Thank you for purchasing the SBS-2003 Digital Hydrometer.
Please read and follow the below instructions carefully.
Warning!
• When measuring toxic substances, be well aware of its
properties and put on protective gloves, goggles, etc.
• The electrolyte contains sulfuric acid and is toxic. Burns
may occur if contact is made on the skin or clothes.
Blindness may occur if contact is made to the eyes. Always
wear protective glasses and rubber gloves. If the electrolyte
is splashed onto the skin or clothing, rinse with sufficient
amounts of cold water. If it gets into the eyes, flush with
cold, clean water for a minimum of 15 minutes and seek
immediate medical attention.
Caution!
• Care should be taken to avoid damage to the prism. Use soft tissue
paper when wiping the prism surface to avoid scratching. Never use
any metal tools or abrasive materials on the prism.
• The instrument is designed to measure the electrolyte solution of lead
acid batteries. Do not use for the measurement of other electrolyte
solutions such as in alkaline batteries. (i.e., NiCd, LIOH, NMH)
• Remove the battery from the instrument before storing for long periods
of time.
1
Note
• Read this manual carefully to understand the operation and functions of the instrument.
• Make sure that the instrument is calibrated correctly by zero-setting daily before taking
measurements.
• Handle the product with care. Do not drop the instrument or expose it to a strong shock.
• SBS will not be held liable for any damages or injury resulting from any improper or
unintended use of the instrument.
• SBS will not be held liable for any damages or injury resulting from the qualities and
characteristics of samples measured with this instrument.
Note about Temperature
Attention to the following note is needed during measurement when the
temperature of the electrolyte varies greatly from the ambient temperature:
The SBS-2003 compensates for the temperature of the sample automatically
by detecting the temperature on the prism in the Sample Chamber. If the
temperature of the sample is different from the prism temperature, wait
approximately 3 seconds for every 1°C℃ (2°F) of the difference between
temperatures before pressing the START key. (If very large differences in
temperatures are suspected, allow time for the sample to warm the prism
before pressing the START key.)
ERRATIC READINGS OR ERRORS
When the SBS-2003 seems to be malfunctioning or out of order, check the battery
power first. Malfunctions can often be solved by simply replacing the battery.
2
Table of Contents
Unpacking℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃4
Names and Functions of Parts℃℃℃℃℃℃℃℃℃℃℃℃℃4
Inserting Sample Tube℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃6
Attaching Wrist Strap7
Inserting Battery8
Zero-setting10
Measuring Sample12
Temperature14
Settings Menu℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃15
Bluetooth Data Logging17
Alternate Method of Bluetooth Data Transfer26
Error Messages29
Maintenance31
Troubleshooting℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃34
Specifications℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃℃37
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
3
1. Unpacking
Make sure that all of the following items are included.
• ℃SBS-2003 (main unit)℃℃℃℃℃℃℃℃ (1)
• ℃Sample tubes℃℃℃℃℃℃℃℃℃℃℃ (3)
• ℃Sample tube caps (℃℃℃℃℃℃℃℃3)
• ℃9V 006P Alkaline battery (℃℃℃℃℃1)
• ℃Wrist strap ℃℃℃℃℃℃℃℃℃℃℃(1)
• ℃Instruction manual (this booklet)℃℃ (1)
• 3" Software CD SBS-2003 (1)
2. Names and Functions of Parts
(Figure 1 and 2 next page)
1. LCD indicator
Measurement values, prism temperature (°F or °C), memory count and menu item are digitally
displayed.
2. SELECT key
Press to move the triangular pointer to indicate between “S.G.(Specific Gravity)”, “Temp.
(temperature)” and “Count.” This button also scrolls through the menu items when in the
menu.
3. ZERO key
Press to zero-set the instrument.
4. START/OFF key
Press to take a measurement. Hold it down for 2 seconds to turn off the instrument. In the
menu pressing the START performs that menu item.
5. Battery cover
Open the lid to insert/remove the battery.
6. Tube pump
Squeeze the pump to create a vacuum and then release to draw a sample into the sample
chamber through the sample tube.
