• Full end-to-end data integrity. 21 CFR Part 11 compliance,
integrated audit-trail, state-of-the-art user management
• Oustanding service support. Integrated status center, integrated
service functions preventive maintenance on accredited
standards
* For all models up to a maximum capacity of 8.2 kg
Product Information
The modular weighing system Cubis® II can be individually configured from different components. The combinability
of the display unit, weighing module, draft shield, software packages for various applications and functions and a
comprehensive range of accessories allows the individual adaptation of the Cubis® II balance to all weighing tasks.
The Cubis® II range of premium laboratory balances with a maximum load between 2.1 g and 70 kg and a readability
between 0.1 μg to 1 g provide the ideal model for every application.
Cubis® II Display and Control Units
TypeMCA
DisplayLarge high end 7” color touch TFT display in 16:9
format with new user interface
SoftwareFactory installed basic set of essential weighing
applications (license free) and Packages with
special weighing applications and function
extensions (license required)
Operation Activated by touch key, touch-free using IR sensor
(draft shield M) or gesture sensor (optional),
learning capability
TypeMCE
DisplayState of the art TFT touch screen operation with
brilliant, readable, display but uncomplex, easy to
operate user interface
SoftwareFactory installed basic set of essential weighing
applications
Operation Activated by touch key, touch-free using IR sensor
(draft shield M) or gesture sensor (optional),
learning capability
Cubis® II
Technical Specifications
Cubis® II Weighing Modules
Ultra-Micro Balances 0.0001 mg
2.7S
Scale interval (d)mg0.0001
Maximum capacity (Max)g2.1
Repeatability at 5 % load
Standard deviation of the load values, tolerancemg0.0002
Standard deviation of the load values, typical valuemg0.00015
Repeatability near Max
Standard deviation of the load values, tolerancemg0.00025
Standard deviation of the load values, typical valuemg0.00018
Linearity Deviation
Tolerancemg0.0009
Typical valuemg0.0007
Deviation at eccentric loading, positions according to OIML R76
Test weightg1
Tolerancemg0.0007
Typical valuemg0.0005
Sensitivity drift between +10 °C and +30 °Cppm/K1
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EUI
Verification scale interval (e) according to Directive 2014/31/EUmg1
Minimum load (Min) according to Directive 2014/31/EUmg0.01
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
External test loadg2
Accuracy class, acc. to OIML R111-1E2
Dimensions
MCE/MCA Weighing module (L × W × H)*mm340 x 139 x 129
MCE Electronics module (L × W × H)mm315 x 240 x 61
MCA Electronics module (L x W x H)mm355 x 240 x 61
Weighing pan sizemmØ 20
Filter weighing panmmØ 50
Weight, approx.*kg6.4/7.1
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II Weighing Modules
Micro Balances 0.001 mg
Deviation at eccentric loading, positions according to OIML R76
Test weightg521
Tolerancemg0.0040.0040.005
Typical valuemg0.0030.0030.003
Sensitivity drift between +10 °C and +30 °Cppm/K 111
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EUIII
Verification scale interval (e) according to Directive 2014/31/EU mg111
Minimum load (Min) according to Directive 2014/31/EUmg0.10.10.1
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
External test loadg1053
Accuracy class, acc. to OIML R111-1E2E2E2
Dimensions
MCE/MCA Weighing module (L × W × H)*mm340 x 139 x 129340 x 139 x 129340 x 139 x 129
MCE Electronics module (L × W × H)mm315 x 240 x 61315 x 240 x 61315 x 240 x 61
MCA Electronics module (L x W x H)mm355 x 260 x 61355 x 260 x 61355 x 260 x 61
Weighing pan sizemmØ 30Ø 30Ø 30
Filter weighing panmmØ 50Ø 50Ø 50
Weight, approx.*kg6.4/7.16.4/7.16.4/7.1
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II3
Cubis® II Weighing Modules
Semi-Micro Balances 0.01 mg
Deviation when load is off-center, positions according to OIML R76
Test weightg1001005050
Tolerancemg0.150.20.150.2
Typical valuemg0.10.10.10.1
Sensitivity drift between +10 °C and +30 °Cppm/K 1111
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EUIIII
Verification scale interval (e) according to Directive 2014/31/EU mg1111
Minimum load (Min) according to Directive 2014/31/EUmg1111
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
External test loadg200200100100
Accuracy class, acc. to OIML R111-1E2E2E2E2
Dimensions
MCE/MCA Weighing module (L × W × H)*mm404 x 240 x 373
MCE Electronics module (L × W × H)mm315 x 240 x 61
MCA Electronics module (L x W x H)mm355 x 240 x 61
Weighing pan sizemm85 x 85
Weight, approx.*kg10.2/11.7
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II Weighing Modules
Analytical Balances 0.1 mg
Deviation when load is off-center, positions according to OIML R76
Test weightg20020020020010050
Tolerancemg0.30.40.30.40.20.2
Typical valuemg0.20.20.20.20.120.12
Sensitivity drift between +10 °C and +30 °Cppm/K 111111
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive
2014/31/EU
Verification scale interval (e) according to
Directive 2014/31/EU
Minimum load (Min) according to Directive
2014/31/EU
mg111111
mg101010101010
IIIIII
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
External test loadg5005002001400014000
Accuracy class, acc. to OIML R111-1E2E2E2E2E2
Dimensions
Weighing module (L × W × H)*mm425 x 240 x 122/284/373425 x 240 x 95
Weighing pan sizemm140 x 140 206 x 206
Weight, approx.*kg5.9/7.5/9.4/10.25.4
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II7
Cubis® II Weighing Modules
Precision Balances
10202S 8202S6202S6202P5202S4202S2202S
Scale interval (d)mg10101010 | 20 | 50101010
Maximum capacity (Max)g1020082006200
1500 | 3000
|6200
520042002200
Repeatability at 5 % load
Standard deviation of the load values,
tolerance
Standard deviation of the load values,
typical value
mg7777677
mg5444244
Repeatability near Max
Standard deviation of the load values,
tolerance
Standard deviation of the load values,
typical value
mg77740677
mg54415244
Linearity Deviation
Tolerancemg20202050102020
Typical valuemg66620566
Deviation when load is off-center, positions according to OIML R76
Test weightg5000500020002000200020001000
Tolerancemg20202050103020
Typical valuemg1010103051010
Sensitivity drift between +10 °C and +30 °C ppm/K 1.5222222
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive
2014/31/EU
Verification scale interval (e) according to
Directive 2014/31/EU
Minimum load (Min) according to Directive
2014/31/EU
mg100100100100100100100
mg10005005005001000500500
IIIIIIIIIIIII
Minimum weight according to USP (United States Pharmacopeia), Chap. 41