schwa-medico EMP 4 PRO Instruction Manual

Page 1
Art.-Nr. 200506
EMP 4 PRO
Four-channel transcutaneous muscle and nerve stimulator
Instruction Manual Art.-No. 101447-V01
Art.-No. 104061
Page 2
schwa-medico EMP 4 PRO
Contents
General Information ...................................3
Intended use .............................................3
Safety instructions .....................................3
Contra indication ........................................3
Side effects ................................................4
Description of the symbols ........................4
Technical specifications ............................5
Pulse form ..................................................5
Description of the EMP 4 PRO ...................6
Description of the display ..........................6
Display of electrical current form ..............7
Operating the EMP 4 PRO ..........................7
Power mode ...............................................7
Using the 220-Volt mains adapter .............7
Using the battery .......................................7
Battery level ...............................................8
Recharging the battery ..............................8
Switching ON the EMP 4 PRO ...................8
Connecting the cables
and the electrodes .....................................8
Selecting a program ...................................8
Selecting an area of stimulation ...............9
Setting the intensity ..................................9
Pausing the stimulation ...........................10
Automatic change of the sequences .......10
Manual change of the sequences ...........10
Stopping the stimulation .........................10
Switching OFF the EMP 4 PRO ................10
Electrode recognition
for enhanced safety .................................11
Overview pre-set programs .....................12
Programs for upper and lower limbs .......13
User programs for own settings ..............13
Overview user programs ..........................14
Setting parameters for user programs ....15
Acoustic signal .........................................16
Information on total
stimulation duration .................................16
Settings, modifications and repairs ........16
Warranty ..................................................16
Guarantee.................................................16
Maintenance and cleaning ......................16
Classification of the device .....................16
Technical check ........................................17
Delivery content .......................................17
Accessories ..............................................18
Stimex self-adhesive electrodes .............18
Silicone electrodes ..................................18
MedicoBack..............................................19
Stimex garment electrodes ......................19
Vaginal probes ........................................20
Cables .....................................................20
Setting of intensity levels ........................20
Principle of dynamic stimulation .............21
Positioning of the electrodes
for dynamic stimulation ...........................21
Short instruction use ................................23
Electrode positioning ...............................24
Page 3
schwa-medico EMP 4 PRO
General Information
Intended use
The EMP 4 PRO device has been designed for the transcutaneous stimulation of muscles and nerves and may not be used for any other purposes.
Safety instructions
Please read the operating instructions carefully before using the EMP 4 PRO.
• The EMP 4 PRO is only to be used for stimulation purposes on one patient at any one time.
• Only use the EMP 4 PRO with original accessories.
• The size of the electrodes should to be not less than 2 cm².
• The electrodes should be positioned in such ways that the current flow will not cross through the heart.
• Keep the EMP 4 PRO away from water or other liquids. Do not drop the EMP 4 PRO, do not use it inappropriately or expose it to extreme temperatures or high levels of humidity (not less than 10 °C or more than 40 °C or a relative humidity of more than 90 %).
• Do not use the EMP 4 PRO if it is not working properly or if it has been damaged in any way.
• Be careful when using the EMP 4 PRO near or on children. Keep the unit away from children.
• Always store the EMP 4 PRO in its case to protect it from damage and dust.
• To avoid reciprocal interference, the EMP 4 PRO should not be operated in the vicinity of other electronic devices. If this is not possible, the various unit functions should be closely monitored during operation so that proper use in accordance with the regulations is ensured.
• The simultaneous connection to the patient of a high frequency surgical unit can lead to burning under the electrodes or around the probe.
• Operating close (for example 1m) to short wave and microwave units or mobile telephone systems can cause fluctuations in the output values of the electric current stimulation unit.
• Do not use the machine during the simultaneous operation of other machines, when driving or when asleep.
Contra indication
Please check with your physician before using the EMP 4 PRO if one of these apply to you:
• Patients with electronic implants such as heart pacemakers, defibrillators or pumps
• Patients with cardiac rhythm disorders
• During pregnancy
Page 4
schwa-medico EMP 4 PRO
• Patients subject to seizure disorder / epilepsy
• Patients with skin disorders in the vicinity of the electrodes
• Patients with malignant diseases in the region of application
Side effects
• Pain caused by the stimulation: The stimulation can be perceived as uncomfortable/ painful if the electric current is too intensive or if the electrodes have been positioned unfavorably. Unpleasant side effects can be avoided by adjusting the intensity, using another program with other parameters and the possible re-positioning of the electrodes.
