Scala SC42TM User guide

Rapid clinical thermometer SC 42 TM
with flexible measuring point and fever alarm
We congratulate you on the purchase of this quality product from Scala. Please read the instructions for use before using for the first time.
Special features: The SC 42 TM is a maximum thermometer, i.e. the highest temperature is displayed after only 30 seconds. The most recent measured value is always
saved. If the temperature registers higher than 37,5°C the fever alarm will sound.
Safety information:
The SC 42 TM is suitable only for measuring human body temperatures. ♦ The thermometer contains small parts (batteries, etc) which could be
swallowed by small children. Please keep the thermometer and its batteries out of the reach of small children. If the battery or other parts are swallowed you should seek medical help immediately.
The enclosed batteries may not be recharged or reactivated by other
methods, may not be dismantled, thrown into fires or short-circuited. Batteries must be disposed of properly. Suitable methods for the disposal of batteries include returning them to the place of purchase or disposal in communal collection points.
Protect the SC 42 TM from direct sunshine and temperatures above 50°C. ♦ Never drop the SC 42 TM, do not bend or open it (with the exception of the
battery compartment for battery replacement).
Operating manual:
Press the 0/I button to switch on. A short signal tone shows that the device is in the "switched on" condition. A segment test takes place simultaneously (checks all display segments and carries out an internal function test (see under "Function test")).
- 1 -
After this, if the temperature measured by the temperature sensor is less than 32°C, the display shows an "L“ and "°C“. After around one second, the "°C“ starts to blink. The „L“ stands for low (less than 32°C).
The current temperature is continuously displayed during measurement, and the "°C" blinks during the entire measurement time. The measurement is completed after around 30 seconds.
If the measurement is taken orally or rectally, the approximate final temperature is shown after about 30 seconds and changes only marginally after this time. A measurement period of between five and 10 minutes must be observed for the underarm measurement method. The signal tone can be sounded several times during this period. Since the contact between the temperature sensor and the armpit is not as good as those for oral and rectal measurements, the temperature rise is slower which causes this effect.
After the measurement is completed, the temperature value measured remains in the display until the SC 42 TM is switched off. To turn the device off after use, press the 0/I button briefly. The signal tone will again be sounded. If the device is not switched off, a signal tone is sounded after around 8 minutes and the device then switches off automatically.
Display of last measured value:
If you keep the 0/I button pressed for longer than three seconds, the last measured value will be shown in the display. As soon as you release the I/O button, the device is ready to take measurements.
Function test:
The thermometer is checked for correct functioning automatically every time it is switched on. All segments are displayed, and then L°C. If a failure (measurement inaccuracy) is recognised, "ERR“ (Error) is shown in the display.
Measured value deviation:
The measured temperatures will always deviate, depending on the measurement location. This is normal, and caused by deviations in temperature distribution on the human body. You could, for example, measure a rectal temperature of 36.6°C, an oral temperature of 36.1°C and an underarm value of 35.6°C.
Fever alarm:
If the temperature registers higher than 37,5°C, th e fever alarm (a quick beeping sound which occurs continuously for about 8 seconds) will sound when the peak temperature has registered..
- 2 -
Measurement locations:
In the mouth (oral)
Small temperature variations in the mouth between measurements are normal. These are caused by varying temperature distribution within the mouth. Try to place the thermometer in one of the two heat pockets under the tongue, to the left and right of the tongue root. The measurement sensor must have good contact with the tissue. The subject should close the mouth and breathe normally through the nose so that the measured result is not affected by breathed air. Measuring period: approx. 1 minute
Heat pockets
In the anus (rectal)
This is the safest method. It is particularly suitable for infants and small children. Guide the thermometer point carefully 2 - 3 cm into the anus. Measuring period: approx. 1 minute.
Under the arm (axillary)
From a medical viewpoint this type of measurement leads to inaccurate values. It is therefore unsuitable for carrying out exact measurements. Measuring period: 5 to 10 minutes.
Measuring the basal body temperature
The device can also be comfortably and practically used for the measurement of fertile or unfertile days in the female cycle using the temperature measurement method.
Changing the battery:
As soon as the symbol appears in the display, the battery has run out and requires replacement. Always replace the discharged battery with one of the same type. To do this, proceed as follows:
- 3 -
Pull the cover off the battery compartment. The battery can now be accessed. Remove the used battery with the help of a pointed object (pincers, small screwdriver etc.) and put the new battery ( LR 41 or SR 41) in. Please make sure it is the right way round. You can purchase batteries from jewellers or electrical shops.
Measurement checks:
Checking the accuracy of the SC 42 TM is not necessary in most cases. The German Medical Product User regulations of 29th June 1998 stipulate a measurement check every two years for devices used in medicinal practices. If you require a measurement check (at your expense), please send the SC 42 TM to our customer service address. Please make sure it is well-packaged and that the correct postage stamps are applied.
Cleaning and disinfection
The thermometer is watertight, and can therefore be fully immersed in one of the cleansing fluids listed below for cleaning.
Name of fluid Manufacturer Application Edisonite Merz GmbH & Co. in accordance with Edisonite (Super) Merz GmbH & Co. Gigasept FF Schülke & Mayr GmbH manufacturer's information D 1 Beiersdorf AG Hamburg Ethyl Alkohol Alkohol Handelskontor for plastic surfaces 59555 Lippstadt We recommend the use of Mucocit-F2000 or HelipurHplus to disinfect the device during use in medical facilities. Both are effective against all hospital-typical pathogens and can be purchased from any dispensing chemist.
Guarantee:
We give a guarantee period of 2 years from the date of purchase. We guarantee perfect design of this article. We will rectify all material or manufacturing errors during the guarantee period. If, contrary to expectations, defects are found, please send the device back to our customer service department. Please enclose the proof of purchase and a description of the defect so that the problem can be rectified as soon as possible. The guarantee excludes damages caused by improper handling, related to the battery or to the packaging.
- 4 -
Disposal information
Batteries must be disposed of properly. Battery disposal containers are provided in all shops selling batteries in addition to communal waste collection points. After the end of your device's service life, please dispose of the device correctly using the communal collection point for old electrical equipment.
Specifications:
Type: maximum thermometer Measuring range: 32.0°C to 43.9°C Temperature below 32.0°C: display "L“ for low. Temperature above 43.9°C: display "H“ for high. Measurement accuracy: 0.1°C between 34.0°C and 42.0 °C at a ambient temperature of 18°C to 28°C. Fever alarm: If the temperature registers higher than 37,5°C, the fever alarm will sound Display: 3-digit liquid-crystal (LCD display). Battery type: 1 no.: LR 41 or SR 41. Battery display: The shows when the battery is discharged. Storage temperature: 0°C to 50°C
Manufacturer:
K-Jump Health Co., Ltd. No. 56 Wu Kung 5th Road Wu Ku Industrial Park Taipei, Taiwan, 248 Factory: Dong Guang, Guangdong
SCALA Electronic GmbH Sensorelektronik Stahnsdorf GmbH Ruhlsdorfer Str. 95 Ruhlsdorfer Straße 95 D-14532 Stahnsdorf D-14532 Stahnsdorf Germany Germany
Importer and distributor: Customer Service:
The device has been certified in accordance with
DIRECTIVE 93/42/EU of 14th June 1993
relating to medicinal products.
- 5 -
rapid
clinical thermometer
SC 42 TM
(30 second thermometer)
with flexible measuring point
and fever alarm
Instructions for use