Sartorius secura,quintix User Manual

98648-020-26
User Manual
Secura®, Quintix
®
26, 35, 65, 125, 125D, 225D, 324
Laboratory Balances
Visit www.balances.com for the full line of Sartorius Laboratory Balances
Visit www.balances.com for the full line of Sartorius Laboratory Balances
Secura, Quintix User Manual 3
Contents
Contents
1. User Information .......................................... 4
1.1 Warning/Danger Symbols.............................4
1.2 Explanation of Symbols...............................4
1.3 Application Advice/Technical Support
................ 5
2. Safety Information
........................................ 6
2.1 Guidelines and General Information.................. 6
2.2 Installation Information .............................. 6
2.3 Intended Use.......................................... 7
3. General View of the Equipment.......................... 8
4. Getting Started
........................................... 9
4.1 Unpacking and Equipment Supplied.................. 9
4.2 Installing the Balance................................10
4.3 Choosing a Location .................................12
4.4 Moving the Balance..................................13
4.5 Power Connection (Power Supply) ..................14
4.6 Warm-up Time .......................................15
4.7 Switching On the Balance
...........................16
4.7.1 Startup Wizard ..............................16
4.8 Anti-theft Locking Device ...........................18
5. Using the Balance
........................................19
5.1 Switching the Balance On/Off (Standby) ............19
5.2 Leveling the Balance .................................20
5.2.1 Leveling with Electronic Level Indicator
(Secura®) .....................................20
5.2.2 Leveling with a Conventional Level Indicator
(Quintix®) ....................................21
5.3 Operating Concept...................................22
5.3.1 Operating and Display Elements ............22
5.3.2 Menu ........................................23
5.3.3 Entering Figures on the Number Pad
........24
5.3.4 Entering Text and Characters (Secura® only) 24
5.3.5 Accessing an Application in the Menu ......25
5.3.6 List of Applications in the Menu ............25
6. System Settings (Setup Menu)
..........................27
6.1 Accessing the Menu and Changing Settings ........27
6.2 List of Available Settings ............................28
6.2.1 Language ....................................28
6.2.2 Date and Time ...............................29
6.2.3 Device Information ..........................30
6.2.4 Calibration/Adjustment......................30
6.2.5 Weighing ....................................30
6.2.6 Printout......................................33
6.2.7 Identifier (Secura® only) .....................34
6.2.8 USB Port .....................................35
6.2.9 Display Brightness ...........................37
6.2.10 Acoustic Signal ..............................37
6.2.11 More Settings................................38
7. Weighing
.................................................40
7.1 Mass Unit Conversion................................41
7.2 SQmin Minimum Weighing (Secura® only) ..........43
7.3 Individual Identifiers (Secura
®
only) .................44
7.4 Mixing................................................45
7.5 Statistics .............................................48
7.6 Totalizing Components ..............................51
7.7 Density ...............................................54
7.8 Percentage ...........................................58
7.9 Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
7.10 Unstable Condition ..................................63
7.11 Checkweighing.......................................65
7.12 Peak Hold ............................................67
7.13 Counting .............................................69
8. Calibration and Adjustment
.............................72
8.1 Calibration/Adjustment Using
an Internal Calibration Weight ......................73
8.2 Calibration/Adjustment Using an External Calibration Weight
......................74
8.3 Automatic Calibration/Adjustment with isoCAL.....75
9. ISO/GLP-compliant Printout
............................76
10. USB Port
..................................................78
10.1 Communication with Peripheral Devices ............78
10.2 Direct Transfer of Data (PC) .........................79
10.3 Interface Specification...............................85
10.3.1 Data Output .................................85
10.3.2 Data Output Formats ........................85
10.3.3 Data Output Format with 22 Characters....87
10.3.4 Data Input ...................................88
10.3.5 Overview of Interface Commands
...........89
11. Status Messages
..........................................90
11.1 Key Fadeout..........................................90
11.2 Application Error Messages ..........................90
12. Transporting the Balance
................................92
12.1 Detaching the Draft Shield/Weighing Pan ..........92
12.2 Packing the Balance .................................93
13. Care and Maintenance
..................................94
13.1 Service ...............................................94
13.2 Cleaning the Balance ................................94
14. Disposal
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
15. Specifications ............................................98
15.1 General Data .........................................98
15.2 Model-specific Data ................................100
15.3 Verified Models with Country-specific
Type Approval Certificate...........................101
15.4 Accessories ..........................................103
15.5 Balance Dimensions.................................104
15.5.1 Secura®, Quintix® ...........................104
15.5.2 Secura® 26 ..................................105
Appendix
....................................................106
Visit www.balances.com for the full line of Sartorius Laboratory Balances
4
User Information
Secura, Quintix User Manual
1. User Information
1.1 Warning/Danger Symbols
Warning and danger symbols used in these instructions:
These notes identify hazards which have a high probability of resulting in death or
serious physical injury if not avoided.