• Skin intolerance which can result from the electrodes, the electrode gel or the electric current impulses. The physician should be consulted in cases of reddening, burning, itching or blistering under the electrodes or in the vicinity of the electrodes. Slight skin reddening of short duration in the area of the electrodes following stimulation is quite normal because blood circulation has been improved by the effects of the stimulation.
• Muscular pain: If the stimulation has been too intense or too long muscular aches in terms of sore muscle can occur. Shorten the treatment time and the intensity of stimulation at the beginning of muscle stimulation to avoid muscle ache.
Description of the symbols
Attention: Read accompanying documents, especially user manual!
BF type application part. Protection against electric shocks.
The year of construction of the product follows this symbol.
The article or order number of the product follows this symbol.
The serial number of the product follows this symbol.
This equipment is marked with the recycling symbol. It means that at the end of the life of the equipment you must dispose of it separately at an appropriate collection point and not place it in the normal domestic unsorted waste stream. This will benefit the environment for all.
By labelling with CE certificate, the manufacturer states that the product meets all active requirements of the regarding EU Directive. A conformation evaluation process has been successfully completed. The code number of the conformation evaluation process is given in accordance with the CE labelling of the involved notified body.
Page 5
schwa-medico EMP 4 PRO
This product works with a nickel metal hydride rechargeable battery. The rechargeable battery must be recycled. At the end of the working life of the rechargeable battery you must dispose of it separately at an appropriate collection point and not place it in the normal domestic unsorted waste disposal. In the interests of environmental protection please observe those laws and regulations of your country which relate to rechargeable battery disposal.
Technical specifications
Four-channel nerve and muscle stimulator with electrically insulated channels, constant current characteristic, output short circuit control element (AKS), 60 integrated pre-set programs and 8 user programs.
Output current: 100mA (with an actual resistor of 1kΩ) Frequency range: 1-120 Hz Pulse width: 50-350 μs Power supply: - mains supply transformer 12V / 1.25A or
- 9.6 V NiMH rechargeable battery Dimensions: 14.5 x 9.2 x 4.2 cm Weight approx. 308 g (incl. rechargeable battery)
Pulse form
Biphasic impulses with AKS
Biphasic symmetric impulses with pause and AKS
NiMH
U
t
I
t
on 1 k real
on ANSI 200
U
t
U
t
I
t
on 1 k real
on ANSI 200
Page 6
schwa-medico EMP 4 PRO
Description of the EMP 4 PRO
1. Display
2. P key: program selection key
3. 4 + key: program start key and simultaneous increase of all channels
4. key: selection of stimulation area / Pause + EDIT function for USER programs
5. Keys to adjust the intensities of each
channels ▲ Increasing intensity
▼ Decreasing intensity
6. ON/OFF key
7. Battery compartment
8. Output socket for the battery charger / main circuit transformer
9. Output sockets for the cables
10. Main circuit transformer / charger
Description of the display
1. Intensity of channel 1
2. Intensity of channel 2
3. Intensity of channel 3
4. Intensity of channel 4
5. Battery level
6. Program in use
7. Stimulation time
8. Sequence number
9. Electrical current form
10. Stimulation area
1
4
2
3
10
6
9
5 7 8
1
5
4
2
3
6
8
7
10
9
8
Page 7
schwa-medico EMP 4 PRO
Display of electrical current form
Stimulation pause
Stimulation rising ramp
Stimulation work phase
Stimulation falling ramp
Biphasic asymmetric impulse
Biphasic symmetric impulse
Biphasic symmetric impulse with pause
Operating the EMP 4 PRO
Power mode
You can use the device directly from the 220 Volt mains or with the rechargeable battery in the device.
Using the 220-Volt mains adapter
• Connect the cable to the central supply socket on the front face of the device between the channel outlets 2 and 3 (number 8).
• Connect the mains adapter to your mains circuit socket.
