Sartorius Signum 1, Signum 2, Signum 3, SIWR, SIWA Service Manual

...
WSI5007-e12036
Service Manual
Sartorius Signum®1 | Signum®2 | Signum®3
Models SIWR | SIWA | SIWS
Signum Series Complete Scales
2
Service Manual Signum
Contents
Introductory Remarks 4
Notes on Using this Manual 4 Printing the Manual 4 Symbols 5 Safety Instructions 6 Model Designation 6 Model Codes: 7 Resolutions: 7 Models with Special Weighing Systems 8 Design 1 - 3 8 Design 4 - 6 9 Additional Tools / Programs, etc. 10 Accompanying Literature 10
Description of the Equipment 11
Display Units 11 Flat Screen Display Unit 11 Tall Display Unit 11 Position Description 12 Overview of the Signum Series Scale Featuring Strain Gauge Technology 13 Overview of the Signum Series Scale Featuring Electromagnetic Force Compensation 14
Activating the Service Mode 15
Activating the Service Mode 15 Additional Menus in the Service Mode 17 In the Text Menu (Setup) 17 Calibration/Adjustment Functions 19
1-9-1External Calibration/Adjustment with Default Weights 19 1-9-3External Calibration/Adjustment with User-Defined Weights 20 1-9-4Calibration/Adjustment with an Internal Weight 2) 22 1-9-6External Linearization with Default Weights 23 1-9-7External Linearization with User-Defined Weights 26 1-9-8Setting the Preload 29 1-9-9Clearing the Preload 30 1-18 Entering Calibration/Linearization
Weights 32
1-18-1 External User-Defined Calibration Weight (ser­vice mode not required) 32 1-18-2 Entering the 1st Linearization Weight 33
3
Service Manual Signum
Replacing the Display Unit 34
Opening the Display Unit 34 Closing the Display Unit 34 Replacing the Front Panel 35 Replacing the Cable 35 Replacing the Display PCB 36
Replacing Components 37
Repairing the Scale 37
Opening a Scale Featuring Strain Gauge Technology 37 Opening a Scale Featuring Electromagnetic Force Compensation 38 Closing the Scale 39 Replacing the Power PCB 40 Replacing the Data Output PCB 41 Replacing the Main PCB 42 Replacing the System PCB 43 Replacing Comonents of SIWAEDG 44
Optional Data Output Ports/Interfaces 46
Installation of the Profibus Module 46 Data Output Plate with IP44 Protection 47 Data Output Plate with IP65 Protection 48
COM 1 Port SIWAEDG 49
Error Codes 50
Additional Tools / Programs, etc. 52 Block Diagram 53
Working with the Service Software 54
Service Switches 54 Function of the Service Switches 54 Menu Access Switch 55 Boot Switch (Flash Switch) 57 SBI/XBPI 57 Close Function 57
Setting the XBPI-232 Protocol 57 Calibration/Adjustment Data for the SIWR 58 Calibration/Adjustment Data for the SIWR 59 Calibration/Adjustment Data for the SIWA 60 Calibration/Adjustment Data for the SIWS 61
4
Service Manual Signum
Introductory Remarks
Introductory Remarks
Notes on Using this Manual
This manual contains information on maintenance, calibration and adjustment of Signum series scales, as well as on error messages.
If you have any comments or require more detailed information, please write to us at: Int.Service@Sartorius.com.
Printing the Manual
To save paper, we recommend configuring the printer settings in the Adobe Acrobat program as follows: File | Print | Properties | Layout: Pages per Sheet: 2
5
Service Manual Signum
Introductory Remark
!
Symbols
The following symbols are used in this manual:
Warning
Indicates safety instructions. Please make sure to observe these instructions; otherwise, physical injury or material damage may result.
Note
The light bulb symbol indicates important information that will help prevent errors, as well as other useful tips.
6
Service Manual Signum
Introductory Remarks
Safety Instructions
The Signum series scale is in conformance with the applicable directives and standards. Use of the scale for other than its intended purpose, however, can result in physical injury or material damage.
Repairs must be performed by service technicians who have been trained at the factory.
Warning
Electric shocks can cause serious burns and life-threatening injuries. To avoid electric shock, always disconnect the power cord from
the wall outlet (mains supply) before beginning repair work or replacement of components. When the scale housing is open, always use an isolating transformer or a ground fault circuit interrupter.
Make sure no voltage is present in the scale before connecting or disconnecting any cables to or from it. Otherwise, assemblies or other components may be destroyed. 1 Press e to switch off the scale. 2 Unplug the power cord. 3 Carry out maintenance/repair work.
For further information, please refer to the operating instructions (WSI6007-e07074.)
!
7
Service Manual Signum
Introductory Remarks
Model Designation
Model Codes:
SIW – Series name R / A / S – Featured technology (R=Regular; A=Advanced with strain gauge technology; S=Supreme with electromagnetic force compensation) 1,2,3 – Application level (display unit); later a 4
th
display unit will be available in addition
X – Indicates explosion-protected (N/A at the initial stage, since it is not available in an explosion-
protected version at this stage)
P – Material (at the initial stage only painted, but will also be available in stainless steel at a later time)
Resolutions:
-R 6,000d< 15,000d
-L 15,.000d
-I 30,.000d
-N multirange n*3,000d
-M multiintervall n*3,000d
-S 50,000d <S<100,000 d
-H >100,000d
-D Dual range high resolution >50,000
-P Polyrange high resolution >50,000
-K Dual range low resolution <= 50,000
-T Polyrange low resolution <=50,000 It is possible to select from various groups of options
(e.g., interfaces, rechargeable battery pack, 2
nd
A/D converter, other IP versions, etc.)
