Sartorius IB 16000 S, IB 31000 P, IB 31 Installation And Operating Instructions Manual

Page 1
Sartorius Industry IB 16000 S, IB 31000 P, IB 31.
Electronic Precision Scales
98648-000-60
Page 2
2
1
Platform cover for scale base
6
Tare control
2
Level indicator (only for IB 16000S)
7
CAL key
3
Power receptacle
8
ON/OFF key
4
Leveling foot
9
Weight display
5
Menu access switch
10
Manufacturer’s label
(for scale operating program)
Page 3
3
Contents
Page
About the Product (Warranty) 4
Storage and Shipping Conditions 5
Equipment Supplied
6
Installations Instructions
7
Ambient Conditions 7
How to Mount the Display Unit
8
Mounting the Display Unit on the Short Side of the Scale Base 9
Mounting the Remote Display Unit 11 Mounting the Display Unit on the
YDH 01 B Support Arm (Option) 13
Startup
16
Connecting the scale to Line Power 16 Voltage Selection 16
Safety Precautions 17
Connecting Electronic Devices (Peripherals) 17
Leveling the Scale using the Level Indicator 17
Operation
18
Turning the Display On and Off 18
Self-Test 19
Weighing 19
Taring 19
Auto Zero 19
Automatic Shutoff 20 Calibration 20
Scale Operating Program 21
Reading a Program Menu Code (List Mode) 21
Changing a Program Menu Code (Change Mode) 21
List of the Programmable Menu Code Settings 23
Troubleshooting Guide 24
Care and Maintenance
25
Cleaning 25
Safety Inspection 25
Accessories
(Options) 26
Interfacing Devices with the Scale (RS Interface 26
Specifications
27
Page 4
4
About the Product (Warranty)
With this Sartorius Scale you have acquired a high-quality electronic Instrument that will ease your daily work load.
Please read these installation and operating instructions carefully be­fore operating your new scale.
Pursuant to the German Directive for the Implementation of Regulations for Prevention of Accidents "Elektrische Anlagen und Betriebsmittel (VBG 4)" [Electrical Installations and Equipment] of April 1986, it is hereby certified that the equipment delivered, "Electronic Precision Scale, series IB," is manufactured and tested in compliance with the following DIN/VDE regulations
DIN IEC 348/VDE 0411 Safety requirements for electronic measuring apparatus
DIN IEC 380/VDE 0806 Safety of electrically energized Office machines
DIN IEC 601/VDE 0750 Safety of medical electrical equipment
and with Article 10 of the Low Voltage Directive 73/23/EEC issued on February 19, 1973 by the European Community When you use electrical equipment in installations and under ambient conditions requiring higher safety Standards, you must comply with the provisions specified in the applicable regulations for installation in your country.
Page 5
5
Do not miss out on the benefits of our full warranty. Please complete the warranty card, indicating the date of installation, and return the card to your Sartorius dealer.
Unpack the scale. Remove the plastic covering and adhesive strips.
Place the platform cover (1) on the scale base.
Storage and Shipping Conditions
Allowable storage temperature: -40°C ...+70°C
-40°F ...+158°F
The packaging of the scale has been designed to ensure that the scale will not be damaged even if it is dropped from a height of 80 cm (about 32 inches).
After unpacking the scale, please check it immediately for any visible damage as a result of rough handling during shipment. lf this is the case, proceed as directed in the section entitled "Safety Inspection."
Save all parts of the packaging. and the box because you may need to ship your scale. Before you pack your scale to ship it, unplug all connected cables to prevent damage.
Do not expose the scale unnecessarily to extreme temperatures, moisture, shocks, blows or vibrations.
Page 6
6
Equipment Supplied
The equipment supplied includes the components shown on the left:
- Scale with platform and display unit
- AG adapter
- Dust cover
- 3 Allen wrenches
- Retainers for mounting the display
- 2 M 4 x 8 screws and washers (for mounting the "Remote Display
Unit")
Page 7
7
Installation Instructions
Ambient Conditions
Unfavourable ambient conditions can influence weighing results.
