Sartorius IS...-X Series,IS2CCE-SX,IS06BBE-SX,IS6CCE-HX,IS6CCE-SX,IS12CCE-SX,IS16EDE-HX,IS64EDESX,IS34EDE-HX,IS300IGG-HX,IS150IGG-HX Operating Instructions Manual

Operating Instructions
Sartorius Intec IS-X
Models IS06BBE-SX, IS2CCE-SX, IS6CCE-HX, IS6CCE-SX, IS12CCE-SX, IS16EDE-HX, IS34EDE-HX, IS64EDE­SX, IS150IGG-HX, IS300IGG-HX and Verifiable Models
Weighing Platforms for Use in Hazardous Areas/Locations
98648-009-42
Contents
Contents
2 Intended Use
3 Operating Limits 4 Tables of Specifications 9 Pin Assignment Chart 10 Troubleshooting Guide 11 Accessories (Options) 12 General View of the Equipment 15 Dimensions (Scale Drawings) 20 Declarations of Conformity 26 Certificates for Operation of the
Weighing Platform in Hazardous Areas
Symbols
The following symbols are used in these nstructions:
§ indicates required steps
$ indicates steps required only under
certain conditions
> describes what happens after you have
performed a certain step
– indicates an item in a list
! indicates a hazard
Make sure you observe the warning and safety information in its entirety during installation and operation, as well as while performing maintenance and repair work on the equipment. It is important that all personnel using the equipment understand this information, and have access to it at all times.
The warning and safety information
is contained in the 11-language installation manual included with the equipment.
The warning and safety information
can be supplemented by the equipment operator. Make sure all operating personnel are informed of any additions to these instructions.
Intended Use
The series IS...-.X models are precision
weighing platforms for weighing
samples in hazardous areas/locations.
Series IS...-.X equipment and connected
accessories are intended to be used exclusively for weighing.
Instructions for installing the
equipment and putting it into operation are contained in the enclosed
11-language installation manual. Important Note: The verifiable models („IS...-.XCE“) can
be verified for use as a scale only when
connected to a display and control unit,
indicator or terminal.
The following features make the
IS...-.X weighing platforms ideal for
use as inspection, measuring and test
instruments in quality management
systems: – isoTEST calibration and adjustment function – ISO/GMP-compliant documentation
functions
IS...-.X weighing platforms save work
and speed up both simple and complex
routine procedures through the
following features: – Ultrafast response times – Connection to a PC in a PLC or other
control system IS...-.X weighing platforms meet the
highest requirements placed on the
accuracy and reliability of weighing
results through the following features: – Efficient filtering-out of vibration – Stable and reproducible results – Rugged, durable weighing system – Weighing platform and accessories
approved for use in hazardous
areas/locations (see the table in the
11-language installation instructions)
All power supplies or AC adapters may
be connected only to an electrical
outlet (mains supply) with at least 90V
and no more than 264V (48 – 62 Hz).
Set up the explosion-protected IS...-
.X series weighing platform and its accessories in accordance with acknowledged technical regulations. You must also comply with the national laws and safety regulations of your country.
Data Output
The weighing platform is equipped
with an RS-485 data interface (RS-232 available on request) which can be used
to connect any of the following: – Printer* – Peripheral device (such as a computer)* – Universal remote switch
* connected through a Zener barrier (see „Accessories“)
Hotline/Technical Advice on Applications
Phone: +49 (0) 40/67960444 Fax: +49 (0) 40/67960474
E-mail: technical.support.hh
@sartorius-intec.com
2
Operating Instructions IS-X
Operating Limits
Operating Limits
Overload Capacity
Thanks to the rugged construction of
Sartorius Intec weighing platforms, the weighing system will not be damaged if the maximum weighing capacity is occasionally exceeded.
The table below lists the maximum
overload capacity of each weighing platform model:
IS06BBE-SX.. IS2CCE-SX.. IS6CCE-SX..
IS6CCE-HX.. 3 kg 12 kg 25 kg
IS16EDE-HX IS150IGG-HX,
IS34EDE-HX IS300IGG-HX IS64EDE-SX 130 kg 600 kg
Shock Resistance
Even though Sartorius Intec weighing
platforms feature highly rugged construction, there are some limits. Avoid exposing the system to falling objects, side impact, or shocks. The weighing platform meets the mechanical requirements of IEC 68 Part 2-27.
