Contents

Contents ....................................3
Intended Use................................. 3
Possible Application............................ 4
Warning and Safety Instructions .................. 5
Installation .................................. 6
Air Buoyancy Correction ........................ 9
Transporting the Climate Module................. 11

Intended Use

The climate module is used to record climate data such as
temperature, air pressure, and humidity. When used in a Sartorius balance near the sample, it can determine the air density and thus perform an air buoyancy correction. This in turn considerably increases the measurement accuracy of the balance. The monitoring of climate limit values is also possible both inside and outside of the weighing chamber.
The climate module is located within the weighing
chamber in balances with a corresponding draft shield. The climate module can also be attached to an external tower outside of the draft shield to record climate data.
3

Possible Application

– YCM20MC:
Climate module, uncalibrated, for all Cubis MCM and MSAxx-PC models
– YCM20MC-DAkkS:
Climate module with DAkkS calibration certificate for all Cubis MCM and MSAxx-PC models
– YCM20MC-Tower:
Tower for climate module, for mounting YCM20MC (-DAkkS), can be connected to the following models MCM10K3 (-DAkkS), MCM40K3 (-DAkkS), MCM60K3 (-DAkkS), MCM60K2 (-DAkkS) and MSAxxx-PC.
4

Warning and Safety Instructions

Installation
Only connect or disconnect the climate
module when the device is unplugged from the power supply.
t Connect the D-SUB plug securely to the
device.
Safety
The climate module should only be
used indoors, and make sure that foreign bodies and liquids cannot penetrate it.
ESD protection measures must be taken
Do not touch conductive parts for
reasons of ESD protection. Therefore, when inserting the climate module, only touch the housing or the side edges! Do not touch the contact surfaces.
Therefore: Take preventative measures.
As a minimum safeguard, you should touch a grounded piece of metal to discharge any static electricity.
5

Installation

Removing the Sticker
Remove the sticker
t
over the recess in the weighing chamber.
Inserting the Climate Module into the Weighing Chamber
t Insert the climate
module into the recess provided on the back of the weighing chamber.
6
Comparator model MCM6.7: Inserting the Climate Module
t Fold up the climate
module holder.
t Insert the climate
module into the recess provided on the climate module holder and then carefully lower the climate module in the holder into the draft shield.
7
YCM20MC-Tower Installation
t Insert the climate
module into the recess provided on the housing of the tower.
t Next, connect the tower
at the rear of the weigh cell: Screw the supplied 5-pin cable into the tower and the weigh cell.