SO-161 Ophthalmic MicroscopeUser Manual Page 3 of 25
Figure 1Scan Optics SO-161 Ophthalmic Microscope
ISSUE NUMBER:5.0
DATE:11/04/04
SUPERSEDES ISSUE: 4.1
DATE:02/01/02
WRITTEN BY:RJK
DATE:05/11/97
CHECKED BY:RDW
DATE:05/11/97
INSTRUCTIONS AND SPECIFICATIONS
SO-161 Ophthalmic MicroscopeUser Manual Page 4 of 25
Please read the following information carefully before installing and using the Scan Optics Ophthalmic
microscope. Scan Optics is responsible for the safety, reliability and performance of the equipment
only if it is used in accordance with these instructions.
PARTS LIST
MAIN ASSEMBLIES
Clamp assembly (includes power supply, clamp, pillar and safety clamp)
Arm assembly (includes horizontal arm and adjustable pantograph arm)
Microscope assembly (includes microscope head, lamphouse and tilt adjuster)
CABLES
Lamphouse cable
Mains power cable
12V dc supply (battery) cable
OTHER
Microscope eyepieces (located in the tool box) (2)
Focus control covers (4)
Binocular assistant microscopeSO-1420
Auxiliary l ight sourceSO-232
Table plateSO-291
ISSUE NUMBER:5.0
DATE:11/04/04
SUPERSEDES ISSUE: 4.1
DATE:02/01/02
WRITTEN BY:RJK
DATE:05/11/97
CHECKED BY:RDW
DATE:05/11/97
UNPACKING AND INSTALLATION
SO-161 Ophthalmic MicroscopeUser Manual Page 5 of 25
1.Remove the clamp assembly from the carton.
2.Check that the power supply is set to the correct mains voltage (110-120V or 220-240V), by
noting the position of the indicator on the power supply near the mains input plug housing (see
figure 2). Severe damage can occur if the setting is incorrect.
MIf the mains voltage indicator is not correctly set, reposition it using a coin.
Figure 2Setting the mains voltage
3.Fix the clamp to the operating table about 40 cm from the head of the table. The clamp may
be fixed on either side of the table. Make sure that the clamp is pressed firmly against the
side of the table before tightening (see figure 3).
MAlternatively, the clamp may be mounted on any horizontal surface that can
be positioned within 60 cm of the working position, such as a mobile trolley.
MIt is important that the mounting surface be free from vibration and
movement.
ISSUE NUMBER:5.0
DATE:11/04/04
SUPERSEDES ISSUE: 4.1
DATE:02/01/02
WRITTEN BY:RJK
DATE:05/11/97
CHECKED BY:RDW
DATE:05/11/97
SO-161 Ophthalmic MicroscopeUser Manual Page 6 of 25
4.Tighten the safety clamp at a point about midway up the pillar (see figure 3).
Figure 3Attaching the clamp to the mounting surface
5. Remove the arm assembly from the carton, and place it on the pillar. Make sure that the arm
assembly rests against the safety clamp (see figure 4).
ISSUE NUMBER:5.0
DATE:11/04/04
Figure 4Attaching the arm assembly
SUPERSEDES ISSUE: 4.1
DATE:02/01/02
WRITTEN BY:RJK
DATE:05/11/97
CHECKED BY:RDW
DATE:05/11/97
SO-161 Ophthalmic MicroscopeUser Manual Page 7 of 25
6.Remove the microscope assembly from the carton and locate it in the end of the arm
assembly. Tighten the collet knob to secure the microscope (see figure 5).
ISSUE NUMBER:5.0
DATE:11/04/04
Figure 5Attaching the microscope assembly
SUPERSEDES ISSUE: 4.1
DATE:02/01/02
WRITTEN BY:RJK
DATE:05/11/97
CHECKED BY:RDW
DATE:05/11/97
SO-161 Ophthalmic MicroscopeUser Manual Page 8 of 25
7.Attach the female socket on the lamphouse cable to the connector on the side of the
lamphouse and plug into the power supply (see figure 6).
ISSUE NUMBER:5.0
DATE:11/04/04
Figure 6Connecting the l amphouse cable
SUPERSEDES ISSUE: 4.1
DATE:02/01/02
WRITTEN BY:RJK
DATE:05/11/97
CHECKED BY:RDW
DATE:05/11/97