Sartorius MSE225S, MSE6.6, MSE225P, MSE125P, MSE324S Operating Instructions Manual

...
Operating Instructions

Sartorius Cubis Series

Electronic Semimicro, Micro, Precision and Analytical Balances MSE Models

1000025315

Contents

Notes on Using these Instructions � � � � � � � � � � � � � � � � 3 Safety Instructions ���������������������������� 4 Intended Use �������������������������������� 6 General View of the Equipment ���������������� 7 Getting Started ������������������������������ 9
Unpacking the Equipment ��������������������� 9 Setup Instructions ��������������������������� 9 Assembly ��������������������������������� 11 Transporting the Balance ��������������������� 17 Storage and Shipping Conditions ���������������� 17 Power Connection ��������������������������� 18 Warm-up Time ����������������������������� 19 Anti-theft Locking Device ��������������������� 20 Modification Options ������������������������ 21 Preparing Below-Cell Weighing ����������������� 26 Using the Cable Opening in the Analytical Balance Draft shield� � � � � � � � � � � � � � � � � � 28
Operation ����������������������������������� 29
Display and Control Panel ��������������������� 29 Basic Weighing Function ���������������������� 30 Leveling the Balance with the Inclination Sensor ���� 31 Turning the Ionizer On/Off (Optional) ������������ 31 Calibration and Adjustment �������������������� 33
Configuration (Operating Menu) ����������������� 36
Key Functions during Configuration ������������� 36 Menu Navigation Example: Setting the Language ��� 36
Menu Structure (Overview) �������������������� 37 Parameter Settings: Overview ������������������ 38
Application Programs �������������������������� 45
Density Determination ������������������������ 45 Counting ��������������������������������� 48 Weighing in Percent ������������������������� 50 Calculation �������������������������������� 52 Animal Weighing/Averaging ������������������� 54 Net-total Formulation ������������������������ 56 Totalizing ��������������������������������� 58 Mass Unit Conversion ������������������������ 60
ISO/GLP-compliant Print-out �������������������� 62 Data Interfaces� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � 63 Error Codes ��������������������������������� 75 Care and Maintenance
(Packing the Balance for Shipping) �������������� 76
Transporting the Balance ����������������������� 80 Disposal ������������������������������������ 86 Specifications �������������������������������� 87 Dimensions (Balance Drawings) ����������������� 100 Accessories (Options) ������������������������� 108 EC / EU Declaration of Conformity ��������������� 110 FCC Supplier’s Declaration of Conformity ��������� 111 CSA Certificate of Compliance ������������������ 112 EC Type Approval Certificate ������������������� 115 Plates and Markings �������������������������� 116
 The enclosed CD-ROM contains the
h
operating instructions as a PDF file� To read and print the file, you require the program Acrobat Reader, which you can download for free from the Adobe website (www�adobe�com)�
Software drivers for configuration of SB interfaces can also be downloaded from the internet: http://www�ftdichip�com/FTDrivers�htm The “installation guides" for these drivers can be found here: http://www�ftdichip�com/Documents/ InstallGuides�htm
2 Cubis MSE Operating Instructions

Notes on Using these Instructions

t Please read these instructions carefully and completely before using the device� t Read the Safety Instructions carefully� t These instructions are part of the product� Keep them in a safe and easily
accessible location�
t If the instructions should be lost or misplaced, please contact Sartorius
for a replacement or download the latest version from our website: www�sartorius�com
Symbols and Signs
The following symbols are used in this manual:
Danger of explosion
1
Potential danger of physical injury or death
Notes on Using these Instructions
Danger of electrical shock
2
Danger of physical injury or death
Warning of potential physical injury,
3
Health risk or risk of property damage
Useful information and tips
h
Notes on use in legal metrology within the scope of validity of
Council Directive 2014/31/EU for instruments for which the conformity assessment procedure has been carried out (verified instruments)�
Note regarding device operation
Explanation of Symbols
The following symbols are used in these instructions:
t Indicates a required action
y Describes what happens after you have performed a particular step
1�, Perform steps in the specified order 2� – Indicates an item in a list
Technical advice / Hotline:
h
Phone: +49�551�308�4440 Fax: +49�551�308�4449
Cubis MSE Operating Instructions 3

