Sartorius MCA324S-2S00-U, MCA3203SE-S00, MCA324S-2S00-I, MCA4202S-2S00-0, MCA36201S0-S00 Data Sheet

...
Cubis® II
Highlights
• Leading performance. Monolithical weighing system, integrated climate sensors, individual sample holders
• Error-free operation. Individual QApp workflows, motorized auto-leveling*
• Full end-to-end data integrity. 21 CFR Part 11 compliance, integrated audit-trail, state-of-the-art user management
• Oustanding service support. Integrated status center, integrated service functions preventive maintenance on accredited standards
* For all models up to a maximum capacity of 8.2 kg
Product Information
The modular weighing system Cubis® II can be individually configured from different components. The combinability of the display unit, weighing module, draft shield, software packages for various applications and functions and a comprehensive range of accessories allows the individual adaptation of the Cubis® II balance to all weighing tasks. The Cubis® II range of premium laboratory balances with a maximum load between 2.1 g and 70 kg and a readability between 0.1 μg to 1 g provide the ideal model for every application.
Cubis® II Display and Control Units
Type MCA Display Large high end 7” color touch TFT display in 16:9
format with new user interface
Software Factory installed basic set of essential weighing
applications (license free) and Packages with special weighing applications and function extensions (license required)
Operation Activated by touch key, touch-free using IR sensor
(draft shield M) or gesture sensor (optional), learning capability
Type MCE Display State of the art TFT touch screen operation with
brilliant, readable, display but uncomplex, easy to operate user interface
Software Factory installed basic set of essential weighing
applications
Operation Activated by touch key, touch-free using IR sensor
(draft shield M) or gesture sensor (optional), learning capability
Cubis® II
Technical Specifications
Cubis® II Weighing Modules Ultra-Micro Balances 0.0001 mg
2.7S
Scale interval (d) mg 0.0001 Maximum capacity (Max) g 2.1
Repeatability at 5 % load
Standard deviation of the load values, tolerance mg 0.0002 Standard deviation of the load values, typical value mg 0.00015
Repeatability near Max
Standard deviation of the load values, tolerance mg 0.00025 Standard deviation of the load values, typical value mg 0.00018
Linearity Deviation
Tolerance mg 0.0009 Typical value mg 0.0007
Deviation at eccentric loading, positions according to OIML R76
Test weight g 1 Tolerance mg 0.0007 Typical value mg 0.0005 Sensitivity drift between +10 °C and +30 °C ppm/K 1
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU I Verification scale interval (e) according to Directive 2014/31/EU mg 1 Minimum load (Min) according to Directive 2014/31/EU mg 0.01
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 0.082 Typical minimum weight mg 0.3 Typical stabilization time s 7 Typical measurement time s 10
Recommended calibration weight
External test load g 2 Accuracy class, acc. to OIML R111-1 E2
Dimensions
MCE/MCA Weighing module (L × W × H)* mm 340 x 139 x 129 MCE Electronics module (L × W × H) mm 315 x 240 x 61 MCA Electronics module (L x W x H) mm 355 x 240 x 61 Weighing pan size mm Ø 20 Filter weighing pan mm Ø 50 Weight, approx.* kg 6.4/7.1
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II Weighing Modules Micro Balances 0.001 mg
10.6S 6.6S 3.6P
Scale interval (d) mg 0.001 0.001 0.001 | 0.002 | 0.005 Maximum capacity (Max) g 10.1 6.1 1.1 | 2.1 | 3.1
Repeatability at 5 % load
Standard deviation of the load values, tolerance mg 0.001 0.001 0.003 Standard deviation of the load values, typical value mg 0.0005 0.0005 0.0005
Repeatability near Max
Standard deviation of the load values, tolerance mg 0.001 0.001 0.005 Standard deviation of the load values, typical value mg 0.0006 0.0006 0.0006
Linearity Deviation
Tolerance mg 0.004 0.004 0.004 Typical value mg 0.003 0.003 0.003
Deviation at eccentric loading, positions according to OIML R76
Test weight g 5 2 1 Tolerance mg 0.004 0.004 0.005 Typical value mg 0.003 0.003 0.003 Sensitivity drift between +10 °C and +30 °C ppm/K 1 1 1
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU I I I Verification scale interval (e) according to Directive 2014/31/EU mg 1 1 1 Minimum load (Min) according to Directive 2014/31/EU mg 0.1 0.1 0.1
