Sartorius MA100,MA50 Operating Instructions Manual

Operating Instructions
Sartorius Moisture Analyzer
Models MA100 | MA50 Electronic Moisture Analyzer
98648-008-50
The MA100 | MA50 Moisture Analyzer can be used for quick and reliable determination of the moisture content of materials of liquid, pasty and solid substances according to the method of thermogravimetry.
The moisture analyzer saves work and speeds up your routine procedures through the following features:
– Fast analysis time, gentle and uniform
sample drying due to the round ceramic IR heating element
– Quick determination of the drying
parameters and easy-to-set drying programs due to automatic determina­tion of the shutoff parameters for semi-automatic analyses (asap mode)
– Setting the fully automatic end-point
recognition parameter for an analysis only requires that you enter the drying temperature
– Optimal adjustment of the moisture
analyzer to other methods of analysis and adaptation to difficult samples due to the semi-automatic end-point recognition
– Quick drying without the risk of scorch-
ing the sample and preheating adapted to the sample’s heat sensitivity by selecting an adequate heating program
– High flexibility for analyzing the widest
variety of samples and storable pro­grams to save time when changing to different types of samples
– User-definable printouts that can
be customized before moisture analysis runs
– Brief instructions on the moisture
analyzer provide quick answers to frequently asked questions
The moisture analyzer is ideal as a measuring and test instrument for in incoming inspection, in-process control and quality control due to the following features:
– Convenient and reliable control of the
accuracy of the moisture analyzer according to the DIN/ISO standards by using the internal calibration weight (MA100 only)
– reproTEST for quick determination of
the standard deviation (MA100 only)
– ISO/GLP-compliant recording capability;
printouts can also be generated with an (optional) internal printer
– Optimal process control and quality
monitoring due to the statistical evalua­tion of up to 9,999 analyses/programs
– Password-protected drying parameters
The moisture analyzer meets the highest requirements placed on the accuracy and reliability of weighing results through the following features:
– High repeatability by limiting the
exposure of the weighing system to vibration during the start of an analysis and better access to the sample cham­ber due to the motorized heating unit
– Excellent readability under any lighting
conditions and backlit display for minimization of reading errors
– Removable sample chamber base plate
for easy cleaning of the sample chamber and protection of the weighing system from debris
Symbols
The following symbols are used in these instructions:
indicates steps you must perform
$ indicates steps you must perform only
under certain conditions
> describes what happens after you have
performed a certain step
– indicates that a list will follow
! indicates a hazard
For technical advice on applications:
Phone: +49.551.308.3160 Fax: +49.551.308.3495
2
Intended Use
2 Intended Use
3 Contents
4 Warning and Safety Information
Getting Started 5 Storage and Shipping Conditions 5 Unpacking
5 Instructions for Recycling 5 Equipment Supplied 6 General View of the Moisture Analyzer 7 Connecting the Moisture Analyzer
to AC Power 8 Leveling the Moisture Analyzer 8 Selecting the Brief Instructions 8 Turning On the Analyzer; Opening and
Closing the Sample Chamber
Operating Design
9 Keys
10 Operation: Analysis and Test Functions 11 Menu Operation and Drying Results 11 Data Output 12 Error Codes 12 Saving Data
Configuring the Moisture Analyzer
13 Setting the Language 14 Setting the Device Parameters 15 Entering or Changing the Password 17 Example: Setting the Time and Date 18 Device Parameters (Overview) 20 Configuring the Printout 24 Device Information
Operating the Moisture Analyzer
25 Weighing Function 25 Basics 27 Preparation 27 Adjustment to an Existing
Measuring System
28 Preparing a Sample 30 Setting the Drying Parameters 30 Drying Parameters (Overview) 35 Displaying Analysis Data 35 Identification Codes 36 Mode 36 Info 36 Statistics 37 Examples
3
Contents
51 “isoTEST” Calibration |
Adjustment Functions 51 Heater Adjustment 51 Weighing System Settings 56 Hardware Tests 58 Adjusting the Heater 58 Heating Unit Adjustment 59 Data Output 59 Output to the Moisture Analyzer 61 Interface Description 62 Data Output Format 63 Data Input Format 65 Digital Input/Output Ports 66 Pin Assignment Chart 67 Cabling Diagram
68 Error Codes
Care and Maintenance
70 Service 70 Repairs 70 Cleaning 70 Replacing the Fan Air Filter 71 Replacing the Fuses 71 Disassembling or Replacing
the Heating Unit 71 Safety Inspection
Overview
72 Specifications 74 Accessories (Options) 75 Declarations of Conformity 78 EC Type-Approval Certificate 79 Plates and Markings 80 Index
Appendix
Entering the General Password
This moisture analyzer complies with the European Council Directives as well as international regulations and standards for electrical equipment, electromagnetic compatibility, and the stipulated safety requirements. Improper use or handling, however, can result in damage and/or injury.
Read these operating instructions thoroughly before using your moisture analyzer to prevent damage to the equipment. Keep these instructions in a safe place.