Safety Instructions

Safety Instructions
This device complies with European Council Directives as well as international regulations and standards for electrical equipment, electromagnetic compatibility, and the stipulated safety requirements� Improper use or handling can, however, result in damage and/or injury� This device should only be operated by trained personnel� Operators must read these installation instructions, particularly the safety information, and must be familiar with the operation of the equipment� The operator is required to supplement these safety precautions as appropriate� The operating personnel must be provided with the appropriate training� The requirements pertaining to applicable installation regulations must be followed when using electrical equipment in systems and environmental conditions with increased safety requirements� Relevant laws, standards, regulations, guidelines and environmental protection laws valid in your country must be observed� Always keep the equipment and balance freely accessible� Any installation work or balance operation that does not conform to the instructions will result in forfeiture of all claims under the manufacturer's warranty�
Danger of Explosion!
Do not use this equipment in hazardous areas in which explosive materials are present�
1
Danger of Electrical Shock!
If the device is not used in accordance with the instructions, this can affect the product liability for the device�
3
Make sure that the voltage rating printed on the AC adapter is identical to your local line voltage� The wall outlet must
2 2
2 2
2
have a protective grounding conductor� The operating instructions included with the AC adapter must be followed�
Do not switch on the equipment if the AC adapter or power cord are damaged� If the balance, the AC adapter or
any of the cables are damaged, disconnect the equipment from the power supply and secure it so that it cannot be turned on�
Chemicals (e�g� gases or liquids) that can corrode and damage the inside or outside of the balance, AC adapter, power cable or
peripherals must be kept away from the equipment�
Do not operate the balance unless the housing and the display are undamaged and the housing is closed so that
inadvertent contact with parts inside the equipment is not possible� Make sure that no liquids penetrate the balance and do not allow conductive parts to touch the balance�
Note on Installation:
The operator shall be solely responsible for any modifications to the equipment and for connecting any cables or equipment not supplied by Sartorius� Information on operational quality is available upon request from Sartorius� You should only use peripherals and options supplied by Sartorius�
Protection Class of the Balance Housing
3
4 Cubis MSE Operating Instructions
– Models with a readability of 10 mg and 0�1 mg: These models fulfill the requirements of protection class IP54� The IP protection only applies to the balance housing and not to the AC adapter� – The IP protection is only guaranteed when the weighing pan is mounted� – The IP protection is restricted if the data interfaces are used�
Retain the cover cap for the data output� When the data output is not in use, reapply the cover cap to protect it against vapors, moisture, and dust or dirt�
Safety Instructions
Note on cable quality:
3
3 3 3
CE conformity is only guaranteed with well shielded cables� The shielding braid of the cable must be made of metal with at least 80% coverage and also covered with metal foil as much as possible� The shielding on both sides on the full metal or metalized connector casing, as well as on the casing inlet, must be high-frequency compliant and have a flat design�
You can only turn off power to the device by pulling the plug or disconnecting the power cord, or by unscrewing the four-
pin plug at the DC supply cable for the balance�
Only use a lightly-moistened cloth for cleaning� See “Care and Maintenance" for cleaning the device� Do not
compromise the IP protection of the balance�
Do not open the balance housing� If the seal is broken, this will result in forfeiture of all claims under the
manufacturer’s warranty� The device may only be opened by specialized personnel trained by Sartorius�
Disconnect the balance from the wall outlet prior to moving the device�
3
Avoid condensation damage to the device� When the balance is turned on,
3
Hazards at the Place of Installation and During Operation
3
3
3
3
3
condensation build-up is not possible� If the device is moved from a cool environment to a warm environment, you must allow the device to acclimatize for approx� 2 hours before connecting to the power supply�
Avoid generating static electricity and establish potential equalization� A 10 kOhm ground is used when connecting
the balance to the power supply� Disconnecting the ground conductor is not permitted�
Models with a separate electronics module (e� g� MSE6�6/3�6/2�7 225S/P and 125P): Do not replace the electronics
module� Replacing with, e� g� models of the ME/SE series will damage the device�
Exercise caution to avoid glass breakage�
There is a danger of lacerations or cuts caused by sharp edges on glass or metal (contact springs), particularly when removing or replacing the glass draft shield panels; during cleaning; if a glass sample container breaks, or when modifying the balance hardware configuration�
Make sure all equipment operators have received proper instructions when working with hazardous or toxic materials�
Failure to follow correct handling procedures can result in breakage of parts or spillage of liquids or other substances� Make sure the necessary protective clothing or equipment is worn, such as protective gloves, clothing, eyewear� All safety regulations applicable in the laboratory must be observed�
Exercise caution to avoid squeezing or crushing fingers when
– Operating the electronic draft shield – Removing or installing the panels – Adjusting the display
Do not expose the device to aggressive chemical vapors or to unnecessarily extreme temperatures, moisture, shocks,
3 3
or vibration�
The device should be setup in areas away from equipment or devices that have strong magnetic properties�
Extreme electromagnetic fields from electrical lines should also be avoided�
Cubis MSE Operating Instructions 5
Safety Instructions
Use69719
87
1
2
3
4
5
6
7
8
9
11
12
13
14
15 16
17
18
19
7
1
3
20
10
10

