Technical Data LA L D1 | N C3
Accuracy class 0.25 0.25 0.04 0.015 % E
max
Minimum dead load lowest limit of specified measuring range E
min
00 0 0% E
max
Maximum capacity highest limit of specified measuring range E
max
s. table s. table s. table s. table
Minimum LC verification minimum load cell verification interval,
v
min
= E
max
/Y Y 5,000 14,000
interval for E
max
= 1,000 kg Y 4,000
for E
max
= 500 kg Y 2,000
Rated output relative output at nominal load Cn16 mA 1 1 1 mV/V
for E
max
= 50 t (LA = 4...20 mA) Cn16 mA 2 2 2 mV/V
Tolerance on rated output permissible deviation from rated output dc< 1.0 < 1.0 < 0.25 < 0.07 % C
n
Zero output signal load cell output signal under unloaded condition S
min
4 mA* < 2.0 < 1.0 < 1.0 % C
n
Repeatability error max. change in load cell output
for repeated loading ε
R
< 0.02 < 0.02 < 0.01 < 0.005 % C
n
Creep, during 30 min. max. change in load cell output
under nominal load d
cr
< 0.05 < 0.05 < 0.03 < 0.015 % C
n
Non-linearity max. deviation from best straight line through zero d
Lin
< 0.25 < 0.25 < 0.03 < 0.01 % C
n
Hysteresis max. difference in load cell output dhy< 0.25 < 0.25 < 0.04 < 0.015 % C
n
between loading and unloading
Temperature effect on S
min
max. change of S
min
/10 K over BTreferred to C
n
TK
Smin
< 0.15 < 0.15 < 0.028 < 0.01 % Cn/10K
Temperature effect on C max. change of C/10 K over BTreferred to C
n
TKc< 0.1 < 0.1 < 0.03 < 0.01 % Cn/10K
Input impedance between supply terminals RLC– 650 + 50 650 ± 6 650 ± 6 O
Output impedance between measuring terminals RO– 610 ± 3 610 ± 1 610 ± 0.5 O
Insulation impedance between measuring circuit and housing at 100 VDCRIS– > 5,000 +106> 5,000 +106> 5,000 +106 O
Insulation voltage between circuit and housing, PR 6201/..E only - 500 500 500 V
Recommended to hold the specified performance Bu20… 28 4… 24 4… 24 4… 24 V
supply voltage
Max. supply voltage permissible for continuous operation U
max
28 32 32 32 V
without damage
Nominal ambient to hold the specified performance BT-10… +55 -10… +55 -10… +55 -10… +55 °C
temp. range
Usable ambient permissible for continuous BTu-30… +55 -40… +95 -40… +95 -40… +95 °C
temp. range operation without damage
Storage transportation and storage BTl-40… +70 -40… +95 -40… +95 -40… +95 °C
temperature range
Permissible eccentricity permissible displacement from nominal load line Sex10 10 10 10 mm
Vibration resistance resistance against oscillation (IEC 68-2-6 Fc) - 20 g, 100 h, 20 g, 100 h, 20 g, 100 h, 20 g, 100 h,
10… 150 Hz 10… 150 Hz 10… 150 Hz 10… 150 Hz
Air pressure effect influence of ambient air pressure on S
min
PK
Smin
up to 2 t 250 250 250 250 g/kPa
3 t to 10 t 320 320 320 320 g/kPa
from 20 t 420 420 420 420 g/kPa
Nominal deflection max. elastic deformation under nominal load s
nom
up to 30 t < 0.5 < 0.5 < 0.5 < 0.5 mm
50 t < 0.8 < 0.8 < 0.8 < 0.8 mm
* Tolerance on zero output signal: ± 2% C
n
Definitions acc. to VDI/VDE 2637
Data for LA version are typical values. The technical data given here serve only as a product
description and must not be interpreted as guaranteed characteristics in the legal sense.
Restoring force
For each mm of movement that the top of
the load cell shifts from the vertical axis, a
horizontal restoring force is generated of:
max. capacity ≤ 10 t: 0.65 %
max. capacity ≥ 20 t: 1.55 %
of the vertical load on the load cell
Load cell housing
Full stainless steel housing, membrane and
measuring element hermetically sealed,
welded, filled with inert gas.
Material-No.1.4301 (DIN 17440), equivalent
to 304 S11/S15 (B.S.)
Ingress Protection
IP68, IEC529 (equivalent to NEMA 6). The
load cell can be submerged in water to a
depth of 1.5 m for 10,000 hours, IP69k
Cable
Robust, flexible, screened
Sheath: TPE Thermopl. Elastomere, grey
(for PR 6201/..E: blue)
Diameter: 5 mm, wires 4 + 0.35 mm
2
Length: 5 m (500 kg - 10 t), 12 m (20 t – 50 t)
Bending radius
Fixed installation: ≥ 50 mm
Flexible installation: ≥ 150 mm
Certificate of conformity
Valid for: PR 6201/..E
Feature:
II 1 G EEx ia IIC T6, II 1D IP65 T 85 °C
Registration number:
PTB 02 ATEX 2059, TÜV 03 ATEX 2301X