Scala SC 6360 User guide

Digital blood pressure monitor SC 6360

for measurements on the upper arm

Operating instructions

Introduction

Thank you very much for your decision to buy the SC 6360 blood pressure monitor. It is most easily operated, as the settings that were still necessary with older models have all been taken care of by the electronic programming (Fuzzy logic). Just press the relevant button and the measurement can start. But this instrument has much more uses. In order to utilize all its options, please read these instructions carefully before you use the instrument for the first time and keep them in a safe place for future reference.

Special features of the SC 6360 blood pressure monitor

The SC 6360 has a special logic for recognizing the
required pumping pressure (Fuzzy logic). If the pumping pressure set by the manufacturer is insufficient, the manometer will recognize it and increase the pressure by further pumping.
The device has 60 memory locations. The date and time
are displayed and stored together with every blood pressure value. To recall the memory, please press “MEM Button” under power-off mode first. Then, the average measurement result will be shown on the LCD first.
You may set 3 different times for the alarm and
measurements.
The instrument switches automatically back into its clock
function 3 min after the blood pressure has been measured.
3

Contents

Introduction .................................................................... 3

Special features of the SC 6360 blood

pressure monitor .......................................................... 3

Safety instructions ......................................................... 4

How to put the SC 6360 into operation /
Changing the batteries ……………………………........... 4

Device Description ……………………………………….. 5

Description of LCD ......................................................... 5

Date / time setting procedure ......................………........ 6

Setting the alarm clock ................................................... 6

Instructions for measuring the blood pressure ............ 8

The blood pressure measurement ............................... 8

Premature discontinuation of the measurement ........... 8

Memory recall and deletion ............................................ 9

What blood pressure values are normal? ................…... 9

Assessment of blood pressure values .......................... 10
Care and maintenance .................................................. 10
Periodic re-calibration .................................................. 11

Error messages and disturbances ................................ 11

Disposal instructions .................................................... 11

Reference to standards ................................................ 12

Warranty ...................................................................... 12
Specifications ............................................................ 13
A colored and illuminated display will assess the blood
pressure just measured.
The cuff is suitable for upper arms with a circumference
between 22 and 34 cm.
Alternatively to battery operations, the SC 6360 may
also be operated with a special power supply unit, which is also available from Customer Services (under 03329/604314).

Safety instructions

The blood pressure monitor SC 6360 enables you to monitor your blood pressure regularly. However, you should never change the treatment which your doctor has prescribed
Self-measurement means monitoring, it does not
imply diagnosis or treatment.
Discuss unusual values with your doctor.
Do not change the medication prescribed by your
doctor.
The heart rate is not a useful indicator of the frequency of a cardiac pace maker. Patients with cardiac arrhythmia should use the blood pressure monitor only after consulting a doctor.

How to put the SC 6360 into operation / Changing the batteries

Put 4 LR6 batteries in the compartment as indicated on the back of the instrument. Preferably, alkali batteries (LR6) should be used. If the batteries are flat, an empty battery will be displayed at the bottom left: .
4

Specifications

Measuring method: Oscillometric
measuring range: pressure display: 30 … 280 mmHg pulse: 40 … 199 per minute measuring accuracy: static pressure: +/- 3 mmHg pulse: +/- 5 %
Measuring value memory: memory with 60 locations for systolic / diastolic / heart rate, date and time display: LCD display/LED display for the blood pressure assessment power supply: 4 x LR6 (alkali batteries),
working temp. range: +10° C … +40° C storage temp. range: -10° C … +50° C weight: approx. 270 g (without batteries) cuff length: for upper arm circumferences between 22 cm and 34 cm
Subject to alterations in the sense of technical improvements