Intended Use

Cubis models are high-resolution balances� They were specially developed for exact determination of material mass in liquid, paste, powder or solid form�
Appropriate containers must be used for each type of sample material� Specific models cover specific weighing ranges: see “Specifications�"
Cubis models are designed specially for use in research, education and day-to-day laboratory tasks in science and industry�
They are designed to be used exclusively indoors�
Cubis models can be operated as standalone, connected to a PC, or on a network�

General View of the Equipment and Equipment Supplied

Balances with a readability of 1 µg
Pos� Description
1 Weighing pan 2 Filter weighing pan d 50 mm 3 Internal draft shield (for MSx2�7S-F only) 4 Optional weighing pan d 75 mm 5 Shield disk 6 Optional filter weighing pan d 90 mm 7 Shield plate 8 Level indicator 9 Leveling foot 10 MSE2�7�� model: bush 11 Draft shield cover
6 Cubis MSE Operating Instructions
Pos� Description
12 Female connector for weigh cell 13 DC jack 14 Display and control unit 15 Serial communications port (PERIPHERALS) 16 Slot for optional interfaces, e� g� 9-pin data output or Bluetooth�
PS2 socket has no function� 17 Lug for attaching antitheft locking device 18 Equipotential bonding conductor terminal 19 Equipotential bonding conductor terminal 20 Draft shield
Not shown: AC adapter, USB cable
General View of the Equipment and Equipment Supplied
9
10
22
3
4
5
5
6
6
12
12
10
10
13
13
6
11
11
4
3
3
14 15 16
17 18
19
20
21
4
23
24
14
16
19
20
21
1
2
3
4
5
6
7
8
General View of the Equipment and Equipment Supplied
Item Description
Analytical and precision balances with a weighing range of up to 15 kg
1 Upper sliding draft shield panel/Handle 2 Back panel 3 Weighing pan 4 Pan support (not for MSx225���/MSx125��� models) 5 Shield plate 6 Pan retainer 7 Right sliding panel/Handle 8 Leveling foot 9 Display 10 Display and control unit 11 Draft shield/Shield disk (only for models with a readability of
1 mg and 10 mg) 12 Below-cell weighing port (on the bottom of the balance) 13 Leveling foot 14 Level indicator 15 Power socket 16 Slot for attaching an anti-theft locking device 17 USB socket for a PC connection 18 Below-cell weighing hook 19 Slot for optional interfaces, e�g�, 9-pin data output (as shown) or Bluetooth� PS2 socket has no function� 20 Communication port (PERIPHERALS) for accessories 21 Menu access switch 22 Left sliding panel/handle 23 Semi-microbalances: Female connector for weigh cell 24 Semi-microbalances: Electronics module (for MSx225���/MSx125��� models)
Not shown: – AC adapter – USB cable – Operating instructions – In-use cover over the display unit – Dust cover (only for balances with a draft shield) – In-use cover for housing (only for models with a readability of
19 mg/100 mg and a weighing capacity of <14 kg)
Cubis MSE Operating Instructions 7
General View of the Equipment and Equipment Supplied
1
3 4 5
7
2
6

