ScanSense TC2000 Operating Manual

USER’S MANUAL
TC2000
PORTABLE TEMPERATURE
CALIBRATOR
A universal stable and accurate temperature well for
calibrating
Manual revisjon: June, 2002 Copyrighted © 2002
Page 1
TC2000 user’s manual
Thank you
for purchasing a Tek Know temperature calibrator. The Tek Know products are manufactured by Scan-Sense AS in accordance with our high quality standards in design, choice of components and workmanship in order to achieve maximum customer
satisfaction and to full our vision to be our customers “First Choice”.
The TC2000 is designed and manufactured by:
Scan-Sense AS Tel: + 47 33 36 30 00 Bekkeveien 163 Fax: + 47 33 36 31 80 N-3173 Vear www.scansense.no NORWAY post@scansense.no
Scan-Sense AS reserves the right to make improvments or alterations
to our products without incurring any responsibility to make the same improvments or alterations to products previously sold.
SCAN-SENSE AS
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form by any means, electronic, mechanical, photocopying, recording, or otherwise without the prior permission of Scan-Sense AS.
All efforts have been made to ensure the accuracy of this handbook. We at Scan-Sense are always striving to improve our products and handbooks, therefore we would greatly appreciate being informed of any errors found in our product or in its manual. The above notwithstanding, Scan-Sense can assume no responsibility for any errors in this handbook or of their consequenses.
Page 2
TC2000 user’s manual
Contents
1 Introduction ...................................................................................................... 5
1.1 Certication ..................................................................................................... 5
1.2 Warranty ......................................................................................................... 5
1.3 Conventions used in this document ................................................................. 5
1.4 Introduction to dry-block calibrators ................................................................. 7
2 Preparation ...................................................................................................... 8
2.1 Unpacking and inspection ............................................................................... 8
2.2 Items supplied with the unti ............................................................................. 8
2.3 Input power connection ................................................................................. 10
3 Getting started ............................................................................................... 11
3.1 Operation controls and connections .............................................................. 11
3.1.1 Software Startup ........................................................................................... 12
3.1.2 LCD display ................................................................................................... 13
3.1.3 Well and insert ............................................................................................... 13
3.1.4 Numerickeypad................................................................................................14
3.1.5 Power on / off ................................................................................................. 14
3.1.6 RS232 Communicatiion ................................................................................. 14
3.1.7 Sensor connection ......................................................................................... 14
3.2 Sensor positioning, connection and calibration accuracy .............................. 15
3.3 Procedure to run calibration .......................................................................... 16
3.4 Procedure to run a liquid bath calibration ...................................................... 17
3.4.1 Calibrating a thermostat .................................................................................18
3.4.2 Calibrating a sensor .....................................................................................18
3.5 Finish the calibration ..................................................................................... 21
3.6 Defaults values .............................................................................................. 22
3.7 Pre-calibration cleaning of well and insert ..................................................... 23
3.7.1 Procedure for cleaning the interior of the temperature well and insert .......... 23
3.7.2 Procedure for cleaning the exterior of the insert ............................................ 24
3.7.3 Instructions for handling Liquid-lled Insert ................................................... 24
4 Operational menus, controls and indicators .................................................. 25
4.1 Introduction .................................................................................................... 25
4.2 Main Menu ..................................................................................................... 25
4.3 Manual Calibration ........................................................................................ 26
4.4 Conguration ................................................................................................. 27
4.4.1 Reference seonsor selection ......................................................................... 27
4.4.2 Reference sensor conguration ................................................................... 27
4.4.3 Reference correction ..................................................................................... 28
4.4.4 Test sensor .................................................................................................... 30
Page 3
TC2000 user’s manual
4.5 System information ........................................................................................32
4.6 Settings ..........................................................................................................32
4.7 Automatic calibration ......................................................................................33
4.8 Semi automatic calibration .............................................................................35
4.9 Calibration report ........................................................................................... 37
4.10 Software upgrade utility ..................................................................................38
5 Technical tips ................................................................................................. 39
5.1 General information regarding accurate dry-block calibrations ..................... 39
5.2 The importance of testing instrument validation ............................................ 40
5.3 The importance of a traceable standard ........................................................ 40
5.4 Correct positioning of the probe in the temperature block ............................. 40
5.5 Calibrating a short-stem probe .......................................................................41
5.6 Using the calibrator as a reference when adjusting measuring devices
(meters/gauges) ............................................................................................ 42
5.7 The Time Constant inuence on temperature-well mass ............................... 42
5.8 When to take the calibration reading ............................................................. 42
5.9 The importance of cleanliness to attain high accuracy results ...................... 43
5.10 Insulating with glass wool to minimize cold/heat loss to surrounding air ....... 43
6 Specications ................................................................................................ 44
7 Electrical specications ................................................................................. 44
Page 4
TC2000 user’s manual
1 Introduction
1.1 Cerication
Scan-Sense AS certies that the TC2000 complies with its published list of specications at the time it was manufactured. Scan-Sense AS also certies that its calibration measurements are traceable to Norwegian Accreditation and to the calibration facilities of other International Standards Organization (ISO) members. Scan-Sense AS conrms that the TC2000 complies with the following standards:
EMC EN61326-1:1997+A1:1998: APP.A, immunity test requirements for equipment intended for use in industrial locations, as amended by EC Directive 9 2/31/Eec and the European Low Voltage Directive IEC 61010-1:90 + A1:92 + A2:95, amended by 93/68/EEC. To ensure compliance, please use screened/shielded serial communication leads (RS232).