The device operates from the mains supply whether or not a rechargeable battery has been inserted.
Using the battery New devices:
The battery is already inserted in new devices. Please fully charge the battery before first use of the EMP 4 PRO.
Otherwise:
Open the battery compartment, plug the battery into the connector inside the compartment (upside right), place the battery in the compartment, close the compartment. The device is ready for battery mode operations. The mains power adapter is only necessary to recharge the battery.
Page 8
schwa-medico EMP 4 PRO
Battery level
The battery symbol on the display indicates the battery level. A completely charged battery shows all bars in the symbol. The fewer the bar symbols, the lower the battery charge level.
Recharging the battery
The battery must be recharged if the battery display indicates empty.
• Switch off the device.
• Ensure that the battery is in its compartment, correctly inserted and the compartment closed.
• Connect the mains adapter to the socket in the centre of the upper part of the device (number 8).
• Connect the mains adapter to the mains circuit.
• The display shows an increasing number of bars. The charging time lasts up to 3 hours. The charger stops automatically once the battery is fully charged.
• To check if the battery is fully charged, disconnect the device from the mains adapter and switch on the device. The battery indicator on the top left side of the screen displays the battery level.
Switching ON the EMP 4 PRO
Press the
key to start the device (number 6). The program shown on the display is the last program used. The device automatically switches OFF after 2 minutes of disuse.
Connecting the cables and the electrodes
• Connect the electrodes with the cable(s) (always 2 electrodes per cable).
• Connect the cables to the EMP 4 PRO on socket no. 9.
• Place the electrodes on the desired area (see electrode placement).
Selecting a program
Select a program by pressing the P key. To scroll through the programs, press the P key as often as necessary until you have selected the program of your choice. To return to the previous program, press the key of channel 1 (far left side).
Page 9
schwa-medico EMP 4 PRO
Selecting an area of stimulation
Depending on the selected program you can choose between upper and lower limbs. The body symbol on the display is half colored, depending on the selected limbs. The pulse width is increased for lower limbs.
Press the
key to select the area of stimulation.
Programs with a possibility of upper and lower limbs stimulation:
P1 - P4, P6, P17, P21 - P26, P32, P38, P40 - P47, P49 - P60
Setting the intensity
Start the stimulation by increasing the intensity with the key of the channel(s) connected to the electrodes. Alternatively you can increase the intensity of all four channels simultaneously by pressing the “4 +” key. You can then adjust the intensity of each channel by using the and keys.
You increase the intensity with the key. You decrease the intensity with the key.
The intensity of each channel is displayed on the screen. The intensity can be set between 0 and 100 mA.
Setting the intensity of agonist/antagonist program (P37 and P67):
When you are using agonist/antagonist programs (P37 and P67), first set the intensity of channel 1 and 2 and then set the intensity of channel 3 and 4. Active channels are identified by flashing intensity values. If the intensity of all channels is not changed over a three seconds period, the device starts with the agonist/antagonist stimulation. Please note that the stimulation is alternating between channel 1 and 2, respectively 3 and 4. If you are stimulating with dynamic programs (P56 to P60 and P65 to P66) or agonist/ antagonist programs (P37 and P67), active channels are identified by flashing intensity values.
Page 10
schwa-medico EMP 4 PRO
10
Pausing the stimulation
You can temporarily stop the stimulation by pressing the
key during the stimulation. The word “PAUSE” then flashes on the display. You can then adjust the position of the electrodes. Press
again to re-start the stimulation. The intensity levels slowly increase to the level
set before the pause.
Automatic change of the sequences
Some programs consist of several sequences. Each sequence uses different parameters. The device changes from one sequence to the other automatically. This change is indicated by an acoustic signal. The sequence number is displayed on the screen at the top right side next to the “S” symbol. The intensity of the stimulation is then automatically adjusted depending on the sequence. It may be necessary to re-adjust the intensities to the desired level.
Manual change of the sequences
If you want to skip one or more sequence(s) manually, first press
to temporarily stop the stimulation (see pausing the stimulation), then press the “4+” key to select the sequence of your choice. The sequence number is displayed on the screen at the top right side next to the “S” symbol.