Notes on Integration into Conveyor
Systems
The weighing platform is suitable for
installation in conveyor systems. Make sure to observe the following notes as well as the equipment dimensions when choosing components to be mounted on your weighing platform
(see „Dimensions (Scale Drawings)“ in
this manual).
Any moving or rotating parts intended
to be permanently attached to the load plate must be designed so that they cannot affect the weighing results. For example, rotating mechanisms must be properly balanced.
The load plate must be clear on all
sides so that any parts or any dirt that falls will not create a connection between the weighing platform and any permanently mounted preload components.
Any cables or tubing between the
weighing platform and other equipment must not apply any force to the weighing platform. Make sure that
cables do not touch the load plate.
Special care must be taken to prevent
the build-up of static electricity caused by moving parts (such as roller conveyors).
Preload Range (Zero-setting Range)
The weight of components permanently
attached to the weighing platform is known as the „preload.“ The preload is electronically compensated up to a defined value. Thus the entire weighing capacity is available, and it is possible to zero or calibrate/adjust the weighing system (with external weights). A preload heavier than the defined value may result in a reduction of the weighing capacity.
The maximum preload that can
be compensated depends on the specification block configured (refer to the tables of specifications below). Components attached to the weighing platform must be installed before the weighing platform is connected to power.
Configuring the Weighing Platform
The weighing platform can be
configured to meet individual requirements, such as weighing in international units, weighing in unfavorable ambient conditions, or external electronic processing of weight values. This means the final configuration of a unit can be performed by the weighing equipment dealer or the end user.
The weighing platform is adapted
for individual requirements using the configuration menu in the connected display and control unit or indicator. The weighing platform can also be configured using a connected
computer running the Sartorius Intec
configuration program
(see „Accessories“).
Options Available for Weighing Range Configuration
SuperRange
This range divides the weighing
capacity of the platform into 50.000 intervals, an exceptionally high resolution. The readability is identical over the entire weighing range
(for example, 1 g).
SingleRange
The weighing range is divided into up
to 50.000 intervals. The readability is identical over the entire weighing range ( for example, 10 g).
Multi-interval/DualRange
Intervals with adjusted accuracy. These functions divide the
weighing capacity into as many 3 weighing ranges, each with a different readability.
The weight readout is shown with
different levels of accuracy in each range. The platform changes from one range to another automatically when the load on the platform is changed
(the readability in each range is listed in
the tables of specifications below).
Once the platform has been tared, the
highest possible resolution is available even if the weighing platform is loaded.
The configuration options available
for each weighing platform are listed below, in the section entitled „Tables of Specifications.“
Operating Instructions IS-X
3
Tables of Specifications
Tables of Specifications
Model IS06BBE-SX IS2CCE-SX IS6CCE-HX IS6CCE-SX IS12CCE-
SX
Readability g 0.001 0.01 0.01 0.1 0.1
.........