8
Service Manual Signum
Introductory Remarks
Models with Special Weighing Systems
Model Design 1 System
Regular weighing system: Strain gauge technology (to the right) Without built-in motorized calibration weight Nominal load 20kg | 50kg | 100kg
Design 2
Advanced weighing system: Strain gauge technology (to the right) Without built-in motorized calibration weight Nominal load 20kg | 50kg | 100kg
SIWAEDG
weight Nominal load 16kg | 35kg | 65kg
Design 3
Supreme weighing system: Narrow forked-lever system (to the right) With built-in motorized calibration weight Nominal load 20kg
AUT22923a.JPG
PW_Sys_DMS_01.eps
AUT22925a.JPG
PW_Sys_GHS_03.eps
9
Service Manual Signum
Introductory Remarks
PW_Sys_GHS_04.eps
PW_Sys_DMS_05.eps
PW_Sys_DMS_06.eps
AUT22923a.JPG
Model Design 4 System
Supreme weighing system: Narrow forked-lever system (to the right) With built-in motorized calibration weight Nominal load 40kg
Design 5
Regular weighing system: Strain gauge technology (to the right) Without built-in motorized calibration weight PW15-EX | Nominal load up to 30kg
Design 6
Regular weighing system: Strain gauge technology (to the right) Without built-in motorized calibration weight PW10-EX | Nominal load up to 100kg
10
Service Manual Signum
Introductory Remarks
Additional Tools / Programs, etc.
Designation Order No.
- Isolating transformer
- RS232 connection cable (25-pin / 25-pin) 7357312
- RS232 connection cable (25-pin / 9-pin) 7357314
- Torx screwdriver set 10/15/20
Accompanying Literature
Designation Order No.
Operating instructions WSI6007-e07074 Service manuals, spare parts lists, etc. can be downloaded from the Internet at
http://iss.sartoserver.de. Registration is required: please contact Malte.Pramann@ Sartorius.com for details.http://iss.sartoserver.de hinterlegt.
11
Service Manual Signum
Description of the Equipment
PW_Bal01.cdr
TitSig.cdr
Kop_in.jpg
Kop_fl_St.jpg
Description of the Equipment
Display Units
Flat Screen Display Unit
These are available with various screen types. The display unit can be hung on the load cell or secured to the scale using a stand or column.
Tall Display Unit
This is available as an option with an A/D converter or a rechargeable battery pack. Position Description
12
Service Manual Signum
Description of the Equipment
PW_Folie01.epsPW_Folie02.epsPW_Folie03.eps
c Clear Function
e On/off
(in standby mode,
Off is displayed).
Signum 2 and 3 only: n If a second weighing platform is connec-
ted, this key toggles the display between the two readouts.
( – Zero the scale
– Cancel calibration/adjustment
) – Tare the scale k Toggle between 1st and 2nd weight unit,
or gross and net values, or normal and 10-fold higher resolution, depending on operating menu settings (depends on model)
J Start calibration or adjustment p – To print: press briefly.
– To print GMP footer: Press and hold (> 2 seconds)
Signum 3 only: I To toggle the scale to Info mode
Signum 3 only: d Identifiertaste zur Eingabe von Bediener-
kennungen
Signum 2 and 3 only:
K Toggle unit between normal and
10-fold higher resolution
Signum 2 and 3 only: L Netto-Bruttowert-Taste
Nur Signum 2 und 3 w Toggle between display modes within an
application program
Signum 2 and 3 only: r Lets you modify reference values
Signum 2 and 3 only: O Saves a value or starts an application
program.
Signum 3 only: R Saves a value in product data memory
Signum 3 only: D Toggle applications
Signum 2 and 3 only: I Press to view either application data or
manual tare values, depending on the key pressed subsequently (e.g., ))
Signum 2 and 3 only: c – Quit an application or delete an input
character
13
Service Manual Signum
Description of the Equipment
The Sartorius Signum series advanced scale (SIWA) with a mechatronic weighing system
Overview of the Signum Series Scale Featuring Strain Gauge Technology
A Power supply B Data output (various interfaces
can be installed as optional
equipment) C Main PCB D Strain gauge technology
– system PCB E Various versions of the strain
gauge system F Standard housing G Without built-in motorized
internal calibration weight
* This cable is suitable only for the
Ethernet and Profibus options
AUT24188.JPG
A
B
C
AUT24197.JPG
System cable Cable to main PCB
D
Cable to power supply Data output cable * Cable to the system PCB Standard data output PCB
G
E
F
A
C
B
Cable to Standard data output PCB
14
Service Manual Signum
Description of the Equipment
The Sartorius Signum Series Supreme Scale (SIWS) with a Monolithic Weighing System Overview of the Signum Series Scale (Featuring Electromagnetic Force Compensation)
A Power supply B Data output (various interfaces
can be installed as optional
equipment) C Main PCB D Electromagnetic force
compensation system PCB E Various versions of the
electromagnetic force
compensation system F Standard housing G Built-in motorized internal
calibration weight
AUT24252.JPG
AUT24188a.JPG
A
B
C
D
G
E
F
A
C
B
Cable to power supply Data output cable Cable to system PCB Optional RS232 or RS485/422 data interfaces; 4-20mA/0-10V or digital I/O
System cable Cable to main PCB
15
Service Manual Signum
Description of the Equipment
Activating the Service Mode
Important note: When performing maintenance on an existing system, you must access the service mode for linearization, to set or clear a preload, or to enter the date of service.
– Switch the Signum scale off (press the e). – Switch the Signum scale back on (press the e). While all segments are
displayed, briefly press the ) key.
APPL is displayed.
– Using the k key, scroll to SETUP.
– Press the ) key. WP-1 is displayed.
– Using the k key, scroll to CODE.
disp_wp-1_d.eps
disp_8888_d.eps
disp_appl_d.eps
disp_setup_d.eps
disp_wp-1_d.eps
16
Service Manual Signum
Working with the Service Mode
– Press the ) key. A blinking cursor is displayed.