Press
again to re-start the stimulation. For safety reason the intensity is set back to zero if you skip a sequence. Adjust the intensity by pressing the “4+” or the “” and “” keys (see setting the intensity).
Important note:
The stimulation time can only be changed for programs with a single sequence. The time cannot be changed for programs with an asterisk
* (see program overview).
Stopping the stimulation
You can stop the stimulation at any time by pressing the P key or the
key. When the therapy time is completed, the stimulation stops automatically. The end of stimulation is indicated by a double acoustic signal.
Switching OFF the EMP 4 PRO
Press the
key to switch OFF the device. This is indicated by an acoustic signal. The
device switches OFF automatically if:
• the power supply is too low (battery mode only)
• the device has not been used for 2 minutes.
Page 11
schwa-medico EMP 4 PRO
11
Electrode recognition for enhanced safety
The EMP 4 PRO constantly monitors the correct connection of electrodes and cables. The EMP 4 PRO automatically stops the stimulation if:
• an electrode detaches itself from the skin
• electrodes are not attached correctly in pairs
• electrodes are not conductive enough
• the connection between the cables and the electrodes is interrupted
• the connection between cables and device is interrupted.
The intensity then returns to 0; this is indicated by a short acoustic signal.
Page 12
schwa-medico EMP 4 PRO
12
Overview pre-set programs
No.
PROGRAMS FOR PAIN MANAGEMENT
1 Gate Control 80 Hz*
2 Gate Control 100 Hz*
3 Endorphine 2 Hz*
4 Endorphine Burst*
5 Modulation
6 Gate Control + Endorphine*
7 Cervicalgy
8 Dorsalgy
9 Lombalgy
10 Ischialgia / sciatic pain
11 Torticollis
12 Gonarthrosis
13 Coxarthrosis
14 Epicondylitis
15 Algodystrophy
16 Hyperalgia
17 Phantom limb pain*
18 Cervicobrachial nevralgia
19 Carpal tunnel syndrome
20 Arthritis of the hand
No. PROGRAMS FOR REHABILITATION
21 Amyotrophy 1st phase*
22 Amyotrophy 2nd phase*
23 Amyotrophy prevention*
24 II A muscle fiber strengthening*
25 II AB muscle fiber strengthening*
26 II B muscle fiber strengthening*
27 Combination: active wrist joint motion
+ pain management
No. PROGRAMS FOR REHABILITATION
28 Combination: active elbow joint motion
+ pain management
29 Combination: active knee joint motion
+ pain management
30 Combination: active ankle joint motion
+ pain management
31 Combination: Hemiplegia of the
shoulder + pain management
32 Spastic limb stimulation*
33 Urology: urge incontinence
34 Urology: mixed incontinence
35 Urology: stress incontinence
36 Urology: pain control
37 Agonist/antagonist stimulation
No. VASCULAR PROGRAMS
38 Venus reflux*
39 Lymphatic drainage (heavy legs)
40 Cramp prevention*
41 Capillarisation*
No. PROGRAMS SPORT
42 Endurance*
43 Resistance*
44 Strength*
45
Explosive strength
*
46
Active recovery
*
47
Cleansing
*
48 Relaxing percussion
Page 13
schwa-medico EMP 4 PRO
13
No. FITNESS PROGRAMS
49
Hypertrophy
*
50
Warm-up
*
51
Toning
*
No. AESTHETIC PROGRAMS
52
Cellulo lipolysis
*
53
Thigh-buttock firming
*
54 Adipose tissue vascularisation 1*
55
Adipose tissue vascularisation 2
*
No. RELAXATION PROGRAMS
56 Slow wave relaxation massage (soft)*
57 Fast wave relaxation massage (soft)*
58 Slow wave relaxation massage
(against muscle spasm)*
59 Fast wave relaxation massage (against
muscle spasm)*
60 Zip wave relaxation massage (soft)*
No. USER PROGRAMS
61 TENS continuous mode
62 Frequency modulation
63 EMS excito-motor stimulation
64 Endorphin burst stimulation
65 Massage through dynamic stimulation
(one direction)
66 Massage through dynamic stimulation
(both directions)
67 EMS agonist/antagonist stimulation
68 2+2 mode
(Ch. 1 & 2: EMS / Ch. 3 & 4: TENS)
Programs for upper and lower limbs
Programs marked with an * are programs with a possibility of upper and lower limbs stimulation
( key).See Page 9.
