last command in the message. Placing them elsewhere will not result in harm to the
instrument, but it may produce results which seem unusual.
Queries which have not finished will be aborted by the receipt of additional commands or queries. This
will result in Query Errors in the IEEE488 interface. If a RP T?, COL?, or SND? query is immediately
followed by another command, it is likely tha t no data will be transmitte d. If COL is followed by another
command, statistics gathering will be halted unless the command *WAI appears between the two
commands.
D. Remote Interface Language Common Commands and Queries
Most commands and queries may be used with either the RS232 interface or the IEEE488 interface. Some
commands, however, work only with the IEEE488 interface. Some commands require that a password
has been entered. A few commands are archaic, but were left in because no harm was done, and they may
be needed again if multi-channel units are constructed again.
The commands which work in both interfaces are presented first , followed by the password protected
commands peculiar to the IEEE488 interface, followed finally by the archaic commands. In the
descriptions which follow, the command (or query) will be presented first, followed optionally by one or
more arguments, separ ated by commas. When the vertical bar "|" is used i n an argument, it denotes the
word "or", and signifies that one and only one of the items separated by bars ma y be inserted into the
PN10190T
38
Page 39
space of the argument. For example, the command: XXXA|B,C has two arguments, the first of which
may be `A' or `B', and the second of which is t he letter `C'. W here a space ch aracter is required by the
syntax, it is represented with an underline character.
Commands in this group are available from both interfaces:
*IDN?
This query takes no arguments, and returns a comma separated collection of four strings, describing
respectively the manufacturer of the instrument (Scientech, Inc.), the model number (S310), the serial
number, and the firmware version number.
*SAV_<dec num>
This is the IEEE488.2 common command for saving the instrument setup. The number may be an integer
from 1 through 4. When used from either remote interface, it saves the current configuration to the
numbered save area. The next time the instrument is powered up or the *RST command is issued with the
same save area as the default, or the RCL command is executed for the same save area, the configuration
will be restored to the values saved.
SENS_PYRO|CAL|PHR|ULTRA
Sets the detector type.
SENS?
Returns the detector type (PYRO|CAL|PHR|ULTRA).
SND?
The SND? query causes the instrument to send the next reading to the remote interface. When using a
calorimeter in the joules mode, the SND? query must be sent to the S310 before the calorimeter is pulsed
by the laser.
RPT?
This query causes the remote interface to begin sending a sequence of comma-separated readings. A new
reading is sent each time one is taken by the instrument. This activity will continue until the remote
interface is interrupted by a new command or query, or the instrument is turned off.
COL _<dec num>
This command causes the instrument in energy mode to begin collecting readings for statistical analysis
indicated by the response COLLECTING... This command continues until <dec num> data points are
collected, or until it is interrupted by another command. If it is interrupted, no statistics are computed. If
it terminates normally, a carriage return and line feed occurs, and it computes values for the mean, max,
standard deviation, and coefficient of variation of the data points collected. Those values may then be
PN10190T
39
Page 40
PHR 2, 5, 9 mm
Long Mode
1
1.000 m
300.0 m
150.0
3.000 m
300.0 μ
2
10.00 m
300.0 m
150.0
30.00 m
300.0 μ
3
100.0 m
300.0 m
150.0
300.0 m
300.0 μ
4
1.000
3.000
150.0
3.000
3.000 m
5
10.00
30.00
150.0
30.00
30.00 m
PHR 2 mm
Short/Int Mode
PHR 5 mm
Short/Int Mode
PHR 9 mm
Short/Int Mode
PHR 5 mm
Painted
PHR 9 mm
Painted
1
3.000 μ
3.000 μ
3.000 μ
3.000 μ
3.000 μ
2
30.00 μ
30.00 μ
30.00 μ
30.00 μ
30.00 μ
3
300.0 μ
300.0 μ
300.0 μ
300.0 μ
300.0 μ
4
300.0 μ
3.000 m
3.000 m
3.000 m
3.000 m
5
300.0 μ
30.00 m
30.00 m
3.000 m
30.00 m
accessed with the queries mean?, max?, sigma?, and cv?. The individual data points may be dumpe d with
the dump? query.
RANGE_<dec num>|auto
Sets the range to the decimal number supplied, or to the closest legal range allowed for the current detector
and configuration. If the word auto is supplied, this command activates the auto-range feature i f it is
supported for the current detector and configuration.
RANGE?
The RANGE query returns the decimal number between 1 and the maximum range of the instrument. The
meaning of the number varies, depending upon the type of detector, and the configuration of the
instrument. (Generally speaking, there are 5 ranges, though not all of them are used for every detector).
Table 4. S310 Range Selection Using Remote Interface Language
Range Cal 25 mm Cal 50 mm Ultra Pyro 25/50 mm
Range
MEAN?
