Sartorius Secura 124, Quintix, Secura 613, Secura 513, Secura 313 User Manual

...
User Manual
Secura®, Quintix®, Practum
®
Laboratory Balances
2
Contents
User Manual for Secura, Quintix, Practum
Contents
Contents ..................................... 2
User Information .............................. 3
Safety Instructions ............................. 5
Intended Use ................................. 6
Getting Started ............................... 8
Unpacking and Equipment Supplied ................ 8
Installing the Balance ........................... 9
Choosing a Location ............................ 11
Electrical Connection ............................ 12
Warm-up Time ................................ 13
Leveling ..................................... 14
Below-balance Weighing ......................... 16
Anti-theft Locking Device ........................ 17
Moving the Balance ............................ 18
Using the Balance ............................. 19
Switching the Balance On/Off ..................... 19
Setting the Language ........................... 20
Operating Concept ............................. 21
System Settings (Setup Menu) .................... 26
Weighing .................................... 38
Mass Unit Conversion ........................... 39
SQmin Minimum Weighing (Secura
®
only) ............ 41
Individual Identifiers (Secura® only) ................. 42
Mixing (Secura® and Quintix® only) ................. 43
Statistics (Secura® and Quintix® only) ................ 46
Totalizing Components (Secura® and Quintix® only) ..... 49
Density ...................................... 52
Percentage ................................... 56
Conversion ................................... 59
Unstable Condition ............................. 61
Checkweighing ................................ 63
Peak Hold .................................... 65
Counting .................................... 67
Calibration and Adjustment ...................... 70
ISO/GLP-compliant Printout ...................... 74
USB Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Communication with Peripheral Devices .............. 76
Direct Transfer of Data (PC) ....................... 77
Connection with Sartorius Lab Printers YDP30, YDP40 . . . 82
Connection for RS-232 Configuration ............... 82
Status Messages ............................... 88
Transporting the Balance ........................ 90
Care and Maintenance .......................... 93
Cleaning the Balance ............................ 93
Disposal ..................................... 98
Specifications ................................. 99
General Data .................................. 99
Model-specific Data ............................100
Accessories ...................................104
Balance Dimensions ............................105
EC Declaration of Conformity .....................109
EC Type-approval Certificate ......................110
Plates and Markings ............................111
3
User Information
User Manual for Secura, Quintix, Practum
User Information
Warning/Danger Symbols
Warning/danger symbols used in these instructions:
This symbol identifies hazards that have a high probability of resulting in death
or serious physical injury if not avoided.
This symbol identifies hazards that can result in moderate or mild injuries
if not avoided.
This symbol identifies hazards associated with the risk of material damage.
Symbols
The following symbols are used in this manual:
Useful information and tips
Notes for use in legal metrology
Note regarding device operation
When individual buttons are displayed, they should be pressed.
4
User Information
User Manual for Secura, Quintix, Practum
The following symbols are used in these instructions:
t Indicates a required action
y Describes what happens after you have performed a particular step
Perform steps in the specified order:
1. First action
2. Second action
3. … – Indicates an item in a list
Conventions for this user manual: – The figures in this manual are based on “standard” balances.
Some displays and printouts shown in the figures may differ from those of verified balances. The differences are explained in detail where they affect the operation of the balances.
Application Advice/Technical Support
Contact addresses for application advice and technical support can be found online at http://www.sartorius.com
5
Safety Information
User Manual for Secura, Quintix, Practum
Safety Information
Guidelines and General Information
– The balance complies with EU Directives and standards for electrical safety and
electromagnetic compatibility.* Improper use or handling can, however, result in damage and/or injury. Any improper use or operation of the balance, i.e., that is not consistent with the instructions, will result in forfeiture of all claims under the manufacturer's warranty.
– Personnel need to have read and understood these installation instructions,
including the safety instructions.
– In the event of use in systems and ambient conditions which have greater
safety requirements, you must observe the requirements and provisions applicable in your country.
– Always keep the equipment and balance freely accessible.
Any installation work or balance usage that does not conform to the instructions in this manual results in forfeiture of all claims under the manufacturer’s warranty.
* = See “Specifications.”
Danger of explosion!
Do not use this equipment in hazardous areas in which explosive materials are present.
Make sure that the voltage rating printed on the AC adaptor is identical to your local mains voltage.