1.2 Warranty
This product is guaranteed free from defects in material and workmanship for one (1) year from the date of shipment. During this warranty period, Scan-Sense AS will, as its option, either repair or replace the TC2000 should it prove defective. The product must be returned to a service facility designated by Scan-sense AS for warranty service or repair (see our worldwide adress le http:// www.scansense.no). The foregoing warranty will not apply to defects resulting from improper or inadequate maintenance by the purchaser, purchaser-supplied software or interfacing, unauthorized modication or misuse, operation exceeding the environmental specications for the TC2000, or improper site preparation or maintenance. No other warranty is expressed or implied by Scan-Sense AS, and Scan-Sense AS shall not be liable for any direct, indirect, special, incidental or consequential damages, whether based on contract, tort, or any other legal theory.
1.3 Conventions used in this document
We have provided this section of the User’s Manual to help you identify noteworthy symbols, terms and conventions used in this manual. Look for the following:
Page 5
TC2000 user’s manual
CAUTION:
Is used when a procedure may cause damage to equipment
or affect operation
WARNING:
Is used when a procedure may cause personal injury to the reader or other persons
A SYMBOL USED AS A WARNING TO DENOTE A HAZARD
CONFORMITÉ EUROPÉENNE
GROUND OR EARTH
Terms and denitions:
The term “insert” is interchangable with “insert adapter” as an equipment identier for the 2-inch diameter, 6-inch long aluminum insert pleced into the temperature well.
We dene calibration as being able to compare the ability of the equipment to perform to a known standard. Temperature calibration provides a means of quantifying uncertainties in temperature measurement in order to optimize sensor and/or system accuracy.
LCD is a term meaning a quarter (1/4) VGA LCD screen. CARE is a memory jogger to ensure best results and safe use of the equipment. Connect only to the correct power voltage/frequency (read the label in the case). Always clean the temperature well and insert before use, and always remove the insert from the temperature well before moving or tran­ sporting the TC2000. Replace worm connection cables (only with approved connection cables). Earthed (or grounded) power sources only should be used.
Page 6
1.4 Introduction to dry-block calibrators
Dry-block calibration units provide the reliable heat (or cold) source needed to verify the accuracy of industrial sensors such as RTDs, thermocouples, thermostats, bi-metal thermometers or liquid­ in-glass thermostats. The TC2000 is designed to provide all the advantages offered by “dry well” temperature calibrators: Convenience of size, weight, speed, portability temperature range, stability and software interaction via RS232 communication port. It is also designed to be “user friendly”.
The TC2000 has a numeric keypad, or softkey controls, for setting the well temperature. Menus are used to dene other data for the TC2000. The display shows the dened “Set temperature” and the resulting “Reference temperature” for the well temperature. Connection points at Channel A and B allow a variety of sensor hook­ ups and an RS232 connection for local PC or remote computer.
The dry well in the calibrator is heat or cold source, and a reference thermometer. That is, the thermometer (or sensor to be tested) is placed in the well and its reading is compared to the reading of the temperature in the calibrator’s display. The calibrator’s display shows both the dened temperature, as set by the operator, and the actual temperature that is measured by the internal reference sensor in the well.
Figure 1: Dryblockschematics
EXTERNAL REF.
CALIBRATOR DISPLAY
INTERNAL REF.
SENSOR TO BE TESTED
PELTIER ELEMENTS
HEATING BLOCK
COMPARISON
Page 7
TC2000 user’s manual
2 Preparation
2.1 Unpacking and inspection
* Inspect the packing container for any sign of damage. If the shipping
cartong is damaged, contact the carrier/shipper.
* Remove the unit from its packing container and inspect it for visible
signs of damage.
* Check that the equipment received matches the inventory. Report
ny discrepancies to Scan-Sense AS’s designated agent.
* Record any discrepancies from the inventory and/or visible damage
to the unit; see Figure 3 TC2000 orientation and location of items.
2.2 Items supplied with the unit
The cables, insert etc. that are used with the TC2000 are located in compartments within the unit. The TC2000 Parts List identies each by part name and part number.
TC2000 PARTS LIST DESCRIPTION OF ITEM: (1 OF EACH ITEM)
* GLASS WOOL, 1 PACK * TOOL, EXTRACTOR * CABLE, MAINSPOWER, 100 VAC DETERMINE TYPE REQUIRED, JAPAN * CABLE, MAINSPOWER, 110 VAC DETERMINE TYPE REQUIRED, USA * CABLE, MAINSPOWER, 230 VAC DETERMINE TYPE REQUIRED, EUR * CABLE, MAINSPOWER, 230 VAC, DETERMINE TYPE REQUIRED, UK * CABLE, TEST, 1 SET * USER’S MANUAL * CALIBRATION CERTIFICATE * RS232 CABLE * CONNECTOR KIT
Page 8
Aluminum inserts
- Dry Well
- Liquid lled (option)
Insert, aluminum, of 2“ x 6” or 50 mm x 150 mm diameter and lenght with: 2 holes of 6.5 mm bore, 2 holes of 3,5 mm bore, 2 holes of 4,5 mm bore, 1 hole of 8 mm bore, 1 hole of 10 mm bore and 1 hole of 12 mm bore.
OPTIONAL ITEMS
Liquid-lled Insert, aluminium, of 2” x 6” or 50 mm x 150 mm diameter and lenght with pop-off cap and impeller at bottom. Insert, aluminum, blank of 2” x 6” or 50 mm x 150 mm ‘ diameter and lenght. Oil to use with the Liquid-lled Insert.
