Sartorius LMA110S,Mark 3 LTE,LMA100P,Mark 3 Operating Instructions Manual

Operating Instructions
Sartorius LMA110S Moisture Analyzer
Model Mark 3 LTE Electronic Moisture Analyzer
98040-699-09 Rev. D
Contents
Warning and Safety Information 4 Sample Pans 6 Safety Precautions 7
Getting Started 8 Storage and Shipping Conditions 8 Unpacking the Moisture Analyzer 8 Instructions for Recycling the Packaging 9 Equipment Supplied General Views of the Moisture Analyzer 10 Assembly 12 Choosing a Suitable Location 13 Connecting the Moisture Analyzer to AC Power 13 Leveling the Moisture Analyzer 14
Operating Design 15 Keypad 15 Display Modes 17 Menu Structure 19
Data Output 20 LED Status Lights 20
Operating the Moisture Analyzer 21 Standby Screen 21 Standby Temperature with a New Program 22 Default Drying Program 23 Recalling a Program 24 Testing a Sample 25
Setup Menu 28 Setting the Device Parameters in Program Manage Menu 28 Edit Current Program 29 Table: Program Settings 30 Units 31 Factors 32 Temperature 1 and 2 33 Time 1 and 2 34 Slope 34 Target Limits 39 Test Mode 40 Program Name 41 Create a New Program 42 Delete Program 42
Sort Programs 42 Configuring the Printout 43 Print Current Program 43 Print Program List 43 Print All Programs 43 Data Log 44 Device Information 45 Modify Operators 46 Graphing 47 Results Printout Format 48 Setting the Time and Date 51 Communication Ports 52 Pin Assignment Chart 52 Serial Command List 53 Conserve Mode 60 Security 61 Activate and Deactivate Security 61 Setup Security Access 62 Clear Data Log 63 Clear Operators 63 Restore Factory Defaults 64 Manage Program Storage 64 Transfer Programs 66
Audio, Video and Language 67 Audio Adjustments 67 Video Contrast Adjustment 68 Video Color Scheme 68
Service 69 System 69
Calibration | Adjustment Functions 70 Calibrate Balance 70 Heater Adjustment 73 Adjust Temperature 73 Verify Temperature 73
Concentration Mode 77 Multi-module Mode 78 Cleaning 80 Dispoal 81 Specifications 82 Accessories (Options) 83 Regulatory Compliance 84
2
3
Introduction
Thank you for choosing the Sartorius Mark 3 LTE moisture analyzer, the inter­national version of the field-proven Mark 3 from Sartorius. This analyzer is a modu­lar analytical instrument that can be con­figured to meet various performance requirements. Each analyzer consists of two parts, a control module and a heater module. The control module is comprised of a keypad, a display and a printer (optional) with the appropriate electron­ics. The heater module consists of the electronic balance and heating cavity. Up to two heater modules may be interfaced with a single control module. On initial power-up, the control module will auto­matically detect how many heater mod­ules have been connected and are pow­ered on. If more than one heater module is detected, the analyzer will run in the Multi-module mode.
The Mark 3 LTE moisture analyzer is used as a quick and reliable means of deter­mining the moisture content in various samples by the principle of loss on drying.
Key features of the Mark 3 LTE include:
§ Sample heating with infrared quartz tubes
§ Precision electronic balance for accurate
weighing and continuous weight moni­toring during analysis
§ Quarter VGA screen for easy viewing
§ Large storage of methods for easy recall
of frequently used drying procedures
§ Modular system for add-on capability of multiple heater modules
Please take the time to read these Operating Instructions to learn the safe operation of this analyzer. The Mark 3 LTE has many features that will be described and can be of benefit to your moisture testing program.
The Operating Instructions contain guidance in the form of descriptions, charts, software flow charts, and pictorial diagrams:
§ Key names are presented in bold:
Start|Stop
§ Safety symbols indicate notes or caution of risks as detailed below.
Descriptions and Symbols
Important safety related instructions are highlighted visually throughout the Operating Instructions with a descrip­tion:
Danger
Warning of a possible danger that may lead to a serious injury.
Caution
Warning of a possible danger that may result in minor injury or damage.
Note
Important guideline on how to use the moisture analyzer correctly.
Instrument labels
Caution:
!