Installation Information
Do not operate the balance if its housing or AC adaptor including any of the
connections are damaged. Immediately disconnect the damaged equipment from the power by pulling the plug to turn off the power.
Do not expose the balance and the accessories supplied by Sartorius to extreme temperatures, aggressive chemical vapors, moisture, shocks, vibration, or strong electromagnetic fields. Observe the conditions of operation described in the Specifications!
Installation note: The operator shall be solely responsible for any modifications to the equipment and for connecting any cables or equipment not supplied by Sartorius. Information on operational quality is available upon request from Sartorius. Only use original Sartorius accessories.
Do not compromise the IP protection of the balance or of the AC adaptor.
Do not allow liquid penetration. The protection class indicates the suitability of devices for different ambient conditions (humidity, foreign bodies).
Before cleaning the AC adaptor or balance, unplug the AC adaptor.
The balance may only be opened by specialized personnel trained by Sartorius.
Do not open the AC adaptor.
6
Safety Information
User Manual for Secura, Quintix, Practum
Hazards during Installation and Operation
If glass breaks, there is a risk of injury posed by cuts on glass edges.
Lay the cables where they pose no risk of causing someone to trip.
Observe the additional safety and danger information in the following chapters.
Intended Use
This high-resolution balance is designed and intended for use only in laboratories and indoor areas under normal atmospheric conditions. It was developed specifically for the exact determination of the mass of materials in liquid, paste, powder, or solid form.
Appropriate containers must be used for each type of sample material.
7
General View of the Equipment
User Manual for Secura, Quintix, Practum
General View of the Equipment
Pos. Name and Function Pos. Name and Function
1 Weighing pan 8 Fastening point for an optional “Kensington” key lock
2 Shield disk 9 Menu access switch: locks various functions and units
when balances are verified
3 Leveling feet 10 USB port for mini AB
4 Touch screen 11 DC jack
5 Level indicator (Quintix
®
and Practum® models) 12 Secura® models with readability of 0.1mg: Pan draft
shield
6 Manufacturer's ID label and additional plate for
verified models
13 Draft shield
7 Lug for attaching anti-theft locking device
6
7
89
10
11
3
4
5
12
1
13
1
2
3
4
5
8
Getting Started
User Manual for Secura, Quintix, Practum
Getting Started
The balance must be disconnected from the mains power for all assembly work.
Unpacking and Equipment Supplied
t Open the packaging, making sure to remove all parts carefully.
t After unpacking the device, check it immediately for any external damage.
y If you detect any damage, proceed as directed in the chapter entitled “Care and
maintenance.”
y Save the box and all parts of the packaging for any future transport. During
shipment, please do not leave cables plugged in!
The following parts are included in the equipment supplied:
Components Balance with Draft Shield Precision Balances
Readability: 0.1 mg 1 mg 10 mg, 0.1 g, and 1 g
Draft shield
99
Weighing pan:
d 90 mm
9
––
Weighing pan:
d 120 mm
9
Weighing pan:
d 180 mm
9
Pan support
99 9
Shield disk (precision balances)
9
Pan draft shield (Secura® only)
9
––
AC adaptor with country-specific power plug
9
9
9
In-use dust cover
99 9
Dust cover
99
Installation and operating instructions
 9
 9
 9
Application guide
99 9
9
Getting Started
User Manual for Secura, Quintix, Practum
Installing the Balance
Balance with Draft Shield
t Install the following parts in the order listed:
– In-use dust cover
– Turn the draft shield lock to “unlocked” .
– Put the draft shield (1) on the balance.
Note:
Use both hands to hold the top struts of the draft shield.
Move the side doors right to the back.
– Turn the draft shield lock to “locked” .
– Put the shield plate (2), shield disk (3), pan support (4), weighing pan (5) and
pan draft shield* (6) on the balance.
Note:
For information about how to clean the draft shield, see the Care and Maintenance chapter.
* = for models with readability of 0.1mg only
Secura® Models:
5 4 2
1
Quintix® and Practum® Models:
10
Getting Started
User Manual for Secura, Quintix, Practum
Inserting (or Removing) Glass Panels
1) Holding it by its rear edge, pull the draft shield cover up and remove it.
2) Push the three glass panels into the draft shield guide, all the way to the front.
3) Then replace the draft shield cover. In doing so, the draft shield cover must lock into position.
