6    FH-M-BA-e-1718 
SET: 
-  1 x Press: Upper Limit value [HidT / Hi Lt]. To change, press: ▲ or ▼ 
(see section High/Low Limit function) 
-   
-  1 x Press: Lower Limit value [LodT]. To change press ▲ or ▼ 
(see section High/Low Limit function) 
-  1 x Press: Minimum limit value to activate memory function at limit value 
measurements (PEAK). [LE.SET / Lo Pe] (also see section „Min Peak Limit 
value function)“. 
-  To change, press: ▲ or ▼ (Only active in „Peak-Mode“, see section limit 
value indication) 
-  1 x Press: Sets the Auto-STOP force value that stops the movement of the 
test stand when reaching a limit force. Adjustment of the force limit. 
-  To change, press: ▲ or ▼ 
   -  1 x Press: Auto-OFF Function. Turns the instrument off after a here defined  
        period of time in sec. [P.OFF]. To change, press: ▲ or ▼  
        (Only active in „battery-Mode“)     
-  1 x Press: Peak-Freeze-Time [PE.2E / A.PE / HoldT]: period of time (in 
sec.) in which a peak value is being shown in the display (PEAK). 
-  To change press: ▲ or ▼ 
-  1 x Press: Optional data transfer [rS232] to PC (PC) or to the printer (Print) 
or (in version U 5.3) to the test stand (stand) 
-  PC: Pressing SET 1x: Saves the Settings 
-  PRINT: Pressing SET 2x: sending data to the printer 
      -  STAND: Pressing SET 3x: sending a signal to the test stand 
         to stop the movement (in version U 5.3)