sauter FH-S, FH Instruction Manual

Page 1
Adaptersockel
Re Set Taste
Set
LCD
Grenzwert
Anzeige
RS 232
Kraftaufnehmer
Se
Sauter GmbH
Ziegelei 1
Tel: +49-[0]7433- 9933-199 D-72336 Balingen
Fax: +49-[0]7433-9933-149 E-Mail: info@sauter.eu
Internet: www.kern-sohn.com
Instruction Manual
FH
FH-S Digitálny silomer(SK)
Thank you for buying a SAUTER force gauge. We hope you are pleased with your high quality force gauge with its big functional range. If you have any queries, wishes or helpful suggestions, do not hesitate to call our service number.
„Sensor Inside“ means the measuring cell is inside the housing.
1.
Included in delivery
- SAUTER FH incl. rechargeable battery
- Carrying Case
- Charger
- Standard Attachments as shown below, 5 pcs M3 x 8 screws
Power plug
Reset button
Display
Pilot lamp
RS-232 port
Force Sensor
Sensor
Inside
Important annotation:
By pressing the RESET key (on the right side of housing, see illustration on the left), individual settings and memorised values can be re-set or erased, in example for a new start of the instrument after an operating error.
Data in mm
2.
Working Conditions
10°C to 30°C / 15% up to 80% humidity
3.
Electrical Power Supply
Either by rechargeable battery or current power supply
Current power supply:
- Connection by power adapter
- Rechargeable batteries are charged simultaneously
Rechargeable battery pack for mobile applications:
- Type: Ni 8.4V / 600 mAh
- Charging time: approx. 1 hour
FH_SI-BA-e-1317 1
Page 2
Sauter GmbH
Ziegelei 1
Tel: +49-[0]7433- 9933-199 D-72336 Balingen
Fax: +49-[0]7433-9933-149 E-Mail: info@sauter.eu
Internet: www.kern-sohn.com
Instruction Manual
FH
SET:
4.
Technical Data
- 1 x Press: Upper Limit [HidT / Hi Lt]. To
- Accuracy: ± 0,5 % of Capacity (measuring range) change press: ▲ or (see section
- Data Sampling Rate: 2.000 Hz High/Low Limit function)
- Weight: 640 g
Hi Lt
- 1 x Press: Lower Limit [LodT]. To change
5. Operation
press ▲ or ▼ (see section High/Low Limit function)
Display
- 1 x Press: Sets the Auto-STOP force value that stops the movement of a test stand.
(1) Measuring Result
Adjustment of the force value. To change
(2) Measuring Units
press: ▲ or ▼ (only active in „Battery-
(3) Activation of PRINT Function
Mode“)
(4) Indication of power charging status (5) PEAK or AUTO-PEAK Mode (6) Average value of stored peak values (7) Force direction
1 x Press: Auto-OFF Function. Turns the
(8) Occupancy of storing spaces
instrument off after a here defined time
(9) AVERAGE- or Saving Mode
period in sec. [P.OFF]. To change press: ▲ or ▼ (Only active in „Battery-Mode“)
Operation keys
ON / OFF:
- 1 x Press: Peak-Freeze-Time [PE.2E /
ON / OFF key
A.PE / HoldT]: Time period (in sec.) in
(For ON, press 1 sec.)