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 0.82 0.82 0.82 Typical minimum weight mg 0.82 0.82 0.82 Typical stabilization time s 5 5 5 Typical measurement time s 8 8 8
Recommended calibration weight
External test load g 10 5 3 Accuracy class, acc. to OIML R111-1 E2 E2 E2
Dimensions
MCE/MCA Weighing module (L × W × H)* mm 340 x 139 x 129 340 x 139 x 129 340 x 139 x 129 MCE Electronics module (L × W × H) mm 315 x 240 x 61 315 x 240 x 61 315 x 240 x 61 MCA Electronics module (L x W x H) mm 355 x 260 x 61 355 x 260 x 61 355 x 260 x 61 Weighing pan size mm Ø 30 Ø 30 Ø 30 Filter weighing pan mm Ø 50 Ø 50 Ø 50 Weight, approx.* kg 6.4/7.1 6.4/7.1 6.4/7.1
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II 3
Cubis® II Weighing Modules Semi-Micro Balances 0.01 mg
225S 225P 125S 125P
Scale interval (d) mg 0.01 0.01 | 0.02 | 0.05 0.01 0.01 | 0.1 Maximum capacity (Max) g 220 60 | 120 | 220 120 60 | 120
Repeatability at 5 % load
Standard deviation of the load values, tolerance mg 0.015 0.015 0.015 0.015 Standard deviation of the load values, typical value mg 0.01 0.01 0.01 0.01
Repeatability near Max
Standard deviation of the load values, tolerance mg 0.025 0.04 0.025 0.06 Standard deviation of the load values, typical value mg 0.02 0.02 0.02 0.02
Linearity Deviation
Tolerance mg 0.1 0.15 0.1 0.15 Typical value mg 0.065 0.1 0.065 0.1
Deviation when load is off-center, positions according to OIML R76
Test weight g 100 100 50 50 Tolerance mg 0.15 0.2 0.15 0.2 Typical value mg 0.1 0.1 0.1 0.1 Sensitivity drift between +10 °C and +30 °C ppm/K 1 1 1 1
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU I I I I Verification scale interval (e) according to Directive 2014/31/EU mg 1 1 1 1 Minimum load (Min) according to Directive 2014/31/EU mg 1 1 1 1
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 8.2 8.2 8.2 8.2 Typical minimum weight mg 13 13 13 13 Typical stabilization time s 2 2 2 2 Typical measurement time s 6 6 6 6
Recommended calibration weight
External test load g 200 200 100 100 Accuracy class, acc. to OIML R111-1 E2 E2 E2 E2
Dimensions
MCE/MCA Weighing module (L × W × H)* mm 404 x 240 x 373 MCE Electronics module (L × W × H) mm 315 x 240 x 61 MCA Electronics module (L x W x H) mm 355 x 240 x 61 Weighing pan size mm 85 x 85 Weight, approx.* kg 10.2/11.7
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II Weighing Modules Analytical Balances 0.1 mg
524S 524P 324S 324P 224S 124S
Scale interval (d) mg 0.1 0.1 | 0.2 | 0.5 0.1 0.1 | 0.2 | 0.5 0.1 0.1 Maximum capacity (Max) g 520 120 | 240 | 520 320 80 | 160 | 320 220 120
Repeatability at 5 % load
Standard deviation of the load values, tolerance mg 0.08 0.08 0.08 0.08 0.07 0.1 Standard deviation of the load values, typical
value
mg 0.04 0.04 0.04 0.04 0.05 0.05
Repeatability near Max
Standard deviation of the load values, tolerance mg 0.1 0.15 0.1 0.1 0.07 0.1 Standard deviation of the load values, typical
value
mg 0.05 0.05 0.05 0.05 0.05 0.05
Linearity Deviation
Tolerance mg 0.4 0.5 0.3 0.5 0.2 0.2 Typical value mg 0.2 0.2 0.2 0.2 0.13 0.13
Deviation when load is off-center, positions according to OIML R76
Test weight g 200 200 200 200 100 50 Tolerance mg 0.3 0.4 0.3 0.4 0.2 0.2 Typical value mg 0.2 0.2 0.2 0.2 0.12 0.12 Sensitivity drift between +10 °C and +30 °C ppm/K 1 1 1 1 1 1
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU
Verification scale interval (e) according to Directive 2014/31/EU
Minimum load (Min) according to Directive 2014/31/EU
mg 1 1 1 1 1 1
mg 10 10 10 10 10 10
I I I I I I
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 82 82 82 82 82 82 Typical minimum weight mg 82 82 82 82 100 100 Typical stabilization time s 1 1 1 1 1 1 Typical measurement time s 3 3 3 3 3 3
Recommended calibration weight
External test load g 500 500 300 300 200 100 Accuracy class, acc. to OIML R111-1 E2 E2 E2 E2 E2 E2
Dimensions
Weighing module (L × W × H)* mm 425 x 240 x 373 Weighing pan size mm 85 x 85 Weight, approx.* kg 8.2/10.0
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II 5
Cubis® II Weighing Modules Precision Balances
5203S 5203P 3203S 2203S 2203P 1203S
Scale interval (d) mg 1 1 | 2 | 5 1 1 1 | 10 1 Maximum capacity (Max) g 5200
1200 | 2400 | 5200
3200 2200 1010 | 2200 1200
Repeatability at 5 % load
Standard deviation of the load values, tolerance
Standard deviation of the load values, typical value
mg 1 1 1 0.7 0.7 0.7
mg 0.6 0.6 0.6 0.5 0.5 0.5
Repeatability near Max
Standard deviation of the load values, tolerance
