Do not use the mirror on the microscope to focus the sunlight through your microscope.
Looking directly at the sunlight can result in irreversible eye damage.
Your microscope is a precision optical instrument and should be handled with care at all times.
Do not drop your microscope as such impacts may damage or misalign the optics in your instrument.
Do not power down your microscope within five minutes of switching it on as this may reduce the effective
lifetime of the lamp.
Avoid using your microscope in direct sunlight, high temperature and humidity, dust and vibration.
This product is designed and intended for use only as a biological microscope. Modifying this product in
any way for use in any situation other than the original and intended product design will automatically
void the warranty.
Setting up your microscope
Remove the microscope and accessories from the box and identify all parts before assembling. When
ready to begin, set the microscope on a stable surface. Your objective lenses should be pre-installed.
If your microscope requires manual installation, please take the following steps:
1. Remove the eyepiece from the packaging.
2. Slide the narrow end of the eyepiece into the eyepiece holder.
3. Plug the power cord into a power outlet.
4. Your microscope is powered by 230V AC power. It is important to ensure the voltage coincides with that
of the microscope.
5. If your microscope is battery powered, please insert the required batteries (not supplied) to get started.
Alternatively you can turn it on with just the AC adapter.
Operating the microscope
Getting started
1. Switch on the power and the light(s) on your microscope should be illuminated. If your microscope has
a light intensity adjustment dial, ensure it is turned up to the brightest.
2. Place and align the specimen / object to the centre of the working stage and where possible,
1. Rotate the nosepiece and place the objective lens with the lowest power (4x) in the optical path. Make
sure you hear a click that confirms it is in place.
2. If needed, loosen the locking screw located in between the stage and the arm to regulate the range
of movement of the stage.
3. Ensure the iris diaphragm is opened fully.
4. Push the stage clips to insert a specimen slide.
5. Looking through the eyepiece(s), turn the coarse focuser knob to focus on the specimen until the
specimen comes into focus.
6. If your microscope has a fine focus knob, re-adjust the focus until the image appears sharp and defined.
Changing magnification
Slowly turn the nosepiece clockwise or anticlockwise to the desired magnification until you hear a click
confirming the selected magnification is secured.
1. Ensure your specimen is sitting directly under the objective lens
2. Rotate the coarse and fine focus until the image comes to focus
Using 100X magnification with immersion oil
To obtain maximum resolution with 100X objective lens, applying immersion oil between the slide and
the front lens is necessary.
1. Place one small drop of immersion oil on top of the cover slip, directly over where light comes up
from the condenser.
2. Slowly rotate the 100X objective lens into place. You should see the oil filling the space between the
lens and the slide. If bubbles appear, they can be removed by slightly rotating the revolving nosepiece
back and forth.
3. Adjust the focus with the fine focus knob only.
4. Always wipe and clean all parts that have come into contact with the immersion oil after use. This is
because oil could dry up on the lens surface and damage the microscope. Always wipe with lens paper
or a soft cotton cloth with the appropriate cleaning solution.
Note: The immersion oil should only be used for the 100X objective lens. Do not use the immersion
oil for any other objective lens as this may cause spherical and chromatic aberrations.
The diameter of the diaphragm can be adjusted by turning the adjustment ring of field diaphragm
Condenser
The 1.25 condenser lens can be adjusted to desired position by adjusting two of the condenser adjustment
thumb screws
Filter holder
Face the front of the microscope and swing the filter holder clockwise towards you. The filter holder can
be found just below the iris diaphragm. Place the coloured filter onto the holder and swing it back to place.
You may now observe the specimen through the eyepiece.
Illumination
Should there be a need to change the light bulb, unplug the electric cord and ensure the microscope
has been switched off for at least 15 minutes.
1. Slowly position the microscope on its side, with the underside of the base facing you
2. Remove the four screw holding the grill
3. Carefully remove the bulb in the microscope
4. Use a clean cloth to hold and insert a new bulb into the socket
5. Attach the grill and screw to tighten
Caution
Do not touch the new bulb with your fingers as it could affect both the transmission of light and the
condition of the bulb.
When handling a halogen bulb, it is recommended to wear gloves or cover the light bulb with protective
sleeve. Oils and other materials from fingers may damage the bulb. If the bulb has been touched,
clean with alcohol and cloth.
Eyepiece WF10x/18mm
Objective DIN 4x, 10x, 40x(r), (Optional) 100x(r)
Condenser N.A 0.65 Pre-centered with built-in illuminator
Head Rotation 360 Degree
Lamp Halogen
Focus Coarse and Fine
Stage Type Plain stage 120mm x 110mm
Nosepiece Triple nosepiece with click stop
Warranty 3 Years Limited Warranty
Power AC Adapter
Weight 2.9kg
What is include in the box
1x saxon Biological Microscope
1x Dust cover
1x AC Adapter
1x set rubber eye guard
1x immersion oil
You can also obtain the e-Manual from our website at www.saxon.com.au
To ensure your microscope performs at its best, regular maintenance of your microscope is recommended.
Cover the microscope with the dust cover whenever it is not in use. This will prevent dust from settling on the
mirror or lens surface.
When moving your microscope, carry it with two hands – one holding the arm and the other supporting the
base. Never pick it up by the focusing knobs.
Do not dismantle or modify any parts of the microscope as either of this will void the warranty. (Unless noted
on instruction manual)
Do not store microscope in direct sunlight or under direct indoor light.
Use only appropriate cleaning tools when cleaning your microscope. We recommend using a combination of
the following:
• A soft brush made from camel hair
• Optical cleaning solution and a soft lint-free cloth
• Special lens paper
Clean the lens surface using an appropriate optical cleaning solution and always remember to dry the lens
after. Do not wipe the lenses when they are dry as they can easily be scratched.
For the condenser and illuminator lens, only clean the top lens surface.