Equipment Supplied and General View of the Equipment

Balances with a weighing range of 20 kg or more
Pos� Bezeichnung
1 DC jack 2 Leveling foot 3 Level indicator 4 Serial communications port (PERIPHERALS)
Pos� Bezeichnung
5 USB interface for PC connection 6 Display and control unit 7 Weighing pan
Not shown: AC adapter, USB cable
8 Cubis MSE Operating Instructions

Getting Started

°C
Getting Started

Unpacking the Equipment

t Open the packaging at the top�
t Use both hands to lift the balance, with the packaging, out of the cardboard
box�
t Place the packaging with the balance on the floor�
t Remove the top part of the packaging�

Installation Instructions

t Select the proper setup location:
– Set up the device on a stable, even surface that is not exposed to vibrations
(e�g� weighing stone)�
– Place the device in a location with enough free space around it so that
excessive heat cannot build up�
– Maintain free access to the device at all times�
Choose a location that is not subject to the following negative influences: – Heat (heater or direct sunlight) – Drafts from open windows, AC systems and doors – Extreme vibrations during weighing – Excessive moisture
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when it is brought into a substantially warmer area� To avoid the effects of condensation, condition the balance for about two hours at room temperature, leaving it unplugged from the power supply� Carefully read all warnings and safety precautions in the respective section of this manual�
Cubis MSE Operating Instructions 9
Getting Started
Balance with Draft Shield
t Remove the packages (containing draft shield panels, weighing pan, pan
support, AC adapter, etc�) from the lower packaging and place them to one side�
t Use both hands to lift the balance out of the packaging�
Exercise caution to avoid glass breakage�
3
t Place the balance on an even surface�
Never lift the equipment by the draft shield panels to remove it from the packaging�
Balance with No Draft Shield
t Use both hands to remove the model without a draft shield from the
packaging�
t Remove the AC adapter and the balance parts from the packaging�
10 Cubis MSE Operating Instructions
Getting Started