Balance with No Draft Shield
t Install the following parts in the given order:
1)
In-use dust cover (not shown)
2) Install the shield disk (1) and turn this counterclockwise until securely in place.
3) Install the pan support (1) and weighing pan (2).
11
Getting Started
User Manual for Secura, Quintix, Practum
Choosing a Location
t Select the right setup location:
– Set up the device on a stable, even surface that is not exposed to vibrations
(e.g., weighing stone).
– Maintain free access to the device at all times.
Choose a location that is not subject to the following negative influences: – Heat (heater or direct sunlight) – Drafts from open windows, AC systems, and doors – Vibrations during weighing – Heavy traffic areas (personnel) – Excessive moisture – Electromagnetic fields
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when it is brought into a warmer area. To avoid the effects of condensation, condition the balance for about two hours, leaving it unplugged from the power supply, before plugging the balance back into the mains.
°C
12
Getting Started
User Manual for Secura, Quintix, Practum
Power Connection (Power Supply)
AC Adaptor Assembly
Using the wrong power plug adaptors may cause fatal electric shock and damage the equipment. Never insert the power plug adaptor separately from the AC adaptor in the socket (danger of electric shock).
t Use the right power plug adaptor for your mains power supply.
A AC adaptor versions: Bag Region/Country Bag Region/Country
a) transparent Europe/EU (except UK) e) black India b) yellow United Kingdom f) turquoise South Africa c) blue USA, Canada, China and Japan g) green Korea d) red Australia h) pink Brazil
i) white Argentina
The packaging of AC adaptor module 6971790 with line adaptors g), h), and i) is labeled with 98671791.
t Push (1) and slide (2) the power plug adaptor (A) required for the power supply
into the opening of the AC adaptor module (B). When doing this, the AC adaptor needs to lock into position.
13
Getting Started
User Manual for Secura, Quintix, Practum
Removing/Replacing the AC Adaptor
t Unlock (1) the AC adaptor and then detach (2) it.
Power Connection/Safety Precautions
– Only use original Sartorius AC adaptors. The AC adaptor has an IP rating of
IP40 in accordance with EN 60529.
– Make sure that the voltage rating printed on the manufacturer's ID label is
identical to that of your local mains voltage.
– If the stated supply voltage or the plug design of the AC adaptor does not
comply with your country's standard, please inform the nearest Sartorius representative.
– The power must be connected in accordance with the regulations applicable
inyour country.
t 1) First of all, plug the AC adaptor into the socket at the back of your balance.
t 2) Afterwards, plug the AC adaptor into an electrical outlet (mains power supply).
Warm-up Time
y To ensure accurate results are delivered, the balance must warm up for at least
30 minutes after initial connection to the power supply. Only after this time will the device have reached the required operating temperature.
When a verified balance of accuracy class k for use in legal metrology is connected to the mains power, it must warm up for at least one hour before operation.
0
30
14
Getting Started
User Manual for Secura, Quintix, Practum
Leveling
The balance must be level to ensure precise weighing results. Leveling the balance compensates for slant or unevenness at the place of installation by twisting the front leveling feet of the balance.
The balance needs to be re-leveled and adjusted each time its setup location is changed.
The leveling function is different for the various balance models: – Electronic level indicator (Secura
®
only)
– Conventional level indicator (Quintix® and Practum®)
Note: The balance needs to be re-leveled and then adjusted each time its location is changed.
Leveling with Electronic Level Indicator (Secura®)
In Secura® models, the balance's position is checked with sensors. Instructions on the display guide the user when leveling the balance.
0.00
Max 2100 g d=0.01 g
LEVEL
y When LEVEL appears in red at the top left of the display, you must level the
balance.
LEVEL
t To start the leveling function, select the LEVEL button.
Leveling
Please turn each foot in the direction of the arrow as shown.
y The balance position is shown as an animation of a level indicator on the
display.
t Twist the front leveling feet of the balance in the specified direction.
A prompt appears every 2 minutes after confirmation of this „Leveling“ status message if the balance has not been leveled.
Leveling
The balance has been leveled.
y When the level indicator animation turns green, the balance is correctly leveled.
t To exit the function, select l.
t Complete the leveling by calibrating the balance.
15
Getting Started
User Manual for Secura, Quintix, Practum
Leveling with a Conventional Level Indicator
(Quintix
®
and Practum®)
L
R
R
L
L
R
The Quintix® and Practum® models are fitted with a conventional level indicator.
