Sartorius Vivaflow 50, Vivaflow 50R, Vivaflow 200 Directions For Use Manual

Page 1
85030-519-25
Directions for Use
Vivaflow 50 | 50R | 200
Page 2
Contents
Vivaflow 50 3
Single Module 3 Linked Modules 4 Diafiltration|Desalting 5 Concentration 6 Recovery 6 System Components 7 Technical Specifications 7 Performance Characteristics 8 Ordering Information 9 Chemical Compatibility 10
Vivaflow 50R and Vivaflow 200 11 Operation of Single Module 11 Operation of Two Modules 12 – 13 Diafiltration|Desalting 14 Concentration 14 Recovery 14 Technical Specifications 15 Cleaning 16 Storage 16 Performance Characteristics 17 – 18 Ordering Information 19 – 20 Chemical Compatibility 21
Page 3
Vivaflow 50
Single module
1. Set up the system as illustrated in Diagram 1. Note the positioning of the flow restrictor on the return line.
2. Vivaflow membranes contain trace amounts of glycerine and sodium azide. To remove these chemicals and to check the security of the tube connections, it is recommended to rinse the module, and to test the system at full pressure before introducing the sample.
3. Place 500 ml deionised water in a suitable reservoir. The Vivaflow sample|diafil­tration reservoir (prod. no. VFA006) is recommended.
4. Pump liquid through the system to purge any air pockets. The recirculation rate should be in the range 200–400 ml/min, and suitable flow should exit the filtrate line. If used, the pressure indicator (prod. no. VFA020), should read approximately
2.5 bar.
5. Allow 400 ml to pass into the filtrate vessel. Check for any leakage at tubing connection points. Drain the system and empty or replace the filtrate vessel, (see recovery section). The system is now ready for use.
Filtrate vessel
Sample|diafiltration reservoir (VFA006)
Pink flow restrictor
Designed for single use only. Anything else constitutes improper use.
Diagram 1: Vivaflow 50 set up
Page 4
Linked modules
1. Set up the system as illustrated in Diagram 2. Link the required number of modules by sliding the tongue and groove edges together. Note the positioning of the flow restrictor on the return line.
2. Use series interconnectors, (prod. no. VFA031), to connect a single row of modules. Detailed ordering information for system components to be found of page 7, Diagram 4.
3. In addition, use T-connectors, (prod. no. VFA030), to connect two rows of modules in parallel.
4. The flow path of the system can be con­figured to suit the membrane|sample combination. Most solutions are better suited by modules connected in parallel. Very low viscosity solutions favour mod­ules connected only in series. Rinse the system as detailed under single module operation.
T-connector
Interconnector
Pressure indicator
Diagram 2: Setting up multiple Vivaflow 50 cassettes
Page 5
Diafiltration|Desalting
1. Set up the system as illustrated in Diagram 3 for linked modules. If only one module is required, connect it in place of the linked modules.
2. Rinse the system as detailed previously.
3. Either place the solution to be purified in the 500 ml sample|diafiltration reservoir (prod. no. VFA006), or concentrate a larger volume to 500 ml or less so that it is fully contained in the 500 ml reservoir. Ensure the reservoir lid is firmly closed.
4. Fill a large feed reservoir with exchange solvent and pump sample through the system as for concentration.
5. As the volume in the 500 ml reservoir decreases, the vacuum created draws exchange solvent through the feed line from the larger reservoir. Over 99% solvent exchange can be accomplished with an exchange volume approximately 5 times the volume of the sample.
Diagram 3: Buffer exchange | diafiltration with Vivaflow 50
Page 6
Concentration
1. Fill the feed reservoir with sample solu­tion. When initial volumes larger than 500 ml are required, place 500 ml in the sample|diafiltration reservoir, the remaining volume in another suit­able container and connect the vessels as detailed in the diafiltration section. Alternatively, use a larger container for the entire sample volume and immerse the feed and return lines directly into the liquid.
2. Pump liquid through the system. The recirculation rate should be in the range 200–400 ml/min, and suitable flow should exit the filtrate line. If used, the pressure indicator, (prod. no. VFA020), should read approximately 2.5 bar.
