Sartorius Sartobind IEX MA 15, Sartobind IEX MA 75, Sartobind IEX MA 100, Sartobind IEX MA Series Operating Instructions Manual

Operating Instructions
Sartobind
®
IEX MA 15 | 75 |100
A Separation Technology Based on Macroporous Membranes
85032-539-13
Read operational instructions carefully before using Sartobind capsules.
!
Important
Use of the product in applications not specified or not described in this manual, may result in improper function, personal injury, or damage of the product or material.
Membrane Adsorber (MA) units should be visually inspected before use.
The units are supplied as non-sterile.
The membrane is dried from glycerol and should be prewashed with equilibration buffer before use.
If you plan to scale up later, preferentially use Sartobind capsules with 4 or 8 mm bed height: e.g. Sartobind Q nano with 1 or 3 ml as downscale.
2
Intended use
The products are intended for ion exchange (IEX) chromatography work in a laboratory for research purposes only.
Sartobind MA 15 units with 3 membrane layers can be used for screening of buffer conditions and quantitative purifications at exceptional high flow rate.
Sartobind MA 75 units contain 15 membrane layers and have been developed for bind and elute and flowthrough applications at high flow rate.
Sartobind MA 100 units with 5 membrane layers can be used for screening of buffer conditions and quantitative work at exceptional high flow rate.
3
4
1. Storage conditions 5
2. Introduction 5
3. Technical data 7
4. Materials and pH stability 8
5. Binding capacity 9
6. Installation 10
7. Operation 11
7.1 Venting 11
7.2 Recommended buffer volumes and flow rates 11
7.3 Buffer conditons 13
7.4
Selection of pH conditions
14
7.5 Contaminant removal from proteins in flow through operation 14
Table of Contents
7.6 Sample preparation and equilibration 15
7.7 Washing 16
7.8 Elution 16
7.9 Regeneration, cleaning and storage 17
7.10 Stability 18
7.11 Operation with peristaltic pump or LC system 19
8. Troubleshooting 20
9. Quality assurance 24
10. Ordering information 25
10.1 MA units 25
10.2 Accessories 26
11. Dimensions and connections 27
5
Sartobind MA units should be stored clean, dry and away from direct sunlight in the box at room temperature.
2. Introduction
Traditional chromatography uses porous particles packed into columns. As liquid flows through the column and around the beads, bio-molecules in the liquid diffuse into the pores of the beads to binding sites on the inner surface of the pores. The limiting factor in low pressure column chromatography is the time required for the molecules to diffuse into and out of the pores. The various steps of equilibration, loading, washing, elution and regeneration can take hours.
Sartobind membranes are made from regenerated and stabilized cel­lulose. The stabilization and cross-linking generate high chemical stability. Conventional ion exchange ligands are covalently attached to the macroporous membrane support. The membrane layers form chromatographic bed between 3 and 15 layers which is incorporated into housings.
1. Storage conditions
Pressure forces the liquid through the pores of the membrane, bringing target substances to direct contact with the binding sites. This direct convection to the binding sites minimizes diffusion limitation of mass transfer and increases the speed of operation to 10 – 30 fold compared to traditional column chromatography.
The Luer-Lok connectors on top and bottom allow you to run the units with a syringe and a prefilter or to connect several units to each other to get a larger bed volume or to run units with different chemistries.
6
7
MA 15 MA 75 MA 100
Membrane Area [cm2] 15 75 100 Number of layers 3 15 5 Bed height [mm] 0.8 4.0 1.4 Membrane volume [ml] 0.41 2.1 2.8 Membrane diameter [mm] 25 25 50 Typical dynamic binding 12 60 80
capacity 10% for Q [mg/unit]* Typical dynamic binding 10.5 52.5 70
capacity 10% for S [mg/unit]* Typical dynamic binding capacity 60
10% for D [mg/unit]** Ligand density [μeq/cm2] 2 – 5 (Q, S), 2 – 4 (D)** Maximum pressure at 20°C 6 bar | 0.6 MPa | 87 psi
* See also section 5. Binding capacity ** D type is available as MA 75 only.