Standard deviation of the load values, typical value
mg 1 1 1 1 1 0.7
mg 0.6 0.6 0.6 0.6 0.6 0.6
Linearity Deviation
Tolerance mg 5 5 5 3 5 2 Typical value mg 2 3 2 2 3 1
Deviation when load is off-center, positions according to OIML R76
Test weight g 2000 2000 1000 1000 1000 500 Tolerance mg 2 2 2 2 3 2 Typical value mg 1 1 1 1 2 1 Sensitivity drift between +10 °C and +30 °C ppm/K 1 1 1 1 1 1.5
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU
Verification scale interval (e) according to Directive 2014/31/EU
Minimum load (Min) according to Directive 2014/31/EU
mg 10 10 10 10 10 10
mg 100 100 100 100 100 100
I I I I I I
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 820 820 820 820 820 820 Typical minimum weight mg 1200 1200 1200 1000 1000 1000 Typical stabilization time s 1 1 1 1 1 1 Typical measurement time s 2 2 2 1.5 1.5 1.5
Recommended calibration weight
External test load g 5000 5000 3000 2000 1000 1000 Accuracy class, acc. to OIML R111-1 E2 E2 E2 E2 E2 E2
Dimensions
Weighing module (L × W × H)* mm 425 x 240 x 122/284/373 Weighing pan size mm 140 x 140 Weight, approx.* kg 5.9/7.5/9.4/10.2
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II Weighing Modules Precision Balances
623S 623P 323S 14202S 14202P
Scale interval (d) mg 1 1 / 2 / 5 1 10 10 | 20 | 50 Maximum capacity (Max) g 620 150 | 300 | 620 320 14200 3500 | 7000 | 14200
Repeatability at 5 % load
Standard deviation of the load values, tolerance
Standard deviation of the load values, typical value
mg 0.7 1 0.7 10 10
mg 0.4 0.4 0.4 5 5
Repeatability near Max
Standard deviation of the load values, tolerance
Standard deviation of the load values, typical value
mg 0.7 1 0.7 10 10
mg 0.5 0.5 0.5 5 5
Linearity Deviation
Tolerance mg 2 5 2 30 50 Typical value mg 0.6 1.5 0.6 10 20
Deviation when load is off-center, positions according to OIML R76
Test weight g 200 200 200 5000 5000 Tolerance mg 2 4 2 20 40 Typical value mg 1 3 1 10 10 Sensitivity drift between +10 °C and +30 °C ppm/K 2 2 2 1.5 1.5
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU
Verification scale interval (e) according to Directive 2014/31/EU
Minimum load (Min) according to Directive 2014/31/EU
mg 10 10 10 100 100
mg 20 20 20 1000 1000
II II II II II
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 820 820 820 8200 8200 Typical minimum weight mg 820 820 820 8200 8200 Typical stabilization time s 0.8 0.8 0.8 0.8 0.8 Typical measurement time s 1 1 1 1.5 1.5
Recommended calibration weight
External test load g 500 500 200 14000 14000 Accuracy class, acc. to OIML R111-1 E2 E2 E2 E2 E2
Dimensions
Weighing module (L × W × H)* mm 425 x 240 x 122/284/373 425 x 240 x 95 Weighing pan size mm 140 x 140 206 x 206 Weight, approx.* kg 5.9/7.5/9.4/10.2 5.4
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II 7
Cubis® II Weighing Modules Precision Balances
10202S 8202S 6202S 6202P 5202S 4202S 2202S
Scale interval (d) mg 10 10 10 10 | 20 | 50 10 10 10
Maximum capacity (Max) g 10200 8200 6200
1500 | 3000 |6200
5200 4200 2200
Repeatability at 5 % load
Standard deviation of the load values, tolerance
Standard deviation of the load values, typical value
mg 7 7 7 7 6 7 7
mg 5 4 4 4 2 4 4
Repeatability near Max
Standard deviation of the load values, tolerance
Standard deviation of the load values, typical value
mg 7 7 7 40 6 7 7
mg 5 4 4 15 2 4 4
Linearity Deviation
Tolerance mg 20 20 20 50 10 20 20 Typical value mg 6 6 6 20 5 6 6
Deviation when load is off-center, positions according to OIML R76
Test weight g 5000 5000 2000 2000 2000 2000 1000 Tolerance mg 20 20 20 50 10 30 20 Typical value mg 10 10 10 30 5 10 10 Sensitivity drift between +10 °C and +30 °C ppm/K 1.5 2 2 2 2 2 2
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU
Verification scale interval (e) according to Directive 2014/31/EU
Minimum load (Min) according to Directive 2014/31/EU
mg 100 100 100 100 100 100 100
mg 1000 500 500 500 1000 500 500
II II II II I II II
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 8200 8200 8200 8200 8200 8200 8200 Typical minimum weight mg 8200 8200 8200 8200 8200 8200 8200 Typical stabilization time s 0.8 1 1 1 0.8 1 0.8 Typical measurement time s 1.5 1.5 1.5 1.5 1 1 1.5
Recommended calibration weight
External test load g 10000 7000 5000 5000 5000 3000 1500 Accuracy class, acc. to OIML R111-1 E2 E2 E2 E2 E2 E2 E2
Dimensions
Weighing module (L × W × H)* mm 425 x 240 x 95
425 x 240 x 122/284/373
425 x 240 x 95