3. Concentrate the sample to the desired volume.
!
Warning: Do not run the same section of
tubing through the pump head for longer than six hours, over use of tubing will result in significant pressure drop and ultimately, failure.
4. When the desired volume has been reached, reduce the recirculation rate to 20–40 ml/min and recirculate the concentrated sample for 1–2 minutes to maximise recovery.
Recovery
1. Disconnect the feed line from the lid of the 500 ml reservoir or when using a different container, remove the feed line from the sample.
2. Pump residual system volume back into the reservoir|container. (When parallel modules are used with viscous solutions, ensure that all modules are empty by pinching the tubing between each of the stacks of modules in turn).
3. For a more complete sample recovery rinse approximately 5–10 ml per module of water or sample buffer through the system, and recover as before.
Page 7
Table 1: Technical Specifications
Dimensions
Overall L|H|W 107|84|25 mm Channel W|H 15 mm|0.3 mm Active membrane area 50 cm
2
Hold up volume (module) 1.5 ml Min recirculation volume < 10 ml Non recoverable hold-up < 0.5 ml
Operating Conditions
Pump flow 200–400 ml/min Maximum pressure 3 bar (45 psi) Maximum temperature 60˚C
Inconnectors (Prod. No. VFA031)
Pink flow restrictor on the return line
T-Connectors for running 2 stacks (Prod. No. VFA030)
Pressure Indicator (Prod. No. VFA020)
Masterflex easy load pump head (Prod. No. VFA012)
Stand (Prod. No. VFA016)
Masterflex pump CE 230V – US 115V (Prod. No. VFP001|VFP002)
System Components
Diagram 4: System componets for setting up multiple Vivaflow 50 units
Page 8
Materials of Construction
Main housing Polycarbonate Flow channel TPX (PMP) Membrane support TPX (PMP) Seals and O rings Silicone Pressure indicator Polypropylene, SS Spring, Flow restrictor Polypropylene Fittings Nylon Tubing PVC (medical grade)
Table 2: Performance characteristics
Time to concentrate up to 20x (min.) at 3 bar
inlet pressure, 20°C
Single Three Solute Recovery % Device Devices
250 ml Start 1 L Start Direct 10 ml rinse Vol. min. Vol. min.
BSA 1.0 mg/ml (66,000 MW)
5,000 MWCO PES 34 49 96% > 99% 10,000 MWCO PES 22 32 94% > 99% 30,000 MWCO PES 22 32 92% 99% 50,000 MWCO PES 20 29 92% 98% γ Globulins 1.0 mg/ml 100,000 MWCO PES 43 62 92% 98% 100,000 MWCO RC 40 58 92% 98% 100,000 MWCO Hydrosart 40 58 92% 98%
Yeast 1.0 mg/ml (S.Cerevisiae)
0.2 µm PES 33 47 92% 98%
Page 9
Ordering Information
Vivaflow 50 include filtrate tube, size 16 peristaltic tubing, Pack Size Prod. No. flow restrictor and fittings and 1 + Series Interconnector
3,000 MWCO (PES) 2 VF05P9 5,000 MWCO (PES) 2 VF05P1 10,000 MWCO (PES) 2 VF05P0 30,000 MWCO (PES) 2 VF05P2 50,000 MWCO (PES) 2 VF05P3 100,000 MWCO (PES) 2 VF05P4
0.2 µm (PES) 2 VF05P7 100,000 MWCO (RC) 2 VF05C4
Vivaflow 50 complete system comprises
Pump (240 V), easy load pump head (size 16), tubing, 1 VFS502 500 ml sample|diafiltration reservoir, module stand, pressure indicator, T connectors, series interconnectors
Pump (115 V), easy load pump head (size 16), tubing, 1 VFS504 500 ml sample|diafiltration reservoir, module stand, pressure indicator, T connectors, series interconnectors
Vivaflow 50 PVC tubing and fittings
Size 16 PVC pump tubing (3 metres, 3.2+ 1.6 mm) VFA004 Flow restrictor set (2+0.4, 0.6, 0.8 mm) VFA009 T connectors for running 2 stacks (2 pieces) VFA030 Series interconnectors (6 pieces) VFA031 Female luer fittings (10 pieces) VFA032 VF50 tubing kit (2+1 m size 16 PVC tubing with inlets fittings, VFA034