Follow the instructions below to ensure safe and trouble-free operation of your moisture analyzer:
! Use the moisture analyzer only for
performing moisture analysis of sam­ples. Any improper use of the analyzer can endanger persons and may result in damage to the analyzer or other material assets
! Do not use this moisture analyzer in a
hazardous area/location; operate it only under the ambient conditions specified in these instructions
– The moisture analyzer may be operated
only by qualified persons who are familiar with the properties of the sample to be analyzed
! Make sure before getting started that
the voltage rating printed on the manufacturer‘s label is identical to your local line voltage (see the section on “Connecting the Moisture Analyzer to AC Power” in the chapter entitled “Getting Started”)
– The device comes with a power supply
that has a grounding conductor
– The only way to switch the power off
completely is to unplug the power cord
– Position the power cable so that it
cannot touch any hot areas of the moisture analyzer
– Use only extension cords that meet
the applicable standards and have a protective grounding conductor
– Disconnecting the ground conductor
is prohibited
– Connect only Sartorius accessories and
options, as these are optimally designed for use with your moisture analyzer
– Protect the moisture analyzer from
contact with liquid
– If there is visible damage to the
moisture analyzer or power cord: unplug the equipment and lock it in a secure place to ensure that it cannot be used for the time being
! Clean your moisture analyzer according
to the cleaning instructions only (see “Care and Maintenance”)
Do not open the analyzer housing. If the seal is broken, this will result in forfeiture of all claims under the manufacturer’s warranty.
In case you have any problems with your moisture analyzer:
$ contact your local Sartorius office,
dealer or service center
Warning: Severe Burns!
– When setting up the moisture analyzer,
leave enough space to prevent heat from building up and to keep your analyzer from overheating: – leave 20 cm (about 8 inches) around
the moisture analyzer
– 1 m (3 ft.) above the device
– Do not put any flammable substances
on, under or near the moisture analyzer, because the area around the heating unit will heat up
– Be careful when removing the sample
from the chamber: the sample itself, the heating unit and the sample pan used can still be extremely hot
– Do not remove the heating unit during
operation: the heating element and its protective glass panels can get extremely hot!
– Prevent excess heat build-up around
the analyzer
Hazards for persons or equipment posed by using specific samples:
Fire Explosion
– Flammable or explosive substances
– Substances that contain solvents
– Substances that release flammable
or explosive gases or vapors during the drying process
In some cases, it is possible to operate the moisture analyzer in an enclosed nitrogen atmosphere to prevent the vapor released during drying from coming in contact with oxygen in the surrounding atmosphere. Check on a case-to-case basis whether this method can be used, because installation of the analyzer in too small an enclosed space can affect its functions (for instance through excessive heat build-up within the analyzer). When in doubt, perform a risk analysis.
The user shall be liable and responsible for any damage that arises in connec­tion with this moisture analyzer.
Poisoning Caustic burns
– Substances containing toxic or caustic
or corrosive substances: These may be dried only under a fume hood. The value for the “lower toxic limit“ in a work area must not be exceeded.
Corrosion:
– Substances that release aggressive
vapors during the heating process (such as acids): In this case we recom­mend that you work with small sample quantities. Otherwise, vapors can condense on cold housing parts and cause corrosion.
The user shall be liable and responsible for any damage that arises in connec­tion with this moisture analyzer.
4
Warning and Safety Information
The moisture analyzer consists of a heating unit, a weighing system, a display and control unit and an optional printer. In addition to the socket for AC power (mains supply), it also has an interface port for connecting peripheral devices, such as a PC, an external printer, etc.
Storage and Shipping Conditions
Allowable storage temperature: 0 ...+40°C +32°F...+104°F
Do not expose the moisture analyzer unnecessarily to extreme temperatures, moisture, shocks, blows or vibration.
Unpacking the Moisture Analyzer
§ After unpacking the moisture analyzer, check it immediately for any visible damage as a result of rough handling during shipment
$ If this is the case, proceed as directed
in the chapter entitled “Care and Maintenance,” under the section on “Safety Inspection”
It is a good idea to save the box and all parts of the packaging until you have successfully installed your moisture analyzer. Only the original packaging provides the best protection for ship­ment. Before packing your moisture analyzer, unplug all connected cables to prevent damage.
Instructions for Recycling the Packaging
To ensure adequate protection for safe shipment, your moisture analyzer has been packaged to the extent necessary using environmentally friendly materi­als. After successful installation of the moisture analyzer, you should return this packaging for recycling because it is a valuable source of secondary raw material. For information on recycling options, including recycling of old weighing equipment, contact your municipal waste disposal center or local recycling depot.
5
Getting Started
Warranty
Do not miss out on the benefits of our full warranty. Please contact your local Sartorius office or dealer for further information. If available, complete the warranty registration card, indicating the date of installation, and return the card to your Sartorius office or dealer.
Equipment Supplied
The equipment supplied includes the components listed below:
– Moisture analyzer
– Power cord
– Pan support
– Shield disk
– Dust cover for keypad
– 80 disposable aluminum sample pans
– 1 pair of forceps
– 3 cards with brief instructions in
6 different languages
Installation Instructions
The moisture analyzer is designed to provide reliable results under normal ambient conditions in the laboratory and in industry. When choosing a location to set up your moisture analyzer, observe the following so that you will be able to work with added speed and accuracy:
– Set up the moisture analyzer on a
stable, even surface that is not exposed to vibrations
– Avoid placing the analyzer in close
proximity to a heater or otherwise exposing it to heat or direct sunlight
– Avoid exposing the moisture analyzer to
extreme temperature fluctuations
– Protect the moisture analyzer from
drafts that come from open windows or doors
– Keep the moisture analyzer protected
from dust, whenever possible
– Protect the moisture analyzerfrom
aggressive chemical vapors
– Do not expose the analyzer to
extreme moisture
– Make sure to choose a place where
excessive heat cannot build up. Leave enough space between the moisture analyzer and materials that are affected by heat.
Conditioning the Moisture Analyzer
Moisture in the air can condense on the surfaces of a cold moisture analyzer whenever it is brought into a substan­tially warmer place. If you transfer the moisture analyzer to a warmer area, make sure to condition it for about 2 hours at room temperature, leaving it unplugged from AC power. Afterwards, if you keep the moisture analyzer con­nected to AC power, the continuous positive difference in temperature between the inside of the moisture analyzer and the outside will practically rule out the effects of moisture condensation.