These notes identify hazards that can result in moderate or mild injuries if not
avoided.
These notes identify hazards associated with the risk of material damage.
1.2 Explanation of Symbols
The following symbols are used in this manual:
Useful information and tips
Notes for use in legal metrology. For instruments the conformity assessment
procedure has been performed (verified balances)
Note regarding device operation
When individual buttons are displayed, they should be pressed.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
5
User Information
Secura, Quintix User Manual
The following symbols are used in these instructions:
t Indicates a required action
y Describes what happens after you have performed a particular step
Perform steps in the specified order:
1. First action
2. Second action
3. … – Indicates an item in a list
Conventions for this User Manual: – The illustrations in these instructions are based on “Standard” balances.
On balances verified for use in legal metrology, some displays and reports may deviate slightly from the figures. Where this is significant for operation, the differences will be explained in the text.
1.3 Application Advice/Technical Support
Contact addresses for application advice and our technical support can be found online at: http://www.sartorius.com
Visit www.balances.com for the full line of Sartorius Laboratory Balances
6
Safety Information
Secura, Quintix User Manual
2. Safety Information
2.1 Guidelines and General Information
– The balance complies with EU Directives and standards for electrical safety and
electromagnetic compatibility*. Improper use or handling, however, can result in damage and/or injury. Any improper use or operation of the balance, i.e., that is not consistent with the instructions, will result in forfeiture of all claims under the manufacturer's warranty.
– Personnel need to have read and understood these installation instructions,
including the safety instructions.
– In the event of use in systems and ambient conditions which have greater safety
requirements, you must observe the requirements and provisions applicable in your country.
– Always keep the equipment and balance freely accessible.
Any installation work or balance operation that does not conform to the instruc­tions will result in forfeiture of all claims under the manufacturer's warranty.
* = see “Specifications”
Danger of Explosion!
Do not use this equipment in hazardous areas in which explosive materials are present.
Make sure that the voltage rating printed on the AC adaptor is identical to your local
mains voltage.
2.2 Installation Information
Do not operate the balance if its housing or AC adaptor is damaged. Immediately
disconnect the damaged device from the power by pulling the plug.
Do not expose the balance, its AC adaptor or the accessories supplied by Sartorius
to extreme temperatures, aggressive chemical vapors, moisture, shocks, vibrations or strong electromagnetic fields. Observe the conditions of operation described in the Specifications.
Note on Installation:
The operator shall be solely responsible for any modifications to the equipment and for connecting any cables or equipment not supplied by Sartorius. Information on operational quality is available upon request from Sartorius. You should only use peripherals supplied by Sartorius.
Note the IP protection class of the balance and its AC adaptor. Do not allow liquid penetration. The protection class specifies the suitability of equipment for various environmental conditions (moisture, foreign bodies).
Before cleaning the AC adaptor or the balance: Unplug from the mains power.
The balance may only be opened by specialized personnel trained by Sartorius. Do not open the AC adaptor.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
7
Safety Information
Secura, Quintix User Manual
Hazards during Installation and Operation:
If glass breaks, there is a risk of injury posed by cuts from glass edges.
Lay the cables where they pose no risk of causing someone to trip.