Assembly

Installing the Analytical Draft Shield Labeled DA, DI, and DU
3 2
1
Assembling the Balance
t Fit parts onto the balance as shown in the picture�
1. Shield plate
2. Pan support (not for MSx225���/MSx125��� models)
3. Weighing pan
t Slide the upper draft shield panel into the guide rails from the rear�
t Slide the left and right draft shield panels into the guide rails from the rear�
Make sure the panels are within the upper and lower guide rails�
t Slide the panels in until they engage�
1. Right draft shield panel
2. Left draft shield panel
y This completes the balance assembly�
Cubis MSE Operating Instructions 11
3
2
1
3
2
1
Getting Started
Installing the Draft Shield Labeled DE for Milligram Balances Assembling the Balance
t Fit parts onto the balance as shown in the picture�
1. Shield plate
2. Pan support
3. Weighing pan
Assembly of Milligram Balances without Glass Draft Shield (DR Option): Balance Assembly
4
3
2
1
t Fit all components listed below onto the balance in the order given:
1. Pan support (see also next section)
2. Shield plate
3. Weighing pan
4. Draft shield frame
t First insert the shield plate�
t Make sure the pin on the pan support is facing toward the front, and slide it
under the clip on the pan retainer�
1. Insert the pan support�
t After inserting the pan support, press it down against the shield plate so that it
lies flat�
3
2. Press down the pan support�
Exercise caution to avoid pinching or crushing fingers�
12 Cubis MSE Operating Instructions
t Place the weighing pan on the pan support�
3. Weighing pan
Getting Started
2
1
2
1
t Slide the upper draft shield panel into the guide rails from the rear while
pressing the locking tab�
t Slide the side draft shield panel into the guide rails from the rear while pressing
the locking tab�
y This completes the balance assembly�
t Remove the panel again if necessary
1. Press on the locking tab�
2. Remove the panel again if necessary
t Place the panel in the storage slot at the back of the balance�
Assembly for Precision Balances without Draft Shield with a Weighing Range
3
2
1
up to 15 kg
t Fit parts onto the balance as shown in the picture�
1. Pan support
2. Shield plate/Draft shield
3. Weighing pan
1
t Place the pan support diagonally and press down lightly�
1. Pan support
Cubis MSE Operating Instructions 13
Getting Started
3
t Carefully turn the pan support clockwise until the two buttons engage�
The pan support is now attached�
t 2. Insert the shield plate/draft shield�
t 3. Place the weighing pan on the pan support�
2
y This completes the balance assembly�
Models with a Weighing Range from 20 kg Upwards:
t Place the weighing pan on the balance�
14 Cubis MSE Operating Instructions
Getting Started
MSE225…, MSx125:
MSE6�6S/3�6P/2�7S:
Connecting the Weigh Cell to the Electronics Module/Evaluation Unit
t Connect the weigh cell and electronics module using the connection cable�
Make sure that the connectors click into place so that both ends are securely connected�
Please check the plug contacts to ensure a proper connection�
3
t Connecting the Weigh Cell to the Evaluation Unit
3
There should be no tension on the connection cable� We therefore recommend placing the device close to the wall�
Make sure that the connectors click into place so that the connection is secure�
Models with a separate electronics module (e�g� MSE6�6/3�6/2�7),
MSE225S/P and MSE125P: Do not replace the electronics module� Replacing them with models from the ME/SE series, for example, can damage the device�
MSE6�6S/3�6P/2�7S:
MSE6�6S-F/2�7S-F:
Microbalances (Models MSE6.6S/3.6P/2.7S): Placing Components on the Weigh Cell
t Install the following parts in the order listed:
5
1� MSE2,7S models: bush
4
3
2 1
2� Shield plate 3� Weighing pan
Note: After inserting the weighing pan, turn it slightly to the left and right,
while pressing it down lightly� 4� Internal draft shield (for MSE2�7S only) 5� Draft shield: Center the hole over the pan (see arrows)
Filter Balances (Models MSE6.6S-F/2.7S-F):
Placing Components on the Weigh Cell
t Place the components listed below inside the weigh cell in the order given:
1� MSE2,7S-F models: female connector 2� Shield plate 3� Internal draft shield 4� Filter weighing pan d 50 mm or weighing pan
(optional filter weighing pan d 75 mm or d 90 mm)
Note: After inserting the pan, turn it slightly to the left and right,
while pressing it down lightly� 5� Draft shield cover
!
Tip: Turn the balance off and then on again after replacing the pan during
operation�
Cubis MSE Operating Instructions 15
Getting Started
Setting up a filter weighing pan for left-handed users:
t Remove the draft shield cover
t Unscrew the pin and move from the right to the left side
t Turn draft shield parts approx� 90 degrees to the left (loosen knurled screw)

Power Connection

t Check the voltage rating and plug design�
– Make sure that the voltage rating printed on the manufacturer’s ID label is
identical to that of your local line voltage�
– If the stated supply voltage or the plug design of the power cord does not
comply with your country’s standard, please inform the nearest Sartorius representative or your dealer�
– The power connection must be made in accordance with the regulations
applicable in your country�
– In order to connect the equipment to the power supply (protection class 1), use
a suitable, correctly installed wall outlet with a protective grounding conductor (PE) and a fuse of a maximum 16 A�
– If a main feeder from the ceiling is required or if installation of a CEE plug is
necessary, these must be carried out by a qualified electrician�
– Only use original Sartorius AC adapters�
16 Cubis MSE Operating Instructions