Therefore, please select a suitable place to set up your scale. lt should not be exposed to the following:
- heat radiation
- drafts
- vibrations
- aggressive chemical atmospheres (the scale must not be used in a hazardous location/area)
Do not expose the scale to extreme moisture over long periods. Moisture in the air can condense on the surfaces of a cold scale whenever it is brought to a substantially warmer area. lf you need to transfer the scale to a warmer area, make sure to condition it for about 2 hours at room temperature, leaving it unplugged. Afterwards, ifyou keep the scale connected to AC power, the continuous positive difference in temperature between the inside of the scale and the outside will practically ruie out the effects of moisture condensation. The components used in the scale are rated to at least Class KSF according to DIN 40040.
Your Sartorius scale provides accurate readouts even when it is exposed to unfavorable ambient conditions. You can adapt the scale to your requirements simply by changing the code settings in the scale operating menu. For more information, please read pages 21 through 23.
Page 8
8
How to Mount the Display Unit
The display unit can be mounted as follows:
base-mounted on the front of the scale
(factory-mounted)
base-mounted on the side of the scale (see next page for direc-
tions)
mounted as a remote display unit (see page 11)
mounted on a support arm as a raised display unit; YDH 01 IB
is an Option (see page 13)
Page 9
9
Mounting the Display Unit on the Short Side of Scale Base
Leave the platform cover on the scale base. To mount the display, turn the scale upside down so that it rests on the platform cover.
Remove the screws from the base plate using the Allen wrench sup­plied (see arrows) and then remove the base plate.
Unscrew the four screws of the angular cover plate, and remove the angular cover plate.
Remove the fastening screws from the display unit using the Allen wrench to untighten them. Remove the cable from the raceway (channel).
Page 10
10
Attach the display unit to the short side of the scale base by tightening the two screws.
Insert the cable in the raceway as shown in the diagram on the left.
Make sure that the rubber seal (see arrow) is not damaged.
Reposition the base plate and fasten it in place by tightening the three M3x8 screws (see the large arrows) and the two M3x10 screws (see the small arrows). Afterwards, fasten the angular cover plate in place by tightening the screws.
Page 11
11
Mounting the Remote Display Unit
Leave the platform cover on the scale. To mount the display, turn the scale upside down so that it rests on the platform cover.
Remove the screws from the base plate (see arrows using the Allen wrench supplied and then remove the base plate.
Unscrew the four screws of the angular cover plate, and remove the angular cover plate.
Remove the fastening screws from the display unit using the Allen wrench to untighten them. Remove the cable from the raceway (channel).
Page 12
12
Attach the display unit to the mounting retainers supplied using the two M4x8 screws and washers (can also be mounted on a wall for use as a wall console).
Thread the cable through one of the three openings in the scale hous­ing.
Make sure that the rubber seal (see small arrow) is not damaged.
Reposition the base plate and fasten it in place by tightening the three M3x8 screws (see the large arrows) and the two M3x10 screws (see the small arrows). Afterwards, fasten the angular cover plate in place by tightening the screws.
Page 13
13
Mounting the Display Unit on the YDH 01IB Support Arm (Option)
Leave the platform cover on the scale. To mount the display, turn the scale upside down so that it rests on the platform cover.
Remove the screws from the base plate (see arrows using the Allen wrench supplied and then remove the base plate.
Unscrew the four screws of the angular cover plate, and remove the angular cover plate.
Remove the fastening screws from the display unit using the Allen wrench to untighten them.
Page 14
14
Remove the four screws from the display unit (see arrows) using the Allen wrench to untighten them.
Unscrew the two screws from the cable clamp. Remove the clamp and then the display holder.
Remove the cable from the raceway in the bottom of the scale.
Slide out the narrow cover plate of the support arm in the direction of the arrow.
Use the Allen wrench to remove the four screws from the rectangular support arm cover plate and remove the cover plate from the support arm base.
Mount the display unit on the support arm using the two retainers and the four screws supplied.
Route the cable through the cable conduit inside the support arm.
Mount the angular retainer using the screw and the spring washer so that the bundle of shielded wires contacts the inside of the support arm.
Insert the cable down through the support arm into opening of the mounting frame. Afterwards, slide the narrow cover plate into the sup­port arm.
