Sartorius YDI01C-WP Installation Instructions Manual

Installation Instructions
Sartorius YDI01C-WP (Option A6|A7)
Installing a Digital Interface in Place of an A/D Converter in Combics Indicators, Models CIS1..|CIS2..|CIS3.. | CISL1..|CISL2..|CISL3..
98647-003-78
The YDI01C-WP digital interface built into the Combics indicator enables connection of any of the following:
– Digital weighing platform
(e.g. IS weighing platform) using the YCC02-R12F6 connecting cable (Option M41: 12-contact round female connector; length: 6 m; open cable ends), or
– Digital weighing instrument (e.g., LA, LP,
ME, EA, EB, CP balance/scale) using the YCC02-D25F6 connecting cable (Option M42: 25-pin D-Sub male connector; length: 6 m; open cable ends), or
– Digital weighing instrument (e.g. QA,
QC, FB, FC scale) using the YCC02-R12M6 connecting cable (Option M43: 12-pin round male connector; length: 6 m; open cable ends).
Note: To have devices communicate with the Combics indicator as plat­forms and scales through the SBI protocol, configure the operating menu as follows: Baud rate: 9600 Handshake mode: Hardware, 1 character after CTS Parity: Odd Number of data bits: 7 Stop bits: 1 Printing: Automatic without stability Identification: With ID code (22 characters)
To have devices communicate with the Combics indicator as platforms and scales over the xBPI protocol, configure the operating menu as follows: Setting: xBPI protocol Interface: RS-232 or RS-485
! Refer to the relevant operating instructions for descriptions of the operating menus. The menu setting in the Combics indicator depends on the particular platform to be connected: Setting: SBI or xBPI protocol Interface: RS-232 or RS-485
Example: For Combics 1 and 2, this setting is configured in the Setup menu; for Combics 3, in the menu under Device parameters WP1
RS-232 or RS-485
SBI: Standard o IS-485 SBI: Trade ADC-485
o IS-232
ADC-232
! Always make sure the equip-
ment is disconnected from power before performing any installation, maintenance or repair work.
! Work that affects IP protec-
tion must be carried out with extreme care.
! Any installation work that
does not conform to the instructions in this manual will void all claims under the manufacturer’s warranty.
! Do not operate the YDI01C-
WP in hazardous areas.
§ Place the Combics indicator on a soft surface and remove the four cap nuts from the front panel.
§ Remove the front panel
Installing the YDI01C-WP PCB
§ Remove the A/D converter PCB
– Unplug the male connector – Remove the 3 screws – Remove the A/D converter
PCB
§ Plug in the YDI01C-WP PCB
– Use the 2 screws to fasten
the PCB
– Plug in the male connector
Printed in Germany on paper that has been bleached without any use of chlorine · W1A000 · KT
Publication No.: WYD6122-e06032
§ Install the power supply cable and connect it as described below (see arrows)
1. To the YDI01C-WP PCB
2. To the power supply PCB
3. To the DC/DC converter PCB
Configuration
Install connectors on the open cable ends and set the jumpers according to the desired operating mode
LV1
Pin Signal
1 Power supply: +15 to 25 V (Line) 2 TXD (RS 232) 3 RXD (RS-232)
4
Supply ground (GND)
LV2
Pin Signal 1 GND 2 RXD-TXD-N (RS-485) 3 RXD-TXD-P (RS-485) 4 GND
Device selection:
Jumper 1
Device Position CIS.1 1.1 CIS.2 1.1 CIS.3 1.1
Interface selection:
Jumper 2
Operating mode Position RS-485 2.1 RS-232 2.2
Operating mode selection in RS-485 mode:
Jumper 3
RS-485 mode Position W/o terminating resistor 3.1 With terminating resistor 3.2
§ Place the front panel on the indicator
§ Place the Combics indicator on a soft
surface and attach the four cap nuts for the front panel.
1
2
7
3
6
4
5
Overview of the YDI01C-WP PCB
1) Configuration jumper: 1
2) Connecting cable for the DC/DC converter
3) Male connector for connecting the power supply PCB (cable enclosed)
4) LV1
5) LV2
6) Configuration jumper: 3
7) Configuration jumper: 2
1
2
3
Loading...