User programs for own settings
The EMP 4 PRO has 8 user programs. Each user program corresponds to a type of stimulation and consists of one single sequence. The stimulation time can be set between 10 and 90 minutes.
Page 14
schwa-medico EMP 4 PRO
14
No. USER Program type Possible parameters
61 TENS continuous mode • Frequency (1-100 Hz)
• Impulse width (50-250 μs)
• Stimulation time (10-90 min)
62 Frequency modulation • Lowest frequency: from 1 to 100 Hz
• Highest frequency: from the lowest frequency + 10 Hz to 100 Hz
• Impulse width (50-250 μs)
• Stimulation time (10-90 min)
63 EMS excito-motor
stimulation
• Frequency (1-60 Hz)
• Impulse width (50-350 μs)
• Rising ramp time: 2 s
• Plateau working time: 6 s
• Falling ramp time: 1 s
• Pause time: 6 s
• Stimulation time (10-90 min)
64 Endorphin burst
stimulation
• Frequency (1-100 Hz)
• Impulse width (50-200 μs)
• Rising ramp: 0.25 s
• Stimulation time (10-90 min)
65 Massage through
dynamic stimulation (one direction)
• Frequency (1-60 Hz)
• Impulse width (50-350 μs)
• Rising ramp time: 2 s
• Plateau working time: 1 s
• Falling ramp time: 1 s
• Stimulation time (10-90 min)
66 Massage through
dynamic stimulation (both directions)
• Frequency (1-60 Hz)
• Impulse width (50-350 μs)
• Rising ramp time: 2 s
• Plateau working time: 1 s
• Falling ramp time: 1 s
• Stimulation time (10-90 min)
67 EMS
agonist/antagonist stimulation
• Frequency (1-60 Hz)
• Impulse width (50-350 μs)
• Rising ramp time: 2 s
• Plateau working time: 3 s
• Falling ramp time: 1 s
• Stimulation time (10-90 min) You must first adjust the intensities of channels 1 and 2 and then channels 3 and 4. The intensities of the active channels flashes on the screen.
Overview user programs
Page 15
schwa-medico EMP 4 PRO
15
No. USER Program type Possible parameters
68 2+2 mode
Ch. 1 & 2: EMS Ch. 3 & 4: TENS
For channels 1 and 2:
• Frequency (half of frequency from channel 3 and 4)
• Impulse width (50-200 μs)
• Rising ramp time: 3 s
• Plateau working time: 1 s
• Falling ramp time: 3 s
• Pause time: 6 s
• Stimulation time (10-90 min)
For channel 3 and 4:
• Frequency (1-90 Hz)
• Impulse width (50-200 μs)
• Stimulation time (10-90 min) The pulse width must be the same for all 4 channels
Setting parameters for user programs
• Select a user program (61 to 68) depending on the desired type of therapy by pressing
the P key (see selecting a program).
• Press the
key for 3 seconds to enter the setting mode.
• The first parameter to be changed flashes.
• Press the and keys of any channel to change the parameters.
• Press
again to go to the next parameter.
• Press the and keys of any channel to change the parameters.
• Press
again to go to the next parameter.
• Press the and keys of any channel to change the parameters.
• Press
after changing the last parameter to save the changes in the user program.
Press P at any time if you want to abort the settings. The changed parameters up to that point are then not saved.
Page 16
schwa-medico EMP 4 PRO
16
Acoustic signal
It is possible to turn off the acoustic signal.
Simultaneously press the
key and the key of channel 3 for 3 seconds to enter the
setting mode.
Press the
key to turn the acoustic signal ON and OFF.
Press the P key to get back to the standard mode.
Information on total stimulation duration
The muscle stimulation unit saves the total stimulation duration.
Simultaneously press the
key and key of channel 4 for 3 seconds to enter the checking
mode.
Press the
key and the P key simultaneously to reset the timer to zero.
Press the P key to get back to the standard mode.
Settings, modifications and repairs
The manufacturer is only responsible for the safety and performance of the EMP 4 PRO when settings, adjustments, alterations and repairs are carried out by authorized persons and when the EMP 4 PRO is used in accordance with the operating instructions.
