Sartorius Vivacon 500 Operating Instructions Manual

85034-538-74
Technical data and operating instructions
Vivacon® 500
2 |
Vivacon® 500 µl – Introduction Operation
Storage conditions | shelf life
Vivacon® ultrafiltration spin columns should be stored at room temperature. The devices should be used before the expiry date printed on the box.
Introduction
Vivacon
®
500 concentrators are disposable ultrafiltration devices optimally suited for DNA and protein concentration. For optimal performance with DNA and protein samples, they are equipped with the patented regen­erated cellulose membrane Hydrosart
®
.
Vivacon
®
500 can be used in a benchtop fixed angle rotor, accepting 1.5 | 2.2 ml centrifuge tubes.
Equipment required for Vivacon
®
500
Centrifuge
Rotor type Fixed angle Minimum 40°
rotor angle Rotor cavity To fit 1.5 | 2.2 ml/11
mm conical bottom tubes
1. When working with DNA samples, select
a molecular weight cut off (MWCO) which retains the fragment size of double stranded DNA (ds DNA) as shown in Table 5. When working with proteins, select a MWCO at least 50% smaller than the molecular size of the protein of interest.
2. Fill concentrator with up to maximum volumes shown in Table 1. (Ensure lid is fully sealed).
3. Insert assembled concentrator into centrifuge.
4. Centrifuge at speeds recommended in Table 2, taking care not to exceed the maximum g force indicated by the MWCO.
5. Once the desired concentration is achieved, (see Table 3 or 4 for guide to concentration times), remove assembly and recover sample by reverse spinning the concentrate into a fresh collection tube. In this procedure remove filtrate tube, invert the concentrator body into new filtrate tube and then spin at up to 2,500 g for 2 minutes (or pulse for 20–30 seconds). The filtrate can be sealed for storage by closing the filtrate tube cap.
Note:
It is not possible to do a backflush with the device.
Table 1: Technical Specifications
Concentrator capacity
Fixed angle rotor 0.5 ml
Dimensions
Total length 45 mm (Concentration)
Total length (Back-spin) 47.5 mm Width 12.4 mm Active membrane area 0.32 cm
2
Hold-up volume < 5 µl Dead stop volume 5 µl (40° rotor)
Materials of construction
Body Polycarbonate Filtrate vessel Polypropylene Membrane
2kDa - 100 kDa Hydrosart
®
125 kDa Cellulose Acetate (CA)
O-ring Medical grade silicone
Table 2: Recommended Spin Speed in Fixed Angle Rotor (x g)
Membrane cut off For DNA For proteins
2 kDa MWCO 7,500 14,000 10 kDa MWCO 7,500 14,000 30 kDa MWCO 5,000 14,000 50 kDa MWCO 5,000 14,000 100 kDa MWCO 3,000 8,000 125 kDa MWCO 2,500* 5,000
Technical Specifications
Desalting | Buffer Exchange
1. Concentrate sample to desired level.
2. Empty filtrate container.
3. Refill concentrator with an appropriate solvent.
4. Concentrate the sample again and repeat the process until the concentration of contaminating microsolute is sufficiently reduced. Typically 3X wash cycles will remove 99% of initial salt concentration.
Vivacon
®
500 Reverse Spinning
| 3
Hydrosart membrane
Reverse spinning concentrate into fresh collection tube
Collection tube fits standard
1.5 | 2.2 ml tube carriers
Filtrate can be sealed for storage
* Spin speed 2,500 + g for DNA samples > 900 bp when using a 125 kDa MWCO.
For DNA samples >650 bp, spin at 1,000 + g.
4 |
Usage Tips
1. Flow Rate
Filtration rate is affected by several parame­ters, including MWCO, porosity, sample con­centration, viscosity, centrifugal force and temperature. Expect significantly longer spin times for starting solutions with over 5% sol­ids. When operating at 4°C, flow rates are approximately 1.5 times slower than at 25°C. Viscous solutions such as 50% glycerine will take up to 5 times longer to concentrate than samples in a predominantly buffer solution.