If stat istics have been collected from the front panel, or by the COL or COL? commands, this command
returns the mean value of the collection. If no statistics have been collected, it returns to zero.
MIN?
If statistics have been co llected from the fro nt panel, or b y the COL or COL? commands, this command
returns the minimum value of the collection. If no statistics have been collected, it returns to zero.
MAX?
If statistics have been collected from the front panel, or by the COL or COL? commands, this command
returns the maximum value of the collection. If no statistics have been collected, it returns to zero.
PN10190T
40
Page 41
SIGMA?
If statistics have been collected from the front panel, or by the COL or COL? commands, this command
returns the standard deviation of the collection. If no statistics have been collected it returns zero.
CV?
If statistics have been collected from the front panel, or by the COL or COL? commands, this command
returns the coefficient of variation of the collection. If no statistics have been collected it returns zero.
PULSES?
Returns the number of pulses collected in the last run.
DUMP?
If statistics have been collected from the front panel, or by the COL or COL? commands, this command
returns all of the data values in the collection. If no statistics have been collected, it does nothing.
BAUD_<dec num>
If the <dec num> takes any of the values 300, 1200, 2400, 9600, or 19200; or any of their abbreviations
3, 12, 24, 96, 192, this command sets the RS232 baud rate to the corresponding value.
PARITY EVEN|ODD|NONE
Sets the parity of the RS232 interface to even parity, odd parity, or no parity, as specified.
HANDS XON|CTS|NONE
Sets the handshake method of the RS232 interface to XON/XOFF,CTS, or NONE as specified.
IO RS232|IEEE
On units with both IEEE and RS232 interfaces, this command makes the named interface the S AVED
active interface. The next time the instrument is powered up, or its configuration is restored from the
configuration save set active when this command was issued, the specified interface becomes the acti ve
remote interface.
*RST
This is the IEEE488.2 common command by the same name. When executed from the RS232 interface,
it has the effect of restoring the saved instrument configuration from the current save area. It has the
additional function in the IEEE488 interface of forcing the interface into the OCIS state and the OQIS
state.
METER ON|OFF
PN10190T
41
Page 42
Activates or deactivates the analog meter.
MODE VOLTS|ENERGY|AVGE|AVGP
If the specified mode is valid for the detector in use, the software changes mode to the one specified.
MODE?
Returns the instrument's operating mode. The possible responses are VOLTS, ENERGY, AVGE,
and AVGP.
*TST?
This is the IEEE488.2 common query. From either interface, it causes ROM checksum, and nondestructive RAM test to be run. If the tests succeed, this query returns zero. If the ROM test fails, a 1 is
returned. If the RAM test fails, a 2 is returned.
*OPC?
This is the IEEE488.2 common query. From either interface, it waits until no overlapping command is in
progress, and then it returns a 1.
*WAI
This is the IEEE488.2 common command. From either interface, it causes the command processor to
wait until any overlapping command is finished before continuing to process commands.
CLR
This command halts RPT?, SND?, or COL? commands in progres s. I t a lso te rmin ate s sta tistics gathering
started by the COL command. This command is issued internally whenever a carriage return that is not
preceded by a command is typed into the RS232 interface. When the IEEE488 interface receives a linefeed
terminated message with no commands, it too generates the CLR command.
*RCL_<dec num>
This is the IEEE488.2 common command. It takes the single numeric argument [1-4], and restores the
instrument configuration to that configuration stored in the corresponding save area. It also places the
IEEE interface into the OCIS and OQIS states.
COUNT_<dec num>
Sets the number of pulses making up each average in average energy mode.
COUNT?
Returns the number of pulses making up each average in average energy mode.
PN10190T
42
Page 43
CALIB_<dec num>
Sets the calibration constant (Volts/Joule) for a pyro detector, (Volt/Watt) for an Ultra Series detector or
the time constant for a calorimeter. Note: This command is context sensitive; the detecto r type mus t be
set by the SENS command before this command is executed.
CALIB?
Returns the constant for a pyro detector, or the time constant for a calorimeter. Note: This command is
context sensitive; the returned value depends on the detector type set by the SENS command.
ATTEN_<dec num>
Sets an attenuation factor for the current detector.
ATTEN?
Returns the attenuation factor for the current detector.
ZERO
This command zeros the power baseline for a calorimeter.
PSWD_<dec num>
Enables password protection commands if the password is entered correctly.
PSPEED HF|BL
Sets the electronic response speed according to the repetition rate capabilit y of the pyroelectric detector
type - a "black" coated absorbing crystal, or uncoated "high frequency" absorbing crystal.
PSPEED?
Returns the pyro speed, possible responses are BL and HF.
CSPEED_<dec num>
Sets the watts mode display response speed for calorimeters and Ultra Series detectors.
CSPEED?