CAUTION: Care should be taken to avoid spillover into the temperature
well when using Liquid-lled Inserts. Use only recommended oil/ liquids to ll the insert. Do not use glass wool with the Liquid-lled Insert.
Page 9
TC2000 user’s manual
Figure 3. TC2000 Orientation and location of items.
2.3 Input power connection
The required power is either 110 VAC or 230 VAC, 50/60 Hz. Always connect the power cable supplied with your unit to a grounded or earthed power outlet.
Page 10
3 Getting started
This section contains a description of the TC2000 operator control and connection points to help you get started quickly. The section also provides instructions for pre-calibration cleaning of the well and inserts (essential to the process) and step-by-step guidance for performing a calibration.
3.1 Operation controls and connections
Very little operator interaction is required to begin calibration. The main operator interaction is via unit’s softkeys (or keypad) and LCD display, or via a (local) PC or other remote computer. Communication with the PC is via the unit’s RS232 serial port connector. To begin a calibration, you must:
* Fit the insert into the well. * Place the sensor(s) into the well and use glass wool to insulate
the sensor.
* Power on the unit ON. * Connect the sensor leads in Channel B. * Congure the sensor to be tested. * Set the temperature using the TC2000’s LCD display, softkeys or
numeric keypad.
Page 11
3.1.1 Software startup
When switching on the calibrator the rst
screen that is shown is a startup display while the internal computer loads the operating system Windows CE 2.12.
A Tast manager program will apear after the CE 2.12 is loaded. The task manager gives two choises
1. Calibration ( which is default) or
2. Firmware upgrade.
TC2000 user’s manual
start and will automatically be selected after 10 sec. of the Task manager appearing. Choise 2. is used when a new software or modication upgrade is to be entered into the CE computer. See seperate section for upgrade...
After choise 1. Calibration is selected either by pressing enter when 1. is highlighted or by default the computer will load the TC2000 application software This will take a couple of miniutes and the status wil be shown at the lower part of the screen viewing a countdown proceedure. After loading is complete the main menu screen will be displayed and the calibrator is ready for calibration tasks.
Choise 1. calibration is the normal
Page 12
TC2000 user’s manual
3.1.2 LCD display
The menus and softkey-controls and information you need to dene para-meters during a calibration are displyed in the unit’s LCD (1/4VGA) display.
3.1.3 Well and insert
The well is an aluminum block with a reference sensor placed at its base. The well heats or cools to the temperature you set. The insert has holes bored to accommodate most sensors.The insert is aluminum and is precision-turned to t the well opening. An extractor tool is provided to remove the insert from the well.
IMPORTANT: The calibrator is calibrated with the external reference sensor
being insulated with supplied insulation material. to acheive the same result use insulation round the sensor undergoing test or calibration.
WARNING: Handle carefully when extracting a Liquid-lled Insert because
the liquid in the insert will be hot and may cause burns to any skin surfaces in contact with it.
CAUTION:
Care should be taken to avoid spillover into the temperature well when using Liquid lled Inserts. Use only recommended oil / liquids to ll the insert. Do not use glass wool with Liquid-lled Insert.
Page 13
3.1.4 Numeric keypad
The numeric keypad is used for entering data into the menus. It can also be used instead of the UP or DOWN softkeys when setting temperatures.
3.1.5 Power on/off
The power cable connection and the power ON or OFF control is provided by a swich­ and-plug connector.
3.1.6 RS232 communication
The RS232 serial communication port is used to connect the TC2000 to a PC during re-calibration and for remote operation when calibrating a sensor.
3.1.7 Sensor connection
Different types of sensors can be connected for testing. Channel A is for the external reference sensor. Channel B is for the sensor under test (test sensor).
TC2000 user’s manual
Page 14
TC2000 user’s manual
3.2 Sensor positioning, connection and calibration accuracy
The calibrator well is always the source for heat or cold for the sensor. However, it is possible to connect the sensor such that the calibrator is the reference and the temperature reading is shown at the QVGA display, or the sensor can be connected to an external measurement device for the readnig.
You must place the sensor and the reference sensor as close to each other as possible to ensure an accurant reading, regardless of what reference and connections you use. To make the calculation accurate when the sensors are placed at some distance from each other; you can use a previously determined offset value to dene the correction factor you need to adjust for the difference-in-distance between sensors.
Note:
Note:
External ref.
External ref.
Note:
at same level
at same level
External ref.
at same level
Page 15
TC2000 user’s manual
3.3 Procedure to run calibration
1. To obtain the best temperature transfer, clean the temperature well and insert before each use. Instructions are provided in section 3.7.
2. Select the insert to be used and lower it into the temperature well.
3. Lower the temperature sensor into the insert. Use glass wool if it is necessary to close off the top of the insert to prevent heat or cold loss. (Glass wool is not to be used in liquid inserts).
4. Power ON the TC2000. The opening menu will appear. If necessary, refere to the chapter 4 Operation menus, controls and indicators.
5. Connect the temperature sensor to your indicating instrument.
6. Select 2 Conguration to select the type of sensor to be tested.
7. On the Conguration menu, select 4. Test Sensor.
8. On the Conguration menu, select the CONFIG softkey to obtain wiring information for Channel B.
9. If you are using an external reference sensor, select 2. Reference sensor.
10. Select the CONFIG softkey to obtain wiring information for Channel A (refernse sensor).
11. If you are using sensor-correction values, select 3. Sensor correction and Table.
12. Select the CONFIG softkey to dene the Read value, True value and tag identier in the table.
13. The default temperature setpoint is 25º Celcius (C). If you want the temperature denored in F or K, see the Settings menu.
