The New Quintix
Redefining Standard.
®
Intuitive application
guideance
Maximum accuracy
at all time
Laboratory Balances
“Plug & Work” connectivity
to printer or computer
AC Adapter
Sartorius AC adaptor module 6971790 with interchangeable country-specific plug-in AC adaptors
Primary 100 – 240 V~, –10% | +10%, 50 – 60 Hz, 0.2 A
Easy cleaning
Secondary 15 V DC, ± 5%, 530 mA (max.) | 8 Watt (max.): 0 to +40°C and
Other data protection class II, in accordance with EN | IEC 60950-1 up to 3000 m above sea level; IP40 as per EN | IEC 60529
Balance
Power supply only via Sartorius AC adaptor module 6971790
Input voltage 12.0 ... 18.0 V DC
Power consumption 2 W (typically)
Ambient Conditions
The specifications apply when the following ambient conditions are in place:
Environment for indoor use only
Ambient temperature * +10°C to +30°C
Operational capacity guaranteed between +5°C and +45°C +45 °C
Storage and shipping –10 °C to +60 °C
Elevation up to 3000 m above sea level
Relative humidity** 15% to 80% for temperatures up to 31°C; non-condensing, decreasing linearly to 50% relative humidity at
Safety of electrical
equipment
Electromagnetic
compatibility
Defined immunity to
interference
Interference emission Class B (suitable for use in residential areas and areas that are connected to a low voltage network that also
15 V DC, ± 5%, 330 mA (max.) | 5 Watt (max.): 0 to +50°C
40°C and 20% at 50°C
in accordance with EN 61010-1/IEC 61010-1. Safety requirements for electrical equipment for mea surement,
control, and laboratory use – Part 1: General requirements
in accordance with EN 61326-1/IEC 61326-1. Electrical equipment for measurement, control, and laboratory
use – EMC requirements - Part 1: General requirements
Suitable for use in industrial areas
supplies residential buildings). The device can therefore be used in both areas.
Balances verified for use in legal metrology comply with the requirements
of Council Directive 2009/23/EC, EN 45501:1992, and OIML R76:2006.
* For balances verified for use in legal metrology in accordance
with EU requirements, refer to the information on the balance.
** For balances verified for use in legal metrology in accordance
with EU requirements, the legal regulations apply.
Technical Specifications
Standard Equipment
Levelling Glass level indicator with air bubble for centering
Calibration Internal calibration isoCAL, External calibration
1
Selectable weight units
Interface mini USB
Display Touch screen with Sartorius graphical user interface optimized
Standard built-in
applications
Special built-in lab
applications
Languages English, German, French, Spanish, Italian, Russian, Chinese,
Protection – Chemical resistant finish of the housing
Password protection Supervisor lock for protection against unintentional changes
Anti-theft lock Kensington lock and lockdown capability for cable or chain
Underfloor weighing Integrated
1
Limited for verified models
Gram, kilogram, carat, pound, ounce, troy ounce, Hong Kong
tael, Singapore tael, Taiwan tael, grain, pennyweights, milligram,
parts per pound, China tael, mommes, Austrian carat, tola, baht,
mesghal and Newton
– Automatic recognition of Sartorius printer models YDP30
or YDP40
– Direct data transfer to Microsoft
®
Windows programs
– Programmable interval for data output
– Data transfer protocols SBI, xBPI, table format, text format
for users in pharmaceutical laboratories
Weighing, Density, Percentage, Checkweighing, Peak Hold,
Counting, Unstable Conditions
Mixing, Components, Statistics, Conversion
Polish, Japanese, Portuguese
– In-use cover
– Dust cover for analytical balances
Analytical Balances Quintix®
Model 224 124
Weighing capacity g 220 120
Readability mg 0.1 0.1
Repeatability (standard deviation) mg 0.1 0.1
Linearity deviation mg 0.2 0.2
Sensitivity drift between +10 and +30°C ± ppm/K 1.5 1.5
Typical stabilization time s 3 2
isoCAL:
– Temperature change
– Time interval
K
h
1.5
4
1.5
4
Display result (depending on the set filter level) s 0.2 0.2
Weighing pan size mm d 90 d 90
Weighing chamber height mm 209 209
Net weight, approx. kg 4.9 4.9
Dimensions, D + W + H mm
Inch
360 + 216 + 320
14.1 + 8.5 + 12.6
Precision Balances Quintix
®
Model 513 313 213 5102 3102 2102 1102 612 5101 5100
Weighing capacity g 510 310 210 5100 3100 2100 1100 610 5100 5100
Readability mg 1 1 1 10 10 10 10 10 100 1000
Repeatability (standard
mg 1 1 1 10 10 10 10 10 100 500
deviation)
Linearity deviation mg 2 2 2 20 20 30 30 30 300 1000
Sensitivity drift between
± ppm/K 3 3 3 3 3 5 5 5 10 10
+10 and +30°C
Typical stabilization time s 1.5 1.5 1.5 1.5 1.5 2 2 2 2 2
isoCAL:
– Temperature change
– Time interval
Display result (depending
K
h
2
6
4
12
4
12
2
6
4
12
4
12
4
24
4
24
4
24
4
24
s 0.1 | 0.2 0.1 | 0.2 0.1 | 0.2 0.1 | 0.2 0.1 | 0.2 0.1 | 0.2 0.1 | 0.2 0.1 | 0.2 0.1 | 0.2 0.1 | 0.2
on the set filter level)
Weighing pan size mm d 120 d 120 d 120 d 180 d 180 d 180 d 180 d 180 d 180 d 180
Weighing chamber height mm 209 209 209 – – – – – – –
Net weight, approx. kg 4.9 4.9 4.9 5.2 5.2 4.7 4.7 4.7 4.7 4.7
Dimensions, D + W + H mm
Inch
360 + 216 + 320
14.1 + 8.5 + 12.6
360 + 216 + 95
14.1 + 8.5 + 3.75