2+50 cm size 16 PVC tubing with 0.6 mm flow restrictors, 1+ series inlet connections)
Flow restrictor 0.6 mm (6 pieces) VFA035
Vivaflow 50 accessories
Masterflex economy drive variable speed peristaltic pump (230 V) VFP001 Masterflex economy drive variable speed peristaltic pump (115 V) VFP002 500 ml sample and|or diafiltration reservoir VFA006 Masterflex easy load pump head – size 15 VFA013 Masterflex easy load pump head – size 16 VFA012 Vivaflow 50 stand VFA016 Pressure indicator (1–3 bar) VFA020
Page 10
10 |
Table 3: Chemical Compatibility for the Vivaflow 50
Solution PES RC
Acetic acid (25%) OK OK Acetone NO NO Ammonium hydroxide (5%) OK OK Ammonium sulfate sat. OK OK DMEM OK OK Ethanol (70%) OK OK Ethyl acetate NO NO Formaldehyde (30%) OK OK Formic acid (5%) OK OK Guanidine HCl (6 M) OK OK Hydrocarbons, Aromatic NO ? Hydrocarbons, Chlorinated NO NO Hydrochloric acid (1 M) OK NO Isopropanol NO NO Lactic acid (5%) OK OK Mercaptoethanol (10 mM) OK OK Methanol (60%) OK OK n-Butanol (70%) OK OK
OK = Acceptable ? = Questionable NO = Not recommended
Solution PES RC
Peracetic acid (0.2%) OK OK Phenol (1%) OK ? Phosphate Buffer (1 M) OK OK Pyridine NO NO RPMI-1640 OK OK Sodium azide OK OK Sodium deoxycholate (5%) OK OK Sodium hydroxide (1 M) NO NO Sodium hydroxide (0.1 M) NO NO Sodium hypochlorite (0.02%) ? ? Sodium nitrate NO NO Sulfamic acid (5%) OK NO Surfactants (0.1%) OK OK Toluene NO NO Trichloroacetic acid (10%) NO ? Trifluoroacetic acid (10%) OK NO Urea (8 M) OK OK
Page 11
Vivaflow 50R and Vivaflow 200
Operation Single Vivaflow 50R or Vivaflow 200 Modules
Assembly and rinsing single module
1. Set up the system as illustrated above in Diagram 5.. Note the positioning of the flow restrictor on the return line. A set of flow restrictors of different gauges (Prod­uct No. VFA009), is available for use with solutions of unusually high or low viscosity.
!
Warning: Ensure Luer connections are
secure before operation.
2. Vivaflow membranes contain trace amounts of glycerine and sodium azide. To remove these chemicals and to check the security of the tube connections, it is recommended to rinse the module, and to test the system at full pressure before introducing the sample.
3. Place 500 ml deionised water in a suitable reservoir.
4. Pump liquid through the system to purge any air pockets. The recirculation rate should be 200–400 ml/min, and suit­able flow should exit the filtrate line. If used, the pressure indicator should read approximately 2.5 bar.
5. Allow 400 ml to pass into the filtrate vessel. Check for any leakage at tubing connection points.
6. Drain the system and empty or replace the filtrate vessel (see Recovery section). The system is now ready for use.
Diagram 5: Vivaflow 50R and Vivaflow 200 set up
| 11
Page 12
12 |
Operating two Vivaflow 50R or two Vivaflow 200 Modules
Assembly and rinsing two modules
1. Set up the Vivaflow 50R system as illus­trated in Diagram 6 and the Vivaflow 200 System as illustrated in Diagram 7 system as illustrated above.
!
Warning: Ensure Luer connections are
secure before operation.
2. Vivaflow 50R: Use two T-connectors (VFA030) to connect to the inlets of both Vivaflow 50R modules.
Vivaflow 200:
Use the Y-connector, (VFA005) to connect to the inlets of both Vivaflow 200 modules. The size 15 tubing connected to the Y-connector is the inlet tubing for the pump. Use a size 15 pump head.