Observe the additional safety and danger information in the following chapters.
2.3 Intended Use
This high-precision balance is designed to be used exclusively indoors under normal atmospheric conditions. It was developed specifically for the exact determination of the mass of materials in liquid, paste, powder or solid form.
Appropriate containers must be used for each type of sample material.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
8
General View of the Equipment
Secura, Quintix User Manual
3. General View of the Equipment
3
4
1
2
5
6
7
8
9
10
11
12 13
14 15
16
No. Name and Function No. Name and Function
1 Cover 9 Leveling feet 2 Grip, upper panel (draft shield) 10 Rear device cover
(only for service)
3 Weighing pan 11 Manufacturer's ID label, for verified models
Additional ID label
4 Shield disk (Secura
®
26) 12 Lug for attaching anti-theft device
5 Collection container 13 Verification access switch: Blocks various functions and
units on verified balances 6 Grip, side panel (draft shield) 14 USB port for mini AB 7 Level indicator (Quintix
®
model) 15 DC jack
8 Touchscreen 16 Fastening point for "Kensington" key lock
(optional)
Visit www.balances.com for the full line of Sartorius Laboratory Balances
9
Getting Started
Secura, Quintix User Manual
4. Getting Started
The balance must be disconnected from the mains power for all assembly work.
4.1 Unpacking and Equipment Supplied
t Open the packaging, making sure to remove all parts carefully.
t After unpacking the balance, check it immediately for any external damage.
y If you detect any damage, proceed as directed in the chapter entitled “Care and
Maintenance.”
y Save all parts of the original packaging for any future transportation. During
shipment, please do not leave cables plugged in!
The following parts are included in the equipment supplied:
Components Secura
®
Quintix
®
Model: 26 225D
125
324 125D6535
Draft shield (4 panels)
99999
Collection container
99999
Weighing pan: d 80 mm
9
99
Weighing pan: d 90 mm
–*
9
**
Weighing pan: d 50 mm
9
----
Shield disk (normal / high)
9
----
AC adapter with country-specific power plug
 9
 9
 9
 9
 9
In-use dust cover
99999
Dust cover
99999
Installation and operating instructions
 9
 9
 9
 9
 9
Application guide
99999
* optional
Visit www.balances.com for the full line of Sartorius Laboratory Balances
10
Getting Started
Secura, Quintix User Manual
4.2 Installing the Balance
Installing the Collection Container
t Insert the collection container into the balance.
Installing the Panels (Draft Shield)
t Press against the front of the cover from both sides and fold it upwards.
1) Inserting the front panel of the draft shield:
t Insert the panel on the top into the guide rails of the draft shield.
t Press the panel gently upwards, while simultaneously pressing downwards into
the slot of the draft shield.
y The panel must click into place in the slot.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
11
Getting Started
Secura, Quintix User Manual
2) Inserting the side panels:
t Mount the guide carriages (1) of the side panel by inserting them into the guide
rails on the top.
t Insert the side panel at the bottom behind the guide rod(2) of the draft shield.
t Make sure that:
– the guide carriage(1) is properly and fully positioned on the guide rail.
– the side panels are positioned behind the guide rod(2).
– the panels can be moved smoothly and freely.
3) Inserting the upper panel:
t Insert the panel on the top of the draft shield guide.
t Push the panel carefully backwards.
t Gently lift the panel by its grip in order to push it below the guide rod(3).
t Close the cover by folding it downwards completely until it locks.
Inserting the Weighing Pan
t Insert the weighing pan into the opening of the collection container.
t Push the upper and side panels forward to close them.
Secura® 26:
t Insert the weighing pan into the opening of the collection container.
t Position one of the two shield disks on the collection container.
t Push the upper and side panels forward to close them.
1
2
3
Visit www.balances.com for the full line of Sartorius Laboratory Balances
12
Getting Started
Secura, Quintix User Manual
4.3 Choosing a Location
t Select the proper setup location:
– Set up the device on a stable, even surface that is not exposed to vibrations
(e.g., a weighing stone).
– Maintain free access to the balance at all times.