Transporting the Balance

Transporting the Device over Short Distances
Exercise caution to avoid glass breakage�
3
Never lift and carry the balance by its draft shield�
Getting Started
t Carry as shown in the illustration�
or
t Carry as shown in the illustration�
Storage and Shipping Conditions
– Permitted storage and shipping temperature: -10 to +60 °C – Unpacked devices can lose their precision if subject to extreme vibrations� – Excessive vibrations may compromise the safety of the equipment� Save the packaging for any future storage or shipment of the balance� Only the original packaging provides optimum protection for the equipment� – Follow all warnings and safety precautions� – Follow the instructions in the chapter: “Transporting the Balance�"
Cubis MSE Operating Instructions 17
Getting Started
2
1
Analytical and precision balances:
1. Plug the DC supply cable of the AC adapter into the power socket of the
balance and tighten the threaded fitting�
Balances with a readability of 0�01 mg
1. Plug the DC supply cable of the AC adapter into the power socket of the
electronics module and tighten the threaded fitting�
2. Plug the other end of the power cord into the AC adapter�
3. Insert the AC adapter power plug into the wall outlet�
y The balance is now ready to use�
Safety Precautions
The output line of the adapter has a terminal (GND) connected to the metal housing of the balance� The data port is also galvanically linked to the balance housing (GND)�
18 Cubis MSE Operating Instructions
Getting Started
93
R
L
L
L
R
R
12
2
Leveling the Balance, Setting the Level Indicator
Leveling the balance compensates for slant or unevenness at the place
h
Leveling the Balance Manually
t The balance is leveled using both front leveling feet� t Screw in both back leveling feet (only for models with back leveling feet)� t Turn the two front feet as shown in the illustration until the air bubble is
y Normally, several leveling steps are required�
t Screw out both back leveling feet until they touch the setup surface
Warm-up Time
y In order to deliver the precise results described in the Technical Specifications,
of installation� The balance must be perfectly horizontal to ensure consistent, reproducible weighing results� All models are equipped with an electronic tilt angle detection feature� If the balance is not level, a warning symbol is shown on the display� Push-button automatic leveling is available on models with motorized leveling feet� For models with manual leveling feet, please follow the instructions on the display:
see “Leveling the Balance with the Inclination Sensor�"
centered within the circle of the level indicator�
(only for models with back leveling feet)�
the balance must warm up for at least 30 minutes after initial connection to
the power supply� Only after this time will the device have reached the required
operating temperature�
6
When a verified balance used in legal metrology (legal-for-trade
applications) is connected to the power, it must warm up for at least two hours before operation�
Cubis MSE Operating Instructions 19
Getting Started
Connecting Electronic Devices (Peripherals)
t Make absolutely sure that the device is unplugged from the power supply
before connecting/disconnecting any peripheral device (printer, scanner, PC) to or from the data port�
A device connected to the power supply should never be opened�
3

Anti-theft Locking Device (Accessory)

Balances with a readability of 0�01 mg
t An anti-theft locking device can be installed to the back of the balance if
required�
Balances with a readability of 1 µg To fasten an antitheft locking device, use the lug located on the back of the weigh cell�
20 Cubis MSE Operating Instructions
Getting Started
2
1
2
1