Diagram 6: Setting up two Vivaflow 50R units
Page 13
| 13
3. Note the positioning of a flow restrictor on the return line from each Vivaflow 200 module. Vivaflow membranes contain trace amounts of glycerine and sodium azide. To remove these chemicals and to check the security of the tube connec­tions, it is recommended to rinse the module, and to test the system at full pressure before introducing the sample.
4. Place 1 litre of deionised water in a suitable reservoir.
5. Pump liquid through the system to purge any air pockets. The recirculation rate should be 500–900 ml/min, and suit­able flow should exit the filtrate line. If used, the pressure indicator should read approximately 2.5 bar.
6. Allow 800 ml to pass into the filtrate vessel. Check for any leakage at tubing connection points.
7. Drain the system and empty or replace the filtrate vessel (see Recovery section). The system is now ready for use.
Diagram 7: Setting up two Vivaflow 200 units
Page 14
Diafiltration|Desalting
1. Set up the system as illustrated in Diagram 8.
2. Rinse the system as detailed previously.
3. Place the solution to be purified in a suitable diafiltration reservoir. For vol­umes up to 500 ml the Vivaflow 500 ml sample|diafiltration reservoir (VFA006), is recommended. Ensure the lid is firmly closed.
4. Fill the large feed reservoir with exchange solvent and pump sample through the system as for concentration.
5. As liquid passes through the membrane, the vacuum created in the diafiltra­tion reservoir draws in exchange solvent through the feed line from the large res­ervoir. Over 99% solvent exchange can be accomplished with an exchange volume of approximately 5 times the sample volume.
Concentration
1. Fill the feed reservoir with sample solution.
2. Pump liquid through the system. The recircu lation rate should be 200– 400 ml/min, (500–900 ml/min for two modules), and suitable flow should exit the filtrate line. If used, the pressure indi­cator should read approximately
2.5 bar.
!
Warning: Do not run the same section
of tubing through the pump-head for longer than 6 hours, overuse of tubing will result in significant pressure drop and ultimately, failure.
3. Concentrate the sample.
4. When nearing the desired volume, reduce the recirculation rate to 20–40 ml/min and recirculate the concentrated sam­ple for 1–2 minutes to increase sample recovery.
Recovery
1. Remove the feed line from the sample.
2. Pump residual system volume back into the reservoir|container.
3. For a more complete recovery, rinse 25–50 ml of water or sample buffer through the system, and recover as before.
14 |
Diagramm 8: Buffer exchange | Desalting with Vivaflow 50R and Vivafow 200
Page 15
| 15
Table 4: Technical specifications Vivaflow 50R and Vivaflow 200
Dimensions Vivaflow 50 R Vivaflow 200
Overall L|H|W 100 | 100 | 24 mm 126|138|38 mm Channel W|H 7.5 | 0.4 mm 10 mm|0.4 mm Active membrane area 50 cm
2
200 cm
2
Hold up volume (module) 1.7 ml 5.3 ml Min. recirculation volume 10 ml < 20 ml Non recoverable hold-up < 0.5 ml < 1 ml
Materials of construction Vivaflow 50R and Vivaflow 200
Main housing Acrylic Flow channel Acrylic Membrane support Polypropylene Seals and O rings Silicone Pressure indicator Polypropylene, SS spring Flow restrictor Polypropylene Fittings Nylon Tubing PVC (medical grade)
Operating conditions Vivaflow 50R and Vivaflow 200
Pump flow 200–400 ml/min Maximum pressure 4 bar (60 psi)
Pressure drop across inlet | outlet 0.5 bar (7 psi)
Maximum temperature 60˚C
Page 16
Cleaning and storage information only for Vivaflow 50R and Vivaflow 200
Cleaning
Vivaflow 50R and 200 modules may be used several times if cleaning and storage instructions are followed.
1. Flush the system with 200 ml of deion­ised water with the filtrate going to waste.
2. Place the feed, return and filtrate lines in a suitable container for the cleaning solution.
3. Prepare cleaning solutions suitable for the membrane.
I. Polyethersulfone membranes:
250 ml of 0.5 mM NaOCl in 0.5 M NaOH
II. Hydrosart membranes: 250 ml of
0.5 M NaOH
4. Recirculate at 50–100 ml/min for 30–40 minutes.
5. Drain the system and recirculate 250 ml of deionised water through the system for 5–10 minutes.
6. Drain and discard rinse solution and flush with a further 500 ml of deionised water with the filtrate going to waste. The system is now ready for further use.