Weighing pan size mm 206 x 206 140 x 140 206 x 206
Weight, approx.* kg 5.4 5.9/ 7.5/9.4/10.2 5.4
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II Weighing Modules Precision Balances
1202S 12201S 8201S 5201S
Scale interval (d) mg 10 100 100 100 Maximum capacity (Max) g 1200 12200 8200 5200
Repeatability at 5 % load
Standard deviation of the load values, tolerance mg 7 50 50 50 Standard deviation of the load values, typical value mg 4 20 20 20
Repeatability near Max
Standard deviation of the load values, tolerance mg 7 50 50 50 Standard deviation of the load values, typical value mg 4 20 20 20
Linearity Deviation
Tolerance mg 20 100 100 100 Typical value mg 6 30 30 20
Deviation when load is off-center, positions according to OIML R76
Test weight g 500 5000 2000 2000 Tolerance mg 20 200 200 200 Typical value mg 10 100 100 100 Sensitivity drift between +10 °C and +30 °C ppm/K 2 4 4 4
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU II II II II Verification scale interval (e) according to Directive 2014/31/EU mg 100 1000 1000 1000 Minimum load (Min) according to Directive 2014/31/EU mg 500 5000 5000 5000
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 8200 82000 82000 82000 Typical minimum weight mg 8200 82000 82000 82000 Typical stabilization time s 0.8 0.8 0.8 0.8 Typical measurement time s 1 1 1 1
Recommended calibration weight
External test load g 700 12000 8000 5000 Accuracy class, acc. to OIML R111-1 E2 F1 F1 F1
Dimensions
Weighing module (L × W × H)* mm 425 x 240 x 95 Weighing pan size mm 206 x 206 Weight, approx.* kg 5.4
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II 9
Cubis® II Weighing Modules High Capacity Balances
32202P 70201S 50201S 36201S 36201P
Scale interval (d) mg 10 | 100 100 100 100 100|1000 Maximum capacity (Max) g 4200 | 32200 70200 50200 36200 10200 | 36200
Repeatability at 5 % load
Standard deviation of the load values, tolerance mg 40 100 100 100 100 Standard deviation of the load values, typical value mg 20 40 40 20 20
Repeatability near Max
Standard deviation of the load values, tolerance mg 40 | 100 100 100 100 100 Standard deviation of the load values, typical value mg 20 | 50 40 40 50 20
Linearity Deviation
Tolerance mg 20 500 500 200 200 Typical value mg 10 150 150 100 100
Deviation when load is off-center, positions according to OIML R76
Test weight g 10000 20000 20000 10000 10000 Tolerance mg 200 500 500 300 300 Typical value mg 100 150 150 200 200 Sensitivity drift between +10 °C and +30 °C ppm/K 2 4 4 4 2
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU II II II II Verification scale interval (e) according to Directive
2014/31/EU Minimum load (Min) according to Directive
2014/31/EU
mg 1000 1000 1000 1000
mg 5000 5000 5000 5000
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 8200 82000 82000 82000 82000 Typical minimum weight mg 8200 82000 82000 82000 82000 Typical stabilization time s 2 1.5 1.5 1.5 1.5 Typical measurement time s 2 1.5 1.5 2 2
Recommended calibration weight
External test load g 30000 70000 50000 30000 30000 Accuracy class, acc. to OIML R111-1 F1 F1 F1 F1 F1
Dimensions
Weighing module (L × W × H)* mm 412 x 400 x 126/159 Weighing pan size mm 400 x 300/Ø 233 Weight, approx.* kg 15.8/17.1
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II Weighing Modules High Capacity Balances
20201S 11201S 70200S 36200S
Scale interval (d) mg 100 100 1000 1000 Maximum capacity (Max) g 20200 11200 70200 36200
Repeatability at 5 % load
Standard deviation of the load values, tolerance mg 100 100 500 500 Standard deviation of the load values, typical value mg 20 20 20 20
Repeatability near Max
Standard deviation of the load values, tolerance mg 100 100 500 500 Standard deviation of the load values, typical value mg 20 20 20 20
Linearity Deviation
Tolerance mg 200 200 1000 1000 Typical value mg 60 60 200 200
Deviation when load is off-center, positions according to OIML R76
Test weight g 5000 5000 20000 10000 Tolerance mg 300 300 1000 1000 Typical value mg 200 200 600 500 Sensitivity drift between +10 °C and +30 °C ppm/K 2 2 3 3
Tare maximum capacity: Less than 100 % of maximum capacity
Accuracy class according to Directive 2014/31/EU II II II II Verification scale interval (e) according to Directive 2014/31/EU mg 1000 1000 10000 1000 Minimum load (Min) according to Directive 2014/31/EU mg 5000 5000 50000 50000
Minimum weight according to USP (United States Pharmacopeia), Chap. 41