14. Press the MAIN MENU softkey to exit the Conguration menu and save the new data.
15. Select 1. Calibration mode and use the UP or DOWN softkeys to dene the temperature.
Page 16
TC2000 user’s manual
3.4 Procedure to run a liquid bath calibration
1. Follow 1-15 in chapter 3.3
2. Choose the right liquid for the temperature range: Water: 0 °C to 98ºC Oil : 40 °C to 150ºC
NOTE: Insulation pad has to be used at the top of the insert
during the calibration.
WARNING: Use eye protection if you are using oil or water above
70ºC to avoide eye injury caused by splashing oil or water bubbles.
WARNÌNG: Be sure that you are using oil (liquid) which is approved for the temperature range that you are working at.
3. Select 1. Calibration mode to select the Impeller speed.
4. On the Calibration menu, select OPTIONS and use the numeric keys to set the impeller speed in %.
5. Impeller speed: In order to get the best circulation in the liquid and thereby avoiding gradients, it’s of great importance to select the correct impeller speed. For example circulated water at too high speed will produce air bubbles and turbulence, which could lead to gradients in the liquid. Gradients may also occur if the circulation is too low. The amount of sensors to be tested at the same time in the liquid insert will also have a signicant impact on the circulation. As a general rule start with the following values if you are testing a single sensor:
Water: 15% and up depending on amount of sensors. Viscosity in the range 0 to 98ºC is almost the same. Oil: 50% and up depending on amount of sensors and temperature range since the viscosity changes a lot over the range - 40 to 150ºC.
Page 17
TC2000 user’s manual
6. TIPS! You will get an extreme stability (typically +/- 0.001ºC) and accuracy as a stirred ice bath (0ºC). Just set the impeller speed to 100% and set point to -3ºC and wait for the freezing point to come.
3.4.1 Calibrating a thermostat:
Use the test cables to connect the thermostat to Channel B according to the wiring information provided in the unit’s QVGA display. Use the UP or DOWN softkeys to dene the temperature, then watch for the display indicator to show if the switch is closed or open.
3.4.2 Calibrating a sensor:
Use the numeric keypad or UP or DOWN softkeys to dene the temperature. The displayed Reference sensor temperature (the temperature measured at the center of the temperature well) will increase to the set value. When the Reference sensor temperature has been stable at +/- 0.2ºC for at least 17 minutes (default), a * will appear in the display to indicate that you can take the reading.
If you require a new setting: Change the tempterature to the new value and wait for the * in the display to indicate that you can take the reading.
Calibration mode graphical:
Page 18
Page 19
TC2000 user’s manual
Page 20
Calibration mode numerial:
3.5 Finishing the calibration
7. When nished, adjust temperature to ambient and allow the temperature well to cool to temperature below 70ºC.
8. Pull the temperature sensor(s) out of the insert.
9. Attach the extractor tool to the insert and pull the insert out of the temperature well.
10. Switch power OFF.
11. Allow the unit to attain temperature below 70ºC before stowing accessories.
Page 21
TC2000 user’s manual
3.6 Defaults values
It is possible to change some of the parameter values displayed in the menus.
Default set temperature = 25ºC (not changeable) Default temperatre units = °C Default reference sensor = Internal sensor (located in the TC2000 well) Default stabilization time = 17 minutes Default stabilization T = ±0.2°C
Information is given in chapter’s 4.4 , 4.5 and 4.6 on how to alter some of these parameters.
The following data is dened during manufacturing and cannot be cannot be changed during its operation.
Serial number TC2000 Well Firmw.version TC2000 Well Serial number TC2000 SIM Firmw.version TC2000 SIM
. On completion of a calibration, the TCCal software will add the following data:
Calibration date TC2000 Well Next calibration date TC2000 Well Calibration date TC2000 SIM Next calibration date TC2000 SIM
Page 22
3.7 Pre-calibration cleaning of well and insert
If necessary, clean the temperature well (interior) and insert (interior and exterior) before use. A dirty or oxidized surface can affect the transfer of heat between the well and the insert and may also cause the insert to stick in the well and be difcult to remove.
WARNING: Use eye protection if you are using compressed air to blow
debris out of the temperature well or insert openings to avoid eye injury.
3.7.1 Procedure for cleaning the interior of the temperature well and insert
1. Insert dry, white paper and turn it so as to remove all dust, dirt or smudges from the interior of the temperature well to ensure optimal heat trensfer to the temperature sensor.
2. If necessary, use compressed air to blow debris from the interior of the temperature well and insert. (A small vacuum hose can be used to extract debris).
3. If necessary, use a cloth with mild soap and water to remove exessive contamination. Alcohol may also be used for this purpose.
4. Use a rod with a small piece of cloth attached to it (e.g. gun­ cleaning kit).
5. When using liquid cleaners, ensure that the well, insert and probe is completely dry before placing the insert to the well.
Page 23
TC2000 user’s manual
3.7.2 Procedure for cleaning the exterior of the insert
1. Insert dry, white paper and rub so as to remowe all dust, dirt or smudges from the exterior surface of the insert to ensure optimal heat or cold transfer from the temperature well.
2. Wipe clean with a clean, dry, lint-free cloth.
3.7.3 Instructions for handling Liquid-lled Insert
Only use clean oil, water or other liquids that have been authorized for use in the Liquid-lled Insert.