– Using the k key, select 2 and press the ) key to confirm. When the cursor
jumps to the second field, enter the next number. Follow these steps to enter the service code 202122.
Important note:
If you enter a number incorrectly, use the ( key to select the field you wish to change. Next, press the ) key repeatedly until the cursor scrolls to the 6
th
place (line 2 blinks).
– Once you have entered the last number, press the ) and ( keys (in this
order) to activate the service mode.
An „S“ in the upper right-hand corner indicates that the service mode is active. If you wish to exit the service code entry mode, continue pressing the ( key until
CODE reappears on the display. Please note: The only way to deactivate the service mode is to switch the scale
off and then on again.
Important note: Once service code 202122 has been input, any customer codes that have already
been entered can only be viewed. You can delete or change a customer code only after entering the general access code 40414243.
disp_202122_d.eps
disp_blin_d.eps
disp_code_s_d.eps
!
17
Service Manual Signum
Working with the Service Mode
Additional Menus in the Service Mode
In the Text Menu (Setup)
In consecutive order following the menu items » DATE « and Code » 202122 « – Service date » S-DATE « – Memory number » MEM-NO « – Service technician can enter a minimum load here » SQMIN-S « – This displays or prints the minimum load » SQMIN «
In the number menu: The setup menu is model-dependent for WP1 (WP-1) and the interfaces COM1 (COM1)
and UniCOM (UNICOM). It has been expanded to allow the following possible settings for configuring weighing platforms WP-1(WP-1) and WP2(WP-2):
1-9 Calibration/adjustment functions:
1-9-1: Ext. calibration/adjustment with a default weight (service mode not
required)
1-9-3: Ext. calibration/adjustment with a user-defined weight (service mode not
required) 1-9-4: Int. calibration/adjustment (service mode not required) 1-9-6:
1)
Internal linearization (only for WP-2 on COM1 and UniCOM) 1-9-7:1) External linearization with user-defined weights 1-9-8: Set preload (service mode not required) 1-9-9: Delete preload (service mode not required) 1-9-10: Key blocked (service mode not required) 1-9-18:1) Determine internal weight
1)
Only accessible in the service mode
18
Service Manual Signum
Working with the Service Mode
Important note:
Once the necessary settings have been configured, set the menu to the particular calibration/adjustment function that the customer wishes to perform.
1-10 Calibration/adjustment sequence
1-10-1: Calibration with automatic adjustment 1-10-2:
3)
Calibration with manual adjustment
1-18 Entry of calibration and linearization weights
1-18-1: Ext. calibration weight of the user ((special value) entry, e.g.:10,000kg) 1-18-2:
1)
Entry of linearization weight 1
1-18-3:
1)
Entry of linearization weight 2
1-18-4:
1)
Entry of linearization weight 3
1-18-5:
1)
Entry of linearization weight 4
1)
Only accessible in the service mode
9-1 Factory settings / menu reset
9-1-1: Restore 9-1-2: Do not restore
19
Service Manual Signum
Working with the Service Mode
Calibration/Adjustment Functions
Important note:
When using a verified weighing platform, open the cover on the back, left-hand side of the indicator housing. Then slide the menu access switch to the right (»free« setting).
– Activate the service mode (see page 15).
1-9-1 External Calibration/Adjustment with Default Weights
– Select weighing platform »WP-1«. – Press the ) key to access the number menu.
– Select menu »1-9-1« (External calibration/adjustment with default weight) by
pressing the ) k ) key repeatedly and then pressing the ) key briefly to confirm.
If this menu has already been selected, switch the indicator off and then on
again to exit the service mode.
display_wps1_serv.eps
display_serv_01.eps
menu_191.eps
20
Service Manual Signum
Working with the Service Mode
– After taring or zeroing the display of the scale, press and hold the J key until
»C.EXT.DEF« is displayed.
Press the J key to continue the calibration/adjustment routine. – After approx. 2 seconds, the required calibration weight is shown in the display.
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration procedure) is displayed.
– To interrupt the calibration/adjustment procedure, press the ( key.
– Press the J key to calibrate/adjust the scale and display the value if menu »1-
10-2« has been selected.
1-9-3 External Calibration/Adjustment with User-Defined Weights
– Select weighing platform »WP-1«.
– Press the ) key to access the number menu.
3)
Only displayed if menu code »1-10-2 (Calibration with manual adjustment) has been selected
cal_def.eps
Display_WP1_Serv.eps
Display_Serv_01.eps
-100000kg.eps
+0.0010.eps
+100000kg.eps
3)
21
Service Manual Signum
Working with the Service Mode
– Press the ) and k ) keys repeatedly to select menu code »1-9-3« (Ex-
ternal calibration/adjustment with a user-defined weight) and confirm by briefly pressing the ) key.
– You can jump to the first columns in the number menu by pressing the ( key. – To store the selected menu, press and hold the ) key.
– After taring or zeroing the display of the scale, press and hold the J key until
»C.EXT.USE« is displayed.
Press the J key to continue the calibration/adjustment routine.
– After approx. 2 seconds, the required calibration weight (user-defined weight) is
shown in the display.
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration) procedure is displayed.
– To interrupt the calibration/adjustment procedure, press the ( key.
– Press the J key to calibrate/adjust the scale.
3)
Only displayed if menu code »1-10-2 (Calibration with manual adjustment) has been selected
cal_use.eps
menu_193.eps
-100000kg.eps
+0.0010.eps
+100000kg.eps
22
Service Manual Signum
Working with the Service Mode
1-9-4 Calibration/Adjustment with an Internal Weight
2)
Select weighing platform »WP1«.