Insert the cable in the raceway as shown in the diagram.
Make sure that the rubber seal (see arrow) is not damaged.
Page 15
15
Fasten the support arm to the rear panel of the scale base using the two M5x16 screws supplied and the star lock washers.
Attach the mounting frame of the support arm to the bottom of the scale using the two M4x8 screws supplied.
Refasten the rectangular cover plate to the support arm using the four screws.
Reposition the base plate and fasten it in place by tightening the three M3x8 screws (see the large arrows ) and the two M3x10 screws (see the small arrows).
Page 16
16
Startup
Connecting the Scale to Line Power
The scale is powered by an AC adapter. Make sure that voltage rating printed on this unit is identical to your local line voltage rating.
lf the line voltage indicated or the plug design of the AC adapter does not match the rating or Standard you use, please contact your Sarto­rius dealer.
Important Note
Only use original Sartorius AC adapters identified by the Sartorius la­bel. Use of AC adapters supplied by other manufacturers, even if these adapters have a registered approval rating from a national test­ing laboratory, requires the consent of a certified Sartorius technician.
Plug the cord of the AC adapter into the power receptacle of the scale. Now plug the AC adapter into a wall outlet.
Voltage Selection
You can select the voltage only if you use our portable power supply (69 71172) that has a European-type plug (rounded prongs).
Page 17
17
Safety Precautions
The power supply or AC adapter rated to Class 2 (double insulated) can be plugged into a wall outlet without taking any additional safety precautions. The negative pole (ground) of the Output voltage (protective extra-low voltage) is connected to the scale housing, which can be grounded for Operation.
The optional interface (see "Interfacing Devices" on page 26 in addi­tion) is also electrically connected to the scale housing (ground).
Connecting Electronic Devices (Peripherals)
Unplug the AC adapter before you connect or disconnect devices to or from the interface ports.
Leveling the Scale using the Level Indicator (only for model IB 16000 S)
At the point of use, level the scale using the leveling feet
(4)
so that
the air bubble is centered within the circle of the level indicator
(2)
.
Page 18
18
Operation
After you have plugged the scale into a wall outlet using the AC adapter, allow for at least 30 minutes' warmup.
The weight display shows the following special Status messages for your Information:
BUSY
The scale processor is still busy processing a function and will not ac­cept any other commands to perform functions at this time.
STANDBY
The display has been turned off with the ON/OFF key
(8)
and the scale is now in the ready-to-operate mode so that it does not require warmup.
POWER OFF
The scale was disconnected from line power (power failure or outage, reconnection to line power after scale was unplugged).
CAL
The calibration function has been activated.
Turning the Display On and Off
Press the ON/OFF key
(8)
to turn the display on or off.
You can also turn it on by pressing the tare control
(6)
.
When the scale is connected to the AC adapter and the power is on, the display will go out whenever you turn off the power using the ON/OFF key. All other electronic circuits will remain energized (indi­cated by STANDBY). This means the scale is immediately ready to operate without requiring warmup the next time you turn it on.
Page 19
19
Self-Test
After the scale is turned on, an automatic self-test of the scale's elec­tronic circuitry is performed.
This self-test ends with the readout 0.0 g/0.000 kg (if the factory code setting 5 1 1/ 5 1 2 is used).
Weighing
Now place your sample or object on the platform
(1)
to determine the
weight. Read off the weight indicated in the display
(9)
as soon as the
weight unit (in this case "g") appears as the stability symbol.
In addition to grams/kilograms, this scale gives you a variety of other menu-definable international weight unit options.
Select the weight unit you need from the table of the menu options for the scale operating program. Set the appropriate code as described in the section "Scale Operating Program."
Taring
lf you wish to weigh into a Container or if the weight display does not indicate 0.0 g/0.000 kg (or the equivalent with the weight unit of your choice), press the tare control to zero the display.
Auto Zero
This scale has an automatic zero tracking function, known as "Auto Zero" (can be turned off by menu code - see "Scale Operating Pro­gram"). Changes off zero 0.5 of a digit per second are set to zero automati­cally.
