Sartorius PR 6201 Technical data

PR 6201 (500 kg… 50 t) Precision Compression Load Cell
Product Profile
The PR 6201 range of load cells is specially designed for weighing silos, tanks and process vessels.
The unique design principle, in combination with the FlexLock installation kits, makes it possible to blance out movements arising from mechanical or thermal ex pansion or contraction of the vessel or its supporting construction.
A particular design characteristic is that the overall height and shape of the load cell remain unchanged, from 500 kg up to 50 t. Alongside this, the unit has an especially high overload capacity of up to 200 %.
At the same time, this range distinguishes itself – in addition to its high measurement accuracy and repeatability - above all for its unmatched reliability, robustness and stability, which offer trouble-free operation without adjustment, year after year.
The pendulum support principle, combined with patented measuring element geometry, ensures that force transmission into the sensor is always at the optimum level and, in this way, the effect on measurement accuracy is minimized. At the same time, the load cell offers a particularly high over­load range, high repeatability and perfect linearity.
There is an especially wide working tempera­ture range attributable to sophisticated resistance strain gauge technology. The hermetically sealed enclosure and special TPE cable allow the unit to be used even under extreme operating conditions in harsh production environments.
The entire measurement chain can be calibrated without the use of reference weights. Due to “matched output” techno ­logy, a damaged load cell can be exchanged without the need for re-calibration. This saves a tremendous amount of time during commissioning.
An explosion-proof (Ex) version of this range of load cells is also available, as an option, for use in intrinsically safe environments.
500 kg… 50 t Type LA | L | D1 | C3
– Easy to install – Well-proven rockerpin design – Easy corner adjustment by matched
load cell output – Full stainless steel housing – Wide temperature range – High overload capacity – Resistant against vibrations – Hermetically sealed, IP68
(depth of 1.5 m for 10,000 hrs.), IP69K – 4 to 20 mA output signal as option
(LA version) – Optimum overvoltage protection – Ex-version available (PR 6201/..E) – W&M aprroval (OIML R60, NTEP)
Technical Data LA L D1 | N C3
Accuracy class 0.25 0.25 0.04 0.015 % E
max
Minimum dead load lowest limit of specified measuring range E
min
00 0 0% E
max
Maximum capacity highest limit of specified measuring range E
max
s. table s. table s. table s. table
Minimum LC verification minimum load cell verification interval,
v
min
= E
max
/Y Y 5,000 14,000
interval for E
max
= 1,000 kg Y 4,000
for E
max
= 500 kg Y 2,000
Rated output relative output at nominal load Cn16 mA 1 1 1 mV/V
for E
max
= 50 t (LA = 4...20 mA) Cn16 mA 2 2 2 mV/V
Tolerance on rated output permissible deviation from rated output dc< 1.0 < 1.0 < 0.25 < 0.07 % C
n
Zero output signal load cell output signal under unloaded condition S
min
4 mA* < 2.0 < 1.0 < 1.0 % C
n
Repeatability error max. change in load cell output
for repeated loading ε
R
< 0.02 < 0.02 < 0.01 < 0.005 % C
n
Creep, during 30 min. max. change in load cell output
under nominal load d
cr
< 0.05 < 0.05 < 0.03 < 0.015 % C
n
Non-linearity max. deviation from best straight line through zero d
Lin
< 0.25 < 0.25 < 0.03 < 0.01 % C
n
Hysteresis max. difference in load cell output dhy< 0.25 < 0.25 < 0.04 < 0.015 % C
n
between loading and unloading
Temperature effect on S
min
max. change of S
min
/10 K over BTreferred to C
n
TK
Smin
< 0.15 < 0.15 < 0.028 < 0.01 % Cn/10K
Temperature effect on C max. change of C/10 K over BTreferred to C
n
TKc< 0.1 < 0.1 < 0.03 < 0.01 % Cn/10K Input impedance between supply terminals RLC– 650 + 50 650 ± 6 650 ± 6 O Output impedance between measuring terminals RO– 610 ± 3 610 ± 1 610 ± 0.5 O Insulation impedance between measuring circuit and housing at 100 VDCRIS– > 5,000 +106> 5,000 +106> 5,000 +106 O Insulation voltage between circuit and housing, PR 6201/..E only - 500 500 500 V Recommended to hold the specified performance Bu20… 28 4… 24 4… 24 4… 24 V
supply voltage Max. supply voltage permissible for continuous operation U
max
28 32 32 32 V
without damage
Nominal ambient to hold the specified performance BT-10… +55 -10… +55 -10… +55 -10… +55 °C temp. range
Usable ambient permissible for continuous BTu-30… +55 -40… +95 -40… +95 -40… +95 °C temp. range operation without damage
Storage transportation and storage BTl-40… +70 -40… +95 -40… +95 -40… +95 °C temperature range
Permissible eccentricity permissible displacement from nominal load line Sex10 10 10 10 mm Vibration resistance resistance against oscillation (IEC 68-2-6 Fc) - 20 g, 100 h, 20 g, 100 h, 20 g, 100 h, 20 g, 100 h,
10… 150 Hz 10… 150 Hz 10… 150 Hz 10… 150 Hz
Air pressure effect influence of ambient air pressure on S
min
PK
Smin
up to 2 t 250 250 250 250 g/kPa 3 t to 10 t 320 320 320 320 g/kPa from 20 t 420 420 420 420 g/kPa
Nominal deflection max. elastic deformation under nominal load s
nom
up to 30 t < 0.5 < 0.5 < 0.5 < 0.5 mm 50 t < 0.8 < 0.8 < 0.8 < 0.8 mm
* Tolerance on zero output signal: ± 2% C
n
Definitions acc. to VDI/VDE 2637 Data for LA version are typical values. The technical data given here serve only as a product description and must not be interpreted as guaranteed characteristics in the legal sense.
Restoring force
For each mm of movement that the top of the load cell shifts from the vertical axis, a horizontal restoring force is generated of: max. capacity 10 t: 0.65 % max. capacity 20 t: 1.55 % of the vertical load on the load cell
Load cell housing
Full stainless steel housing, membrane and measuring element hermetically sealed, welded, filled with inert gas. Material-No.1.4301 (DIN 17440), equivalent to 304 S11/S15 (B.S.)
Ingress Protection
IP68, IEC529 (equivalent to NEMA 6). The load cell can be submerged in water to a depth of 1.5 m for 10,000 hours, IP69k
Cable
Robust, flexible, screened Sheath: TPE Thermopl. Elastomere, grey (for PR 6201/..E: blue) Diameter: 5 mm, wires 4 + 0.35 mm
2
Length: 5 m (500 kg - 10 t), 12 m (20 t – 50 t)
Bending radius
Fixed installation: 50 mm Flexible installation: 150 mm
Certificate of conformity
Valid for: PR 6201/..E Feature: II 1 G EEx ia IIC T6, II 1D IP65 T 85 °C Registration number: PTB 02 ATEX 2059, TÜV 03 ATEX 2301X