which a peak value is being shown in the display in sec. To change press: ▲ or ▼
UNIT:
- 1 x Press: Data Transfer [rS232] to PC
- Press shortly: Select unit:
P.OFF
HoldT
StoP
(PC) or to the printer (Print) or (in version U
N, kg or lb
5.1) to the test stand
- Press for 2 sec.: Display return
rS232
ZERO:
Lo dT
Lo Pe
- 1 x Press: Minimum limit to save peak values in the instrument’s memory. [LE.SET / Lo Pe]. Please see section „Min
Limit Peak Save“. To change press: ▲ or ▼ (Only active in „Peak-Mode“) (see
section min. limit peak save)
- PC: Pressing SET once: Saves the Settings
Three functions:
- PRINT: Pressing SET twice: sending data to the printer
- Zeros the measuring result (Tara function)
- Cleans the peak value (in Peak mode)
- STAND: Pressing SET three times: sending a signal to
- Saves a setting (in SET mode) the Test Stand to stop its movement (in version U 5.1)
BACK LIGHT:
FH_SI-BA-e-1317 2
Page 3
Sauter GmbH
Ziegelei 1 D-72336 Balingen E-Mail: info@sauter.eu
Tel: +49-[0]7433- 9933-199 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Instruction Manual
FH
PEAK:
Three functions available:
- Track mode (continuous measurement)
- Peak mode (capture of maximum values)
- Auto-Peak mode, same as Peak-mode, only
without the „Min limit peak save“ function
MEMORY:
Saves the peak values to calculate the average value (please see section “Memory mode”
DELETE Function
Deletes stored peak values (only in „Memory modeactive)
PRINT:
Sends the stored peak values to a PC or Printer (please see section „Data Output”)
High / Low limit function
LEDs to display OK / NOT OK Tests
Lower than lower limit
Lights if the STOP value is reached Higher than highest limit
This function allows efficient testing of OK / NOT OK’S measurements of similar or identical testing objects. A lower and an upper limit value can be defined. The instrument compares the individual measuring results with these limit values and shows the OK or NOT OK result by green or red light diode and by sound. To set these limit values, please see the SET Menu in
section „Operating keys“.
Measurement (Track Modus)
Display (1) shows the continuous force in a defined direction (6)
To zero the display, press:
Peak-Hold Function (Peak Mode)
Please press:
Auto-Peak-Hold-Function (Auto-Peak Mode)
Please press:
Min. limit peak function to activate storage of measurements
This function allows to eliminate unwanted „Pre-Peak values“ that are lower than the main peak value (Fp). The “Min limit peak save” value (Fo) takes care, that these “Pre-Peak values” are not saved. The „Min limit peak save“ function is only possible in Peak-
Mode. To set this Min limit value, please see the SET Menu in
section „Operating keys“.
Memory mode and average value (from up to 10 peak values)
Saving peak values in the instrument
Activating the „AUTO PEAK Function“ by PEAK key
Deactivating the „Average Function“ by MEMORY key
Now, all peak values are stored automatically in the
instrument
To browse through the stored values, please use the ▲
or ▼ keys. (The values will be shown in the upper
display segment)
By pressing the MEMORY key, the average value of the
stored peak values can be shown (in the upper display segment)
To delete every stored value, press the▼-key in the
AVERAGE-Mode
FH_SI-BA-e-1317 3
Page 4
Sauter GmbH
Ziegelei 1
Tel: +49-[0]7433- 9933-199 D-72336 Balingen
Fax: +49-[0]7433-9933-149 E-Mail: info@sauter.eu
Internet: www.kern-sohn.com
Instruction Manual
FH
6. Configuration of RS 232
SUB-D 9pm
Pin
Signal
Illustration
2
TxD
Output signal
3
RxD
Input signal
5
GND
Ground
6
+1.6 to + 2 V
Over upper limit
7
+1.6 to + 2 V
Lower than lower limit
8
+1.6 to + 2 V
OK
6.1 Output Protocol
RS-232 Parameter Baud rate: 9600 Data-Bit: 8 Parity: none Stop-Bit: 1
The measured value is requested by the PC by the ASCII
Sign “9”.
The measured value that comes from the instrument has this format:
e.g. 0011.70 means -11,70 Newton, if Newton is the selected unit
||-----| | |____> the other 6 places describe the
measured value as ASCII-Signs
|_______> the first place describes the
direction of the force (0 = minus = Pressure;
1 = plus = Tension)
or: 1021.15 means +21,15 N (Tension)
|_______> the first place describes the
direction of the force (0 = minus = Pressure;
1 = plus = Tension)
or: 1021.15 means +21,15 N (Tension)
7. Warning
Intended use
The instrument you have acquired serves to determine the measuring value of the material to be measured. It is intended to be used as a “non-automatic“ instrument, i.e. the material to be measured is manually and carefully attached at the instrument. The measuring value can be
read off after a stable measurement value has been obtained.