Page 20
20
Automatic Shutoff
Make sure the "Automatic Shutoff" code is set to "ON" when using a non-rechargeable or rechargeable battery. This will increase the amount of time you can operate the scale before having to recharge the battery (see "Scale Operating Program"), lf you do not press a key or if a weight readout in the display does not change for at least 2 minutes, "STANDBY" will start flashing in the display. The scale will shut itself off if it has not been operated after another 2 minutes.
Calibration
— This is only possible with an accurate calibration weight (10 kg) — See "Accessories" -
Unload the scale and press the CAL key
(7)
for at least 3 seconds un-
til the calibration weight readout appears in the display.
Center the calibration weight on the platform. Now the weight unit symbol is displayed. An acoustic signal indicates the end of the calibration procedure.
You can block access to the calibration function -to find this menu code, refer to the "Scale Operating Program." This function is acces­sible when the scale operating program is unlocked using the menu access switch
(5)
.
Page 21
21
Scale Operating Program
The scale operating program lets you adapt your scale to various am­bient conditions and to different weighing requirements, and select weight units commonly used in your country. At the factory, we have set the codes for a Standard program which is protected by a locking function to prevent accidental changes.
The "
menu code
" contains the information of the operating program, lt consists of three digits, known in "computerese" as the page (1st digit), the line (2nd digit) and the word (3rd digit).
Reading a Program Menu Code (List Mode)
How to access the menu of the scale operating program:
With the display turned off (
STANDBY
state), hold down the tare con-
trol
(6)
and briefly press the ON/OFF key
(8)
. Upon completion of the
automatic self-test, release the tare control as soon as "
CH5
" is dis­played. The Status of the scale operating program is indicated in the weight display:
"L" Stands for the list mode. In this mode, you can read a menu code setting, but you cannot change it.
Changing a Program Menu Code (Change Mode)
lf you wish to change a program menu code, you must first unlock the menu access switch to access the program menu.
To do so, remove the protective cover located on the right of the dis­play unit, and slide the menu access switch
(5)
in the direction of the
arrow.
The display will indicate "C", which Stands for the change mode, meaning you can now change the menu code settings.
Page 22
22
After you have accessed the menu of the scale operating program, the display will show a continuous numerical sequence from 0 to 5 for the "page" or the first digit of the code, in addition to the Status code letter "L" or "C". When the first digit of the code you wish to check or change appears, press the tare control
(6)
. The "page" code number now stops in the display, and a series of numbers for the 2nd digit or ''line" will begin to cycle. Press the tare control again to stop the code number of your choice in the display. Next, the numbers for the "word" (last digit) will cycle in the display. Repeat the procedure to enter the last digit of the code.
The "o" symbol that appears indicates the actual setting.
To change any settings ("C" mode), press the tare control as soon as the appropriate numeric code is displayed.
Brief display of BUSY and the "o" symbol confirms your selection, fol­lowed by a return to "zero" for the 2nd digit or "line."
How to return to the weighing mode: Press the tare control each time a 0 appears in the numerical se­quence (word, line, page). lf you have changed a menu code, it will be stored as soon as the display returns to the weighing mode. Lock the scale operating program using the menu access switch ("L" readout) and replace the protective cover.