– Press the ) key to access the number menu.
– Press the ) and k ) keys repeatedly to select menu code »1-9-4«
(Internal calibration/adjustment
2)
). Then briefly press the ( ) key to
confirm.
– You can jump to the first columns in the number menu by pressing the ( key. – To store the selected menu, press and hold the ) key. – After taring or zeroing the display of the scale, press and hold the J key until
»C.INTERN« is displayed.
Press the J key to continue the calibration/adjustment routine. – The motorized, internal calibration weight is placed on the scale and then
removed.
– After approx. 2 seconds, the difference in the current calibration value is
displayed.
– Press the J key to store the current calibration weight. Afterwards, the scale
returns to the normal weighing mode.
2)
Only performed on scales with an internal calibration weight.
3)
Only displayed if menu code »1-10-2 (Calibration with manual adjustment) has been selected
cal_def.eps
Display_WP1_Serv.eps
Menu-194.eps
cal_ci.eps
+00015.eps
+00000.eps
3)
23
Service Manual Signum
Working with the Service Mode
1-9-6 External Linearization with Default Weights
– Activate the service mode (see page 15). – Select weighing platform »WP1«.
– Press the ) key to access the number menu.
– Press the ) and k)) keys repeatedly to select menu code »1-9-6« (Ext.
Linearization with Default Weight). Then briefly press the ) key to confirm.
– Press and hold the ) key to store the selected menu and return the scale to
the weighing mode.
If this menu has already been selected, switch the indicator off and on again to
exit the service mode.
– After taring or zeroing the display of the scale, press the J key until »L.EXT.
DEF« is displayed.
display_serv_01.eps
menu_196.eps
display_wps1_serv.eps
lin-ext-def.eps
24
Service Manual Signum
Working with the Service Mode
– After approx. 2 seconds, the 1st required linearization weight (defined weight)
is displayed. You can enter the weights for linearization via menus »1-18-2« through »1-18-5«2).
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration) procedure is displayed.
– Press the ( key if you need to interrupt the calibration/adjustment procedure. – Press the J key to store the 1st linearization weight. Next, the 2nd required
weight is displayed.
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration) procedure is displayed.
– Press the J key to store the 2
nd
linearization weight. Now the 3rd required
weight is displayed.
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration) procedure is displayed.
*)
Weights displayed depend on the scale model
2)
Only possible in the service mode
3)
Only displayed if menu code »1-10-2« (Calibration with manual adjustment) has been selected.
2.5kg.eps
display_serv_02.eps
5kg.eps
display_serv_03.eps
7.5kg.eps
display_serv_04.eps
*
*
*
3)
3)
3)
25
Service Manual Signum
Working with the Service Mode
– Press the J key to store the 3rd linearization weight. The 4th required weight is
now displayed.
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration) procedure is displayed
– Press the J key to store the 4th linearization weight. The scale now prompts a
zero point (remove all weights from the scale).
– Once the zero point is stored, the scale will automatically return to the weighing
mode.
*)
Weights displayed depend on the scale model.
3)
Only displayed if menu code »1-10-2« (Calibration with manual adjustment) has been selected.
10kg.eps
display_serv_06.eps
display_serv_04.eps
0kg.eps
*
*
3)
26
Service Manual Signum
Working with the Service Mode
1-9-7 External Linearization with User-Defined Weights
– Activate the service mode (see page 15). – Select weighing platform »WP1«.
– Press the ) key to access the number menu.
– Press the ) and k ) keys repeatedly to select menu code »1-9-7« (Ext.
linearization with user-defined weight) and confirm by briefly pressing the ( key.
– To store the selected menu, press and hold the ) key. The scale now returns to
the weighing mode.
If this menu has already been selected, switch the indicator off and on again to
exit the service mode.
– After taring or zeroing the display of the scale, press the J key until »L.EXT.
USE« is displayed.
display_serv_01.eps
menu_197.eps
display_wps1_serv.eps
cal_ext_use.eps
27
Service Manual Signum
Working with the Service Mode
– After approx. 2 seconds, the 1st required linearization weight (user-defined
weight) is displayed. You can enter the linearization weights via menus »1-18-2« through »1-18-5«2.
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration) procedure is displayed
– Press the ( key if you need to interrupt the calibration/adjustment procedure. – Press the J key to store the 1st linearization weight. Next, the 2nd required
weight is displayed.
– After a brief pause, the difference between the current value and the value from
the last adjustment (calibration) procedure is displayed
– Press the J key to store the 2nd linearization weight. Now the 3rd required
weight is displayed.
– After a brief pause, the difference between the current value and the value from
the last adjustment (calibration) procedure is displayed.
*)
Weights displayed depend on the scale model
3)
Only displayed if menu code »1-10-2« (Calibration with manual adjustment) has been selected.
– Press the J key to store the 3rd linearization weight. The 4th required weight is
2.5kg.eps
display_serv_02.eps
5kg.eps
display_serv_03.eps
7.5kg.eps
display_serv_04.eps
*
*
*
3)
3)
3)
28
Service Manual Signum
Working with the Service Mode
now displayed.
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration) procedure is displayed
– Press the J key to store the 4th linearization weight. The scale now prompts a
zero point (remove all weights from the scale).
– Once the zero point is stored, the scale will automatically return to the weighing
mode.
*)
Weights displayed depend on the scale model.
3)
Only displayed if menu code »1-10-2« (Calibration with manual adjustment) has been selected.
1-9-8 Setting the Preload
10kg.eps
display_serv_06.eps
display_serv_04.eps
0kg.eps
*
*
3)
29
Service Manual Signum
Working with the Service Mode
Important note:
Remove the cover on the back, left-hand side of the indicator housing and slide the menu access switch to the right (»free« setting, see page 19 for more information).