Weighing capacity g 620 2200 6200 6200 12,000 Max. overload capacity kg 3 12 12 25 25 Tare range (subtractive) g – 620 – 2200 – 6200 – 6200 – 12000 Preload (can be electronically g 93 110 1240 1200
compensated without reducing the weighing capacity) Max. preload at start of isoTEST g 110 1300 5200 6400 10000
calibration/adjustment
(platform must be zeroed) Repeatability <±g 0.001 0.01 0.01 0.05 0.05 Linearity <±g 0.002 0.02 0.02 0.1 0.2 Sensitivity drift within +10 to +30 °C <±/K 2 • 10 Response time (average) s 1.5 1.5 1.5 1 1 External calibration weight g 500 (E2) 2000 (F1) 5000 (E2) 5000 (F2) 5000 (F1)
(of at least accuracy class...) Load plate dimensions mm Ø 130 218 x 200 218 x 200 218 x 200 218 x 200 Platform dimensions (W x D x H) mm 240x294x86 240x294x86 240x294x86 240x294x86 240x294x86 Net weight, approx. kg 7 7.3 8.4 7.3 7.3 Dust and water protection rating for the housing acc. to EN 60529 IP44 AC power source Power supply model YPS02-X or AC adapter model YPS02-Z (100-240V)
(must be connected by the weighing equipment dealer or a Sartorius Intec service tech-
nician) Frequency Hz 48 – 62 Allowable ambient operating temperature 0 °C to +40 °C (273 °K to 313 °K; 32 °F to 104 °F) Operating temperature range +10 °C to +30 °C (or 0 °C to +40 °C; depends on configuration) Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels Display update (depends on filter level selected) 0.1 - 0.1 - 0.2 - 0.4 (normal output rate) Power consumption Average: 25VA Selectable weight units: Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwanese taels, grains, pennyweights, milligrams, parts per
pound, Chinese taels, mommes, Austrian carats, tola, baht and mesghal Built-in RS-485 interface (standard equipment; xBPI protocol):8 bits, odd parity; transmission rate: 9600 -
38,400 baud; half-duplex; convertible to RS-232 (SBI protocol): 7 bits, Parity: even,
odd, mark, space; transmission rates: 150 — 19,200 baud; 1 or 2 stop bits, software/
hardware handshake; factory settings: 1200 baud, odd parity, 1 stop bit, hardware
handshake with 2 characters after CTS
Model series Marking (ATEX) Use in IP rating Markings in acc. with FMRC Use in
Zone approval for USA and CSA
approval for Canada
IS.. BBE-.X.. 1, 2 (gas) IP 54 IS CL I, DIV 1, GR A,B,C,D, T4 Class I, Division 1 (gas)
IS.. CCE-.X.. 1, 2 (gas) IP 54 IS CL I, DIV 1, GR A,B,C,D, T4 Class I, Division 1 (gas)
IS… EDE-.X.. 1, 2 (gas) IP 65 IS CL I, DIV 1, GR A,B,C,D, T4 Class I, Division 1 (gas)
20, 21, 22 IP 65 IS CL II,III, DIV 1, GR E,F,G, T4 Class II,III Division 1
IS64EDE-HX.. 1, 2 (Gas) IP 44 IS CL I, DIV 1, GR A,B,C,D, T4
h II 2 G EEx ib IIC T4 CL l, ZONE 1 AEx ib IIC T4 Class I, Zone 1 (gas)
h II 2 G EEx ib IIC T4 CL l, ZONE 1 AEx ib IIC T4 Class I, Zone 1 (gas)
h II 2 G EEx ib IIC T4 CL l, ZONE 1 AEx ib IIC T4 Class I, Zone 1 (gas)
h II 1 D T135°C (dust) (dust+ fibers) h II 2 G EEx ib IIC T4 CL l, ZONE 1 AEx ib IIC T4 Class I, Zone 1 (gas)
-6
2 • 10-6 2 • 10-6 4 • 10-6 4 • 10
Class I, Division 1 (gas)
-6
IS… IGG-HX 1, 2 (gas) IP 65 and IP67 IS CL I, DIV 1, GR A,B,C,D, T4 Class I, Division 1 (gas)
(dust) (dust + fibers)
h II 2 G EEx ib IIC T4 CL l, Zone 1 AEx ib IIC T4 Class I, Zone 1 (gas) h II 1 D T135°C 20, 21, 22 IP 65 and IP 67 IS CL II,III, DIV 1, GR E,F,G, T4 Class II,III Division 1
* = factory setting
4
Operating Instructions IS-X
Tables of Specifications
HX-Mode IS16EDE-HX IS34EDE-HX IS64EDE-SX IS64EDE-HX IS150IGG-HX IS300IGG-HX Readability g 0,1 0,1 1 0,1 1 2 Weighing capacity kg 16 34 64 64 150 300 Max. overload capacity kg 130 130 130 130 600 600 Tare range (subtractive) kg – 16 – 34 – 64 -64 – 150 -300 Preload (can be electronically compensated kg 4 4 13 13 20 60