Setting up the Moisture Analyzer
§ Position the components listed below in the order given:
– Dust cover over the keypad
– Shield disk
– Pan support; turn to the left or right,
press slightly until it stops and snaps into place
– Disposable sample pan
No. Designation Order no.
for replacement
1 Internal printer (option) YDS01MA 2 Leveling foot 69MA0091 3 Level indicator 4 Brief instructions 5 Fuse 6 Power socket 7 Interface port 8 Keys for numeric input 9 Toggle key for alphabetic input
10 Key for opening and closing the
sample chamber by the heating unit 11 CF key (clear function) 12 Line feed; press the key to advance
the paper by one blank line 13 6 function keys (soft keys)
No. Designation Order no.
for replacement
14 “isoTEST” key (calibration/
adjustment functions) 15 On/off key 16 “Setup” key 17 Print key 18 Keypad 19 Display 20 Metrological Data
(only for models MA100.-0CE) 21 Shield disk 69MA0093 22 Pan support 69MA0092 23 Disposable sample pan 6965542 (80 units)
Not shown: Dust cover for keypad 6960MA01 Forceps 69MA0072
6
General View of the Moisture Analyzer
1
11
2
3
5
2
6
7 8 910
12
14
13
15
16
17
18
19
20
21
22
4
§ Check the voltage rating and the plug design
$ The heating unit of the moisture analyzer has been factory-set to 230 or 115 volts for
technical reasons. The voltage has been set as specified on your order. The voltage setting is indicated on the manufacturer’s label (see the bottom of the analyzer), for example: – 230 volts: MA50C-...230.. – 115 volts: MA50C-...115..
! If they do not match: To have the voltage setting changed, contact your local Sartorius
office or dealer, and do not operate your moisture analyzer in the meantime!
Use only – Original power cords – Power cords approved by a certified electrician/Sartorius service technician – If you need to connect an extension cord:
Use only a cable with a protective grounding conductor
§ Connecting the moisture analyzer, rated to Class 1, to AC power (mains supply):
The moisture analyzer must be plugged into a properly installed wall outlet which has a
protective grounding conductor (PE)
Safety Precautions
If you use an electrical outlet that does not have a protective grounding conductor,
make sure to have an equivalent protective conductor installed by a certified electrician as
specified in the applicable regulations for installation in your country.
The protective effect must not be negated by using an extension cord without a protective
grounding conductor.
Information on Radio Frequency Interference
Note: This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference in which case the user will be required to correct
the interference at his own expense. Changes or modifications not expressly approved
by Sartorius AG could void the user‘s authority to operate the equipment.
The AC adapter rated to Class 2 can be plugged into any wall outlet without requiring
any additional safety precautions. The ground or earth terminal is connected to the
housing, which can be additionally grounded, if required. The data interface is also
electrically connected to the scale housing (ground).
Connecting Electronic Devices (Peripherals)
§ Make absolutely sure to unplug the moisture analyzer from AC power before you connect
or disconnect a peripheral device (printer or PC) to or from the interface port.
Warmup Time
To deliver exact results, the moisture analyzer must warm up for at least 30 minutes after
initial connection to AC power or after a relatively long power outage. Only after this time
will the moisture analyzer have reached the required operating temperature.
Using Verified Moisture Analyzers as Legal Measuring Instruments in the EU*: $ Make sure to allow the analyzer to warm up for at least 24 hours after initial connection to
AC power or after a relatively long power outage
* including the Signatories of the Agreement on the European Economic Area
Charging the Built-in Rechargeable Battery
Before initial operation, leave the moisture analyzer connected to the line current
(mains supply) for at least 10 hours to charge the built-in rechargeable battery. When the
analyzer is disconnected from line current, the analysis data is retained in memory for
approximately three months.
7
Connecting the Moisture Analyzer to AC Power
Leveling the Moisture Analyzer
Purpose:
– To compensate for unevenness at the place of installation
– To achieve perfectly horizontal positioning of the moisture analyzer
for consistent repeatability
– This is necessary especially for testing liquid samples that need to be at a uniform level
in the disposable sample pan
Always level the moisture analyzer again any time it is moved to a different location.
Only the 2 front feet are used for leveling.
§ Retract the both rear feet
§ Turn the 2 front feet as shown in the illustration until the air bubble is centered
within the circle of the level indicator
> Several leveling steps are usually required
§ Extend the 2 rear feet until they touch the surface on which the moisture analyzer rests
Selecting the Brief Instructions
The moisture analyzer has separate cards with brief instructions on its most important functions. The following languages are included:
– English/Dutch – German/Italian – French/Spanish
To exchange the brief instruction card:
§ Pull out the brief instruction card on the right (you need to use a little force to overcome the resistance which is used to keep the card in place)
To insert the card:
§ Slide the selected card into the slot
$ Pull out the card on the right to remove it, if necessary
Turning On the Analyzer; Opening and Closing the Sample Chamber
§ To turn on the analyzer: Press the e key
§ To open or close the sample chamber: Press the k key
> The motor opens or closes the sample chamber
Setting the Language
– See the section on “Setting the Language” in the chapter entitled
“Configuring the Moisture Analyzer”
Setting the Date and Time
– See the section on “Entering User Data” in the chapter entitled
“Configuring the Moisture Analyzer”
8
Operation of the moisture analyzer follows a standardized “philosophy” which is described below.
Keys
The moisture analyzer is operated either using the keys on the display and control unit or via a connected PC. Operation using the keys is described in the following.
Labeled Keys
These keys always have the function indicated, but are not available at all times. Availability of these functions depends on the current operating status of the moisture analyzer and its menu settings.