– Activate the service mode (see page 15).
– Select weighing platform »WP1«.
– Press the ) key to access the number menu.
– Select menu »1-9-8« (Setting the preload) by pressing the k ) key repeatedly
and then pressing the ( key briefly to confirm.
– Press and hold the ) key to store the selected menu and return the scale to
the weighing mode.
– Tare the scale (by pressing the ) key) or zero the display (by pressing the (
key).
After the scale has been tared, „NET“ might remain in the display.
– Place the preload weight on the scale.
– Press the J key until »SET Vor« appears in the display.
display_serv_01.eps
menu_198.eps
display_wps1_serv.eps
0kg.eps
vorl1.eps
30
Service Manual Signum
Working with the Service Mode
– Press the ) key. After a brief pause, the Signum indicator returns to the
weighing mode.
1-9-9 Clearing the Preload Important note:
Remove the cover on the back, left-hand side of the indicator housing and slide the menu access switch to the right (»free« setting, see page 19 for more information).
– Activate the service mode (see page 15).
– Select the weighing platform »WP!«.
– Press the ) key to access the number menu.
– Select menu »1-9 -9 « (Clear preload) by pressing the k ) key repeatedly
and then pressing the ) key briefly to confirm.
– Press and hold the ) key to store the selected menu and return the scale to
the weighing mode.
– Remove the preload weight from the scale. A weight with a preceding „minus“
sign is displayed (e.g., -0.335kg).
0kg.eps
display_serv_01.eps
display_wps1_serv.eps
menu_199.eps
set_vor.eps
31
Service Manual Signum
Working with the Service Mode
– Press the J key until »CLr Vor« is displayed.
– Press the ) key. The preload is cleared, and after a brief pause, the scale
returns to the weighing mode.
Clr_vor.eps
0kg.eps
vorl4.eps
32
Service Manual Signum
Working with the Service Mode
1-18 Entering Calibration/Linearization Weights
1-18-1 External User-Defined Calibration Weight (service mode not required)
– Activate the service mode (see page 15). – Select weighing platform »WP1«.
– Press the ) key to access the number menu.
– Select menu code »1-18-1 « (Ext. user – defined calibration weight) by pressing
the ) key and the k ) key repeatedly).
– Access the input mode for the user-defined weight by briefly pressing the )
key. The first number blinks.
– Enter the user-defined weight (10,000 kg in this example); press the following
keys in order: ) k ) p ) ) ) k ). Now the last number entered blinks.
– Press the ) key again to store the user-defined weight.
display_wps1_serv.eps
display_serv_01.eps
menu_1181.eps
user1.eps
user2.eps
menu_1181.eps
33
Service Manual Signum
Working with the Service Mode
1-18-2 Entering the 1st Linearization Weight
– Scroll to menu code »1-18-2 « (by pressing the k key).
Important note: The service mode must be active.
– Access the input mode for the 1
st
linearization weight by briefly pressing the )
key. The first number blinks.
– Enter the 1st linearization weight (2,500 kg in this example); press the following
keys in the order indicated: 3*k ) p ) 6*k ) k ) k. Now the last number entered blinks.
– Press the J key to store the 1st linearization weight. Then the scale returns to
menu code setting »1-18-2 «.
– Using the k key, scroll to the menu code settings »1-18-3 «, »1-18-4 «, and
»1-18-5 « and enter the values for Lin. 2, Lin. 3 and Lin. 4 according to the instructions above.
If you do not require all linearity positions, enter „0“ in the unused fields to blank
these lines in the display.
– Once all entries have been made, menu code setting »1-18-5 « must be
displayed.
menu_1182.eps
user1.eps
userl2.eps
menu_1182.eps
menu_1185.eps
34
Service Manual Signum
Repairing the Display Unit
Repairing the Display Unit
Opening the Display Unit
Warning
Disconnect the cable from the power source.
1 You must loosen four nuts (see the picture on the left) to open the display unit. 2 Cut through the warranty stickers (seals). 3 Open the housing. 4 Now all parts are accessible.
Closing the Display Unit
1 Plug all connectors back into the unit. 2 Close the display unit. 3 Reconnect the unit to line voltage. 4 Check the functions of the display unit to ensure that they are working properly.
Important note:
Once the scale has been repaired and checked to ensure its proper functioning, new warranty stickers (seals) must be applied.
AUT23103.JPG
!
35
Service Manual Signum
Repairing the Display Unit
Replacing the Front Panel
In the case of a defective keypad overlay, keys, or display, the entire front panel must be replaced (see the picture on the left).
– Open the display unit (see page 34)
- Replace the entire front panel with a new one
- Close the display unit (see page 34)
Replacing the Cable
– Open the display unit (see page 34)
- Unplug all connectors from the cable
- Loosen the screw
- Replace the cable with a new one
- Close the display unit (see page 34)
– After replacing the cable, use a torque wrench to tighten the cable gland to
3 Nm.
AUT23152.JPG
AUT23301a.JPG
36
Service Manual Signum
Repairing the Display Unit
Replacing the Display PCB Important note:
Do not connect or disconnect live power cables to or from the equipment; always disconnect the power cable from the wall socket (mains supply) first. Otherwise, components could be destroyed.
– Open the display unit (see page 34)
– Unplug the cable from the ST1 connector and the overlay cable from the keypad. – If an A/D converter or rechargeable
battery pack is being used, unplug the cable from connector ST2 also.
– Remove the screws – Remove the display PCB (X)
– Replace the display PCB with a new one
– Connect the cable
- Close the display unit
(see page 34)
AUT24201.JPG
AUT24205.JPG
BU1
St2
St1
BU1
AUT23302a.JPG
AUT23359a.JPG
X
37
Service Manual Signum
Replacing Components
Repairing the Scale
Opening a Scale Featuring Strain Gauge Technology
IP44 IP65
Warning:
Disconnect the cable from the power source.