Returns the watts mode display response speed for calorimeters and Ultra Series detectors.
CDELAY_<dec num>
Sets the "Calorimeter Delay" function of the meter.
PN10190T
43
Page 44
CDELAY?
Returns the "Calorimeter Delay" function of the meter.
E. IEEE488 Specific Commands
These commands may be used only from the IEEE488.2 interface. They are all members of the collection
of the so-called "common Commands" described in the IEEE488.2 standard.
*CLS
Clears the Standard Event Status Register and forces the device into Operation Complete Command Idle
state.
*ESR?
Returns a decimal number which is the value of the Standard Event Status Register. Reading the register
clears it.
*ESE_<dec num>
Sets the bits of the Standard Event Status Enable Register to the binary representation of the decimal
integer supplied.
*ESE?
Returns a decimal number representing the contents of the Standard Event Status Enable Register.
*SRE_<dec num>
Sets the bits of the Standard Request Enable Register to the binary representation of the decimal integer
supplied.
*SRE?
Returns a decimal number which represents the contents of the Service Request Enable Register.
*OPC
Sets the "Operation Complete" event bit in the Standard Event Status Enable Register when pending
device operations have been completed.
*STB?
Returns a decimal number which is the value of the IEEE488.1 status byte and the Master Summary Status
message.
PN10190T
44
Page 45
SCIENTECH CALIBRATION SERVICE
Scientech recommends that a complete calibration be performed annually to verify system accuracy.
Please visit Scientech’s website at www.scientechinc.com to obtain an RMA (returned material
authorization) number and complete the form noting that you are requestin g calibration of your detector.
You may also contact our Product Service Department at (800)525-0522 or (303)444-1361 or Fax
(303)444-9229 or email inst@scientechinc.com. Be prepared to provide model number, serial number
along with contact information. to arrange for a NIST traceable, factory calibration. Scientech calibrates
the detector with its power meter for the same price.
LIMITED WARRANTY
Scientech warran ts and r epres ents t hat th e las er p ower m easur ement p rodu ct w ill b e free f rom d efe cts in
design, materials and workmanship and conform with applicable Scientech product specifications for a
period of three (3) years. The product warranty period begins on the date of shipment from Scientech.
Scientech warrants that its products shall conform to applicable Scientech specifications and drawings and
will meet all the functional and performance requirements when properly installed, operated, and
maintained in accordance with Scientech’s operating manual. Warranty does not extend to any Scientech
products that have been subjected to misuse, abuse, or accidents, or improper installation, maintenance or
applications, repaired by unauthorized personnel, or Products in which the tamper proof sticker has been
removed or broken.
During the warranty period, Scientech will repair, or at its option replace at no charge, components that
prove to be defective. The product must be returned, shipping prepaid, to Scientech's authorized repair
facility. Products repaired by Scientech’s authorized repair personnel/facilities will be warranted against
defects in the repaired component and workmanship for a period of 365 days from the date of shipment
of the repaired Product.
RETURN MATERIAL PROCEDURE
Should it become necessary to return any product to Scientech for any reason including calibration, please
visit Scientech’s website at www.scientechinc.com to obtain an RMA (returned material authoriz ation)
number and complete the form. You may also contact our Product Service Department at (800)525-0522
or (303)444-1361 or Fax (303)444-9229 or email inst@scientechinc.com. Be prepared to provide model
number, serial number, and a description of the problem along with contact information. Frequently we
can provide self-help information which will eliminate the need for returning the product.
If product return is required, please pack the items in the original box and packing material. As an alternate,
place the equipment in a snug-fitting box, and then pack that box in a larger box with at least four inches
of packing material. Scientech does not assume responsibility for products damaged during shipping and
shipping damage will not be treated as a warranty repair. Please include a point of contact, email address,
and phone number of the person we should contact regarding repair questions.
Normally, products are repaired and shipped within five (5) business days following receipt of the product
at the authorized service facility. The repair turn-around time could vary depending on the workload.
Shipping Address: Scientech, Inc.
PN10190T
45
Page 46
Product Service Department
5649 Arapahoe Ave.
Boulder, Colorado 80303 U.S.A.
DISPOSAL OF ELECTRONIC EQUIPMENT
Scientech recommends the following for disposal of electrical and electronic equipment:
1. The best option is to reuse the equipment in its entirety.
2. Where the equipment cannot be reused in its entirety, priority should be given to reuse of
its subassemblies and components.
3. Where reuse is not appropriate, electrical and electronic equipment, including batteries,
should be recycled according to local ordinances.
4. Waste electrical and electronic equipment should never be mixed with municipal waste.
PN10190T
46
Page 47
SCHEMATICS
PN10190T
47
Page 48
PN10190T
48
Page 49
PN10190T
49
Page 50
PN10190T
50
Page 51
PN10190T
51
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.