This is located to the left of the display. The position of the air bubble in the indicator shows whether or not the balance is correctly leveled.
t Twist both of the front leveling feet of the balance to move the air bubble so
that it is centered in the circular marking.
The figure shows which leveling foot should be twisted in which direction. Generally, both feet must be adjusted in order to position the air bubble exactly in the center of the circular marking.
Examples: – If the air bubble is too far to the lower right, twist the right leveling foot
counterclockwise (see figure).
– If the air bubble is too far to the upper right, twist the left leveling foot
clockwise.
– If the air bubble is too far to the left, twist the left leveling foot
counterclockwise and the right leveling foot clockwise.
t Check the level indicator after each adjustment and correct if necessary.
t Complete the leveling by calibrating the balance.
16
Getting Started
User Manual for Secura, Quintix, Practum
Below-balance Weighing
A port for a below-balance weighing hook is located on the bottom of the balance.
Install a draft protection shield.
t Lift cover plate out of the bottom of the balance.
When doing this, put the balance to one side and not on the weighing pan side. Placing the balance on the weighing pan side may lead to measurement inaccuracies and damage to the balance!
t Attach a wire (or similar) to the sample and hang it on the notched hook.
If the below-balance weighing port is no longer being used, close the opening with the cover plate again, so that the weighing results are not adversely affected by drafts.
The below-balance weighing port may not be opened or used on balances used for legal metrology.
17
Getting Started
User Manual for Secura, Quintix, Practum
Anti-theft Locking Device
t A “Kensington” key lock cable can be installed at the fastening point on the
back of the balance if required.
t The balance can also be secured in place at the installation location with a
chain or lock, for example.
18
Moving the Balance
User Manual for Secura, Quintix, Practum
Moving the Balance
Moving the Balance over Short Distances in the Lab
Avoid glass breakage, shocks, and vibrations: Never lift or carry the balance when holding it by its draft shield!
t Hold the balance under the housing, lift it up carefully, and carry it to its new
location.
The balance needs to be re-leveled and calibrated each time its setup location is changed.
19
Using the Balance
User Manual for Secura, Quintix, Practum
Using the Balance
Switching the Balance On/Off (Standby)
0.00
Max 2100 g d=0.01 g
y The Leveling screen is displayed first after any model is switched on.
Level the balance if necessary. The actual weighing screen is only displayed after confirmation.
– When the balance is switched on, the following message appears on the display:
Secura®: Level The balance has been leveled (if leveled). Quintix® and Practum®: Level Please check leveling.
– If the balance needs to be leveled, a corresponding message appears on the
display (see “Leveling the Balance” on page14).
– The balance must be regularly calibrated and adjusted to ensure correct
weighing results (see “Calibration and Adjustment” on page70).
– The display text is given in English as a default. The language can be changed
(see “Setting the Language” on page20).
t To switch the balance to standby mode, select the Menu key at the bottom left
of the display on the Weighing screen.
ComponentsStatistics
MixingWeighing
y The menu is displayed.
t Select the j button in the Application menu.
y The balance switches to standby mode.
t To switch the balance back on, select j on the display.
The balance starts in the application that was most recently used before it was switched off.
20
Using the Balance
User Manual for Secura, Quintix, Practum
Setting the Language
The display text is given in English as a default. To change the language, proceed as follows:
t In the Weighing screen, select the Menu key at the bottom left of the display.
ComponentsStatistics
MixingWeighing
y The menu is displayed.
t Select the i button (Setup) in the menu.
Settings
Language
Date and time
Device information
Calibration/Adjustment
y The Settings window appears.
t Select Language or the top entry in the list.
Language
Deutsch
Français
English
y The Language settings window appears.
t Select the desired language, such as Deutsch (German).
t Select l to confirm.
y The display immediately switches to the desired language.
t Select ; to return to the menu.
21
Using the Balance
User Manual for Secura, Quintix, Practum
Operating Concept
This section describes the operational possibilities of your balance.
Operating and Display Elements on the Display
Sharp or pointed instruments (such as ballpoint pens) can damage the device.
– The touch screen should only be operated by lightly touching it with your
fingertips. This can be done while wearing laboratory gloves.