without reducing the weighing capacity)
Max. preload at start of kg ca. 19 ca. 21 ca. 45 45 -
isoTEST calibration/adjustment
(platform must be zeroed) Repeatability <±g 0,05 0,1 0,5 0,2 1 2 Linearity <±g 0,2 0,2 1 1 4 8 Sensitivity drift
within +10 to +30 °C <±/K 2 • 10
-6
2 • 10-6 3 • 10
-6
3 • 10
-6
2,5 • 10
-6
2,5 • 10
-6
Response time (average) s 1,5 External calibration weight kg 10 (F1) 10 (F1) 10 (F2) 10 (F1) 50 (F2) 50 (F2
(of at least accuracy class...) Operating temperature range °C 10...30 10...30 0...40 0...40 0...40 0...40 Load plate dimensions mm 300 x 400 300 x 400 300 x 400 300 x 400 800x600 800x600 Net weight, approx. kg 16 16 16 16 70 70 Dust and water protection rating of the
housing acc. to EN 60529 IP65 IP65 IP65 IP44 IP67 IP67 AC power source Power supply model YPS02-X or AC adapter model YPS02-Z (100-240V) (must be
connected by the weighing equipment dealer or a Sartorius Intec service technician) Frequency Hz 48 – 62 Allowable ambient operating temperature 0 °C to +40 °C (273 °K to 313 °K; 32 °F to 104 °F) Operating temperature range +10 °C to +30 °C (or 0 °C to +40 °C; depends on configuration) Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels Display update (depends on filter level selected) 0.1 - 0.1 - 0.2 - 0.4 (normal output rate) Power consumption Average: 25VA Selectable weight units: Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwanese taels, grains, pennyweights, milligrams,
parts per pound, Chinese taels, mommes, Austrian carats, tola, baht and mesghal Built-in RS-485 interface (standard equipment; xBPI protocol): 8 bits, odd parity, transmission rate:
9600 - 38,400 baud; half-duplex; convertible to RS-232 (SBI protocol): 7 bits,
Parity: even, odd, mark, space; transmission rates: 150 — 19,200 baud; 1 or 2
stop bits, software/ hardware handshake; factory settings:
1200 baud, odd parity, 1 stop bit, hardware handshake with 2 characters after CTS
* = factory setting
Operating Instructions IS-X
5
Tables of Specifications
IS „-.XCE“ Series Verifiable Models Model IS06 BBE-SXCE IS2 CCE-SXCE IS6 CCE-HXCE Readability g 0.001 0.01 0.01 Weighing capacity g 620 2200 6200 Max. overload capacity kg 3 12 12
Tare range (subtractive) g – 620 – 2200 – 6200 Preload (can be electronically compensated) g 93 110
.........
without reducing the weighing capacity)
Max. preload at start of g 110 1300 5200
isoTEST calibration/adjustment (weighing inst. must be zeroed)
Repeatability <±g 0.001 0.01 0.01
.........
Linearity <±g 0.002 0.02 0.02 Sensitivity drift within +10 to +30 °C <±/k 2 • 10 Response time (average) s 1.5 1.5 1.5 External calibration weight g 500 (E2) 2000 (F1) 5000 (E2)
(of at least accuracy class...) Accuracy class K K K
Type . BD BF BF Verification scale interval g 0.01 0.1 0.1
Minimum capacity acc. to g 0.02 0.5 0.5
prepackage regulations* Operating temperature range °C 10 to 30 (50 to 86°F) 10 to 30 (50 to 86°F) 10 to 30 (50 to 86°C) Load plate dimensions mm Ø 130 218 x 200 218 x 200 Platform dimensions (WxDxH) mm 240x294x86 240x294x86 240x294x86 Net weight, approx. kg 7 7.3 8.4 Dust and water protection rating for the housing acc. to EN 60529 IP44 AC power source Power supply model YPS02-X or AC adapter model YPS02-Z (100-240V);
YRB02-Z (rechargeable battery), YPS02-XV24 (24V) Frequency Hz 48 – 62 Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels Display update
(depends on filter level selected) 0.1 - 0.1 - 0.2 - 0.4 (normal output rate) Power consumption Average: 25VA Built-in RS-485 interface (standard equipment; xBPI protocol): 8 bits, odd parity,
transmission rate: 9600 - 38,400 baud; half-duplex; convertible to