Optimum minimum weight mg 82000 82000 820000 820000 Typical minimum weight mg 82000 82000 820000 820000 Typical stabilization time s 1.5 1.5 1 1 Typical measurement time s 2 2 1.2 1.2
Recommended calibration weight
External test load g 20000 10000 70000 30000 Accuracy class, acc. to OIML R111-1 F1 F1 F1 F1
Dimensions
Weighing module (L × W × H)* mm 412 x 400 x 126/159 Weighing pan size mm 400 x 300/Ø 233 Weight, approx.* kg 15.8/17.1
* depending upon weighing pan size, filter weighing pan and draft shield
Cubis® II 11
Accessories
Cubis® II Power Supply Unit
Only by Sartorius AC Adapter YEPS03-15V0 Unit Value
Primary
AC voltage V Frequency Hz Current consumption, maximum A 1.0
Secondary
DC voltage at 2 A output current V 14.25–15.75 Power, maximum W 30 Short circuit protection: Electronic Protection class according to IEC 62368-1 I Pollution level according to IEC 61010-1 2 Overvoltage category according to IEC 606641-1 II Installation site according to IEC 62368-1, maximum altitude above sea level m 5000
Temperature
In operation °C 0 – +40 During storage and transport °C -20 – +80 Connector for mains connection according to IEC 60320-1 / C14: 3-pin
100–240 (±10%) 50–60 (±5%)
Power supply cable
Power supply cable according to IEC 60320-1 / C14: Country-specific, 3-pin, two-sided plug Other data: See label on AC adapter
Safety of Electrical Equipment
According to EN 61010-1 / IEC 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General Requirements
Electromagnetic Compatibility
Interference resistance
Suitable for use in industrial areas
Transient emissions
Class B Suitable for use in residential areas and areas that are connected to a low voltage network that also supplies residential buildings.
Materials
Housing: Die-cast aluminum, plastic PBT, Optiwhite float glass and stainless steel 1.4401/1.4404, PA handles, aluminum trim Control unit: Die-cast aluminum, painted, float glass and plastic PBT, PP
Integrated Clock
Maximum deviation per month (RTC): 30s
Backup Battery
Lithium battery: type CR2032 Service life at room temperature, minimum: 10 Years
Alibi Memory Value
Maximum number of data records: 150,000
Interfaces
Specifications for the COM-RS232 Interface
Type of interface: Serial interface Interface operation: Full duplex Level: RS232 Connection: D-sub connector, 9-pin
Pin assignment: Pin 1: Not assigned
Pin 2: Data output (TxD) Pin 3: Data input (RxD)
1
6
Specifications for the USB-A Interface
Communication: USB host (master) Connectable devices: Sartorius printers, USB sticks with software update
Specifications for the USB-B Interface
Communication: USB device (slave) Type of interface: Virtual serial interface (virtual COM-port, VCP) and “PC direct” communication
5
9
Pin 4: Not assigned Pin 5: Internal ground Pin 6: Not assigned Pin 7: Clear to Send (CTS) Pin 8: Request to Send (RTS) Pin 9: Not assigned
Cubis® II 13
Ambient Conditions
Unit Value
Installation Site
AC voltage Frequency
V Hz
100–240 (±10%) 50–60 (±5%)
Current consumption, maximum A 1.0
Secondary
Standard laboratory rooms Installation site according to IEC 60259-1, maximum altitude above sea level m 3000 For indoor use only
Temperature
In operation °C +5 – +40 In operation for conformity-assessed devices: See information on the device’s ID plate During storage and transport °C -20 – +60
Relative humidity
At temperatures of up to 31 °C % 80 Then linear decrease from 80 % at 31 °C to 50 % at 40 °C
No heat from heating systems or direct sunlight No drafts from open windows, AC systems, or doors No vibrations No “heavy traffic” areas (personnel) No electromagnetic fields No dry air
Meteorological Data
Code Item
SØØ Standard version non-verified, all units S01 Standard version non-verified, metric units only CCN Balance with Type Approval Certificate for China CEU Verified balance with EC Type Approval Certificate (for EU except France) CFR Verified balance with EC Type Approval Certificate for France only OBR Balance with Type Approval Certificate for Brazil OIN Balance with Type Approval Certificate for India OJP Balance with Type Approval Certificate for Japan ORU Balance with Type Approval Certificate for Russia
Draft Shields
Code Item
0 Flat, stainless steel weight pan with no draft shield for weighing modules. A Automatic, glass motorized draft shield with learning capability for user-friendly operation and easy customization to
the changing requirements of different applications.