Warranty
Legal right of warranty is applied according to German Civil Code.
Guarantee
The manufacturer issues a guarantee of 12 months from the date of purchase. The guarantee does not apply in the following cases:
- damage due to improper handling
- defects the customer is aware of on the date of purchase
- damage caused by the customer
- for wearing parts and consumable supplies like, for instance, electrodes, batteries and cables.
Page 17
schwa-medico EMP 4 PRO
17
Maintenance and cleaning
No special cleaning or care agents are required for the EMP 4 PRO. Clean the unit with a soft, fluff-free cloth. Please ensure that no moisture permeates the unit. If moisture does permeate the unit, a technical check must be carried out before re-use.
Classification of the device
In accordance with the law on medical devices, the EMP 4 PRO is classified as being a Class IIa medical device.
Technical check
We recommend a technical check on the EMP 4 PRO every 24 months. This includes:
1. Confirming that user instructions have been included in the accompanying documentation.
2. Checking the equipment for completeness.
3. Visual check:
- for mechanical damage
- for damage to all cables and sockets
4. Functional Safety
- checking the output signals with a load resistance of 1 k real current and voltage
- checking the frequency
- checking the pulse width
These technical checks may only be carried out by suitably qualified persons. The results must be documented with the date and name of the person carrying out the check.
Delivery content
Ref. no. Article Description
104061 EMP 4 PRO 450781-0026 Rechargeable battery 104709 Main circuit adapter / charger 104769 Electrode cables set, type 7 (black, red, grey, yellow) 283400 STIMEX self-adhesive electrodes 50x50 mm 283600 STIMEX self-adhesive electrodes 50x90 mm Transport case 101447 Instruction manual
The EMP 4 PRO may be used together with all accessories mentioned in the chapter „Accessories“
Page 18
schwa-medico EMP 4 PRO
18
Accessories
The following accessories can be combined with the EMP 4 PRO.
Stimex self-adhesive electrodes
Art. No. Article Pieces
281000 Stimex 32 mm round 4 282000 Stimex 50 mm round 4 283400 Stimex 50x50 mm 4 281032 Stimex 50x50 mm 20 283600 Stimex 50x90 mm 2 281007 Stimex 50x90 mm 4 283000 Stimex 50x130 mm 2 283100 Stimex 80x130 mm 2 281027 Stimex sensitive 50x50 mm 4
Silicone electrodes (to be used with contact gel)
Art. No. Article Pieces
107090 Silicone electrode 20 mm round 2 107060 Silicone electrode 25 mm round 2 107075 Silicone electrode 40x28 mm 2 107035 Silicone electrode 56x28 mm 2 107020 Silicone electrode 75x30 mm 2 107055 Silicone electrode 90x35 mm 2 107011 Silicone electrode 38x45 mm 2 107010 Silicone electrode 48x48 mm 2 107050 Silicone electrode 70x65 mm 2 107070 Silicone electrode 70x140 mm 2 108000 Contact gel for silicone electrodes (60 g) 1
Page 19
schwa-medico EMP 4 PRO
19
MedicoBack
For the treatment of low back pain
Low back support bandage with pelotte and integrated electrodes to be used with pain management programs of the EMP 4 PRO. MedicoBack P-type offers patients lumbar vertebrae stabilization and support as well as easy stimulation provided by the integrated electrodes. The pelotte with its massage nubs enhances the pain relieving effect. MedicoBack P-type may be used for slight or acute pain as well as for chronic pain in the lower back area.
Art. No. Article Size Length
107034 MedicoBack P-type S 80 - 90 cm 107036 MedicoBack P-type M 90 - 100 cm 107037 MedicoBack P-type L 100 - 110 cm 107038 MedicoBack P-type XL 110 - 120 cm 107039 MedicoBack P-type XXL 120 - 130 cm
Stimex garment electrodes: gloves and socks
The gloves and socks used in combination with the EMP 4 PRO provide stimulation of the entire hand and/or foot and ankle and avoid the sometimes tiresome attachment of self-adhesive electrodes on the hand or the foot.