1 Remove the load cell.
2 Loosen and remove the two screws (Q)
from the support plate.
3 Remove the four screws (R)
from the O-ring (see picture on the right).
4 Remove the ten screws (P) from the cover.
Cut through the warranty stickers (seals). Remove the O-ring (on units with IP65
protection).
5 The scale housing is now accessible.
6 Remove the six screws (O) from the data
output plate (U).
Remove the gasket (on units with IP65
protection).
!
1
2
4
5
6
O
P
Q
Explo00a.jpg
Explo01z.jpg Explo01e.jpg
U
1
2
3
4
5
P
Q
R
Explo00a.jpg Explo01y.jpg
38
Service Manual Signum
Replacing Components
Opening a Scale Featuring Electromagnetic Force Compensation
Warning
Disconnect the cable from the power source.
1 Remove the load cell plate.
2 Loosen the two screws (Q) on the load cell
and remove the load cell.
3 Place the load cell to the side
(see picture on the right).
4 Remove the ten screws (P) from the cover.
5 Cut through the warranty stickers (seals). 6 The scale housing is now accessible.
AUT24260a.JPG
AUT24261.JPG
AUT24252.JPG
!
Q
P
AUT24262.JPG
39
Service Manual Signum
Replacing Components
AUT24263a.JPG
Closing the Scale
1 Plug the connectors back in. 2 Reassemble the parts of the housing in order. 3 If necessary, be sure to include the all O-rings and gaskets. 4 Replace and tighten the screws (P). 5 Plug the unit back into AC power. 6 Check the scale to ensure that it is functioning properly.
Important note:
Once the scale has been repaired and checked to ensure its proper functioning, new warranty stickers (seals) must be applied.
40
Service Manual Signum
Replacing Components
Replacing the Power PCB
Warning
Disconnect the cable from the power source.
Important note:
Do not connect or disconnect live power cables to or from the equipment; always disconnect the power cable from the wall socket (mains supply) first. Otherwise, components could be destroyed.
– There are three different power supply versions: (B1) 24V and (B2) 230V and (B3)
Zone 2.
– Loosen and remove six screws from the data output plate (U). – Cut through the warranty stickers (seals). – Fold down the data output plate (U).
– Pry the protective cap (A1) from the power PCB (B1).
To do this, insert the end of a slotted-head screwdriver into each of the four
openings on the protective cap and carefully push the retainer clips, located further down, to one side.
– Remove the protective cap (A1) and disconnect the wire (red/brown).
– Unplug connector X, remove the 2 screws (V) and replace the power supply (B1). – Make sure to return the protective cap to its original position.
– Apply new warranty stickers (seals) to the unit.
AUT22922a.JPG
AUT22922b.JPG
B2
!
A1
V
W
V
AUT24188.JPG
B3
AUT24186.JPG
B3
B1
X
X
X
X
U
41
Service Manual Signum
Replacing Components
!
AUT22916.JPG
- +
Z
AUT22917b.JPG
O1
U
O
AUT24188.JPG
Z
U
Replacing the Data Output PCB
Warning
Disconnect the cable from the power source.
Important note:
Do not connect or disconnect live power cables to or from the equipment; always disconnect the power cable from the wall socket (mains supply) first. Otherwise, components could be destroyed.
Different versions of this PCB (see pages 59-60) can be installed in the unit.
– Loosen and remove six screws (O) from the data output plate (U). – Cut through the warranty stickers (seals). – Remove the data output plate (U).
– When you replace the data output PCB (Z) make sure that the connectors are
plugged in correctly.
– If you are replacing the battery, it is imperative that you install it with the
correct polarity (+ / -).
42
Service Manual Signum
Replacing Components
Replacing the Main PCB
Warning Disconnect the cable from the power source.
Important note: Do not connect or disconnect live power cables to or from the
equipment; always disconnect the power cable from the wall socket (mains supply) first. Otherwise, components could be destroyed.
Different versions of this PCB (see picture on the left) can be
installed in the unit.
– Loosen and remove six screws (O) from the data output plate (U). – Cut through the warranty stickers (seals). – Remove the data output plate (U).
– When you replace the digital PCB, be sure that all cables are reconnected.
AUT23155a.JPG
!
AUT24188.JPG
Z
U
43
Service Manual Signum
Replacing Components
AUT24194.JPG
AUT24192.JPG
AUT24197a.JPG
AUT24192.JPG
Replacing the System PCB
Important note:
Please refer to page 37, „Opening a Scale Featuring Strain Gauge Technology
– Strain gauge scale with system PCB – When you replace the system PCB, make sure that all cables have been
reconnected.
Important note:
See page 38, „Opening a Scale Featuring Electromagnetic Force Compensation
– Electromagnetic force compensation scale with system PCB – When you replace the system PCB, make sure that all cables have been
reconnected.
44
Service Manual Signum
Component Replacement SIWAEDG
Replacing the Power Supply PCB Opening the Display Unit (
see page 34 )
Caution: Unplug the power cord
– The protective cover needs to be lifted when replacing the power supply PCB. To do this, insert a regular screwdriver into the four openings of the protective cover
and carefully press the clips located at the bottom to the side.
– Remove the protective cover and disconnect the power cord on the right (blue/brown). – Remove the left plug (red/blue), remove the screws and replace the power supply.
Do not forget to reattach the protective cover.
Replacing the Main PCB
– Loosen and remove the ribbon cable and power supply cable – Remove the screws, cable and replace the PCB – Once the main PCB has been replaced, make sure that the plug connections are properly
reconnected.
45
Service Manual Signum
Component Replacement SIWAEDG
Replacing the System Board
Caution: Unplug the power cord
– Remove the ribbon cable and system plug – Remove the screws and replace the PCB
– Check to make sure that the plug connections are connected properly after replacement.