Page 23
23
List of the Programmable menu Code Settings
C 1 3 1
Page Line Word
Menu of the Balance Operating Program
(Active Parameters)
Code
Ambient Conditions
Code
Program Lock
C 1 1 1
Very stable
C 4 1 1
OFF
C 1 1 2
Stable
C 4 1 2
ON
C 1 1 3
Unstable
C 1 1 4
Very unstable
Code
Acoustic Signal
C 4 3 1
ON
Code
Stability Range
C 4 3 2
OFF
C 1 2 1
0,25 digit
C 1 2 2
0,5 digit
Code
Weight Units
1
)
C 1 2 3
1 digit
C 5 1 1
Grams g

C 1 2 4
2 digits
C 5 1 2
Kilograms kg
C 1 2 5
4 digits
C 5 1 3
Carats ct
C 1 2 6
8 digits
C 5 1 4
Pounds lb
C 1 2 7
16 digits
c 5 1 5
Ounces oz
C 1 2 8
32 digits
c 5 1 6
Troy ounces ozt
C 1 2 9
64 digits
c 5 1 7
Parts/pound o
c 5 2 1
Hong Kong taels tl
Code
Display Format
c 5 2 2
Singapore taels tl
C 1 3 1
Last decimal ON
c 5 2 3
Taiwan taels tl
C 1 3 2
Last decimal OFF
c 5 2 4
Grains gr
C 1 3 3
Last decimal at stability
c 5 2 5
Pennyweights dwt
C 1 3 4
All decimals at stability
c 5 2 6
Mommes o
c 5 2 7
Milligrams o
Code
Tare parameter
C 5 2 8
Austrian carats o
C 1 4 1
Without stability
C 1 4 2
At stability
C 5 2 0
Call Program Line
Code
Auto Zero
C 5 0
Call Program page
C 1 5 1
ON
C 1 5 2
OFF
C 0
End of programming
Code
Calibration
C 1 6 1
Accessible
C 1 6 2
Accessible blocked
Additional Parameters for the data Output format at the interface port and for calculation programs are available on request.-Please refer to the “Accessories."
1)
You can choose any weight unit as long as it can be displayed in the particular weighing range you
selected (for example do not set the code for "kg" when you are using a 0.1 mg balance).
Page 24
24
Troubleshooting Guide
Problem ... Causes ... Remedy
No segments are indicated in the weight display
(9)
- No line voltage available
- The AC adapter is not plugged
in
- Check power supply
- Plug in the AC adapter
-
Weight display shows
“L”
or
“CH2”
- The platform cover
(1)
is not in
place
- An object is pressed against the platform cover or has gotten underneath it
- Place platform cover on the scale base
- The platform cover must rest properly on the scale base without any interfering objects in between
Weight display shows
“H”
- Load exceeds the capacity of the scale
- Unload the scale
The weight readout changes constantly or the special message
“BUSY”
does not go out in the
weight display
- Unstable ambient conditions
- Too much vibration or there is
a draft
- Sample does not have a stable weight
- Set up scale in another area
- Access the menu to adapt
scale to the particular type of weighing environment
The weight display shows
“CE”
- A zero readout was not indicated in the weight display when the CAL key
(11)
was
pressed to calibrate
- The scale is loaded
- Press tare control
(9)
and re-
press the CAL key
The code
“CC“
in the display does
not go out
- The CAL key
(7)
has not been pressed when the display read zero
- The scale is loaded
- Press the tare control
(6)
and
re-press the CAL key
- Unload the scale The weight readout is obviously wrong
- Scale has not been calibrated
- Scale has not been tared
before weighing
- The air bubble of the level indicator
(2)
is not within the
circle(only for IB 16000 S)
- Calibrate scale
- Tare before weighing
- Level scale
“POWER OFF”
blinks in the display
- The operating voltage is too loo (battery discharged)
- Connect the scale or the external rechargeable battery to line voltage using the AC adapter
Page 25
25
Care and Maintenance
Cleaning
Please do not use any aggressive cleaning agents (solvents or similar agents). Instead, use a piece of cloth that has been wet with a mild detergent to clean the scale. Make sure that no liquid enters the scale housing. After cleaning, wipe down the scale with a soft, dry piece of cloth.
Safety Inspection
lf there is any indication that safe Operation of the scale with the AC adapter/power supply is no longer warranted, turn off the scale and disconnect the equipment from AC power immediately. Lock the equipment in a secure place to ensure that it cannot be used for the time being.
In this case, notify the Sartorius Service Center or International Tech­nical Support Group based in Goettingen, Germany. Only service technicians who are authorized by Sartorius and have access to the required manuals are allowed to perform maintenance and repair work on the equipment.
Safe Operation of the scale with the AC adapter/power supply is no longer ensured when
there is visible damage to the AC adapter/power supply the AC adapter/power supply no longer functions properly the AC adapter/power supply has been stored for relatively
long periods under unfavourable conditions
the AC adapter/power supply has been exposed to rough
handling during shipment
We recommend that the scale along with the AC adapter/power sup­ply be inspected by a qualified Sartorius service technician according to the checklist given below.