Max 220 g d=0.0001 g
41.9765
Component
1
+
Net
LEVEL
G
SQmin 3.0500 g
1
2
3 4
5
6
7
8
9
13
12
10
11
1 Toolbar with currently available buttons:
Zero, Tare, and possibly Print (data output) and GLP printout
2 Metrology line (additional info for minimum capacity Min and the verification
scale interval e appear on verified balances)
3 Warning symbol for calculated values, negative gross values, or values that are
smaller than the minimum initial weight (SQmin) 4 Current measurement value 5 Unit and stability indicator; set weight unit and display accuracy (see
page39) 6 Run application, such as saving the weight value of a component 7 Settings for the selected application 8 Menu key: switch to menu and functional selection 9 SQmin minimum weighing (Secura® only; see page41) 10 Bar graph: scaled measurement value display (percentage weighing capacity) 11 Preceding symbol (±) for weight value 12 Display value ID, e.g. “G” for gross value, “Net” for net value; switch display
value for application, such as net value, total value, weight value, or calculated
value 13 Field for status and warning displays, and activation of isoCAL and leveling
22
Using the Balance
User Manual for Secura, Quintix, Practum
The Menu
All balance application and settings can be accessed via the balance menu.
t To access the menu, select the Menu key located at the bottom left of the
display in any application.
y The menu appears.
ComponentsStatistics
MixingWeighing
1
5
23
7
6
8
4
1 Selected application (such as Mixing) 2 Application selection area: symbols for all available applications
(see “List of Applications in the Menu” on page 24) 3 Function area 4 Back: run the most recent application. 5 Setup menu: access the balance system settings
(see “System Settings (Setup Menu)” on page26). 6 Access the balance leveling function via electronic level indicator (Secura® only)
(see “Leveling with Electronic Level Indicator (Secura®)” on page14). 7 Access the calibration and adjustment options
(see “Calibration and Adjustment” on page70). 8 Switch the balance to standby mode.
23
Using the Balance
User Manual for Secura, Quintix, Practum
Enter Figures on the Keypad
g
Min
123
456
789
+/- 0 .
C
+
0000000
Numerical values can be entered for many applications or system settings (such as
a minimum value for the Checkweighing application). A keypad appears on the display, which can be used for this purpose.
t To enter numbers or values, touch the corresponding numbers in order.
Theentered numbers appear at the top of the display.
t To correct the most recent number, selectk.
The number is deleted. Then enter the correct number.
t To delete the entire value, select C. t To confirm the entry, selectl.
Text and Character Input (Secura® only)
A keyboard will appear whenever you have to enter text or characters. The cursor is
located in the line above the keyboard.
t Select the desired character simply by touching them.
y The entered text will appear in the input line.
t Press the shift button to change between uppercase and lowercase letters.
t Press the 123 key to switch from letters to numbers.
t Press the ABC key to switch from numbers to letters.
t The backspace key is used to delete the character to the right.
t The l key ends the process and saves the characters entered.
24
Using the Balance
User Manual for Secura, Quintix, Practum
Accessing an Application in the Menu
The right-hand side of the display shows available applications for various weighing tasks which can be carried out with the balance.
You can scroll up and down through the menu to select the desired application.
ComponentsStatistics
Mixing
Weighing
PercentageDensity
t To scroll through the menu, place a finger on the display and drag it slowly up
or down. y The symbols in the menu move in the corresponding direction.
ComponentsStatistics
Mixing
Weighing
PercentageDensity
t Select a symbol to start that particular application.
Max 2100 g d=0.01 g
0.00
Reference
%
10.0
y The desired application appears on the display.
List of Applications in the Menu
The menu provides the following applications for various weighing tasks:
Weighing (see page38)
This is the standard application that appears when the balance is first switched on. Use this application to determine the weight of a sample within the device's specific weighing range.
Mixing (Secura® and Quintix® only; see page43)
Use this application to weigh up to 99 components one after the other for a mixture or formula in one container. The balance is automatically tared after each component is weighed. The weight value of an individual component or the total weight can be displayed as desired.
25
Using the Balance
User Manual for Secura, Quintix, Practum
Statistics (Secura® and Quintix® only; see page46)
Use this application to save weight and calculated values and statistically analyze them. You can save up to 99 components.
Components (Secura® and Quintix® only; see page49)
Totalize weight values. You can save up to 99 components that are weighed in various individual containers. Each container can be tared before each component is weighed.