Gloves
Art. No. Article Size Pair
107014 Stimulation gloves S 1 107021 Stimulation gloves M 1 107022 Stimulation gloves L 1
Socks
Art. No. Article Size Pair
107023 Stimulation socks M 1 107024 Stimulation socks L 1
Page 20
schwa-medico EMP 4 PRO
20
Vaginal probes
To be used with URO programs only
Art.-No. Article Pieces
101471 Vaginal probe V21B (type 7) 1
V21B
25 mm Ø
10 10
Länge 95 mm
length
Cables
104765 Pair of electrode cables type 7 (black) 104766 Pair of electrode cables type 7 (red) 104768 Pair of electrode cables type 7 (grey) 104767 Pair of electrode cables type 7 (yellow)
Setting of intensity levels
Do not try to achieve a higher and higher level of intensity. Set the intensity in such a way that you get a comfortable feeling during the stimulation. Increase the intensity carefully to a maximum tolerable and then reduce it slightly step by step to a comfortable level. Note that the intensity settings change depending on the stimulation area and the time. This is normal and can be explained as follows:
• Resistance of the skin: Dry skin has a lower conductivity level than damp skin (due to sweating). The skin does
not show the same resistance to linear current on all parts of the body. For example, the resistance level of horny skin can be twice that of the skin in the hollow of the knee. The innervation differs greatly according to the area of the body. This results in a difference of sensitivity.
• Volume of the muscle mass: If you stimulate muscles with a high volume you will automatically need to set a higher
intensity than for the stimulation of small muscles.
Page 21
schwa-medico EMP 4 PRO
21
• Muscle fatigue: A tired muscle will only tolerate a lower intensity.
• Age of electrodes: Self-adhesive electrodes do not last forever. You can use them up for up to 60
stimulation sessions. This depends on the quality of the electrodes but also on the skin of the user. Old electrodes have a higher resistance and their conductivity levels decline. It is important to replace electrodes regularly.
• Adaption to the current: The nerves adapt rapidly to the current. It is normal to feel “less current” after a few minutes
of stimulation with specific parameters. In such a case feel free to increase the intensity in order to get the best results.
• Frequency and pulse width: Frequencies and pulse widths differ depending on the programs. There-fore the intensity
can not be set at the same level for all the programs even if the electrodes are placed in the same position. The higher the frequency is set, the lower you will be able to set the intensity.
The higher the pulse width is set, the lower you will be able to set the intensity.
Principle of dynamic stimulation
The stimulation is not administered simultaneously on all channels. The current flows from one electrode to the other just like a wave undulating on the body. This kind of stimulation is far more comfortable than traditional neuro-muscular stimulation. Dynamic stimulation can be used with high frequency parameters for better pain control or with low frequency parameters for optimized muscle stimulation. This treatment method also has an enhanced effect on lymphatic drainage stimulation.
Positioning of the electrodes for dynamic stimulation
Sideways Sideways movement: E.g gluteus, stomach muscles
Channel 1 Channel 2 Channel 3 Channel 4
Page 22
schwa-medico EMP 4 PRO
22
Vertical movement E.g. back, legs Falling
In order to apply a falling wave, the channels must be arranged as follows: channel 1 (black) channel 2 (red) channel 3 (grey) channel 4 (yellow)
Rising
In order to apply a rising wave, the channels must be arranged as follows: channel 4 (yellow) channel 3 (grey) channel 2 (red) channel 1 (black)
Dynamic stimulation on legs:
Place one electrode of each channel on each leg (size 5x9 cm).
Dynamic stimulation on the back:
Place one electrode of each channel on each side of the spinal column (size 5x9 cm).
Important note:
If you use electrodes with two connectors, always connect them to the same pole color of the cables (either both white or both red).
Channel 1
Channel 2
Channel 3
Channel 4
Channel 1 Channel 2 Channel 3 Channel 4
Channel 4
Channel 3
Channel 2
Channel 1
Channel 4
Channel 3
Channel 2
Channel 1
1
1
4
4
33
2
2
1
1
2
2
3
3
4
4
Page 23
schwa-medico EMP 4 PRO
23
Short instruction use
1. Connect the cables with the electrodes.
2. Plug in the cables into the sockets of the
EMP 4 PRO.