E Manual glass draft shield for precision balances. F Manual stainless steel draft shield for weighing filters with diameters of up to 50 mm (75 mm and 90 mm pans
optional).
I Identical to the A draft shield, but also includes an integrated ionizer to eliminate interfering electrostatic charges on
samples and sample containers.
M Automatic, motorized, round 100% glass draft shield with learning capability for ultra-microbalance and microbalances. R Flat, stainless steel weighing pan draft shield (removable, with no glass components) for all precision balances U Manual glass analytical draft shield chamber, with smooth-action doors that open wide and provide unimpeded access
to the weighing chamber without interfering braces.
Software
Code Item
QP1 Q-App Package Pharma QP2 Q-App Package Advanced Applications QP3 Q-App Package Utilities QP4 Q-App Package Connectivity
Accessories
Printers and Communication Quantity Order number
Thermal transfer | thermal printer for GLMP printouts on continuous paper and labels 1 YDP30 Standard paper and ink ribbon, set, 90 m, for YDP30 1 69Y03285 Self-adhesive paper and ink ribbon, 90 m, for YDP30 1 69Y03286 Standard thermal paper, 24 m roll, for YDP30 | YDP40 5 69Y03287 Self-adhesive thermal paper, 13 m roll, for YDP30 5 69Y03288 Self-adhesive labels for YDP30 58 mm × 100 mm 350 69Y03094 58 mm × 76 mm 500 69Y03093 58 mm × 30 mm 1000 69Y03092 Display cable, 3 m, for separate installation of display and weighing units, installation by Sartorius
1 On request
Service or at the factory RS232C connection cable, 9-pin, 3 m, for connection to a PC with 9-pin COM interface 1 On request Sartorius Wedge, software for data communication between the PC and balance 1 YSW02
Cubis® II 15
Accessories (continued)
Displays and Input/Output Elements Quantity Order number
MCE Display and control unit with color TFT display, touch screen, and keys 1 69ME0212 MCA Display and control unit with high-contrast display, touch screen, and keys 1 69MA0218 Motion sensor for triggering a maximum of 4 functions via gesture control, selection via menu 1 YHS02MS Display stand for weigh cells with scale interval of 100 mg | 1 g and weighing capacity > 20 kg for
raising the operating unit Display stand for weigh cells with scale interval of 10 mg | 100 mg for raising the operating unit 1 YDH03MS
Hardware for Pipette Calibration (Semi-Micro and Analytical balances) Quantity Order number
Pipette calibration kit (hardware). Consists of moisture trap and all required adapters 1 YCP04MS
Density Determination Kits Quantity Order number
Density determination kit for solids and liquids for 0.1 and 0.01 mg weighing modules 1 YDK03MS Density determination kit for solids and liquids for 1 mg weighing modules 1 YDK04MS
Filter Pans, Ionizer and Weighing Scoops Quantity Order number
Grid pan for model with a scale interval of 10 mg or 100 mg for weighing in laboratory hoods, safety barries and workbenches, reduced wind attack surface of the weighing pan, replaces standard pan
Anti-static weighing pan, 100 mm diameter, for weighing module for semi-mnicrobalance and analytical balances with 0.1 mg or 0.01 mg scale interval
Filter weighing pan made of Titanium, diameter 52 mm, for ultra-micro and micro-balances only together with F draft shield
Filter weighing pan made of Titanium, diameter 75 mm, for ultra-micro balance or micro balance models only together with F draft shield
Filter weighing pan made of Titanium, diameter 90 mm, for ultra-micro balance or micro balance models only together with F draft shield
Ionization blower for electrostatically charged samples 1 YIB01-ODR Ionizer with u-shaped electrode for 230 V 1 YIB02-230V Ionizer with u-shaped electrode for 115 V 1 YIB02-115V Stat-Pen ionization pen for discharging electrostatically charged samples 1 YSTP01 Aluminum weighing scoop, 4.5 mg for ultramicrobalance and microbalance models 250 6565-250 Aluminum weighing scoop, 52 mg for ultramicrobalance and microbalance models 50 6566-50 Weighing scoop made from chrome-nickel steel, L 90 mm x W 32 mm x H 8 mm 1 641214
1 YDH04MS
1 YWP07MS
1 YWP04MS
1 YSH34
1 YSH35
1 YSH36
Other Accessories Quantity Order number
Below-balance weighing hook for precision balances with scale interval of 100 mg | 1 g and weighing capacity > 20 kg, not for verified models
Weighing Tables Quantity Order number