9
Operating Design
The keys have the following functions:
e On/off key
Turns the moisture analyzer on/off. The moisture analyzer remains in standby mode
T isoTEST
Calibration/adjustment of the weighing system and hardware tests can be carried out
s Configuring the moisture
analyzer Access to the Setup menu; exit Setup
l Line Feed
(Optional) printer advances the paper by one line
p Data Output
Press this key to output displayed data via the interface port or to generate printouts using the (optional) printer
c Clear Function
Deletes keypad input Interrupts calibration/adjustment routines
k Arrow Key
Opens or closes the sample chamber
0 ... 9. Numeric Keys
see the section entitled “Text Input”
a Alphabetic Keys
see the section entitled “Text Input”
Numeric Input
To enter numbers: Press the 0 1 9 . keys
To store numbers entered: Press the corresponding soft key
To interrupt/cancel numeric input digit by digit: Press the c key
Text Input
To enter numbers: See the section entitled “Numeric Input”
To enter letters or characters: Press the a key
> Letters are displayed in the bottom line
for selection
To select a different letter: Press the corresponding soft key to change the letter shown
To select the letter/character shown: Press the corresponding soft key
> The selected letter is shown on
the display
$ Enter the next letter/character, if
desired, as above
$ To exit the letter input mode (i.e. if the
last character entered is a letter): Press the a key
To store a word: Press the corresponding soft key (e.g.,
ID)
To delete an input or character: Press c key
To delete user data: Enter . “decimal point” or a space “ “ and confirm by pressing Enter
F5F6 F4 F3 F2 F1
Prog. Info Stat. ID Mode Tara
Function Keys (Soft Keys)
The current function of a soft key is indicated in the bottom line of the display (footer).
Texts (as abbreviations) or symbols can be displayed in the example shown below.
Texts (Examples)
Info:
Information about the “Phase Drying” program
Prog.:
Select/configure the drying program
Stat.:
Statistics display/delete
ID:
Enter the ID number
Mode:
Change the parameters
Tare:
Tare the sample pan
The function keys are numbered (F1) through (F6), from right to left.
Symbols The bottom line shows the following symbols:
ooReturn to Setup menu
(in the Setup menu: exit the Setup program)
o Go back to the higher
selection level
O Show sub-items under the
active item
Q Move upward in the input/
output window
q Move downward in the input/
output window
l Set the selected menu parameter
There are two fundamentally different types of display:
– display for analysis and test functions
– display for menu parameter settings
(e.g., SETUP, Mode, ID) and final results (Info, Statistics)
Prog. Info Stat. ID Mode Tara
F5F6 F4 F3 F2 F1
Operation
Analysis and Test Functions
This display is divided into nine sections.
Example: Moisture Analysis
Info Drying Program Line
The following information is displayed here:
– Program name with 10 characters max.
(factory setting), e.g. BUTTER or
– Number of the selected drying program,
e.g. P1 (Configuration: Drying para­meter: Mea. No. # with automatic
Counter: On) – Temperature settings – End of measurement criteria
Bar Graph:
The bar graph indicates the percentage
by which the weighing system’s capacity
is “used up” by the current sample on
the pan
The bar graph is shown if you have
selected
minimum and
maximum initial weight
or target value,
tolerance in %.
The following symbols may be
displayed here:
0% Lower load limit for initial
sample weight
100% Upper load limit for initial
sample weight
Bar graph showing 10% intervals
- Minimum tolerance
= Target value
+ Maximum tolerance
Info drying program/Test function Bar graph
Measured value line
Text line Soft key labels
Plus/minus sign Stability indicator
Unit
Symbol for drying
Symbol for printout
Plus/Minus Sign: A plus or minus sign (F or H) is shown here for a weight value (e.g., a calculated value, when weighing in percent).
Measured Value Line: This section shows the weighed or calculated value or alphanumeric input.
Unit and Stability: When the weighing system reaches stability, the weight unit or calculation unit is displayed here.
Symbol for Drying: During the drying program, the following symbol is displayed here:
Drying in progress
Symbol for Printout: During the printout of the analysis results and other data, the following symbols appear in this column:
S
Print
Text Line: Additional information is displayed here (e.g.,operating state, operator guidance prompts, analysis temperature and measuring time, etc.)
Text line example for operator guidance (e.g., “TARE: Tare sample pan”)
Soft Key Labels: The current functions (abbreviations) of the arrow keys (soft keys) are indicated here. Please note that when we say “press the X soft key,” we are referring to the actual key indicated below the soft key label displayed.
g
d= 0.01g
100%
000
F S
Info Prog. Stat. ID Mode
A H L S T
P2 80/105/120 C Auto.
0%
TARE: Tare sample pan
a d
0015
Tare
10
Menu Operation and Drying Results
This display is divided into three sections.
Line for Operating State:
The line for operating state indicates the function of the current screen page. In the Setup menu, the current menu path is shown here.
Example for Setup, Language:
Input and Output Window: This window contains either detailed infor­mation (e.g., on the active application) or a pick list. A selected item is displayed inversely (white letters on a black back­ground). You can also enter information in an active field in this window using the alphanumeric keys.
Example for Setup, Parameter Settings:
The following symbol in the input or output window indicates:
d this symbol marks the saved
menu setting
Soft Key Labels:
See the “Function Keys (Soft Keys)” on the previous page
o Normal vibration
Strong vibration
SETUP LANGUAGE
Line for Operating State
Input and Output Window
Soft key labels
11
Parameter Settings
The parameters are configured in menus. These menus have several levels.