Choose a location that is not subject to the following negative influences: – Heat (heater or direct sunlight) – Drafts from open windows, AC systems, and doors – Extreme vibrations during measurement – Heavy traffic areas (personnel) – Extreme humidity – Electromagnetic fields – Extremely dry air
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when it is brought into a substantially warmer area. You should therefore let a device that has been disconnected from its power source acclimatize for approximately 2 hours before reconnecting it to the power.
°C
Visit www.balances.com for the full line of Sartorius Laboratory Balances
13
Getting Started
Secura, Quintix User Manual
4.4 Moving the Balance
Moving the Balance over Short Distances in the Lab
Avoid glass breakage, shocks and vibrations: Never lift or carry the balance by its draft shield!
t Hold the balance under the housing, lift it up carefully, and carry it to its new
location.
The balance needs to be re-leveled and then adjusted each time its setup location is changed.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
14
Getting Started
Secura, Quintix User Manual
4.5 Power Connection (Power Supply)
AC Adaptor Assembly
Using the wrong power plug adaptors may cause fatal electric shock and damage the equipment. Never plug the power plug adaptor into the socket when it is disconnected from the device (danger of electrical shock).
t Use the right power plug adaptor for your mains power supply.
A AC adaptor versions: Bag Region/Country Bag Region/Country
a) transparent Europe/EU (except United Kingdom) e) black India b) yellow United Kingdom f) turquoise South Africa c) blue USA, Canada, China, and Japan g) green Korea d) red Australia h) pink Brazil
i) white Argentina
The packaging of AC adaptor module 6971790 with line adaptors g), h), and i) is labeled with 98671791.
t Push (1) and slide (2) the power plug adaptor (A) required for your power supply
into the opening of the AC adaptor module (B). When doing this, the AC adaptor needs to lock into position.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
15
Getting Started
Secura, Quintix User Manual
Removing/Replacing the AC Adaptor
t Unlock (1) the AC adaptor and then detach (2) it.
Power Connection/Safety Precautions
– Only use original Sartorius AC adaptors. The AC adaptor has an IP rating of IP40
in accordance with EN 60529.
– Make sure that the voltage rating printed on the manufacturer's ID label is
identical to that of your local mains voltage.
– If the stated supply voltage or the plug design of the AC adaptor does not comply
with your country's standard, please inform your nearest Sartorius representative.
– The power must be connected in accordance with the regulations applicable in
your country.
t 1) First, plug the AC adaptor into the socket at the back of your balance.
t 2) Then connect the AC adaptor to the wall outlet (mains voltage).
4.6 Warm-up Time
y To ensure accurate results are delivered, the balance must warm up for at least
30 minutes after initial connection to the power supply. Only after this time will the balance have reached the required operating temperature.
When a verified balance of accuracy class
k for use in legal metrology is connected
to the mains power, it must warm up for at least one hour before operation.
0
30
Visit www.balances.com for the full line of Sartorius Laboratory Balances
16
Getting Started
Secura, Quintix User Manual
4.7 Switching On the Balance
t Touch j on the display to switch on the balance.
y When the balance is switched on for the first time or when it has been reset to
the factory settings, the startup wizard appears.
4.7.1 Startup Wizard
The default language for the display text is English. When the balance is switched on for the first time, the startup wizard is automatically activated. Follow the instructions on the interactive display to configure your selections for:
– Language – Date format/date – Time format/time
y The Language settings window appears.
t Touch the language you would like to select, such as Deutsch.
t Touch l to confirm.
y The display changes immediately to the desired language.
y The Getting Started settings window appears.
The display format, date, and time can be set here.
t Touch START.
y The Date Format settings window appears:
t Select how the date should be displayed and printed out.
DD-MMM-YYYY:
The date is displayed in the order of day, month and then year.
MMM-DD-YYYY:
The date is displayed in the order of month, day and then year.
YYYY-MM-DD (ISO):
The date is displayed in the order of year, month and then day. (The time is displayed in 24-hour format with this setting.)
t Touch l to confirm.