Read manual before operating
Danger: Electrical hazard
Caution: Hot surface
Warning and Safety Information
This moisture analyzer complies with the European Council Direc­tives as well as international regu­lations and standards for electrical equipment, electromagnetic com­patibility, and the stipulated safety requirements. Improper use or handling, however, can result in damage and/or injury.
Read these operating instructions thoroughly before using your mois­ture analyzer to prevent damage to the equipment. Keep these instruc­tions in a safe place.
Follow the instructions below to ensure safe and trouble-free opera­tion of your moisture analyzer:
!
Use the moisture analyzer only for performing moisture analysis of samples. Any improper use of the analyzer can endanger persons and may result in damage to the ana­lyzer or other material assets.
!
Samples or containers (e.g., beakers or graduated flasks) that are too large or bulky to permit complete closure of the sample chamber hood must not be placed in the analyzer.
!
Do not use this moisture analyzer in a hazardous area/location; oper­ate it only under the ambient con­ditions specified in these instruc­tions.
The moisture analyzer may be
operated only by qualified persons who are familiar with the properties of the sample to be analyzed.
!
Class A Warning
Warning: This is a class A product. In a domestic environment this product may cause radio interfer­ence in which case the user may be required to take adequate measures.
4
!
Make sure before getting started that the voltage rating printed on the manufacturer‘s label is identical to your local line voltage (see the section on “Connecting the Mois­ture Analyzer to AC Power” in the chapter entitled “Getting Started”)
The analyzer comes with a power
supply that has a grounding con­ductor
The only way to switch the power
off completely is to unplug the power cord.
Position the power cable so that it
cannot touch any hot areas of the moisture analyzer.
Use only extension cords that meet
the applicable standards and have a protective grounding conductor.
Disconnecting the ground conduc-
tor is prohibited.
Connect only Sartorius accessories
and options, as these are optimally designed for use with your mois­ture analyzer.
Protect the moisture analyzer from
contact with liquid.
If there is visible damage to the
moisture analyzer or power cord: unplug the equipment and lock it in a secure place to ensure that it cannot be used for the time being.
!
Clean your moisture analyzer according to the cleaning instruc­tions only (see “Cleaning”).
Do not open the analyzer housing. If the seal is broken, this will result in forfeiture of all claims under the manu­facturer’s warranty.
In case you have any problems with your moisture analyzer:
$ contact your local Sartorius office,
dealer or service center
Warning: Severe Burns!
– When setting up the moisture analyzer,
leave enough space to prevent heat from building up and to keep your analyzer from overheating: – leave 20 cm (about 8 inches) around
the moisture analyzer
– 1 m (3 ft.) above the device
– Do not put any flammable substances
on, under or near the moisture analyzer, because the area around the heating unit will heat up.
– Be careful when removing the sample
from the chamber: the sample itself, the heater module and the sample pan used can still be extremely hot.
– Do not remove the heater module unit
during operation: the heating element and its protective glass panels can get extremely hot!
– Prevent excess heat build-up around the
analyzer.
5
Hazards for persons or equipment posed by using specific samples:
Fire Explosion
– Flammable or explosive substances
– Substances that contain solvents
– Substances that release flammable or
explosive gases or vapors during the drying process
In some cases, it is possible to operate the moisture analyzer in an enclosed nitrogen atmosphere to prevent the vapor released during drying from coming in contact with oxygen in the surrounding atmosphere. Check on a case-to-case basis whether this method can be used, because installation of the analyzer in too small an enclosed space can affect its functions (for instance through excessive heat build-up within the analyzer). When in doubt, perform a risk analysis.
The user shall be liable and responsible for any damage that arises in connec­tion with this moisture analyzer.
Poisoning Caustic burns
– Substances containing toxic or caustic
or corrosive substances: These may be dried only under a fume hood. The value for the “lower toxic limit” in a work area must not be exceeded.
Corrosion:
– Substances that release aggressive
vapors during the heating process (such as acids): In this case we recom­mend that you work with small sample quantities. Otherwise, vapors can con­dense on cold housing parts and cause corrosion.
The user shall be liable and responsible for any damage that arises in connec­tion with this moisture analyzer.