Example of the Setup Menu:
To select a parameter: Press the s key
To move within a menu level: Press the Q or q soft keys
To select a menu item (submenu): Press the O soft key
To set a parameter:
Press the Q or q soft keys repeatedly until the desired setting is selected (displayed inversely)
Confirm your selection by pressing the
l soft key
To change the numeric value of a parameter:
Press the Q or q soft keys repeatedly until the desired setting is selected (displayed inversely)
Enter the desired number using the 0 1 ... 9 . keys or the a soft key and enter the desired letters
Confirm your selection by pressing the
l soft key
To exit Setup: Press s or the
oo soft key
Data Output
You can choose between:
– internal printer (option)
– interface port for:
– Sartorius printer
(such as the YOP03-01)
– computer (PC)
– process logic controller (PLC)
– universal remote control switch
(Internal/External) Printer
You can configure the print functions to meet your individual requirements by selecting the corresponding menu code in Setup. The printouts can be generated as standard or ISO/GLP-compliant printouts.
ISO: International Organization for Standardization
GLP: Good Laboratory Practice
You can have printouts generated auto­matically, or by pressing p; printout generation can be dependent on or inde­pendent of the stability or time parameters (for example, automatic when a drying program starts, at specific time intervals, at the end of a drying program).
See the section on “Data Output Func­tions” in the chapter entitled “Operating the Moisture Analyzer” for a detailed description of data output options.
Interface Port
You may choose to connect a different peripheral device, such as one of the following, to the interface port instead of or in addition to the internal printer:
– external printer
– status indicators with digital input ports
– process logic controller with digital
input/output port
– a computer (PC) with a communications
port
The moisture analyzer can be monitored and remote-controlled via the interface port.
For a detailed description, see the section on “Data Output Functions” in the chapter entitled “Operating the Moisture Analyzer.”
Error Codes
If you press a key that has no function, or which is blocked at a certain point in an application program, this error is indicated as follows:
– a double-beep is sounded as
an acoustic signal if the key has no function
– invalid input is indicated by an
error message
– incorrect operation is indicated by an
error code or error message
The response to an operator error is identical for all operating modes. See the chapter entitled “Error Codes” for a detailed description.
Saving Data Storing Parameter Settings
The parameter settings in the Setup menu and for drying programs are active when you switch on the moisture analyzer. The parameter settings, selected with the
Mode soft key dur-
ing an analysis, are not saved (excep­tion: limits for the control function). In addition, the factory settings can be restored.
Saving Parameter Settings
You can assign passwords in order to block access to:
– user-configured drying programs
– setting mode for device parameters
– printout configuration
If no password has been assigned, anyone can access drying programs, “SETUP: Device parameters” and “Printout configuration” without entering a password.
If you assign a password and then forget what the word is, you can use the General Password (see Appendix) to access these menus.
12
Purpose
You can configure your moisture analyzer to meet individual require­ments by entering user data and setting selected menu parameters in the Setup menu.
The Setup menu is divided into the sections:
– Language
– Device parameters
– Printout configuration
– Device information
13
Configuring the Moisture Analyzer
Setting the Language
You can choose from 5 languages for the information display:
– German
– English (factory setting)
– English with U.S. date/time format
– French
– Italian
– Spanish
Example: Selecting the language: “U.S. Mode”
Step Key (or Instruction) Display/Printout
1. Select Setup menu s
2. Confirm the language O soft key
3. Set the language “U.S. Mode” Q soft key
4. Store the language l soft key
5. Exit Setup oo soft key
Setting the Device Parameters (DEVICE)
Purpose
Device configuration, i.e., to meet individual requirements by selecting predefined menu parameters in the Setup menu. You can block access to the menu by assigning a password.
Features
The device parameters are combined
in the following groups: – Password to the Setup menu – User ID – Weighing parameters – Interface – Internal printer (option) – Keypad – Display – Clock – Extra functions – Factory settings
You can view, enter or change the
following parameters:
Password – Password for access to the SETUP
menu: “Device parameters,” “Printout
configuration” and “Drying programs”
(8 characters max.)
User ID – ID codes:
User ID (20 characters max.)
Weighing parameters
– Adjustment to the ambient conditions
– Weight set number for
calibration/adjustment: W ID (weight ID; 14 characters max.)
– Exact calibration weight value for
calibration/adjustment of the analyzer, such as for adjustment according to a DKD certificate (see the section on “Calibration/Adjustment” in the chapter entitled “Operating the Moisture Analyzer”)
Interface
– SBI operating mode
Simple record of analysis results for PC or external printer, factory setting for YDP03-OCE printers Format: baud rate, number of data bits, parity, stop bits, handshake
– xBPI operating mode
Function-oriented interface with clear data transmission Network address: enter a number from 0 to 31; factory setting: 0
– Sartonet via RS-485 interface
Network address: enter a number from 0 to 31; factory setting: 1
Keys
– CF function: delete entire input
or last character
– Block key functions
Display
– Background
– Contrast/angle of the display (enter a
number from 0 to 4; factory setting: 2)
Clock
– Time (hh.mm.ss; hh can be entered
without a preceding zero)
– Date (dd.mm.yy or mm.dd.yy when
you select “English with U.S. date/time” as the language”)
Extra Functions
– Acoustic signal on or off
– Functions for external universal remote
control switch, extra keyboard or bar code scanner
Factory Settings
Parameters: The factory-set configu­rations are identified by an “o” in the list starting on page 18.