CAUTION: Care should be taken to avoid spillover into the temperature
well when using Liqued-lled Inserts. Use only recommended oil/liquids to ll the insert. Do not use glass wool when using Liquid-lled Inserts.
WARNING: Handle carefully when extracting a Liqued-lled Insert
because the liquid in the insert will be hot and cause burns to any skin surfaces in contact with it.
Page 24
4 Operational menus, controls and indicators
4.1 Introduction
The TC2000 is highly automated and requires minimal operator interaction during a calibration. Normally, you are only required to set the temperature necessary for the calibration and to congure the sensor to be tested.
4.2 Main Menu
The Main Menu contains sub-menus to main available functions. Menu entries are listed with a unique number. To select Main Menu functions:
Press the UP or DOWN softkey, and then press the ENTER softkey, or Select the function-number on the numeric keypad (Do not press ENTER when using a number from the numeric keypad).
The selected menu line is inverted
The  VGA LCD display will show Information regarding interaction, controls and the available menus.
NOTE:
The TC2000 uses the dened conguration values until they are
changed. Therefor, it is important to ensure that the values are cor­ rectly dened before starting a calibration.
When power is swiched on, default opening screens will show until the calibrator has booted and the main menu screen in visible.
Page 25
TC2000 user’s manual
4.3 Manual Calibration
Select choice 1. Manual calibration, to enter the manual calibration mode, the Conguration menu will then be displayed. This will enable you to congure your reference sensor and the type of sensor to be tested.
See chapter on Conguration for more details on how to congure and select a sensor. Press Continue to enter into the manual calibration mode.
The set point temperature can either be set using the numerical keypad + enter or using the softkeys. The softkeys will give setpoint increments of 10°C
Pressing the MAIN MENU softkey will take you back to the main menu.
Selecting the OPTIONS softkey in the manual calibration mode will allow you to:
* Set the temperature resolution in the graphical display using the ZOOM IN and ZOOM OUT softkeys * Toggle between the graphic and the numeric displays. * Dene the speed of the liquid insert impeller in % of full speed ( 0 to 100 % in steps of 1 ).
Default values are:
* Set temperature = 25 ºC * Swich test = OPEN * Impeller speed = OFF (0%) * Graphical resolution = 2 °C / horizontal line
Page 26
4.4 Conguration
Important input connection information to remember: Channel A is the connector for the external “Reference sensor”. Channel B is the connector for the sensor to be tested “Test sensor”.
4.4.1 Reference sensor selection
The NEXT softkey will allow you to navigate and change each setting by stepping up or down in the menu. Use the SELECT softkey to select the available settings.
Select 1. Internal, to use the calibrators internal reference sensor. In the manual calibration mode window the reference sensor will be labeled with Internal.
Select 1. Pt100, to use an external reference sensor. Then press CONFIG to setup the external reference sensor.
4.4.2 Reference sensor conguration
The external reference sensor can be congured as either a 2, 3 or 4 wire sensor. Use the UP and Down softkeys to select the appropriate conguration.
Please note that in each conguration the sensor will be terminated differently to the 8-pin connector. See advisory text giving pin connections.
Page 27
TC2000 user’s manual
4.4.3 Reference correction
If you have selected an external sensor as the reference sensor, its correction values can be placed into the sensor correction table. Sensor correction will show the value as None (internal sensor or no correction data), and Table (sensor correction data). The default selection is None.
Reference sensors that have calibration data values which have been determined during a calibration, can be placed in a table to allow correction of the reference sensor so that the result will be that the reference sensor works in true temperature.
1. Use NEXT softkey to move to 2. Sensor correction. The selected “cell“ 2. Sensor correction will be inverted.
2. Press SELECT softkey to change your options.
3. Press the CONFIG softkey and use the UP or DOWN and LEFT/RIGHT softkeys to dene the correction information and tag number for the Pt100. Press EXIT when nished. The correction values that are placed in the correction table are normally produced during calibration of the sensor. For example, when a sensor reads 21.0 temperature (Read value), the display shows the correct value as 20.0 (True value). Read value is what the sensor is reading, True value is what it should be reading.
4. Press EXIT to return to the 3. Sensor correction menu.
5. Move to another item in the menu by pressing NEXT, or press MAIN MENU to save your changes.
When the Sensor correction table for the Pt100 is selected, it will
Page 28
always be used whenever the Pt100 is dened as the reference sensor for this conguration , until you change the conguration.
NOTE: It is very important to remember which reference sensor is
being used, and whether or not you need to use the sensor correction table. If you are using the internal temperature reference, return to 2. Reference sensor, select INTERNAL and OK.
Page 29
4.4.4 Test sensor
You can use 3. Test sensor: to select the type of sensor to be tested and to obtain information for connecting it to Channel B. The default selection is Pt100.
To make a selection:
Use the NEXT softkey to move to
3. Test sensor: The selected “cell”
3. Test sensor is inverted. Press SELECT softkey to change your options.
1. Pt100 selection : Select Pt100 using SELECT softkey and press CONFIG softkey to obtain Channel B connection information. In CONFIG, use the UP or DOWN softkeys to select the wiring for obtaining the connection information for the Pt100. Press the OK softkey when nished. Then press CONTINUE softkey in Conguration to start manual calibration.