RS-232 (SBI protocol): 7 bits, parity: even, odd, mark,
space; transmission rates: 150 — 19,200 baud; 1 or 2
stop bits, software/hardware handshake; factory settings:
1200 baud, odd parity, 1 stop bit, hardware handshake with 2 characters after CTS
-4
2 • 10-4 2 • 10-4
* = factory setting
6
Operating Instructions IS-X
Tables of Specifications
Verifiable Models, IS Series IS16EDE-HXCE IS34EDE-HXCE IS64EDE-SXCE IS150IGG-HXCE IS300IGG-HXCE Readability in g 0,1 0,1 1 1 20
Weighing capacity in kg 16 34 64** 150** 300** Max. overload capacity in kg 130 130 130 600 600 Tare range (subtractive) in kg – 16 – 34 – 64 – 150 -300 Preload (can be electronically compensated in kg 4 4 13 20 20
without reducing the weighing capacity)
Max. preload at start of ca. 19 ca. 21 ca. 45 -
isoTEST calibration/adjustment
(weighing inst. must be zeroed) in kg Repeatability <±g 0,05 0,1 0,3 1 5 Linearity <±g 0,2 0,2 1 4 8 Sensitivity drift within
+10 to +30 °C <±/K 2 • 10-6 2 • 10-6 3 • 10
-6
2,5 • 10
-6
2,5 • 10
-6
Response time (average) s 1,5 1,5 1,5 1,5 1,5 External calibration weight g 10000 (F1) 10000 (F1) 10000 (F2) 50000 (F2) 100000(F2)
(of at least accuracy class...) Accuracy class K K K K K
Type BF BF HC BF BF Verification scale interval g 1 1 10 10 20
Minimum capacity acc. to g 5 5 50 50 100
prepackage regulations* Mindestlast nach FPV g 150 150 - - ­ Operating temperature range °C 10...30 10...30 10...30 10...30 10...30
(50 to 86°F) (50 to 86°F) (50 to 86°F) (50 to 86°F) (50 to 86°F) Load plate dimensions mm 300 x 400 300 x 400 300 x 400 800x600 800x600 Net weight, approx kg 16 16 70 70 70
Dust and water protection rating of the
housing acc. to EN 60529 IP65 IP65 IP65 IP67 IP67
AC power source
Power supply model YPS02-X or AC adapter model YPS02-Z (100-240V); YRB02-Z (rechargeable battery), YPS02-XV24 (24V)
Frequency Hz 48 – 62 Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels Display update (depends on filter level selected) 0.1 - 0.1 - 0.2 - 0.4 (normal output rate) Power consumption Average: 25VA Built-in RS-485 interface (Standard equipment; xBPI protocol): 8 bits, odd parity,
transmission rate: 9600 - 38,400 baud; half-duplex; convertible to
RS-232 (SBI protocol): 7 bits, parity: even, odd, mark,
space; transmission rates: 150 — 19,200 baud; 1 or 2
stop bits, software/hardware handshake; factory settings:
1200 baud, odd parity, 1 stop bit, hardware handshake with 2 characters after CTS
* = factory setting ** = variable weight unit kg, t
Operating Instructions IS-X
7
Pin Assignment Chart
Pin Assignment Chart
Female Interface Connector:
14-contact round female connector with screw lock hardware for cable gland
Pin Assignments
Platform: Zener barrier (output in safe area) YDI01-Z, YDI02-Z, YDI03-Z:
14-pin 12-pin Signal RS-232
1
) Signal RS-485 Round male connector Barrier (SBI and xBPI) (xBPI) G A Control output „heavier“ Control output „heavier“ K B Data output (TxD) RxD – TxD – N J C Data input (RxD) RxD – TxD – P N D Data terminal ready (DTR) M E Signal GND Signal GND F G
3
) Control output „lighter“ Control output „lighter“ A H Clear to send (CTS) E J
3
) Control output „equal“ Control output „equal“ O Universal switch 2) Universal switch 2) D L
3
) Control output „set“ Control output „set“ Provide a low-resistance connection between shield and connector casing.
1
) RS-232-interface available on request
2
) Function: see „External Universal Remote Switch“ in the section entitled „Extra Functions.“
3
) YDI03-Z only
Important:
! Only electrical apparatus with a max. voltage Um f 250V may be connected to the Zener barrier. The Zener voltage UZ is 12V.