Language
Deutsch
Français
English
Installation
Initializing date and time.
START
Date format
DD-MM-YYYY
MM-DD-YYYY
YYYY-MM-DD (ISO)
Visit www.balances.com for the full line of Sartorius Laboratory Balances
17
Getting Started
Secura, Quintix User Manual
y The current date is displayed.
t If the date is displayed correctly, select l to confirm.
t To adjust the date manually, if required, touch t, enter the current date and
confirm with l.
y The Time Format settings window appears:
t Select how the time is displayed and printed out.
24h:
The time is displayed in 24-hour format. (This is the only available setting when the date is set to ISO format.)
12h (AM/PM)
The time is displayed in 12-hour format. The hours before noon are displayed with AM and the hours after noon are displayed with PM.
t Touch l to confirm.
y The current time is displayed.
t If the time is displayed correctly, touch l to confirm.
t To adjust the time manually, if required, touch t, enter the current time and
confirm with l.
y The
Leveling settings window appears:
t Follow the instructions on the interactive display.
Follow the instructions in Chapter “Leveling the Balance” on page 20.
If required, you can subsequently modify your selection in the “Settings” menu
(see Chapter “Accessing the Menu and Changing Settings” on page 27).
Date
2014-10-06
Time format
24 h
12 h (AM/PM)
Time
15.55
Leveling
Please turn each foot in the direction of the arrow as shown.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
18
Getting Started
Secura, Quintix User Manual
4.8 Anti-theft Locking Device
t A “Kensington” key lock can be installed at the fastening point on the back of the
balance if required.
t The balance can also be secured at the installation location with a chain and lock,
for example.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
19
Using the Balance
Secura, Quintix User Manual
5. Using the Balance
5.1 Switching the Balance On/Off (Standby)
0.00000
Max 120 g d=0.00001 g
y The leveling screen is displayed first after any model is switched on.
Level the balance if necessary. The actual weighing screen is only displayed after confirmation.
– When the balance is switched on, the following message appears on the display:
Secura®: Level The balance has been leveled (if leveled). Quintix®: Level Please check leveling.
– If the balance needs to be leveled, a message appears on the display (see
“Leveling the Balance” on page 20).
– The balance must be regularly calibrated and adjusted in order to achieve
accurate weighing results (see “Calibration and Adjustment” on page 72).
– The default language for the display text is English.
You can change the language (see “Language” on page 28).
t To switch the balance to standby, select the menu key at the bottom left of the
weighing screen on the display.
ComponentsStatistics
MixingWeighing
y The menu is displayed.
t Select the j button in the application menu.
y The balance switches to standby mode.
t To switch the balance on again: Select j on the display.
The balance starts in the application most recently used before the it was last switched off.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
20
Using the Balance
Secura, Quintix User Manual
5.2 Leveling the Balance
The balance must be leveled in order to enable it to provide accurate weighing results. Leveling the balance compensates for slant or unevenness at the place of installation by twisting the front leveling feet of the balance.
Leveling is always required if the balance is set up at a different location or is moved at its original location.
The leveling function is different for the various balance models: – Electronic level indicator (Secura® only) – Conventional level indicator (Quintix®)
Note: The balance needs to be re-leveled and then adjusted each time its setup
location is changed.
5.2.1 Leveling with Electronic Level Indicator (Secura®)
In Secura® models, the balance's position is checked with sensors. Instructions on the display guide the user when leveling the balance.
0.00000
Max 120 g d=0.00001 g
LEVEL
y When LEVEL appears in red at the top left of the display, you must level the
balance.
LEVEL
t To start the leveling function, select the LEVEL button.
Leveling
Please turn each foot in the direction of the arrow as shown.
y The balance position is shown as an animation of a level indicator on the display.
t Twist the front leveling feet of the balance in the specified direction.
A prompt appears every 2 minutes after confirmation of this “Leveling” status
message if the balance has not been leveled.