Replacing the Data Output PCB
– Remove the screws, cable and replace the PCB – Once the data output PCB has been replaced, make sure that the plug connections are
properly connected.
– If the battery is being replaced, make sure that the polarity (+ / -) is correct.
46
Service Manual Signum
Optional Data Output Ports/Interfaces
Installation of the Profibus Module
Profibus Module Plate with IP65 Protection
Important note:
After installation of the Profibus module, the cable connectors must
be equipped with a flat ferrite and glued with silicone.
(see pictures on the rigth).
Data Output Plate with IP65 Protection
DSC03656a.JPG
DSC03654a.JPG
AUT23155a.JPG
Bu1a
DSC03658.JPG
DSC03653a.JPG
DSC03651a.JPG
Bu2a
St2a
St1a
FF
FF
SI
SI
47
Service Manual Signum
Optional Data Output Ports/Interfaces
Data Output Plate with IP44 Protection
Com1: RS232 Com1: RS232 + Option COM2
Com1: RS232 Com1: RS232 + Option COM2
Com1: RS232+ Com1: RS232+ + Option COM2
Option Clock Com1: RS232++ Com1: RS232++ + Option COM2 Com1: RS232++ + Option COM2
Option Clock + PS/2
Ethernet
48
Service Manual Signum
Optional Data Output Ports/Interfaces
Data Output Plate with IP65 Protection
Com1: RS232 Com1: RS232 + Option COM2
Com1: RS232 Com1: RS232 + Option COM2
Com1: RS232+ Com1: RS232+ + Option COM2
nicht benutzt
nicht benutzt
nicht benutzt
49
Service Manual Signum
Com1 Interface SIWAEDG
Installing the Data Output Cable
– Open the display unit (see page 34)
- Connect the cable according to the assignment chart
- Close the display unit (see page 34)
– After installing the cable, tighten the cable gland using a torque of 3 Nm.
J1
1
Screw Terminal Assignment
Terminal Function Terminal Function 1 CTS 11 LOAD PRINT 2 DTR 12 RESET OUT 3 RxD 13 GND 4 TxD 14 GND 5 GND 15 5V OUT 6 Universal 16 5V switched 7 Minor (<) 17 Keyboard data 8 PARES (=) 18 Keyboard clock 9 MAJOR (>) 19 Shield 10 SET 20 LINE OUT
50
Service Manual Signum
Error Codes
Error codes are shown on the main display. Err codes are shown continuously; Inf messages are shown for 2 seconds, after which the program returns automatically to the weighing mode.
Display Cause Solution
Err 101 - 104 Key is stuck Release key or Key pressed at power on Contact your local Sartorius Service Center
Err 320 Program memory defective Contact your local Sartorius Service Center Err 335 Verified weighing platform not compatible Connect a compatible weighing platform
with the connected terminal Err 340 Operating parameter memory (EEPROM) defective Turn the scale off and then on again
If Err 340 is still displayed, contact your local Sartorius Service Center
Err 341 Data lost from RAM; Leave the scale power on for at least 10 hrs. battery needs to be recharged Err 343 Data lost from the memory module Contact your local Sartorius Service Center
for transaction numbers in external alibi memory
Inf 01 Data output not compatible with output format Change the menu settings Inf 02 Calibration/adjustment condition Calibrate only when zero is displayed.
not met; for example, not tared or Press ) to tare there is a load on the weighing pan Unload the scale
Inf 03 Adjustment could not Allow the scale to warm up again and then be completed within a certain time repeat the adjustment process
Inf 06 Built-in calibration weight defective Contact your local Sartorius Service Center
51
Service Manual Signum
Error Codes
Display Cause Solution
Inf 07 Function not allowed in scales Contact your local Sartorius Service Center verified for use in legal metrology for details on changing settings
Inf 08 The load on the scale is too heavy to zero the readout Check whether “Tare/zero at power on” is set (1.12) Inf 09 Taring is not possible when the gross weight Zero the scale
is a minus value Inf 10 Tare key is blocked when there The data stored in the 2nd tare memory
is data in the tare memory (Combics 2 only) must be deleted (clear the memory) before taring
Inf 22 Error in storing reference value, load is too light Put a heavier weight on the scale Inf 23 Error in initializing an application Contact your local Sartorius Service Center Inf 29 Minimum load not reached Change menu, or perform “Close” function check
the interface and cable of the connected devicen Contact your local Sartorius Service Center
Inf 30, 31 Indicator is in the xBPI-mode Define a lower value for the minimum load (in the Application settings, menu item 3.6)
Inf 71 Cannot store the current weight value None (e.g., if control limits are too low or too high)
Inf 72 Cannot store the current weight value None (e.g., the transaction counter has reached its limit)
Inf 73 Data not found or unreadable Contact your local Sartorius Service Center
52
Service Manual Signum
Error Codes
Display Cause Solution
Inf 74 Function is blocked (e.g., menu is locked) None Inf 98 No weighing platform connected Connect weighing platform Inf 99 No weighing platform connected Connect weighing platform NO WP No weighing platform connected Connect weighing platform
In addition to standard tools, you will need the following special tools and programs to work on Signum series scales
Designation Order No.
- Sartorius Service Software (H8S) 6740-88
with USB dongle
*)
- Sartorius Service Software (H8S) 6740-89
with LPT dongle
*)
*)
Registration in ISSS required
53
Service Manual Signum
Block Diagram
Block Diagram
54
Service Manual Signum
Working with the Service Software
Service Switches
There are two sliding switches on the data output PCB for use by service technicians. These switches are located on the back of the scale and can be activated through the borehole using a thin object.