Leakage current <0.05 mA measured by a properly calibrated
multimeter
Insulation resistance >7 megohms measured with a constant
voltage of at least 500 V at a500kohmload.
The duration and number of measurements should be determined by a qualified Sartorius service technician according to the particular ambient conditions and operational conditions of the AC adapter. Such inspection should be done at least once a year.
Page 26
26
Accessories (Options)
Data printer with date/time and statistics functions
YDP 02-0D
Print speed approx. lines/sec. 1,5
Printer housing (W x D x H) in mm
in inches
150 x 138 x 43
5.5 x 5.4 x 1.7
Interface YDO 01 IB “Data Input” (incl. interface) YDI 11 IB Dedicated applications kit for the YDI 11 IB (incl. interface) Replace & with the letter code: C for parts counting F for formulation A for animal weighing U for the PLUS Performance Package
YAK 11 I-000&
Support arm (for raised display unit YDH 01 IB
Rechargeable battery pack, external; approx. 10 hours of operation; can be recharged by the AC adapter (detailed information for powering the scale is available in our Service Information bulletin, no. 15/88)
YRB 02 Z
Adapter for hook-up to a 12-Volt battery
YCC 01-0007
Roller track conveyer plate YRT 01 IB-0001
Remote display (can be connected to the interface port) — LCD, reflective — for overhead projectors, transmissive
7371 01A 7371 02A
Calibration weight ( 1 x 10 kg) 7072 18
Dust cover 69 60I310
Interfacing Devices with the Scale
Please note that the interface port is electrically connected to the pro­tective grounding conductor of the scale housing. The cabling sup­plied as accessory components is shielded and electrically connected on both ends to the cases of the connectors. This electrical connec­tion may result in interference caused by ground loops or by transient currents if you have grounded the housing or connected the protective grounding conductor for AC power. lf necessary, connect an equipotential bonding conductor to the scale.
Page 27
27
Specifications
Model IB 16000 S IB 31000 P IB 31
Capacity kg 16,1 4/8/16/31 31 Readability
g 0,1 0,1/0,2/0,5/1 1
Maximum overload capacity kg 75 75 75
Tare range (by subtraction) kg 16,1 31 31
Standard deviation
g ±0,05 ±0,1/0,1/0,25/1 ±0,5
Max. linearity deviation g
±0,2±0,2/0,2/0,5/1±1
Stabilization time (typical) s
2
Adaption to ambient conditions and application requirements
by selection of one of four digital filter levels
Display update rate s 0,1 – 0,8 (selectable)
Stability range d 0,25 … 64 (selectable)
Ambient temperature range 0 . oC.. +40 oC (32oF … 104oF) Moisture-proof rating according
to DIN 40040
Class F, non-condensing Dust and water protection rating in compliance with DIN 40050/IEC 529
IP 65 (scale) = NEMA 4
IP 20 (AC adapter) Sensitivity drift within +10 ... +30oC
±2,5∙10
-6
±4∙10
-6
±1010-6 Platform size mm 417 x 307 (16.4 x 12.1 in.) Scale base (W x D x H) mm 417 x 307 x 113
(16.4 x 12.1 x 4.4 in.) (without display unit)
Net weight of the scale kg 15 (33 lbs.)
Net weight of the display unit kg 1 2.2 lbs.) Power requirements (voltage +
frequency: 50 – 60 Hz)
115 or 230 V depending on the AC adapter used
Allowable voltage fluctuation -20% ... +15%
Power consumption VA 9 (typical) Interface (optional)
RS 232 C/V24 – 28, RS 423/V10; 7-bit; parity: even, mark, odd, space; transmissions rate: 150 ... 9600 Baud
Page 28
Sartorius AG
B 37070 Göttingen P W eender Landstraße 94–108, 37075 Göttingen T (0551) 308-0, F (0551) 308-3289
Internet: http://www.sartorius.com Copyright by Sartorius AG, Göttingen, Deutschland.
All rights reserved. No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartorius AG.
Sartorius AG reserves the right to make change to the technology, features, specification and design of the equipment without notice.
Loading...