Storage
To store Vivaflow 50R and Vivaflow 200 after cleaning, fill module with deionised water and 10% ethanol. Seal inlet, outlet and filtrate ports and refrigerate at approximately 4°C.
16 |
Page 17
| 17
Table 5: Performance characteristics Vivaflow 50R
Time to concentrate up to 20x (min)
at 3.0 bar inlet | 2.5 bar outlet pressure, 20°C start volume Average flux Recovery % 250 ml ml/min Direct 25 ml rinse Lysozyme 0.25 mg/ml (14,000 MW)
5,000 MWCO Hydrosart 70 3.4 96% 98% 10,000 MWCO Hydrosart 23 10.3 94% 96%
BSA 1.0 mg/ml (66,000 MW)
10,000 MWCO Hydrosart 24 9.9 98% > 99% 30,000 MWCO Hydrosart 15 15.8 97% > 99% γ Globulins 1.0 mg/ml (150,000 MW) 100,000 MWCO Hydrosart 46 5.2 97% > 99%
Start volume 1 litre (one Vivaflow 50R at 3 bar) 10,000 MWCO Hydrosart
BSA 1.0 mg/ml 95 10.0 98% > 99%
Start volume 1 litre (two Vivaflow 50R in parallel at 3 bar) 10,000 MWCO Hydrosart
BSA 1.0 mg/ml 48 19.8 98% > 99%
Page 18
18 |
Table 6: Performance characteristics Vivaflow 200
Time to concentrate up to 20x (min.) at 3 bar
inlet pressure, 20°C start volume Average flux Recovery % 1 litre ml/min Direct 25 ml rinse Lysozyme 0.25 mg/ml (14,000 MW)
2,000 MWCO Hydrosart 160 6 97% > 99% 3,000 MWCO PES 180 5 97% > 99%
BSA 1.0 mg/ml (66,000 MW)
5,000 MWCO PES 29 33 98% > 99% 5,000 MWCO Hydrosart 70 14 98% > 99% 10,000 MWCO PES 23 41 96% > 99% 10,000 MWCO Hydrosart 35 27 98% > 99% 30,000 MWCO PES 25 38 96% 99% 30,000 MWCO Hydrosart 20 48 96% > 99% 50,000 MWCO PES 22 43 96% 98% γ Globulins 1.0 mg/ml (average 160,000 MW) 100,000 MWCO PES 54 18 96% 99%
Yeast 1.0 mg/ml (S. Cerevisiae)
0.2 µm PES 11 86 92% 98%
Dilute solute concentration, start volume 1 litre at 3 bar, 10,000 MWCO PES
BSA 0.001 mg/ml 18 52 90% 98% BSA 0.01 mg/ml 20 47 92% 98% BSA 0.1 mg/ml 21 45 94% 99%
Start volume 5 litres (two Vivaflow 200 in parallel at 3 bar) 10,000 MWCO PES
BSA 1.0 mg/ml (66,000 MW) 67 70 97% > 99%
Page 19
| 19
Ordering Information
Vivaflow 50R modules include pressure indicator, flow restrictor Qty per Prod. no. and size 16 pvc peristaltic tubing and fittings box
5,000 MWCO Hydrosart 1 VF05H1 10,000 MWCO Hydrosart 1 VF05H0 30,000 MWCO Hydrosart 1 VF05H2 100,000 MWCO Hydrosart 1 VF05H4
Vivaflow 200 modules include pressure indicator, flow restrictor Qty per Prod. no. and size 16 pvc peristaltic tubing and fittings box
3,000 MWCO PES 1 VF20P9 5,000 MWCO PES 1 VF20P1 10,000 MWCO PES 1 VF20P0 30,000 MWCO PES 1 VF20P2 50,000 MWCO PES 1 VF20P3 100,000 MWCO PES 1 VF20P4
0.2 µm PES 1 VF20P7 2,000 MWCO Hydrosart 1 VF20H9 10,000 MWCO Hydrosart 1 VF20H0 30,000 MWCO Hydrosart 1 VF20H2 100,000 MWCO Hydrosart 1 VF20H4
Vivaflow 50R | 200 complete system comprises
Pump (240 V), Easy load pump head (size 16), tubing, 1 VFS202 500 ml sample|diafiltration reservoir
Pump (115 V), Easy load pump head (size 16), tubing, 1 VFS204 500 ml sample|diafiltration reservoir