The illustration shows two individual switches:
- Menu access switch (L)
- Flash switch (M) —> see page 20 for details.
Function of the Service Switches
Menu Access Switch
Warning
Only open this switch for service purposes. At all other times, it should remain in the „closed“ position.
This switch must be open in order to work with the Sartorius Service Software. If the switch remains closed, the error message „ACCESS SWITCH LOCKED“ will be displayed in the service program.
Men-Swi-Li.eps
Upper switch right closed
Upper switch Left open
Men-Swi-Re.eps
!
AUT23003a.JPG
M
L
55
Service Manual Signum
Working with the Service Software
Boot Switch (Flash Switch)
Warning
Activate this switch only if an attempt to program the application­data memory with the PPLOADER program fails (programming routine stops responding).
Procedure for returning the PCB to bootable mode: – Disconnect the scale from AC power. – Slide the boot switch to the right and plug the scale back into AC power. – Afterwards, slide the switch to the left again. After a short time, the PCB is
ready to boot (reload the application software using PPLOADER).
SBI/XBPI
Important note: The COM 1 interface must be set to XBPI (eXtended Binary
Processor Interface) protocol before you can perform calibration/
adjustment procedures or program the scale using the Sartorius Service Software (see page 22).
Close Function
After using the service software, select the „Close“ function to reactivate write­protection and return the weighing system to SBI mode for data output. When the SBI mode is active the scale uses the SBI protocol, which is required for communication with the printer.
!
Men-Swi-Li.eps
Lower switch left open
Lower switch right closed
Men-Swi-Li.eps
56
Service Manual Signum
Working with the Service Software
Setting the XBPI-232 Protocol
You need to set the scale to the XBPI protocol (BPI-232) in order to perform calibration/adjustment procedures or program the scale using the service software.
– Switch the Signum scale off (press the e key). – Switch the Signum scale back on (by pressing the e key). While all
segments are displayed, briefly press the ) key.
APPL appears in the display.
– Using the k key, scroll to SETUP.
– Press the ) key. WP-1 is displayed.
disp_wp-1_d.eps
disp_8888_d.eps
disp_appl_d.eps
disp_setup_d.eps
57
Service Manual Signum
Working with the Service Software
– Mit der Taste k bis COM-1 scrollen. – Use the k key to scroll to COM-1.
– Press the ) key. DATPROT is displayed.
– Press the ) key again. BPI-232 appears in the display.
Important note:
The X is not displayed when BPI-232 is selected.
disp_xbpi_d.eps
disp_com-1_d.eps
disp_datprot_d.eps
58
Service Manual Signum
Calibration/Adjustment Data
Calibration/Adjustment Data for the SIWR
Example for the order number of a Signum full range scale: SIWRDCP-1-15-L or SIWRDCP-3-15-NCE
Model type | Sensor | Platform | Type of/ | Application- | Capacity | Display | Verifiable/ | technology dimensions material level (kg) resolution verified (in mm) versions SIW R
1)
DC P
4)
1 15 N NCE
1)
= SIWR: „Regular“ standard weighing system - (strain gauge technology)
4)
= Painted
59
Service Manual Signum
Calibration/Adjustment Data
Calibration/Adjustment Data for the SIWR
Example for the order number of a Signum full range scale: SIWRDCP-2-60-R oder SIWRDCP-3-60-RCE
Model type | Sensor | Platform | Type of/ | Application- | Capacity | Display | Verifiable/ | technology dimensions material level (kg) resolution verified (in mm) versions SIW R
1)
DC P
4)
2 15 R RCE
1)
= SIWR: „Regular“ standard weighing system - (strain gauge technology)
4)
= Painted
just_Signum_290508.xls
60
Service Manual Signum
Calibration/Adjustment Data
Calibration/Adjustment Data for the SIWA
Example for the order number of a Signum full range scale: SIWADCP-2-35-S
Model type | Sensor | Platform | Type of/ | Application- | Capacity | Display | Verifiable/ | technology dimensions material level (kg) resolution verified (in mm) versions SIW A
2)
DC P
4)
2 35 S
2)
= SIWA: „Advanced“ mechatronic weighing system (strain-gauge technology)
4)
= Painted
Signum SIWAEDG
SIWAEDG-3-65-S 65 kg 1 g 20 kg 1 g 20 kg 2 g
F2
20 kg 20 kg 1 g - kg 0 - Max kg 3 g 8 n
SIWAEDG-3-35-S 35 kg 0,5 g 10 kg 0,5 g 10 kg 2 g
F2
10 kg 10 kg 0,5 g - kg 0 - Max kg 1,5 g 6 n
SIWAEDG-3-16-S 16 kg 02 g 5kg 0,2 g 5kg 0,8 g
F2
5kg 5kg 0,2 g - kg 0 - Max kg 0,6 g 6 n
61
Service Manual Signum
Calibration/Adjustment Data
Calibration/Adjustment Data for the SIWS
Example for the order number of a Signum full range scale: SIWSDCP-3-16-H oder SIWSDCP-3-16-HCE
Model type | Sensor | Platform | Type of/ | Application- | Capacity | Display | Verifiable/ | technology dimensions material level (kg) resolution verified (in mm) versions SIW S
3)
DC P
4)
1 15 N NCE
3)
= SIWS: „Supreme“ monolithic weighing system
4)
= Painted
just_Signum_290508.xls
Sartorius Weighing Technology GmbH Weender Landstraße 94–108 37075 Goettingen, Germany Telephone (0551) 308-4440 Fax (0551) 308-4449 Internet: http://www.sartorius-mechatronics.com E-mail: Int.Service@Sartorius.com
Copyright by Sartorius, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius.
The status of the information, specifications and illustrations in this manual is indicated by the date given below. Sartorius reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
Status: March 2012 Sartorius, Goettingen, Germany