Made from synthetic stone, with vibration dampening 1 YWT03 Made from wood with synthetic stone 1 YWT09 Wall console 1 YWT04
1 69EA0040
Accessories (continued)
Climate Modules (available July 2019) Quantity Order number
Climate module, uncalibrated, for draft shield A or I and user interface MCA 1 YCM20MC Calibration of a climate module YCM20MC with DAkkS calibration certificate 1 YCM20DAkkS Climate module with DAkkS calibration certificate for draft shield A or I and user interface MCA 1 YCM20MC-DAkkS Tower for climate module, for mounting YCM20MC; can be ported to all Cubis II weighing modules
with user interface MCA
Flexible Sample Holder (available July 2019) Quantity Order number
For weighing vessels and filters with diameters of up to 120 mm, replaces the original weighing pan, for semi-microbalance and analytical balance
Titanium Holders for Ultra-Micro Balances and Micro Balances (available July 2019) Quantity Order number
For coronary stents (up to 38 mm) 1 YSH10
For save-lock tubes, 1.5 ml - 2 ml 1 YSH13
Titanium Holders for Semi-Micro Balances and Analytical Balances (available July 2019) Quantity Order number
Q-Grid weighing pan for 10 mg or 100 mg balances 1 YWP07MS For save-lock tubes, 1.5 ml – 2 ml
1 YCM20MC Tower
1 YFH01MS
1
YSH15
For save-lock tubes up to 5 ml
For vials
For weighing boats
For filters, 150 mm diameter
For titration vessels and round bottom flasks
For syringes, vertical
1
1
1
1
1
1
YSH19
YSH23
YSH26
YSH30
YSH37
YSH46
Cubis® II 17
Balance Dimensions
Ultra-Micro Balance
All dimensions are given in millimeters
129
121
80
67
35
50
70
244
110
139
246
MCA
24 0
211
MCE
211
315
355
240
Micro Balance
84
All dimensions are given in millimeters
345
246
109
115.5
MCA
240
211
355
129
94
78
35
50
70
244
19
110
139
MCE
211
315
240
Cubis® II 19
Semi-Micro Balance
MCE
All dimensions are given in millimeters
252
167
206
360
MCA
261
373
56
170
217
211
240
450
300
206
173
211
240
92
261
373
360
125
128
210
404
217
300
56
211
211
240
173
240
92
Analytical Balance
MCE
All dimensions are given in millimeters
252
167
206
MCA
373
359
240
205
261
92
170
125
210
217
470
105
240
320
206
211
240
240
205
261
373
359
128
425
217
320
105
240
211
240
92
Cubis® II 21
Precision Balance
MCA
MCE
All dimensions are given in millimeters
214.5 206
MCA
214.5 206
95
272
470
214.5 206
320
213173
105
240
211
240
249
MCE
214.5 206
95
128
227
425
213
320
211
240
249
105
240
High-Capacity Balance
All dimensions are given in millimeters
126 ±1
365
400
240
MCA
33 265
31
16
300
±2
444
400
240
31
16
300
412 ±2
MCE
Cubis® II 23
Sales and Service Contacts
For further contacts, visit www.sartorius.com
Europe
Germany
Sartorius Lab Instruments GmbH & Co. KG Otto-Brenner-Strasse 20 37079 Goettingen
Phone +49.551.308.0 Fax +49.551.308.3289
France & Suisse Romande
Sartorius France 2, rue Antoine Laurent de Lavoisier ZA de la Gaudrée 91410 Dourdan
Phone +33.1.70.62.50.00 Fax +33.1.64.59.76.39
Austria
Sartorius Austria GmbH Modecenterstrasse 22 1030 Vienna
Phone +43.1.7965760.0 Fax +43.1.7965760.24
Belgium
Sartorius Belgium N.V. Rue Colonel Bourg 105 1030 Bruxelles
Phone +32.2.756.06.90 Fax +32.2.481.84.11
Finland & Baltics
Sartorius Biohit Liquid Handling Oy Laippatie 1 00880 Helsinki
Phone +358.9.755.951 Fax +358.9.755.95.200
Hungary
Sartorius Hungária Kft. Kagyló u. 5. 2092 Budakeszi
Phone +3623.457.227 Fax +3623.457.147
Ireland
Sartorius Ireland Ltd. Unit 41, The Business Centre Stadium Business Park Ballycoolin Road Dublin 11
Phone +353.1.8089050 Fax +353.1.8089388
Italy
Sartorius Italy S.r.l. Via Torino 3/5 20814 Varedo (MB)
Phone +39 0362 5557.11 Fax +39 0362 5557.99
 www.sartorius.com
Netherlands
Sartorius Netherlands B.V. Phone +31.30.60.53.001
Fax +31.30.60.52.917 info.netherlands@sartorius.com
Poland
Sartorius Poland sp.z o.o. ul. Wrzesinska 70 62-025 Kostrzyn
Phone +48.61.6473830 Fax +48.61.6473839
Russian Federation
LLC “Sartorius RUS” Vasilyevsky Island
th
5
line 70, Lit. A
199178 St. Petersburg Phone +7.812.327.53.27
Fax +7.812.327.53.23
Spain & Portugal
Sartorius Spain, S.A. Avda. de la Industria, 32 Edificio PAYMA 28108 Alcobendas (Madrid)
Phone Spain +34.913.586.095 Phone Portugal +351.800.855.800 Fax Spain +34.913.589.623 Fax Portugal +351.800.855.799
Switzerland
Sartorius Mechatronics Switzerland AG Ringstrasse 24a 8317 Tagelswangen (ZH)
Phone +41.44.746.50.00 Fax +41.44.746.50.50
U.K.