3. Place the electrodes on the skin.
4. Select a program by pressing the
P key. Scroll back in the program list by pressing the key of channel 1.
5. Select an area of stimulation by pressing
the
key. See upper or lower limb on the
display.
6. Increase the intensity of all channel
simultaneously by pressing the 4+ key.
7. Adjust the intensity of each channel
by pressing the and key of each channel.
Page 24
schwa-medico EMP 4 PRO
24
Electrode positioning
Electrode positioning for pain therapy (P1 to P7, P10 to P20)
Neck Trapezius Deltoid
1r
1b
2r
2b
1r
1b
2r
2b
1r
1b
2r
2b
Elbow Hip Knee
1r
1b
1r
1b
2r
2b
1r
1b
2r
2b
r = red (anode) b = blue (cathode)
Page 25
schwa-medico EMP 4 PRO
25
Sciatic pain, Ischialgia Ankle Wrist
1r
1b
2r
3r
4r
2b
3b
4b
1r
1b
2r
2b
1r
1b
r = red (anode) b = blue (cathode)
Page 26
schwa-medico EMP 4 PRO
26
Electrode positioning for pain therapy using dynamic stimulation
Back Lower back
1r
1b
2r
2b
3r
3b
4r
4b
1r
1b
2r
2b
3r
3b
4r
4b
r = red (anode) b = blue (cathode)
Page 27
schwa-medico EMP 4 PRO
27
Electrode positioning for rehabilitation (P21 to P37), vascular disease (P38 + P41), sports (P42 to P47), fitness (P49 to P51), aesthetics (P53 to P55)
Hand flexor Hand extensor Biceps
1r
1b
1r
1b
1r
1b
Trapezius Back Latissimus
r = red (anode) b = blue (cathode)
1r
1b
2r
2b
1r
1b
2r
2b
1r
1b
2r
2b
Page 28
schwa-medico EMP 4 PRO
28
Stomach muscle Gluteus Quadriceps
1r
1b
2r
2b
3r
3b
4r
4b
1r
1b
2r
2b
3r
3b
4r
4b
1r
1b2r2b
Flexor Foot flexor Foot extensor
1r
1b
2r
2b
1r
1b
1r
1b
2r
2b
r = red (anode) b = blue (cathode)
Page 29
schwa-medico EMP 4 PRO
29
Electrode positioning for urological rehabilitation (P33 to P36)
Positioning on the pubic bone
Anode (red) Cathode (blue) Anode (red) Cathode (blue)
Training with a probe
Vaginal probe Anal probe
Positioning on the perineum
Place the anode (red) on the pubic bone
Cathode (blue) Cathode (blue)
Page 30
schwa-medico EMP 4 PRO
30
Electrode positioning for anti-cellulitis treatment using dynamic stimulation (P52)
Gluteus Stomach muscle Thigh
1r
1b
2r
2b
3r
3b
4r
4b
1r
1b
2r
2b
3r
3b
4r
4b
1r
2r
3r
4r
1b
2b
3b
4b
r = red (anode) b = blue (cathode)
Page 31
schwa-medico EMP 4 PRO
31
Electrode positioning for vascular desease (P39 + P48), relaxation (P56 to P60), dynamic stimulation (P65 + P66)
Flexor Calf Flexor and calf
1r
1b
2r
2b
3r
3b
4r
4b
1r
1b
2r
2b
3r
3b
4r
4b
1r
1b
2r
2b
3r
3b
4r
4b
Back Quadriceps Low back
1r
1b
2r
2b
3r
3b
4r
4b
1r
1b
2r
2b
3r
3b
4r
4b
1r
1b
2r
2b
3r
3b
4r
4b
r = red (anode) b = blue (cathode)
Page 32
schwa-medico · Export Department Wetzlarer Strasse 41-43 · 35630 Ehringshausen · Germany Tel. +49 6443 8333-113 · Fax +49 6443 8333-119 Email export@schwa-medico.de · www.schwa-medico.com
Manufacturer: Pierenkemper GmbH Hoernsheimer Eck 19 · 35578 Wetzlar · Germany
Art.-Nr. 451600-0339
Stand: 2010-11-03
Loading...