Preparation
Display existing “Device parameters”
§ Select the Setup menu: Press the s key
> SETUP is displayed:
§ Select “Device parameters”: Press the q and O soft keys
If no password has been assigned, anyone can access “SETUP: Device parameters” and “Printout configura­tion” without entering a password
If you have already assigned a password:
> The password prompt is displayed
$ If access is blocked by a password:
enter the password using the alpha­numeric keys
$ If the last character of the password is a
letter: conclude input by pressing a
§ Confirm the password and display “Device parameter”: Press the l soft key
> Device parameters are displayed:
14
Entering or Changing the Password
– Password for access to the following
functions, entered with 8 characters max.:
– SETUP parameters
– Data record configuration
– Edit or copy drying program
§ Select the Setup menu: Press the s
> SETUP is displayed
§ Select parameters: Press the q und O soft keys
If you have already assigned a password:
> The password prompt is displayed:
$ Enter the password
$ Confirm the password and
display the parameters: Press the l soft key
§ Write down the password here for easy reference:
Password = .............................
If you assign a password and then forget what the word is:
$ Enter the General Password
(see Appendix)
$ Confirm the password and display
the parameters: Press the l soft key
> Parameters are displayed
§ Select the password-setting function: Press the q or Q soft key repeatedly and O soft key until
> Password: and any existing
password are displayed:
15
§ New password: Enter the numbers and/or letter of the new password (8 characters max.)
If “none” is displayed as a password: this means no password has been assigned To delete the user password: Enter . and confirm
§ To confirm your entry: Press the l soft key
§ Exit the Setup menu: Press the oo soft key
> Restart the application
Extra Functions
§ Exit the Setup menu: Press the oo soft key
> Restart the application
§ Printout the parameter setting:
– If “Device parameters” are displayed:
Press the p soft key
> Printout (example)
Printouts with more than 20 characters are cut off
-------------------­SETUP
DEVICE
--------------------
User ID
User ID:
ABC123456
Weighing parameters
Adapt filter
Normal vibration
Calibration/Adjust
Wt.ID (W ID):
123
Cal./Adj. wt.:
50.000 g
Interface
SBI
Baudrate
1200 baud
Number of data bit
7 bit
Parity
Odd
Number of stop bit
1 stop bit
Handshake mode
Hardware 1 char
etc.
Reset the device parameters to the factory settings: See the section on selecting the factory settings in the chapter entitled “Device Parameters (Overview)”
Example 1:
Adapt filter to the ambient condition: “Strong vibration”
Step Key (or Instruction) Display/Printout
1. Select Setup menu s
2. Select and confirm q soft key, then the “Device parameters” O soft key
3. Select and confirm q soft key twice, “Weigh. parameters” then the O soft key
4. Confirm menu item O soft key “Adapt filter” and select next menu level
5. Select menu item q soft key “Strong vibration”
6. Confirm menu item l soft key “Strong vibration”
7. Select other menu items, if desired q Q soft key(s)
8. Store the setting and oo soft key exit Setup menu
16
Example 2:
Set time and date
Step Key (or Instruction) Display/Printout
1. Select Setup menu; s, then select “Device parameters” q and O soft keys
2. Set clock q and
O soft keys repeatedly
3. Enter the time 1 1 . 1 2
. 3 0
4. Set the time according to your local clock l soft key
5. Enter the date 1 3 . 0 3
. 9 9
6. Store the date l soft key
7. Enter other date, if desired q Q soft key(s)
8. Exit Setup menu oo soft key
17
Device Parameters (Overview)
ο factory settinguser-defined setting
Setup – Password empty Enter 8 Device characters max. parameters User ID empty Enter 20
characters max.
Weighing Adapt filter ο Normal vibration parameters Strong vibration
Calibration/ Weight set number None Enter 14 adjustment (weight ID) characters max.
Calibration/adjustment 50.000g Enter exact weight value weight value
Interface SBI Baud rate 150 baud
300 baud
ο 1200 baud
2400 baud 4800 baud 9600 baud 19200 baud
Number of data bits ο 7 bit
1
)
8 bit
Parity None
2
)
ο Odd
Even
Number of stop bits ο 1 stop bit
2 stop bits
Handshake Software handshake mode ο Hardware handshake
1 character after CTS
xBPI Network address 0 Enter any number
from 0 through 31
YDP01IS
YDP01IS label (Label printout)
Sartonet Network address 1 Enter any number
from 1 through 31
YDP02IS
For keys, display and clock see next page
1
) not if “None” parity is selected
2
) only if 8 data bits selected
18
Factory setting
Factory setting
Factory setting
Setup – Internal printer ο Off Device (option) On parameters
Keys CF function Delete entire
for input input
ο Delete last
character
Block ο All keys unblocked key functions All keys blocked except
for s, e Alphanumeric keys blocked
Display Background ο White
Black
Contrast Contrast (0 to 4) Selection: 0
1
ο 2
3 4
Clock Time: Enter
hh.mm.ss
Date: Enter dd.mm.yy or
mm.dd.yy (01.01.97)
Extra Acoustic signal ο On functions Off
Function: ο p key external switch F1 function key
(right-hand soft key) c key Open or close sample chamber Bar code scanner / Extra keyboard
Factory No settings Yes
19
Werkseinstellung
Factory setting
Factory setting
Configuring the Printout (CONFIG)
Purpose
You can configure individual printout formats for each application. All analysis printouts have basic factory settings.
You can block access to the “Printout configuration” data by assigning a password.
Features
– Maximum items in a data record: 30
– Header, footer,
intermediate results, statistics and information records can be configured separately
– Printout header is output when starting
moisture analysis
– Intermediate records are output
during the moisture analysis runs by pressing p or at each printout interval
– Footer printed out at the end of
a moisture analysis
– Statistics are printed out while the
statistics are displayed by pressing p (MA100 only):
– Information on phase drying is printed
by pressing p while the information is displayed
– Printout items can be deleted
individually
– “Form feed” in a printout footer:
Select the “YDP01IS label” print mode to configure automatic form feed up to the beginning of the next label
Extra Functions
§ Exit “Printout configuration”: Press the oo soft key
> Restart the application
Printout LIST or SELECT
LIST: Output of the current printout
list SELECT: Print currently selectable items
§ When the select bar is on LIST or SELECT: Press p
> Printout (example)
PRINT HEADER
LIST ==================== Blank line GLP header Program name Heating param. Standby temp. Start parameter End parameter Initial weight
-------­====================
etc.