3. Technical data
Housing and basic membrane Housing MA 15, 75, 100 Polysulfone Membrane matrix Stabilized reinforced cellulose,
nominal pore size >3 μm
Membrane ligands Quaternary ammonium (Q) Strong basic anion exchanger R-CH2–N+(CH3)
3
Strong acidic cation exchanger Sulfonic acid (S)
R-CH2–SO
3
Weak basic anion exchanger Diethylamine (D)
R-CH
2
–N(C2H5)
2
8
4. Materials
9
Data are based on dynamic binding capacity measurements 10% using MA15 run at 10 ml/min.
Membrane Typical dynamic binding Reference protein
capacity 10% (mg/cm2) and buffer
Q 0.8 BSA (bovine serum
albumin) in 20 mM Tris/HCl, pH 7.5
S 0.7 Lysozyme in 10 mM
potassium phosphate, pH 7.0
D 0.8 BSA (bovine serum
albumin) in 20 mM Bis Tris, pH 6.0
5. Binding capacityy
The contents of the packages are described in chapter 10.1.
The Sartobind Membrane Adsorber (MA) units are ready-to-use devices and can be used out of the box. They can be operated with a syringe (chapter 7.1 to 7.8), peristaltic pump or liquid chromatogra­phy (LC) system (follow chapter 7.11 first).
The unit should be installed in an upright position in the process flow. The female Luer connector is the inlet and the male Luer con­nection is the outlet of the unit. The flow is guided from top through the 3 to 15 membrane layer to the outlet. The units can also be connected in series to increase the binding capacity.
Fig. 1 Flow direction of buffer
10
6. Installation
11
7.1 Venting
It is important to remove air from the unit completely.
Fill a 10 – 20 ml Luer syringe with equilibration buffer and connect to Luer female inlet, then hold unit upright (outlet is up) and expel air until the first fluid is seen at the outlet. If you still detect any air in the filled unit, close it at the outlet with enclosed cap, flip the syringe and the MA so that the syringe is on top and the unit is on the bottom, hold the syringe up and move the plunger a few strokes up and down that air bubbles can ascend into the syringe. Very small air bubbles observed directly below the inlet do not disturb separa­tions. The unit will function normally as long as the small air bub­bles remain outside of the membrane bed.
7.2 Recommended buffer volumes and flow rates
Membranes are dried from glycerol to avoid shrinking. For pre-wash­ing with e.g. equilibration buffer or 1 M NaCl the following flushing volumes are recommended. Lower flow rates than the recommended ones can also be used.
7. Operation
Tab. 1: Buffer consumption of Sartobind MA units [ml]
MA 15 MA 75 MA 100
Equilibration 10 15 20 Wash 10 – 20 10 – 20 10 – 20 Elution 5 – 20 5 – 20 5 – 20 Regeneration 5 – 10 10 – 20 10 – 20
Tab. 2: Flow rate recommendation
MA 15 MA 75 MA 100
Recommended flow rate [ml/min] 10 – 30 10 – 30 20 – 50
12
13
7.3 Buffer conditions
In the majority of applications an equilibration buffer concentration of 10 mM provides sufficient buffering capacity and prevents the protein of interest from precipitation. The ionic strength should be kept as low as possible to avoid reduction of binding capacity. The buffer should have a pKa within 0.5 pH units of the pH used. It should be filtered with 0.2 μm filters before use and the quality of water and chemicals should be of high purity.
!
Important
Do not apply pure water through the units as it may lead to a reversible swelling and decrease of the flow rate of the membrane. The buffering ion should carry the same charge as the ion exchange ligand.
Binding | equilibration buffer examples
To bind BSA on Sartobind Q 20 mM Tris/HCl pH 7.5 To bind BSA on Sartobind D 20 mM Bis Tris pH 6.0 To bind lysozyme on Sartobind S 10 mM Potassium phosphate
buffer pH 7.0
7.4 Selection of pH conditions
In ion exchange chromatography a charged molecule is bound to oppositely charged groups attached to the insoluble matrix. This binding is reversible by application of salt ions to the elution buffer. The pH value at which a biomolecule has no net charge is the isoelectric point (pI). In buffers below the pI (at least 1 pH unit), a protein, for example, carries a positive net charge and will bind to a cation exchanger (Sartobind S). In buffers above its pI (at least 1 pH unit), it will bind to an anion exchanger (Sartobind Q or D).
7.5 Contaminant removal from proteins in flow through operation
For contaminant removal from proteins such as monoclonal anti­bodies, pH conditions in the range of pH 6 to 8 are used in order to bind highly negatively charged DNA, endotoxins, contaminating proteins, some host cell proteins and viruses. The product of interest, the monoclonal antibody with pI of 8 – 9.5 for example, will not bind and pass through Sartobind Q or D.