– Sample Pans (Delivered with the
Analyzer): The reusable metal pans are not made of stainless steel. For this reason, they are not resistant to corrosion. These pans can be used only for powder and granules (e.g. plastics) with very low moisture content. If the moisture content is higher, the disposable aluminium pans (90 mm diameter) must be used (see chapter “Accessories”).
6
Caution
Do not touch the metal surfaces inside the drying chamber while removing or placing a sample in the analyzer since the surfaces are very hot.
Do not touch the heater hood ventila­tion area at any time because this area can be hot during and after a test.
§ Do not test flammable or toxic materials.
§ Use the analyzer in a fume hood if the samples emit fumes which could be harmful.
§ Know where the fire extinguisher is located. Use only an extinguisher rated for use with electrical fires.
§ Keep the analyzer clean. Always unplug the analyzer and cool it thoroughly before cleaning or performing service.
§ Do not block the ventilation areas of the heating chamber.
§ If necessary, press the Start|Stop key during a test to stop the test.
§ Locate the analyzer away from flammable materials.
§ If liquid is spilled into the analyzer, unplug the Heating Module from the electrical supply immediately.
§ Analyzer is suitable for continuous operation
Safety Precautions
Danger
Use of this product in a manner not specified by the manufacturer may impair any safety protection provided by the equipment!
Every attempt has been made to make this analyzer safe and easy to use. However, like any laboratory instrument respect must be given to the operation of the analyzer due to environmental conditions, the nature of the samples being tested and of the chemicals which might be near the analyzer. To avoid personal injury or damage to the analyzer, please observe the following precautions: Read all instructions in the Operat­ing Instructions prior to operating your analyzer.
Warning
Control module contains 3V Lithium Battery. Please dispose of in accor­dance with local regulations and conventions.
7
Getting Started
Thank you for choosing the Sartorius LMA110S moisture analyzer. The LMA110S is a modular analytical instru­ment designed to be configured to meet various performance requirements. Each analyzer consists of two parts, a control module and a heater module. The con­trol module contains the keypad, dis­play and a printer (optional) with the appropriate electronics. The heater module consists of an electronic bal­ance and heating cavity. Up to two heater modules may be interfaced with a single control module. On initial power-up, the control module will automatically detect how many heater modules have been connected and are powered on. If more than one heater module is detected, the analyzer will run in the multi-module mode.
Storage and Shipping Conditions
Allowable storage temperature: –10 to +50°C –14°F to +122°F
Allowable shipping temperature: –40 to +70°C –40°F...+158°F
Unpacking the Moisture Analyzer
The moisture analyzer is delivered in a set of two custom-made boxes specif­ically designed for this precision instru­ment to provide optimum protection during transportation.
Note
Retain the original packaging in order to ship the analyzer if it needs to be transported or stored at some future time.
8
Follow the instructions carefully when unpacking the analyzer:
§ Unpack the analyzer carefully and gently.
§ Remove all the contents of the carton. Check carefully to make sure you have removed all accessory items.
§ The LMA110S consists of two modules, a control module and a heater module. Up to two heater modules may be connected to a single control module.
§ Inspect each module for physical dam­age and report any damage immediately to Sartorius or the distributor that you purchased the analyzer from.
§ Verify that you have received the following accessories:
Instructions for Recycling the Packaging
To ensure adequate protection for safe shipment, your moisture analyzer has been packaged to the extent necessary using environmentally friendly materi­als. After successful installation of the moisture analyzer, you should return this packaging for recycling because it is a valuable source of secondary raw material. For information on recycling options, including recycling of old weighing equipment, contact your municipal waste disposal center or local recycling depot.
