– Unmatched reliability with the world-class weighing
instrument – Overload protection for years of reliability – Supervisor Lock – Fast and easy data transfer

Product Information

Get best value for your money, without compromises in precision and reliability. Rely on consistent readings and excellent repeatability ensured by Sartorius quality and technology developed and designed in Germany.

Technical Specifications

AC Adapter Sartorius AC adaptor
module Primary 100 – 240 V~, –10% | +10%, 50 – 60 Hz, 0.2 A Secondary 15 V DC, ± 5%, 530 mA (max.) |
Other data protection class II, in accordance with
6971790 with interchangeable country­specific plug-in AC adaptors
8 Watt (max.): 0 to +40 °C and 15 V DC, ± 5%, 330 mA (max.) | 5 Watt (max.): 0 to +50 °C
EN | IEC 60950-1 up to 3000 level; IP40 as per EN | IEC 60529
m above sea
Balance Power supply only via Sartorius AC adaptor module
6971790 Input voltage 12.0 ... 18.0 V DC Power consumption 2 W (typically)
Laboratory Balances
Ambient Conditions The specifications apply when the following ambient conditions
are in place: Environment for indoor use only Ambient
+10 °C to +30 °C
temperature* Operational capacity guaranteed between +5 °C and +45 °C Storage and shipping –10 °C to +60 °C Elevation up to 3,000 m above sea level Relative humidity** 15% to 80% for temperatures up to 31 °C;
non-condensing, decreasing linearly to 50% relative humidity at 40 °C and 20% at 50 °C
Safety of electrical equipment
in accordance with EN 61010-1 | IEC 61010-1. Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements
Electromagnetic compatibility
in accordance with EN 61326-1 | IEC 61326-1. Electrical equipment for measurement, control, and laboratory use – EMC requirements – Part 1: General requirements
Defined immunity
Suitable for use in industrial areas
to interference Interference emission Class B (suitable for use in residential
areas and areas that are connected to a low voltage network that also supplies residential buildings). The device can therefore be used in both areas.
Verified balances in accordance with EU requirements comply with the requirements of Council Directive 2009/23/EC with EN 45501:1992 and OIML R76:2006.
* For verified balances in accordance with EU requirements,
refer to the information on the balance.
** For verified balances in accordance with EU requirements,
the legal regulations apply.
Standard Equipment Levelling Glass level indicator with air bubble for
centering Calibration External calibration Selectable
weight units
Gram, kilogram, carat, pound, ounce,
troy ounce, Hong Kong tael, Singapore tael,
Taiwan tael, grain, pennyweights, milligram,
parts per pound, China tael, mommes, Aus-
trian carat, tola, baht, mesghal and Newton Interface mini USB
– Automatic recognition of Sartorius
printer YDP40
– Direct data transfer to Microsoft
®
Windows programs – Programmable interval for data output – Data transfer protocols SBI, table format,
text format
Display Touch screen with Sartorius graphical
user interface
Standard built-in applications
Weighing, Density, Percentage, Checkweighing, Peak Hold, Counting, Unstable Conditions (Animal Weighing)
Languages English, French, German, Hungarian,
Italian, Polish, Portuguese, Russian, Spanish, Turkish, Chinese, Japanese, Korean
Protection – Rugged, easy-to-clean housing
– In-use cover – Dust cover for analytical balances
Password protection Supervisor lock for protection against
unintentional changes
Anti-theft lock Kensington lock and lockdown capability
for cable or chain
Underfloor
Integrated
weighing
Design 1 Design 2 Design 3

Standard Models

Model 224-1x
1)
124-1x
1)
64-1x
1)
513-1x
1)
313-1x
1)
213-1x
1)
Design 1 1 1 2 2 2
Weighing capacity g Readability mg Repeatability
mg
220 120 60 510 310 210
0.1 0.1 0.1 1 1 1
0.1 0.1 0.1 1 1 1
(standard deviation) Linearity deviation mg
Sensitivity drift between
± ppm/K
0.2 0.2 0.2 2 2 2 2 2 2 3 3 3
+10 °C and +30 °C Typical stabilization time s 2 2 2 1 1 1 Weighing pan size mm
d 90 d 90 d 90 d 120 d 120 d 120 Weighing chamber height* mm 209 209 209 209 209 209 Net weight, approx. kg 4.5 4.5 4.5 4.9 4.9 4.9 Dimensions, D × W × H mm
Inch
Model 3102-
360 × 216 × 320
14.1 × 8.5 × 12.6
2102-
1x
1)
1x
1)
1102-
1)
1x
612-
1)
1x
412-
1)
1x
6101-
1)
1x
5101-
1)
1x
2101-
1)
1x
6100-
1)
1x
5100-
1)
1x
Design 3 3 3 3 3 3 3 3 3 3 Weighing capacity g 3,100 2,100 1,100 610 410 6,100 5,100 2,100 6,100 5,100 Readability mg 10 10 10 10 10 100 100 100 1,000 1,000 Repeatability
mg 10 10 10 10 10 100 100 100 500 500
(standard deviation) Linearity deviation mg 30 30 30 30 30 300 300 300 1,000 1,000 Sensitivity drift between
± ppm/K 4 4 4 4 4 8 8 8 8 8
+10 °C and +30 °C Typical stabilization time s 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1.5 1 1 Weighing pan size mm
d 180 d 180 d 180 d 180 d 180 d 180 d 180 d 180 d 180 d 180 Net weight, approx. kg 3.1 3.1 3.1 3.1 3.1 3.1 3.1 3.1 3.1 3.1 Dimensions, D × W × H mm
Inch
360 × 216 × 95
14.1 × 8.5 × 3.75
* Upper edge of the weighing pan to the lower edge of the upper draft shield panel
1)
Possible terms for country-specific models: x = S: Standard balances without country-specific additions x = SAR: Standard balances with country-specific additions for Argentina x = SJP: Standard balances with country-specific additions for Japan x = SKR: Standard balances with country-specific additions for South Korea
Laboratory Balances 3