Page 20
20 |
Vivaflow 50R | 200 accessories
Masterflex economy drive variable speed peristaltic pump (230V) VFP001 Masterflex economy drive variable speed peristaltic pump (115V) VFP002 500 ml sample and|or diafiltration reservoir VFA006 Masterflex easy load pump head – size 16 VFA012 Masterflex easy load pump head – size 15 VFA013
(only for operating two Vivaflow 200) Size 15 pvc pump tubing and Luer fittings (3 m, 4.8 + 2.6 mm) VFA003 Size 16 pvc pump tubing and Luer fittings (3 m, 3.2 + 1.6 mm) VFA004 Flow restrictor set (2 + 0.4, 0.6, 0.8 mm) VFA009 Female luer fittings size 16 (10 pieces) VFA032 Flow restrictors 0.6 mm (6 pieces) VFA035 Female luer fittings size 15 (10 pieces) VFA036 Flow restrictor set (2 + 0.4, 0.6, 0.8 mm) VFA009 Female luer fittings size 16 (10 pieces) VFA032 Flow restrictors 0.6 mm (6 pieces) VFA035 Female luer fittings size 15 (10 pieces) VFA036
Vivaflow 200 tubing and fittings for operation two modules
Y connector (size 15 to 2 + size 16, Luer fittings) VFA005 Masterflex easy load pump head – size 15 VFA013
Vivaflow 50R tubing a fittings for operation of two modules
T-Connector VFA030
Page 21
| 21
Solution PES HY
Peracetic acid (0.2%) OK NO Phenol (1%) OK NO Phosphate Buffer (1 M) OK OK Pyridine NO NO RPMI-1640 OK OK Sodium azide OK OK Sodium deoxycholate (5%) OK ? Sodium hydroxide (1 M) OK OK Sodium hydroxide (0.1 M) OK OK Sodium hypochlorite (0.02%) OK NO Sodium nitrate NO OK Sulfamic acid (5%) OK ? Surfactants (0.1%) OK OK Toluene NO NO Trichloroacetic acid (10%) NO OK Trifluoroacetic acid (10%) OK OK Urea (8 M) OK OK
Table 7: Chemical Compatibility for Vivaflow 50R and Vivaflow 200
Solution PES HY
Acetic acid (25%) OK OK Acetone NO NO Ammonium hydroxide (5%) OK OK Ammonium sulfate sat. OK ? DMEM OK OK Ethanol (70%) OK OK Ethyl acetate NO NO Formaldehyde (30%) OK OK Formic acid (5%) OK OK Guanidine HCl (6 M) OK OK Hydrocarbons, Aromatic NO NO Hydrocarbons, Chlorinated NO NO Hydrochloric acid (1 M) OK OK Isopropanol NO NO Lactic acid (5%) OK OK Mercaptoethanol (10 mM) OK OK Methanol (60%) OK OK n-Butanol (70%) OK OK
OK = Acceptable ? = Questionable NO = Not recommended
Page 22
22 |
Page 23
| 23
Page 24
Sartorius Stedim Lab Ltd Sperry Way Stonehouse Park Gloucestershire GL10 3UT, UK
www.sartorius.com
Copyright by Sartorius Lab Instruments GmbH & Co. KG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius Lab Instruments GmbH & Co. KG. The status of the information, specifications and illustrations in this manual is indicated by the date given below. Sartorius Lab Instruments GmbH & Co. KG reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
Status: June 2016, Sartorius Lab Instruments GmbH & Co. KG, Goettingen, Germany
Printed in the EU on paper bleached without chlorine.
Publication No.: SLU6097-e160609 Ver. 06 |2016
Loading...