Equipment Supplied
– Inter-module cable
– Power module cable
– Pan support
– Pan shield
– Power cord
– Tweezers
– Sample pans
– Operating instructions manual
9
General Views of the Moisture Analyzer
No. Designation
Heater Module
1 Pan shield 2 Pan support 3 Cylindrical quartz heating tubes 4 RTD temperature sensor 5 Heater adjustment tool connector 6 LED status lights 7 Service panel, heaters & RTD
temperature sensor 8 Level vial 9 Adjustable leveling feet
10 Inter-module cable receptacle 11 Main power switch 12 Main cord receptacle 13 Fuse 1 (250 V~5A T) 14 Fuse 2 (250 V~5A T) 15 Module power cable receptacle
No. Designation
Control Module
16 Storage Bin 17 VGA display 18 Soft keys and display overlay 19 Keypad 20 Inter-module cable receptacle 21 Module power cable receptacle 22 Ethernet socket: only for Sartorius
Service
23 USB socket: only for Sartorius
Service
Cables and Power cord
24 Serial connector 25 Inter-module cable 27 Main power cord
10
16
17
18
19
4
3
8 7 5
1 2
6
9
Control Module Connector Panel
Heater Module Connector Panel
Cables Connected to Control and Heater Modules
11
23
20
21
11
15
12
21 20
25 26
27
24
22
10
13
14
15
10
Assembly
The analyzer does not come fully assembled. The control and heater modules are placed next to each other and may be placed on either side of each other. Assemble the modules and individual components in the following order:
1. Open the heater hood using the handle and install the pan shield so that the center hole fits over the center ring of the cavity base plate.
2. Next, place the stem of the pan support through the center ring of the base plate and into the hole of pan receiver.
3. Remove the foam from the heater chamber.
4. Connect the power module cable between the control module and the heater module, observing the pin orien­tation.
5. Connect the inter-module cable between the control module (MOD COMM 1 connector) and the heater module, observing the connector orien­tation.
6. If multiple heater modules are used, install the additional inter-module cables from each heater module to the control module inter-module cable receptacle (MOD COMM 1-2). Refer to the chapter in this manual on multi­mode operation for instructions specific to operating multiple heater modules.
12
Choosing a Suitable Location
In order to ensure that the analyzer functions properly, select a location that meets the following requirements:
§ Permissible ambient environment – Temperature:
+15 to +40°C (59°F to 104°F)
– Relative humidity:
10 – 85%, non-condensing
– Altitude:
0 to 2,000 meters (0 to ~6562 ft)
§ The analyzer is intended for indoor use.
§ Place the analyzer on a rigid, horizontal
surface, preferably free of vibration.
§ Avoid drafts and excessive temperature fluctuation.
§ Locate the analyzer away from flammable materials.
Connecting the Moisture Analyzer to AC Power
This Class I Equipment (grounded type) analyzer contains a universal self-adjusting power supply rated to 100-120/200-250V~5/2.5A 50/60 Hz. Check that this voltage matches your local line voltage. If your voltage is not within this rating, do not connect the power cord to the power supply.
Danger
! The analyzer may only be operated
using the original power cord supplied. Please do not use it with an extension cord. The wall socket must be grounded. The AC Power mains supply fluctuations are not to exceed +/–10% of the nomi­nal supply voltage. The (Over Voltage) Category II for transient over voltages is EN 61010-1. The pollution Degree 2 is supplied EN 61010-1. The analyzer is rated to Normal Protection (not protected against harmful ingress of moisture).
§ Insert the power cord that with the analyzer into the power input receptacle on the back of the heater module.
§ Then plug the power cord into a wall socket.
§ Turn the analyzer on by moving the power switch on the back of the analyzer to the On position. The analyzer will briefly proceed through a self-diagnostic routine and then display the standby screen. It will automatically enter a warm-up period as noted on the screen. Should the analyzer display any warn­ings, consult the troubleshooting sec­tion of this manual, or call Sartorius.
13
Leveling the Moisture Analyzer
In order to function properly, the heater module should be horizontal. The heater module has a level vial. You can tell when the analyzer is level by checking whether the bubble is centered in the vial.
§ Check that all four feet are in the fully inserted position.
§ Observing the level vial, rotate one foot at a time to level.
Adjust control module so that it does not rock.
Note
The heater module must be re-leveled each time the analyzer is moved.
LMA110S Keypad
Warm-up
To enable the analyzer to function optimally, plug it in and leave it pow­ered on for a minimum of 15 minutes. The analyzer is preprogrammed to alert the operator that the device is in the warm-up period. It is best to allow the analyzer to complete this warm-up period before testing samples; however, this may be overridden by pressing the Enter key.
14
Numeric keys
Soft keys
Navigational keys
Operating Design
The Mark 3 LTE analyzer is operated through the keypad in conjunction with the quarter VGA screen. The Mark 3 LTE is preprogrammed with operational and setup software that enables the analyzer to be custom configured to meet the particular needs of the application, environ­ment or user ability.