Density (see page52)
Use this application to determine the density of solid samples using a density set based on the buoyancy method. The density is determined using Archimedes' Principle. The upward buoyant force exerted on a body immersed in a fluid is equal to the weight of the fluid the body displaces.
Percentage (see page56)
This application is used to determine the percentage share or the percentage difference of the sample related to a reference weight.
Conversion (Secura® and Quintix® only; see page59)
Use this application to multiply the weight value by a user-defined factor. The selected factor is saved to protected memory.
Unstable condition (see page61)
Use this application with moving samples (such as live animals) and for weighing in unstable environments. A measurement cycle is automatically carried out with a defined number of measurements for each object to be weighed. The individual measurements are averaged, and this average is displayed as the result.
Checkweighing (see page63)
Use this application to check whether a weight value falls within the specified tolerances. This application also makes it easy to fill sample materials to a specified target weight.
Peak hold (see page65)
Use this application to calculate the maximum weight value of a sample (peak value). The value remains on the display for five seconds after the sample has been removed from the balance. Example: Use for measuring release force during an experiment, or use when a load is so big that the balance display is hidden from view during weighing.
Counting (see page67)
Use this application to determine the number of parts of approximately equal weight. The weight of a counted reference sample is calculated and then the objects with an unknown piece count are weighed. The balance displays the number of parts and the piece weight.
26
System Settings (Setup Menu)
User Manual for Secura, Quintix, Practum
System Settings (Setup Menu)
The Settings (Setup menu) contain all basic settings for the balance.
Not all functions/settings are available on verified balances.
Accessing the Menu and Changing Settings
t Select the Menu key in any application.
ComponentsStatistics
MixingWeighing
y The menu appears.
t To access the balance system settings, select i (Setup) in the menu.
Settings
Language
Date and time
Device information
Calibration/Adjustment
y The Settings (Setup menu) appear.
Settings
Printout
Date and time
Device information
Calibration/Adjustment
Weighing
t To scroll through the menu, place a finger on the display and drag it slowly up
or down. y The settings in the menu move in the corresponding direction.
A gray scrollbar appears on the right of the display when scrolling, indicating
the current location in the list of options.
t Select a setting and make the desired changes. Information about available
settings can be found from page27.
t Select l to confirm your changes.
t Select ; to return to the menu. The changed settings become active when you
return to the first level of the menu.
27
System Settings (Setup Menu)
User Manual for Secura, Quintix, Practum
List of Available Settings
This section contains information on all basic balance settings which can be defined in the Settings menu.
Language
Language
Deutsch
Français
English
Set the display language here. English is the default language (see “Setting the
Language” on page20).
The following languages are available: – English – Deutsch (German) – Français (French) – Español (Spanish) – Italiano (Italian) – 日本語 (Japanese) – Pусский (Russian) – ㅩ昝/ 㖶措 (Chinese) – Polski (Polish) – Português (Portuguese) – ዽ૑࿌/ Ⴎ໓൥ (Korean) – Turkish – Hungarian
Date and Time (Secura® and Quintix® only)
Date and time
Date
Date format
Time
Time format
Set the date, time, and display format here.
Date menu option:
t To set the date, select t, enter the current date, and confirm withl.
Date format menu option:
t Choose how the date is displayed and printed.
DD-MMM-YYYY:
Displayed in order of day, month, and then year.
MMM-DD-YYYY:
Displayed in order of month, day, and then year.
YYYY-MM-DD (ISO):
Displayed in order of year, month, and then day. (The time is given in 24-hour format when this setting is selected.)
Time menu option:
t To set the time, select t, enter the current time, and confirm withl.
Time format menu option:
t Choose how the time is displayed and printed.
24h:
The time is displayed in 24-hour format. (This is the only possible setting when the date format is set to ISO.)
12h (AM/PM)
The time is displayed in 12-hour format. Times shown before noon are marked with AM, and times shown after noon are marked with PM.
28
System Settings (Setup Menu)
User Manual for Secura, Quintix, Practum
Device Information
Device information
Manufacturer: Sartorius Model: SECURA2102 Serial number: 0012345678 Version BAC: 00-50-02.xx CN:1701 Version APC: 01-70-02.xx CN:6587
The manufacturer, model, serial number, and software version of your balance are
displayed here.
Calibration/Adjustment
Calibration/Adjustment
isoCAL
Calibration report
With Secura® models, it is possible to set how the automatic calibration/adjustment
function isoCAL is started.