Leveling
The balance has been leveled.
y When the level indicator animation turns green, the balance is correctly leveled.
t To exit the function, select l.
t Complete the leveling by calibrating the balance.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
21
Using the Balance
Secura, Quintix User Manual
5.2.2 Leveling with a Conventional Level Indicator (Quintix®)
The Quintix® models are fitted with a conventional level indicator.
This is located to the left of the display. The position of the air bubble in the indicator shows whether or not the balance is correctly leveled.
t Twist both of the front leveling feet of the balance to move the air bubble so that
it is centered in the circular marking.
The figure shows which leveling foot should be twisted in which direction. Generally, both feet must be adjusted in order to position the air bubble exactly in the center of the circular marking.
Examples: – If the air bubble is too far to the lower right, twist the right leveling foot
counterclockwise (see figure).
– If the air bubble is too far to the upper right, twist the left leveling foot
clockwise.
– If the air bubble is too far to the left, twist the left leveling foot counterclockwise
and the right leveling foot clockwise.
t Check the level indicator after each adjustment and correct if necessary.
t Complete the leveling by calibrating the balance.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
22
Using the Balance
Secura, Quintix User Manual
5.3 Operating Concept
This section explains the operating options available to you so that you can familiarize yourself with the balance.
5.3.1 Operating and Display Elements
Sharp or pointed instruments (such as ballpoint pens) can damage the device!
– The touchscreen should only be operated by lightly pressing it using the tips of your
fingers. This can be done while wearing laboratory gloves.
1
+
Net
LEVEL
G
SQmin 3.05000 g
Component
41.97650
1
2
3 4
5
6
7
8
9
13
12
10
11
Max 120 g d=0.00001 g
1 Toolbar with currently available buttons: Zero, tare, GLP printout, print (data
output)
2 Metrology line (additional info for minimum capacity Min and the verification
scale interval e appear on verified balances)
3 Warning symbol to indicate calculated values, negative gross values, or values
that are smaller than the minimum sample quantity (SQmin) 4 Current weight or value measured 5 Unit and stability indicator; set weight unit and display accuracy (see page 41) 6 Run application, such as saving the weight value of a component 7 Settings for the selected application 8 Menu key: Switch to menu or function selection 9 SQmin minimum weighing (Secura® only; see page 43) 10 Bar graph: Scaled display showing capacity usage (in percent) 11 Preceding symbol (±) for weight value 12 Display value ID, e.g., “G” for gross value, “Net” for net value; switch display value
for application, such as net value, total value, weight value, or calculated value 13 Field for status and warning displays, and activation of isoCAL and leveling
Visit www.balances.com for the full line of Sartorius Laboratory Balances
23
Using the Balance
Secura, Quintix User Manual
5.3.2 Menu
All applications and settings for the balance can be accessed via the menu.
t To access the menu, select the menu key at the bottom left of the display.
y The menu appears.
ComponentsStatistics
MixingWeighing
1
5
23
7
6
8
4
1 Selected application (such as Mixing) 2 Application selection area: Symbols for all available applications
(see “List of Applications in the Menu” on page 25) 3 Function area 4 Back: Run the most recent application. 5 Setup menu: Access the balance system settings
(see “System Settings (Setup Menu)” on page 27). 6 Access the balance leveling function via electronic level indicator (Secura® only –
see “Leveling with Electronic Level Indicator (Secura®)” on page 20). 7 Access the calibration and adjustment options
(see “Calibration and Adjustment” on page 72). 8 Switch the balance to standby mode.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
24
Using the Balance
Secura, Quintix User Manual
5.3.3 Entering Figures on the Number Pad
g
Min
123
456
789
+/- 0 .
C
+
0000000
Numerical values can be entered for many applications or system settings (such as a
minimum value for the Checkweighing application). A number pad appears on the display, which is always used in the same way.
t To enter numbers, select the desired numbers one after the other. The entered
numbers appear at the top of the display.
t To correct the most recent number, select k.
The number is deleted. Then enter the correct number.
t To delete the entire value, select C. t To confirm the entry, select l.