Keypad
The analyzer has two membrane type keypads: the main keypad and a second that also serves to sealed the display. The keypad consists of three dedicated keys, numeric keys, navigational keys and soft keys. The following is a description of each of the key functions:
Dedicated keys:
Each key has a particular function. Pressing the key once will initiate the desired action.
Initiates the testing of a sample
Stops a test in progress
u Press to finalize the entry of
a numeric value
– Selects the highlighted menu
selection in a list
– From the Standby screen will print
the last result
Start|Stop
Numeric and Decimal Point Keys
Numbers from 0 to 9 and the deci­mal point are pressed either to enter a specific numeric value or to make a selection from a list of menu choices preceded by a number.
Navigational Keys
Four keys are intended to move the highlighted cursor on the screen in a particular direction.
Moves the cursor to the right
Moves the cursor to the left
Moves the cursor up
Moves the cursor down
15
Soft keys:
The functionality of these keys will change depending on what is on the display. A label on the display just above the key will describe the key function at the moment. Some examples of the soft keys are as follows:
Prints various items corresponding to the menu selection
Toggles a highlighted selection between on and off
Used to select a drying program from the Program menu
Access to all setup parameters from the Setup menu.
Setup
Recall
On/off
Print
Used to select an operator. Name will be printed on the results printout.
Display will return to the previous menu.
Display will return to the Standby menu.
Deletes the single highlighted character when entering alphanu­meric program names and operators.
Sets analyzer to the weighing mode.
Weigh
Delete
Users
16
Display Modes
The large quarter VGA display is designed to make operation and setting up the analyzer easy through of detailed information in clear descriptions, choices or menus. The analyzer will display information in several different formats as described below:
Warm-up:
This mode is entered whenever the analyzer is pow­ered on. The warm-up period ensures that the ana­lyzer is properly conditioned before testing is begun. During warm-up the screen will be mostly blank.
Standby (or Test Done):
In this mode the analyzer is in an ideal status. This is the period between testing samples. The temperature of the chamber is controlled to a particular level based on the Standby temperature set for the last test performed. The screen will display the program name, date and time, temperature and the last result, if a test has been run.
Test:
This mode is entered when the Start|Stop key is pressed to initiate a test. The screen is similar in design to the Standby mode but also includes prompts to guide the operator through the steps of adding a sample, entering an ID, if desired, and displaying the result when the test is completed.
Default
05/01/2011 02:14 PM
WARMIN G UP
Ver. 1.00.00
Mark 3 LTE Moisture Analyzer
Recall
Setup
Weigh
Users
Press ENTER to override warm up
Default
05/01/2011 02:14 PM
WARMIN G UP
Ver. 1.00.00
Mark 3 LTE Moisture Analyzer
Recall
Setup
Weigh
Users
Press ENTER to override warm up
Add sample to pan
DRY MILK POWDER
05/01/2011 02:14 PM
Module 1
0.000
g
Zero
{ }
0%
Standby
Current
60 57
17
Graph:
The graph mode can be entered during or after a test to show a graph of the weight versus time, temperature versus time, or result versus time.
Data Log:
The data log mode is used for reviewing past samples that have been tested. The log of samples can be sorted, or samples can be manually selected to perform statistical calculations.
18
Temperature / time
DRY MILK POWD ER
05/01/2011 02:14 PM
Print Result
210
30
0.5 1.0 1.5 2.0 2.5 3.0
3.5 4.0 4.5 5.0 min
25.67 %
View Data Log
Down Up
Top
Dry milk power 05/01/2011 02:10
3.45%M 6.7min 672001 Dry milk power 05/01/2011 02:02
3.50%M 6.8min 672002 Dry milk power 05/01/2011 01:53
3.47%M 7.0min 672003
05/01/2011
02:18 PM
Menu Structure
The Mark 3 LTE set up software is structured according to a series of menus. The Setup menu lists several topics of instrument setup. By drilling down into these topics, sub-menus will allow specific choices to be selected. See the Setup section for details.
How to navigate in the menu structure:
§ To select a menu and to drill down to more specific information: – Press the preceding number key of this listing or – Use the navigational arrow keys to highlight the
desired listing first, then press the Enter key
§ To return to a preceding menu level, press the arrow soft key. Subsequent presses will return you to the Standby screen.