Modification Options

Setting Up the Display and Control Unit at the Place of Use
The display and control unit can be removed for all models to enable the operator to customize the work space�
Removing the Retainer with the Display and Control Unit
1
2
t Remove all items (such as weights) from the draft shield�
3
t Carefully remove all parts as shown in the illustration�
1. Weighing pan
2. Pan support (not for MSE225… MSE125… models)
3. Shield plate/Draft shield
t Keep all parts in a safe place�
t Remove the panels (right and left)�
Models with the analytical draft shield:
1. Press on the locking tab�
2. Remove the panel�
Models with the milligram draft shield:
1. Press on the locking tab�
2. Remove the panel�
t Keep all parts in a safe place�
Cubis MSE Operating Instructions 21
Getting Started
2
1
2
1
t Remove the upper draft shield panel�
Models with the analytical draft shield (left illustration)�
1. Press on the locking tab�
2. Remove the panel�
Models with the milligram draft shield (right figure):
1. Press on the locking tab�
2. Remove the panel�
t Keep all parts in a safe place�
t Turn over the balance and place it on a soft surface�
3
t Use a 2�5 mm Allen wrench to remove the two screws from the display and
t Remove the display and re-insert both screws back into their holes�
t Lengthen the cable and position the display and control unit as desired�
t Turn the balance over and place it on an even surface�
3
2
1
t Carefully place all parts on the balance:
1. Shield plate/Draft shield
2. Pan support (not for MSE225… /MSE125… models)
3. Weighing pan
Exercise caution to avoid breaking the glass on models with a draft
shield�
control unit retainer bracket�
22 Cubis MSE Operating Instructions
Getting Started
1
1. Upper draft shield panel
2. Right draft shield panel
3. Left draft shield panel
t Replace the upper and side shield panel�
3
t Level the balance�
Leveling the balance compensates for slant or unevenness at the place
h
2
of installation� The balance must be perfectly horizontal to ensure consistent, reproducible weighing results� All models are equipped with an electronic tilt angle detection feature� If the balance is not level, a warning symbol is shown on the display� The display shows a graphical representation of a level indicator and text prompts that guide the user through the leveling process: see “Leveling the Balance with the Inclination Sensor�"
R
L
t The balance is leveled using both front leveling feet�
t Screw in both back leveling feet (only for models with back leveling feet)�
L
t Turn the two front feet as shown in the illustration until the air bubble is
centered within the circle of the level indicator�
y Normally, several leveling steps are required�
R
1
2
t Screw out both back leveling feet until they touch the setup surface
(only for models with back leveling feet)�
Removing the Display and Control Unit from Precision Balances without a Draft Shield and with a Weighing Range of up to 15 kg
t Carefully remove all parts as shown in the illustration�
1. Weighing pan
3
2. Shield plate/Draft shield
3. Pan support
t Keep all parts in a safe place�
t Turn over the balance and place it on a soft surface�
Cubis MSE Operating Instructions 23
Getting Started
2
1
3
t Remove the two retaining screws�
t Remove the display and re-insert both screws back into their holes�
t Carefully pull the cable connected between the display and control unit from
the retainer�
t Determine the required cable length�
t Return the balance to an upright position and fit the parts onto the balance�
1. Attach the pan support�
2. Shield disk (only for models with a readability of 10 mg)
3. Weighing pan
t Level the balance�
Removing the Display and Control Unit from Models with a Weighing Range of 20 kg or more
Remove the weighing pan before unscrewing the control unit so that
3
t Turn over the balance so that the pan side is facing down� t Remove the two fixing screws using a screwdriver� t Remove the control unit and re-insert both screws into their holes� t Carefully remove the connection cable from its holder�
y Long connection cables should only be installed by a Sartorius technician�
the weighing pan does not fall and cause injury�
24 Cubis MSE Operating Instructions
Getting Started
Semi-microbalances: Attaching the Display and Control Unit to the Electronics Module (MSE225..., MSE125... models)
The display and control unit can also be attached to the electronics module if required for operation�
t Turn over the balance and place it on a soft surface�
Remove the connection cable from the cable channel:
t Remove two screws from beneath the weigh cell and detach the plate�
t Remove the connection cable plug�
t Then reattach the plate to the slot�
t Remove the display and control unit from the weigh cell:
Remove two retaining screws�
t Remove the display and control unit�
Cubis MSE Operating Instructions 25
Getting Started
t Attach the display and control unit to the electronics module:
Reattach the two retaining screws�
Open the slot for the connection cable on the electronics module:
t Remove the screws from beneath the electronics module and detach the plate�
Connect the display and control unit to the electronics module:
t Plug in the connection cable�
t Then reattach the plate to the slot�
t Slide the cable that sticks out into the cable channel�
Preparing Below-Cell Weighing
Balances with a readability of 0,01 mg:
1
A port for a below-cell weighing hook is located on the bottom of the balance�
Analysis and Precision Balances with a Weighing Range up to 15 kg:
1. Remove the hook for below-cell weighing from the clip on the bottom of the
balance�
26 Cubis MSE Operating Instructions
Getting Started
2. Push the cover of the below-cell weighing port to one side�
2
3. Carefully screw on the below-cell weighing hook�
Do not screw it in too tightly, as this could damage the thread
3
or the balance�
Install a draft protection shield�
h
t Attach the sample (e�g�, using a suspension wire) to the hook�
3
4. When weighing has been completed, unscrew the hook and return it to the clip�
t Close the cover of the below-cell weighing port�
Balances with a readability of 1 µg
t Remove both screws from beneath the weigh cell and detach the cover plate�
t Attach a wire, for example, to the sample and hang it on the notched hook�
Install a draft protection shield�
h
Balances with a Weighing Range of 20 kg or more:
t Use a suitable screwdriver to unscrew the cover plate from the bottom of the
balance�
t Request the hook directly from Sartorius� The hook is available as an accessory�
Install a draft protection shield�
h
The below-cell weighing port may not be opened or used on balances
used for legal metrology�
Cubis MSE Operating Instructions 27
Getting Started
Using the Cable Opening of the Manual Analytical Balance Draft Shield
Models with a manual analytical draft shield have an opening for passing a cable (for example, for a temperature sensor) through to the interior of the weighing chamber�
1. Lift the locking tab on the rear panel of the analytical draft shield�
2. Lift the panel out of the draft shield�
t Turn the panel clockwise 180° so that the opening is at the bottom�
t Install the desired sensor�
t Insert the panel in the guide rail�
t Lift the locking tab and gently press the panel into position�
t Press the locking tab down and close it�
t You can now begin weighing�
28 Cubis MSE Operating Instructions