Output of All Printout Settings:
– When the select bar is on printout
overview (SETUP CONFIG.): Press p
> Printout (example)
--------------------
13.07.1999 13:08 Mod. MA100C Ser. no. 90706913 Ver. no. 01-38-07 ID
-------------------­SETUP
CONFIG.
-------------------­Printout header
Blank line GLP header Program name Heating param. Standby temp. Start parameter End parameter Initial weight
--------
Intermediate result
Analysis time Analysis result
Printout footer
-------­Ending time Final weight Final result
--------
etc.
§ Reset “Printout Configuration” to the factory settings: See “Printout configuration”
Set Printouts to Factory Settings and confirm with YES.
20
Data Items for the Printout:
Parameter Display text Print- Inter- Print- Statistics Info Printout (Example)
out mediate out header result footer
Blank line 1) Blank line +* ++* +* +* Dotted line 1) -------- +* ++* +* +* -------------------- GLP header
GLP header +* +* +* --------------------
13.07.1999 13:06 Mod. MA100C Ser. no. 90706913 Ver. no. 01-38-07 ID WORKSTAT 234
--------------------
GLP footer
GLP footer +* +* 13.07.1999 14:06
Name:
--------------------
Date/time Date/time +++++13.07.1999 13:06 Time with seconds Time +++++ 13:06:45 User ID
(from Setup: Device) User ID +++ID WORKSTAT 234 Identification code 1 ID1 +++++ID1 SARTORIUS Identification code 2 ID2 +++++ID2 GOETTINGEN Identification code 3
ID3 +++++ID3
WEENDER LANDSTRASSE
Identification code 4 ID4 +++++ID4 LOT 15 Name in program memory Program name +* +* +* Prg 1 BUTTER Heating program with parameters
Heating param. +* Heating STANDARD
Fin.temp 105 'C
Standby temperature Standby temp. +* Stdby temp. OFF Start parameter Start parameter +* Start W/STABIL. End parameter End parameter +* End AUTOMATIC Preset tare weight Preset tare + PTare 0.000 g Initial weight Initial weight +* IniWt + 5.712 g Number of current analysis Analys no. + #1 Current weight CurrWt + CurrWt+ 5.1357 g Analysis time according
to current display Analysis time ++* Time 1.0 min Analysis result according
to current display Analys. res. ++* Res + 0.91 %L Analysis time and result
according to current display Analys time/res. +* 1.0 + 0.91 %L Form feed Form feed + Space for signature, ID
Name +++Name:
Final weight Final weight +* FinWt + 5.1357 g Time at end of analysis Ending time + Time 15.0 min Final result according
to current display2) Final result + Res + 9.85 %L Analysis time and result
according to current display2) Final time/res. +* 15.0 + 9.85 %L Phase drying Res1 + 4.45 %L
intermediate results
Phase results ++* Res2 + 3.15 %L
Res3 + 2.25 %L
* Factory setting
1
) Print items can be selected more than once
2
) “asap” result additionally printed at the end
of analysis in the “asap” mode
21
asap Difference 1.2 % Interval 24 s ec
or
asap
Canceled
Parameter Display text Print- Inter- Print- Statistics Info Printout (Example)
out mediate out header result footer
Text line “Statistics” Statistics ID + STATISTICS Number of analyses Number of analys. +* n5 Mean value Mean value +* Avg. + 4.84 %L Standard deviation Std. deviation +* s 0.05 %L Minimum Minimum +* Min + 4.80 %L Maximum Maximum +* Max + 4.90 %L Text line “ANALYSIS INFO” Info ID +* ANALYS.INFO
* Factory setting
22
23
Example:
Adding the Item “Phase Results” to Configure the Printout Footer
Step Key (or Instruction) Display/Printout
1. Select Setup menu; s, then Select “Printout q soft key twice, configuration” then O soft key
2. Select printout footer q soft key twice,
then O soft key
3. Use the select bar to define q soft key repeatedly the position for printout items “Phase results” on the LIST
4. Change to Select O soft key
5. Select “Phase results” q soft key repeatedly
6. Include phase results l soft key in the list
7. Select or delete other q Q l soft key printout items, if desired or
oq Q Delete soft key(s)
8. Configure other analysis o q Q O soft key(s) printouts, if desired
9. Exit Setup menu oo soft key
10. Perform moisture analysis p and press --------------------
Time 15.0 min FinWt + 9.5819 g Res1 + 12.05 %L Res2 + 7.12 %L Res3 + 4.96 %L C-Res + 24.13 %L
--------------------
Device Information
Purpose
Display of device information
Display Device Information
§ Select Setup menu: Press the s soft key
> “SETUP” is displayed:
§ Select “INFO: Device information”: Press q soft key three times, then O soft key
> Device information is displayed:
§ Printout information: Press p
> Printout (example)
--------------------
13.07.1999 13:02 Mod. MA100C Ser. no. 90805355 Ver. no. 01-38-07
(Operating program version)
ID BECKER123
(User ID)
-------------------­SETUP
INFO
-------------------­Versions No:
01-38-07
(Operating program version)
Wgh.sys ver.
00-25-03
(Wgh.sys.program version)
Model:
MA100C
Serial no.:
90805355
--------------------
§ Reset to SETUP Overview: Press the o soft key
§ Exit SETUP: Press the oo soft key
> Original settings are restored
24
Weighing Function
Purpose
The MA100/MA50 moisture analyzer can be used for quick and reliable determination of the moisture content of materials of liquid, pasty and solid substances according to the method of thermogravimetry.