§ To return to the Standby screen, press the bent arrow if one is available from that screen. To choose
a specific selection option or to enter a value:
A selection can be made by: – Pressing the preceding number key of this listing, or – Using the navigational arrow keys to highlight the
desired listing first, then press the Enter key.
§ How to make a numeric input: – Use the numeric keypad to enter a value fist, then
press the Enter key.
19
1)Program manage
2)Data log
3)Modify operators
4)Results printout format
5)Time and date
6)Communication ports
7)Conserve mode
8)Security
9)Audio & video
0)Module port scan
. ) Service
Setup Menu
05/01/2011 02:18 PM
Data Output
LED Status Lights
The Mark 3 LTE heater module has two LED lights on the front left corner that visually indicate the status of the ana­lyzer. The left LED shows the status of the balance. The right LED indicates the status of the heating function.
Each LED may have three colors: red, green or amber. Red is generally associated with a warning state and shows that corrective action must be taken. Green is generally associated with a positive condition, indicating you can proceed. Yellow generally indicates a transitional period or that the test is in progress.
20
Warmup
Balance LED
Heater LED
Warming up
Rapidly ashing RED Rapidly ashing RED
Ready to test
Steady GREEN Steady GREEN
Heater in conserve mode
Steady GREEN Steady YELLOW
Standby
Balance LED Heater LED
Sample in chamber heating cavity (chamber)
Flashing YELLOW Flashing GREEN
Test mode Balance LED
Heater LED
Test in progress Steady YELLOW
Steady YELLOW
Test done
Flashing YELLOW Flashing GREEN
Sample loading
Below target
Flashing GREEN
Steady GREEN
Within target Flashing YELLOW
Steady GREEN
Above target Flashing RED
Steady GREEN
Diagnostics Balance LED
Heater LED
Failed diagnostics
Flashing RED
Flashing RED
The Mark 3 LTE Moisture Analyzer is designed to test moisture content in a large variety of materials including powders, pastes and liquids. The perform­ance of the analyzer is dependent upon selecting optimum drying analysis conditions for your specific sample type. Sartorius will support you in finding the set of testing parameters (referred to as a pro­gram) you need for your specific tests. On request, we will provide a chart of the parameters and test results we have compiled by testing samples in our laboratory and comparing the results to those obtained by reference methods.
If our chart of applications shows a drying program that is not already programmed into your analyzer, you may do this yourself by following the instruc­tions under Setup in this manual.
Standby Screen
During warm-up, your analyzer will enter the Stand­by mode and the Standby screen will be displayed. It is from the Standby (or Test Done) screen that all functions of operating the analyzer, including test­ing samples and setting parameters, take place.
The standby screen shows the following information:
§ Program name
§ Time and date
§ Standby temperature and actual temperature
§ Various soft keys for options that are available
depending on the security setup options engaged
§ Warnings that will need to be addressed before testing
21
Operating the Moisture Analyzer
DRY MILK POWDER
Standby
Recall Setup
Weigh Users
Press START key to begin test
Standby
Current
60 59
05/01/2011 02:18 PM
Ver. 1.00.00
Mark 3 LTE Moisture Analyzer
Functions available from the Standby screen:
§ Press Start|Stop key to initiate a test
§ Press Recall soft key to change the program to
be used for testing a sample
§ Press User soft key to select an operator name
(feature must be turned on in Setup)
§ Press Setup soft key to enter the Setup menu in
order to make all parameter changes
§ Press Graph soft key to display a graph
§ Press Weigh soft key to use as a balance
Standby-Temperature
!
with a New Program
When creating a new program, the standby temperature setting from the previous drying program is used automatically by the analyzer.
Example: You have created a new program for drying a temperature-sensitive sample at low tempera­ture. The analyzer uses the standby tempera­ture, e.g. 165°C, from the previous program. You must call up the standby temperature parameter and change the temperature value. If you forget performing this setting, your ana­lyzer is always too hot before starting the test.
22
Default Drying Program
Your analyzer is delivered with a default drying pro­gram. The drying program contains all the specific parameters to test a sample including temperature, endpoint, unit of measure, standby temperature, etc. You will notice if you are in the default program or a specific material program by the program name at the top left of the display.
