Sartorius Vivacon 500 Operating Instructions Manual

85034-538-74
Technical data and operating instructions
Vivacon® 500
2 |
Vivacon® 500 µl – Introduction Operation
Storage conditions | shelf life
Vivacon® ultrafiltration spin columns should be stored at room temperature. The devices should be used before the expiry date printed on the box.
Introduction
Vivacon
®
500 concentrators are disposable ultrafiltration devices optimally suited for DNA and protein concentration. For optimal performance with DNA and protein samples, they are equipped with the patented regen­erated cellulose membrane Hydrosart
®
.
Vivacon
®
500 can be used in a benchtop fixed angle rotor, accepting 1.5 | 2.2 ml centrifuge tubes.
Equipment required for Vivacon
®
500
Centrifuge
Rotor type Fixed angle Minimum 40°
rotor angle Rotor cavity To fit 1.5 | 2.2 ml/11
mm conical bottom tubes
1. When working with DNA samples, select
a molecular weight cut off (MWCO) which retains the fragment size of double stranded DNA (ds DNA) as shown in Table 5. When working with proteins, select a MWCO at least 50% smaller than the molecular size of the protein of interest.
2. Fill concentrator with up to maximum volumes shown in Table 1. (Ensure lid is fully sealed).
3. Insert assembled concentrator into centrifuge.
4. Centrifuge at speeds recommended in Table 2, taking care not to exceed the maximum g force indicated by the MWCO.
5. Once the desired concentration is achieved, (see Table 3 or 4 for guide to concentration times), remove assembly and recover sample by reverse spinning the concentrate into a fresh collection tube. In this procedure remove filtrate tube, invert the concentrator body into new filtrate tube and then spin at up to 2,500 g for 2 minutes (or pulse for 20–30 seconds). The filtrate can be sealed for storage by closing the filtrate tube cap.
Note:
It is not possible to do a backflush with the device.
Table 1: Technical Specifications
Concentrator capacity
Fixed angle rotor 0.5 ml
Dimensions
Total length 45 mm (Concentration)
Total length (Back-spin) 47.5 mm Width 12.4 mm Active membrane area 0.32 cm
2
Hold-up volume < 5 µl Dead stop volume 5 µl (40° rotor)
Materials of construction
Body Polycarbonate Filtrate vessel Polypropylene Membrane
2kDa - 100 kDa Hydrosart
®
125 kDa Cellulose Acetate (CA)
O-ring Medical grade silicone
Table 2: Recommended Spin Speed in Fixed Angle Rotor (x g)
Membrane cut off For DNA For proteins
2 kDa MWCO 7,500 14,000 10 kDa MWCO 7,500 14,000 30 kDa MWCO 5,000 14,000 50 kDa MWCO 5,000 14,000 100 kDa MWCO 3,000 8,000 125 kDa MWCO 2,500* 5,000
Technical Specifications
Desalting | Buffer Exchange
1. Concentrate sample to desired level.
2. Empty filtrate container.
3. Refill concentrator with an appropriate solvent.
4. Concentrate the sample again and repeat the process until the concentration of contaminating microsolute is sufficiently reduced. Typically 3X wash cycles will remove 99% of initial salt concentration.
Vivacon
®
500 Reverse Spinning
| 3
Hydrosart membrane
Reverse spinning concentrate into fresh collection tube
Collection tube fits standard
1.5 | 2.2 ml tube carriers
Filtrate can be sealed for storage
* Spin speed 2,500 + g for DNA samples > 900 bp when using a 125 kDa MWCO.
For DNA samples >650 bp, spin at 1,000 + g.
Loading...
+ 5 hidden pages