2. Pre-rinsing
Membranes fitted to Vivacon
®
concentrators contain trace amounts of glycerine. Should these interfere with analysis, they can be removed by rinsing fill volume of buffer solution or deionised water through the con­centrator. Decant filtrate and concentrate before processing sample solution. If you do not want to use the pre-rinsed device imme­diately, store it in the refrigerator with buffer or water covering the membrane surface. Please do not allow the membrane to dry out.
3. Sterilisation of Vivacon
®
Devices
Vivacon
®
devices should not be autoclaved as high temperatures will substantially increase membrane MWCO. To sterilise, use a 70% ethanol solution or sterilising gas mixture.
4. Chemical Compatibility
Vivacon
®
concentrators are designed for use with biological fluids and aqueous solutions.
For chemical compatibility details, please refer to Table 6.
5. Retention and Recovery
The membranes used in Vivacon
®
are characterized by a molecular weight cut off (MWCO). For proteins, it corresponds to their ability to retain 90% of a molecule with this nominal molecular weight. For achieving better recovery, use a MWCO which is 1 to 2 of the species weight you need to concentrate.
For nucleic acid applications, strand length is the most useful parameter for selecting the Vivacon
®
device appropriate for a specific application. However, other parameters including DNA concentration, the magnitude of the driving force (g-force) and the salt concentration all act in concert to affect DNA recovery. For characteristic recoveries and concentration times, see Table 3 and 4. For the correlation between MWCO and nucleotide cut-off (bp), see Table 5.
Performance Characteristics
Table 3: Performance Characteristics Vivacon® 500 for DNA
Start volume 0.5 ml, sample concentration 50 ng/ml Sample size (bp)
Time to concentrate up Concentrate g-force (xg) to 30x [min.] at 20°C recovery %
2,000 MWCO 10 60 min 93% 7,500 10,000 MWCO 30 25 min 94% 7,500 30,000 MWCO 50 18 min 88% 5,000 50,000 MWCO 300 18 min 91% 5,000 100,000 MWCO 600 10 min 87% 3,000 125,000 MWCO 650 12 min 85% 2,000 125,000 MWCO 900 9 min 94% 3,000
Table 4: Performance Characteristics Vivacon
®
500 for Proteins
Start volume 0.5 ml, sample and concentration of proteins as specified in table Sample
Time to concentrate up Concentrate g-force (xg) to 30x [min.] at 20°C recovery %
2,000 MWCO 0.25 mg/ml 30 min 95% 14,000 cytochrome c
10,000 MWCO 0.25 mg/ml 15 min 92% 14,000 cytochrome c
30,000 MWCO 1.0 mg/ml BSA 10 min 95% 14,000 50,000 MWCO 1.0 mg/ml BSA 10 min 92% 14,000 100,000 MWCO 1.0 mg/ml 11 min 90% 8,000
bovine IgG 125,000 MWCO 1,0 mg/ml 10 min 81% 8,000
bovine IgG
Table 5: Conversion Table for Hydrosart
®
MWCO to Nucleotide Cut-off
Membrane MWCO Double-Stranded Nucleotide Cut-off (bp)
Hydrosart
®
2 kDa > 10
Hydrosart
®
10 kDa > 30
Hydrosart
®
30 kDa > 50
Hydrosart
®
50 kDa > 300
Hydrosart
®
100 kDa > 600
Cellulose Acetate 125 kDa >650
| 5
6 |
Table 6: Chemical Compatibility (2hr contact time) Hydrosart
®
Cellulose Acetate