TC2000 user’s manual
2. 4-20 mA selection : Select 4-20mA using SELECT softkey and press CONFIG softkey to obtain Channel B connection information. In CONFIG, use the UP or DOWN softkeys to dene the miliamphere to temperature information from the 4-20 mA sensor, and the wiring connection information for the 4-20 mA. Press OK when nished. Then press CONTINUE softkey in Conguration to start manual calibration.
Page 30
3. Volt selection : Select V using SELECT softkey and press CONFIG softkey to obtain Channel B connection information connection information. In CONFIG, use the UP or DOWN softkeys to dene the voltage to temperature information from the Volt sensor, and the wiring connection information for the Voltage snesor. Press OK when nished.
4. Thermocouple selection : Select TC using SELECT softkey and press CONFIG softkey to obtain Channel B Thermocouple Conguration In Thermocouple Conguration select the type of thermocouple using the UP or DOWN softkeys.Press OK softkey when nished .
5. Other sensor selection : Select OTHER using SELECT softkey then press CONTINUE softkey to start the maual calibration mode without using any of the inputs on channel B. Other selection is used when the test sensor is connected to external instrumentation.
Page 31
TC2000 user’s manual
4.5 System information
The TC2000 contains the following data that you cannot alter during operation of the unit:
* Serial number TC2000 Well TC2000 SIM * Firmware version TC2000 Well TC2000 SIM * Calibiration date TC2000 Well TC2000 SIM * Temperature range TC2000 Well TC2000 SIM
On completion of a re-calibration of TC2000, the TCCal software will add the following data:
* Next calibration TC2000 Well TC2000 SIM
4.6 Settings
By highlighting either of the sections can be set :
Temperature unit °C °F °K Keypad beep ON OFF Stabilization time 0 to 1hour Stabilization span 0 to 20°C Date and time can be set
Page 32
4.7 Automatic calibration
Select choice 2. Automatic calibration,
to enter the automatic calibration mode, the Calibration Sets menu will then be displayed. This will enable you to set up your calibrator to perform an automatic calibration task.
Each calibration set may consist of up to 20 calibration temperatures. Up to 10 calibration sets may be programmed. Each set is identied with a tag number. Press continue to enter the conguration menu for your calibration set.
Select the conguration you want for your calibration set. See chapter 4.4 on sensor conguration. Note the calibration set tag no. is displayed in the top left corner of the screen. Press continue to enter Calibration Points menu.
TC2000 user’s manual
Up to 20 different calibration point temperatures can be set using the numeric keypad. The temperature range is from
-20°C to + 150°C Press continue to start the automatic calibration mode.
Page 33
When the automatic calibration has started the sensor temperatures will be displayed in a graphical screen. the calibration will continue automatically through every set point temperature that was set in the Calibration Set.
TC2000 user’s manual
one can dene to simulate at stable state in order to force the TC2000 to make a measurement and proceed to the next set point temperature.
Pressing View List in the soft key menu displays the calibration set points that have alredy been performed.
After the last calibration set point temperature has been completed the Calibration nished menu will show. In this menu there are options for dumping the calibration data to a computer or printing a calibration report to a serial printer. See section for serial port setup.
At any stage in the calibration process
Page 34
4.8 Semi Automatic calibration
Select choice 2. Automatic calibration,
to enter the automatic calibration mode, the Calibration Sets menu will then be displayed. This will enable you to set up your calibrator to perform an automatic calibration task.
Each calibration set may consist of up to 20 calibration temperatures. Up to 10 calibration sets may be programmed. Each set is identied with a tag number. Press continue to enter the conguration menu for your calibration set.
Select the conguration you want for your reference sensor. See chapter 4.4 on sensor conguration. For test sensor select Other to enter the Semi Automatic Mode. Press continue to enter Calibration Points menu.
TC2000 user’s manual
Up to 20 different calibration point temperatures can be set using the numeric keypad. The temperature range is from
-20°C to + 150°C
Page 35
Press CONTINUE start the semi-automatic calibration. The temperature of the well will be shown on the LCD screen. At any time you may choose to similate a stable condition of your test sensor by selecting the SIM.STABLE softkey option.
When the temperature and sensor is at a stable state the calibrator will promt you for a temperature entry of your test sensor. Enter this using the numerical keypad on the TC2000. The calibrator will then continue to the next set point temperature.
TC2000 user’s manual
At any time during the semi-automatic calibration the Calibration View menu can be activated to show the progress in the calibration.
After the last set point temperature is stable and a entry has been made the Calibration Finished menu will be displaed. from here you have the choise to dump the calibration data to an other computer via RS232 or send a calibration report to a serial printer.
Page 36
TC2000 user’s manual
4.9 Calibration report
T E M P E R A T U R E
C A L I B R A T I O N R E P O R T
Printing date: 23-Jan-02 14:51 Page 1 of 1
---------------------------------------------------------------------------------------------------------------------------------
Type of instrument : Temp node1 Model : Proto Manufacturer : ScanSense AS Serial number : 01 Tag/Id Number : 4 Temperature range : 0 -150 C Output signal : 0-10 V
Accuracy : 0.5%
---------------------------------------------------------------------------------------------------------------------------------
No Set point Ref value Test value Deviation Date/time 1 28.00 C 28.00 C 29.24 C 1.24 C 9-Apr-02 14:02
2 30.00 C 30.00 C 31.06 C 1.06 C 9-Apr-02 14:19
---------------------------------------------------------------------------------------------------------------------------------
Type of instrument : Temperature Calibrator Model : Tek Know TC2000 Sensor Interface Module : Input Ch.B OTHER Manufacturer : Scan-Sense AS Serial number : (SIM serial no.) Temperature range : (i henhold til input valg) Accuracy : (i henhold til input valg)
Calibration cert.no. :
---------------------------------------------------------------------------------------------------------------------------------
Sensor : TC2000 Internal Reference Manufacturer : Scan-Sense AS Serial number : 1072 Temperature range : +/- 0.18% F.S.