14
15
To remove contaminating proteins and aggregates with Sartobind S in flow through mode, process impurities have to be charged posi­tively to bind while the target protein stays negative. At the pH of the buffer above the pI, the protein product flows through without binding.
7.6 Sample preparation and equilibration
The sample should be adjusted to the starting buffer and be pre­filtered through a 0.2 μm membrane filter (not included). For small volumes in the ml range use a Minisart
®
with Luer outlet (order no.
16534-K).
Fig. 2: Inline filtration with a Minisart coupled to the inlet of Sartobind MA unit
Sartobind
Minisart
L
L
The Minisart shall be coupled directly to inlet of the MA unit during use as shown in Fig. 2 to avoid clogging of the adsorber. This is recommended also for usage of MA units with LC systems.
!
Important
Unfiltered feed will block the Membrane Adsorber and lead to capacity loss and increased back pressure. We recommend inline filtering during operation. With increase of pressure replace filter and restart.
7.7 Washing
When using MA units in bind and elute mode, wash with equilibra­tion buffer (see Tab. 1 and 2) after the sample has been loaded.
7.8 Elution
Elute the target protein with elution buffer (Tab. 1 and 2), with increased salt concentration or shifted pH.
Elution buffer examples
To elute BSA from Sartobind Q or D 1 M NaCl in binding buffer To elute lysozyme from Sartobind S 1 M NaCl in binding buffer
16
17
7.9 Regeneration, cleaning and storage
After elution, wash the unit with equilibration buffer. If necessary, use 1 N NaOH, 1 N HCl or 70% ethanol for 1 hour and store in 20% ethanol in equilibration buffer. Do not store units in water. Do not freeze the units. The MA 15, 75 or 100 can be cleaned in place directly after unpack­ing with 1 N NaOH, 1 hour at 20°C.
Example for cleaning of a Sartobind MA 75, bed volume 2.1 ml 1 1 N NaOH at the flow rate of 1 bed volume (BV)/min at 20°C
for1 h (in total ~130 ml) 2 10 BV (~20 ml) 1 M NaCl 3 10 BV (~20 ml) equilibration buffer
Do not autoclave.
7.10 Stability
The MA 15, 75 and 100 units are stable against all commonly used buffers in chromatography, e.g. 8 M urea, 8 M guanidine hydrochlo­ride, ethanol and acetone. They can be sanitized after prewashing with 1 N NaOH, 1 hour at room temperature (20°C). Do not use oxidizing agents such as hypochlorite or H
2O2
.
pH Stability
QS D Short term* 2 – 14 3 – 14 2 – 14 Long term** 2 – 12 4 – 13 4 – 13
* Short term refers to cleaning in place and regeneration procedures during
operation of units
** Long term: Storage of units e.g. overnight and longer. Preferably store units
in 20% ethanol in buffer within pH limits
18
19
7.11 Operation with peristaltic pump or LC system
After the unit is filled completely with equilibration buffer, close the outlet of the MA 15, 75 or 100 and remove the syringe. Start your LC system or peristaltic pump at a low flow rate. When fluid emerges, stop the pump and connect the tubing to the inlet of the unit. Make sure that no air is introduced. Remove the cap from outlet. Run the pump until fluid emerges from the outlet of the unit and stop it. Then connect the outlet of the unit via Luer adapter to the LC detector and proceed with loading. If your system pressure is too high, refer to your LC system manual to remove any flow restric­tor after the UV cell, as the system may generate a pressure above the allowed maximum pressure. As Membrane Adsorbers run typical­ly at much higher flow rates than columns, there is no risk of bubble formation in the UV cell when removing the restrictor.
Problem Possible cause Action
High back Material has Prefilter with 0.2 μm filter pressure not been before processing through the during sample filtered unit. loading
Material has Proteins can form aggregates been filtered within hours or during but was stored operation. Thus we recommend before to prefilter inline by attaching a purification 0.2 μm filter in front of the adsor-
ber. When you observe again pressure built up, replace the filter.