5.3.4 Entering Text and Characters (Secura® only)
A keyboard will appear whenever you have to enter text or characters. The cursor is
located in the line above the keyboard.
t Select the individual characters simply by touching them.
y The entered text will appear in the input line.
t Use the Shift key to switch the keyboard display between uppercase and
lowercase letters.
t Use the 123 key to switch the keyboard display from letters to numbers.
t Use the ABC key to switch the keyboard display from numbers to letters.
t Use the backspace key to delete the last character.
t The l key ends the process and saves the characters entered.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
25
Using the Balance
Secura, Quintix User Manual
5.3.5 Accessing an Application in the Menu
The right-hand side of the display shows available applications for various weighing tasks which can be carried out with the balance.
You can scroll up and down through the menu to select the desired application.
ComponentsStatistics
Mixing
Weighing
PercentageDensity
t To scroll, place your finger on the display and drag it slowly either up or down.
y The symbols in the menu move in the corresponding direction.
ComponentsStatistics
Mixing
Weighing
PercentageDensity
t Select a symbol to start that particular application.
%
10.0
0.00000
Max 120 g d=0.00001 g
Reference
y The desired application appears on the display.
5.3.6 List of Applications in the Menu
The menu provides the following applications for various weighing tasks:
Weighing (see page 40)
This is the standard application that appears when the balance is first switched on. Use this application to determine the weight of a sample within the device's specific weighing range.
Mixing (see page 45)
Use this application to weigh up to 99 components one after the other for a mixture or formula in one container. The balance is automatically tared after each component is weighed. The weight value of an individual component or the total weight can be displayed as desired.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
26
Using the Balance
Secura, Quintix User Manual
Statistics (see page 48)
Use this application to save weight and calculated values and statistically analyze them. You can save up to 99 components.
Components (see page 51)
Totalize weight values. You can save up to 99 components that are weighed in various individual containers. Each container can be tared before each component is weighed.
Density (see page 54)
Use this application to determine the density of solid samples using a density set based on the buoyancy method. The density is determined using Archimedes' Principle. The upward buoyant force exerted on a body immersed in a fluid is equal to the weight of the fluid the body displaces.
Percentage (see page 58)
This application is used to determine the percentage share or the percentage difference of the sample related to a reference weight.
Conversion (see page 61)
Use this application to multiply the weight value by a user-defined factor. The selected factor is saved to protected memory.
Unstable condition (see page 63)
This application is used for moving samples (e.g., live animals) and for weighing in unstable environments. A measurement cycle is automatically carried out with a defined number of measurements for each object to be weighed. The individual measurements are averaged, and this average is displayed as the result.
Checkweighing (see page 65)
Use this application to check whether a weight value falls within the specified tolerances. This application also makes it easy to fill sample materials to a specified target weight.
Peak hold (see page 67)
Use this application to calculate the maximum weight value of a sample (peak value). The value remains on the display for five seconds after the sample has been removed from the balance. Example: Use for measuring release force during an experiment, or use when a load is so big that the balance display is hidden from view during weighing.
Counting (see page 69)
Use this application to determine the number of parts of approximately equal weight. The weight of a counted reference sample is calculated and then the objects with an unknown piece count are weighed. The balance displays the number of parts and the piece weight.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
27
System Settings (Setup Menu)
Secura, Quintix User Manual
6. System Settings (Setup Menu)
The Settings (Setup) menu contains all basic settings for the balance.
On balances verified for use in legal metrology, not all functions and settings are
available.
6.1 Accessing the Menu and Changing Settings
t Select the menu key in any application.
ComponentsStatistics
MixingWeighing
y The menu appears.
t To access the system settings of the balance, select i (Setup) in the menu.
Settings
Language
Date and time
Device information
Calibration/Adjustment
y The Settings (Setup) menu appears.
Settings
Signals
Date and time
Device information
Calibration/Adjustment
Printout
t To scroll, place your finger on the display and drag it slowly either up or down.
y The settings in the menu move in the corresponding direction.