Sartorius UK Ltd. Longmead Business Centre Blenheim Road, Epsom Surrey KT19 9QQ
Phone +44.1372.737159 Fax +44.1372.726171
Ukraine
LLS “Sartorius RUS” Post Box 440 “B” 01001 Kiev, Ukraine
Phone +380.44.411.4918 Fax +380.50.623.3162
Americas
USA
Sartorius Corporation 5 Orville Drive, Suite 200 Bohemia, NY 11716
Phone +1.631.254.4249 Toll-free +1.800.635.2906 Fax +1.631.254.4253
Argentina
Sartorius Argentina S.A. Int. A. Ávalos 4251 B1605ECS Munro Buenos Aires
Phone +54.11.4721.0505 Fax +54.11.4762.2333
Brazil
Sartorius do Brasil Ltda Avenida Senador Vergueiro 2962 São Bernardo do Campo CEP 09600-000 - SP - Brasil
Phone +55.11.4362.8900 Fax +55.11.4362.8901
Canada
Sartorius Canada Inc. 1173 North Service Road West, D4 Oakville, ON L6M 2V9
Phone +1.905.569.7977 Toll-Free +1.800.668.4234 Fax +1.905.569.7021
Mexico
Sartorius de México, S.A. de C.V. Libramiento Norte de Tepotzotlan s/n, Colonia Barrio Tlacateco, Municipio de Tepotzotlan, Estado de México, C.P. 54605
Phone +52.55.5562.1102 Fax +52.55.5562.2942
leadsmex@sartorius.com
Asia|Pacific
Australia
Sartorius Australia Pty. Ltd. Unit 5, 7-11 Rodeo Drive Dandenong South Vic 3175
Phone +61.3.8762.1800 Fax +61.3.8762.1828
China
Sartorius (Shanghai) Trading Co., Ltd.
rd
3
Floor, North Wing, Tower 1 No. 4560 Jinke Road Zhangjiang Hi-Tech Park Pudong District Shanghai 201210, P.R. China
Phone +86.21.6878.2300 Fax +86.21.6878.2882
Hong Kong
Sartorius Hong Kong Ltd. Unit 1012, Lu Plaza 2 Wing Yip Street Kwun Tong Kowloon, Hong Kong
Phone +852.2774.2678 Fax +852.2766.3526
India
Sartorius Weighing India Pvt. Ltd. #69/2-69/3, NH 48, Jakkasandra, Nelamangala Tq 562 123 Bangalore, India
Phone +91.80.4350.5250 Fax +91.80.4350.5253
Japan
Sartorius Japan K.K.
th
4
Fl., Daiwa Shinagawa North Bldg. 8-11, Kita-Shinagawa 1-chome Shinagawa-ku, Tokyo, 140-0001 Japan
Phone +81.3.3740.5408 Fax +81.3.3740.5406
Malaysia
Sartorius Malaysia Sdn. Bhd Lot L3-E-3B, Enterprise 4 Technology Park Malaysia Bukit Jalil 57000 Kuala Lumpur, Malaysia
Phone +60.3.8996.0622 Fax +60.3.8996.0755
Singapore
Sartorius Singapore Pte. Ltd 10 Science Park Road, The Alpha #02-13/14, Singapore Science Park II Singapore 117684
Phone +65.6872.3966 Fax +65.6778.2494
South Korea
Sartorius Korea Ltd.
th
8
Floor, Solid Space B/D, PanGyoYeok-Ro 220, BunDang-Gu SeongNam-Si, GyeongGi-Do, 463-400
Phone +82.31.622.5700 Fax +82.31.622.5799
Thailand
Sartorius (Thailand) Co. Ltd. 129 Rama 9 Road, Huaykwang Bangkok 10310
Phone +66.2643.8361-6 Fax +66.2643.8367
Specifications subject to change without notice. Printed in the EU on paper bleached without chlorine. |
Ver. 1 | 2019 Jan | EN