With Secura® and Quintix® models, saved reports about the most recent calibration
procedures, including deviations identified, can be viewed and output to a PC or Sartorius lab printer.
isoCAL menu option (Secura
®
only):
t Select the desired start option for isoCAL.
Off: isoCAL is switched off.
Info, manual start: A note appears on the display when predefined time
intervals or temperature values are exceeded. isoCAL can be started manually.
Automatic: isoCAL is switched on. Calibration/adjustment starts
automatically when predefined time intervals or temperature values are exceeded.
Calibration report menu option (Secura® and Quintix® only):
t Select the desired report by date.
If there is more than one calibration report per date, use the < and > buttons to
navigate through the reports.
t Select a to print the displayed report.
A maximum of 99 entries are saved per day. Calibration reports are deleted after 30 days.
Weighing
Weighing
SQmin
Application
Ambient conditions
Safety level
A number of basic settings for the weighing functions can be defined here. Options
are limited for verified balances. All settings given in the menu are permitted.
SQmin menu option (Secura
®
only):
If the SQmin function is set up on the balance, the function can be switched On or
Off here (see “SQmin Minimum Weighing (Secura
®
only)” on page41).
Calibration report
2012-10-04: 2 Protocols
2012-10-01: 2 Protocols
2012-09-25: 3 Protocols
2012-09-24: 3 Protocols
29
System Settings (Setup Menu)
User Manual for Secura, Quintix, Practum
Safety Level
The Secura® models monitor the following at all times: – whether the balance is leveled correctly – whether adjustment is necessary – whether minimum sample quantity requirements are met according to USP.
Safety level menu option (Secura
®
only):
t Select the desired safety level for SQmin, isoCAL, and LEVEL
If a verified balance has not been leveled, the following functions will be disabled
immediately (as is the case with the “High” safety level): – Printout – Startup and saving of application Here, it does not matter what settings this menu item has.
High (factory setting): If one of the three conditions has not been met, a warning is
displayed straight away. No data is transmitted.
The following functions are disabled immediately:
– Printout – Startup and saving in applications Displays: – Information about red marking of display element (LEVEL, isoCAL or SQmin) – Weight value displayed in gray with a black warning symbol A warning is displayed after 60 seconds in the event of non-observance.
Standard: If one of the three conditions has not been met, a warning is displayed
straight away. Displays: – Information about red marking of display element (LEVEL, isoCAL or SQmin) – Weight value displayed in gray with a black warning symbol A warning is displayed after 60 seconds in the event of non-observance. Printout: – Weight values are marked with “!”
Low: The following information is displayed:
– SQmin value has not been achieved -> SQmin value is marked red – Balance not leveled -> Red marking on “LEVEL” display element – Balance not adjusted due to temperature or time -> Red marking on “isoCAL”
display element
Weighing
SQmin
Application
Ambient conditions
Safety level
30
System Settings (Setup Menu)
User Manual for Secura, Quintix, Practum
Ambient conditions menu option:
t Choose whether the conditions at the balance location are Stable or Unstable.
If Unstable is selected, adverse ambient conditions (drafts, vibrations) are
filtered by changing the measurement time for the weight values.
Application menu option:
t Choose whether the balance will be used for Weighing or Dosing.
This setting is used to compensate for load fluctuations on the display.
Weighing: The display responds very quickly to fast load changes.
Thedisplay responds more slowly to small changes in weight (in the digit range). This setting is suitable for normal weighing.
Dosing: The display responds quickly to small changes in weight, making
itfaster and more accurate for additional dosing and container filling functions.
Stability signal menu option:
t Choose whether stability is shown with High accuracy, Medium accuracy,
orFast on the display.
The balance stability is displayed as soon as the weighing result is constant
within a defined range. Until stability is reached, the measured values is shown
in gray on the display and only turns black once the balance is deemed stable.
Zero / Tare menu option:
t Define settings for zeroing and taring.
Zero/Tare function: Choose whether the balance is zeroed and tared
With stability or Without stability. If Without stability is selected, the
balance will automatically be tared when A or D is selected. If With stability is selected, the balance is tared the next time stability is reached after A or D is selected.
Automatic zero (zero tracking function): When this option is active,
changes of a set fraction of scale intervals per second from the display zero point are zeroed automatically.
Zero/Tare at power on: When this option is active, the balance is
automatically zeroed or tared when switched on.