Notes on Models MA100C-0CE230V1, MA100H-0CE230V1:
The models MA100C-0CE230V1, MA100H-0CE230V1 are delivered with initial calibration ex factory. Do not exceed the given ambient temperature range (+15° to +25°C) stated on the manufacturer’s ID label. The verifiable weighing program is designated by the Q symbol in the display.
To ensure reliable measurements, allow the analyzer to warm up before switching to the verifiable weighing program. An internal calibration/adjustment must be conducted before each measurement series (see page 54). This applies until the analyzer has cooled down to ambient temperature.
25
Operating the Moisture Analyzer
Basics
The moisture of a material is often mistakenly equated with its water content. In fact, the moisture of a material includes of all the volatile components which are given off when the sample is heated, resulting in a decrease in sample weight. Among
such volatile substances are: – water – fats – oils – alcohols – organic solvents – flavorings – products of decomposition (when a
sample is overheated)
There are many methods to determine
the moisture content of a substance.
Basically, these methods can be divided
into two categories:
When absolute methods are used, the
moisture content is directly determined
(for example, as a weight loss registered
during the drying routine). These
methods include oven drying, infrared
drying, and microwave drying. All
three of these methods are thermogravi-
metric.
When deductive methods are used, the moisture content is indirectly determined. A physical property, which is related to the moisture in the substance, is measured (e.g., absorption of electromagnetic rays). These methods include Karl-Fischer titration, infrared spectroscopy, microwave spectroscopy, etc.
Thermogravimetry is the process of deter­mining the loss of mass that occurs when a substance is heated. In this process, the sample is weighed before and after being heated, and the difference between the two weights is calculated.
In a conventional drying oven, circulating hot air warms the sample from the outside to the inside. Efficiency is lost during drying because as the moisture evaporates, it cools the sample surface.
By contrast, infrared rays (IR rays) pene­trate a sample without being impeded. Having reached the interior of a sample, they are converted into heat energy, which stimulates evaporation, thus drying the sample. A small part of the IR rays is reflected from the surface of the substance.
IR-Strahlung
Substanz
Wärmeenergie
IR rays
thermal energy
substance
The quantity of reflected IR rays depends to a great extent on whether the substance is light or dark-colored.
Dark Substance Light Substance Low Reflection High Reflection
How the rays penetrate the sample depends on the sample’s degree of light­transmitting capacity. If the degree of light-transmitting capacity is low, the rays can penetrate only the uppermost layers of the sample. The heat conductivity of the sample dictates the degree to which the heat can be transmitted to the underlying layers. The higher the conductivity, the faster and more uniformly the substance is heated.
The substance should be applied to the sample pan in a thin, even layer. A height of approximately 2–5 mm for 5–15 g substance weight has proved to be ideal. Otherwise, the sample will not be dried completely or the analysis time will be unnecessarily extended, a crust/skin will form on the surface of the sample or the sample will scorch, and the analysis results obtained will not be reproducable, and therefore, cannot be used.
Incorrect Application of a Sample
When preparing a substance for analysis, you should use methods that do not generate heat so that the sample does not lose moisture before it is analyzed.
Perform initial analysis of a new substance to test how the IR rays are absorbed by the sample and converted into heat. The printout of the intermediate values of the drying process provides you with this information at an early stage.
Experience has shown that the temperature setting selected during the infrared drying is usually lower than the tempera­ture setting used when working with a drying oven.
In many cases, the fully automatic shutoff mode will meet your requirements. If the final result is higher or lower than expected, try varying the temperature setting before resorting to a different shutoff parameter.
When analyzing samples that lose their moisture only very slowly or when operat­ing a cold moisture analyzer, the fully automatic mode may end the drying rou­tine too early, if it does not detect any analyzable progress in the drying routine under these conditions. In this case, pre­heat the moisture analyzer for 2–3 minutes before starting the drying routine or select a different shutoff parameter.
The Sartorius Moisture Analyzer Applications Guide will provide you with important information on the use of your moisture analyzer.
26
Preparation
Before drying a sample, you must carry out the following preparations:
– Adjustment to the available measuring
system (if required)
– Sample preparation
– Setting the parameters for drying
program
27
Adjustment to an Existing Measuring System
A moisture analysis method often replaces another drying method (e.g., the oven drying
method), because it is simple to use and requires shorter analysis time. In this case,
you should adapt this method to that of the moisture analyzer in order to obtain values
comparable to those yielded by your standard reference method.
Perform parallel measurements: take a fresh sample and divide it in half
Determine the moisture content of the first half using your standard method of analysis
Analyze the second half of the sample in the moisture analyzer.
Use the following settings:
– fully automatic mode for the shutoff parameter
– lower temperature settings than for the oven drying method
– temperature setting for organic substances: 80 – 100°C
– temperature setting for inorganic substances: 140 – 160°C
$ If the result for the second part does not correspond to that of the first:
– first, repeat the analysis using a different temperature setting
– then use the semi-automatic mode for the shutoff parameter (such as 5 mg/30 s
or the asap mode)
asap stands for “automatic searching/automatic programming”. The asap mode monitors
the drying process and, at the press of a key, calculates a semi-automatic shutoff
parameter for the expected results of the analysis. It then saves the parameter in a
program routine.
$ Vary the shutoff parameter, if required:
– Increase end-point recognition: set the parameter to 2 mg/30 s or 5 mg/60 s
– Decrease end-point recognition: set the parameter to 10 mg/30 s or 5 mg/10 s
Loading...
+ 61 hidden pages