Calibration cert.no. :
---------------------------------------------------------------------------------------------------------------------------------
Air temperature : 22 C Air humidity : 70 %
------- -------------------------------------------------------------------------------------------------------------------------
Date calibrated : 9-Jan-02 14:02 Company : ScanSense AS
INSTRUMENT (SENSOR) CALIBRATED
CALIBRATION POINTS
INTERMEDIATE INSTRUMENT
REFERENCE INSTRUMENT
CALIBRATION ENVIRONMENT
CALIBRATION STATUS
Calibrated by : Certied by :
Page 37
4.10 Software upgrade utillity
In order to be able to upgrade the application a program called TC2000 Upgrade Utillity must be run on a PC and the TC20000 must be connected to the PC through the serial port.
In the start up proceedure of the calibrator select Firmware upgrade when the Task manger appears on the screen. This will set up the calibrator so that it can receive a new application software from the PC.
TC2000 user’s manual
The transfere time and staus will be displayed on the PC TC2000 Upgrade Utillity program. Use a standard TC serial cable for transfere.
After the transfere is complete the calibrator will automatically install the software internal and go to the Main Menu.
Page 38
TC2000 user’s manual
5 Technical tips
We dene “calibration” as being able to compare the ability of the equipment to perform to a known standard. Temperature calibration provides a means of quantifying uncertainties in temperature measurement in order to optimize sensor and/or system accuracy.
This chapter provides general, informative information about dry-block calibration and tips/hints to help you obtain accurate calibrations with the TC2000 dry-block calibrator. Dry-block calibration is recommended for calibrating industrial sensors such as RTD (Resistive Temperature Device), thermocouples, thermistors, bi-metal thermostats or liquid-in-glass thermometers, etc., because the dry­ block provides the reliable heat (or cold) source needed to verify the accuracy of these devicec.
The dry-block calibrator offers a combination of accuracy, portability and price for industrial calibration applications. Most dry-well calibrators will produce a calibration accuracy of better than +/- 0.5ºC. For comparison of sensors in a dry-well’s temperature block, hole-to­ hole uniformity is typically +/- 0.05ºC. These uncertainties are well matched to the typical uncertainties of industrial temperature sensors.
5.1 General information regarding accurate dry-block calibrations
The dry -block in the calibrator is often used as heat/cooling source and a reference thermometer - i.e., the thermometer (or sensor to be tested) is placed in the block and its reading is compared to the reading of the temperature in the dry-block calibrator’s display. The calibrator’s display shows both the dened temperature, as set by the operator, and the actual temperature that is measured by the block’s internal sensor. To ensure an accurat calibration, always observe the following when using this type of calibration:
* Ensure the internal reference sensor and the sensor being tested or calbrated is positioned close togehter in the insert. To do otherwise will mean you can only guess the accuracy of the temperature shown at the display.
* Ensure that display-accuracy includes traceable calibration report with the dry-block calibrator.
* Use accurate reference thermometer during your calibrations.
Page 39
TC2000 user’s manual
5.2 The importance of testing instrument validation
Validation of measuring instrument (as well as material measurement) comprises testing and stamping by responsible calibration authorities (in accordance with calibration directives). This validation testing ensures that the measuring instrument meets the demands made on its characteristic data and its maesuring technology properties, more especially it determines whether the measurement deviation contributions re within the error limit(s). Stamping (or marking) the instrument documents that it has fullled these requirements at the time of testing, and declares that the condition of the measuring instrument is expected to remain within the tolerance range until the date for revalidation (provided it is handled in accordance with the rules of technology). Laws regulate which measuring instruments are subject to mandatory validation and which instruments are not.
5.3 The importance of a traceable standard
A treceable standard is a procedure by which the “measurand”, indicated by a measuring instrument (or a material measurement), is compared with a national standard. In each level of standard, the deviation in measurement has been determined beforehand by calibration to a higher evel standard. This gives a calibration hieratchy as shown in Figure 2 Calibration hierarchy.
5.4 Correct positioning of the probe in the temperature block
The block in the TC2000 Temperature Calibrator can be used as either a source of heat or cold, or as a thermometer reference. In other words, the sensor can be placed into the temperature block and its reading compared to the reading on the TC2000 display or it can be compared with the reading from an external thermometer.
Page 40
How the sensor is placed in the insert is of crucial importance to obtaining accurate calibration measurements. Regardless of what type of reference is used (i. e. internal or external) it is crucial that the sensor and the reference be at an equal distance when placed into the calibrator’s insert. For example, immersing a 2-inch sensor assembly into a 6-inch well could yield an error up to 10ºC. This error is inherent in all dry-well calibrators.
SENSOR TO
BE TESTED
NOTE: INCORRECT WAY DOING CALIBRATION
INTERNAL REF. SENSOR
5.5 Calibrating a short-stem probe
We recommend the use of a comparison technique when calibrating a short-stem probe. Do not compare the test reading to the dry-block display; it will not give you the best results. It is more accurate to use a reference probe of similar size and diameter. Size and diameter greatly affect the amount of heat lost to ambient through the probe stem. Therefore, the closer the sizes match, the more accurate the comparison.