8
Operating Instructions IS-X
Troubleshooting Guide
Troubleshooting Guide
Problem Possible cause(s) Solution No segments are shown No AC power is available Check the AC power supply; con-
nect on the display The power supply or AC adapter the power supply/AC adapter to power is not connected
Device not used > 5 minutes Turn on the display and control (now in stand-by mode) unit/indicator using the On/Off switch
Weight display shows “ H“ The load exceeds the weighing capacity Unload the weighing platform
Weight display The load plate is Place the load plate
shows “L“ or “ Err 54“ not in position on the platform
Weight display Data output not compatible Change the menu settings
shows “Err 01“ briefly with output format in device setup menu
Weight display One of the conditions for calibration Calibrate only when zero is dis-
played shows “Err 02“ briefly was not met Zero the weighing platform
Load on weighing platform Unload the weighing platform
Weight display Calibration/adjustment could not Allow the scale to warm up and
then shows “Err 03“ briefly be performed within repeat calibration/adjustment the required time period
Error code “Err 08“ The load on the weighing platform Check whether your configuration
is too heavy to allow the platform to be zeroed meets the zero-setting point requirements
Error code “Err 09“ Gross value < zero and weighing Zero the weighing platform
platform is unloaded, but cannot be tared
Error code “Err 10“ Data in application tare memory Clear the application tare
when attempted to clear lower tare memory
Error code “Err 12“ Attempt to write data in application Check configuration memory first
tare memory when blocked (menu item 2 2 x)
Value not permitted for Check value entered manual tare input
Error code “Err 17“ Internal calibration/adjustment not possible Reduce preload or
because preload is too high Select a different configuration
Error code “Err 19“ The available weighing capacity Reduce preload or
is too low; max. capacity select a different configuration reduced by high preload
The special code “A“ remains Press a key on the display; none of the keys has been pressed since the equipment was switched on
The weight readout Unstable ambient conditions Set up the weighing platform
changes constantly (too much vibration, or the in a different area, or weighing platform is exposed access the Setup menu to change to a draft) device configuration
A foreign object is caught Remove the foreign object between the load plate and the weighing platform housing
The weight readout is The weighing platform has not been calibrated Perform calibration/adjustment
obviously wrong The platform was not tared before weighing Tare before weighing
If any other errors occur, please contact your local Sartorius Intec office or dealer or the Sartorius Intec Service Center.
Operating Instructions IS-X
9
Accessories
Accessories (Options)
Order no.: Power supply, flameproof, for use in hazardous area/location 100 - 240V EU YPS02-XDR GB YPS02-XGR USA/CDN YPS02-XKR
AC adapter for use outside the hazardous area/location 100 - 240V EU YPS02-ZDR GB YPS02-ZGR USA/CDN YPS02-ZKR
In the safe area (outside the hazardous area/location), accessories can be connected through a Zener barrier.
Order no.:
Data output: RS-232 for IS-X YDO55IS-X Data output: TTY/10mA for IS-X YDO01F-X Zener barrier YDI01Z (2 RS-485 data lines) for connection in a bus network Zener barrier YDI02Z (4 RS-232 data lines) Zener barrier YDI03Z (4 RS-232 data lines and 4 control lines) Configuration program (on diskette) for IS-X weighing platform YAD01IS (PC, DOC) without interface cable Sartonet cable 6906926 Sartonet connection box 72583 Round plug-in connector 69Y03166
For customized cable solutions or accessories, please contact the Sartorius Intec.
10
Operating Instructions IS-X
General View of the Equipment
2
3
4
6
8
10
11
1
5
7
9
3
6
10
11
7
12
9
8
IS06BBE-SX
General View of the Equipment
IS2CCE-SX, IS6CCE-HX, IS6CCE-SX, IS12CCE-SX
Pos. Designation 1 Draft shield cover 2 Glass draft shield 3 Load plate 4 Pan support 5 Shield ring 6 Lug for attaching an antitheft locking device 7 Leveling foot 8 DC jack 9 Data interface (RS-485) 10 Menu access switch (under protective threaded cap) 11 Level indicator 12 Pan draft shield (models IS2CCE-SX, S6CCE-HX only)
Operating Instructions IS-X
11
General View of the Equipment
IS16EDE-HX, IS34EDE-HX, IS64EDE-SX
Pos. Designation
1 Level indicator 2 DC jack 3 Data interface (RS-485) 4 Leveling feet 5 Weighing platform 6 Load plate
12
Operating Instructions IS-X
IS150IGG-HX, IS300IGG-HX
1
3
2
4
5
7 6
General View of the Equipment
Pos. Designation
1 Load plate 2 Manufacturer’s ID label 3 DC jack (optional) 4 Leveling feet 5 Level indicator 6 DC jack 7 Data interface (RS-485)
Operating Instructions IS-X
13
Dimensions
Dimensions (Scale Drawings)
IS06BBE-SX
All dimensions given in millimeters
14
Operating Instructions IS-X
IS2CCE-SX, IS6CCE-HX
Dimensions
All dimensions given in millimeters
Operating Instructions IS-X
15