23
Program name: Default Temp1=105C Time1=Off Temp2=Off, Time2=Off Slope=0.050%/1 minute, Actual Start delay=0 second Standby temp=60C, ­Ideal weight=5g, Lock=Off Target limits=Off Mode=Standard
Recalling a Program
Drying programs are stored for easy access at a later time. This eliminates the need to change parameters for frequently tested samples. Up to 60 drying pro­grams can be stored. Creating and storing a program is described under Developing a Drying Program.
To recall a stored program, perform the following steps:
1. From the Standby screen, press the RECALL soft key. The screen will change to the first page of stored programs as show below.
2. To display programs on the next pages, press the Down soft key.
3. Press the number(s) key followed by the Enter key for the number that precedes the program name. As an alternative, use the arrow keys to highlight the desired program name, then press Enter. The display will change to the Standby screen with the new program in the top left corner. This program will be used for the next test. If a blank program is entered, the analyzer will use the last valid program that was entered.
Note
As a shortcut, it is not necessary to have the desired program listed on the display to recall it. Simply enter the desired number from your memory on the first page and press the Enter key.
24
11)
2) Whey Protein 12) )3
1 )3
)4
1
)
4
)5
1
)
5
)6
1
)6
)71 )7
)8
1
)
8
)91
)9
)0
2
)01
Programs
Program #:
Down
1)
Dry Milk Powder
Up
_
Top
05/01/2011 02:18 PM
Testing a Sample
A test can only be initiated from the Standby or Test Done screen. The program that will be used is shown in the upper left corner.
To test a sample, perform the following:
1) Open the hood of the heater module using the handle on the front of the analyzer.
Caution
The heater hood will be hot during testing and the standby mode. Only touch the heater hood by the handle.
2) Place one pan onto the pan support and close the heater hood.
3) From the Standby screen, press the Start|Stop key. The display will change to the test mode as shown on the left. The balance will automatically tare and show zero weight on the screen.
Note
If you forgot to place the pan onto the balance ahead of time, you may open the heater hood now, add the pan, close the heater hood and manually tare by pressing the Zero soft key.
25
DRY MILK POWDER
Standby
Recall Setup
Weigh Users
Press START key to begin test
Standby
Current
60 59
05/01/2011 02:18 PM
Ver. 1.00.00
Mark 3 LTE Moisture Analyzer
Sampling
DRY MILK POWDER
Module 1
0.000
g
Zero
{ }
0%
Standby
Current
60 57
05/01/2011 02:19 PM
4. At the “Add sample to pan” prompt, open the heater hood and begin adding sample to the pan. A bar graph on the display will guide the operator to place the correct amount of sample as set by the ideal weight parameter of the drying program. Continue adding sample until the beeper sounds, indicating that the sample amount is within 10% of the ideal weight. The exact weight is also displayed on the screen.
5. At the “Close heater hood” prompt, close the heater hood. The balance will capture the initial weight of the sample “Sampling” and then show that the test is in progress. The display will change to show cur­rent information including the elapsed time and the current percentage weight loss in the units as speci­fied in the drying program.
Note
To stop the test in progress, press the Start|Stop key.
The optional printer will begin to print as the test begins. Printout information can be selected in Setup menu under Printer.
At the end of the test the display will show “Test Done.” The result will stay on the screen as well as be printed. The analyzer will enter the Standby mode until the next test is initiated.
26
Dry Milk Powder
Test in Progress
0.01
% Moisture
Elapsed time: 00.10
Graph
Press STOP key to stop test
Temp 1
Current
105
71
05/01/2011 02:18 PM
Press START key to begin test
Dry Milk Powder
Test Done
3.47 % Moisture
Result time: 7.0 min
ID: 126399.1
Target: Pass
Recall Setup
Graph User
Warning: Sample on pan
Standby
Current
105 105
05/01/2011 02:18 PM
The display will prompt you, the operator to remove the sample before the next sample can be test­ed. Open the hood and carefully remove this sample using tweezers, and while the hood is open, add the sample pan for the next sample to be tested.
Caution
!
The sample is hot during and after the test. Remove the sample with care by using the tweezers provided with the analyzer. Do not touch any internal part of the heating chamber or pan shield as they will be very hot.
Press the Start|Stop key to begin the next test using the same program.
Note
The analyzer balance must be calibrated before putting the unit into routine service. See the section on Calibrate Balance in this manual.
27
Loading...
+ 61 hidden pages