Your objective should be to achieve identical heat properties inside the block to ensure that both sensors are sensing the same temperatures in the same way. Any deviation will cause further error. Therefore, for a satisfactory result, we recommend that you immerse the reference probe and the test probe, both of which should be identical in size, into their bore holes in the insert to exact same depth.
Our new “Liquid-lled Insert” eliminates this problem and can be used to calibrate short probes of any length using the probe’s internal sensor as the reference.
Page 41
TC2000 user’s manual
5.6 Using the calibrator as a reference when adjusting measuring devices (meter/gauges) The temperature calibrator can
also be used as a calibrated reference when adjusting measuring devices such as meters or gaughes. Using a multi-hole insert allows multiple meters/gauges to be adjusted at the same temperature.
NOTE: RECOMMENDED CALIBRATION OF SHORT-STEM PROBES
SENSOR TO BE TESTED
INTERNAL REF. SENSOR
EXTERNAL REF. SENSOR
5.7 The Time Constant inuence on temperature-well mass
The sensor being tested has a large time constant due to larger mass and longer distance to the cooling/heating elements. Consequently, the temperature curve for the internal reference sensor and the sensor being tested (center of insert) will be different (see Figure 3 Temperature reference). All Scan-Sense temperature calibrators have an internal reference sensor mounted in Figure 3. Temperature reference the temperature block to react quickly to temperature changes and thus reduce the resulting time constant.
5.8 When to take the calibration reading
With the TC2000, when the REF temperature, that is the temperature at the center of the well, has been stable +/- 0.2ºC for at east 10 minutes (default value), the display will show the message “Temperature stable, take reading”. Some temperature overshoot will occur when approaching the SET temperature. However, this only applies to the internal reference sensor. The sensor being tested has
Page 42
a larger mass bacause of the insert, so it reacts more slowly to
temperature changes; see Figure 3 Temperature reference.
5.9 The importance of cleanliness to attain high accuracy results
The ability of any dry-block calibrator to provide accurate calibrations and performance depends on factors such as; sensor length, sensor diameter, thermal-conductivity, the sensor position in the insert, etc. cleanliness of the sensor, temperature well and insert is crucial to the resulting accuracy of each calibration, you must ensure that all surfacec are clean and free of dust, dirt, smudges and oxidaization.
5.10 Insulating with glass wool to minimize cold/heat loss to surrounding air
Loss of cold or heat to the surrounding air will occur when calibrating or testing a long sensor thet is partly exposed when it is placed into the calibrator. When the sensor has a large exposed mass, this cold or heat loss can affect the accuracy of the calibration by as much as 0.5ºC to 1ºC. As a general rule, the effect of cold or heat loss is minimal if the length of the part of the sensor that is exposed is less than (<) 1/4 of its length. We recommend that the exposed part of the sensore be insulated. Further, we recommend the use of glass wool for insulation because, although there are different types of insulating materials available, the material of choice must have good insulating characteristics and be able to withstand extreme operating temperatures. It must also be exible enough to be tucked around the sensor, hence the choice of glass wool. Also bear in mind that drafts can affect the stability of the calibrator. Please contact ScanSense for further information.
Page 43
TC2000 user’s manual
6 Specications
TEMPERATURE RANGE
Temperature range -40ºC* to +150ºC * Rel. ambient Accuracy ±0.3ºC Resolution 0.01ºC Gradients at 95ºC w/water 0.06ºC Stability (30 min. after normal stabilisation): * As a dry block ±0.02ºC * As a stirred liquid bath ±0.005ºC * As a stirred ice bath ±0.001ºC Uncertainty (best) <0.03ºC (depending on ext. ref.) Well depth 155mm Well diameter 50mm Bath volume 261cm³ Thermostat test yes Serial interface RS232 Power supply 110V(90-130V), 50/60Hz 230V(180-250V), 50/60HZ Operating temp. 0 to +40°C Dimensions 406x254x356 mm Weight 15.5 kg without inserts
7 Electrical specications
INPUTS PORT A: RANGE RES. ACCURACY (% of reading)
Pt100 2, 3 and 4-wire* -50 to 800ºC 0.01ºC ±0.05ºC ±0.009%
Inputs Port B:
Pt100 2, 3 and 4-wire* -50 to 800ºC 0.01ºC ±0.05ºC ±0.009% Type K -100 to 1370ºC 0.1ºC ±0,25ºC ±0,03% Typr J -200 to 1200ºC 0.1ºC ±0.25ºC ±0.03% Type E -200 to 1000ºC 0.1ºC ±0.25ºC ±0.04% Type T -200 to 400ºC 0.1ºC ±0.25ºC ±0.03% Type S 0 to 1760ºC 0.1ºC ±0.25ºC ±0.03% Type R 0 to 1760ºC 0.1ºC ±0.25ºC ±0.03% Type B +400to 1820ºC 0.1ºC ±0.25ºC ±0.03% Type N -200 to 1300ºC 0.1ºC ±0.25ºC ±0.05% Type L Voltage** 0 to 10V 1µV ±0.25ºC ±0.02% Current** 0 to 22mA 1µA ±2µV ±0.02% Cold junction*** 0 to 40ºC 0.1ºC ±2µA ±0.25%
* Pt100 conforms to IEC751 and temperature scalev of ITS-90 TC input temperature drift 90ppm ** All thermocouple inputs are cold junction compensated V and mA input temperature drift 27ppm *** Will in addition indicate negative inputs but these are not calibrated Published specications may change without notice!
Page 44
Loading...