7. Sample chamber hook
Used to hold the sample chamber in place.
4
8. Sample chamber
Samples come in contact with the prism and are measured.
9. Sample tube
Sample is drawn through the sample tube.
10. Sample tube cap
Used to protect the tip of the sample tube.
Figure 1Figure 2
R
E
T
E
M
O
R
D
Y
H
L
TA
I
G
I
D
u
n
e
M
.
p
m
Te
t
un
o
C
.
S.G
J
=
p
SBS-2003
_
p
N
O
P
Q
R
6
7
8
9
10
5
3. Inserting Sample Tube
The sample tubes come detached from the main unit.
Follow the instructions below to attach them.
(1) Push the sample tube into the
hole located at the bottom of the
sample chamber until it stops.
(Figure 3)
Figure 3
(2) Fit the sample tube into the
groove on the side by pushing it in
with the thumb.(Figure 4)
Test Equipment Depot - 800.517.8431 - 99 Washi
TestEquipmentDepot.com
Figure 4
ngton Street Melrose, MA 02176
6
4. Attaching Wrist Strap
Follow the instructions below to attach the included wrist strap. Use the strap to
prevent dropping the instrument as shown on the front cover of this manual.
Figure 5
(1) Put the strap through the rectangular
opening at the bottom of the instrument.
(Figure 5)
(2) Loop the strap through the strap adjuster
buckle. (Figure 6)
Figure 6
7
5. Inserting Battery
Warning!
• Do not touch the battery with wet fingers to avoid shock.
Caution!
• Insert the battery with the correct polarity (plus + and
minus ℃terminals). Operating with a battery with incorrect
polarity may cause a failure of the instrument.
(1) Turn the screw on the battery cover
lid counterclockwise with screwdriver
until the slot is aligned with the “OPEN”
position marker. (Figure 7)
Figure 7
G
I
D
R
E
T
E
M
O
R
D
Y
H
L
TA
I
u
n
e
M
.
p
m
Te
t
n
u
o
C
.
G
.
S
J
=
SBS-2003
8
(2) Insert the battery with the correct polarity.
(Figure 8)
Note: Have a replacement battery ready
at all times. The remaining life of the
battery included with the instrument may
be shorter than expected.
(3) Close the lid and turn the screw clockwise until the
slot is aligned with the “CLOSE” position marker.
Figure 8
=
R
D
Y
H
L
TA
I
IG
D
.
p
m
Te
S.G.
R
E
T
E
M
O
u
n
e
M
Count
SBS-2003
J
p
_
p
• The SBS-2003 does not have a power switch. Once the battery is
inserted the electric circuit of the SBS-2003 is on. Remove the battery
from the instrument before putting it into long-term storage.
• Any generic 9V battery can be used.
• Always zero-set the instrument after the battery is replaced.
• When the voltage of the battery is low, replace with a new one.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
9
6. Zero-setting
Caution!
• The SBS-2003 needs to be zero-set at the
beginning of each day. Moreover, each time the
ambient temperature changes while in use, it
needs to be zero-set.
• Zero-setting is also needed whenever the
battery is replaced.
• Allow 30 to 60 minutes for the instrument to be
acclimated to the ambient temperature.
1. Remove the sample tube cap.
2.Have distilled or de-ionized water ready.
3.Press the START key to turn on the SBS2003, then press the SELECT key to
move the triangular pointer to “Temp.” or
“S.G.”(Figure 9)
4.Submerge the tip of the sample tube in
water. Squeeze and then release the tube
pump. Water is drawn up into the sample
chamber. (Figure 10)
Figure 9
DIGITAL HYDROMETER
Temp.Menu
S.G.Count
SBS
-200
3
Figure 10
R
E
T
E
M
O
R
D
Y
H
L
TA
I
G
I
D
u
n
e
M
.
p
m
Te
t
n
u
o
C
.
G
.
S
SBS-2003
Note: Confirm that the sample chamber
is completely filled with water. If any air
bubbles are present, pump water again.
10