While you are scrolling, a gray scroll bar appears on the right of the display,
indicating where you are in the list of options.
t Select a setting and make the desired changes. Information about the available
settings can be found from page 28.
t Select l to confirm your changes.
t Select ; to return to the menu. The changed settings are active when you reach
the first level of the menu again.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
28
System Settings (Setup Menu)
Secura, Quintix User Manual
6.2 List of Available Settings
This section contains information on all basic balance settings which can be defined in the Settings menu.
6.2.1 Language
Language
Deutsch
Français
English
The language of the display text can be set here. English is the default language for
the balance.
The following languages are available: – English – German – French – Spanish – Italian – Japanese – Russian – Chinese – Polish – Portuguese – Korean – Turkish – Hungarian
Setting the Language
The default language for the display text is English. To change the language, proceed as follows:
t Select the menu key at the bottom left of the weighing screen on the display.
ComponentsStatistics
MixingWeighing
y The menu is displayed.
t Select the j (Setup) key in the menu.
Settings
Language
Date and time
Device information
Calibration/Adjustment
y The Settings window appears.
t Select Language or the top entry in the list.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
29
System Settings (Setup Menu)
Secura, Quintix User Manual
Language
Deutsch
Français
English
y The Language settings window appears.
t Touch the language you would like to select, such as English.
t Touch l to confirm.
y The display changes immediately to the desired language.
t Select ; to return to the menu.
6.2.2 Date and Time
Date and time
Date
Date format
Time
Time format
The date, time, and display format can be set here.
Under Date:
t To set the date, select t, enter the current date and confirm with l.
Under Date format:
t Select how the date should be displayed and printed out.
DD-MMM-YYYY:
The date is displayed in the order of day, month and then year.
MMM-DD-YYYY:
The date is displayed in the order of month, day and then year.
YYYY-MM-DD (ISO):
The date is displayed in the order of year, month and then day. (The time is displayed in 24-hour format with this setting.)
Under Time:
t To set the time, select t, enter the current time and confirm with l.
Under Time format:
t Select how the time is displayed and printed out.
24h:
The time is displayed in 24-hour format. (This is the only available setting when the date is set to ISO format.)
12h (AM/PM)
The time is displayed in 12-hour format. The hours before noon are displayed with AM and the hours after noon are displayed with PM.
Visit www.balances.com for the full line of Sartorius Laboratory Balances
30
System Settings (Setup Menu)
Secura, Quintix User Manual
6.2.3 Device Information
Device information
Manufacturer: Sartorius Model: SECURA225D Serial number: 0012345678 Version BAC: 00-50-05.xx CN:15B2 Version APC: 01-71-02.xx CN:8152
The manufacturer, model, serial number and software version of the balance are
displayed here.
6.2.4 Calibration/Adjustment
Calibration/Adjustment
isoCAL
Calibration report
You can configure how the automatic calibration/adjustment function isoCAL starts. You can view the saved reports on recently conducted adjustments, including the
deviations detected, and send the output to a PC or a Sartorius laboratory printer.
Under isoCAL:
t Select the desired start option for isoCAL.
Off: isoCAL is switched off. – Info, manual start: A note appears on the display when predefined time
intervals or temperature values are exceeded. isoCAL can be started manually.
Automatic: isoCAL is switched on. Calibration/adjustment starts
automatically when predefined time intervals or temperature values are exceeded.
Calibration report
2014-10-04: 2 Protocols
2014-10-01: 2 Protocols
2014-09-25: 3 Protocols
2014-09-24: 3 Protocols
Under Calibration report:
t Select the desired report by date.
If there is more than one calibration report per date, use the < and > buttons to navigate through the reports.
t Select a to print the displayed report.
Up to 99 entries per day are saved.
Calibration reports older than 30 days are deleted.
6.2.5 Weighing
Weighing
SQmin
Application
Ambient conditions
Safety level
A number of basic settings for the weighing functions can be defined here. Options
are limited for verified balances. All settings given in the menu are permitted.
Under SQmin (Secura® only): If the SQmin function is set up on the balance, the function can be switched On or
Off here (see “SQmin Minimum Weighing (Secura® only)” on page 43).
Visit www.balances.com for the full line of Sartorius Laboratory Balances
Loading...
+ 82 hidden pages