LC system Remove restrictor after the generates high UV cell. pressure
The MA 15, 75 Perform a regeneration cycle. or 100 adsorber is clogged
8. Troubleshooting
20
21
Problem Possible cause Action
Target mole- Conditions for Decrease conductivity, control other cule is not binding are process parameters as type of buffer bound insufficient and pH
Proteins or Perform a regeneration cycle contaminants are still bound from last cycle
Binding capa- Improper Prefilter with 0.2 μm filter before city decreases filtration processing through the unit. after several uses
Some molecule The binding capacity of the unit species binds depends on the nature of sample tightly (e.g. DNA and sample preparation, on Sartobind Q) prefiltration as well as proper and cannot be regeneration and application. removed with Tightly bound molecules may be 1 N NaOH, 1 h removable by specific enzymes,
or use unit only once.
Problem Possible cause Action
Proteins or Run a 1 M NaCl buffer step to contaminants elute tightly bound proteins are still bound quantitatively. Then regenerate from last cycle adsorber by loading with 1 N NaOH
and keep it for 1 hour at room temperature (20°C).
Wrong storage Do not store in sodium hydroxide
containing buffers. Store long term* in 20% ethanol-buffer (e.g. equili­bration buffer) solution and do not use oxidative chemicals in buffers.
22
23
Problem Possible cause Action
Early Binding Use larger adsorber device, or break- capacity is connect two adsorbers (same size) through of not sufficient in series (i.e. connect outlet of first protein adsorber to inlet of second) to
achieve higher binding capacity. As a rule of thumb the pressure doubles when the flow rate is kept constant and the number of membrane layers is doubled. We do not recommend to run two adsorbers in parallel.
* Long term storage is defined as storage of units when not in operation,
e.g. storage overnight and longer
Sartobind membranes have been tested for protein binding capacity and flow rate.
Sartobind MA units are manufactured in a controlled environment and have been tested for protein binding capacity and flow rate.
The product meets all Sartorius Stedim Biotech standards for traceability, production and specifications as given here or exceeded them.
9. Quality assurance
24
25
10.1 MA units
Order number Description Quantity
Sartobind MA units (reusable) 93IEXQ42GB-12--A Sartobind Q 15, Luer female and 4
male connectors
93IEXS42GB-12--A Sartobind S 15 Luer female and 4
male connectors
93IEXQ42DB-12--V Sartobind Q 75, Luer female and 2
male connectors
93IEXS42DB-12--V Sartobind S 75, Luer female and 2
male connectors
93IEXD42DB-12--V Sartobind D 75, Luer female and 2
male connectors
93IEXQ42BC-12 Sartobind Q 100, Luer female and 1
male connectors
93IEXS42BC-12 Sartobind S 100, Luer female and 1
male connectors
10. Ordering information
10.2 Accessories
Order number Description Quantity
1ZA- - -0004 Luer male adapter to 1
UNF 10-32 female, PEEK
1ZA- - -0005 Luer female adapter to 1
UNF 10-32 female, PEEK
17002---140 2 + M6 female to female Luer and 1
2 + M6 female to male Luer Lok Adapter, black Tefzel
26
27
MA 15 MA 75 MA 100
Membrane Area [cm2]1575100 Membrane volume [ml] 0.41 2.1 2.8 Dimensions 25 + 36 28 + 36 31 + 66
(height + diameter) [mm] Frontal surface area [cm2]5510 Connector inlet Luer female Connector outlet Luer male Approximate weight [g] 9 10 35
11. Dimensions and Connections
Sartorius Stedim Biotech GmbH August-Spindler-Strasse 11 37079 Goettingen, Germany
Phone +49.551.308.0 Fax +49.551.308.3289 www.sartorius-stedim.com
Published
28. February, 2011 Sartorius Stedim Biotech GmbH, Goettingen, Germany
Copyright by Sartorius Stedim Biotech GmbH, Goettingen, Germany.
All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written per­mission of Sartorius Stedim Biotech GmbH. The status of the information, specifications and illustrations in this manual is indicated by the date given below. All products are sold subject to the terms and conditions ruling at the time of delivery. Sartorius Stedim Biotech GmbH reserves the right to make changes to the technology, features, specifications and design of the equipment or discontinue products described at any time without notice or obligation. Sartobind and Minisart are trademarks of Sartorius Stedim Biotech GmbH, Luer-Lok is a trademark of Becton, Dickinson and Company.
Printed in Germany on paper that has been bleached without any use of chlorine W4A000 Sartobind IEX MA 15|75|100 Publication No.: SL-6171-e11011 Ver. 02 | 2011
Loading...