Operation

Operation
22
21
3
20 19
18 17
16 15
14 13

Overview of Display and Control Panel

Position Description 1 Weight units
2 Calculated-value indicator: not a weight value 3 Tare 4
5 Symbol for “Printing mode active" 6 Symbol for “GLP printing mode active" 8 Symbol for “Application program active" 9 Data output: Press this key to send readout values
10 Level balance, open/close draft shield or turn on/
11 Start an application program 12 Select an application program | Access the menu 13 Symbols for an active application
symbol flashes: Balance is not level, leveling is required; n: open/close draft shield using key *; m: ionizer active
to the built-in data interfaces�
off ionizer
(Z, L, V, +, /, u, o)
1
2 3
4 5 6 7 8
9
10 11 12
Position Description 14 Clear Function
This key is generally used to cancel functions: – Quit application program
– Cancel calibration/adjustment routine |
Exit the menu – GLP-compliant printout 15 Start the calibration process 16 On/off Key 18 Display: “isoCAL:" Calibration/adjustment function 20 Symbols for zero range (verified models only) 21 Metrological Data 22 Weight value displayed in selected weight unit
Symbol: 19 tt Exit the menu 19 t Select previous menu level 17 q Select menu item 7 u Select next item in current menu level 7↵ Confirm menu item
Cubis MSE Operating Instructions 29
Operation

Basic Weighing Function

Features
– Tare the balance – Print weight value
Using a verified balance in legal metrology:
The Type-Approval Certificate for verified balances is only valid for non-automatic weighing instruments� For automatic operation with or without additional, integrated equipment, please follow the applicable national regulations for the installation location�
§ Before use in legal metrology, the balance should be adjusted via the internal
motorized calibration weight at the installation location: see the “Calibration and Adjustment" section in this chapter�
$
The temperature range (°C) listed on the ID label should not be exceeded during operation�
Example: MSX K +10/+30°C +5°C/+40°C isoCAL
Preparation
§ Turn balance on: Press the A key
§ Level the balance if required
§ Tare the balance if necessary: Press the J key
$ If necessary, change the configuration settings: see “Configuration" $ If desired, load the factory settings: see “Configuration"
Additional functions: $ Turn balance off: Press the A key
Quick Guide: First Weighing
Step Press key Display/Printout
1� Turn balance on A 0.0 g Self-test runs�
Then the balance performs an automatic initial tare�
2� Place container on weighing pan (in this example : 11�5 g)� + 11.5 g Close the draft shield (if available)�
3� Tare the balance J 0.0 g The balance is tared and the displayed value is zero�
4� If required, open the draft shield and carefully place the sample in the container on the weighing pan - in this example : 132 g� + 132.0 g Close the draft shield (if available)� You can read the measured value as soon as the weight value stops changing and the unit is displayed�
5� Print weight value P N + 132.0 g
30 Cubis MSE Operating Instructions