Inappropriate use
Do not use the instrument for medical measurements. In the event that small quantities are removed or added to the material to be measured, incorrect measuring results can
be displayed due to the “stability compensation“ in the
instrument. (Example: Slow draining off of liquid from a container suspended from the instrument). Do not attach a continuous load. This can damage the measuring unit as well as the parts, relevant to safety.
Prevent jolts, torsion and oscillation (e.g. by appending slopingly) of all kinds. Be sure to prevent overloading the instrument in excess of the stated maximum load (max.), minus any tare weight that may possibly exist. This could damage the instrument (risk of breakage).
Important:
Always make sure that there are no people or materials below the load that could be injured or damaged!
The instrument is not suitable for measuring people. Do not use as baby scales!
The instrument does not comply with the medical product law (MPG). Never operate the instrument in hazardous locations. The series design is not explosion-proof. Structural alterations may not be made to the instrument. This can lead to incorrect measuring results, faults concerning safety regulations as well as to destruction of the instrument. The instrument may only be used in compliance with the described guidelines. Varying areas of application/ planned use must be approved by SAUTER in writing.
Guarantee
The guarantee is not valid in following cases:
Non-observation of our guidelines in the operating instructions
Use outside the described applications
Alteration to or opening of the device
Mechanical damage and damage caused by media,
liquids
Natural wear and tear
Inappropriate erection or electric installation
Overloading of the measuring equipment
FH_SI-BA-e-1317 4
Page 5
Sauter GmbH
Ziegelei 1 D-72336 Balingen E-Mail: info@sauter.eu
Tel: +49-[0]7433- 9933-199 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Instruction Manual
FH
Monitoring the test substances
8. CE Declaration of Conformity
The metrology features of the instrument and any possible available adjusting weight must be checked at regular intervals within the scope of quality assurance. For this purpose, the answerable user must define a suitable interval as well as the nature and scope of this check. Information is available on the home page (www.KERN-sohn.com) with regard to the monitoring of instrument test substances and the test weights required for this. Test weights and instruments can be adjusted
quickly and at a reasonable price in KERN’s accredited
DKD calibration laboratory (return to national normal).
Fundamental safety information
Do not use the hanging instrument to transport loads. Prevent jolts, torsion and oscillation (e.g. by appending slopingly) of all kinds. Never use the hanging instrument over the maximum permitted weight (!! Danger of breaking!!). Always make sure that there are no living beings or materials below the load that could be injured or damaged. The hanging electronic instruments from the SAUTER instrument are only suitable for hand-held use or use in a test stand. They are not suitable for hanging from a mechanical hook, e.g. a crane hook. Observe the information in the operating instructions Please read the operating instructions carefully before erecting and commissioning, even if you already have experience with SAUTER instruments.
Staff training
The device may only be operated and looked after by trained members of staff.
FH_SI-BA-e-1317 5
Page 6
Sauter GmbH
Ziegelei 1
Tel: +49-[0]7433- 9933-199 D-72336 Balingen
Fax: +49-[0]7433-9933-149 E-Mail: info@sauter.eu
Internet: www.kern-sohn.com
Instruction Manual
FH
FH Adjustment Procedure
1
Start the
Press ON/OFF instrument
2
Enter the
Directly after pressing adjustment
ON/OFF, press PEAK procedure
and PRINT together for more times and very short intervals until the red light is on.
3
Instrument
Press SET Type
Directly after pressing PEAK + PRINT
a
(Back to the
(If you are in the normal
operation mode again, mode ???)
the instrument has to be shut off. Start again with
1. Eventually pressing the buttons faster)
4
Select the
Press ▼▲ instrument type
5
Save
Press SET instrument type
6
Enter the
Press MEMORY adjustment sequence
7
Choose the
Press UNIT and adjustment
with▼▲ specify the weight
calibration weight in Newton (X kg * 9,81)
8
Save the
Press SET and UNIT adjustment
simultaneously weight
9
Attach the
Press ZERO weight to the instrument (by hanging)
finished
The green light is on
The lower red light is on XXX N is shown
The correspondi ng value to the instr. appears in the display
The upper red light is on
The weight is written in the display
The instrument changes to operation mode. Adjustment procedure is
FH_SI-BA-e-1317 6
Loading...