Compatible pH range pH 1-9 pH 4-8 Acetic Acid (25.0%) OK NO Acetone (10.0%) NO NO Acetonitrile (10.0%) NO NO Ammonium Hydroxide (5.0%) OK OK Benzene (100%) NO NO Chloroform (1%) OK OK Dimethyl Formamide (10.0%) NO NO Dimethyl Sulfoxide (5.0%) NO NO EDTA (1.0 M) Yes Yes Ethanol (70.0%) OK OK Ethyl Acetate (100%) NO NO Formaldehyde (30%) OK OK Formic Acid (5.0%) OK ? Glycerine (70%) OK OK Guanidine HCI (6 M) OK ? Hydrocarbons, aromatic NO NO Hydrocarbons, chlorinated NO NO Hydrochloric Acid (1 M) OK NO Isopropanol (70%) OK OK Lactic Acid (5.0%) OK NO Mercaptoethanol (1.0 M) OK NO Methanol (60%) OK OK Nitric Acid (10.0%) NO NO Phenol (1%) OK OK Phosphate Buffer (1.0 M) OK OK Sodium Chloride (5 M) Yes Yes Sodium Dodecylsulfate (0.1 M) OK OK Sodium Hydroxide (1.0 M) NO NO Sodium Hypochlorite (200 ppm) NO NO Sodium Nitrate (1.0%) OK ? Tetrahydrofuran (5.0%) NO NO Toluene (1.0%) NO NO Trifluoroacetic Acid (10%) OK NO TRIS Buffer, pH 7.2 - 9 (1.0 M) Yes Yes Tween 20 (0.1%) OK OK Triton X-100 (0.1%) OK OK Urea (8 M) OK ?
OK = Acceptable ? = Questionable NO = Not recommended
FAQ
– DNA recovery is lower than expected If the DNA sample contains a high salt concentration, dilute the sample. Run the device at the recommended g-force.
– Can proteins be concentrated with Vivacon
®
?
Proteins can be concentrated with Vivacon
®
, using the guidelines on page 5 to choose the
correct MWCO. However, we recommend Vivaspin
®
500 for protein concentration due to faster
concentration achieved with a vertical membrane design for protein applications.
Sample runs to dryness. Spinning for much longer than the recommended spin times can allow samples to go to dryness. To recover the sample, add 10 µl of water or buffer to the device, vortex gently for up to1 min and then recovery as normal.
– DNA recovery too low. Spinning the sample at 1000 + g may result in higher DNA recoveries, when working close to the membrane cut off limits, e.g. with a 650 bp DNA sample with a 125 kDa membrane.
No DNA signal visible after PCR reaction. Using membrane cut offs smaller than 125 kDa can in some cases lead to concentration of PCR inhibitors along with the sample DNA. Use the 125 kDa membrane cut off and spin your samples at 2000 + g for optimal DNA recovery and sequencing results. Additionally, 1-2 washes with buffer may be needed to remove the inhibitors.
Ordering Information Vivacon
®
500 Qty. per box Prod. No.
2,000 MWCO 25 VN01H91 2,000 MWCO 100 VN01H92 10,000 MWCO 25 VN01H01 10,000 MWCO 100 VN01H02 30,000 MWCO 25 VN01H21 30,000 MWCO 100 VN01H22 50,000 MWCO 25 VN01H31 50,000 MWCO 100 VN01H32 100,000 MWCO 25 VN01H41 100,000 MWCO 100 VN01H42 125,000 MWCO 25 VN01H81 125,000 MWCO 100 VN01H82 125,000 MWCO 500 VN01H83
Vivacon
®
500 Qty. per box Prod. No.
Sample Kit L (4 units each of 2, 10, 30 K) 12 VN01HL12 Sample Kit H (4 units each of 30, 50, 100 K) 12 VN01HH12
| 7
Sartorius Stedim Lab Ltd Sperry Way Stonehouse Park Gloucestershire GL10 3UT, UK
www.sartorius.com
Copyright by Sartorius Lab Instruments GmbH & Co. KG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius Lab Instruments GmbH & Co. KG. The status of the information, specifications and illustrations in this manual is indicated by the date given below. Sartorius Lab Instruments GmbH & Co. KG reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
Status: June 2016, Sartorius Lab Instruments GmbH & Co. KG, Goettingen, Germany
Printed in the EU on paper bleached without chlorine. Vivacon · W Publication No.: SLL6002-e160607 Ver. 06 | 2016
Loading...